Ion relaxation plays an important role in a wide range of
phenomena involving the transport of charged biomolecules. Ion
relaxation is responsible for reducing sedimentation and diffusion
constants, reducing electrophoretic mobilities, increasing intrinsic
viscosities, and, for biomolecules that lack a permanent electric
dipole moment, provides a mechanism for orienting them in an external
electric field. Recently, a numerical boundary element method was
developed to solve the coupled Navier-Stokes, Poisson, and ion
transport equations for a polyion modeled as a rigid body of arbitrary
size, shape, and charge distribution. This method has subsequently been used to compute the electrophoretic mobilities and intrinsic
viscosities of a number of model proteins and DNA fragments. The
primary purpose of the present work is to examine the effect of ion
relaxation on the ion density and fluid velocity fields around short
DNA fragments (20 and 40 bp). Contour density as well as vector field diagrams of the various scalar and vector fields are presented and
discussed at monovalent salt concentrations of 0.03 and 0.11 M. In
addition, the net charge current fluxes in the vicinity of the DNA
fragments at low and high salt concentrations are briefly examined and discussed.
 |
INTRODUCTION |
The ion atmosphere surrounding a charged
particle, or polyion, in solution can be perturbed from its equilibrium
value in a number of ways. In a sedimentation velocity experiment,
where the driving force is a centrifugal force, the ion atmosphere
tends to lag behind the translating charged particle (Booth, 1954
). This ion relaxation, in turn, retards the motion of the polyion resulting in a larger translational friction coefficient, or smaller sedimentation constant, relative to an equivalent uncharged particle. Similarly, a dilute suspension of charged particles placed in a shear
field exhibits a larger viscosity than the corresponding suspension of
uncharged particles (Booth, 1950b
; Sherwood, 1980
; Sherwood, 1981
;
Allison, 1998
). In this article, steady state shall refer to a polyion
that translates with constant velocity in a viscous fluid due to the
presence of a constant external driving force. In steady-state free
solution electrophoresis, the driving force is a constant external
electric field. In this case, the ion atmosphere is distorted both by
the electric field and by the transport of the charged particle through
the fluid (Booth, 1950a
; Wiersema et al., 1966
; O'Brien and White,
1978
; Stigter, 1978a
,b
). Ion relaxation reduces the electrophoretic mobility, and this relaxation effect becomes more significant as the
charge on the polyion increases. Over the last few years, boundary
element (BE) methods that account for ion relaxation have been
developed, and this has made it possible to estimate the free solution
electrophoretic mobility of rigid model polyions of arbitrary size and
charge distribution (Allison, 1996
). This approach has been applied to
lysozyme (Allison and et al., 1997
) and short DNA fragments (Allison
and Mazur, 1998
). When ion relaxation is included, calculated
mobilities of 20-bp DNA fragments in 0.11 M KCl agree with experimental
mobilities (Laue et al., 1996
) to within a few percent. For a 27-bp
fragment in 0.04 M Tris acetate, calculated and experimental
(Stellwagen et al., 1997
) mobilities agree to within 10-15%.
Furthermore, ion relaxation reduces the mobility at 0.11 M KCl by 25%
relative to the predicted mobility in its absence (Allison and Mazur,
1998
). A final electrooptical phenomenon worthy of mention involves
orienting macroions in solution by external electric fields and
monitoring this by optical birefringence or dichroism. Partial
orientation occurs as a result of the interaction of the external field
with the electric dipole moment of the macroion. Whether the dipole is
permanent or induced can be distinguished by field reversal techniques
(Tinoco and Yamaoka, 1959
) or by the field strength dependence of
birefringence/dichroism (Stellwagen, 1981
). DNA fragments have been
extensively studied by these methods (Elias and Eden, 1981
; Stellwagen,
1981
; Hagerman, 1981
; Diekmann et al., 1982
; Porschke, 1994
), and the
dipole is primarily, though not exclusively, induced (Elias and Eden,
1981
; Stellwagen, 1981
). In this example, the induced dipole moment
develops as a consequence of ion relaxation (Fixman and Jagannathan,
1981
). The above cases serve to illustrate that ion relaxation plays an
important role in biophysics and the transport of macroions in general.
As discussed above, it is now possible to account for ion relaxation in
modeling the transport of rigid macroions of arbitrary shape and charge
distribution, and this approach has been used to calculate
