Surface-based binding assays are often influenced by the
transport of analyte to the sensor surface. Using simulated data sets,
we test a simple two-compartment model to see if its description of
transport and binding is sufficient to accurately analyze BIACORE data.
First we present a computer model that can generate realistic BIACORE
data. This model calculates the laminar flow of analyte within the flow
cell, its diffusion both perpendicular and parallel to the sensor
surface, and the reversible chemical reaction between analyte and
immobilized reactant. We use this computer model to generate binding
data under a variety of conditions. An analysis of these data sets with
the two-compartment model demonstrates that good estimates of the
intrinsic reaction rate constants are recovered even when mass
transport influences the binding reaction. We also discuss the
conditions under which the two-compartment model can be used to
determine the diffusion coefficient of the analyte. Our results
illustrate that this model can significantly extend the range of
association rate constants that can be accurately determined from
BIACORE.
 |
INTRODUCTION |
Until recently, obtaining reliable equilibrium
and rate constants for interacting biomolecules was often difficult and
time consuming. Optical biosensors now offer a rapid way to determine equilibrium and rate constants without the need to label the
interacting biomolecules (Garland, 1996
; Silin and Plant, 1997
).
Because of these features these instruments, such as BIACORE (Biacore,
Uppsala, Sweden), have gained wide use and have been employed in the
study of the interactions of a variety of biomolecules, including
proteins, nucleic acids, lipids, and carbohydrates (reviewed in Szabo
et al., 1995
; Raghavan and Bjorkman, 1995
; van der Merwe and Barclay, 1996
; Myszka, 1997
; Schuck, 1997a
; Fivash et al., 1998
).
In BIACORE instruments, one of the reactants is immobilized on a sensor
chip. We will call it the receptor, in analogy to a receptor on a cell
surface, although in the literature it is often referred to as the
immobilized ligand. The other reactant, called the analyte, flows past
the chip. The immobilization of the receptor is usually accomplished by
coupling it to a thin dextran layer that extends ~100 nm out from the
sensor surface, 0.2% of the height of the flow chamber. Detection of
binding is based on the optical phenomenon of surface plasmon resonance
(SPR). The SPR response is used to detect changes in the index of
refraction caused by mass changes at the sensor surface. These changes
are brought about by the binding of analyte to receptor (Jönsson et al., 1991
; Malmqvist, 1993
; Garland, 1996
). Thus continuous monitoring of the SPR signal allows the kinetics of binding to be
followed in real time. After binding, buffer alone may be introduced to
monitor the dissociation kinetics.
As the preceding description suggests, the SPR signal in a typical
experiment is not simply a report of the progress of the chemical
processes of association and dissociation at the surface. Rather, the
signal is the result of a combination of these chemical processes and
the transport processes of diffusion and flow. Thus obtaining estimates
of the intrinsic forward and reverse rate constants of the
analyte-receptor reaction requires some model that will, when needed,
account for transport. The model most often used assumes that after a
brief transient, during which analyte is transported to the sensor
surface, no correction is needed, because the free analyte
concentration remains uniform in space and constant in time, kept so by
the continuous influx of new analyte. We will refer to this model as
the rapid mixing model because, after the brief transient, the binding
at the sensor surface is the same as for a well-mixed system with
constant analyte concentration.
Systematic errors in the estimates of the rate constants can arise if
the rapid mixing model is not a reasonable description of the binding
kinetics (Chaiken et al., 1992
; Malmborg et al., 1992
; Ito and
Kurosawa, 1993
; Morton et al., 1994
; Wohlhueter et al., 1994
). A number
of authors have discussed this problem and presented conditions under
which the assumption is expected to break down (Glaser, 1993
; Karlsson
et al. 1994
; Schuck, 1996
; Schuck and Minton, 1996
; Yarmush et al.,
1996
; Christensen, 1997
). There are two possible sources for the
breakdown: 1) the transport (diffusion and flow) of analyte in solution
to the dextran layer and 2) the transport (primarily diffusion) of
analyte within the dextran layer. If one or both of these influence the
kinetics of binding, then the free analyte concentration cannot be
taken to be constant and uniform.
In this paper we consider only experiments in which the properties of
the dextran layer do not influence the binding kinetics, i.e., where
the binding can be treated as if the receptors are coupled directly to
the sensor surface. BIACORE binding studies have been performed both
with receptors coupled to dextran layers and with receptors coupled
directly to the sensor surfaces (Karlsson and Fält, 1997
; Parsons
and Stockley, 1997
). These experiments demonstrate that under
appropriate experimental conditions (i.e., low surface capacity), the
dextran layer has no significant effect on the binding kinetics.
