Biophys J, March 2000, p. 1578-1588, Vol. 78, No. 3
Scanning Electrochemical Microscopy as a Local Probe of Oxygen
Permeability in Cartilage
Marylou
Gonsalves,*
Anna L.
Barker,*
Julie V.
Macpherson,*
Patrick R.
Unwin,*
Danny
O'Hare,
and
C. Peter
Winlove
*Department of Chemistry, University of Warwick, Coventry CV4 7AL;
Department of Pharmacy, University of Brighton,
Brighton BN2 4GJ; and
Physiological Flow
Studies Group, Department of Biological and Medical Systems, Imperial
College of Science Technology and Medicine, London SW7 2BY, United
Kingdom
 |
ABSTRACT |
The use of scanning electrochemical microscopy, a
high-resolution chemical imaging technique, to probe the distribution
and mobility of solutes in articular cartilage is described. In this application, a mobile ultramicroelectrode is positioned close (~1
µm) to the cartilage sample surface, which has been equilibrated in a
bathing solution containing the solute of interest. The solute is
electrolyzed at a diffusion-limited rate, and the current response measured as the ultramicroelectrode is scanned across the sample surface. The topography of the samples was determined using
Ru(CN)64
, a solute to which the cartilage matrix was
impermeable. This revealed a number of pit-like depressions
corresponding to the distribution of chondrocytes, which were also
observed by atomic force and light microscopy. Subsequent imaging of
the same area of the cartilage sample for the diffusion-limited
reduction of oxygen indicated enhanced, but heterogeneous, permeability
of oxygen across the cartilage surface. In particular, areas of high permeability were observed in the cellular and pericellular regions. This is the first time that inhomogeneities in the permeability of
cartilage toward simple solutes, such as oxygen, have been observed on
a micrometer scale.
 |
INTRODUCTION |
Permeability is a key factor governing transport
rates in biological membranes and tissues (Fournier, 1999
). Articular
cartilage is a specialized biological tissue for which permeability and fluid transport have been particularly well-studied (Maroudas, 1975
;
Bernich et al., 1976
; Allhands et al., 1984
; Mow et al., 1984
), due to
the possible association between mass transport characteristics in the
tissue and diseases such as rheumatoid and osteoarthritis (Keuttner et
al., 1992
; Muehleman and Arsenis, 1995
).
Cartilage is a connective tissue which provides a smooth,
shock-absorbing, cushioning surface at load-bearing diarthrodial joints. It is composed of an extracellular matrix (ECM) containing a
gel of negatively charged proteoglycan molecules, cells (chondrocytes), and interstitial water, embedded in a porous supporting framework of
collagen fibers (Shrive and Frank, 1994
). Under normal physiological conditions, the water content constitutes ~70% of the tissue weight. The chondrocytes, though relatively sparse (<1% of the cartilage volume), are responsible for synthesizing, maintaining, and
metabolizing the cartilage matrix components. Because cartilage is
aneural and avascular, intercell communication and the transport of
nutrients and waste products must be effected by diffusion through the
matrix. Knowledge of the transport of solutes through the cartilage
matrix is therefore crucial in understanding the physiological
functioning of the tissue, both in the healthy and diseased states. In
particular, it is thought that disturbances in fluid and solute
transport through the cartilage matrix may contribute to, or result
from, biochemical or structural changes in the tissue during the
progression of osteoarthritis (Urban and Hall, 1992
). However,
traditional experimental methodologies have, as yet, been unable to
probe mass transport with adequate spatial resolution to detect such changes either in vitro or in vivo.
A number of studies have been made on the bulk diffusion of solutes
through cartilage (Maroudas, 1970
; Bernich et al., 1976
; Roberts
et al., 1996
; Torzilli et al., 1997
, 1998
). However, it is known that
cartilage is structurally heterogeneous, evident from various
microscopy studies (Horky, 1993
; Jurvelin et al., 1996
). Some workers
have addressed this issue by taking samples (50-200 µm thick) at
varying depths from the articular surface (Bernich et al., 1976
;
Torzilli et al., 1997
, 1998
), but data were still averaged across
the surface of the slice, thereby neglecting any lateral
inhomogeneities in transport rates. More recently, magnetic resonance
imaging has emerged as a useful tool for the study of biological
tissues, and a number of measurements of solute and water transport in
cartilage have been reported (Burstein et al., 1993
; Fischer et al.,
1995
; Knauss et al., 1996
; Potter et al., 1997
). This technique has the
advantage of being non-invasive, with in vivo capabilities, and
currently has the potential to achieve resolution on a scale of tens of
micrometers (Potter et al., 1997
). The technique is, however, limited
to the study of paramagnetic species.
Scanning electrochemical microscopy (SECM) is a powerful technique for
examining the diffusive, convective, and migratory transport of
solutes. In SECM, an ultramicroelectrode (UME), attached to
piezoelectric positioners, is mobile in three dimensions. The UME can
be positioned close to an interface with submicron precision, and can
probe the topography, reactivity, or permeability of that interface
with high spatial resolution (Bard et al., 1991b
; Barker et al., 1999
).
SECM has been applied to the study of a number of synthetic membranes
and biomaterials including skin (Bath et al., 1998
; Scott et al., 1991
;
1993a
, b
; 1995
), dentine (Macpherson et al., 1995a
, b
; Unwin et al.,
1997
), and bilayer lipid membranes (Matsue et al., 1994
). SECM has the
advantage over scanning ion conductance microscopy, which has found
some application in the investigation of membrane transport (Hansma et
al., 1989
; Korchev et al., 1997
), in that it can selectively detect
both neutral and charged species, rather than measure total ion currents.
