| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, May 2000, p. 2597-2605, Vol. 78, No. 5
Departments of Biochemistry and Biophysics and of Radiology, University of Rochester School of Medicine and Dentistry, Rochester, New York 14642 USA
| |
ABSTRACT |
|---|
|
|
|---|
Phosphorescence quenching of certain metalloporphyrins is
used to measure tissue and microvascular pO2. Oxygen
quenching of metalloporphyrin triplet states creates singlet oxygen,
which is highly reactive in biological systems, and these
oxygen-consuming reactions are capable of perturbing tissue
oxygenation. Kinetics of photochemical oxygen consumption were measured
for a Pd-porphyrin in two model systems in vitro over a range of
irradiances (1.34-134 mW cm
2). For a given irradiance,
and, after correction for differing porphyrin concentrations, rates of
oxygen consumption were similar when the Pd-porphyrin was bound to
bovine serum albumin and when it was taken up by tumor cells in
spheroids. At irradiances comparable to those used in imaging
superficial anatomy, rates of oxygen consumption were sufficiently low
(2.5 µM s
1) that tissue oxygenation would be reduced by
a maximum of 6%. An irradiance of 20 mW cm
2, however,
initiated a rate of oxygen consumption capable of reducing tissue
pO2 by at least 20-40%. These measured rates of
consumption impose limitations on the use of phosphorescence quenching
in thick tissues. The irreversible photobleaching of the Pd-porphyrin was also measured indirectly. The bleaching branching ratio, 23 M
1, is significantly lower than that of porphyrin
photodynamic agents.
| |
INTRODUCTION |
|---|
|
|
|---|
Remarkably high yields of room-temperature
triplet state phosphorescence have been demonstrated for a number of
metalloporphyrins (Eastwood and Gouterman, 1970
; Gouterman et al.,
1975
; Vanderkooi et al., 1987
). Porphyrin triplets are efficiently
quenched by molecular oxygen
(3O2), which is a ground
state triplet, through the scheme credited to Kautsky (Foote, 1968
),
T + 3O2
S + 1O2, where T and
S are the porphyrin triplet state and ground singlet state,
respectively, and 1O2 is
singlet oxygen, the lowest lying electronic excited state of dioxygen.
Thus, the lifetime of the metalloporphyrin phosphorescence provides an
optical reporter of the oxygen concentration of the ambient medium. The
use of phosphorescence quenching in biological systems was introduced
by Vanderkooi et al. (1987)
, and it has been implemented by that group
and by several other investigators to measure the oxygen tension in a
variety of biological systems in vitro and in vivo. Examples of the use
of this method in vivo include pO2 measurements
in subcutaneous rodent tumor models (Wilson and Cerniglia, 1992
;
Cerniglia et al., 1997
), exposed organs (Rumsey et al., 1988
; McIlroy
et al., 1998
), the retina of the eye (Shonat et al., 1992
), and in thin
window chamber tissue preparations (Torres Filho and Intaglietta, 1993
;
Torres Filho et al., 1994
; Buerk et al., 1998
).
Recently, Vinogradov and his colleagues have developed and
characterized a series of longer-wavelength absorbing and emitting metalloporphyrins to take advantage of the more favorable optical properties of tissue in the red and near infrared spectral regions (Vinogradov and Wilson, 1995
; Vinogradov et al., 1996
). Between ~600
and 1300 nm, absorption of soft tissue is significantly reduced compared to that on either side of this window (Wilson and Jacques, 1990
). Light scattering by tissue, which follows a weaker wavelength dependence, is also reduced in this region compared to shorter wavelengths (for example, Parsa et al., 1989
). The phosphor Green 2W,
with an absorption maximum at 636 nm and a phosphorescence emission
maximum at 790 nm (Vinogradov et al., 1996
), is well designed to allow
exploitation of these favorable optical properties in the attempt to
image oxygenation of relatively deep tissue structures, such as
malignant tumors of the breast and of the extremities, for example, in
which hypoxia has been linked to a poor response to ionizing radiation
therapy and to increased probability of appearance of metastases
(Okunieff et al., 1993
; Brizel et al., 1996
).