To improve the analysis of bulk transport effects on the SPR signal, a
simple model has been proposed that treats binding as a two-step
process: transport of analyte to the sensor surface, followed by
reaction of analyte with receptors on the surface (Myszka et al.,
1997
). This two-compartment model is attractive because it can be
formulated in terms of a simple system of ordinary differential
equations and because it is only slightly more complicated than the
rapid mixing model. The two-compartment model has been used to describe
several experimental systems studied with BIACORE instruments (Myszka
et al., 1996
, 1997
; Morton and Myszka, 1998
). This is encouraging, but
uncertainty remains about how well the two-compartment model describes
BIACORE binding kinetics. What is provided in this paper is a
demonstration that the two-compartment model is sufficiently accurate
to analyze BIACORE data. In particular, we show that the numerical
parameter values obtained by fitting the two-compartment model to noisy
SPR data are accurate estimates of the underlying physical parameter
values.
The approach we use to demonstrate the validity of the two-compartment
kinetic scheme for analyzing BIACORE binding data is as follows:
1. We present a computer model that can simulate BIACORE binding
experiments involving monovalent analytes binding to monovalent receptors, i.e., that accounts for laminar flow and diffusion of the
analyte, chemical reactions at the sensor surface, and geometry of the
BIACORE. This computer model is similar to that developed by Glaser
(1993)
, the difference being that we include diffusion in the flow
direction.
2. We use the computer model to simulate BIACORE binding data, choosing
parameter values that lead to transport-influenced binding kinetics as
well as parameter values where such effects are negligible.
3. We fit the two-compartment binding model to the simulated data and
show that we recover the original values of the rate constants (the
values used to simulate the data).
4. We then add random noise at realistic levels to the simulations and
demonstrate that good estimates of the input parameters are recovered
when the two-compartment model is used to analyze the noisy simulated
data.
 |
OBTAINING SIMULATED DATA SETS |
The mathematical model
In BIACORE instruments, analyte is transported by diffusion and
flow to the sensor surface, where it reacts with immobilized receptors.
The flow channel of the BIACORE has a rectangular cross section (length
l, height h, and width w). We let
t be the time, and we introduce Cartesian coordinates with
the origin at the inflow port, the x axis parallel to the
direction of flow, and the y axis normal to the
receptor-coated surface (illustrated in Fig.
1). (In BIACORE instruments, the sensor
surface is actually on top of the flow cell. For convenience we draw it
on the bottom.) Because the instrument's flow chamber is 10 times
wider than it is high, variations in concentrations across the width of
the chamber can be neglected.

View larger version (11K):
[in this window]
[in a new window]
|
FIGURE 1
Schematic of a BIACORE flow chamber, where the
variation along the width of the chamber has been neglected and the
sensor surface has been placed on the bottom rather than the top. For a
standard instrument, h = 5 × 10 3 cm, l = 0.24 cm, and the width
equals 5 × 10 2 cm. The figure is not to scale. The
true length is 48 times longer than the height.
|
|
Along the flow chamber, laminar flow is fully developed essentially
over its entire length (Brody et al., 1996
). The velocity profile is
therefore parabolic, equal to zero at the top
(y = h) and bottom
(y = 0) boundaries and rising to a maximum,
vc, in the center, i.e., the velocity
v(y) at a height y above the sensor surface is (Batchelor, 1967
)
|
(1)
|
(Because Eq. 1 is sometimes written in terms of the average
velocity,
, we note that
= 2vc/3.)
The dependent variables of the model are the bulk concentration of free
analyte, C(t, x, y), and
the surface density of bound analyte,
B(t, x) (with units of mass/volume
and mass/area, respectively). The net analyte flux due to convection
and diffusion is
|
(2)
|
where J is the flux vector, D is the analyte
diffusion coefficient, and i and j are the unit
vectors in the x and y directions. The component
of J normal to any surface, integrated over that surface,
equals the flux through the surface, e.g., the number of particles per
second passing through the surface. From conservation of mass (the
continuity equation), the following partial differential equation (PDE)
is obtained:
|
(3)
|
The boundary conditions for Eq. 3 are as follows. Because the top
boundary is impenetrable and unreactive, the mass flux of analyte must
vanish at this boundary:
|
(4a)
|
In contrast, the analyte mass flux at the sensor surface should
equal the time rate of change in the amount bound at the surface, i.e.,
|
(4b)
|
where C(t, x, 0) is the free analyte
concentration at position x just above the sensor surface.
R(t, x) = RT
B(t, x) is the free surface
receptor concentration at x. RT is
the total receptor concentration and is constant with respect to both
position and time. The rate constants of the reaction are
ka and kd. Note that
these are the intrinsic microscopic forward and reverse rate constants
for binding of the analyte to the receptor.
At the entry port to the flow cell (x = 0), the
concentration is constant and equal to the injection concentration,
CT, i.e.,
|
(4c)
|
At the end of the sensor, where analyte exits (x = l), we adopt a simple continuation condition that assumes
the exit of analyte is due entirely to flow:
|
(4d)
|
Hence they leave the computational domain and have a negligible
effect on the dynamic processes happening inside the instrument.
