| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, October 1999, p. 2226-2236, Vol. 77, No. 4
*Department for Cell Physiology, Max-Planck Institute for Medical Research, Heidelberg, Germany and #Botanical Institute, Ludwig-Maximilians University, München, Germany, and §High Resolution Optical Microscopy Group, Max-Planck Institute for Biophysical Chemistry, Göttingen, Germany
| |
ABSTRACT |
|---|
|
|
|---|
The signal and limitations of calcium florescence imaging using nonresonant multiphoton absorption of near-infrared femto- and picosecond laser pulses were examined. The fluorescence changes of various Ca2+-indicators induced by transient increases of the intradendritic calcium concentration were evaluated by evoking physiological activity in neocortical neurons in rat brain slices. Photodamage was noticeable as irreversible changes in the parameters describing the calcium fluorescence transients. At higher two-photon excitation rates, a great variety of irregular functional and structural alterations occurred. Thus, signal and observation time were limited by phototoxic effects. At lower excitation rates, photodamage accumulated linearly with exposure time. Femtosecond and picosecond laser pulses were directly compared with respect to this cumulative photodamage. The variation of the pulse length at a constant two-photon excitation rate indicated that a two-photon excitation mechanism is mainly responsible for the cumulative photodamage within the investigated window of 75 fs to 3.2 ps. As a direct consequence, at low excitation rates, the same image quality is achieved irrespective of whether two-photon Ca2+-imaging is carried out with femto- or picosecond laser pulses.
| |
INTRODUCTION |
|---|
|
|
|---|
When imaging in scattering tissue, nonresonant
two-photon excitation (TPE) fluorescence microscopy reportedly is
superior to its single-photon excitation (SPE) counterpart (Denk et
al., 1990
, 1996
). The quadratic dependence of the excitation on the illumination intensity confines the excitation to the focal region so
that the mere generation of fluorescence in a raster fashion establishes a three-dimensional image. Importantly, ballistic and
scattered emitted photons contribute equally well to the total signal.
This allows the collection of a large fraction of fluorescence photons
by external detectors resulting in a much higher collection efficiency.
Moreover, TPE of visible dyes requires excitation wavelengths in the
near-infrared where scattering coefficients are lower (Svaasand and
Ellingsen, 1983
) and, with rare exceptions, linear absorption is
negligible. Weak endogenous absorption and spatial confinement of TPE
inflicts considerably lower phototoxic stress than SPE, which is
invaluable for live-cell imaging (Wokosin et al., 1996
; Maiti et al.,
1997
). However, the nonlinear nature of TPE requires much higher
intensities. The high intensities may well induce photodamage and
irreversible sample modifications.
To provide an estimate for the intensities and power levels involved,
let us first review briefly the power and intensity levels in SPE and
TPE microscopy, assuming typical wavelengths of
exc = 450 nm and
exc = 870 nm,
respectively, and a numerical aperture NA = 1.0. The SPE rate
is given by k1 =
1I/(hc/
)
whereby h is Planck's constant, and c the speed
of light (Lakowicz, 1983
). For a typical cross section of
1 = 10
16 cm2, an
intensity I = 200 kW/cm2 is sufficient to
reach k1 = 5 × 107
s
1, which is close to saturation by ground state
depletion. This intensity corresponds well to a continuous wave average
power of ~140 µW at the focus. To obtain the same rate,
k2 =
2[I/(hc/
)]2, with nonresonant TPE,
for a typical TPE cross section of
2 = 10
49 cm4s (Xu and Webb, 1996
), a much higher
intensity of 5.4 GW/cm2 is required. Given the same lenses
and the wavelength needed for TPE is
exc = 870 nm,
this intensity corresponds to an average power of 12 W at the focus.
Although efficient continuous wave TPE microscopy is feasible for
highly concentrated staining and large TPE cross sections (Booth and
Hell, 1998
; Hell et al., 1998
), TPE microscopy usually requires pulsed
lasers such as the mode-locked Ti:Sapphire, providing pulses of
100 fs or
1 ps pulse duration, at a pulse
repetition rate f
80 MHz. For pulsed illumination,
the n-photon absorption rate is given by
|
(1) |
1/f, the evaluation of the
integral in the paraxial focusing approximation gives
|
(2) |
|
|
and
denote
the time-averaged intensity and time-averaged power, respectively. The
constant g(n) is the time-zero
n-order temporal coherence. For hyperbolic-secant-squared shaped pulses, g(2) = 0.59. When
considering photodamage issues, it is important to bear in mind that
the time-averaged values are not determining the absorption rate alone.
They are normally used because they can be easily measured by a (slow)
detector. The relationship between the peak intensity
Ipeak and the time-average intensity
is Ipeak = (
f)
1
. The
proportionality of kn to
(Ipeak)n
1 reflects the fact that
(only) nonlinear (n > 1) processes benefit from high
peak intensities.
Compared to the CW TPE case, pulsing reduces the required
time-averaged power by a factor of (
f)
1/2
but, at the same time, increases the peak intensity by the same factor.