Although light at these wavelengths can penetrate several centimeters
in tissue, it is rapidly attenuated, and efforts to extract optical
signals from phosphorescent probes in thick tissues must address this
practical reality. Using typical values of soft tissue absorption and
transport scattering coefficients, 0.01 and 1.0 mm
1, respectively, the value of the
characteristic effective attenuation length in the diffusion
approximation is approximately 5.7 mm. Because both the excitation
light and the luminescence emission will be attenuated as they
propagate from and to the tissue surface, higher excitation irradiances
will be needed for signal recovery from deeper structures. This fact
has been acknowledged by Vinogradov et al. (1996)
, who suggested that
excitation lamp intensities could be "increased 10- to 100-fold" in
efforts to extend phosphorescence-quenching measurements to "tissue
thicknesses of 5 cm or greater." We have also suggested the
feasibility of using metalloporphyrin phosphorescence to detect regions
of hypoxia in tumors residing several centimeters from the skin surface
(Hull et al., 1998
). In considering implementation of higher excitation
irradiances in such measurements, it is necessary to evaluate carefully
the consequences of the photochemical kinetics. Because
1O2 is highly reactive in
biological systems, the quenching of metalloporphyrin triplets by
3O2 can, under conditions
of sufficiently high irradiance, result in significant rates of
photochemical oxygen consumption, which could potentially perturb the
oxygen concentration that is being measured. These
1O2 reactions are also
capable of inducing undesirable biological damage (Weishaupt et al.,
1976
) and may lead to the self-sensitized photobleaching of the
metalloporphyrin (Moan and Berg, 1991
). Little experimental work has
been reported that investigates the possible limitations that these
phenomena may impose on the application of the phosphorescence
quenching technique. The purpose of this study was to perform detailed
oxygen-consumption measurements during irradiation of a widely studied
metalloporphyrin in environments relevant to the in vivo situations in
which luminescence-quenching measurements are performed. In one series
of experiments, the metalloporphyrin was bound to bovine serum albumin
(BSA) in aqueous solution to mimic the environment encountered during
intravascular pO2 measurements. In the second,
multicell tumor spheroids were incubated with porphyrins to approximate
the situation in which the phosphorescent probe leaves the blood
vessels and is taken up by the surrounding cells. We also evaluated the
rates and possible mechanism of photobleaching of this compound.
| |
METHODS |
|---|
|
|
|---|
Chemicals
Pd-meso-tetra(4-carboxyphenyl)porphine (PdTCPP) was obtained from Porphyrin Products Inc. (Logan, UT) and was used as received. To simulate two types of environments in which metalloporphyrin phosphorescence has been used to report O2 concentrations, O2 consumption experiments were performed on aqueous solutions of PdTCPP bound to BSA in cuvettes and on multicell tumor spheroids preincubated with PdTCPP. Cell culture media, fetal calf serum, and Hank's balanced salt solution were purchased from GIBCO (Grand Island, NY). Unless otherwise noted, other chemicals and reagents were obtained from Sigma Chemical (St. Louis, MO).
Solution experiments
To simulate intravascular O2 concentration
measurements, PdTCPP (0.7 mg ml
1) was added to
a solution of BSA (60 mg ml
1) in physiological
saline buffered to pH 7.4. The PdTCPP concentration we used was
determined on the basis of in vivo experiments reported by Wilson and
Cerniglia (1992)
and Cerniglia et al. (1997)
, in which 4 mg PdTCPP were
administered to tumor-bearing rats via intravenous tail vein injection.
