| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, September 1998, p. 1144-1162, Vol. 75, No. 3
*Department of Medicine, A difficulty of using confocal microscopy to study
Ca2+ sparks is the uncertainty of the linescan position
with respect to the source of Ca2+ release. Random
placement of the linescan is expected to result in a broad distribution
of measured Ca2+ spark amplitudes (a) even if
all Ca2+ sparks were generated identically. Thus variations
in Ca2+ spark amplitude due to positional differences
between confocal linescans and Ca2+ release site are
intertwined with variations due to intrinsic differences in
Ca2+ release properties. To separate these two sources of
variations on the Ca2+ spark amplitude, we determined the
effect changes of channel current or channel open time Calcium (Ca2+) "sparks" are
brief, spatially localized Ca2+ release events resulting
from the opening of one or a cluster of sarcoplasmic reticulum (SR)
Ca2+ release channels. The combination of laser scanning
confocal microscopy and the fluorescent Ca2+ indicator,
fluo-3, has revealed Ca2+ sparks in single cardiac cells
(Cheng et al., 1993 The observation of spontaneous Ca2+ sparks (Cheng et al.,
1993 While the trigger for SR Ca2+ release is understood, how
the release is controlled is still unclear. A myriad of factors appear to control SR Ca2+ release. For example, 1) SR load affects
both the probability of Ca2+ spark occurrence and the
Ca2+ spark amplitudes (Satoh et al., 1997 To better understand the mechanisms underlying the control of
Ca2+ release we need to estimate the current through the SR
Ca2+ release channel, but this current cannot be measured
directly in an intact cell and must be estimated from the amplitude of the Ca2+ spark. This estimate is complicated by 1) the
kinetics and capacity of Ca2+ buffering by endogenous
Ca2+ buffers and exogenous Ca2+ buffers, such
as the Ca2+ indicator fluo-3 (Balke et al., 1994 Pratusevich and Balke (1996) To understand the relationship between the measured
Ca2+ spark properties and the underlying events we need to
simulate the processes that influence the formation of the
Ca2+ spark. These processes are (1) the release of
Ca2+ from the SR by the opening of a Ca2+
release channel; (2) the diffusion of Ca2+ into the
cytoplasm; (3) the reaction of Ca2+ with endogenous
buffers, such as troponin-C, and the fluorescent Ca2+
indicator fluo-3; (4) the formation of the optical image of the Ca2+-bound fluo-3 dye; (5) the generation of a linescan
image from the optical signal; (6) the generation of random
fluctuations of the fluorescent signal due to photon and other sources
of noise; and (7) the detection of the Ca2+ spark.
Reaction-diffusion equations
Processes (1), (2), and (3) are captured in the set of partial
differential equations describing the reaction of Ca2+ with
buffers and the diffusion of Ca2+ in the cytoplasm. The
solution of these equations gives the 3-dimensional distribution of the
Ca2+-bound fluo-3 as a function of time.
The chemical species included in the model equations are
Ca2+ (concentration denoted by C), immobile
endogenous buffers both free (Fb) and bound to
Ca2+ (Gb), mobile fluo-3 (free
Fm, bound Gm), and
immobilized fluo-3 (free Fi, bound
Gi). Mass transport of Ca2+ and
mobile fluo-3 is assumed to follow Fick's law and the reaction rates
(Rj) are governed by mass action kinetics. Thus
the reaction rate between Ca2+ and any buffer is given by
![]()
ABSTRACT
Top
Abstract
Introduction
Methods
Results
Discussion
References
collectively
called the source strength, 
had on the measured Ca2+
spark amplitude histogram, N(a). This was done by 1)
simulating Ca2+ release, Ca2+ and fluo-3
diffusion, and Ca2+ binding reactions; 2) simulation of
image formation of the Ca2+ spark by a confocal microscope;
and 3) using a novel automatic Ca2+ spark detector. From
these results we derived an integral equation relating the probability
density function of source strengths, f
(
),
to N(a), which takes into account random positional variations between the source and linescan. In the special, but important, case that the spatial distribution of Ca2+-bound
fluo-3 is Gaussian, we show the following: 1) variations of
Ca2+ spark amplitude due to positional or intrinsic
differences can be separated, and 2) f
(
)
can, in principle, be calculated from the Ca2+ spark
amplitude histogram since N(a) is the sum of shifted
hyperbolas, where the magnitudes of the shifts and weights depend on
f
(
). In particular, if all
Ca2+ sparks were generated identically, then the plot of
1/N(a) against a will be a straight line.
Multiple populations of channels carrying distinct currents are
revealed by discontinuities in the 1/N(a) plot. 3) Although
the inverse relationship between Ca2+ spark amplitude and
decay time might be used to distinguish Ca2+ sparks from
different channel populations, noise can render the measured decay
times meaningless for small amplitude Ca2+ sparks.