By rearranging Eq. 2, one can calculate that, for the focusing
conditions and assumed TPE rate k2 = 5 × 107 s
1, pulsed-mode TPE imaging with
pulses of
= 1 ps, and 100 fs would keep the time-averaged
power
down to 140 mW and 45 mW, respectively, at the
expense of involving peak intensities of Ipeak = 800 GW/cm2 in the 1 ps,
and even higher intensities of 2.5 TW/cm2, in the 100 fs mode.
Standard Ti:Sapphire systems emit between 0.5 and 2 W of time-averaged power. Thus, the laser usually provides ample power to saturate nearly all dyes by TPE. In fact, the available power would allow an increase in the focal intensity by another order of magnitude. Evidently, dyes with 100 times lower cross sections could still be excited with saturating intensities. However, Eq. 2 indicates that the high peak intensities of >1 TW/cm2 foster nonlinear (n > 1) processes, some of which might be detrimental to the fluorophore or the sample. Such intensities can overwhelm the light specimen interaction before saturation by ground-state-depletion is reached.
Such evidence for irreversible fluorophore damage induced by TPE has
indeed been found in single-molecule experiments (Eggeling et al.,
1998
). Photodamage in TPE microscopy has also been reported by several
researchers in the field, often as a sudden localized onset of
increased luminescence (König et al., 1999
; Straub and Hell,
1998
). It is reasonable to expect the limiting mechanisms to vary with
the sample and fluorophore. In fact, such limits may well be a
combination of linear and nonlinear processes, with each of them
prevailing at a given power and intensity range. It will be important
to establish the dependence of the damage on the excitation intensity
as a function of the power order n of the (non-)linearity.
According to Eq. 2, photodamage by linear absorption (n = 1) depends solely on the average intensity
and
not on the peak intensity Ipeak. If linear
photodamage dominated, safe TPE would be possible only for high
transient intensities such as can be obtained by minimizing the pulse
width. If cubic or higher order (n
3) processes
would dominate photodamage, increasing Ipeak by
reducing the pulse width would be disadvantageous. In the special case
of a quadratic dependence, the photodamage and TPE rate will just be
proportional to each other, and the attainable TPE rate would not
depend on the pulse width.
As in the earlier stages of TPE microscopy, linear absorption was
considered to be the limiting factor, so, keeping the pulses as short
as possible was considered of particular importance (Denk et al.,
1995
). Prechirping units were applied to compensate for group velocity
dispersion in the microscope (Soeller and Canell, 1996
; Brakenhoff et
al., 1995
), so that the pulse could reach the specimen at its shortest
possible duration (~100 fs). However, recent two-photon imaging with
increased
through picosecond (7 ps) pulsed
(Bewersdorf and Hell, 1998
) and CW-excitation (
= 200 mW) (Hell et al., 1998
) revealed nonlinear optical effects as
the major limitations and indicated that the reduction of the pulse
duration is of secondary importance.
In a more recent study, a large population of live unstained Chinese
hamster ovarian cells were exposed to Ti:Sapphire laser illumination
and the impairment of cell replication quantified as a measure for
photodamage (König et al., 1999
). For pulses ranging from 2.2 ps
down to 240 fs, the photodamage followed explicitly a quadratic power
law; this rendered the pulse width irrelevant for optimizing for least
photophysical damage. Whereas, for the 240-fs pulses, an average power
of
= 7.3 mW was sufficient to impair cell
viability, for the 2.2-ps pulses, an average power of
= 20 mW was required to produce the same
effect. Taking into account the NA = 1.35, the corresponding focal
intensities Ipeak were 0.35 TW/cm2
for the femtosecond and 0.1 TW/cm2 for the picosecond
pulses. Although these results are very valuable, the use of chirped
laser pulses and the complexity of the damage mechanisms does not allow
a general conclusion for other applications.
Due to its high penetration depth, TPE has been particularly successful
in the imaging of Ca2+ dynamics in small neuronal
compartments of live brain tissue (Svoboda et al., 1996
; Yuste and
Denk, 1995
; Schiller et al., 1998
; Koester and Sakmann, 1998
; Denk et
al., 1995
; Yuste et al., 1999
). Although the relevance of
characterizing phototoxic effects has been recognized, information
about the relationship between excitation power and intensity and
photodamage in TPE microscopy is scarce. Finding out the optimal
operational parameters for multiphoton imaging is of paramount
importance to the proper use of this powerful imaging mode. Therefore,
in our study, we investigated the limiting factors of
TPE-Ca2+ imaging in rat brain slices using various
fluorescent Ca2+ indicators and pulse widths. In
particular, we provide a quantitative description of the observed
photodamage and investigate the order of nonlinearity of the damage.
| |
MATERIAL AND METHODS |
|---|
|
|
|---|
Slice preparation
Acute neocortical slices were prepared from 13 to 14-day-old
Wistar rats as described previously (Markram et al., 1997
).
Layer 5 pyramidal neurons in the somatosensory cortical area were
identified using gradient-contrast (Dodt et al., 1998
) infrared
videomicroscopy. The bath solution contained 125 mM NaCl, 25 mM
NaHCO3, 2.5 mM KCl, 1.25 mM
NaH2PO4, 1 mM MgCl2, 25 mM Glucose,
and 2 mM CaCl2 (Biometra, Goettingen, Germany). The bath
temperature was 32-34°C. All experiments were performed on basal
dendrites 30-80 µm away from the soma. To keep the attenuation of
excitation laser power comparable, only structures 30-70 µm below
the slice surface were selected.