It is useful to write Eq. 3 and the boundary conditions, Eqs. 4a-4d,
in nondimensional form. To do this we introduce the following nondimensional time and lengths, concentrations, and rate constants:
|
(5a)
|
|
(5b)
|
|
(5c)
|
and nondimensional parameters:
|
(5d)
|
Equations 3 and 4a-4c become
|
(6)
|
|
(7a)
|
|
(7b)
|
|
(7c)
|
|
(7d)
|
If diffusion in the x direction is ignored (i.e., the
2 term in Eq. 6 is set equal to zero), the model reduces
to what has now become a standard description for the BIACORE flow cell
(Lok et al., 1983
; Glaser, 1993
; Christensen, 1997
). Yarmush et al. (1996)
have also used this approximation, but in the context of a
nonnegligible dextran layer. For a standard flow cell,
= 0.021. This suggests that such an approximation is justified, although for y <
, diffusion in the x direction
will dominate flow. We have not made this approximation, because we are
using the model as an absolute standard for generating simulation data,
and therefore it is desirable to include as much detail as possible.
Because we are solving Eq. 6 numerically, there is no necessity to do otherwise.
The numerical method
We have solved Eq. 6 by using a general, conservative, and
well-tested hydrodynamics transport code based on the method of finite
elements (Dembo, 1994a
,b
). Shown in Fig.
2 is a small computational grid (8 × 8) with 49 node points and 32 surface elements. In the x
direction the length is divided uniformly into eight parts. In the
y direction the height first is divided into four equal parts. The top and bottom sections are then divided in half, so there
are six sections in all. Finally, the new top and bottom sections are
divided in half, forming the eight sections shown in Fig. 2. The height
of the bottom and top sections is h/16. Refinements of this
grid are generated by subdividing each of the quadrilaterals of the
course mesh into four equal parts. For example, the top and bottom
elements of a 64 × 64 grid are h/128 in height and
l/64 in length. In the BIACORE flow cell, except at very
short times after the initiation of binding and dissociation, the
concentration of analyte is uniform far from the sensor surface. Thus
there is no reason for generating a fine grid near the top surface.
This was done only because the code we adapted required a grid that was
symmetrical about h/2. We have since developed a code for
which this is not required (unpublished result).
In trial simulations we compared the time course for the average
concentration of bound analyte, calculated using a 32 × 32 and a
64 × 64 grid. The average concentration of the bound ligand was
calculated by summing the bound ligand concentration over each bottom
surface element and dividing the sum by the number of bottom surface
elements, n + 1 for an n × n
grid. There was no significant difference between the values obtained
using the two grids. Nevertheless, in the Results section, where we
test the two-compartment model's ability to recover accurate parameter values, all simulations were done on a 64 × 64 grid. For this grid the first set of grid points above the sensor surface is at
y = h/128
400 nm, or about four
times the thickness of the dextran layer for a standard CM5 sensor
surface (Stenberg et al., 1991
).
Simulations
To illustrate the information that can be obtained from solving
Eq. 6 numerically, subject to the boundary conditions in Eqs. 7a-7d,
we simulate a BIACORE binding study using parameter values previously
determined for interleukin-2 (IL-2), flowing past, and interacting
with, its immobilized low-affinity receptor, IL-2R
(Myszka et al.,
1996
). In the simulation we took the association rate constant
ka = 8 × 106 M
1
s
1 and the dissociation rate constant
kd = 0.2 s
1. Analyte was
injected for 50 s, after which buffer was introduced and
dissociation proceeded for an additional 50 s. Fig.
3 shows the average concentration of
bound analyte and the average free analyte concentration just above the
sensor surface plotted as a function of time. We see from Fig. 3 that
the free analyte concentration near the sensor surface is not constant
during the association phase or zero during the dissociation phase, as
is often assumed. Fig. 4 presents
"snapshots" of the free analyte concentration, as a function of
position throughout the flow cell, at four successive time points (two
during the association phase and two during the dissociation phase).
Note that at 25 s and 75 s there is substantial variation in
the free analyte concentration along the sensor surface. As a
consequence, the bound analyte concentration will also be nonuniform,
as illustrated in Fig. 5. In BIACORE
instruments the SPR detector is centered in the middle of the flow cell
and monitors ~75% of the total flow path. Thus the output is a
measure of the average bound analyte over this region.

View larger version (23K):
[in this window]
[in a new window]
|
FIGURE 3
Predicted time course for the average concentration of
bound analyte ( ) and average free analyte concentration at the
sensor surface (- - -). In the simulation, 25 nM analyte was
injected from 0 to 50 s (CT = 25 nM), then the injected concentration was set to zero
(CT = 0) and dissociation was followed for
50 s. vc = 5 cm/s, corresponding to a
flow rate of 50 µl/min, and the analyte diffusion coefficient
D = 1 × 10 6 cm2/s.