![]()
INTRODUCTION
Top
Abstract
Introduction
Methods
Results
Discussion
References
; López-López et al., 1995
), cardiac
trabeculae (Wier et al., 1997
), skeletal muscle cells (Tsugorka et al.,
1995
; Klein et al., 1996
), and smooth muscle cells (Nelson et al.,
1995
).
) and the description of the voltage dependence of evoked
Ca2+ sparks (López-López et al., 1995
; Cannell
et al., 1995
) provided important experimental support for the local
control theory of cardiac excitation-contraction coupling (Stern,
1992
). It is now generally accepted that Ca2+ current
through L-type Ca2+ channels locally triggers SR
Ca2+ release. Evidence for this comes indirectly from
measurements of Ca2+ sparks under whole cell voltage clamp
(López-López et al., 1994
, 1995
; Santana et al., 1996
) and
more directly from measurements of Ca2+ sparks localized to
the region under a cell-attached membrane patch (Shorofsky et al.,
1998
).
; Györke et
al., 1997
; Song et al., 1997
); 2) the number of SR Ca2+
release channels opening to generate a Ca2+ spark might be
variable (Lipp and Niggli, 1996
) and produce Ca2+ sparks of
different amplitudes (Parker and Wier, 1996
); 3) the SR
Ca2+ release channel appears to have a subconductance that
results in two populations of Ca2+ sparks with different
amplitudes (Cheng et al., 1993
; Xiao et al., 1997
); 4) Ca2+
sparks can trigger other Ca2+ sparks (Klein et al., 1996
;
Parker et al., 1996
; Blatter et al., 1997
), that might appear as
population of Ca2+ sparks with different amplitudes (Klein
et al., 1996
); and 5) the proportion of Ca2+ sparks in
different amplitude populations appears to be altered in certain
disease states (Shorofsky et al., 1996
, 1997
).
; Smith et
al., 1996
); 2) the diffusion properties of free Ca2+ and
mobile Ca2+ buffers (Wagner and Keizer, 1994
; Smith et al.,
1996
); and 3) the uncertainty of the distance between the site of
Ca2+ release and the position of the confocal linescan
(Pratusevich and Balke, 1996
; Shirokova and Ríos, 1997
).
showed that the random placement of the
confocal linescan relative to the Ca2+ release sites
results in a broad distribution of measured Ca2+ spark
amplitudes even if all Ca2+ sparks were generated
identically. Thus it becomes difficult to distinguish variations in
Ca2+ spark amplitude due to intrinsic variations
in SR Ca2+ release channel current from
positional variations arising from varying distance between
the confocal linescan and the site of Ca2+ release. In this
paper we show, in theory, that variations in Ca2+ spark
amplitude arising from intrinsic and positional changes can be
separated. We do this by deriving an integral equation that gives the
relationship between the probability distribution of source strengths
(SR Ca2+ release channel current or channel open time) and
the Ca2+ spark amplitude histogram. This integral equation
takes into account the effect of positional variations on the
Ca2+ spark amplitude histogram. The integral equation can
be solved analytically, allowing us to explicitly solve for the
probability distribution of source strengths when given the measured
spark amplitude histogram. In this way the variations in
Ca2+ spark amplitude due to intrinsic and positional
changes, which are intertwined in the amplitude histogram, can be
separated.
![]()
METHODS
Top
Abstract
Introduction
Methods
Results
Discussion
References
where j = i, m, or b and
kj+ and kj
(1)
are
the forward and reverse rate constants, respectively.
We also make the following assumptions: (1) Ca2+ released from the SR is approximated by a point source; (2) both reaction and diffusion occur radially symmetrically; (3) the diffusion coefficients of the Ca2+-bound mobile fluo-3 and the free mobile fluo-3 are identical; and (4) before the opening of the Ca2+ channel, all chemical species are at their steady-state value and there are no spatial gradients.
Under these assumptions the reaction-diffusion equations are
|
(2) |
|
(3) |
|
(4) |
|
(5) |
|
(6) |
|
(7) |
2 for the radially
symmetric domains is
|
(8) |
(r), where
(r) is
the Dirac delta-function. JSR is related to the
Ca2+ release channel current ISR by
JSR = ISR/(z
) where z = 2 is the Ca2+ valence and
is Faraday's constant.
A typical value of ISR is 1.4 pA.
The reason there are no equations for Gj is
because under assumption (3) the sum Hj
satisfies the linear diffusion equation
Hj/
t = Dj
2Hj, and under
assumption (4) the initial condition satisfies
Hj(r, 0) =
j = constant, so the diffusion equation has the solution
Hj(r, t)
j.