In a recent study we used SECM to image osmotically driven convective
transport through laryngeal cartilage (Macpherson et al., 1997
) and
were able, for the first time, to correlate local convective fluxes
with sample topography on a microscopic scale. The study of spatially
resolved localized diffusion, however, has proved more difficult. To
investigate the related transport properties of permeability and
diffusion, we have recently introduced an SECM induced-transfer
(SECMIT) approach (Barker et al., 1998
). In SECMIT, a solute is
partitioned between two phases at equilibrium. A UME, positioned in one
phase close to the interface, is biased so as to deplete the local
concentration of solute at the tip. This perturbs the equilibrium,
driving the transfer of solute from the second phase to the tip. Hence,
measurements of diffusion in the second phase can be made without the
need to enter or contact the second phase, which is particularly
advantageous for measurements in biological tissues where sample
integrity is essential. However, this technique has not, as yet, been
used to image variations in permeability across an interface.
In this paper we describe the use of SECMIT to probe the diffusive
transport of solutes through cartilage in order to further understanding of the relationship between tissue structure and local
permeability. Particular attention is given to oxygen, due to its
general biological relevance and specific role in cartilage metabolism
(Stockwell, 1983
). The high-resolution capabilities of the technique
offer the possibility of probing transport processes occurring at the
level of a single cell. In future studies this will be particularly
useful for monitoring the metabolic rates of chondrocytes. To examine
viable (living) cartilage, a preliminary study is made here to
investigate transport properties in non-metabolizing ECM tissue.
 |
EXPERIMENTAL DETAILS |
Materials
Bovine articular cartilage, from the metacarpal phalangeal
joints of mature animals, was obtained fresh from the abattoir and
stored at
20°C before use. Full-depth plugs of cartilage were
removed from thawed joints using a 5 mm diameter cork borer and cut
into 50-µm-thick sections parallel to the articular surface on a
microtome (Model 5030, Bright Instruments, Huntington, UK).
Before SECM experiments, the sections were equilibrated in a 0.2 mol
dm
3 potassium chloride (analytical reagent,
Fisons, UK) solution, phosphate-buffered to pH 7.0 ± 0.2. SECM
approach curves, imaging measurements, and atomic force microscopy
(AFM) were carried out with the cartilage sample bathed in a solution
containing 5 × 10
3 mol
dm
3 potassium hexacyanoruthenate (II)
(K4Ru(CN)6) (Alfa, Royston, UK) and 0.2 mol dm
3 potassium chloride,
phosphate-buffered to pH 7.0 ± 0.2. Chronoamperometric measurements were made in 0.02 mol dm
3
K4Ru(CN)6 and 0.2 mol
dm
3 potassium chloride, phosphate-buffered to
pH 7.0 ± 0.2. Samples were allowed to soak in the bathing
solution for at least 1 h before measurements. Given the thinness
of the samples, this was sufficient time for the system to equilibrate.
All solutions were prepared under ambient conditions, and measurements
were made at room temperature (22 ± 2°C). Representative
cartilage sections were stained for histological analysis using
standard protocols (Carleton, 1980
): van Gieson's stain for collagen,
hematoxylin, and eosin for cell nuclei and toluidine blue for proteoglycans.
Instrumentation
The general SECM set-up has been described previously
(Macpherson et al., 1995a
). For tip approach experiments, the UME was a
25-µm-diameter platinum disk electrode, embedded in a glass insulating sheath with the overall tip diameter 10 times that of the Pt
disk, i.e., 250 µm. A 5-µm-diameter platinum disk electrode, embedded in a glass capillary, with an overall tip diameter of 50 µm,
was used for all imaging and chronoamperometric measurements. The UME
served as the working electrode in a conventional two-electrode set-up,
with a silver wire quasi-reference electrode (AgQRE), against which all
potentials are quoted. For chronoamperometric measurements, the
current-time transients were recorded on a digital oscilloscope
(Model NIC-310, Nicolet Technologies, Milton Keynes, UK).
The 50-µm-thick cartilage sections were fixed on a glass disk
(diameter 12.7 mm, thickness 1.6 mm) using a 1:1 volume mixture of nail
varnish and cyanoacrylate glue applied around the circumference of the
sample, while ensuring that the sample remained flat and hydrated
during the fixing procedure. The glass disk was then mounted in the
base of an SECM cell so that the disk was perpendicular to the UME tip
axis. Optical micrographs of the samples were taken using an Olympus
BH2 microscope equipped with a 3-CCD color video camera (model
KY-F55BE, JVC Professional, London, UK) coupled to a computerized data
acquisition system (Image Grabber/PCI, Neotech Ltd., London, UK).
AFM images were made in contact mode under solution using a Nanoscope E
atomic force microscope and fluid cell (Digital Instruments, Santa
Barbara, CA). The silicon nitride AFM probe had a nominal spring
constant of 0.06 N m
1.
Procedures
Tip approach measurements
Tip approach curves were acquired by holding the UME, of radius
a = 12.5 µm, at a fixed x-y location (Fig.
1 A), and the current response
for the diffusion-limited electrolysis of the solute measured as the
UME was translated in the z-direction from a position in
bulk solution to one close to the interface. The tip potential was held
at a value where electrolysis of the solute was diffusion-limited, identified by recording a steady-state voltammogram. This was 1.2 V for
Ru(CN)64
oxidation and
0.5 V for
O2 reduction. The UME was then scanned toward the
interface at a velocity of 1 µm s
1, and the
limiting current response, i (normalized with respect to the
steady-state current with the tip positioned at an effectively infinite
distance from the sample interface, i(
)), recorded as a
function of tip-sample separation, d. By analyzing the
i/i(
)-d curves, the transport properties of
the solutes of interest in cartilage were analyzed using a theory
outlined in full elsewhere (Barker et al., 1998
).

View larger version (33K):
[in this window]
[in a new window]
|
FIGURE 1
Schematics depicting the different modes of SECM
operation for (A) tip approach measurements and
(B) imaging experiments (not to scale).
|
|
SECM imaging
SECM images were obtained by holding the UME tip
(a = 2.5 µm) at a constant z position and
scanning in the x-y plane over an area of interest (Fig. 1
B). For a solute to which the cartilage was impermeable,
changes in the diffusion-limited current response were due to
variations in the sample-tip separation, from which topographical
information on the sample surface could be obtained (Mirkin et al.,
1992
). When the same sample area was subsequently scanned, using a
solute toward which cartilage was permeable, the current response
provided a permeability map of the area.
For Ru(CN)64
oxidation the tip was held at a
potential of 1.2 V. After recording the steady-state current,
i(
), with the tip positioned far from the surface, the
height of the UME tip above the cartilage sample was adjusted until
i/i(
) attained a value of 0.25. Based on the data herein,
this corresponded to a sample-tip separation of ~1 µm. The UME was
then scanned over a 100 µm × 100 µm area at 5 µm
s
1 in unidirectional lines with a separation of
5 µm between line scans, and the tip current recorded as a function
of lateral tip position. At the end of the scan, the UME was retracted
200 µm from the cartilage surface and the bulk solution limiting
current verified.
For oxygen permeability scans the UME was moved to the same initial
position as for the Ru(CN)64
oxidation scans, and
repolarized to
0.5 V for diffusion-limited oxygen reduction. The UME
was scanned over the same 100 µm × 100 µm area at the same
scan rate.
Chronoamperometry
For chronoamperometry, the UME was positioned over an area of
interest (identified from SECM imaging), close to the tissue/bathing solution interface, and the current recorded as a function of time
after jumping the potential from a value where no current flowed to one
where solute electrolysis was diffusion-limited. All chronoamperometric
measurements were made at a 5 µm diameter Pt UME with
Ru(CN)64
as the solute.
 |
RESULTS AND DISCUSSION |
The results presented herein are for a single cartilage sample,
but are typical of >20 samples studied.
Structural and chemical characterization of the cartilage surfaces
Cartilage is known to be structurally and chemically heterogeneous
on a local (submicron) scale (Shrive and Frank, 1994
). Fig.
2 shows a light micrograph of an area of
cartilage comparable to that scanned by the SECM. A number of pit-like
features can be clearly observed, with diameters ranging from 15 µm
to 25 µm. An AFM image of a neighboring field, recorded under
solution, is shown in Fig. 3. Again, a
number of recessed regions are clearly visible, with diameters of ~20
µm and depths of 2-3 µm. These recesses correspond to cells or,
more usually, groups of two or more cells. Although the pericellular
matrix may have been distorted during the freezing/thawing process, the
nucleus is still intact in the majority of cells. In the
interterritorial matrix, surface irregularities are of the order of a
micron in depth.

View larger version (83K):
[in this window]
[in a new window]
|
FIGURE 3
Atomic force microscopy image of a cartilage section,
recorded under solution (5 × 10 3 mol
dm 3 K4Ru(CN)6 and 0.2 mol
dm 3 KCl, phosphate-buffered to pH 7 ± 0.2).
|
|
The rate of water and solute transport through the cartilage matrix is
expected to be greatly influenced by the local biochemical composition,
based on observations for other tissues (Weinberg et al., 1997
).
Micrographs of histologically stained cartilage sections are shown in
Fig. 4. On this scale, the nuclei and
cell boundaries are clearly visible. Collagen staining is light in the
cellular and pericellular regions, and most intense in the interterritorial region, some distance from the cells. In the case of
proteoglycan, the staining is lightest in the area surrounding the
nucleus and more intense in the pericellular and interterritorial regions. This pattern of staining is similar to that observed in
previous studies (Hunziker et al., 1997
).

View larger version (75K):
[in this window]
[in a new window]
|
FIGURE 4
Light micrographs of histologically stained cartilage
sections: (a) hematoxylin and eosin for cell nuclei,
(b) toluidine blue for proteoglycans, and
(c) van Gieson's stain for collagen.
|
|
Tip approach measurements
In order to determine the surface topography of the cartilage
sample under SECM conditions, a solute was required that did not
permeate the matrix. An initial estimate of the permeability of solutes
in the cartilage matrix was obtained by analyzing the i/i(
) response of the UME tip as it approached the
interface. By using a relatively large UME, the tip response is
averaged over a sizeable area and approaches that representative of the "bulk" behavior of the cartilage. As the topographical features on
the cartilage surface are on the order of a few microns in depth, the
surface may be considered as approximately planar (for the approach by
a large UME).
Ru(CN)64
oxidation
Ru(CN)64
is a large, highly charged anion, and
so would not be expected to significantly permeate the matrix, as
cartilage is negatively charged at physiological pH (Maroudas, 1975
).
Typical approach curves for Ru(CN)64
oxidation at a
25 µm diameter UME approaching both a flat glass disk and cartilage
tissue are shown in Fig. 5. The
experimental responses are seen to be in excellent agreement with the
theoretical prediction for an impermeable planar substrate (Kwak and
Bard, 1989
) in both cases. The tip is able to attain a slightly closer distance to the glass surface than the cartilage surface due to the
irregularities in the topography of the latter structure. Nevertheless, the tip is able to approach the cartilage surface to
within a micron, and the data clearly demonstrate that
Ru(CN)64
shows negligible induced-diffusion through
the cartilage, making it a suitable solute for topographical
measurements.

View larger version (16K):
[in this window]
[in a new window]
|
FIGURE 5
Approach curves of normalized current versus sample-tip
separation for the diffusion-controlled oxidation of
Ru(CN)64 at a 25 µm diameter Pt UME approaching
cartilage ( ) and a flat glass surface ( ). The dashed line
represents the theoretical response for approach to an impermeable
substrate (Kwak and Bard, 1989 ).
|
|
Oxygen reduction
Oxygen is a small, neutral molecule that permeates tissues and
membranes (Bicher and Bruley, 1973
). Fig.
6 is a typical voltammogram for the
reduction of oxygen at a 25-µm-diameter tip, clearly showing that the
electrolysis process is diffusion-limited at a potential of
0.5 V,
used for the approach curve measurements shown in Fig. 7. The experimental response in Fig. 7
for the cartilage surface does not fit the theory for approach to an
impermeable substrate, found when the same experiment is carried out at
a flat glass surface, the data for which are also shown in Fig. 7.
Rather, the current is significantly higher than the theoretical
response at all values of d. There is still a significant
current (i/i(
) ~ 0.5) at the point where the UME
contacts the cartilage surface. We therefore deduce that the UME
promotes diffusion of O2 through the cartilage
matrix to the tip: local depletion of O2 in the tip-substrate gap induces transfer from the sample to solution to
restore equilibrium. Hence, the approach curves demonstrates that
oxygen shows appreciable permeability in cartilage.

View larger version (14K):
[in this window]
[in a new window]
|
FIGURE 6
Linear sweep voltammogram for the reduction of oxygen
at a 25-µm-diameter Pt UME in a solution containing 5 × 10 3 mol dm 3
(K4Ru(CN)6) and 0.2 mol dm 3
potassium chloride, phosphate-buffered to pH 7.0 ± 0.2, recorded
at a scan rate of 20 mV s 1.
|
|

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 7
Approach curve of normalized current versus sample-tip
separation for the diffusion-controlled reduction of oxygen at a
25-µm-diameter Pt UME approaching cartilage ( ) and a flat glass
surface ( ). The short-dashed line represents the theoretical
response for approach to an impermeable substrate (Kwak and Bard,
1989 ). The long-dashed lines represent the theoretical response for
induced diffusion for different values of
Ke (Barker et al., 1998 ): 0.6 (top
line) and 0.5 (bottom line).
|
|
The relative permeability of a solute in cartilage, compared to the
bathing solution, may be expressed as the product,
Ke
(Maroudas, 1975
), where
Ke is the partition coefficient of the solute in the cartilage (ratio of the concentration in the cartilage to
that in the contacting solution) and
is the ratio of the solute diffusion coefficient in cartilage to that in solution. Theoretical simulations (Barker et al., 1998
) of i/i(
)
vs. d at different
Ke
values indicate that a
value of Ke
between 0.5 and
0.6 best describes the experimental data, as indicated by the dashed
lines in Fig. 7. This is similar to the value of
Ke
measured previously for
oxygen in laryngeal cartilage (Macpherson et al., 1997
).
SECM imaging
Sample topography
For the reasons outlined above, Ru(CN)64
was
used to image the cartilage topography. A typical contour plot of
normalized current as a function of lateral probe position for
Ru(CN)64
oxidation is shown in Fig.
8. The dark areas indicate higher i/i(
) values. Clearly, the i/i(
) response
is heterogeneous over the scanned area. There are well-defined circular
regions of enhanced current. Under steady-state conditions the
sample-tip separation, d, can be evaluated from the
following relationship between i/i(
) and d
(Mirkin et al., 1992
):
|
(1)
|
Although this equation strictly applies to a planar interface, it
provides useful semi-quantitative information on sample topography
(Bard et al., 1994
). For the electrode used in this study the residence
time in the vicinity of a spot on the sample, tres, is 1 s (where
tres =
2a/
tip, and
tip is the tip scan speed). This is more
than sufficient time for steady-state conditions to be established
(Bard et al., 1991a
).

View larger version (98K):
[in this window]
[in a new window]
|
FIGURE 8
Contour plot showing the normalized current map for a
100 × 100 µm area of cartilage, imaged using the
diffusion-controlled oxidation of Ru(CN)64 at a
5-µm-diameter Pt UME at an initial sample-tip separation of ~2
µm.
|
|
Fig. 9 shows the topography image
derived, using Eq. 1, from the normalized current data in Fig. 8. It
can be seen that the circular regions of current in Fig. 8 correspond
to pits in the cartilage surface of diameter 15 to 25 µm and depths
between 2 and 3 µm. These dimensions are in excellent agreement with
the topographical features observed by optical microscopy and AFM (Figs. 2 and 3), demonstrating that for this particular surface, hindered diffusion imaging provides a good means of determining sample
topography.

View larger version (100K):
[in this window]
[in a new window]
|
FIGURE 9
Contour plot showing the corresponding topography map,
calculated from Eq. 1, for the area of cartilage imaged in Fig. 8.
|
|
For the permeability studies that follow, it was essential that the
Ru(CN)64
data gave a true representation of the
sample topography. As an additional check on the impermeability of
cartilage to Ru(CN)64
, and as a complement to the
approach curve measurements made with the large electrode,
high-resolution chronoamperometric measurements were made at defined
positions close to the interface. In particular, we were interested in
further demonstrating that the electrochemical imaging method provided
an accurate picture of the topography of the recessed regions.
Chronoamperometry has been shown to be particularly sensitive to the
partition coefficient and relative diffusion coefficients of solutes in
two-phase systems (Barker et al., 1998
). For these studies, an area of
cartilage was initially imaged with Ru(CN)64
to
identify a single pit, followed by chronoamperometric measurements at a
fixed point in the recess. Fig. 10
shows typical results, made at d = 0.6 µm, presented
as i/i(
) vs. t (A) and
t
1/2 (B) so as to
emphasize the long and short time behavior, respectively, for
Ru(CN)64
oxidation close to the cartilage surface.
Also shown is the theoretical response at an impermeable substrate for
a log(d/a) value of
0.6. The observation of close
agreement between theory and experiment over a wide time scale further
confirms that, under the conditions of our SECM experiments,
Ru(CN)64
can be considered as having negligible
permeability in the cartilage matrix.

View larger version (11K):
[in this window]
[in a new window]
|
FIGURE 10
Plots of i/i( ) vs.
(A) t and (B)
t 1/2 for the diffusion-controlled
oxidation of Ru(CN)64 at a 5-µm-diameter Pt UME
recorded at d = 0.6 µm from the cartilage surface
( ). The theoretical response for an impermeable substrate with the
tip at a log(d/a) of 0.6 is also shown ( -).
|
|
Oxygen permeability
Fig. 11 shows a diffusion-limited
current map for oxygen reduction recorded with the tip in the same
z position and over the same area considered for Figs. 8 and
9. Similar circular regions of enhanced current can be seen, as were
found in Fig. 8. However, the overall i/i(
) values are
higher than those observed for the Ru(CN)64
oxidation
image (Fig. 8), which can be attributed to the induced diffusion of
oxygen through the cartilage matrix. Using a computer simulation
(Barker et al., 1998
), the effect of permeability, defined as
Ke
, on i/i(
)
was evaluated as a function of sample-tip separation, from which the
relationship between i/i(
), d, and Ke
was derived. In this way,
the normalized current data for O2 reduction was
processed to take account of the varying tip-substrate separation (Fig.
9), yielding a map of oxygen permeability in the cartilage sample. A
permeability plot for the i/i(
) data in Fig. 11 is shown
in Fig. 12. It is evident that the
permeability is not uniform. Significant variations are observed: the
value of Ke
ranges from
~0.4 over most of the interterritorial matrix to 0.7 over the
recesses in the cartilage surface, corresponding to the cells.
Considering the data in Fig. 12 together with the histochemical
composition of cartilage (Fig. 4), it is evident that there is high
oxygen permeability in the cellular and pericellular regions. The
intensity of toluidine blue staining, seen in Fig. 4 b,
indicates a higher fixed charge density in the pericellular matrix.
Whether this arises from differences in the type, organization, or
accessibility of the proteoglycans is unclear. However, the higher
staining intensity does not correspond to an increase in oxygen
permeability, favoring the latter possibility. Evidence from various
sources indicates that glycosaminoglycans have only a small effect on
the diffusivity of small solutes (Gribbon et al., 1998
). There is a
strong relationship, however, between permeability and the intensity of
collagen staining by van Gieson's stain (Fig. 4 c), with
the lowest permeability observed in regions of intense staining.
Although factors such as collagen type and fiber organization, as well
as collagen content, influence the staining intensity, it is probable
that collagen content affects the diffusivity and/or distribution of
the analytes, both of which contribute to the measured permeability.
This is the first time that lateral heterogeneities of solute
permeability in cartilage have been observed on a micrometer scale.

View larger version (107K):
[in this window]
[in a new window]
|
FIGURE 11
Contour plot showing the normalized current map for
the diffusion-controlled reduction of oxygen at a 5-µm-diameter Pt
UME, in the area of cartilage considered in Fig. 8.
|
|
The analysis of the oxygen permeability data assumes a planar
interface. In order to ascertain that the large enhancements in oxygen
reduction currents observed over the recessed regions were not simply
due to the non-planar geometry of the interface in these locations, a
new SECM model was developed to predict the steady-state response with
a UME close to and directly over a symmetrical recess. The simulations
and theoretical results, summarized in the Appendix, showed that for
the Ke
values of interest, a
planar substrate model suffices to a good approximation for the
situation where the tip is directly above a depression in the surface.
While there may be higher-order effects from the heterogeneous nature
of the substrate topography, the results in Fig. 12 represent a
good description of the variation in oxygen permeability in the surface
region of cartilage.
 |
CONCLUSIONS |
This study has shown SECMIT to be a powerful tool for probing
diffusion processes in cartilage tissue, specifically allowing oxygen
diffusion through the cartilage matrix to be correlated with tissue
morphology. Enhanced oxygen diffusion has been detected in areas that
are low in collagen content. This is the first time that lateral
inhomogeneities in tissue permeability have been observed on a
micrometer-length scale.
Following this preliminary study on non-metabolizing cartilage tissue,
we should now be able to examine metabolic processes occurring in
viable cartilage tissue. As previously demonstrated (Macpherson et al.,
1997
), SECM measurements of transport can be made in cartilage both in
the presence and absence of an applied pressure. Hence, measurements
should be possible under conditions similar to those of in vivo loaded
cartilage, which is particularly important for the metabolic studies.
We also plan to examine transport processes in both healthy and
diseased cartilage. Finally, it should be noted that the measurement of
spatially resolved diffusion processes using the SECM protocol
described in this paper has considerable potential application to a
wide range of biological tissues and membranes.
 |
APPENDIX |
Effect of substrate geometry on the SECM tip current response
The SECM induced-transfer (SECMIT) technique can be used to
probe the permeability of a target solute in a sample not in direct contact with the UME. Details of the numerical model developed for this
approach with a planar surface have been given previously (Barker et
al., 1998
). The following brief account outlines the formulation of a
numerical model applicable to substrates with more complex geometry; in
particular, for surfaces that contain pit-like depressions, which more
closely resemble the biological samples studied in this paper.
The principle of the SECMIT methodology and the coordinate system used
to define the new model are illustrated in Fig.
13. The UME is positioned in the
aqueous phase (phase 1) close to the surface of the substrate (phase
2), directly above a circularly symmetric pit in the surface, such that
the combined tip/substrate geometry is axisymmetric. The two phases
contain a common electroactive species, A. The interface between the
two phases is assumed to be sharply defined and the transfer of species
A across the interface is not kinetically limited. Initially, with the
partitioning of A across the interface in dynamic equilibrium, there is
zero net flux of species A across the interface and each phase has a
uniform bulk concentration of A, c*i,
where i = 1 or 2 denotes the phase. A potential step is
subsequently applied to the UME sufficient to electrolyze species A at
a diffusion-controlled rate. This perturbs the interfacial equilibrium,
inducing transfer of species A across the interface from phase 2 to
phase 1. The flux of species A to the UME, and hence the tip-current
response, are dependent on the rate of mass transfer of the species in
each phase.

View larger version (82K):
[in this window]
[in a new window]
|
FIGURE 13
The coordinate system used to define the model
outlined in the Appendix. The coordinates r and
z are in the directions radial and normal to the
electrode surface, respectively. The electrode radius is denoted by
a, rg is the distance from
the center of the electrode to the outermost edge of the insulating
glass sheath surrounding the electrode, rp
is the radius of the circularly symmetric pit in the surface of the
substrate (phase 2), d is the distance from the surface
of the electrode to the bottom of the pit, and
dp is the height of the pit. Phase 2 is
considered to extend a semi-infinite distance in the
z-direction. Species A is the common electroactive
mediator in the two phases, while B denotes the product of the
electrode reaction.
|
|
Formulation of the problem
Time-dependent diffusion equations, appropriate to the
axisymmetric SECM geometry, can be written for the species A in each phase.
|
(A1)
|
where r and z are the coordinates in the
directions radial and normal to the electrode surface, respectively,
measured from the center of the electrode;
ci and
Di are the concentration and diffusion
coefficient of the electroactive species, A, in phase i; and
t is time. As shown in Fig. 13,
rg denotes the radius of the probe
(electrode plus insulating sheath), rp
is the radius of the pit, d represents the separation
between the end of the probe and the bottom of the pit, and
dp is the depth of the pit. The
calculation of the tip current response involves solving the diffusion
equation A1, subject to the boundary and initial conditions of the
system. Before the potential step, the initial condition is
|
(A2)
|
Following the potential step, A is electrolyzed at a
diffusion-controlled rate at the electrode, but is inert with respect to the insulating glass sheath surrounding the electrode, and remains
at bulk concentration values beyond the radial edge of the tip. The
exterior boundary conditions may be summarized as follows:
|
(A3)
|
|
(A4)
|
|
(A5)
|
This latter condition is valid provided that RG = rg/a
10, where a is the
radius of the electrode (Kwak and Bard, 1989
).
The axisymmetric SECM geometry implies there is no radial flux of
species A at the cylindrical axis of symmetry:
|
(A6)
|
At a semi-infinite distance from the electrode, in phase 2, the
electroactive species attains its bulk concentration,
c*2.
|
(A7)
|
The final internal boundary conditions apply to the surface of
the substrate:
|
(A8)
|
|
(A9)
|
for
where ci,int is the concentration of
A at the interface in phase i and
Ke = c*2/c*1.
To formulate a general solution, the diffusion equations and boundary
conditions were cast into dimensionless form through the introduction
of the following normalized terms:
|
(A10)
|
|
(A11)
|
|
(A12)
|
|
(A13)
|
|
(A14)
|
The aim of the calculation was to determine the tip current
response as a function of time and tip/interface separation for particular Ke
values. The UME
current is related to the flux of c1
at the electrode surface,
|
(A15)
|
where n is the number of electrons transferred per
redox event and F is Faraday's constant.
The normalized current ratio is given by:
|
(A16)
|
where i(
) is the steady-state diffusion-limited
current at an inlaid disk electrode positioned at an effectively
infinite distance from the interface (Saito, 1968
),
|
(A17)
|
The problem was solved numerically using the alternating
direction implicit finite-difference method (ADIFDM). General details on the application of this method to solve a variety of SECM problems have been given previously (see, for example, Barker et al., 1998
). The
modifications required to treat the present problem are straightforward and will not be discussed further. The ADIFDM method evaluates the
current-time response of the UME; however, for the present problem only
the steady-state current, derived from the chronoamperometric data in
the long time limit, was of interest.
Simulations were performed for parameter values appropriate to the
experimental system under study, using
= 0.5 and
Ke = 1 and a UME characterized by
RG = 10. The plot in Fig.
14 shows the steady-state current as a
function of normalized pit radius (rp/a) for a fixed value of
d/a = 0.8 (which for a UME with a radius of
2.5 µm corresponds to d = 2 µm) for three different pit wall depths. For a large pit radius, i.e.,
rp/a
10, the steady-state current for all pit depths approaches that predicted for a
planar substrate at a distance of d/a = 0.8 from the electrode, plotted as the dashed line in Fig. 14. As the
radius of the pit is decreased the steady-state current decreases below
this limit. The effect is particularly pronounced as the pit depth,
dp is increased. For the recesses
observed in the cartilage surface in this study,
rp/a can approach 5, and it
is clear that under these conditions the assumption of a planar
interface in deriving permeability data is a good approximation for the
typical tip-substrate separations employed.

View larger version (14K):
[in this window]
[in a new window]
|
FIGURE 14
Simulated normalized steady-state current as a
function of normalized pit radius
(rp/a) for = 0.5, Ke = 1, and a distance
d/a = 0.8. Normalized pit depths,
dp/a, take the values
(a) 0.2, (b) 0.4, (c) 0.6. The dashed line represents the steady-state current predicted for a UME
at a distance d/a = 0.8 above a
planar surface.
|
|
 |
ACKNOWLEDGMENTS |
The authors thank Rosemary Ewins for preparation and histological
staining of the samples.
M.G. gratefully acknowledges financial support from the Wellcome Trust.
A.L.B. and J.V.M thank the EPSRC for support.
 |
FOOTNOTES |
Received for publication 10 June 1999 and in final form 14 December 1999.
Address reprint requests to Prof. P. R. Unwin, Department of
Chemistry, University of Warwick, Coventry CV4 7AL, UK. Tel.:
+44-24-7652-3264; Fax: +44-24-7652-4112; E-mail:
P.R.Unwin{at}warwick.ac.uk.
 |
REFERENCES |
-
Allhands, R. V.,
P. A. Torzilli, and F. A. Kallfelz.
1984.
Measurement of diffusion of uncharged molecules in articular cartilage.
Cornell Vet.
74:111-123[Medline].
-
Bard, A. J.,
G. Denuault,
R. A. Freisner,
B. C. Dornblaser, and L. Tuckerman.
1991a.
Scanning electrochemical microscopy: theory and application of the transient (chronoamperometric) SECM response.
Anal. Chem.
63:1282-1288[Medline].
-
Bard, A. J.,
F.-R. F. Fan, and M. V. Mirkin.
1994.
Scanning electrochemical microscopy.
In
Electroanalytical Chemistry, Vol. 18.
A. J. Bard, editor. Marcel Dekker, New York. 243-373.
-
Bard, A. J.,
F.-R. F. Fan,
D. T. Pierce,
P. R. Unwin,
D. O. Wipf, and F. Zhou.
1991b.
Chemical imaging of surfaces with the scanning electrochemical microscope.
Science.
254:68-74.
-
Barker, A. L.,
M. Gonsalves,
J. V. Macpherson,
C. J. Slevin, and P. R. Unwin.
1999.
Scanning electrochemical microscopy: beyond the solid/liquid interface.
Anal. Chim. Acta.
385:223-240.
-
Barker, A. L.,
J. V. Macpherson,
C. J. Slevin, and P. R. Unwin.
1998.
Scanning electrochemical microscopy (SECM) as a probe of transfer processes in two-phase systems: theory and experimental applications of SECM-induced transfer with arbitrary partition coefficients, diffusion coefficients and interfacial kinetics.
J. Phys. Chem. B.
102:1586-1598.
-
Bath, B. D.,
R. D. Lee,
H. S. White, and E. R. Scott.
1998.
Imaging molecular transport in porous membranes. Observation and analysis of electroosmotic flow in individual pores using the scanning electrochemical microscope.
Anal. Chem.
70:1047-1058.
-
Bernich, E.,
R. Rubenstein, and J. S. Bellin.
1976.
Membrane transport properties of bovine articular cartilage.
Biochim. Biophys. Acta.
448:551-561[Medline].
-
Bicher, H. I., and D. F. Bruley, editors.
1973.
Oxygen Transport to Tissue. Instrumentation, Methods and Physiology. Plenum Press, London.
-
Burstein, D.,
M. L. Gray,
A. L. Hartman,
R. Gipe, and B. D. Foy.
1993.
Diffusion of small solutes in cartilage as measured by nuclear magnetic resonance spectroscopy and imaging.
J. Orthop. Res.
11:465-478[Medline].
-
Carleton, H. M.
1980.
Carleton's Histological Technique. Oxford University Press, Oxford.
-
Fischer, A. E.,
T. A. Carpenter,
J. A. Tyler, and L. D. Hall.
1995.
Visualisation of mass transport of small organic molecules and metal ions through articular cartilage by magnetic resonance imaging.
Magn. Reson. Imaging.
13:819-826[Medline].
-
Fournier, R.
1999.
Basic Transport Phenomena in Biomedical Engineering. Taylor and Francis, Philadelphia.
-
Gribbon, P. M.,
A. Maroudas,
K. H. Parker, and C. P. Winlove.
1998.
Water and solute transport in the extracellular matrix: physical principles and macromolecular determinants.
In
Connective Tissue Biology. Integration and Reductionism.
R. K. Reed, and
K. Rubin, editors. Portland Press, London. 95-123.
-
Hansma, P. K.,
B. Drake,
O. Marti,
S. A. C. Gould, and C. B. Prater.
1989.
The scanning ion-conductance microscope.
Science.
243:641-643[Medline].
-
Horky, D.
1993.
The submicroscopic structure of articular cartilage in the adult pig.
Acta Vet. Brno.
62:9-18.
-
Hunziker, E. B.,
M. Michel, and D. Studer.
1997.
Ultrastructure of adult human articular cartilage matrix after cryotechnical processing.
Microscopy Research and Technique.
37:271-284[Medline].
-
Jurvelin, J. S.,
D. J. Müller,
M. Wong,
D. Studer,
A. Engel, and E. B. Hunziker.
1996.
Surface and subsurface morphology of bovine humeral articular cartilage as assessed by atomic force and transmission electron microscopy.
J. Struct. Biol.
117:45-54[Medline].
-
Keuttner, K. E.,
R. Schleyerbach,
J. G. Peyron, and V. C. Hascall, editors.
1992.
Articular Cartilage and Osteoarthritis. Raven Press, New York.
-
Knauss, R.,
G. Fleischer,
W. Grunder,
J. Kärger, and A. Werner.
1996.
Pulsed field gradient NMR and nuclear magnetic relaxation studies of water mobility in hydrated collagen II.
Magn. Reson. Med.
36:241-248[Medline].
-
Korchev, Y. E.,
C. L. Bashford,
M. Milovanovic,
I. Vodyanoy, and M. J. Lab.
1997.
Scanning ion conductance microscopy of living cells.
Biophys. J.
73:653-658[Abstract].
-
Kwak, J., and A. J. Bard.
1989.
Scanning electrochemical microscopy. Theory of the feedback mode.
Anal. Chem.
61:1221-1227.
-
Macpherson, J. V.,
M. A. Beeston,
P. R. Unwin,
N. P. Hughes, and D. Littlewood.
1995a.
Scanning electrochemical microscopy as a probe of local fluid flow through porous solids.
J. Chem. Soc. Faraday Trans.
91:1407-1410.
-
Macpherson, J. V.,
M. A. Beeston,
P. R. Unwin,
N. P. Hughes, and D. Littlewood.
1995b.
Imaging the action of fluid flow blocking agents on dentinal surfaces using a scanning electrochemical microscope.
Langmuir.
11:3959-3963.
-
Macpherson, J. V.,
D. O'Hare,
P. R. Unwin, and C. P. Winlove.
1997.
Quantitative spatially resolved measurements of mass transfer through laryngeal cartilage.
Biophys. J.
73:2771-2781[Abstract].
-
Maroudas, A.
1970.
Distribution and diffusion of solutes in articular cartilage.
Biophys. J.
10:365-379[Medline].
-
Maroudas, A.
1975.
Biophysical chemistry of cartilaginous tissue with special reference to solute and fluid transport.
Biorheology.
12:233-248[Medline].
-
Matsue, T.,
H. Shiku,
H. Yamada, and I. Uchida.
1994.
Permselectivity of voltage-gated alamethicin ion channel studied by microamperometry.
J. Phys. Chem.
98:11001-11003.
-
Mirkin, M. V.,
F.-R. F. Fan, and A. J. Bard.
1992.
Scanning electrochemical microscopy, part 13. Evaluation of the tip shapes of nanometer size microelectrodes.
J. Electroanal. Chem.
328:47-62.
-
Mow, V. C.,
M. H. Holmes, and W. M. Lai.
1984.
Fluid transport and mechanical properties of articular cartilage: a review.
J. Biomech.
17:377-394[Medline].
-
Muehleman, C., and C. H. Arsenis.
1995.
Articular cartilage, part II: the osteoarthritic joint.
J. Am. Podiatr. Med. Assoc.
85:282-286[Abstract].
-
Potter, K.,
R. G. S. Spencer, and E. W. McFarland.
1997.
Magnetic resonance microscopy studies of cation diffusion in cartilage.
Biochim. Biophys. Acta.
1334:129-139[Medline].
-
Roberts, S.,
J. P. G. Urban,
H. Evans, and S. M. Eisenstein.
1996.
Transport properties of the human cartilage endplate in relation to its composition and calcification.
Spine.
21:415-420[Medline].
-
Saito, Y.
1968.
Theoretical study on the diffusion current at the stationary electrodes of circular and narrow band types.
Rev. Polarogr.
15:177-187.
-
Scott, E. R.,
A. I. Laplaza,
H. S. White, and J. B. Phipps.
1993a.
Transport of ionic species in skin
contribution of pores to the overall skin conductance.
Pharm. Res.
10:1699-1709[Medline]. -
Scott, E. R.,
J. B. Phipps, and H. S. White.
1995.
Direct imaging of molecular transport through skin.
J. Invest. Dermatol.
104:142-145[Abstract].
-
Scott, E. R.,
H. S. White, and J. B. Phipps.
1991.
Scanning electrochemical microscopy of a porous membrane.
J. Membr. Sci.
58:71-87.
-
Scott, E. R.,
H. S. White, and J. B. Phipps.
1993b.
Ionophoretic transport through porous membranes using scanning electrochemical microscopy
application to in vitro studies of ion fluxes through skin.
Anal. Chem.
65:1537-1545[Medline]. -
Shrive, N. G., and C. B. Frank.
1994.
Articular cartilage.
In
Biomechanics of the Musculo-Skeletal System.
B. Nigg, and
W. Herzog, editors. J. Wiley and Sons, Chichester. 79-105.
-
Stockwell, R. A.
1983.
Metabolism of cartilage.
In
Cartilage, Vol. 1. Structure, Function and Biochemistry.
B. K. Hall, editor. Academic Press, London. 253-280.
-
Torzilli, P. A.,
J. M. Arduino,
J. D. Gregory, and M. Bansal.
1997.
Effect of proteoglycan removal on solute mobility in articular cartilage.
J. Biomech.
30:895-902[Medline].
-
Torzilli, P. A.,
D. A. Grande, and J. M. Arduino.
1998.
Diffusive properties of immature articular cartilage.
J. Biomed. Mater. Res.
40:132-138[Medline].
-
Unwin, P. R.,
J. V. Macpherson,
M. A. Beeston,
N. J. Evans,
N. P. Hughes, and D. L. Littlewood.
1997.
New electrochemical techniques for probing phase transfer dynamics at dental interfaces in vitro.
Adv. Dent. Res.
11:548-559[Medline].
-
Urban, J., and A. Hall.
1992.
Physical modifiers of cartilage metabolism.
In
Articular Cartilage and Osteoarthritis.
K. E. Keuttner,
R. Schleyerbach,
J. G. Peyron, and
V. C. Hascall, editors. Raven Press, New York. 393-406.
-
Weinberg, P. D.,
S. L. Carney,
C. P. Winlove, and K. H. Parker.
1997.
The contributions of glycosaminoglycans, collagen and other interstitial components to the hydraulic resistivity of porcine aortic wall.
Connect. Tissue Res.
36:297-308[Medline].
Biophys J, March 2000, p. 1578-1588, Vol. 78, No. 3
© 2000 by the Biophysical Society 0006-3495/00/03/1578/11 $2.00