The immobilized receptor concentration RT = 7.5 × 1011 receptors/cm2 = 1.25 nM cm.
(Because 1 RU = 10 10 g/cm2, when all
sites are filled with IL-2, MW = 14,000, there is an increase of
175 RU over baseline.) A standard flow cell was assumed, with
h = 5 × 10 3 cm and
l = 0.24 cm. ka = 8 × 106 M 1 s 1 and
kd = 0.2 s 1, as reported in
the literature for IL-2 binding to IL-2R (Myszka et al., 1996 ). A
32 × 32 grid was used.
|
|

View larger version (76K):
[in this window]
[in a new window]
|
FIGURE 4
Predicted free analyte concentration in the BIACORE
flow cell for the simulation parameters given in Fig. 3. Different
colors indicate different free analyte concentrations in 0.5 nM
increments. The contours separating two colors are lines of fixed
concentration. The value of the concentration at any point on the grid
is obtained by interpolating between the calculated values of the two
nodes the grid line connects. Note that in the association phase
(t = 25 s and 50 s) the contours
separating the two colors nearest the top surface are not smooth. This
indicates that in this region, more node points are required to improve
accuracy. This simulation was done on a 32 × 32 grid. Despite the
apparent lack of spatial resolution far from the boundary, when we
compared calculations using a 32 × 32 and a 64 × 64 grid,
the quantity we are most interested in, the average bound analyte
concentration, , changed only slightly. Between
10 s and 90 s, the difference was always less than 1%.
|
|

View larger version (16K):
[in this window]
[in a new window]
|
FIGURE 5
Spatial variation in the bound analyte concentration,
B(t, x). Shown are time
courses for the bound analyte concentration in RU at a distance of
l/4 ( ), l/2 (- · -), and
3l/4 (·····) along the sensor surface. The
simulation parameters used are the same as in Fig. 3. The results
demonstrate that the bound analyte concentration varies with the
distance from the entrance to the flow cell.
|
|
The parameter values used in Figs. 3-5 were chosen to demonstrate how
transport affects the kinetics of binding and dissociation in a BIACORE
instrument. In Fig. 6 we show that as the
association rate is decreased, transport effects become negligible,
i.e., the concentration of free analyte becomes constant in time, and the rapid mixing model becomes a good description of the binding.

View larger version (15K):
[in this window]
[in a new window]
|
FIGURE 6
The time course of the average free analyte
concentration just above the sensor surface, for different values of
the forward rate constant, ka. Shown are
time courses for the average free analyte concentration for
ka = 8 × 104
M 1 s 1 ( ), 8 × 105
M 1 s 1 (- · -), and 8 × 106 M 1 s 1 (·····).
All other simulation parameters are the same as in Fig. 3.
|
|
In the next section we discuss a proposed method of analyzing BIACORE
binding data that are influenced by mass transport. To test this method
we simulate additional BIACORE binding experiments under a variety of
conditions and see if the method can recover the kinetic parameters
used in the simulations.
 |
ANALYZING MASS TRANSPORT-INFLUENCED BINDING DATA |
A two-compartment model
Obtaining kinetic parameters from binding data requires fitting a
model to the data. At present, it is impractical to do this by using
the mathematical description of the previous section, a PDE (Eq. 6),
and its boundary conditions, Eqs. 7a-7d. A simpler mathematical model
is required, one that can accurately calculate the average
concentration of bound analyte at the sensor surface. This model need
do nothing more. For example, it does not have to describe the spatial
variation of the analyte concentration throughout the flow cell or of
the bound analyte over the sensor surface.
A two-compartment model, consisting of a set of coupled ordinary
differential equations, has been used to analyze a variety of BIACORE
binding experiments that have been shown to be influenced by mass
transport (Myszka et al., 1996
, 1997
; Morton and Myszka, 1998
). First
we review this model, and then in the following section we test it, to
see if accurate rate constants can be recovered from simulated BIACORE
experiments.
The model handles the variation in analyte concentration in the BIACORE
flow cell (see Fig. 4) by dividing the flow chamber into two
compartments as shown in Fig. 7. Within
each compartment the concentrations are uniform in space but may change
in time. These concentrations are averages over the length of the
BIACORE flow cell. It is assumed that the analyte concentration in the outer compartment is constant and equal to the injection concentration, CT. The model ignores the brief transition
periods that occur when binding or dissociation is initiated and the
concentration in the outer compartment rises or falls to
CT. At the typical flow rates used in kinetic
experiments (50-100 µl/min, corresponding to
vc = 5
10 cm/s), the time needed for an
analyte molecule that is not near the walls of the flow cell to travel
0.24 cm, the length of the flow chamber, is less than 0.05 s. The
concentration in the inner compartment, C, changes because
analyte is transported between compartments and because analyte binds
to, and dissociates from, immobilized receptors on the sensor surface.