The rat blood plasma volume is ~5-6 ml, thus the intravascular
concentration in vivo of the metalloporphyrin was approximately 0.7 mg
ml
1. Metalloporphyrin solution, 150 µl, was
placed in a UV quartz cuvette (NSG Precision Cells Inc., Farmingdale,
NY) with inside dimensions of 3×3×24 mm (height). A Clark-style
microelectrode (Diamond General, Ann Arbor, MI) with a tip diameter of
~10 µm and a response time of <1 s was immersed into the sample
solution. Sample irradiation was performed using the 514-nm output of a cw argon ion laser (Coherent Inc., Santa Clara, CA). The laser light
was delivered to the surface of the cuvette through an optical fiber
terminated with a gradient index lens (General Fiber Optics, Fairfield,
NJ). The optical density of the sample at 514 nm attenuated the
incident beam 10-fold across the 3-mm cuvette, producing a gradient of
photochemical oxygen consumption. Thus, to ensure reproducibility, care
was taken to position the electrode tip inside the cuvette as closely
as possible to the surface at which the laser beam was incident. The
incident irradiation spot size was 4.9 cm2, so
that the laser beam overfilled the area of the cuvette occupied by the
sample. The electrode current, which is directly proportional to the
O2 concentration, was recorded during the
irradiation period. We observed a monotonic decrease in
O2 concentration of the solution with the onset
of irradiation. As indicated in Fig. 1,
this O2 depletion results from the irreversible
reaction of 1O2 with
substrates, which, in the case of the solution experiments, are amino
acids of BSA. These reactions compete with the monomolecular decay of
1O2 to the ground triplet
state, resulting in a progressive depletion of O2
from the medium as irradiation proceeds. The reported
O2 consumption rates are averages calculated from
three separate experiments at each of five incident irradiances. In
each experiment, the O2 consumption rate was
determined from the initial slope of the electrode data.
|
Spheroid experiments
To determine O2 consumption rates under
conditions of intracellular metalloporphyrin distribution, experiments
were performed with multicell tumor spheroids. Our electrode technique
is a time-dependent extension of the method developed by
Mueller-Klieser (1984)
. The details of the method have been described
previously (Nichols and Foster, 1994
). Briefly, EMT6/Ro spheroids
(500-µm diameter) were incubated in a 100-mm suspension-tissue
culture dish containing 20 ml Eagle's basal medium with 10% fetal
calf serum and 10 µg ml
1 PdTCPP at 37°C in
a humidified 5% CO2 and 95% air atmosphere for
approximately 24 hrs. After the incubation period, a single spheroid
was selected and placed in an open dish containing 20 ml of Hank's
balanced salt solution. The spheroid was then immobilized with a thin
glass needle on top of a pedestal, which has an
O2 permeable membrane. The tip of the Clark-style
microelectrode was placed at the edge of the spheroid using a
micropositioning device. A lensed optical fiber coupled to an argon ion
laser (Ion Laser Technology, Salt Lake City, UT) delivered 514 nm light
to the surface of the pedestal. The spheroids were irradiated using three different irradiances, and the results presented are averages computed from three separate experiments at each irradiance. The microelectrode current represents the changes in the
O2 concentration taking place in the spheroid.
The irradiation was continuous for approximately 500 s.
Consistently, we observed a rapid initial decrease in the
O2 concentration followed by a slight increase due to metalloporphyrin photobleaching. The initial
O2 consumption rate,
0,
and the ratio of two photophysical rate constants,
kp/kot, were determined by fitting numerical solutions to a pair of
diffusion-with-reaction equations (see below) to the first 40 s of
data using a Levenberg-Marquardt nonlinear least squares fitting
algorithm. After these two parameters were determined from the initial
O2 transients, the entire time-dependent data
were analyzed to obtain the ratio
kos/koa[A],
as described below. All rate constants are defined in the next section
and are depicted in Fig. 1.
Spheroid data analysis
Details of the method developed to analyze time-dependent
O2 concentrations in spheroids during
photodynamic therapy have been published elsewhere (Nichols and Foster,
1994
; Georgakoudi et al., 1997
). Briefly, the spatial and temporal
distribution of the O2 concentration within and
in the proximity of a sensitized multicell spheroid is described by
solutions to the time-dependent diffusion equations,
|
(1) |
|
(2) |
(r, t) is the total rate of
O2 consumption within the spheroid, which, during
irradiation, includes contributions from both photodynamic and
metabolic processes. Therefore, for a sensitized spheroid during
irradiation,
|
(3) |
met is the rate of
O2 consumption due to cellular metabolism, and
PDT(r, t) represents
the rate of photochemical O2 consumption. The
explicit functional form of
PDT is derived from the kinetic equations governing the photophysical processes, as
described in detail by Nichols and Foster (1994)
PDT is written as
|
(4) |
|
(5) |
denotes the fraction of triplet quenching
collisions with 3O2 that
result in 1O2 formation,
t denotes the photosensitizer triplet yield,
Ia(t) denotes the rate of
photon absorption, kot is the
bimolecular rate of triplet quenching by
3O2,
koa is the rate of chemical reaction
between 1O2 and unspecified
substrate [A], kos is the rate of
chemical reaction between
1O2 and the
metalloporphyrin, and kd and
kp are the monomolecular decay rates
of 1O2 and the
metalloporphyrin triplet, respectively.