Accordingly, Fj and Gj
satisfy the algebraic relationship Gj = Hj
Fj =
j
Fj.
For simplicity, we did not include a Ca2+ pump in the model
because others (Gómez et al., 1996
) found that 80% of the
decline of the Ca2+ fluorescence signal could be accounted
for by diffusion and buffering.
The diffusion coefficient for Ca2+ was set to 6 × 10
6 cm2/s. The apparent diffusion coefficient
of fluo-3 is found to be 0.2 × 10
6
cm2/s in frog skeletal muscle, which is about a factor of 5 times smaller than predicted from its molecular weight (Harkins et al., 1993
). To account for this difference Harkins et al. (1993)
estimated that 78% of the dye is bound to immobile myoplasmic constituents and
only 22% is freely mobile. In our simulations the ratio of concentrations (
) of immobile to mobile fluo-3,
was set to 5 or
in some cases 2. This latter value comes from prior determination of
the ratio of immobile to mobile fura-2 in guinea pig heart cells
(Blatter and Wier, 1990
). Dm was set to 0.9 × 10
6 cm2/s. The free mobile fluo-3
concentration was fixed to 50 µM in all simulations.
The rate-limiting step of most reactions involving Ca2+ is
the dehydration of the calcium ion and is ~200-700/µM s (Hague,
1977
). We chose a value for the forward rate constant
km+ to be near the middle of the range,
400/µM s. The reverse rate constant, km
= 160/s, was calculated using the dissociation constant value of 400 nM.
The total concentration of endogenous buffers
b was set to 123 µM (Berlin et al.,
1994
). The forward rate constant kb+ was
chosen to be 100/µM s and the reverse rate constant of
kb
= 100/s to give the endogenous buffer
dissociation constant of 1 µM, close to the value (0.96 µM) found
by Berlin et al. (1994)
.
The experimental parameters were determined at room temperature
(20-25°C) except for Harkins et al.'s estimate of
= 5, which was measured at 16°C. No temperature compensation was made.
Numerical solution
Eqs. 2-5 were solved using Facsimile (AEA Technologies, Harwell, UK), which solves each of the equations in the time variable on a workstation (RS 6000, IBM Corp., Armonk, NY). The spatial domain extending from 0 < r < L was discretized into N compartments of equal lengths h = L/N. The ith compartment is the spherical shell bounded by ri = ih and ri+1 = (i + 1)h. The time rate of change of the concentration in the ith compartment is
|
(9) |
|
i2h2 is the surface area of the
sphere of radius ri and
Vi = 4
h3(i2 + i + 1/3) is the volume between the spheres of radius
ri and ri+1.
Ri is the reaction term.
Ji is given by Fick's law Ji =
(D/h)(ci
ci
1). The differential equation for the Nth compartment is derived similarly, but imposing the
zero-flux boundary condition that requires the fictitious point
cN+1 = cN. For
i = 0, which contains the point source, material
balance yields
|
(10) |
h3/3.
We used L = 6 µm and N = 600. This code was tested on the linear problem obtained by setting all reaction terms to zero for which the analytic solution was available for comparison. Except at very early times following channel opening (~10 µs) the relative error was within 5% even at the smallest resolvable distance of r = 0.015 µm. In all simulations the concentration at r = 6 µm did not vary over the short (<200 ms) integration time so any of the usual boundary conditions (Dirichlet, Neumann, or Robin) would give essentially identical results.
Formation of the optical image
The Ca2+ spark is the optical image of the distribution of Ca2+-bound fluo-3, Gm(r, t) + Gi(r, t). Any optical instrument forms a blurred image of the object and the extent of the blurring is given by the instrument's point spread function (PSF). We used a 3-dimensional Gaussian function as the PSF of the confocal microscope
|
(11) |
xy2
z
3/2)
1
normalizes the integral of the PSF over all space to unity. The
standard deviation
is related to the confocal full-width at
half-maximum (FWHM) by
= FWHM/[2(log 2)1/2]. A
typical value for the lateral FWHM is FWHMxy = 0.4 µm.