If Vi is the volume of the inner compartment,
S is the surface area of the flow chamber, and
kM is the transport coefficient describing
diffusive movement of analyte between the compartments, then
|
(8a)
|
|
(8b)
|
The initial conditions for the binding phase of the experiment are
that at t = 0, CT equals the
injection concentration, C = 0 and B = 0. The initial conditions for the dissociation phase are that at the
time of dissociation, tdiss,
CT = 0, and C and B equal
the values they attained at t = tdiss. If the association phase is sufficiently
long so that a steady state is reached, then at t = tdiss, C = CT and B = KCTRT/(1 + KCT), but this is not required.

View larger version (10K):
[in this window]
[in a new window]
|
FIGURE 7
A two-compartment model of the BIACORE flow cell. In
the binding phase the average analyte concentration near the sensor
surface, C, equals zero initially and grows with time
until it equals the injection concentration
CT or dissociation is initiated and it
decays to zero. The analyte concentration far from the sensor surface
remains constant and equal to CT during
binding, and constant and equal to zero during dissociation.
B and R are the bound and free receptor
concentrations, averaged over the sensor surface.
|
|
The rapid mixing model assumes that after a brief transient, the free
analyte concentration is uniform throughout the flow chamber. This
model is described by Eq. 8b with C = CT.
As discussed in the Appendix, kM is the
diffusion-limited forward rate constant, averaged over the sensor
surface. Transport effects will influence the kinetics of binding when
the reaction rate is fast compared to the rate of transport, i.e., when
kaRT
kM. (To evaluate this inequality, the units must
be consistent, so for example, if ka is in
M
1 s
1, then RT
should be in M cm and kM in cm/s.) As discussed
by Lok et al. (1983)
and Sjölander and Urbaniczky (1991)
, it can
be shown that, to a good approximation,
|
(9)
|
Because the model does not specify the height of the inner
compartment, there are four unknown parameters that enter Eqs. 8a and
8b, ka, kd,
kM, and hi = Vi/S. However, for the range of parameter values that describe BIACORE experiments, the solutions to
Eqs. 8a and 8b are insensitive to the value of
hi. It can be shown (see Appendix and Fig. 3)
that at the sensor surface, once binding or dissociation is initiated,
C changes rapidly over a short period of time and slowly
thereafter. Furthermore, while this rapid change in C is
occurring, there is a negligible change in B. This behavior
means that a quasi-steady-state approximation can be made where we set
dC/dt = 0 (Segel and Slemrod, 1989
). When
this is done, we see from Eq. 8a that hi drops
out of the equations, which explains why the choice of
hi does not affect the solution. Once the
quasi-equilibrium approximation is made, the model is identical to the
often discussed effective rate constant model (Karlsson et al., 1994
).
Returning to Eqs. 8a and 8b, we define
= B/hi,
= R/hi,
T = RT/hi, and
M = kM/hi. In terms of these
quantities, Eqs. 8a and 8b take the familiar form
|
(10a)
|
|
(10b)
|
When these equations are used to analyze binding data, some care
must be taken in the choice of units. If we measure R,
RT, and B in RU; C and
CT in M; ka in
M
1 s
1; kd in
s
1; and kM in cm/s, this is
equivalent to setting hi = 1 RU/M. (Because the
data are insensitive to the value of hi, we are
free to choose any value for it we wish. Taking
hi = 1 RU/M avoids having to divide the data
values by hi before fitting the data. It is a convenient choice having nothing to do with the physical height of the
inner compartment.) To obtain kM in cm/s from
M in 1/s, the parameter determined from
the least-squares fit of the data, one must multiply by
10
7 and divide by the molecular weight of the analyte,
i.e.,
|
(11)
|
where we have used the conversion factor 1RU = 10
10 g/cm2.
Data fitting approach
We briefly summarize the data fitting approach we employ, because
it has been discussed in detail elsewhere (Myszka, 1997
; Myszka et al.,
1997
; Morton and Myszka, 1998
). The rate constants ka and kd,
M (related to the transport coefficient
kM by Eq. 11), and the surface receptor density
RT are taken to be free parameters. The best fit
values of these parameters are determined from nonlinear least-squares
fitting of the data, where the calculated values of the average bound
analyte concentration for each time point are determined by solving
Eqs. 10a and 10b numerically (Morton et al., 1995
). (The software
program used in this analysis (CLAMP) may be downloaded for no charge
at the following web site:
http://www.hci.utah.edu/cores/biacore/docs/clamp.html.)
