| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, April 2000, p. 2127-2137, Vol. 78, No. 4
Cell Biophysics Laboratory, Imperial Cancer Research Fund, London WC2A 3PX, England
| |
ABSTRACT |
|---|
|
|
|---|
Global analysis techniques are described for frequency domain fluorescence lifetime imaging microscopy (FLIM) data. These algorithms exploit the prior knowledge that only a limited number of fluorescent molecule species whose lifetimes do not vary spatially are present in the sample. Two approaches to implementing the lifetime invariance constraint are described. In the lifetime invariant fit method, each image in the lifetime image sequence is spatially averaged to obtain an improved signal-to-noise ratio. The lifetime estimations from these averaged data are used to recover the fractional contribution to the steady-state fluorescence on a pixel-by-pixel basis for each species. The second, superior, approach uses a global analysis technique that simultaneously fits the fractional contributions in all pixels and the spatially invariant lifetimes. In frequency domain FLIM the maximum number of lifetimes that can be fit with the global analysis method is twice the number of lifetimes that can be fit with conventional approaches. As a result, it is possible to discern two lifetimes with a single-frequency FLIM setup. The algorithms were tested on simulated data and then applied to separate the cellular distributions of coexpressed green fluorescent proteins in living cells.
| |
INTRODUCTION |
|---|
|
|
|---|
Fluorescence lifetime imaging microscopy (FLIM)
allows the lifetimes of one or more fluorophores to be spatially
resolved and can be used to provide information about the state of the fluorescent species and their immediate molecular environment. The
fluorescence lifetime is sensitive to environmental conditions such as
pH, and excited state reactions such as fluorescence resonance energy
transfer (FRET) or collisional quenching, properties that have been
exploited for resolving physiological parameters in the cell (Lakowicz
et al., 1992
, 1994
; Gadella et al., 1993
, 1999
; Gadella and Jovin,
1995
; Sanders et al., 1995
; Szmancinski and Lakowicz, 1995
; Bastiaens
and Jovin, 1996
; French et al., 1997
; Bastiaens and Squire, 1999
; Ng et
al., 1999
; Pepperkok et al., 1999
). Fluorescence lifetime imaging in a
wide-field or confocal microscope has been implemented, using
time-domain and frequency-domain techniques (Lakowicz and Berndt, 1991
;
Buurman et al., 1992
; Clegg et al., 1992
; Gadella et al., 1994
; Dong et
al., 1995
; Draaijer et al., 1995
; So et al., 1995
; Clegg and Schneider,
1996
; Periasamy et al., 1996
; Carlsson and Liljeborg, 1997
; Schneider
and Clegg, 1997
; Sytsma et al., 1998
; Squire and Bastiaens, 1999
).
Until recently, fluorescence lifetime imaging has been limited to
single lifetime estimates at each pixel of the sample image. However,
in many cases a mixture of different molecular species or molecules in
different biochemical states is present in each resolvable image
element of the sample. As a result, the time-resolved response is a
complex decay that is not described accurately by a single exponential.
In such cases, a single lifetime estimation is only a semiquantitative
measurement of the decay kinetics of the sample. Instead, the lifetimes
and populations of each species should be resolved. For time domain
measurements, this can be achieved by sampling the response to a light
pulse and fitting a multiexponential function to the measurements. This
approach was implemented in a microscope by Scully et al. (1996
, 1997
). In the frequency domain, the phase shifts and modulations must be
recorded at multiple frequencies and the parameters of a
multiexponential decay can then be derived using a nonlinear analysis
algorithm (Gratton et al., 1984
; Lakowicz et al., 1984
). The latter
technique was recently implemented by us in a microscope; we call this
multiple-frequency fluorescence lifetime imaging microscopy (mfFLIM)
(Squire et al., 1999
). The mfFLIM technique opens the possibility of
resolving a complex decay function and determining the lifetimes and
populations of a number of species in each pixel. The signal-to-noise
ratio of the data, however, is critical for obtaining a reliable
result. Improving the signal-to-noise ratio is especially difficult for data acquired with fluorescence microscopy, because the exposure time
is limited by factors such as photobleaching and, in the case of living
cells, imaging speed. Similar limitations arise with the use of time
domain techniques.
In this paper these problems are solved by applying a global analysis
method to frequency domain FLIM image sequences in a fashion similar to
techniques used for the analysis of fluorescence measurements in
cuvettes (Knutson et al., 1983
; Beechem et al., 1983
, 1985
; Beechem and
Brand, 1986
; Beechem, 1992
; Gratton et al., 1984
; Lako-wicz et al.,
1984
). There, the results of multiple experiments are analyzed
simultaneously, assuming that some of the parameters are identical for
each experiment, while the remaining parameters may differ per
experiment. The classical example is the resolution of a mixture of two
fluorescent species by lifetimes. The results of lifetime measurements
are analyzed in terms of a biexponential model, resulting in the
lifetimes and the fractional contributions to the steady-state
fluorescence. This experiment can be repeated at different wavelengths,
yielding data sets with different fractions, but the same lifetime
values. It can be shown that simultaneous analysis of all data sets,
using the knowledge that the lifetimes are invariant, yields vastly
superior results compared to separate analysis of each set. The same
lifetime invariance principle can be applied to the analysis of
fluorescence lifetime images. Assuming that there is a given number of
molecular species with different lifetimes, it is reasonable to assume
that these lifetimes do not vary spatially. Thus, instead of analyzing
each pixel separately, we analyze all pixels simultaneously, yielding images of the populations of each molecular species and the discrete lifetime values. In this paper we show that this approach leads to a
dramatic improvement in the accuracy and precision of FLIM data
analysis. In addition, we show that global analysis allows two
lifetimes to be resolved by the use of a single-frequency measurement,
which is fundamentally impossible by conventional data analysis
techniques. We test the proposed analysis techniques on simulated data
and use them to separate the fluorescence distributions of coexpressed
green fluorescent proteins in living cells.
| |
THEORY |
|---|
|
|
|---|
Fluorescence lifetime measurements in the frequency domain
The theory of fluorescence lifetime measurements in the frequency
domain has been described extensively (Clegg et al., 1992
; Gratton and
Limkeman, 1983
; Gratton et al., 1984
; Lakowicz et al., 1984
; Clegg and
Schneider, 1996
; Gadella et al., 1994
; Squire and Bastiaens, 1999
).
Here we repeat the most important results. The fluorescence response
R(t) of a sample to impulse excitation is a sum of
Q exponential decays:
|
(1) |
q and
q are the fractional
contributions to the steady-state fluorescence and the lifetime from
the qth emitting species, respectively. From the fractional
contributions the relative populations aq of
each species can be found by dividing by the quantum yield of the
fluorophore, corrected for the instrument response, and renormalizing
the sum of the populations to one. If the spectra of all species are
the same, this is equivalent to dividing by the lifetimes and
renormalizing. Multiplying the populations aq by
the steady-state fluorescence yields the concentration of each species
up to a constant multiplier.
In the frequency domain approach to FLIM the sample is illuminated by a
periodically modulated excitation field that can be represented by a
Fourier series:
|
(2) |
is the fundamental circular frequency of the periodic
excitation signal, N is the number of harmonics that are
present, E0 is the average intensity, and
En and
E,n are the modulation and
the phase of the nth harmonic. The fluorescence emitted by the sample is proportional to the convolution of the excitation light
E(t) with the response function R(t):
|
(3) |
|

n and modulation
Mn are given by
|
(4a) |
|
(4b) |
|
(5a) |
|
(5b) |
q and lifetimes
q, the modulations Mn and phase
shifts 
n must be measured for a number of frequencies
and fitted to Eqs. 4 and 5.
To measure the modulations Mn and phase shifts

n in a microscope, heterodyne or homodyne phase
detection techniques can be employed, using high-speed image
intensifier devices (Gadella et al., 1993
; Kume et al., 1988
). These
devices convert an image on the photocathode surface into electrons
across the faces of a multichannel plate (MCP). These are then
converted back to photons after they hit a photoluminescence surface. A
repetitive high-frequency voltage modulation across either the MCP or
the photocathode results in high-frequency modulation of the gain of
the device. This makes it possible to modulate the detected
fluorescence image at a frequency that is close (heterodyne detection)
or equal (homodyne detection) to that of the excitation light. The
resulting image only contains the low-frequency components of the
modulated image, because high frequencies are filtered out by the slow
response of the photoluminescence surface at the output of the intensifier.
The time-dependent gain of the modulated intensifier can be represented
by a Fourier series:
|
(6) |
' is the fundamental circular frequency of the gain
modulation. G0 is the average gain, and
Gm and
G,m are the amplitude and
the phase of the mth harmonic. The phase of G(t)
can be controlled by adjusting k
). The resulting image
is given by the low-frequency components of the result of multiplying
Eq. 3 by Eq. 6. For homodyne detection the output image can be written
as
|
(7) |
R,n =
G,n
E,n is the reference phase,
which must be obtained from an independent measurement on a sample with
a known lifetime. D0 is the average intensity,
which depends on factors such as photon detection efficiency, intensity
of excitation, and fluorophore quantum yield. D0
is proportional to the steady-state fluorescence from the sample. The
signal D(k) is time-independent and can be phase-sampled at
discrete phases k
. Equation 7 can be transformed to a
linear equation (Gadella et al., 1994
n (Squire and Bastiaens, 1999Given the modulations Mn and phase shifts

n, it is possible to compute the fractions
q and lifetimes
q, using Eqs. 4 and 5. For the case of only a single lifetime (Q = 1,
Q = 1), the lifetime value can be calculated
directly from either modulation or phase shift:
|
(8a) |
|
(8b) |


,n =
M,n, and the values
are independent of the frequency n
. Differences in the
lifetime estimations from phase shift and modulations indicate that the
decay kinetics of the sample are more complex.
In the case of a multiexponential decay model, a nonlinear fit to Eqs.
4 and 5 can be used to obtain the populations and lifetimes (Gratton et
al., 1984
; Lakowicz et al., 1984
; Squire et al., 1999
).
Global analysis of phase-modulation image data
The analysis that was outlined above can be applied on a pixel-by-pixel basis to a phase-dependent sequence of FLIM images. For microscope samples, the need to minimize photobleaching and acquisition time limits the number of frequencies at which measurements are taken. This can lead to a signal-to-noise ratio that is low, leading to inaccurate results. A solution to this problem is to reduce the number of variables that have to be estimated by employing prior knowledge about the sample.
First we note that the
q are fractional contributions to
the steady-state fluorescence and that their sum equals one. Therefore, we eliminate one variable at each pixel by substituting
Q = 1
q=1Q
1
q. This allows us to rewrite Eq. 5:
|
(9a) |
|
(9b) |
q, and if the
lifetime values are known, the populations can be found by a linear
least-squares method, for instance, a singular value decomposition.
An important form of prior knowledge is that in many applications the
lifetimes are spatially invariant. Each pixel in the image represents a
mixture of different fluorophores or molecules in different biochemical
states. This gives rise to a multiexponential decay whose lifetimes do
not vary spatially. One simple approach to making use of that knowledge
was implemented by us previously and termed the "lifetime invariant
fit" (Squire et al., 1999
). In this approach, the intensities of each
image in the FLIM sequence are averaged to a single value. This gives a
phase-dependent data series with an increased signal-to-noise ratio at
the expense of spatial information. The lifetimes of each fluorescent
species can then be obtained by estimating the phase shifts and
modulations from this averaged series, followed by a nonlinear
analysis, as described in the previous section. The fractional
contributions in each pixel can then be recovered by substituting these
lifetime values back into Eq. 9.
A second approach utilizes global analysis (Knutson et al., 1983
;
Beechem et al., 1983
, 1985
; Beechem and Brand, 1986
; Beechem, 1992
;
Gratton et al., 1984
; Lakowicz et al., 1984
), where all pixels are
analyzed simultaneously and the lifetimes are constrained to be the
same in each pixel. This is accomplished by nonlinear minimization of
the following
2 measure, summing over N
harmonic frequencies, and M pixels in the
image:
|
(10) |
|
n,m are calculated from the measured
phase shift and modulation in the mth pixel for the
nth frequency, using Eq. 4, and
(An,m) and
(Bn,m)
are estimates of their standard deviation, calculated by error
propagation from the errors in the phase shift and modulation. The
expressions for An,m (
q,m,
q) and Bn,m (
q,m,
q) are given by Eq. 9, where the subscript m
was added to indicate that the fractions differ at each pixel. The
parameters that are varied to minimize this
2 measure
are the Q lifetime values
q and the
M sets of Q
1 fractions
q,m. Minimization of Eq. 10 is a separable nonlinear least-squares problem that can be solved by dedicated algorithms (Golub
and Pereyra, 1973
n,m
images are calculated in a manner identical to that of a conventional
analysis, where lifetime estimates are calculated on a pixel-by-pixel
basis, using Eq. 8.
The two approaches described for implementing the spatial invariance
constraint are fundamentally different. The lifetime invariant fit
improves the signal-to-noise ratio to get an accurate estimation of the
lifetime values. With these values, an accurate estimation of the
fractions can be made. In contrast, the global fit makes use of the
spatial variation in fractions, information that is lost in lifetime
invariant fit because of the averaging operation. This difference has
important consequences for the number of lifetimes that can be
estimated for a given number of frequencies. For a successful fit, the
number of parameters to be estimated must be less than or equal to the
number of independent measurements. In a conventional fit, Q
lifetimes, with Q
1 populations, must be estimated
from the phase shifts and modulations for N frequencies.
Thus the following inequality must hold:
|
(11) |
1 fractions in each pixel, from N phase shifts and
modulations in M pixels. The requirement for a successful
fit then becomes
|
(12) |
| |
RESULTS |
|---|
|
|
|---|
Simulated data
The two different approaches for fitting FLIM data were tested on
simulated data. A biexponential decay model was assumed, with a long
lifetime of 2 ns, and a shorter lifetime, which, for different
simulations, was either fixed or varied. An image of 64 × 64 pixels was generated with fractions for the short lifetime species
according to the formula
1,m = m/M,
where m was the number of the pixel, counting in a
one-dimensional fashion from the upper left to the lower right corner
of the image, and M was the number of pixels. The fractions
for the longer lifetime were
2,m = 1
1,m. These fractions were the same in all simulations.
Multiple-frequency data with a fundamental frequency of 80 MHz and four
harmonics was generated using Eq. 7, with reference phases
R,1 = 0,
R,2 = 0.1,
R,3 = 0.2,
R,4 = 0.3, and
reference modulations MR,1 = 1, MR,2 = 0.8, MR,3 = 0.6, MR,4 = 0.4. In addition, single-frequency FLIM data at 80 MHz were generated with
R,1 = 1. In
both cases the number of phase sampling points was 32. Poisson noise
was then added to the phase-dependent images, where the average
intensity D0 was varied to control the
signal-to-noise ratio.
Fig. 1 shows the results of the different fit approaches for a simulated image, with the short lifetime equal to 1.5 ns and an average intensity over all phase images of D0 = 1000 counts at each pixel. Shown are images of the original and recovered fractions along with the estimated lifetimes. The top row shows the results for the multiple-frequency data. Both the lifetime invariant fit and the global fit give accurate results; however, the conventional pixel-by-pixel fit fails completely. The bottom row shows the result for the single-frequency data. The global fit was capable of recovering the lifetimes and fractions, whereas the conventional pixel-by-pixel and lifetime invariant fits fail to recover the correct values, as expected for an underdetermined fit.
|
Fig. 2 shows the estimated lifetime values as a function of the short lifetime value, where the long lifetime is fixed at 2 ns. The simulations were repeated 25 times with different realizations of a set amount of Poisson noise to compute average values and standard deviations. For the case of a conventional fit, the estimated values were the averages over the pixels, where the estimated value was between 0 ns and 2.5 ns. Clearly the conventional pixel-by-pixel fit performs very badly, except for large differences between the true lifetimes. The lifetime invariant fit performs much better as long as the true lifetime values are not too close. The global fit performs even better for lifetime values separated by as little as 0.2 ns. Fig. 3 shows the same results for data with a single harmonic. In this case the conventional fit and the lifetime invariant fit are not capable of estimating the correct values, whereas the global fit performs more or less as well as the global fit of the data with four harmonics. The results for the conventional and spatial invariant fits are not as bad as one might expect for these underdetermined fits and show a trend similar to that of the multifrequency results. Apparently, for this choice of lifetimes and frequency, the nonlinear least-squares minimization converges to a point that is not very far from the true values. However, the large standard deviations indicate that in general the results for the conventional and spatial invariant fits of single-frequency data cannot be expected to be correct.
|
|
To investigate the dependence on the signal-to-noise ratio of the data, simulations were performed, where the average intensity of the data was varied to control the amount of noise. Because the mean of the Poisson process is the intensity of the simulated phase image, an increased average of the phase-dependent data means an increase in detected photons and thus an increase in the signal-to-noise ratio. Fig. 4 shows the estimated lifetime values and standard deviations as a function of the average intensity for the different fit approaches using four harmonics. The true lifetime values were 1.5 ns and 2 ns. Average values and standard deviations were calculated from 25 repetitions of the simulation. In the case of the conventional fit the plotted lifetimes are the average of the pixels with estimated lifetime values between 0 ns and 2.5 ns. It can be seen that the conventional pixel-by-pixel fit fails. The global fit performs much better than the lifetime invariant fit, especially for low average intensity (low signal-to-noise ratio) values.
|
In all of these simulations the conventional pixel-by-pixel fit performs badly, indicating that for a typical FLIM measurement the signal-to-noise ratio is too low for an adequate pixel-by-pixel fit. For multiple-frequency data, the spatial invariant fit performs much better if the true lifetime values are well separated. The global fit is clearly superior to the two other methods and is the only approach that can be used to accurately fit a two-component model to single-frequency data.
Experimental data
The global analysis algorithm was used to disentangle the cellular
distributions of coexpressed green fluorescent proteins with different
lifetimes in living cells. Vero cells were plated on Matek Petri dishes
(Matek Corp.) in minimum essential medium supplemented with 5% fetal
calf serum. These were microinjected in the nucleus with cDNA encoding
plasmids for the yellow fluorescent protein (YFP5), which distributes
evenly throughout the cytosol and nucleus, and a green fluorescent
fusion protein (NLS-GFP5) that remains largely located in the nucleus
(Pepperkok et al., 1999
). The cells were washed and submerged in
CO2 independent medium (GibcoBRL) before they were measured
at 37°C. The same procedure was used to coexpress a Golgi resident
green fluorescent fusion protein (NA-GFP5) and YFP5 in Vero cells.
Phase-dependent images were collected using the mfFLIM setup that was
described in detail by Squire et al. (1999)
. Excitation was with a
488-nm argon/krypton laser line. The fluorescence emission was detected
through a filter block containing a dichroic beam-splitter 505LP in
combination with either a High Q bandpass filter 515/10 BP for the Hela
cells or with a 540/50 BP for the Vero cells (Delta Light and Optics).
The sample was illuminated and the fluorescence collected with a Zeiss
Plan Apochromat 100×/1.4 NA ph3 oil objective. The frequency of the
fundamental harmonic was 39.175 MHz. A Fourier analysis of mfFLIM
reference data from a reflecting sample showed that at least five
significant harmonics were present, and therefore five harmonics were
assumed for determining the modulations and phase shifts in the data.
The first three harmonics had a modulation depth larger than 25%, and
only these were used in the further global analysis of the data. For
each sample 16 phase-dependent images at 22.5° phase steps with
200-ms exposure were collected. To compensate in first order for
photobleaching, a second sequence of 16 images was collected by phase
sampling backward, which were added to the first set (Gadella et al.,
1994
). For our data this first-order correction was sufficient for the
relative short exposure times that were used. Before analysis, a
background correction was applied to each phase-dependent image by
subtracting the mean pixel value within a flat region outside the cell.
A mask was then generated by thresholding the first image in the FLIM
sequence. Only pixels within the mask were used in the global analysis. The mfFLIM data were also used to test the concepts of global analysis
of single-frequency data. For each of the first three harmonics a
single-frequency global analysis was performed, which is equivalent to
global analysis of data acquired with a single-frequency FLIM
instrument. In addition these three harmonics were used in a
multiple-frequency global analysis of the data.
The lifetime values that are found for the different GFP variants by
each analysis are given in Table 1. The
value of ~3.6 ns that was found for YFP5 does not vary much for
different frequencies, and the same value is recovered by single- and
multiple-frequency analysis. This indicates that the decay kinetics of
YFP5 are well described by a monoexponential decay. In contrast, the
values found for NLS-GFP5 and NA-GFP5 vary strongly with frequency.
This indicates that the decay kinetics of these GFPs are complex. These observations are in agreement with previous measurements by Pepperkok et al. (1999)
at a single frequency (80 MHz), where the lifetimes were
calculated from phase shift and modulation, using Eq. 8. These results
indicated that YFP5 is indeed homogeneous, with a lifetime of ~3.6
ns. The decays of NLS-GFP5 and NA-GFP5 were found to be heterogeneous,
with estimations of 1.92 ns and 2.05 ns from the phase, and 2.25 ns and 2.40 ns from the modulation.
|
Single-frequency measurements are most sensitive at values of
1/
and thus detect the fluorescence of short lifetime components more efficiently as the frequency increases. The estimated lifetime values of NLS-GFP5 and NA-GFP5 in Table 1 decrease with increasing frequency, indicating that the measurements become sensitive for a
short component in the decay kinetics of these GFP variants. Thus, for
higher modulation frequencies, the analysis is more strongly influenced
by the choice of an incorrect two-component model, and consequently the
estimated populations are more likely to be erroneous. The
multiple-frequency analysis is also sensitive to such model errors,
because it makes use of information obtained at multiple harmonics. By
choosing a sufficiently low frequency in a single-frequency analysis,
one can filter out short lifetime components to obtain an adequate fit
of the long component. This approximation works well because the
integrated fluorescence contribution from the short component can be neglected.
The fluorescence intensity from each species of molecule can be found by multiplying the steady-state fluorescence by the fractions obtained by global analysis. Fig. 5 gives the steady-state fluorescence intensity of the data with the calculated intensities of NLS-GFP5 and YFP5 for the single-frequency calculation at 39.175 MHz and for the multiple-frequency analysis. The intensities of the NLS-GFP5 and YFP5 are disentangled effectively by the global analysis and show that the NLS-GFP5 is located mainly in the nucleus, while the YFP5 is distributed evenly throughout the cell as expected. Because of the heterogeneous nature of NLS-GFP5, the two-component model used by in the analysis is not correct, and the multifrequency analysis should be more sensitive to this model error than the single-frequency measurement at 39.175 MHz. Despite this, the intensity images from the two analysis approaches do not appear to differ.
|
Similar results can be seen in Fig. 6, where the intensities of NA-GFP5 and YFP5 are separated. The global analysis effectively separates the Golgi from the rest of the cell. In particular note the Golgi at the top of the image that is not clearly visible in the average intensity image because it is hidden by the high intensity of YFP5. YFP5 diffuses throughout the cytosol and nucleus, but it does not enter the Golgi. Thus the intensity of YFP5 should be lower in the Golgi, although it will not be zero, because of out-of-focus contributions from other parts of the cell. Fig. 7 focuses on the lower part of the images in Fig. 6. Fig. 7 A shows that in the single-frequency analysis at 39.175 MHz the intensity of YFP5 is indeed lower in the Golgi, while in the multiple-frequency analysis (Fig. 7 B) this is not the case. The difference is emphasized in Fig. 7 C, which shows the division of Fig. 7 A and Fig. 7 B. The lifetime value for NA-GFP5 found by the multifrequency analysis is a weighted average of the true lifetimes that does not characterize well the complex decay kinetics of the NA-GFP5 fluorescence. This model error propagates to the estimation of the fractional fluorescence of each species. In the single-frequency analysis at 39.175 MHz this model error appears to be much smaller. For comparison, the NA-GFP5 fluorescence is shown in Fig. 7 D, indicating that at 39.175 MHz the single-frequency analysis adequately disentangles the fluorescence distributions.
|
|
| |
CONCLUSIONS |
|---|
|
|
|---|
We have shown that the application of global analysis algorithms can significantly improve the accuracy and precision of lifetime and population estimations from FLIM and mfFLIM data. We employ the prior knowledge that each pixel contains a mixture of different fluorophores or fluorescently tagged molecules in different biochemical states, which gives rise to a multiexponential fluorescence decay, the lifetimes of which do not vary spatially. The parameters of interest are the lifetimes of each molecular species and the populations of all species in each pixel. Two approaches to making use of the spatial invariance of the lifetimes were implemented. The first approach averages the phase-dependent data spatially to increase the signal-to-noise ratio at the cost of lost spatial information. From these averaged data an accurate estimation of the lifetimes can be made. The estimated lifetimes from this lifetime invariant fit are used to recover the populations in each pixel. The second approach uses global analysis algorithms that analyze all pixels simultaneously, while constraining the lifetimes in all pixels to be equal.
It was shown that for a given number of harmonics the global fit is capable of fitting twice the number of lifetimes compared to a conventional pixel-by-pixel fit or a spatial invariant fit. This is an important property because applications in fluorescence microscopy often require that exposure times be minimized, which can be achieved by decreasing the number of frequencies. Another important consequence is that it becomes possible to resolve two lifetimes with a single-frequency FLIM setup. This is of importance because these types of lifetime microscopes are simpler to implement and more widespread than multiple-frequency setups.
Using simulations, it was found that both approaches are far superior to a conventional pixel-by-pixel fit. It was also found that the global analysis was superior to the lifetime invariant fit. This can be explained by the fact that the global analysis approach can make use of spatial information, whereas the spatial invariant fit just improves the estimation of the lifetimes by increasing the signal-to-noise ratio at the cost of that spatial information. The global analysis performs much better, particularly at low signal-to-noise ratios, which is important if the noise in the FLIM sequence is larger, for instance, in applications that require rapid imaging of living cells.
The global analysis algorithm was applied to disentangle the
fluorescence intensities of coexpressed green fluorescent proteins in
living cells. It was found that this is possible, even if the assumption that each fluorophore has monoexponential decay kinetics is
only an approximation. Short components in a complex decay can be
filtered out in a single-frequency measurement by carefully choosing a
modulation frequency where its contribution to the fluorescence is
small. Adequate results can then be obtained if the decay behaves
effectively as a monoexponential with a lifetime approaching the long
component of the complex decay. A multifrequency analysis also makes
use of the information contained in the higher harmonics and is
therefore sensitive to short components in a complex decay. As a
result, the analysis of multifrequency measurements may be more
sensitive to errors in the decay model that is assumed in the analysis.
In the case where one or more fluorophores are present with a complex
decay, it is appropriate to fit mfFLIM data with the correct
multiexponential model to obtain more accurate results. However,
fitting more than two components is a far more complex problem than
fitting a two-component model (Gratton et al., 1984
; Lakowicz et al.,
1984
). This problem might be solved by an algorithm that exploits the
knowledge that a fluorophore has a multiexponential decay with fixed
relative amplitudes. Resolution of fluorophores with complex decays and
separation of more than two fluorophores continues to be a topic of research.
In this work, the global analysis technique was applied to FLIM in the
frequency domain; however, we expect that similar gains can be achieved
with systems that operate in the time domain. Another interesting
application would be to photobleaching data (Jovin and Arndt-Jovin,
1989
), which follow exponential decay kinetics with decay times
that are inversely proportional to the lifetime of the fluorophore.
Fluorescence energy resonance transfer (FRET) is an important application that is currently under investigation in our laboratory. FRET can be measured by monitoring the decrease in lifetime of the donor fluorophore as it nonradiatively transfers energy to an acceptor that is typically within a distance of 10 nm. In applications where FRET occurs because of the formation of a protein complex, the spatial configuration of the donor and acceptor is fixed upon binding, which determines the decrease in the donor lifetime. Thus, when FRET is imaged with FLIM, the decay model may be assumed to be biexponential with spatially invariant lifetimes, and global analysis should provide a powerful tool for evaluating the populations and properties of bound and unbound protein states in cells.
| |
ACKNOWLEDGMENTS |
|---|
PJV was supported by a long-term fellowship from the European Molecular Biology Organization (EMBO).
| |
FOOTNOTES |
|---|
Received for publication 29 October 1999 and in final form 29 December 1999.
Address reprint requests to Dr. Philippe I. H. Bastiaens, Imperial Cancer Research Fund, Cell Biophysics Laboratory, 44 Lincoln's Inn Fields, London WC2A 3PX, England. Tel.: 44-171-269-3082; Fax: 44-171-269-3094; E-mail: p.bastiaens{at}icrf.icnet.uk.
| |
REFERENCES |
|---|
|
|
|---|
I.
Proc. Natl. Acad. Sci. USA.
93:8407-8412[Abstract].
a stereochemical model for tyrosine kinase receptor activation.
J. Cell Biol.
129:1543-1558[Abstract].
spatial resolution of microstructures on the nanosecond time-scale.
Biophys. Chem.
48:221-239
activation in cells.
Science.
283:2085-2089[Abstract/Full Text].
a truncated Newton minimization package for large-scale problems. I. algorithm and usage.
ACM Trans. Math. Soft.
18:46-70.
Biophys J, April 2000, p. 2127-2137, Vol. 78, No. 4
© 2000 by the Biophysical Society 0006-3495/00/04/2127/11 $2.00
This article has been cited by other articles:
![]() |
S. Padilla-Parra, N. Auduge, M. Coppey-Moisan, and M. Tramier Quantitative FRET Analysis by Fast Acquisition Time Domain FLIM at High Spatial Resolution in Living Cells Biophys. J., September 15, 2008; 95(6): 2976 - 2988. [Abstract] [Full Text] [PDF] |
||||
![]() |
G.-J. Kremers, E. B. van Munster, J. Goedhart, and T. W. J. Gadella Jr. Quantitative Lifetime Unmixing of Multiexponentially Decaying Fluorophores Using Single-Frequency Fluorescence Lifetime Imaging Microscopy Biophys. J., July 1, 2008; 95(1): 378 - 389. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Digman, V. R. Caiolfa, M. Zamai, and E. Gratton The Phasor Approach to Fluorescence Lifetime Imaging Analysis Biophys. J., January 15, 2008; 94(2): L14 - L16. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. T. Neves-Petersen, S. Klitgaard, A. S. L. Carvalho, S. B. Petersen, M. R. Aires de Barros, and E. P. e Melo Photophysics and Photochemistry of Horseradish Peroxidase A2 upon Ultraviolet Illumination Biophys. J., March 15, 2007; 92(6): 2016 - 2027. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Esposito, H. C. Gerritsen, and F. S. Wouters Fluorescence Lifetime Heterogeneity Resolution in the Frequency Domain by Lifetime Moments Analysis Biophys. J., December 1, 2005; 89(6): 4286 - 4299. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Keese, R. J. Magdeburg, T. Herzog, T. Hasenberg, M. Offterdinger, R. Pepperkok, J. W. Sturm, and P. I. H. Bastiaens Imaging Epidermal Growth Factor Receptor Phosphorylation in Human Colorectal Cancer Cells and Human Tissues J. Biol. Chem., July 29, 2005; 280(30): 27826 - 27831. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Stepinac, S. R. Chamot, E. Rungger-Brandle, P. Ferrez, J.-L. Munoz, H. van den Bergh, C. E. Riva, C. J. Pournaras, and G. A. Wagnieres Light-Induced Retinal Vascular Damage by Pd-porphyrin Luminescent Oxygen Probes Invest. Ophthalmol. Vis. Sci., March 1, 2005; 46(3): 956 - 966. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Parsons, J. Monypenny, S. M. Ameer-Beg, T. H. Millard, L. M. Machesky, M. Peter, M. D. Keppler, G. Schiavo, R. Watson, J. Chernoff, et al. Spatially Distinct Binding of Cdc42 to PAK1 and N-WASP in Breast Carcinoma Cells Mol. Cell. Biol., March 1, 2005; 25(5): 1680 - 1695. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Peter, S. M. Ameer-Beg, M. K. Y. Hughes, M. D. Keppler, S. Prag, M. Marsh, B. Vojnovic, and T. Ng Multiphoton |