Electrophysiology
Cells were filled with dyes by whole-cell recording. Patch
pipettes were filled with 115 µM K-gluconate, 20 µM KCl, 10 µM HEPES, 4 µM ATP-Mg, 10 µM phosphocreatine, 0.3 µM GTP, and a dye, which was either Calcium Green-1 (CG-1, 100 µM), Oregon Green 488 BAPTA-1 (OGB-1, 100, 200, or 500 µM), Oregon Green 488 BAPTA-2 (OGB-2, 500 µM), Magnesium Green (MG, 500 µM), or Fluorescein (500 µM). All dyes were purchased from Molecular Probes (Eugene, OR). Dyes
were allowed to equilibrate for
20 min after establishing whole-cell
recording before experiments started. Electrical recordings were made
with a patch-clamp amplifier (EPC-7, List-Electronics, Darmstadt,
Germany) operated in the current clamp mode. Initial access resistances
were <6 M
. Action potentials (AP) were evoked by brief current
injection (1 nA for 4-7 ms) into the soma.
Two-photon laser scanning microscopy
A modified galvanometer scanning unit (TCS 4D, Leica Microsystems, Heidelberg, Germany) was adapted to an upright microscope (BX50WI, Olympus Optical Co., Tokyo, Japan) equipped with a 60× objective (LUMPlanFL 60xW0.9IR, Olympus). For excitation femtosecond laser pulses and picosecond pulses that were nearly transform-limited in the focal plane were used. In a few cases, we also used femtosecond pulses chirped to picoseconds. The femtosecond laser pulses originated from a Ti:Sa-Laser (MIRA 900F, Coherent, Santa Clara, CA) pumped by a large-frame argon ion laser (Sabre ML14, Coherent). The nearly transform-limited picosecond pulses originated from a picosecond Ti:Sa-Laser (Tsunami, Spectra-Physics, Mountain View, CA). Both lasers were operated at 870 ± 5 nm. Insertion of a mirror allowed a convenient change between the lasers during the experiment. When imaging in the line scan mode, signals were collected during both scan directions. External detectors were placed behind the objective and the condensor for signal collection. Transmission- and epifluorescence signals were recorded by photomultiplier tubes (PMT) (R6357, Hamamatsu Photonics, Herrsching, Germany) and digitized into 8-bit values. The signals were averaged off-line. In control experiments, a bandpass filter (525 ± 25 nm, Leica Microsystems) was inserted into the epifluorescence pathway.
Analysis of calcium fluorescence transients
Fluorescence line scan images were analyzed using dedicated
software. A line was scanned every 2.27 ms. Pixels of two lines between
two positions enclosing the dendrite examined were averaged to obtain
one time point. This resulted in a temporal resolution of 4.54 ms.
Stimulation protocols began 150 ms after the start of the line scan
(512 lines). Before stimulation, fluorescence was averaged for 100 ms
to obtain the basal fluorescence, F0. A region
distant from any indicator-containing structure was chosen for
determination of the background fluorescence,
FB, which was subtracted. Fluorescence traces
F(t) of a Ca2+ transient evoked by an AP were
fitted with a single exponential using a least-square fit routine
(IGOR, Wave- Metrics, Lake Oswego, OR). This fit rendered the amplitude
of the fluorescence increase AF and decay time
constant
. Relative fluorescence changes were calculated as
F(t)/F0 = (F(t)
F0)/(F0
FB).
AR was the amplitude of the relative fluorescence transient.
Pulse compression and diagnostics
Pulse width control
An external prism compressor of two SF10 glass prisms placed 82 cm apart was used for correcting first-order group velocity dispersion (GVD) of the femtosecond pulses (Kafka and Bear, 1987
|
Interferometric autocorrelation
To determine the pulse width of the femtosecond pulses in the focal plane, the laser beam was split and directed into the two arms of a Michelson interferometer. The combined beam from the interferometer was fed into the multiphoton laser scanning microscope (Soeller and Canell, 1996Background-free autocorrelation
Since the GVD induced by the microscope is not expected to change the pulse width of the chirped pulses or the picosecond pulses to a significant extent, we used an external, background-free autocorrelator (Model 409, Spectra-Physics) to obtain a better signal-to-noise ratio. Autocorrelation traces were monitored using a digital oscilloscope (DL 708, Yokogawa, Tokyo, Japan) that allowed convenient data storage. A Gaussian fit using least-square fit routines (IGOR, WaveMetrics) was performed and the FWHM was determined. We used the correction factor of 0.707 for Gaussian pulses for calculating the actual pulse widths from autocorrelation traces.Laser power determination
Average laser power was determined by a photodiode placed behind the objective. Diode measurements were calibrated with a power meter (Lasermate, Coherent) placed behind an aperture stop with the diameter and position of the pupil plane of the objective used. The transmission of the objective at 870 nm was taken into account.Determination of the photodamage rate r at different pulse lengths
Illumination protocol
A line 5 µm in length was scanned repetitively every 2.23 ms for 400 s orthogonal to a basal dendrite. Taking the duty cycle of the scan (0.5) and the size of a basal dendrite into account (about 1 µm), cell structures were exposed for about 100 µs every 1.115 ms. All data presented are given in terms of average laser power measured behind the objective. The incident intensities at the focus were attenuated by the scattering tissue. For estimating the TPE rates, the intensities have to be scaled by I(d) = I0e
d, where d is the
tissue depth. The parameter
was, to our knowledge, not yet
determined for brain slices.