To obtain accurate values for the kinetic rate constants and the
transport coefficient, the data that are being fit must be sensitive to
these parameters. Recall that as the receptor density on the sensor
surface is increased, the binding reaction at the sensor surface speeds
up, and the binding kinetics become transport limited. As a result, at
high surface receptor densities, the binding kinetics are dependent on
kM. In the other limit, as the receptor density
is reduced, transport effects decrease and these data become
insensitive to kM and strongly dependent on the
reaction rate constants. To try to ensure that the data depend on both the transport and the reaction rate constants, we simultaneously fit
data sets for different surface receptor densities. The procedure is
sometimes called global analysis and has been extensively used to
analyze BIACORE binding data (Morton et al., 1995
; Myszka et al., 1996
,
1997
; Morton and Myszka, 1998
).
 |
RESULTS |
In this section we report the results of fitting the
two-compartment model to simulated data sets generated under a variety of conditions. Before considering the effects of noisy data on parameter estimation, we see how well the model does at recovering parameter values in the absence of noise.
We start by analyzing data where the reactions are not influenced by
mass transport. Fig. 8 illustrates two
sets of binding experiments simulated with identical rate constants
(ka = 5 × 104 M
1
s
1 and kd = 0.04 s
1)
and flow rates (100 µl/min), but with different receptor densities (RT1 = 50 RU and RT2 = 15 RU). Note that when the reaction is not influenced by transport,
for the same analyte concentration, the binding responses are identical
for the two surfaces when normalized with respect to the maximum
response, i.e., a plot of the fraction of receptor sites bound versus
time is independent of RT. Under these
conditions, the rapid mixing model (Eq. 8b, with C = CT) gives an excellent fit to the data (not
shown) and returns correct parameter values, as shown in Table
1A, i.e., the best fit values of the
parameters match the values used to simulate the data. The standard
deviation of the residuals equals 0.054 RU, which is small but nonzero.
Because the ODE models used to fit the data are always approximations
to the PDE model used to simulate the data, even in the absence of
noise, the standard deviation of the residuals is nonzero. As shown
in Table 1B, applying the two-compartment model to these same data sets
gives essentially the same results. The additional parameter,
M, has little effect on the quality of
the fit (std residuals = 0.053 RU), the returned parameter values,
standard deviations, or the correlation coefficients. However, the
value returned for
M of 2.9 × 1012 s
1, which from Eq. 11 corresponds to
kM = 9.7 cm/s for MW = 3 × 104 Da, is grossly in error. From Eq. 9, the predicted
value for kM is 2.6 × 10
3
cm/s. This error is not surprising. The association rate is so slow
that the binding is unaffected by transport, and so the data contain no
information for the determination of kM.

View larger version (17K):
[in this window]
[in a new window]
|
FIGURE 8
Binding response simulated for a reaction occurring
under conditions where the binding is reaction limited. The parameter
values used in the simulations were ka = 5 × 104 M 1 s 1,
kd = 0.04 s 1,
D = 1 × 10 6 cm2/s,
and vc = 10 cm/s. The surface receptor
concentrations were 0.167 nM cm (a) and 0.05 nM cm
(b), which, for an analyte with MW = 3 × 104, correspond to a maximum response of 50 RU and 15 RU,
respectively. The ligand concentrations used in the simulation were
8000, 2670, 890, 297, and 99 nM. The results of the fit are given in
Table 1.
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 1
Parameter estimates, standard deviations, and correlations
for the fits of the rapid mixing model and the two-compartment model to
data simulated in Fig. 8 with ka = 5 × 104 M 1 s 1 and
kd = 4 × 10 2
s 1
|
|
Next we carried out the same fitting procedures on data simulated with
a 10-fold higher association rate constant
(ka = 5 × 105
M
1 s
1). All other parameters were the same.
Again a good fit was obtained with the rapid mixing model (std
residuals = 0.062 RU), yielding parameter values that are close to
the input values (Table 2A). As shown
in Table 2B, using a two-compartment model improves the quality of the
fit somewhat (std residuals = 0.055 RU). Furthermore, the best fit
value
M = 8.9 × 108
s
1, which corresponds to kM = 3.0 × 10
3 cm/s, is greatly improved, being within
13% of the expected value. The improvements in the residual standard
deviation, rate constant estimates, and the value of
kM suggest that under these conditions, transport is beginning to influence the binding kinetics.
View this table:
[in this window]
[in a new window]
|
TABLE 2
Parameter estimates, standard deviations, and correlations
for the fits of the rapid mixing model and the two-compartment model to
data simulated with ka = 5 × 105 M 1 s 1 and
kd = 4 × 10 2
s 1
|
|
Fig. 9 illustrates what happens to the
shapes of the binding responses when the association rate constant is
increased to 1 × 108 M
1
s
1 and the data become strongly influenced by transport.
(Note that the off-rate constant was increased to 0.08 s
1, so that the shapes of the progress curves would
remain similar.) Without doing any data fitting, one can see (compare
Fig. 9 c with 9 d) that the binding is transport
limited because, unlike in Fig. 8, the shapes of the binding curves for
the two surfaces are not identical. Now equilibrium is approached more
rapidly on the surface with the lower receptor density. When the
binding is influenced by transport, the fraction of receptor sites
bound is no longer independent of RT. Receptors
no longer act independently. In the association phase they compete for
analyte, whereas in the dissociation phase rebinding occurs.