electrophoretic mobilities and intrinsic viscosities for DNA fragments
and lysozyme. The purpose of the present work is to provide visual
demonstration of the relaxation effect by looking at various scalar and
vector fields around short DNA fragments. The present study is similar
to that of Stigter's 1980
paper on spherical colloid particles
(Stigter, 1980
). We shall focus on free solution electrophoresis, but
by doing so, it is also possible to examine the sedimentation problem
as well. Provided the external electric field is sufficiently weak that
the electrophoretic mobility is independent of field strength, one can
view electrophoresis as a superposition of two transport cases
(O'Brien and White, 1978
). In case 1, the polyion translates at
constant velocity through a stationary fluid in the absence of an
electric field. This, however, corresponds to sedimentation velocity.
In case 2, the polyion is held stationary but is subjected to a
constant electric field. Steady-state electrophoresis corresponds to
that superposition of the two cases that results in a net force of zero
acting on the polyion.
 |
METHODS |
The "primitive" model (McQuarrie, 1976
) is employed in which
the solvent is characterized as a continuum with a dielectric constant
of 78 and a viscosity of 0.010 poise corresponding to water at 20°C.
The polyion is modeled as a rigid body the surface of which is
subdivided into a large number of connected triangular plates. In the
present study, models of DNA fragments 20 and 40 bp in length are
considered. They are modeled as capped cylinders of length (in nm) 0.34 Nbp (Nbp = number of base
pairs) and diameter equal to 2 nm. As discussed previously (Allison and
Mazur, 1998
), these models accurately reproduce the translational and
rotational diffusion constants of the actual fragments. The charge
distribution is modeled as a uniform line charge extending from the
origin of one hemispherical cap to the other. The total fragment charge is
2 Nbp q where q is
the protonic charge. The polyion concentration is assumed to be low
enough so that interactions between different polyions can be ignored.
The primitive model is also employed to represent the counterion and
co-ion distributions about a single macroion. In the absence of an
electric or flow field, the equilibrium electrostatic potential of mean
force at position r,
0(r), is related to the equilibrium local charge density
0(r) by the equilibrium Poisson equation:
|
(1)
|
where
is the local dielectric constant. In the fluid domain,
|
(2)
|
where q is the protonic charge,
z
is the valence of ion
,
n
0(r) is the local concentration
of ion
,
= (kBT)
1
(kB is Boltzmann's constant and T is
the absolute temperature), and c
0 is the
concentration of ion
far from the polyion. The physical
significance of
0(r) is that if we placed a
test charge qt at position r, then
the average electrostatic potential energy of our test charge would be
qt
0(r) (McQuarrie, 1976
). A variety of numerical methods are now available to solve the
equilibrium Poisson equation for complex macromolecules (Zhou, 1994
;
Holst and Saied, 1995
).
When a polyion is subjected to a perturbing electric and/or flow field,
the mobile ion distribution is distorted from its equilibrium value,
and this distortion is called ion relaxation. It shall be assumed that
the perturbation is sufficiently weak so that only terms linear in the
perturbing field(s) need to be accounted for. Including ion relaxation
at the level of the primitive model requires solution of the linearized
Navier-Stokes (for solvent flow), Poisson (for charge/ion
distributions), and ion transport equations. It should be emphasized
that these equations are coupled and must be solved simultaneously.
This can be achieved in general by an iterative boundary element
procedure, which is described in detail elsewhere (Allison, 1996
). In
the present work, we shall concentrate on only a few key features of
the method in an attempt to keep the mathematics to a minimum. Let
j
(r) denote the average local
current density (in ions per second per unit area) in some convenient
frame of reference. It can be written as the sum of convective,
diffusive, and "direct force" currents:
|
(3)
|
where n
(r) is the
nonequilibrium ion concentration, v(r) is the
local fluid velocity, D
is the diffusion
constant of an
-ion, and f
(r) is the local force on an
-ion. Consider the simple example of a
polyion in equilibrium so that, by definition, we have no average current or fluid flow. Then Eq. 3 reduces to
|
(4)
|
The equilibrium distribution of ion
about the polyion can be
viewed as a dynamic equilibrium where a direct force current (which,
for example, drives counterions toward the polyion surface) is balanced
by an opposing diffusive current. From Eqs. 2 and 4, it is easy to
deduce that f
0 =
qz

0. At equilibrium, the force on an
-ion is its charge
(qz
) times the local electric field
(

0) produced by the polyion and its ion
atmosphere. When an external electric or flow field is present, however, the situation becomes more complicated. In what follows, we
shall explore how a perturbing electric/flow field alters the ion
distribution, fluid flow, and ion currents around the DNA fragments.
As mentioned previously, we shall restrict ourselves to short fragments
in a water plus simple salt solution at 20°C. The simple salt is
chosen to be KCl and the diffusion constants
(D
in Eq. 3) of K+ and
Cl
are estimated from limiting molar conductivities
(Allison and Mazur, 1998
). The hydrodynamic radii of the K+
and Cl
ions are calculated to be 0.1242 nm.
Scalar and vector fields are first created using software of our own
making. The commercial program Mathematica is then used in preparing
the figures. The first seven figures are two-dimensional contour
diagrams constructed through the command "ListDensityPlot". The
remaining figures represent the merging of three-dimensional solid
objects and the projection of a three-dimensional vector field onto a
plane. The vector fields are constructed through the
"ListPlotVectorField3D" command.
 |
RESULTS AND DISCUSSION |
Perturbed ion densities
When a DNA fragment or some other polyion is placed in a
centrifugal (sedimentation) or electric (electrophoresis) field, the
ion atmosphere is distorted from its equilibrium value. At sufficiently
low centrifugal or (external) electric field strengths, the extent of
distortion, or ion relaxation, varies linearly with the strength of the
imposed field and vanishes at zero field strength. Consider first a
20-bp DNA fragment in 0.03 M KCl at 20°C translating at steady state
in a direction parallel to the cylinder axis at a velocity of 5.00 × 10
4 cm/s in response to a centrifugal force. If the
fragment was in a centrifuge cell located 6.5 cm from the axis of
rotation, a spin rate of approximately 2000 rpm would be necessary to
achieve this velocity. Under these conditions, a plot of the counterion density n+(r) versus position
r would be virtually identical to a plot of the
corresponding equilibrium counterion density
n+0(r). Shown in Fig.
1 is the difference plot in the
counterion density,
n+ = n+
n+0. The DNA is
translating in the direction of the arrow and white/black shades
correspond to maximal excesses/deficiencies of counterions, respectively. A neutral shade of gray corresponds to
n+ = 0 and is also used to represent the DNA
interior for reference. It is seen that the greatest deficiency occurs
at the leading edge of the polyion (toward the bottom of the figure)
and the greatest excess is immediately behind it. At equilibrium, DNA is surrounded by an atmosphere of predominantly positive ions. As it
moves, however, the front end continually enters regions of the fluid
in which the positive ion concentration falls below this equilibrium
value, and this accounts for the deficiency in front of the polyion. A
complementary argument can explain the excess of counterions behind. In
a reference frame that moves with the polyion, a steady-state
distribution of mobile ions prevails, which reflects a balance of ion
flow due to convection, diffusion (against a concentration gradient),
and direct forces as summarized by Eq. 3. We shall return to this point
later. From Fig. 1,
n+ is seen to be dipolar
in character and falls off gradually with distance. The relative
difference in charge density, 
=
0, is
shown in Fig. 2 where
= q(n+
n
).
All quantities have been scaled to give relative extrema that are about
the same in the two figures. Although similar, it is seen that the
charge density approaches the equilibrium value more quickly with
distance from the polyion surface than the counterion density. The
co-ion atmosphere also contributes to ion relaxation, particularly at intermediate distances from the polyion surface. Fig.
3 is similar to Fig. 2 (a contour plot of

) except that the driving force is now an electric field directed
opposite the polyion flow direction and is chosen sufficiently strong
(1 V/cm) so that the polyion translates at the same velocity as in
Figs. 1 and 2. It is clear on comparing Figs. 2 and 3 that the
perturbed steady-state charge densities are very similar regardless of
the nature of the driving force. It should be emphasized that the
driving force in Figs. 2 and 3 are not identical in magnitude, but the
steady-state velocities are.

View larger version (64K):
[in this window]
[in a new window]
|
FIGURE 1
n+ for sedimentation of
20-bp DNA in 0.03 M KCl. The fragment is translating along the cylinder
axis in the direction of the arrow. White shading corresponds to the
greatest positive deviation in counterion concentration from
equilibrium and black to the greatest negative deviation. Neutral gray
(which also shades the fragment interior) corresponds to
n+ = 0. The range in
n+/c+0 is
~±9 × 10 6. Ten different gray shadings appear in
the figure that reflect this range. The bar at lower right denotes
 1 = 1.74 nm.
|
|

View larger version (58K):
[in this window]
[in a new window]
|
FIGURE 2
 for sedimentation of 20-bp DNA in 0.03 M KCl.
The fragment is translating along the cylinder axis in the direction of
the arrow. White shading corresponds to the greatest positive deviation
in charge density from equilibrium and black to the greatest negative
deviation. Neutral gray (which also shades the fragment interior)
corresponds to  = 0. The range in
 /qc+0 is ~±9 × 10 6. The bar at lower right denotes  1 = 1.74 nm.
|
|

View larger version (53K):
[in this window]
[in a new window]
|
FIGURE 3
 for electrophoresis of 20-bp DNA in 0.03 M KCl.
As in Fig. 2, the polyion translates in the direction of the arrow, but
the driving force is a constant electric field (opposite to the flow
direction). The DNA is in steady-state motion. White shading
corresponds to the greatest positive deviation in charge density (from
equilibrium) and black to the greatest negative deviation. Neutral gray
(which also shades the fragment interior) corresponds to  = 0. The range in  /qc+0 is ~±3 × 10 4. The bar at lower right denotes  1 = 1.74 nm.
|
|
The contour plots shown in the first three figures come from BE
calculations of model polyions made up of 144 triangular plates, and
what is being plotted is a two-dimensional cut in a plane that includes
the cylinder axis. As the charge distribution has been approximated by
a line charge, the scalar and vector fields considered in this work
should be axially symmetric to a good approximation provided the
external flow/electric field is collinear with the cylinder axis.
Careful inspection of these density plots show that they are not
entirely symmetrical about the rod axis as they should be. This is due
to noise in the computations. The similarity between ion relaxation in
sedimentation (Fig. 2) and electrophoresis (Fig. 3) is expected, in
part at least, because both scalar fields represent solutions of the
same coupled equations, but subject to different boundary conditions.
Also, Onsager relations have been derived that relate the two processes
of electrophoresis and sedimentation (de Groot et al., 1952
), and these
have been partially confirmed (Stigter, 1980
). However, it is not true
that u/u0 = µ/µ0
where u and u0 are the (magnitudes of
the) polyion sedimentation velocities in the presence and absence of
ion relaxation subject to the same centrifugal force, whereas µ and
µ0 are the electrophoretic mobilities in the presence and
absence of ion relaxation but subject to the same external electric
field strength. From calculations on a 144-plate model (20-bp DNA at
20°C at 0.03 M KCl and the polyion translating end-on as shown in
Figs. 1-3), u/u0 = 0.937 and
µ
/µ
0 = 0.720. These results are
qualitatively consistent with the predictions for sedimentation (Booth,
1954
) and electrophoresis (Booth, 1950a
) of charged spheres. The
orientationally averaged electrophoretic mobility predicted by the
model is
4.05 × 10
4 cm2/V s, and the
experimental value under very similar salt conditions is
3.71 × 10
4 cm2/V s (Wooll, 1994
). Closer agreement
between model calculations and experiment can be achieved by doing a
series of calculations on models with varying number of platelets and
extrapolating to the limit of infinite plate number (Allison and Mazur,
1998
).
It should be emphasized that the magnitude of the ion relaxation seen
in Figs. 1-3 is directly proportional to the magnitude of the polyion
velocity (u), or external electric field (e). In Fig. 1 where u = 5 × 10
4 cm/s, the
range in
n+/c+0
(c+0 equals the counterion concentration far
from the polyion) is ~±9 × 10
6. The range seen
in Fig. 2 is approximately the same as this as the perturbation of
greatest magnitude occurs near the polyion surface where the co-ion
concentration is very small. In Fig. 3, the polyion translates at the
same velocity as in Fig. 2 but is also subjected to an e-field of 1 V/cm in a direction opposing the flow direction of the polyion, and
this corresponds to steady-state electrophoresis. The range in

/qc+0 in this case is ~±3 × 10
4, which, in terms of absolute magnitude, is
substantially larger than the relaxation effect seen in the case of
sedimentation. The fact that Figs. 2 and 3 appear very similar is due
to the much narrower spacings of the contours in Fig. 2 relative to
Fig. 3. Fig. 4 is similar to Fig. 3, but
the KCl concentration has been increased from 0.03 to 0.11 M. The close
similarity between the perturbed charge densities seen in these two
figures shows that the relative ion relaxation effect is similar at the
two salt concentrations for fragments of this length. However, the range in 
/qc+0 at 0.11 M is ~±6 × 10
5, which is 5 times smaller than the range seen at 0.03 M. Thus, although the relative relaxation effect does not vary strongly with salt concentration, the absolute relaxation effect does and decreases with increasing salt concentration. A curious feature of
electrophoresis at these two salt concentrations is that the absolute
range in 
is approximately the same.

View larger version (58K):
[in this window]
[in a new window]
|
FIGURE 4
 for electrophoresis of 20-bp DNA in 0.11 M KCl.
The polyion translates in the direction of the arrow, but the driving
force is a constant electric field (opposite to the flow direction).
The DNA is in steady-state motion. White shading corresponds to the
greatest positive deviation in charge density and black to the greatest
negative deviation. Neutral gray (which also shades the fragment
interior) corresponds to  = 0. The range in
 /qc+0 is ~±6 × 10 5. The bar at lower right denotes  1 = 0.91 nm.
|
|
In addition to contour density plots of DNA translating parallel to the
cylinder axis, similar plots for DNA translating perpendicular to the
cylinder axis exhibit dipolar character similar to what is seen in
Figs. 1-4 when the contour plane bisects the midpoint of the cylinder.
Fig. 5 shows an example of this for the
electrophoresis of 20-bp DNA at 0.11 M KCl. Figs.
6 and 7
show 
for 40-bp DNA at 0.03 and 0.11 M KCl, respectively. The
fragments are undergoing steady-state electrophoresis in an external
field of 1 V/cm directed parallel to the axis of the cylinder. Arrows
at the center of the polyion indicate the direction of polyion flow.
The magnitude of the drift velocities are 5.58 (0.03 M) and 5.09 × 10
4 cm/s (0.11 M), respectively. Somewhat different
contour shadings/spacings are used in the 40-bp plots compared with the
20-bp plots to emphasize the differences at the two salt
concentrations. Similar to the 20-bp plots, Figs. 6 and 7 show that ion
relaxation is largely dipolar in character with a nodal plane (
= 0) that bisects the center of the cylinder and is perpendicular to the
cylinder axis. It is interesting to note from Fig. 6 that ion
relaxation is not confined to the very ends of the fragment, but
extends quite far along the chain from the ends. Keep in mind that the cylinder diameter is 2 nm and the reciprocal of the Debye-Huckel screening length, 
1, equals 1.74 nm and 0.91 nm at
0.03 M and 0.11 M KCl, respectively. Qualitatively, it would appear
that ion relaxation is significant at distances up to several

1 from the ends of the fragment in steady-state
electrophoresis.

View larger version (128K):
[in this window]
[in a new window]
|
FIGURE 5
 for electrophoresis of 20-bp DNA in 0.11 M KCl.
Similar to Fig. 4, but the cylinder model of DNA translates (in the
direction of the arrow) in a direction perpendicular to the long axis
of the polyion. In this contour density plot, one is looking down the
long axis of the DNA at a plane that bisects the center of the
cylinder. The polyion is in steady-state motion. White shading
corresponds to the greatest positive deviation in charge density and
black to the greatest negative deviation. Neutral gray (which also
shades the fragment interior) corresponds to  = 0. The bar at
lower right denotes  1 = 0.91 nm.
|
|

View larger version (58K):
[in this window]
[in a new window]
|
FIGURE 6
 for electrophoresis of 40-bp DNA in 0.03 M KCl.
The polyion translates in the direction of the arrow along the long
axis of the cylinder, but the driving force is a constant electric
field (opposite to the flow direction). The DNA is in steady-state
motion. White shading corresponds to the greatest positive deviation in
charge density and black to the greatest negative deviation. Neutral
gray (which also shades the fragment interior) corresponds to  = 0. The bar at lower right denotes  1 = 1.74 nm.
|
|

View larger version (57K):
[in this window]
[in a new window]
|
FIGURE 7
 for electrophoresis of 40-bp DNA in 0.11 M KCl.
The polyion translates in the direction of the arrow, but the driving
force is a constant electric field (opposite to the flow direction).
The DNA is in steady-state motion. White shading corresponds to the
greatest positive deviation in charge density and black to the greatest
negative deviation. Neutral gray (which also shades the fragment
interior) corresponds to  = 0. The bar at lower right denotes
 1 = 0.91 nm.
|
|
At the same ambient salt concentration, the absolute range in 
for the electrophoresis of DNA translating parallel to the cylinder
axis for the 40-bp fragment is approximately the same as the 20-bp
fragment. This does not necessarily imply that the induced electric
dipole moment for a rod translating parallel to its axis varies
linearly with length as ion relaxation effects are evidently diffuse in
character and not confined to the very ends of the rod. At the present
time, we are investigating the induced electric dipole moment of DNA
fragments and the related problem of their electric polarizability
(Fixman and Jagannathan, 1981
). The extended nature of the perturbed
ion densities seen in these figures does raise concern about how long
range end effects are in transport properties such as the
electrophoretic mobility of short DNA fragments. For a rod translating
parallel to its axis in an external field, it has been argued that end
effects on the electrophoretic mobility should become negligible in the limit of a very long rod (Stigter, 1978b
). For DNA modeled as a rod in
0.11 M KCl and translating parallel to the rod axis, BE calculations
have shown that ion relaxation reduces electrophoretic mobility by
approximately 32%, 25%, and 20% for 20-bp, 40-bp, and 60-bp
fragments, respectively (Allison and Mazur, 1998
). Thus, end effects
may very well become negligible for very long rods as proposed by
Stigter, but their contribution to parallel electrophoresis appears to
be quite long range.
Velocity fields
In addition to the perturbed ion and charge density plots
considered previously, it is worthwhile to also consider how ion relaxation influences various vector fields such as the actual fluid
velocity v(r) in the vicinity of a sedimenting or
electrophoresing particle. First of all, consider the velocity field
v0(r) of a sedimenting 20-bp DNA
fragment in the absence of ion relaxation translating with velocity
u0 in a direction parallel to the cylinder axis.
The velocity field is shown in Fig. 8 for
a plane of fluid that passes through the cylinder axis. The reference
frame is chosen to be stationary relative to the laboratory, and the
instantaneous velocity is downward in the figure. The actual value of
u0 is unimportant provided its magnitude is
sufficiently small to insure low Reynolds number flow conditions
(Happel and Brenner, 1963
). A value of 5 × 10
4 cm/s
is used here as a reference value. We shall also assume stick boundary
conditions hold, which means v0(r) = u0 for points r on the polyion
surface. Also, the upper and lower portions of the capped cylinder
models have been cut away in Fig. 8 to illustrate more clearly the flow
pattern near the particle. These fields are generated numerically by a BE procedure described elsewhere (Allison, 1996
). As can be seen, the
velocity field falls off rather slowly with distance away from the
particle, and the pattern is similar to the velocity field profile for
a sphere (Stigter, 1980
). For a sedimenting DNA with ion relaxation,
the flow pattern is similar to that shown in Fig. 8. If we compare two
DNA fragments subject to exactly the same centrifugal force but one
with ion relaxation accounted for and the other with ion relaxation
ignored, the sedimentation velocity of the fragment with ion relaxation
included would be smaller. In our reference example, ion relaxation
would reduce the sedimentation velocity from 5.00 × 10
4 to 4.68 × 10
4 cm/s in 0.03 M KCl
and from 5.00 × 10
4 to 4.86 × 10
4 in 0.11 M KCl. This is due to the fact that the
sedimenting fragment in the absence of ion relaxation is entirely
equivalent to the sedimentation of an uncharged particle of the same
size and shape. The charged particle with ion relaxation sediments more
slowly as it experiences additional frictional drag brought about by the tendency of the atmosphere to lag behind the polyion itself. In
what follows, the difference velocity field, or relaxation field, shall
be plotted:
|
(5)
|
where the 0 subscript refers to the analogous unrelaxed case. The
difference fields for sedimentation are shown in Figs. 9 (0.03 M) and 10 (0.11 M KCl). At the
particle surface, the difference fields point in a direction opposite
to that shown in Fig. 8, which is due to the fact u is in
the same direction as u0 but smaller in
magnitude. For our reference example, the magnitude of
v
is 0.32 × 10
4 and 0.14 × 10
4
cm/s at 0.03 and 0.11 M KCl, respectively. Thus, the absolute magnitude
of the vectors at the polyion surface are 6.4% (Fig. 9) and 2.8%
(Fig. 10) as large as those of the
surface vectors in Fig. 8. Comparing the relaxation field plots with
the Stokes field plot of Fig. 8, it is clear that the relaxation fields
fall off more quickly with distance and that they also exhibit a
substantial salt dependence. At high salt, the relaxation field falls
off more rapidly with distance than at low salt. Figs.
11 and
12 show the corresponding relaxation
fields for steady-state electrophoresis at 0.03 and 0.11 M KCl,
respectively. The DNA fragment is placed in an electric field of 1 V/cm
(directed in the direction of
v at the polyion surface).
In the absence of relaxation, the DNA translates along the long axis
under steady-state conditions with a velocity of 5 × 10
4 cm/s. Ion relaxation reduces these velocities to
3.60 × 10
4 and 4.10 × 10
4 cm/s
at 0.03 and 0.11 M KCl, respectively. Thus, the magnitude of the
relaxation fields at the polyion surface in Figs. 11 and 12 are 28%
and 18% as large as the magnitude of the Stokes velocity field at the
polyion surface (Fig. 8). Although similar to the previous figures for
sedimentation, the corresponding relaxation fields for electrophoresis
fall off even more quickly with distance. Fig.
13 is similar to Fig. 12 but represents
a blow-up of the relaxation velocity field in the vicinity of one end
of the polyion. Also note in Fig. 13 that there is a small reversal of
the relaxation field at some distance from the polyion surface. Keep in
mind, however, that these relaxation velocities are not actual
velocities but difference velocities according to Eq. 5. Consequently,
the actual velocity field does not reverse itself. At 0.11 M, the fall-off in the relaxation field appears sufficiently rapid that thin
double layer theories (O'Brien, 1983
; Solomentsev et al., 1993
) may
actually be appropriate.

View larger version (29K):
[in this window]
[in a new window]
|
FIGURE 8
Stokes velocity field for 20-bp DNA. Fluid velocity
field of an uncharged, or unrelaxed model translating parallel to the
cylinder axis. Portions of the rigid model have been cut away to
display the flow pattern more clearly. The structure translates at
constant velocity in a fluid that is otherwise at rest.
|
|

View larger version (23K):
[in this window]
[in a new window]
|
FIGURE 9
Relaxation field for sedimentation of 20-bp DNA in
0.03 M KCl. The relaxation field is defined v = v v0 where v is
the velocity field in the presence of ion relaxation and
v0 is the velocity field in the absence of ion
relaxation. In both cases, the centrifugal force on the particles is
the same and both v and v0 are
directed toward the bottom of the figure. The relative magnitude of the
vectors at the polyion surface are the same as in Fig. 8; however, the
absolute value of the surface vectors in this figure are 6.4% as large
as those in Fig. 8.
|
|

View larger version (17K):
[in this window]
[in a new window]
|
FIGURE 10
Relaxation field for sedimentation of 20-bp DNA in
0.11 M KCl. Similar to Fig. 9, but at the higher salt concentration.
The relative magnitude of the vectors at the polyion surface are the
same as in Fig. 8; however, the absolute value of the surface vectors
in this figure are 2.8% as large as those in Fig. 8.
|
|

View larger version (20K):
[in this window]
[in a new window]
|
FIGURE 11
Relaxation field for electrophoresis of 20-bp DNA in
0.03 M KCl. See the caption of Fig. 9 or the text for the definition of
the relaxation field. The electric field is the same in both the
relaxed and unrelaxed cases and is directed upward, in the direction of
v at the model surface. The relative magnitude of the
vectors at the polyion surface are the same as in Fig. 8; however, the
absolute value of the surface vectors in this figure are 28% as large
as those in Fig. 8.
|
|

View larger version (14K):
[in this window]
[in a new window]
|
FIGURE 12
Relaxation field for electrophoresis of 20-bp DNA in
0.11 M KCl. Similar to Fig. 11, but at the higher salt concentration.
The relative magnitude of the vectors at the polyion surface are the
same as in Fig. 8; however, the absolute value of the surface vectors
in this figure are 18% as large as those in Fig. 8.
|
|

View larger version (62K):
[in this window]
[in a new window]
|
FIGURE 13
Close-up view of the relaxation field for
electrophoresis of 20-bp DNA in 0.11 M KCl. Similar to Fig. 12, but the
relaxation field near the tip of the polyion has been magnified. The
relative magnitude of the vectors at the polyion surface are the same
as in Fig. 8; however, the absolute value of the surface vectors in
this figure are 18% as large as those in Fig. 8.
|
|
Current fluxes
Finally, we would like to examine current fluxes around 20-bp DNA.
Shown in Fig. 14 is the difference
current, jd = j+
j
at 0.002 M KCl in the case of steady-state
electrophoresis. The frame of reference is chosen to be stationary
relative to the polyion. In the lab frame, the DNA moves downward in
the figure and the electric field points upward. Near the DNA, the
difference current is dominated by the flux of counterions
j+, and it is worth considering which term(s) on
the right side of Eq. 3 dominate the transport process. It cannot be
the convective term as in a frame of reference stationary with respect
to the polyion, n
v would vanish at
the polyion surface and reach some plateau value at some distance. This, however, is inconsistent with Fig. 14. The direct force term (third term on the right side of Eq. 3) is expected to be dominated by
the + ions interacting with the e field and be greatest near
the polyion surface. This, however, would yield a flow pattern directed
upward (along e) in Fig. 14. Close to the sides of the DNA,
the flow pattern is seen to be just the opposite of this. This leaves
the second term, which corresponds to diffusion against a concentration
gradient. From Fig. 3, we know that there is an excess of counterions
at the top end of the fragment and a deficiency at the other end. Also,
the counterion concentration will be greatest near the polyion surface.
Thus, the flow pattern near the sides of the DNA in Fig. 14 is entirely
consistent with what one would expect if Eq. 3 were dominated by the
diffusion term. Shown in Fig. 15 is
exactly the same vector field but at the higher salt concentration of
0.11 M. It is worth noting that the sign of jd
close to the sides of DNA is reversed relative to Fig. 14 despite the
fact the e values are the same in the two cases and the
u values are at least in the same direction. Based on our
previous discussion, it is clear that the direct force term in Eq. 3
dominates the ion flux at 0.11 M salt whereas the diffusive term
dominates at low salt. Similar studies over a range of salt
concentrations reveal that the general flow pattern seen in Fig. 15
occurs except at very low salt (less than ~0.005 M monovalent salt).

View larger version (45K):
[in this window]
[in a new window]
|
FIGURE 14
jd for the electrophoresis of
20-bp DNA in 0.002 M KCl. The electric field is directed upward in a
direction opposing the direction of jd seen in
the figure close to the sides of the polyion. The DNA translates
downward. The reference frame is stationary with respect to the DNA.
|
|

View larger version (46K):
[in this window]
[in a new window]
|
FIGURE 15
jd for the electrophoresis of
20-bp DNA in 0.11 M KCl. Similar to Fig. 14, but at a higher salt
concentration.
|
|
We can offer the following simple physical interpretation of this
phenomenon. In general, a DNA fragment undergoing steady-state electrophoresis and moving parallel to the long axis of the molecule will have one end that is rich in counterions and one end that is poor
in counterions. At low salt, the primary source of counterions for the
poor end is diffusion, against a concentration gradient, of counterions
from the rich end (and sides) to the poor end. At high salt, the
primary source of counterions for the poor end is the bulk solution
where they are now present in abundance. As the + ions in the bulk
solution tend to move in the general direction of e, this is
reflected in the flow field for jd.
 |
CONCLUSIONS |
The primary objective of this work has been the graphic
illustration of the phenomenon of ion relaxation in the transport of
short DNA fragments. Contour plots of counterion and/or charge density
have been presented for both 20- and 40-bp fragments at 0.03 and 0.11 M
monovalent salt (KCl). Ion relaxation extends out distances of order

1 from the ends of the fragment for the DNA
translating parallel to its cylinder axis. The relaxation effect is
similar for sedimentation and electrophoresis, but there are
differences. In addition, vector plots of the Stokes field and the
relaxation velocity field are presented for both sedimentation and
electrophoresis of 20-bp DNA. The relaxation effect falls off rapidly
with distance, and this fall-off is more rapid at higher salt. These
results are qualitatively similar to an earlier study of charged
spheres (Stigter, 1980
). Finally, the charge current flux around the
20-bp fragment is examined at 0.002 and 0.11 M KCl. At the lower salt
concentration, the charge flux is dominated by diffusion against a
concentration gradient, but at the higher salt concentration, the
dominant term involves the force on the counterions by the external
electric field. The charge current flux plays a central role in the
theory of conductance of polyelectrolytes (Stigter, 1979
), and the
results presented here will be expanded upon in future work.
This work was supported in part by National Science Foundation
grants MCB-9807541 (to S.A. Allison) and MCB-9807550 (to T.M. Laue).
Address reprint requests to Dr. Stuart Allison, Department of
Chemistry, Georgia State University, University Plaza, Atlanta, GA
30303. Tel.: 404-651-1986; Fax: 404-651-3099; E-mail:
chesaa{at}panther.gsu.edu.