Determination of r
We monitored the basal fluorescence F0 from 20 line scans every 20 s during 400 s of constant light exposure. To correct for the differences in optical path length and excitation rates, we calculated the relative changes in basal fluorescence
F0(t)/F0(t = 0) = F0(t)
F0(t = 0)/F0(t = 0) and fitted this with a line to
obtain the rate of relative change in basal fluorescence,
|
) = A ·
x, where A and the
exponent x were free parameters. When using a high-affinity
dye (CG-1 or OGB-1), the changes in amplitude and decay time constant
were also monitored.
Comparing picosecond and femtosecond laser pulses in single cells
Experiments were performed on basal dendrites of labeled layer 5 cells. Dendrites were selected to be nearly at the same tissue depth (<10 µm variation). The rate of relative increase in basal fluorescence, r, was determined at different time-averaged laser powers,
. At a given pulse length, normally
3-6 values of
were used in random order. For each
trial of 400 s exposure, a location on the dendrite was chosen proximal
to the trials executed before. This was done to avoid possible changes
in AP-propagation and dye diffusion through damaged structures. The
number of trials conducted in a single cell was limited by the lifetime
of a dye-filled cell (60-90 min at the bath temperature of 35°C).
Only two different pulse lengths were tested per cell, and the first
pulse length was tested only at three different laser powers because of
the limited lifetime. A cell was classified as stable when no
significant changes in resting membrane potential, excitability, or
calcium fluorescence dynamics at undamaged locations occurred. Trials were taken into account only as long as the cell was stable. Pulse length was varied across trials in a random order. We mainly used 78 ± 4 fs pulses and 1.4-3.2 ps pulses, both nearly
transform-limited in the focal plane. In some experiments, we also used
chirped femtosecond pulses that had passed through 36-cm SF6, resulting in pulse lengths 1.4-2.8 ps. Pulse length of the picosecond laser systems was determined before and after each series of exposures (at
3-5 different laser powers) within 20 min. Femtosecond laser pulses
were controlled only in the initial experiments before and after each
cell because of the apparent stability in pulse length.
We assumed
|
(3) |
|
|
|
(4) |
Detection scheme
We used detectors placed at three different positions. One detector was the confocal, descanned detector with the pinhole removed. A second detector for capturing nondescanned epifluorescence was placed behind the water immersion objective (LumPlanFl 63W0.9IR, NA 0.9, Olympus). A third, external detector was placed behind the high NA = 1.4 oil condensor (see Fig. 2). We used the same type of PMT (R6357, Hamamatsu), supplied with the same voltage for all three detector positions. Because the sensitive area of the side-window PMTs is only 8 × 24 mm, they required additional optics to achieve the scan-fields desired. For demagnification, we placed lenses with focal length of f = 12 mm (for the transmission detector) and f = 20 mm (epifluorescence external detector) in the corresponding planes. The virtual scanfield of the condensor PMT depends on the aperture (Fig. 2). In the system used, the virtual scanfield was about 1.2 mm. External detection through the NA = 1.4 oil condensor and the NA = 0.9 water immersion objective and descanned detection were compared using a fluorescence solution layer (Fluorescein). The number of photons and gain was calculated from the histograms to exclude variation of gain and dark current between the PMTs. The signal collected by the descanned detector was about 70% of the signal collected by the epifluorescence nondescanned detector. With the same fluorescence solution, the signal collected by the transmission detector was 215 ± 65% (mean ± SD, n = 18) of the descanned epifluorescence signal. Because the situation might be different in highly scattering tissue, we also tested dye-filled (OGB-1, 200 µM) neurons in brain slices. Here, the number of photons collected by the transmission external detection compared to the descanned epifluorescence detection ranged up to 450% (n = 2).
|
| |
RESULTS |
|---|
|
|
|---|
Calcium dynamics of single back-propagating AP and cumulative photodamage
One major use of two-photon laser scanning microscopy is calcium
imaging in small compartments of cells in highly scattering tissue.
Therefore, we investigated the fluorescence dynamics in layer 5 neocortical neurons in rat brain slices filled with a fluorescent
calcium indicator dye. Upon evoking an AP, the physiologically relevant
transient increase of intradendritic free calcium was observed as a
transient increase in fluorescence of the calcium indicator (Fig.