Values for the axial FWHM range from 0.41 µm (Parker et al., 1997The intensity contributions to the image of the Ca2+-bound fluo-3 at any point (x, y, z) measured from the point source at the origin is proportional to the convolution
|
(12) |
|
Generating a linescan image
Generation of the linescan image of a Ca2+ spark
starts by choosing the linescan position (y*, z*) in the
y
z plane, which is perpendicular to the linescan
direction along x. Then for each time point
tj values of I(x, y*, z*,
tj) for all x are collected. Stacking these
one-dimensional arrays for all the computed times (0 < tj < 180 ms) produces the linescan image of the
Ca2+ spark. The length of the linescan image along
x is 4 µm and 180 ms in time. This small linescan image is
embedded in a larger array whose values are set to the image value of
bound fluo-3 (Gm + Gi) at
equilibrium. Additionally, multiple Ca2+ sparks can be
embedded in the large array at random positions, with the constraint
that Ca2+ sparks do not overlap. The result of this
embedding is an image that looks qualitatively like a linescan image
from a real confocal microscope. A sample image in which signal
fluctuations have been added is shown in Fig. 1
B.
|
We created realistic linescan images because we wanted observers to identify these simulated Ca2+ sparks in order to study the role of subjective factors on Ca2+ spark identification.
Random fluctuations of the fluorescent signal
Random fluctuations of the fluorescent signals arise from the
intrinsic granularity of photons and from electronic noise (Pawley, 1995
). To accurately model noise in simulated confocal linescan images,
we measured the noise properties in linescan images made from the
homemade confocal microscope (Parker et al., 1997
). The mean signal
level and the standard deviation were calculated in a 10 × 10-pixel sample area in three regions: background, where the
fluorescence intensity was low and uniform; regions containing a narrow
band of elevated fluorescence at the site of the t-tubules (Shacklock
et al., 1995
; Klein et al., 1996
); and at Ca2+ sparks. Fig.
2 shows that the standard deviation of
the values in the sample areas increase linearly (slope = 0.3)
with the mean fluorescence, and this linear relationship holds
regardless of the sample region. Moreover, the distribution of noise
values is approximately Gaussian (data not shown). We therefore added to the value of each point in our simulated linescan image a random number from a Gaussian distribution whose standard deviation was 0.3 times the value at that point.
|
Automatic detection of Ca2+ sparks
We developed a program to automatically identify Ca2+ sparks in linescan images. This program relieves the tedium of manually identifying Ca2+ sparks and ensures a more consistent choice of Ca2+ sparks than might be achieved by observers. This program identifies as Ca2+ sparks regions that have a sufficiently high density of pixels that exceed some threshold level. Identification of Ca2+ sparks starts by creating a binary image in which all pixels in an image whose value is less than the background level + threshold are set to zero, and all other pixels to unity. The threshold equals the standard deviation of the background signal times a factor (typically 1.4) that can be varied by the user.
High density regions of non-zero pixels are identified by using the following procedure iteratively. At every pixel in the image (i, j), the number of non-zero pixels within a square neighborhood of size Nsize centered on (i, j) are counted. If this number is less than Nlive, then the (i, j) pixel is set to 0 (i.e., the pixel "dies"); otherwise it is set to 1 (i.e., the pixel "survives" or is "born"). This procedure is repeated Ngeneration times for every pixel. As the notation suggests, this algorithm is based on ideas gleaned from modeling density-dependent population growth using cellular automata. Although the procedure appears slow and tedious, it in fact runs quickly with the array-oriented programming language IDL (Research Systems, Inc., Boulder, CO). The number of "live" neighbors a pixel has is found by doing a boxcar averaging of size Nsize × Nsize (typically Nsize = 7) on the binary image. This smoothed array is thresholded-setting all pixels whose value is less than Nlive to 0 and 1 otherwise.
Before processing actual linescan images, the prominent horizontal lines seen in many images (see Fig. 1 A) are removed to avoid being identified as potential Ca2+ sparks by the detection program. This is done by setting the zero frequency component (corresponding to time) of the image's Fourier transform to zero. The linescan image without horizontal lines is recovered by inverting the modified transform.
| |
RESULTS |
|---|
|
|
|---|
Confocal images of Ca2+ sparks
Fig. 1 A shows a linescan image of a rat ventricular
cell obtained by using our homemade confocal microscope system (Parker et al., 1997
). This system has a lateral FWHM of 0.31 µm and an axial
FWHM of 0.41 µm. The bottom panel shows a simulated linescan in which
both axial and lateral FWHM values were set to 0.35 µm. Setting the
axial FWHM equal to the lateral FWHM greatly simplifies the analytic
calculations without sacrificing much accuracy.
Before going to a quantitative description of the simulated
Ca2+ sparks we point out two features of the simulated
linescan. First, 1 A shows prominent horizontal lines
that are spaced ~2 µm apart vertically; they originate on the
t-tubules (Shacklock et al., 1995
) and may arise from inhomogeneous
distribution of dye. These lines are absent in the simulated image as
we have assumed that the dyes, both mobile and immobile, are initially
homogeneously distributed. Second, apart from the absence of the
streaks, the simulated linescan image looks qualitatively
like the linescan from an actual experiment.