View larger version (38K):
[in this window]
[in a new window]
|
FIGURE 9
Binding response simulated for a reaction occurring
under conditions where the binding is transport limited. The parameter
values used in the simulations were ka = 1 × 108 M 1 s 1 and
kd = 0.08 s 1. All other
parameters were the same as in Fig. 8. The surface receptor
concentrations correspond to a maximum response of 50 RU
(a, c) and 15 RU (b,
d) for a 3 × 104 Da analyte. The
analyte concentrations used in the simulation were 8, 2.67, 0.89, 0.297, and 0.099 nM. Shown are simultaneous fits (red
lines) of the rapid mixing model (a,
b) and the two-compartment model (c,
d) to the simulations. The results of the fits are given
in Table 3.
|
|
The fit of the rapid mixing model to the simulated data (Fig. 9,
a and b) is poor (residual std = 1.52 RU),
whereas the fit of the two-compartment model (Fig. 9, c and
d) is excellent (residual std = 0.07 RU).
Furthermore, in Table 3A we see that
the two-compartment model returns parameter values that are close to
those used in the simulations. The best fit value
M = 7.72 × 108 RU
s
1, or equivalently, kM = 2.6 × 10
3 cm/s, is within 1% of the predicted value. These
results demonstrate that in the absence of noise, the two-compartment
model returns accurate estimates for both the intrinsic reaction rate
constants and the transport coefficient when the binding kinetics are
influenced by transport.
View this table:
[in this window]
[in a new window]
|
TABLE 3
Parameter estimates, standard deviations, and correlations
for the fits of the two-compartment model to data simulated with
ka = 1 × 108 M 1
s 1 and kd = 8 × 10 2 s 1 in the absence of noise (Fig. 9) and
the presence of noise (Fig. 10)
|
|
The accurate determination of kM suggests that
BIACORE can be used to determine the diffusion coefficient of the
analyte by calculating D from Eq. 9. The value of
M in Table 3A gives D = 9.85 × 10
7 cm2/s, which is remarkably
close to the value used in the simulation of 1.0 × 10
6 cm2/s. However, because Eq. 9 is an
approximate result, to determine D from
kM, one must ensure that the experiments are
done in a parameter range in which Eq. 9 is sufficiently accurate (see
Discussion).
Finally, to study the effects of noise on the best fit values of the
parameters, we added 0.5 RU of pseudo-normally distributed noise to the
data sets of Fig. 9. This level of noise is consistent with what is
seen in typical experiments done on BIACORE 2000 (Morton and Myszka,
1998
). Fig. 10 shows the results of
fitting the two-compartment model to these data. The residual standard deviation was 0.504 RU, which is consistent with the level of noise
added to the data. The best fit parameter values are given in Table 3B.
Surprisingly, this level of random noise had little effect on the
determined parameter values or the correlation coefficients, although
as expected, the standard errors were increased. To demonstrate that
the linear approximation statistics returned for the nonlinear analysis
were valid, we performed Monte Carlo analysis to generate confidence
probability distributions for the estimated parameters (Straume and
Johnson, 1992
). One thousand data sets were simulated with noise and
solved as before with varied starting parameters. The results of this
analysis confirmed that the confidence intervals were normally
distributed (data not shown), and the standard deviations for the
confidence intervals (Table 3C) were similar to those determined from
the linear approximation statistics (Table 3B). The Monte Carlo
analysis routines are readily available in the CLAMP program.

View larger version (23K):
[in this window]
[in a new window]
|
FIGURE 10
Simulations of the binding response of Fig. 9 in the
presence of noise. Random noise with a standard deviation of 0.5 RU was
generated and added to the simulations of Fig. 9. As in Fig. 9, the
surface receptor concentrations correspond to a maximum response of 50 RU (a) and 15 RU (b) for a 3 × 104 Da analyte. Shown are simultaneous fits (red
lines) of the two-compartment model to the simulations. The
results of the fits are given in Table 3.
|
|
 |
DISCUSSION |
The binding kinetics in a BIACORE flow cell are the result of
diffusion and flow, which bring the analyte to, and take it away from,
the sensor surface, and the chemical reaction that occurs between
analyte and receptor at the sensor surface. We have modeled these
processes for the case of a monovalent analyte interacting with a
monovalent receptor, when the conditions under which the receptor is
immobilized do not affect the binding kinetics. If, for example, the
receptor is coupled to the sensor surface through a dextran layer, then
our model is restricted to conditions where the layer has no influence
on the binding kinetics. As discussed in the Appendix, numerous authors
have considered this case and argued that a two-compartment model gives
a reasonable description of the binding kinetics when the data are
influenced by transport (Glaser, 1993
; Karlsson et al., 1994
; Schuck
and Minton, 1996
; Christensen, 1997
), although its use in analyzing
such data has been questioned (Schuck, 1997b
). To see if, in fact, a
two-compartment model can be used to extract accurate parameter values
from transport-influenced BIACORE binding data, we generated test sets
of simulated data. The obvious advantage of using simulated data sets
is that the values of the underlying parameters are known exactly. We
extended the model of Glaser (1993)
to include diffusion in the
direction of flow, obtaining a partial differential equation with
appropriate boundary conditions that accounts for the processes that
occur in a flow cell: convection, diffusion, and reversible chemical reaction at the sensor surface. The simulated data sets were generated by solving this model numerically, to obtain noiseless data, and then
adding noise to the numerical solutions. These simulated data sets
served as our standard for testing the two-compartment model. We showed
that when the two-compartment model was fit to the simulated data,
values for the rate constants were returned with surprisingly high
accuracy.
The data we simulated corresponded to optimal experimental conditions
currently available on BIACORE 2000, a flow rate of 100 µl/min and
surface receptor densities as low as 15 RU. Our data analysis method
involved simultaneously fitting the two-compartment model to two sets
of simulated data, corresponding to two different receptor densities,
15 RU and 50 RU. The idea behind the approach, known as global analysis
(reviewed in Myszka et al., 1997
), is to obtain data that are sensitive
to both the chemical rate constants (data from the low-capacity
surface) and the transport coefficient (data from the high-capacity
surface). In our fits we showed that even though the maximum response
from one of the surfaces was only 15 RU, the addition of the expected
level of experimental noise (0.5 RU) did not affect our ability to
accurately recover the reaction rate constants.
It has been pointed out that if receptors are immobilized in a dextran
layer, the binding kinetics can be influenced by the transport of the
analyte within the layer (Schuck, 1996
; Yarmush et al., 1996
). It is
expected that such effects will become pronounced at high receptor
densities. At such densities the two-compartment model will no longer
give a reasonable description of the binding kinetics. Elsewhere we
discuss how to reformulate the model when the thickness of the layer
cannot be neglected (Wofsy et al., manuscript in preparation). There we
show that the effects of the layer will be negligible when the layer
"looks" thin to the analyte, i.e., when the average distance the
analyte travels in the layer before it binds to a receptor (its mean
free path) is long compared to the height of the layer, d.
The mean free path equals
,
where RT/d is the three-dimensional receptor concentration and Di is the diffusion
coefficient in the layer. Thus one can ignore the layer and treat the
binding as if it is directly to the sensor surface when 1
. (For the low receptor densities, we expect Di
D. In our simulations the 50 RU surface corresponded to
RT = 0.167 nM cm and D = 1 × 10
6 cm2/s. For a standard flow cell,
d
1 × 10
5 cm. For these values
the inequality breaks down when ka
1 × 108 M
1 s
1.)
In addition to determining the rate constants, we showed that the
two-compartment model can be used to determine D, the
diffusion coefficient of the analyte. This was done by determining the
transport coefficient, kM, by fitting the
simulated data, and then calculating D from Eq. 9. Because
Eq. 9 is an approximate result, it can be used to determine
D only under conditions in which it is highly accurate.
Equation 9 overestimates kM. It is derived by
ignoring the parabolic flow velocity profile and taking the velocity to rise linearly with the height above the sensor surface. It also ignores
the upper boundary of the flow cell. Analyte that diffuses to this
boundary is not reflected. However, if the time it takes to travel the
length of the flow cell, l/
= 3l/(2vc), is short compared to the
time it takes to diffuse to the boundary,
h2/(4D), or equivalently, if 1
6lD/(h2vc),
then we expect Eq. 6 to be valid. Thus to determine D from Eq. 9, one should use a fast flow velocity for the binding studies.
We have shown that the two-compartment model can be used to determine
accurate values of the rate constants and transport coefficients for
data that are influenced by flow and diffusion. From the transport
coefficient the analyte diffusion coefficient can be determined as
well. Using the two-compartment model and analyzing the data globally
extends the range of reaction rates that can be determined with
instruments like BIACORE. At present, this approach seems to be the
best available for analyzing experiments when the effects of mass
transport cannot be eliminated from the binding kinetics. This model
has been incorporated in BIAevalution 3.0 analysis software available
from BIACORE, Inc. It has already been used to analyze such
experimental data, yielding excellent fits to the data and improved
accuracy of the returned rate constants (Myszka et al., 1996
; Myszka et
al., 1997
; Morton and Myszka, 1998
).
The authors acknowledge helpful discussions with C. Wofsy.
This work was supported by National Institutes of Health Grant GM35556
and performed under the auspices of the U.S. Department of Energy.
Address reprint requests to Dr. Byron Goldstein, Theoretical Biology
and Biophysics Group, Theoretical Division, T-10, MS K710, Los Alamos
National Laboratory, Los Alamos, NM 87545. Tel.: 505-667-6538; Fax:
505-665-3493; E-mail: bxg{at}lanl.gov.