3 A). A single exponential
function satisfactorily fitted the decay. The decay times varied
between 50 and 400 ms, depending on the buffering by the dye (Helmchen
et al., 1996
). The calcium fluorescence dynamics can be described by
the following set of parameters: the basal fluorescence
F0, the amplitude of the fluorescence change
AF, and the decay time constant
. Changes in
these parameters during prolonged laser light exposure were elected as
measures of photodamage.
|
The relative fluorescence change
F/F(t) was calculated
and the amplitude AR was monitored during up to
400 s of laser light exposure. Whereas the basal fluorescence
F0 and decay time
increased with time, the
amplitude of the relative fluorescence change AR decreased (see Fig. 3, B-E). At low average laser powers,
that is <3-7 mW (depending on the depth of the cell within the slice) for femtosecond laser pulses and <8-24 mW for picosecond laser pulses, all changes were observed to be linear with exposure time. The
changes were restricted to the illuminated area (see Fig. 4 A); they were cumulative
and irreversible. This type of changes is further referred to as
cumulative photodamage.
|
To rule out loading effects, the average laser power and pulse length
were randomly varied. Also, the results did not correlate with loading
time or distance to the soma. Most experiments were performed on
dendrites close to the soma, where the dye concentration reached
equilibrium within a few minutes (Helmchen et al., 1996
). This is well
below the time we waited after establishing the whole-cell recording
before starting the experiments (>20 min).
The increase of the basal fluorescence was also observable without back-propagating APs (OGB-2, n = 2 and MG, n = 18 with n representing the number of trials) and when a dye was used that was not sensitive to calcium (Fluorescein, n = 8). The decrease of AR could be explained by the increased basal fluorescence, because the amplitude of the absolute fluorescence change of the calcium fluorescence transient, AF, remained constant (Fig. 3 D). The change of the basal fluorescence was observable with all dyes used. Qualitatively, these changes did also occur when intermitted laser light exposure instead of continuous exposure was used (e.g., 1 s exposure every 10-20 s). These characteristics allowed us to use the rates of change as quantitative measures of photodamage. The changes in basal fluorescence exhibited the highest S/N-ratio; therefore, the rate of relative change in basal fluorescence, r (see Methods), was taken as a measure for cumulative photodamage.
Photodamage at high excitation rates
Aside from the cumulative photodamage described above, a broad
spectrum of phototoxic effects was observed. When calcium-sensitive dyes were used, sharp increases in F0
occasionally occurred in the absence of any electrophysiological
signal. These sudden increases were irregular, depended nonlinearly on
the applied exposure time, and were not localized to illuminated parts
of the cell. In fact, they extended some micrometers along the
dendrite. These effects were regularly observed at
> 8 and
> 30 mW for femtosecond and
picosecond pulses, respectively. Increased photobleaching and other
morphological alterations of spines and dendrites occurred at these
laser powers, such as the formation of vesicular dye-filled structures
(Fig. 4 B). For determination of r, we
restricted the laser power intensity to lower values so that only the
linear changes in basal fluorescence were observed. Experiments in
which one of the other types of artifacts were observed were excluded from data. The damage rate r ranged up to 1.6%/s for
< 8 mW (femtosecond laser pulses) and
< 30 mW (picosecond laser pulses).
Emission spectrum
A possible explanation for an increased level of fluorescence is
the generation of a new chromophore from cellular compounds. Therefore,
the emission spectrum was examined for possible changes. In some
experiments, we compared the fluorescence from throughout the
detectable spectrum (350 nm <
em < 600 nm)
with the fluorescence emitted at 525 ± 25 nm and calculated the
corresponding photodamage rates r. The dyes used have their
emission maximum in the wavelength interval around 510-540 nm. The
bandpass filter was used only for the epifluorescence signal. In this
way, it was possible to compare fluorescence at 525 nm with the
detectable fluorescence signal. The comparison was carried out for
OGB-1 (100, 200, and 500 µM) and MG (500 µM). No significant
difference between the two rates r could be observed. On
average, the r at 525 nm was 95 ± 6%
(n = 15, ± SE) of the r calculated from the
total fluorescence. This indicates that the increase in basal
fluorescence is not caused by the generation of a new chromophore.
Furthermore, no changes could be detected in the background
fluorescence FB (all dyes and laser powers,
n > 100) supporting this point of view.
Dependence of the photodamage rate r on dye concentration
Cells at similar depth at close quarters were patched and filled by whole-cell recording with the same indicator at different concentrations (OGB-1, 200 and 500 µM). In this way, it was possible to choose cell structures at very close quarters (<15 µm) and at equal depth containing different concentrations of dye. At a given average laser power, and thus, a constant TPE rate, the damage rate r was determined in both structures, and thus, at different dye concentrations. On average, we found r(500 µM)/r(200 µM) = 104 ± 15% (n = 7, ± SE). Despite the 2.5-fold difference in concentration, no significant difference between the damage rates was found. This indicates that the measure of cumulative photodamage used did not depend on dye concentration; it suggests that r is a robust measure for photodamage.
Photodamage and diffusion inhibition
The slower kinetics of the calcium fluorescence transients suggest
changes in mobility in damaged areas. Therefore, we tested the
diffusion properties of the dye upon evoking photodamage. For this
purpose, cells (n = 4) were filled with fluorescein
(500 µM) by whole-cell recording. At a location on a basal dendrite (e.g., the location marked by * in Fig.