0 is a
constant that, in our experiments, describes the initial rate of
photochemical O2 consumption, i.e., the rate of
consumption before the onset of significant O2
depletion or photobleaching. We note that treating
0 in this way implies an assumption of a
uniform distribution of the metalloporphyrin throughout the spheroid.
In the event that there is a radial concentration gradient, the value
of
0 that we extract will underestimate the
rate of O2 consumption near the outer rim of the
spheroid, and it will overestimate this rate near the center of the
spheroid. On the basis of the above definitions and the energy level
diagram of Fig. 1, the ratio kp/(kot
[3O2]) defines a
branching ratio for the metalloporphyrin triplet state. When this ratio
is equal to unity, the triplets are as likely to undergo monomolecular
decay to the ground state as they are to be quenched through collisions
with 3O2. Thus, the ratio
kp/kot,
which has dimensions of a concentration, is a measure of the
3O2 dependence of the
dye-sensitized 1O2
formation process. At O2 concentrations higher
than
kp/kot, quenching by O2 is the predominant decay route
for the triplets, whereas, at lower concentrations, the quenching
efficiency is limited by O2 availability. As
noted above, numerical solutions to Eqs. 1 and 2 are fitted to the
first 40 s of the data to obtain best estimates of
0 and
kp/kot.
Then, solutions are fit to the entire data set (approximately 500 s of recordings) to obtain the ratio
kos/koa[A],
which provides a measure of the self-sensitized photobleaching.
Selection of irradiances
The rationale for the specific irradiances used in these studies
was based on the published literature, with particular emphasis on
phosphorescence-quenching experiments performed in vivo. The lowest
value was chosen to match the average irradiance used in imaging
studies of superficial anatomy. Shonat et al. (1992)
used flash-lamp
pulse fluences of 39 µJ cm
2 for blue light
excitation and 52 µJ cm
2 for green
excitation. Typically, for each data acquisition delay time, 8 frames
are averaged, and the total time required to obtain an image is
approximately 1 s (Wilson and Cerniglia, 1992
; Vinogradov et al.,
1996
; Cerniglia et al., 1997
). Thus, the time between successive
flash-lamp excitations is 125 ms, and the average irradiance incident
on the tissue surface is ~0.3 mW cm
2. The
extinction coefficient of PdTCPP is much greater in the blue region
than it is at 514 nm, where our experiments were performed. Using the
measured absorption spectrum of PdTCPP (see Fig. 5) and accounting for
the difference in photon energies at the two wavelengths, we determined
the 514-nm intensity that would match the rate of photon absorption
that resulted from 0.3 mW cm
2 delivered at 418 nm.
These considerations combine such that the appropriate 514-nm
irradiance is ~1.3 mW cm
2. The higher
irradiances were selected because of the recent availability of longer
wavelength absorbing and emitting metalloporphyrins (Vinogradov and
Wilson, 1995
; Vinogradov et al., 1996
), which have created interest in
the possibility of imaging oxygen at depths of several centimeters in
tissue. On the basis of the suggestion of Vinogradov et al. (1996)
that
the flash intensity could be increased 10-100-fold, we selected a
maximum irradiance of 134 mW cm
2.
Porphyrin concentration in spheroids
To make a quantitative comparison between the spheroid and BSA
solution oxygen-consumption rates, it is necessary to measure the
concentration of metalloporphyrin in the spheroids. To accomplish this,
20 spheroids (500 µm diameter) were incubated with 10 µg ml
1 PdTCPP for 24 hrs as described above. After
incubation, the spheroids, along with the media, were placed in a 15-ml
tube and centrifuged for 5 min. After centrifugation, the medium above
the spheroid pellet was aspirated, and the spheroids were then
dissociated and the cells dissolved using 25% Scintigest in 10×
trypsin without phenol red. We then measured the absorption spectrum of
this sample. The metalloporphyrin concentration was calculated from
these absorption measurements using a calibration curve generated from
the absorbance of known PdTCPP concentrations in the same solvent.
| |
RESULTS |
|---|
|
|
|---|
In the solution experiments, the choice of the lowest irradiance,
1.34 mW cm
2, was based on an approximation to
the average incident irradiance used in the in vivo experiments
described by Shonat et al. (1992)
. As described in the Methods section,
this average was computed on the basis of the fluence per pulse and the
pulse repetition rate reported by those authors. Examples of the
photochemically-induced changes in the
3O2 concentration observed
using the microelectrode during sample irradiation with this and with
several other irradiances are shown in Fig.
2. From three experiments performed with
separate samples at the lowest irradiance, the
3O2 consumption rate was
found to be 2.5 ± 0.1 µM s
1. Typical
image-acquisition times reported in papers that describe results of in
vivo phosphorescence imaging of superficial anatomy are 1-1.5 s
(Wilson and Cerniglia, 1992
; Cerniglia et al., 1997
). Oxygen tensions
in normal tissue are typically in the range of 40-50 torr (~60-80
µM). Thus, it may be concluded that the conditions under which the
phosphorescence-quenching measurements are performed for superficial
anatomy produce a modest reduction in the tissue 3O2 concentration of 6% or
less, provided that there is sufficient time between image acquisitions
to allow for 3O2 recovery.
|
As described earlier, the rationale for the particular choices of the
higher irradiances was based on suggestions published by Vinogradov et
al. (1996)
and on our own experimental experience with detection of
sources of luminescence buried in scattering media (Hull et al., 1998
).
With irradiances of 20, 60, 100, and 134 mW
cm
2, the O2 consumption
rates were found to be considerably higher, as shown in Fig. 2 and
summarized in Fig. 3. The rates reported in Fig. 3 were determined from the initial slopes of the curves in Fig.
2 and represent means and standard deviations of three separate
experiments. At 20 mW cm
2, assuming the same
range of tissue O2 concentrations and
image-acquisition times as mentioned above, the data indicate that the
measurement would perturb the O2 concentration by
a minimum of 20% and by as much as 40%. Of course, higher irradiances
would have an even greater effect. The curves of Fig. 2 illustrate that
the rate of depletion is diminished at the lowest
O2 concentrations. This diminished rate occurs as
the concentration approaches that described by the ratio
kp/kot
(see Methods) and results from the O2 dependence of the rate of 1O2
formation. Thus, the magnitude of the perturbation imposed by the
phosphorescence-quenching technique depends in part on the ambient
O2 concentration.
|
Whereas the solution experiments, where the metalloporphyrin is bound
to BSA, represent a reasonable approximation to the experiments in vivo
that seek to measure intravascular O2
concentrations, the spheroid provides a test of the
O2 consumption that would be expected under
conditions where the phosphorescent probe leaks from the vasculature
into the surrounding tissue. Examples of the initial changes in the
3O2 concentration that we
record at the edge of PdTCPP-sensitized spheroids for irradiances of
20, 40, and 60 mW cm
2 are shown in Fig.
4. The solid lines through the curves
represent the best fits of numerical solutions of Eqs. 1 and 2 to the
data. In this analysis, both
0 and
kp/kot
are extracted as fitting parameters. From a total of nine experiments
(n = 3 for each irradiance), we find that the ratio of
kp/kot
for PdTCPP is 8.3 ± 3.8 µM. Figure 5 summarizes the values of the means and
standard deviations of
0 determined from the
same nine experiments.
|
|
A comparison of Figs. 3 and 5 indicates that, under the conditions of
these experiments, the photochemical O2
consumption rate in spheroids was significantly greater than that
observed in solutions containing BSA at the same irradiance. To compare these results on a quantitative basis, we determined the concentration of the metalloporphyrin in a spheroid after the same 24-hr incubation period used in the O2 consumption experiments.
From a series of three independent solubilization and optical
absorption measurements, we calculated the PdTCPP concentration in the
spheroids to be approximately 1.1 mg ml
1. The
absorption spectrum of PdTCPP in aqueous solution of trypsin and the
solubilizing agent Scintigest is shown in Fig.
6 along with the calibration curve used
to determine the metalloporphyrin concentration. The porphyrin
concentration in the spheroids was greater than that used in the
solution experiments (0.7 mg ml
1). When the
O2 consumption rates measured in BSA solution are corrected for this difference in porphyrin concentration, they approach
those measured in the spheroids.
|
When porphyrins are subjected to sustained irradiation in biological
systems, irreversible photobleaching is observed (Mang et al., 1987
;
Moan et al., 1988
). Bleaching may be measured indirectly with
O2 electrodes through the gradual increase in the
O2 concentration that results as the rate of
light absorption by the porphyrin decreases (Georgakoudi et al., 1997
).
To determine the rate and mechanism of PdTCPP bleaching, we analyzed
time-dependent O2 concentration data recorded
continuously for approximately 500 s at the edge of spheroids
irradiated using three different fluence rates. Representative data are
shown in Fig. 7. The best fits of
solutions to Eqs. 1 and 2 to the data are depicted by the solid lines.
The particular form used for the rate of photochemical
O2 consumption,
PDT, was
derived on the basis of a bleaching mechanism in which the porphyrin is
degraded through self-sensitized
1O2 reaction. Fits to the
data allow determination of a ratio, kos/koa[A],
which is a measure of the relative bleaching efficiency. From the nine
experiments conducted using the three fluence rates mentioned above, we
find that the value of this ratio for PdTCPP is 23.0 ± 7.0 M
1 (mean ± SD). This value is
significantly lower than that measured previously by Georgakoudi et al.
(1997)
for the photosensitizer Photofrin (76 ± 12 M
1) and by Georgakoudi and Foster (1998)
for
-aminolevulinic acid-induced protoporphyrin IX (90 ± 15.9 M
1) using this method.
|
| |
DISCUSSION |
|---|
|
|
|---|
The findings of this study illustrate potential limitations
imposed by O2 consumption during measurement of
tissue pO2 by the metalloporphyrin
phosphorescence-quenching technique. We have demonstrated the severity
of the problem by determining the O2 consumption
rates when the metalloporphyrin PdTCPP was irradiated in two
environments in vitro. In the first case, the PdTCPP was bound to BSA
in solution, with the metalloporphyrin concentration chosen to be as
close as possible to that used in the intravascular pO2 measurements in vivo of Cerniglia et al.
(1997)
and Wilson and Cerniglia (1992)
. In these in vivo experiments,
the metalloporphyrin is bound to BSA, so that the probe is confined to
the blood volume and so that the calibration parameters calculated from
measurements in vitro are relevant in vivo (Sinaasappel and Ince, 1996
;
Vinogradov et al., 1996
). In the second case, multicell spheroids were
incubated with PdTCPP to approximate the environment encountered when
the porphyrin leaks out of the vessels into the tissue (Torres Filho et
al., 1994
; Buerk et al., 1998
).
The results of our solution experiments demonstrate that, for most if
not all published reports in which phosphorescence quenching has been
used to measure O2 concentrations in superficial
structures in vivo and in applications involving intravital microscopy,
photochemical oxygen consumption is of minimal importance. However, if
the average excitation irradiance is increased to just 20 mW
cm
2, the perturbation of the local oxygen
concentration would be significant, even if one assumes that the total
image acquisition time is held to the 1-1.5 s range typically used in
imaging superficial anatomy. Because the detection of luminescence
signals originating at depths of several centimeters in an attenuating
medium will require both relatively higher irradiances and longer
data-acquisition times, our results suggest that great caution must be
exercised in applying the phosphorescence-quenching technique to the
imaging of oxygenation in all but the most shallow layers of tissue. In our own experience with imaging highly fluorescent targets buried in a
scattering emulsion using a sensitive, cooled CCD camera and 160 mW
cm
2 incident irradiance, signal-acquisition
times varied from 10 s for source depths of approximately 6 mm to
245 s for source depths of approximately 44 mm (Hull et al.,
1998
).
The plots of Fig. 2 show that the rate of photochemical oxygen consumption depends on the O2 concentration of the medium, with the rate being diminished at lower concentrations. This phenomenon follows simply from the fact that, as the O2 concentration is lowered, the probability increases that the metalloporphyrin triplet will undergo direct, monomolecular decay to the ground state. Thus, relatively fewer triplets are quenched by oxygen, leading to a reduced rate of 1O2 formation. The characteristic O2 concentration at which this effect becomes appreciable is described by the ratio of porphyrin triplet state rate constants, kp/kot, which were previously defined. This ratio identifies the O2 concentration at which triplet monomolecular decay and quenching by 3O2 are equally likely. Our analysis of the electrode recordings obtained at the edge of irradiated spheroids indicates that, for the PdTCPP compound in the intracellular environment, this characteristic concentration is approximately 8 µM. A practical and perhaps somewhat counterintuitive consequence of this aspect of the photochemistry is that phosphorescence quenching is more likely to perturb tissue oxygenation at relatively higher tissue oxygen tensions. From the perspective of imaging thick tissue, then, it appears unlikely that a scheme could be devised to correct for the effects of photochemical oxygen consumption without prior knowledge of the detailed oxygen distributions.
When the data from the two in vitro systems that we investigated are
corrected for the difference in their PdTCPP concentrations, the rates
of photochemical oxygen consumption are found to be comparable. It is
well known that the intracellular environment presents many targets for
1O2 reaction, and these
data demonstrate quantitatively that BSA provides a similar effective
concentration of substrate. This finding is consistent with those of
Reddi et al. (1984)
, who showed that the tryptophan, tyrosine, and
histidine residues of human serum albumin were readily photo-oxidized
in the presence of hematoporphyrin derivative. Indeed, in the paper
that introduced the phosphorescence-quenching method, Vanderkooi and
her colleagues (1987)
described BSA as "a sink" for reactive oxygen
species. On the basis of our results, therefore, one may conclude that
the potential problems posed by photochemical oxygen consumption are at
least as likely to occur during intravascular phosphorescence-quenching
measurements as they are in those situations where the porphyrin is
allowed to distribute out of the intravascular space.
The limitations imposed by photochemical oxygen consumption on the
applications of metalloporphyrin phosphorescence quenching here appear
to be fundamental to the method rather than specific to the PdTCPP
compound. Therefore, it is not immediately apparent what, if any, means
could be used to diminish or overcome this effect. No significant
photochemical differences should be anticipated through the use of the
longer wavelength-absorbing metalloporphyrin Green 2W. The peak of the
phosphorescence emission of this compound was reported by Vinogradov et
al. (1996)
to be 790 nm. Thus, the Green 2W triplet state is
sufficiently energetic to sensitize the formation of
1O2, which is 1270 nm above
the ground state. It is true that, for a source of phosphorescence
buried several centimeters from the tissue surface, the incident
irradiance would be attenuated significantly, and this will reduce the
rate of photochemical O2 consumption at that
site. However, because the phosphor cannot be expected to localize
perfectly in the target tissue volume, the oxygen concentration in the
superficial tissue would be perturbed significantly. As mentioned
previously, without detailed prior knowledge of the oxygen distribution
in tissue calculation of a correction, does not seem feasible. Other
imaging strategies not based on triplet-state quenching could be
considered. For example, oxygen quenching of singlet-state fluorescence
has been reported (Chong and Thompson, 1985
). Although fluorescent
probes are certainly useful in a variety of fundamental studies, their lifetimes are sufficiently short that it is unlikely that they could be
useful reporters of physiologically relevant hypoxia in tissue.
Therefore, we are left to conclude that, although the very long-lived
porphyrin triplet-state luminescence provides a sensitive measurement
of tissue pO2, the mechanism through which the
triplet reports the O2 concentration imposes
important and quite fundamental limits to the situations in which
phosphorescence quenching can be applied.
An interesting recent study by Buerk et al. (1998)
compared the tissue
oxygen tensions reported by 5-µm tip-recessed electrodes and the
phosphorescence-quenching method in a hamster skinfold model. They
concluded that, under the conditions of their intravital microscopy
experiment, the phosphorescence technique lowered the oxygen
concentration only minimally, even when the irradiation was performed
for as long as 1 min. Although the pulse energies used in that study
were not reported, it is likely that their findings are not
inconsistent with our results. In several respects, intravital
microscopy represents the most favorable situation in vivo for the
application of the phosphorescence method. The thicknesses of tissue
that are interrogated are minimal, typically 200-300 µm, thereby
enabling the use of very low excitation pulse energies and short
acquisition times. In the particular case of the Buerk et al. study, in
which avascular regions of the hamster skin fold were investigated, the
metalloporphyrin concentration that was illuminated was also very low,
because only a small amount of the albumin-bound compound will
extravasate. Further, because the illuminated field diameter is
restricted to approximately 140 µm in these experiments (Torres Filho
and Intaglietta, 1993
), any oxygen depletion during irradiation is
replenished from outside the irradiated area.
The photobleaching of porphyrins in biological systems is a
well-established phenomenon. Although the phosphorescence-quenching technique is based on emission lifetimes rather than amplitudes, the
bleaching kinetics and mechanism of the metalloporphyrins are
nevertheless of interest, especially in the context of proposals to use
the method to image in thick systems. Photobleaching during irradiation
reduces the rate of light absorption and would thereby progressively
reduce the rate of photochemical oxygen consumption. Also, if
photobleaching is mediated through a predominantly oxygen-dependent mechanism, as has been reported for various porphyrins (Krieg and
Whitten, 1984
; Spikes, 1992
), one might imagine a scenario in which a
period of preirradiation preferentially bleached the compound in
relatively well-oxygenated regions, thereby possibly enhancing the
optical detection of hypoxic malignant tumors.
Photobleaching is reported indirectly in the spheroid microelectrode
measurements as a gradual increase in the measured oxygen concentration
with prolonged irradiation (Georgakoudi et al., 1997
). Analysis of the
traces obtained during irradiation of spheroids incubated with PdTCPP
demonstrates that this compound is significantly more stable with
respect to bleaching than the other porphyrins that we have studied. As
noted earlier, this relative stability is reported through the ratio
kos/koa[A],
which defines the reciprocal of the porphyrin concentration at which
1O2 reactions with the
porphyrin are as likely as those with other nearby targets. This
property, along with the efficiency with which it sensitizes the
formation of reactive oxygen species, may make PdTCPP an interesting
candidate for biological and medical studies of photodynamic action.
| |
ACKNOWLEDGMENTS |
|---|
The authors thank Mr. Jim Havens for assistance in maintaining spheroid cultures. This work was supported by National Institutes of Health grants CA68409 and CA36856.
| |
FOOTNOTES |
|---|
Received for publication 20 September 1999 and in final form 25 February 2000.
Address reprint requests to Thomas H. Foster, Department of Radiology, 601 Elmwood Avenue, Box 648, University of Rochester, Rochester, NY 14642. Tel: 716-275-1347; Fax: 716-273-1033; E-mail: thfoster{at}optics.rochester.edu.
| |
REFERENCES |
|---|
|
|
|---|
Biophys J, May 2000, p. 2597-2605, Vol. 78, No. 5
© 2000 by the Biophysical Society 0006-3495/00/05/2597/09 $2.00
This article has been cited by other articles:
![]() |
A. S. Golub and R. N. Pittman PO2 measurements in the microcirculation using phosphorescence quenching microscopy at high magnification Am J Physiol Heart Circ Physiol, June 1, 2008; 294(6): H2905 - H2916. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. Golub, M. C. Barker, and R. N. Pittman Microvascular oxygen tension in the rat mesentery Am J Physiol Heart Circ Physiol, January 1, 2008; 294(1): H21 - H28. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Stepinac, S. R. Chamot, E. Rungger-Brandle, P. Ferrez, J.-L. Munoz, H. van den Bergh, C. E. Riva, C. J. Pournaras, and G. A. Wagnieres Light-Induced Retinal Vascular Damage by Pd-porphyrin Luminescent Oxygen Probes Invest. Ophthalmol. Vis. Sci., March 1, 2005; 46(3): 956 - 966. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |