| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, April 2000, p. 1703-1713, Vol. 78, No. 4

*EVOTEC BioSystems AG, D-22525 Hamburg, Germany, and
Institute of Experimental Biology, Harku 76902, Estonia
| |
ABSTRACT |
|---|
|
|
|---|
A method of sample analysis is presented which is based on fitting a joint distribution of photon count numbers. In experiments, fluorescence from a microscopic volume containing a fluctuating number of molecules is monitored by two detectors, using a confocal microscope. The two detectors may have different polarizational or spectral responses. Concentrations of fluorescent species together with two specific brightness values per species are determined. The two-dimensional fluorescence intensity distribution analysis (2D-FIDA), if used with a polarization cube, is a tool that is able to distinguish fluorescent species with different specific polarization ratios. As an example of polarization studies by 2D-FIDA, binding of 5'-(6-carboxytetramethylrhodamine) (TAMRA)-labeled theophylline to an anti-theophylline antibody has been studied. Alternatively, if two-color equipment is used, 2D-FIDA can determine concentrations and specific brightness values of fluorescent species corresponding to individual labels alone and their complex. As an example of two-color 2D-FIDA, binding of TAMRA-labeled somatostatin-14 to the human type-2 high-affinity somatostatin receptors present in stained vesicles has been studied. The presented method is unusually accurate among fluorescence fluctuation methods. It is well suited for monitoring a variety of molecular interactions, including receptors and ligands or antibodies and antigens.
| |
INTRODUCTION |
|---|
|
|
|---|
Fluorescence intensity distribution analysis
(FIDA; Kask et al., 1999
; independently Chen et al., 1999
) has been
recently developed as a complement to fluorescence correlation
spectroscopy (FCS; Magde et al., 1972
). The general idea of FIDA was
introduced in 1990 (Qian and Elson, 1990a
, b
), but then realized as a
less powerful method of moment analysis (MAFID). As in most common versions of FCS, in first realizations of MAFID and FIDA, a single detector was used to monitor fluorescence from a microscopic sample volume. However, fluorescence fluctuation spectroscopy does not have to
be restricted to the use of a single detector. For different purposes,
two detectors have been used in some applications of FCS. In the
nanosecond time domain, the two-detector system has been applied simply
because the dead time of the detector would otherwise not permit
correlation studies on such a short time scale (Kask et al., 1985
). In
rotational correlation studies, two detectors have been used for
monitoring light of different polarization, which helps one to
distinguish fluctuations of light intensity due to rotational motion of
particles from intensity fluctuations of other origin (Kask et al.,
1989
). Furthermore, in some cross-correlation studies two detectors
monitor light of different colors originating from different
fluorophores, enabling one to estimate to what extent two labeled
molecules are bound to each other (Schwille et al., 1997
). The present
paper introduces the theory of FIDA with two detectors. The
two-dimensional FIDA (2D-FIDA), if used with a polarization cube, is a
tool that can distinguish fluorescent species with different specific
polarization ratios. Alternatively, if two-color equipment is used, the
2D-FIDA can determine concentrations and specific brightness values of fluorescent species corresponding to individual labels as well as their complex.
The presented method is well suited for monitoring molecular interactions including receptors and ligands or antibodies and antigens, which are both of great relevance in the life sciences. The method is extremely accurate if the assay of interest can be designed with a significant contrast in specific brightness between bound and unbound states. Furthermore, it is characterized by single molecule sensitivity, the ability to resolve different species and determine their absolute concentrations, and to detect coincidences of different molecules in time and space. It is a homogeneous method, i.e., washing steps are not required. It is fast, well miniaturizable, and as a confocal technique, insensitive to surface adsorption. Therefore, 2D-FIDA is expected to find a wide variety of applications. Our pharmaceutical applications already cover the scale from detailed biochemical assay development to primary drug screening.
| |
THEORY |
|---|
|
|
|---|
The expected joint distribution of photon count numbers
To elaborate a simple theory that enables one to express the expected two-dimensional distribution of the number of photon counts, it is favorable to use the same assumptions used in one-dimensional FIDA and its predecessor, the moment analysis of photon count number distribution. The assumptions are as follows. 1) Coordinates of particles are random and independent of each other; 2) contribution to fluorescence intensity from a particle can be expressed as a product of a specific brightness of the particle and a spatial brightness profile function characteristic of the optical equipment; and 3) a short counting time interval T is selected, during which the brightness of fluorescent particles does not significantly change due to translational diffusion.
We shall first express a joint distribution of count numbers from a
single fluorescent species and a single small open volume element
dV. The latter is a small fraction of the microscopic observable volume, where the spatial brightness is considered to be
constant. Let us characterize the volume element by coordinates r and spatial brightness B(r), and the
fluorescent species by its specific brightness values
q1 and q2. By
q1 and q2 we have denoted
the mean photon count rates by two detectors from a particle situated
at a point where B(r) = 1. The spatial brightness
function describes the varying excitation and detection conditions
across the observed volume. A convenient choice is to select a unit of
B, as usual in FCS, by the equation
1 =
2, where
k =
Bk
(r)d3r. If the volume element
contains m particles, then the expected mean photon count
numbers per time interval T from the volume element are
mq1TB(r) and
mq2TB(r), while the joint distribution of
the numbers of photon counts from m particles is Poissonian
for both detectors independently:
|
(1) |
|
(2) |
|
(3) |
|
|
(4) |
|
(5) |
|
|
= ei
, then the distribution P(n2,
n2) and its generating function
G(
1,
2) are interrelated by a
two-dimensional Fourier transform. What makes the generating function
attractive in photon count number distribution analysis is the
additivity of its logarithm: logarithms of generating functions of
photon count number distributions of independent sources, like different volume elements and different species, are simply added for
the calculation of the combined distribution. Therefore, the generating
function of the overall distribution of the number of photon counts can
be expressed in a closed form:
|
(6) |
1 by detector 1 and
2 by
detector 2, and contributions from different fluorescent species,
denoted by the subscript i. Numeric integration according to
Eq. 6 followed by a fast Fourier transform is to our knowledge the most
efficient means to calculate the theoretical distribution
P(n1, n2) of a given sample (i.e.,
given concentrations and specific brightness values of
fluorescent species).
The spatial brightness function
The spatial brightness function is accounted for through the
spatial integration on the right side of Eq. 6. In the same way as in
1D-FIDA, the three-dimensional integration can be reduced to one
dimension by replacing the three-dimensional coordinates r
by a one-dimensional variable, a monotonic function of the spatial
brightness B(r). A convenient choice of the variable
is x = ln[B(0)/B(r)]. A sufficiently
flexible model of the one-dimensional spatial brightness profile is
presented in the following expression:
|
(7) |

would
yield.) Empirical values of the constants a1 and
a2 of our equipment are in the vicinity of
a1
0.4 and a2
0.08. However, the values of a1 and
a2 are highly correlated, and therefore it is
recommended to present at least one of them with a higher than the
nominal accuracy of their determination (which is typically 5-10%).
Weights, statistical errors, and data simulation algorithm
In the interval of obtained count numbers, the probability of
obtaining a particular pair of count numbers usually varies by many
orders of magnitude. Consequently, the variance of the experimental
distribution also has a strong dependence on the count numbers. To
determine weights for least-squares fitting, one can assume for
simplification that coordinates of particles in all counting intervals
are randomly selected. (Thus we ignore correlations of the coordinates
in consecutive counting intervals.) In this assumption, we have a
problem with distributing M events over choices of different
pairs of count numbers n1, n2, each particular outcome having a given probability of realization, P(n1, n2). Covariance matrix
elements of the distribution can be expressed as follows:
|
(8) |
|
For a further simplification, one may ignore the second term on the
right side of Eq. 8, which can be interpreted as a consequence of
normalization. In this case, the weights are simply equal to the
inverse values of the diagonal covariance matrix elements
|
(9) |
2 are usually close to unity.
However, we have verified that the error values of estimated parameters
returned by the linearized least-squares fitting algorithm are
underestimated. The factor of underestimation is typically in the range
of 1.5 to 4, depending most significantly on the ratio of the mean
translational diffusion time of the molecule to the width of the
counting time interval. The reason for the error underestimation is
definitely the simplifying assumption behind Eq. 9, which ignores
correlations between the photon count numbers measured in consecutive
counting time intervals. As a reliable method of error determination,
we have used the following algorithm. It involves, first, the
simulation of a series of at least 30 random distributions at identical
conditions; second, fitting them; and finally, the determination of
statistical errors from scattered values of estimated parameters. The
algorithm of data simulation does not ignore correlations between the
photon count numbers of consecutive intervals, but includes random walk simulation of individual molecules. We have used this algorithm of
error determination whenever error values are of special interest, despite its clumsiness and slowness. The error values presented in this
paper are determined by this method. In other applications (in
particular if the speed of analysis is of a higher interest than the
exact error values), the error values corresponding to weights by Eq. 9
are often used multiplied by a roughly determined factor of underestimation.
Accuracy of 2D-FIDA versus 1D-FIDA
In Table 1 theoretical errors of FIDA and 2D-FIDA simulations are presented in two selected cases for two fluorescent species. In both cases the ratio of specific brightness values of the two species is three. Throughout the simulations, a data collection time of 8 s, a counting time interval of 40 µs, a diffusion time of 400 µs for both species, a speed of scanning (or flow) of 11000 1/e2-radius values per second, and a background count rate of 1 kHz were assumed. Note that the statistical error values of the estimated parameters are significantly lower in the 2D-FIDA example. This result is related to the design principle of the two color experiments recommending that the spectral sensitivities of the two detectors (A and B) are tuned to different species. The values of parameters in the 2D-FIDA example are selected symmetrically. The error values would also be equal in pairs if the number of realizations was significantly higher than 30.
|
Two-dimensional moment analysis of fluorescence intensity fluctuations
In the same manner as one-dimensional FIDA can be generalized to
the two-dimensional case, it can be done with MAFID. Factorial moments
of the distribution P(n1, n2) are
defined as
|
(10) |
|
(11) |
|
(12) |
1 =
2,
1 has the
meaning of the sample volume, denoted by V, and Eq. 12 can be written as
|
(13) |
denotes a series of constants characterizing the
brightness profile:
|
(14) |
In Table 2 statistical errors of 2D-FIDA and 2D-MAFID are presented as determined by generating a series of 30 random distributions of count numbers, simulated for identical "samples," thereafter applying 2D-FIDA and 2D-MAFID and determining the variance of estimated parameters in both cases. It is evident that the advantages of 2D-FIDA compared to 2D-MAFID increase with the number of parameters to be estimated.
|
In Table 3 the relative deviation of mean values (i.e., bias) of estimated parameters are presented for 2D-FIDA and 2D-MAFID. In each case bias is determined from analysis of a series of 30 simulated random distributions of count numbers. Three cases were analyzed. In the first case, models used in data simulations and data analysis were identical. In the second case, the distributions of count numbers were simulated assuming that particles of the second species are not equivalent, but are distributed by their individual brightness with a relative half-width of 20%. However, we intentionally ignored this phenomenon in analysis. Of course, applying a slightly inadequate model for analysis produces bias of estimated parameters. The third case is similar to the second one except that the relative half-width of the individual brightness distribution of the second species is 50%, which is a usual value for vesicular preparations. It is worth noting that methodological deviations are noticeable when mapping weighted residuals of 2D-FIDA in cases two and three. However, 2D-FIDA still returns meaningful results, while 2D-MAFID is much more sensitive to model deviations.
|
| |
EXPERIMENTS |
|---|
|
|
|---|
Equipment
The central optical part of a 2D-FIDA experiment is a confocal
microscope as it is used in fluorescence correlation spectroscopy (Koppel et al., 1976
). For excitation of fluorescence, a beam from a
continuous-wave laser is attenuated by neutral filters, passes a beam
expander, and is directed to the microscope objective by a dichroic
mirror. In a number of experiments with slowly diffusing particles,
e.g., vesicles, beam scanning in combination with sample scanning is
used, as a tool known from laser scanning microscopy. Fluorescence is
collected by the same objective through the dichroic mirror and is
focused to a confocal pinhole, which serves to reject the out-of-focus
light. The light, which passes the pinhole, is divided by a
beamsplitter for detection by two detectors. Depending on the general
type of a 2D-FIDA experiment, the beamsplitter is either a polarization
cube or a dichroic mirror. In the first case, a common spectral
band-pass filter is used, while in the case of two-color FIDA, each
detector has a different band-pass filter positioned in front of it
(Fig. 1). The photon counting detectors are silicon avalanche photodiode modules (SPCM-AQ-131, EG&G
Optoelectronics, Vaudreuil, Quebec, Canada). The TTL pulses from the
detectors are collected continuously by a two-channel counter
constructed at EVOTEC as a computer plug-in card that calculates the
count number distributions in real time from the 32 MB onboard buffer.
By feeding the detector outputs to a correlator, FCS measurements can
be performed in parallel with FIDA experiments.
|
The levels of background count rate for both detectors are determined in a separate experiment on bidistilled water. The main contributor to the nonfluctuating background light intensity is Raman scattering from water.
The radius of the monitored sample volume can be adjusted by selecting
an appropriate expansion factor of the original laser beam. The focal
beam radius of ~0.6 µm is used, yielding diffusion times for simple
organic dye molecules (e.g., 5'-(6-carboxytetramethylrhodamine (TAMRA))
of ~260 µs, which is considered long compared to the 40 µs dwell
time of counters, so that the assumption of constant molecular
brightness during the counting interval is well-founded. The excitation
intensity is adjusted as a compromise between a high count rate per
molecule and low population of the triplet state. In our experiments we
have kept the triplet state population below 15% because higher
triplet population values might significantly distort the apparent
spatial brightness profile (Kask et al., 1999
).
Test experiments
For methodological test experiments two different dyes were
selected, TAMRA and rhodamine red X (RRX). These dyes have different emission spectra and different extinction coefficients at the excitation wavelength of 543.5 nm. In these experiments, a wideband 40/60 beamsplitter was used in front of the detectors. The spectral filter of the "red" channel has the central wavelength of 605 nm
and FWHM of 50 nm, while the corresponding figures for the "yellow-green" channel are 575 nm and 30 nm. The dyes were diluted in distilled water so that the average number of molecules in the
observation volume was in the range of 0.5-2.0, corresponding to
concentrations between 0.23 and 0.92 nM. For each experiment ~20 µl
of the sample solution was placed on a coverslip separating the sample
from the water immersion objective (Zeiss C-Apochromat 40 × 1.2 W
Korr). All the dyes were measured separately for 60 s, as were
their mixtures with two different concentration ratios (~1:1 and
1:2.5). The parameters describing the spatial brightness profile were
determined from adjustment experiments on TAMRA and were fixed for the
subsequent analysis of other samples at values a1 =
0.405 and a2 = 0.0772.
Results of test experiments
As an example, Fig. 2 visualizes a joint count number distribution of an ~0.5 nM TAMRA solution. The results of the above-described test experiments are summarized for multiple realizations in Table 4. We have not specified samples by concentration values calculated from dilution factors of the preparation because adsorption of dye molecules to glass surfaces may influence the real concentrations in the sample volume. Therefore, the determined concentration values vary slightly from realization to realization. However, the specific brightness values are well-reproduced throughout all experiments.
|
|
It is worth noting that distributions measured with mixtures are qualitatively different from those measured for pure dyes. Fig. 3 visualizes weighted residuals of fitting a distribution measured with a mixture of TAMRA and RRX. The top graph (A) corresponds to the adequate analysis when two species were assumed to be present; residuals are scattered quite randomly and uniformly. The bottom graph (B) corresponds to the assumption that only a single species is present. Here a significant difference between the measured and the calculated distribution is evident.
|
| |
APPLICATIONS |
|---|
|
|
|---|
In the following section we shall present in detail two essentially different examples of 2D-FIDA from biochemical assay development illustrating the broad applications of the method. As outlined in the introduction of this paper it is essential that the two detectors monitor different qualities of fluorescent species. Therefore, in the first example the two detectors monitor the two polarization components of fluorescence, while in the second example the two detectors have different spectral responses, thus monitoring different labels.
Fluorescence polarization studies
In conventional fluorescence polarization studies, average
intensities of two polarization components of fluorescence are directly
measured. Fluctuations of the intensities are not of direct interest,
but are considered rather as a source of statistical errors. The
conventional fluorescence polarization method has proven to be a
powerful tool in the study of molecular interactions (Checovich et al.,
1995
; Jameson and Sawyer, 1995
; Jolley, 1996
). Changes in the
fluorescence polarization values of a sample containing a fluorescently
labeled binding partner reflect changes in molecular volume and, hence,
provide direct information on equilibrium binding. Fluorescence
polarization measurements can also be performed in real time, allowing
the kinetic analysis of association and dissociation reactions. One of
the most widely used fluorescence polarization applications is the
competitive immunoassay used for the detection of therapeutic and
illicit drugs. The method of fluorescence polarization has been used
for clinical immunoassays for more than a decade (Jolley, 1981
). The
homogeneous FPIA (fluorescence polarization in immunoassays) has
well-accepted advantages over conventional heterogeneous immunoassays
like RIA or ELISA. However, it fails if multiple-binding step reactions
are to be investigated because the separation of individually polarized
species is impossible. Therefore, ligand-binding curves demonstrate
only the overall decrease of polarization, meaning that the mechanistic
binding constants cannot be determined. Further limitations are seen in sample volume and in mass restrictions.
With 2D-FIDA the full content of information usually buried in fluorescence anisotropy can be used, thereby overcoming the limitations mentioned above. 2D-FIDA directly determines two specific quantities per fluorescent species in one measurement: the fluorescence intensity per molecule and the anisotropy of a given model. Based on this supplementary information, the delineation of all participating species and even the quantification of the binding behavior are possible. 2D-FIDA anisotropy is an ideal tool for the quantitative description of systems exhibiting multiple binding steps, aggregation, and multimerization phenomena as demonstrated with the following example.
The theophylline/anti-theophylline antibody interaction
Theophylline therapy has been a cornerstone of asthma therapy for several years and, therefore, there is a strong demand for assaying and fine-tuning the theophylline level in serum (Hinds et al., 1984
|
|
Two-color approach of 2D-FIDA for vesicle-based binding assays
For the two-color approach of 2D-FIDA, the two detectors are spectrally tuned to monitor fluorescence from two labels of different color. In the assay type described below, ligand molecules are labeled in "green." Because each vesicle carries a large number of receptors, vesicles in samples with a low binding degree can be distinguished from vesicles in samples with a high binding degree by a significantly higher specific brightness in "green." Vesicles are additionally stained in "red." Specific brightness of vesicles in "red" is not altered by binding of ligand molecules, but staining in "red" is a means to increase contrast between free ligand molecules (which are nearly invisible in "red") and vesicles (which in the case of extremely low binding may be of nearly the same brightness in "green" as free ligand molecules). Contributions from the two fluorescent species of a single sample to the measured joint distribution of the numbers of photon counts are very different in this assay type indeed, and therefore the analysis is extremely reliable.
Binding of somatostatin-14 to the human type-2 high-affinity somatostatin receptor
To demonstrate the advantages of 2D-FIDA the binding of TAMRA-labeled somatostatin-14 (SMS-14-5TAMRA) to small membrane vesicles carrying the human type-2 high-affinity somatostatin receptor SSTR-2 (Schoeffter et al., 1995
|
|
|
| |
CONCLUSIONS |
|---|
|
|
|---|
In this paper we extended the theory of fluorescence intensity distribution analysis (FIDA) to two dimensions. By comparing theoretical with experimental data we could prove that the collected joint photon count number distributions are in agreement with the theoretically predicted ones. The introduction of the generating function facilitates data evaluation and makes the method applicable even in high throughput drug screening, where it has already been successfully applied (EVOTEC/Novartis collaboration).
Compared to many other fluctuation methods, which often suffer from low precision, the presented method can be extremely accurate. In fact, the accuracy depends on the skill of assay design; a very important nuance is that fluorescence from different species was split at different intensity ratios. Only in the case of the worst design, when fluorescence from different species is split between the two detectors at the same ratio, is the resolving power of 2D-FIDA equal to that of 1D-FIDA. The higher the contrast between species in the intensity ratio, the more one can gain from using 2D-FIDA.
2D-FIDA has proven to be a method of versatile applicability. It offers
new insights into polarization studies. Compared to a number of other
polarization methods, it is superior due to its ability to separate
different components and determine their absolute concentrations. Among
different fluorescence methods we have compared with 2D-FIDA, only the
polarization analysis by burst integrated fluorescence lifetime (BIFL;
Schaffer et al., 1999
) has both of these properties. However, BIFL is
restricted to significantly lower concentrations than 2D-FIDA.
The other branch of 2D-FIDA, the two-color analysis, offers the possibility to work under near-to-ideality conditions because here assays can be designed or selected with a pronounced contrast between the bound and non-bound states. The closest method to two-color 2D-FIDA is its counterpart in FCS, the cross-correlation method. Between them, 2D-FIDA seems to be a more favorable method again, because FIDA is directly focused to separate the absolute concentrations of different components, whereas in FCS only the product of the concentration and the square of the specific brightness of each component can be directly separated.
| |
ACKNOWLEDGMENTS |
|---|
The authors gratefully acknowledge NOVARTIS PHARMA AG for providing the SMS model. Dr. Nicholas Hunt is acknowledged for critically reading the manuscript.
| |
FOOTNOTES |
|---|
Received for publication 8 September 1999 and in final form 12 January 2000.
Address reprint requests to Dr. Karsten Gall, EVOTEC BioSystems AG, Schnackenburgallee 114, D-22525 Hamburg, Germany. Tel.: 49-40-560-81-0; Fax: 49-40-560-81-222; E-mail: gall{at}evotec.de.
| |
REFERENCES |
|---|
|
|
|---|
a new tool for cell and molecular biology.
Nature.
375:254-256[Medline].
measurement by fluorescence correlation spectroscopy.
Phys. Rev. Lett.
29:704-708.
identity with SSTR-2 receptors.
Eur. J. Pharmacol.
289:163-173[Medline].
Biophys J, April 2000, p. 1703-1713, Vol. 78, No. 4
© 2000 by the Biophysical Society 0006-3495/00/04/1703/11 $2.00
This article has been cited by other articles:
![]() |
S. Saffarian, Y. Li, E. L. Elson, and L. J. Pike Oligomerization of the EGF Receptor Investigated by Live Cell Fluorescence Intensity Distribution Analysis Biophys. J., August 1, 2007; 93(3): 1021 - 1031. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. N. Hillesheim, Y. Chen, and J. D. Muller Dual-Color Photon Counting Histogram Analysis of mRFP1 and EGFP in Living Cells Biophys. J., December 1, 2006; 91(11): 4273 - 4284. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Moger, P. Gribbon, A. Sewing, and C. P. Winlove The Application of Fluorescence Lifetime Readouts in High-Throughput Screening J Biomol Screen, October 1, 2006; 11(7): 765 - 772. [Abstract] [PDF] |
||||
![]() |
B. Wu, Y. Chen, and J. D. Muller Dual-Color Time-Integrated Fluorescence Cumulant Analysis Biophys. J., October 1, 2006; 91(7): 2687 - 2698. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Palo, U. Mets, V. Loorits, and P. Kask Calculation of Photon-Count Number Distributions via Master Equations Biophys. J., March 15, 2006; 90(6): 2179 - 2191. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. N. Hillesheim and J. D. Muller The Dual-Color Photon Counting Histogram with Non-Ideal Photodetectors Biophys. J., November 1, 2005; 89(5): 3491 - 3507. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Eggeling, P. Kask, D. Winkler, and S. Jager Rapid Analysis of Forster Resonance Energy Transfer by Two-Color Global Fluorescence Correlation Spectroscopy: Trypsin Proteinase Reaction Biophys. J., July 1, 2005; 89(1): 605 - 618. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Chen, M. Tekmen, L. Hillesheim, J. Skinner, B. Wu, and J. D. Muller Dual-Color Photon-Counting Histogram Biophys. J., March 1, 2005; 88(3): 2177 - 2192. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. McHale, A. J. Berglund, and H. Mabuchi Bayesian Estimation for Species Identification in Single-Molecule Fluorescence Microscopy Biophys. J., June 1, 2004; 86(6): 3409 - 3422. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Jager, N. Garbow, A. Kirsch, H. Preckel, F. U. Gandenberger, K. Herrenknecht, M. Rudiger, J. P. Hutchinson, R. P. Bingham, F. Ramon, et al. A Modular, Fully Integrated Ultra-High-Throughput Screening System Based on Confocal Fluorescence Analysis Techniques J Biomol Screen, December 1, 2003; 8(6): 648 - 659. [Abstract] [PDF] |
||||
![]() |
P. J. Rothwell, S. Berger, O. Kensch, S. Felekyan, M. Antonik, B. M. Wohrl, T. Restle, R. S. Goody, and C. A. M. Seidel Multiparameter single-molecule fluorescence spectroscopy reveals heterogeneity of HIV-1 reverse transcriptase:primer/template complexes PNAS, February 18, 2003; 100(4): 1655 - 1660. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Haupts, M. Rudiger, S. Ashman, S. Turconi, R. Bingham, C. Wharton, J. Hutchinson, C. Carey, K. J. Moore, and A. J. Pope Single-Molecule Detection Technologies in Miniaturized High-Throughput Screening: Fluorescence Intensity Distribution Analysis J Biomol Screen, February 1, 2003; 8(1): 19 - 33. [Abstract] [PDF] |
||||