5), a significant change in basal
fluorescence was induced by illumination. The patch pipette with
fluorescein was withdrawn and the cell repatched, this time without any
dye in the patch pipette. Washout of the dye molecules lead to a
decrease of fluorescence in the cell. At a location nearby (<10 µm
distance) and proximal to the damaged site, the fluorescence decreased
exponentially with a time constant close to that anticipated for the
dye to diffuse back through the dendrite into the pipette (
= 30-70 s, depending on the distance to the soma). However, in
comparison to that, the fluorescence from within the damaged area
decreased very slowly with a decay time >100-fold larger (
> 6000 s, in all four cells examined). This drastic decrease in
mobility indicates that a part of the dye molecules was bound and was
no longer able to diffuse. Indeed, even the very slow decrease may have
been caused mainly by photobleaching that is normally masked by
diffusion of unbleached dye molecules into the observed area.
|
An important detail was revealed by another observation. In two experiments (of four cells), the decay of fluorescence distal to the damaged site was significantly slower than its proximal counterpart: 288 versus 65 and 150 versus 72 ms, in spite of close (10 µm) proximity between the locations. That means that unaffected dye diffused more slowly through damaged areas. This difference is consistent with the hypothesis that damaged areas were changed structurally in correspondence with the observed alterations in calcium fluorescence kinetics.
Dependence of the photodamage rate r on pulse length
Changing the pulse length is well suited for estimating the
predominant order of the damage process at a given TPE rate. By adjusting the pulse length and peak intensity so that
· Ipeak2 is kept constant, the TPE rate can be
kept constant while higher- and lower-order processes are being
changed. Conversely, if the damage rate does not depend on
· Ipeak2, then the damaging
process must be a two-photon mechanism. The influence of the pulse
duration on the damage rate r was tested (see Methods) by
altering the pulse width from 75 fs to 0.8-3.2 ps. The excitation
rates used were kept low to avoid the additional, irregular
photodamaging mechanisms occurring at higher excitation rates. At these
excitation rates, r ranged up to 0.25%/s. For 22 cells,
r was measured as a function of the average laser power for
femtosecond and picosecond laser pulses within a small power interval.
For equal r, the exponent x could be derived by
means of Eq. 4. No difference was found between chirped
(n = 5) and nearly transform-limited laser pulses
(n = 17) regarding the calculated exponents; therefore
all data were pooled. We found an average exponent of x = 1.997 ± 0.38 (n = 22, ± SD, Fig.
6), indicating a two-photon-induced
photodamaging mechanism at low excitation rates.
|
Dependence of the photodamage rate r on excitation rate
It has been speculated that higher-order processes like
three-photon excitation contribute to photodamage at the peak intensity and power levels used for imaging. One might ask if, at the low excitation rates used, potential third-order components were masked by
noise. Therefore, we tested the dependence of the photodamage on the
excitation rate for femto- and picosecond laser pulses, separately. For
a given pulse length, r was measured at different average
laser powers for femtosecond (Fig.
7 A) or picosecond laser
pulses (Fig. 7 B). The dependence on
was
fitted to determine the exponent x (see Methods) using Eq. 3. We tested 25 cells. Since no differences between the calculated
exponents using different calcium dyes (MG: n = 10;
OGB-1: 100 µM, n = 2; 200 µM, n = 3; 500 µM, n = 2; OGB-2: 500 µM, n = 3) and chirped (n = 11) and nearly transform-limited
(n = 13) picosecond laser pulses were observed, all
data were pooled. For femtosecond laser pulses, the average exponent
was x = 2.5 ± 1.0 (mean ± SD,
n = 16). For picosecond laser pulses, the average
exponent was x = 2.2 ± 1.0 (mean ± SD,
n = 12). Since calcium indicators were used in these experiments, it was possible that the changes in basal fluorescence resulted from changes in resting calcium levels. Therefore, control experiments with fluorescein (500 µM) were performed. The average exponent was 2.5 ± 0.5 (mean ± SD, n = 5)
for femtosecond laser pulses. This indicates that the results did not
depend on whether a calcium indicator or an unspecific dye was used.
|
| |
DISCUSSION |
|---|
|
|
|---|
To fully exploit the benefits of two-photon laser scanning
microscopy, detailed knowledge is required about power limitations and
potential photodamage. Various criteria have been proposed to quantify
photodamage, strongly depending on the particular biological
application of multiphoton imaging. In multiphoton imaging of cultured
cells, photodamage has been related to the decrease in cell viability
and quantified through the observation of the metabolic state (Sako et
al., 1997
; König et al., 1999
). In the case of calcium imaging of
neuronal cells, photodamage can be observed as nonphysiological
functional and morphological changes. At low excitation rates, changes
in AP-evoked calcium fluorescence transients occurred. The observed
changes are independent of the occurrence of APs, clearly localized to
areas exposed to the laser, proportional to the time elapsed, and
irreversible. The fact that the basal fluorescence increased linearly
with time allowed us to define the rate of basal fluorescence increase
as a dependable measure of photodamage. The damage rates are presented in terms of average laser power. The quantification in terms of TPE
rates would be useful but compromized by several assumptions, such as
scattering in brain slices, refractive mismatch-induced abberations,
etc. Nevertheless, we want to give an estimate of the orders of
magnitudes involved; this is done in the Appendix.
A possible explanation of the increase in the basal level of
fluorescence is that some dye molecules undergo changes binding them to
the cellular matrix or membranes. Hence the basal fluorescence increased, because the mobile fraction of dye equilibrated again, raising the absolute concentration of dye. In contrast, the absolute fluorescence increase AF evoked by an AP remained constant,
indicating that the bound dye molecularly altered and no longer
reported to calcium changes. This is supported by a previous report for another dye, fura-2, where a similar intermediate photoproduct with
altered characteristics was found (Becker and Fay, 1987
). Based on
these assumptions, one can calculate that the maximal number of photons
a dye molecule emits before it binds is <106 (see
Appendix). In our case, the generation of a new chromophore is unlikely
because we did not observe a significant spectral difference between
the fluorescence from the photodamaged and undamaged areas. We also did
not observe an unexpected onset of fluorescence in cell structures that
did not contain dye molecules. The steady increase in basal
fluorescence used for quantification of cumulative photodamage should
not be confused with the abrupt increases of basal fluorescence
observed at higher intensities when a calcium-sensitive dye was used.
The latter might be caused by calcium entering through a leaking
external or internal membrane.
In addition, we found a hampered diffusion of dye molecules through damaged areas and a slower decay of the calcium fluorescence transients. This might be explained by a reduced mobility of the dye molecules. If this interpretation is right, it indicates much more severe changes than only the binding of dye molecules to intracellular structures. One possible interpretation is that the intracellular environment was heavily changed and became viscous, presumably by the binding of many proteins to each other and the cytoskeleton, comparable to what happens in fixation procedures. Such a severe change would impair the physiology of the affected cell structure because proteins cannot fulfil their functions anymore.
Several observations indicate the physiological relevance of the
measure used for the cumulative photodamage. The increase in the decay
time of the calcium fluorescence transients and the reduced diffusion
through and out of damaged cell parts suggest that the photodamage
affects physiological parameters that are independent of the
fluorophore. Furthermore, the laser intensities of excitation correlate
to those reported where photodamage in unstained cells has been
examined (König et al., 1999
).
The investigation of the role of the pulse length is of particular
interest because it reveals the dependence of the photodamage mechanism
on the excitation intensity. If we just consider the TPE rate (Eq. 2),
we find that, to achieve the same two-photon fluorescence as the
femtosecond (
1 = 100 fs) illumination, the picosecond (
2 = 2 ps) pulse train requires a
time-averaged power
, or time-averaged intensity
that is higher by a factor of
4.5 but a peak power,
or peak intensity, that is lower by the same factor. (Note that, in Eq. 2, the TPE rate is proportional to the product of average and peak
power). Our experiments show that this variation in pulse length,
average, and peak power can really be carried out and that calcium
signals can be equally well observed with picosecond laser pulses (Fig. 8). This is in accordance with previous
findings (Hänninen et al., 1994
; Hell et al., 1998
; Bewersdorf
and Hell, 1998
; Schönle and Hell, 1998
; Jenei et al., 1999
;
König et al., 1999
) and sufficient to prove that the mechanism
responsible for the cumulative photodamage is not caused by a linear
absorption mechanism such as heating through single-photon absorption.
In fact, at low excitation rates, the measurements revealed a quadratic
exponent, x = 2.0 ± 0.4, indicating that
cumulative photodamage is dominated largely by two-photon absorption.
Hence, for a given photodamage rate r, the femtosecond
pulses involved a higher peak power Ipeak in
connection with a lower average power, whereas the picosecond pulses
used a higher average power in conjunction with a lower peak power Ipeak, just as suggested by Eq. 2. As a good
rule of thumb, the rate of photodamage induced is determined by the
product of the squared intensities times the pulse length:
· Ipeak2
(P · NA2)2/
. This relationship
indicates that higher time-averaged power can be afforded if the
aperture is lower or the pulse length is longer. In other words, for
low excitation rates, the photodamage is just proportional to the
two-photon fluorescence rate.
|
Although, at low excitation rates, we found a quadratic nature for the damaging mechanism, at higher rates, we measured exponents >2.0. This might indicate that higher-order effects that are too weak to be observed at low intensities may become more important at higher excitation rates. The data presented are indeed consistent with this assumption, but this interpretation is difficult to test due to the onset of additional mechanisms of damage. These irregular artifacts observed at higher excitation rates could not be quantified and might have affected the estimation of the exponents when evaluating the dependence of the photodamage rate r on the average laser power.
In summary, the aim of this work was to clarify to what extent possible light-specimen interactions influence physiologically significant parameters in multiphoton calcium imaging. Furthermore, we investigated the role of peak intensity, excitation rate, and pulse length in photodamage. For low excitation rates, we established the rate of increase of basal fluorescence as an index of photodamage. This parameter enabled us to examine the order of the damaging process by changing the pulse length. The observed exponent of 2.0 ± 0.4 indicates two-photon absorption as the predominant damaging mechanism. This means that photodamage in TPE imaging is proportional to the signal and, thus, that signal can only be obtained at the expense of photodamage. When imaging over a longer time, low excitation rates must be used to avoid too heavy changes. This results unfortunately in a lower S/N-ratio. One way to overcome this is to optimize the collection efficiency (see Methods: Detection scheme). One should keep this in mind when applying multiphoton imaging to live cells. For low excitation rates, the results indicate that, within a window of 75 fs to 3.2 ps, the pulse width is irrelevant for two-photon calcium imaging.