Fig. 3 A illustrates how the simulated Ca2+ sparks are generated. The point source of Ca2+ release is assumed to be at the origin in panels A-C. The circles in the figure show the randomly chosen positions of the linescan in the y-z plane. Each circle corresponds to at least one Ca2+ spark (some positions may be chosen more than once) in the 200 linescan images made for this particular simulation. The triangles indicate the position of those Ca2+ sparks that were detected by the Ca2+ spark detection program. The arc encloses the region where 90% of the Ca2+ sparks were detected; the arc radius R90 is 0.56 µm. The linescan positions of Ca2+ sparks labeled a and b in Fig. 1 are correspondingly labeled in Fig. 3 A. As expected, the bright Ca2+ spark (a) arose when the linescan was near the point source and the dim Ca2+ spark (b) arose when the linescan was far away.
|
Performance of the Ca2+ spark detection program
As shown in Fig. 3 A a large number of Ca2+ sparks in the linescan images go undetected. The sensitivity of the program to pick out dim Ca2+ sparks can be altered by changing the detection parameters Nsize, Nlive, and Ngeneration. Decreasing Nsize or Nlive increases the sensitivity of the program allowing detection of dimmer Ca2+ sparks, but at the expense of making more false identifications. Increasing Ngeneration has only a small effect on the sensitivity but reduces the number of false identifications. We could check the false identification rate because the positions of all Ca2+ sparks in the linescans were known. Note that falsely identified Ca2+ sparks were excluded from our measurements. The program parameters were adjusted empirically to achieve a balance between sensitivity and low false identification rate. We found that by using Nsize = 7, Nlive = 12, and Ngeneration = 3 the program identified all Ca2+ sparks correctly identified by observers and correctly identified dim Ca2+ sparks not identified by observers, while maintaining a false identification rate of ~2-5%. The number of dim Ca2+ sparks found by the program that was not detected by observers varied between observers but the program typically found ~50% more of the dimmest detectable Ca2+ sparks. The processing time for 100 linescan images, 166 pixels (500 ms) by 256 pixels (25.6 µm) in size, is ~2 min on an IBM RS-6000 workstation.
Properties of simulated Ca2+ sparks
Ca2+ sparks shown in Fig. 1 B were
generated using a channel current of 1.4 pA, a channel open time of 10 ms (Rousseau and Meissner, 1989
; Lukyanenko et al., 1996
) with
= 5. We also ran an identical simulation except with
= 2. Ca2+ spark characteristics from both simulations are shown
in Table 1.
|
The time for the fluorescence (that is, Gi + Gm) to decrease from peak value (measured at the
brightest point of the Ca2+ spark) to half its value to the
baseline is t1/2. The peak ratio, or
Ca2+ spark amplitude a, equals F/Fo where F is
the peak fluorescence value and Fo is the baseline fluorescence value.
The mean Ca2+ spark amplitude is given by
F/Fo
and
the maximum ratio [occurring when (y, z) = (0, 0)] is
F/Fo(max). The spatial spread of the Ca2+ spark at the time
of the peak fluorescence is characterized by the FWHM. Because of the
large variation in t1/2 and FWHM, we also
calculated these values [t1/2(bright),
FWHM(bright)] using only the 10 brightest Ca2+ sparks.
Typical t1/2 values for Ca2+
sparks from heart cells is ~20 ms (Cheng et al., 1993
), which is
close to that found when
= 5 but not when
= 2. Note that the
standard deviations are quite large, about half the mean
t1/2 value. The reason for this large variation
is shown in Fig. 4 A where
t1/2 is plotted against the Ca2+
spark amplitude. The variation in t1/2 is fairly
small for the large amplitude Ca2+ sparks but is large for
the low amplitude Ca2+ sparks because of noise.
|
Fig. 4 B shows a plot of t1/2 against amplitude for the same set of simulations in Fig. 4 A but in the absence of noise. Since the Ca2+ sparks were generated identically, amplitude variations are due solely to variations in distance between linescan and Ca2+ spark origin. The decay time of identically generated Ca2+ sparks is controlled by the diffusion of Ca2+ into the scanned volume, so it increases with distance and, equivalently, decreases with Ca2+ spark amplitude. Viewed in isolation, Fig. 4 B suggests that the decay time could be used to distinguish whether a Ca2+ spark has a small amplitude because the linescan was far from the source or because the source strength was small. The results in Fig. 4 A cautions against such a method as virtually any decay time may be obtained for small amplitude Ca2+ sparks.
The mean Ca2+ spark amplitude is almost identical for
= 2 and 5 and is typical for experimentally measured
Ca2+ sparks. F/Fo(max) values are also similar for the two
values of
, indicating that despite the larger amount of dye
available when
= 5 the amount of Ca2+ released is
sufficient to saturate the dye.
The FWHM values for the 10 brightest Ca2+ sparks is ~2
µm, which is about half the value reported by Gómez et al.
(1996)
for rat ventricular cells. Simulations carried out with longer
open times or larger channel currents did not greatly alter the FWHM values.
Effect of changing microscope's FWHM
The triangles in Fig. 3 A showing the linescan
positions at which the Ca2+ spark could be detected are
symmetrically distributed around the origin, as expected since the
axial and lateral FWHM values are equal. To study how this distribution
changes when the blurring kernel is asymmetric, we increased the
z/
xy ratio to 3 (Fig. 3 B) and
6 (Fig. 3 C) where
xy was fixed to 0.2 µm.
Note that the confocal parameters are different from those used to
generate Fig. 3 A. The case where
z/
xy = 1 is not shown since the
distribution of detected Ca2+ sparks is symmetric, as in
Fig. 3 A. (Fig. 3, A-C may be interpreted in two
equivalent ways: the point source is at the origin and the circles
represent the linescan positions, or the linescan is fixed at the
origin and the circles mark the point source locations. We take the
latter viewpoint now so we can talk about the distribution of
detected Ca2+ sparks instead of the more unwieldy
distribution of linescan positions at which the Ca2+ spark
was detected.) Fig. 3 B shows, surprisingly, that the
distribution of detected Ca2+ sparks is still symmetric
about the origin despite the axial FWHM being 3 times larger than the
lateral FWHM. The distribution of detected Ca2+ sparks
becomes asymmetric, however, when
z/
xy = 6, as shown in Fig. 3 C.
We were initially surprised to see the distribution in Fig. 3
B because we had expected to see an ellipsoidal distribution that parallels the elongation of the PSF along the z-axis.
With increases in the depth of field (increasing
z)
comes a loss in contrast of the Ca2+ spark, making it more
difficult to detect the Ca2+ spark. This decrease in
contrast with increases in
z is shown in Fig. 3
D (right to left) where the same Ca2+ spark is
imaged with
z equaling 0.2, 0.6, and 1.2 µm,
respectively. (No microscope to date has achieved an axial resolution
of ~0.2 µm, but we have used this value for illustration.) The F/Fo
values for these three cases are 2.19, 2.64, and 3.12 (left to right) yielding Ca2+ concentration values of 312, 447, and 664 nM
(Cheng et al., 1993
). Thus the simple act of opening the confocal
pinhole, which increases both axial and lateral FWHM, can reduce the
Ca2+ concentration estimates.
The physical reason for the decrease in contrast as
z
increases is that because the light energy is spread over a larger volume, the intensity must be lower to maintain energy conservation. Mathematically, this constraint is expressed in the larger denominator (~
z) in the normalization factor of the Gaussian
kernel in Eq. 11.
Ca2+ spark amplitude distribution
Ca2+ spark amplitude distributions obtained from simulated linescan images are shown in Fig. 5. All Ca2+ sparks were generated with a channel current of 1.4 pA and channel open time of 10 ms; only the linescan positions were varied randomly. In the absence of noise, panel A, the Ca2+ spark amplitude distribution decreases monotonically except for statistical sampling variations. [Nonmonotonicity in the Ca2+ spark amplitude distribution due to sampling variation can be distinguished from intrinsically multimodal distributions (for example when the SR Ca2+ release channels are arranged on a lattice, see below) by increasing the sample size or by changing the seed value of the random number generator. Intrinsically multimodal Ca2+ spark amplitude distributions are unaffected by these changes.]
|
These graphs illustrate the inherent difficulty in assessing the source
strength distribution. Although all Ca2+ sparks in the
linescan images were generated identically, because of the arbitrary
placement of the linescan relative to the source, there is a broad
distribution of measured Ca2+ spark amplitudes instead of a
single narrow bin or narrow Gaussian distribution. A Gaussian
distribution has been interpreted to indicate that Ca2+
sparks have stereotypic origins. However, Fig. 5 shows that, in our
model, Ca2+ sparks generated identically do not generate a
narrow Ca2+ spark amplitude distribution. This result is
similar to that obtained by Pratusevich and Balke (1996)
.
One way that a monotonically decreasing Ca2+ spark
amplitude distribution, Fig. 5 A, might be transformed into
a Gaussian-like distribution is suggested by Fig. 5 B. In
the presence of noise, Ca2+ sparks whose amplitude was
<~1.2 were not detected by the detection program. Moreover, more
Ca2+ sparks whose amplitudes were in the range 1.3-1.4
were detected than those Ca2+ sparks with amplitudes of
1.2-1.3. When noise is present the Ca2+ sparks of low
amplitude are not detected with the same reliability as the
large amplitude Ca2+ sparks. Thus, although there were
actually more low amplitude Ca2+ sparks in the linescan
images (Fig. 1 B) these Ca2+ sparks are masked
by noise and appear to occur less frequently. The difference
in reliability is quantified by a visibility function proposed by
Pratusevich and Balke (1996)
. The sigmoidal visibility function gives
the probability of detecting a Ca2+ spark of a given
amplitude and ranges from 0 for amplitudes near 1 and rises to unity as
the Ca2+ spark amplitude increases. The Ca2+
spark amplitude distribution that is measured is then the product of
the "ideal" amplitude distribution, obtained by a perfect detector in the absence of noise (Fig. 5 A), and the visibility
function. Multiplying an appropriately shaped visibility function with
an amplitude distribution such as in Fig. 5 A can give a
Ca2+ spark amplitude distribution that is Gaussian-like and
similar to those reported in the literature (Klein et al., 1996
;
Shorofsky et al., 1996
, 1997
; Shirokova and Ríos, 1997
; Xiao et
al., 1997
; Wier et al., 1997
).
The key question is whether the intrinsic properties of the SR Ca2+ release channel, not detector characteristics, produce these experimentally measured Ca2+ spark amplitude distributions. To answer this question we need to establish the relationship between the Ca2+ spark amplitude distribution and the underlying source strength distribution.
Relationship between the Ca2+ spark amplitude distribution and source strength distribution
Let fa(a) be the probability
density function (pdf) of Ca2+ spark amplitudes. That is,
the probability of finding a Ca2+ spark whose amplitude is
between a
a/2 and a +
a/2 is
fa(a)
a. Likewise, let
f
(
) be the pdf of the source strength. The source strength
may refer to the SR Ca2+ release
channel current for a fixed channel open time or to the open time for a
fixed channel current.
To establish the link between fa and
f
consider a simple and intuitive example.
Suppose that a light bulb located at the origin flashes with intensity
' with probability p(
') and flashes with intensity
" with probability p(
") = 1
p(
'). The light
intensity a that an observer measures depends on his/her distance r from the lamp and the lamp intensity
, and is
given by the observation function, g(
, r)
|
(13) |
' and the observer measures intensity
a1 = g(r',
'). If the observer moves
randomly then the mean number of times that he/she measures an
intensity a1 is proportional to the probability of being at a distance r', p(r'), times the probability that
the lamp flashed with intensity
', that is
|
(14) |
" and his/her distance
r" is adjusted accordingly to give a1 = g(r",
"). The appropriate distance is given by
|
(a,
) can always be found provided
the observation function g(
, r) is a strictly monotonic
function of r. Thus the probability
pa of measuring intensity
a1 becomes
|
(15) |
, r) < a.
Fa(a) is the cumulative distribution function
|
(16) |
|
Although
and r are independent random variables, the
values of
and a constrain the lower limit of integration
of r. In order to satisfy g(
, r) < a, the
lower bound of r must be
(
, a). Thus,
|
(17) |
compatible with a given a
is given by
max = g
1(a, r).
Thus Eq. 17 becomes
|
(18) |
|
(19) |
(
) is the main result. We now need to
find specific forms of f
, fr, and
. If the linescan can be at any position between 0
r
R with equal probability, then
fr(r) = 2r/R2. Note that
the use of r and not (y, z) comes from the
implicit assumption that the blurring along the lateral dimensions
x and y is the same as along the
z-axis. Another assumption implicit in the use of
r is that the diffusion is radially symmetric.
Explicit form for the observation function
g(
, r)
Because of the nonlinear buffer reactions, the observation
function cannot be found analytically. We determined
g(
, r) empirically using the following procedure.
Linescan images (100-200) containing a total of ~150-300
Ca2+ sparks were generated with a set of parameters for the
reaction-diffusion simulations and a channel current of
, say 1.4 pA, and a fixed channel open time (10 ms). Ca2+ sparks were
found using the Ca2+ spark detection program and their
amplitudes (a = F/Fo) calculated. Since the
(y, z) coordinates of each linescan were known, the amplitude at the distance r = (y2 + z2)1/2 could be calculated. The pairs of
(a, r) were fit to the function
|
(20) |
), B(
), and C(
).
The observation functions for four different channel currents are shown
in Fig. 6 A. The solid curve
shows the best fit to the data and, for clarity, data points are only
shown for
= 0.7 pA and
= 2.8 pA. A(
) was fit to
the hyperbolic function
|
(21) |
) to the line
|
(22) |
) and
C(
); they were simply chosen for simplicity.
B(
) was essentially independent of
varying between
1.20 and 1.25. This is expected since B should only reflect the sensitivity of the detection program and the amount of added noise.
Moderate (3-5-fold) reduction of km+,
km
, ki+, and
ki
values did not affect the functional
form of the observation function or of A(
) and
C(
).
|
In the next section we will derive a specific relationship between
f
and fa that will
allow us to examine the effects that different source strength
distributions have on the Ca2+ spark amplitude
distribution. The specific relationship between f
and fa depends, of
course, on our assumption that the observation function is Gaussian.
Different observation functions yield different relationships between
f
and fa. Thus, it is
worthwhile to examine the range of conditions under which the
observation function is likely to be Gaussian. When Ca2+
release comes from a point source and the source strength is sufficiently weak so that the dye does not saturate, then the Ca2+ bound fluo-3 distribution is approximately Gaussian.
We assumed that the PSF is Gaussian, which well approximates the actual
PSF for a correctly aligned confocal microscope with a fairly small pinhole. The convolution of the Gaussian Ca2+ bound fluo-3
concentration profile with the Gaussian PSF gives a Gaussian image; the
observation function is the profile of this convolution. We note that
the spatial profiles of many Ca2+ sparks are approximately
Gaussian (Parker et al., 1996
; Gómez et al., 1996
).
The observation function will deviate from a Gaussian when the source
is extended [see Smith et al. (1998)
for a discussion of extended
sources], when the dye is saturated, or the confocal microscope is
poorly aligned. Under these conditions the observation function must be
amended. Later we will see the effect of dye saturation on
fa.
Explicit relationship between f
and
fa
From Eq. 20 it follows that
(a, r) is given by
|
(23) |
|
(24) |
(
) =
(
o), where
is the Dirac delta-function. In
this case Eq. 19 becomes
|
(25) |
is
amin(
o) = g(
o, R) and the largest is
amax(
o) = g(
o, 0). Using Eqs. 23, 24, and
fr(
) = 2
/R2, the
explicit expression for fa(a) in Eq. 25 is
|
(26) |
|
|
x) is a step function that equals 1 for
x
a and 0 otherwise.] Between these limits
fa(a) ~ (a
B)
1.
Equation 26 is one of the key results of this paper. It implies that if all Ca2+ sparks were generated identically and if the observation function were Gaussian (Eq. 20), then the resulting Ca2+ spark amplitude histogram as measured by confocal microscopy should be hyperbolic, not Gaussian. Accordingly, a plot of 1/fa(a) against a yields a straight line.
Relationship between fa and the Ca2+ spark amplitude histogram N(a)
Let N(a) be the number of Ca2+ sparks
having amplitudes between a
/2
a
a +
/2, where
is the binwidth. Then
|
(27) |
|
(28) |
/N(a) (squares), where
N(a) is the data from Fig. 5 A, Ntotal = 176 Ca2+ sparks, and
= 0.1. The solid line is the theoretical fa calculated using Eq. 26 with C(
= 1.4) = 2.86 and R = R90 = 0.80 µm; this line is the best descriptor of
the data points as it is virtually coincident with the best fit line
(dashed line). This agreement between simulation and
theoretical results is important because it provides a check on the
derivation of the relationship between fa and
f
. Thus we can simulate the distribution of
Ca2+ spark amplitudes in a new way. Instead of making
linescan images, detecting the Ca2+ sparks, and then
calculating their amplitudes, we used the following method. The
confocal linescan position was chosen randomly in the y-z
plane and its distance r from the Ca2+ spark at
the origin was calculated. The amplitude was then calculated using the
observation function. With this new method we could simulate conditions
that would be extremely tedious or impossible by the old method.
Estimating
from the Ca2+ spark amplitude
histogram
In this instance
was known so the theoretical line could be
calculated. In practice
is unknown but can be calculated from the
information available in the Ca2+ spark amplitude histogram
as follows. If the plot 1/N(a) against a falls on
a single straight line then the data are consistent with a
delta-function source strength pdf,
(
o). (See
below for fa when f
is
more complicated than a single delta-function.)
o is
calculated using the largest measured Ca2+ spark amplitude
using Eqs. 20 (with r = 0) and 21. In this case amax = 2.85, which gives
o = 2.0, precisely the value used in the simulations. Having calculated
o, R can be calculated for each a
using Eqs. 22 and 20. The calculated values will naturally depend on
the simulation parameters such as the amount of buffer available and
their kinetics of reaction with Ca2+.
fa of more complicated
f
Suppose instead of f
being a single
Dirac
-function, f
is the weighted sum of
-functions
|
(29) |
i gives the probability of the source
strength being
i so the
values satisfy
i
i = 1. Since Eq. 26 holds for all
it
follows that
|
(30) |
|
|
) and
amax(
) are increasing functions of
,
fa is the sum of terms (a
B)