Nevertheless, the use of multiphoton imaging provides several advantages over confocal microscopy in highly scattering tissue. The possibility of using ballistic photons for signal generation allows the capture of a much larger fraction of the fluorescence photons than in confocal microscopy, and the spatial confinement of excitation inflicts considerably less phototoxic stress to the cell. By carefully adjusting safe intensity levels for imaging, these benefits can be fully exploited.
| |
APPENDIX: AN ESTIMATE OF THE PHOTOSTABILITY |
|---|
|
|
|---|
This model is based on the assumption that the increase in basal
fluorescence is caused by binding of dye molecules to intracellular structures or membranes. Then, two pools of dye molecules exist, one of
free, mobile dye molecules repetitive excitation will shift a dye
molecule to the pool of bound, immobile dye molecules. The pool of the
mobile molecules is assumed to equilibrate by diffusion within a time
(about ms) much shorter than the observation time (seconds), and its
concentration is, therefore, approximated by a constant value. It would
be useful to have an estimate of the average number of excitations
before a dye molecule is shifted from the mobile pool to the immobile
pool. Cross sections are difficult to measure for TPE, thus, in most
cases, the product of cross section and quantum efficiency
2qe is given (Xu and Webb, 1996
).
For this reason, we can also calculate only the number of TPE
fluorescence photons that will be emitted before a dye molecule gets
bound. The estimate is based on several assumptions, like the
attenuation of excitation light by tissue and collection efficiency,
but it will give a feel for the orders of magnitude involved.
The model assumes that a cylindrical dendritic element of 1 µm
diameter and 0.5 µm length is illuminated. The volume V =
(0.5 µm)3 = 4 · 10
19 m3 contains B = c · V = 5 · 104 dye molecules at a dye concentration of
c = 200 µM. We assume typical values encountered in
the experiments: a gray value of G = 90 per pixel, a
pixel dwell time of tp = 4 µs, the time
the structure is illuminated during one cycle:
ti = 100 µs (see Methods), and a line
repetition rate of f = 440 Hz. At the voltage used for
the PMTs, HV = 920 V, the gain was about w = 4.0. Based on this, the number of photons collected in the
epifluorescence detection channel is
nc = (ti/tp) · (G/w) · f = 2.5 · 105
s
1. Given the detection efficiency of
=
· Qe = 0.1 · 25% = 0.025, where
is the collection efficiency of the system and
Qe the quantum efficiency of the PMTs, the
number of photons emitted per molecule per second is
nem = (nc/
) · (1/B) = 2 · 102 s
1. Another way to get an estimate
for nem is to use Eq. 2. A typical value of
= 3 mW that is attenuated to 50%,
2qe = 10
50
cm4s is assumed. Using Eq. 2, the number of photons emitted
per molecule per second is nem = 4 · 103 s
1. This number deviates from the first
number by one order of magnitude. Nevertheless, both numbers are used
as upper and lower limits.
If dye molecules in both pools exhibit the same quantum efficiency and
cross section, the number of photons emitted before a dye molecule
changes pool is given by nem/r. Given
a typical value for r = 0.4% s
1 found
for
= 3 mW, the average number of photons that
a dye molecule emits before it binds is 0.5 · 105 to
1 · 106. This number of
nem/r is independent of
because nem and r scale both with
2.
This simple model overestimates the number of emitted photons because
photobleaching is not considered. Photobleaching will reduce the number
of the dye molecules in the bound, immobile pool. The rate r
is then decreased by the rate by which the pool of immobile, but still
fluorescent, dye molecules is depleted by photobleaching. Consequently,
the real r is higher than the observed one, and thus,
nem/r is lower. In fact, the number
of useful photons might be very low. Under certain conditions
photobleaching was observed in our experiments and it will therefore
play a role. Fluorescein is known to have a high bleaching rate (about
3 · 10
5, (Hirschfeld, 1976
)). This might explain
the observation that the damage rates r observed with
fluorescein were lower than the damage rates observed with other dyes.
| |
ACKNOWLEDGMENTS |
|---|
We thank Dr. B. Sakmann for continuous support, Dr. K. Schaller (Munich, Germany) for technical assistance in optics, Dr. W. Zinth (Munich, Germany) and collaborators for helping us to set up the laser system, M. Dyba for reviewing some computations, Dr. W. F. Colmer, Dr. N. Urbian, J. Bollmann, and T. Nevian for reading the manuscript, H. Bohnet and C. Koch for programming the scanner.
| |
FOOTNOTES |
|---|
Received for publication 5 April 1999 and in final form 30 June 1999.
Address reprint requests to Helmut J. Koester, Abteilung Zellphysiologie, Max-Planck Institut für medizinische Forschung, Jahnstr. 29, D-69120 Heidelberg, Germany. Tel.: +49-6221-486-434; Fax.: +49-6221-486-459; E-mail: hkoester{at}mpimf-heidelberg.mpg.de.
| |
REFERENCES |
|---|
|
|
|---|
Biophys J, October 1999, p. 2226-2236, Vol. 77, No. 4
© 1999 by the Biophysical Society 0006-3495/99/10/2226/11 $2.00
This article has been cited by other articles: