Biophys J, June 2000, p. 3178-3185, Vol. 78, No. 6



*Department of Molecular and Cell Biology;
Life
Sciences Division and
Physical Biosciences Division,
Lawrence Berkeley National Laboratory; §Graduate Group in
Biophysics; and ¶Experimental Systems Group, Advanced
Light Source, Lawrence Berkeley National Laboratory, University of
California, Berkeley, California 94720 USA
| |
ABSTRACT |
|---|
|
|
|---|
The x-ray exposure at which significant radiation damage occurs has been quantified for frozen crystals of bacteriorhodopsin. The maximum exposure to ~11-keV x-rays that can be tolerated for high-resolution diffraction experiments is found to be ~1010 photons/µm2, very close to the value predicted from limits that were measured earlier for electron diffraction exposures. Sample heating, which would further reduce the x-ray exposure that could be tolerated, is not expected to be significant unless the x-ray flux density is well above 109 photons/s-µm2. Crystals of bacteriorhodopsin that contain ~1011 unit cells are found to be large enough to give ~100 high-resolution diffraction patterns, each covering one degree of rotation. These measurements are used to develop simple rules of thumb for the minimum crystal size that can be used to record x-ray diffraction data from protein microcrystals. For work with very small microcrystals to be realized in practice, however, it is desirable that there be a significant reduction in the level of background scattering. Background reduction can readily be achieved by improved microcollimation of the x-ray beam, and additional gains can be realized by the use of helium rather than nitrogen in the cold gas stream that is used to keep the protein crystals frozen.
| |
INTRODUCTION |
|---|
|
|
|---|
Synchrotron sources provide a number of well-recognized advantages for protein crystallography experiments. Among the key features one can mention are extremely rapid data collection, routinely higher resolution, the opportunity to tune the x-ray wavelength to an absorption edge for multiple anomalous dispersion (MAD) phasing, and the ability to use much smaller crystals than is practical with a laboratory source.
The ability to conduct diffraction experiments with protein
microcrystals can open up a number of important opportunities. Microcrystals can be used to evaluate the quality of diffraction at an
early stage, before extensive effort is invested in improving the
crystal size. If microcrystals are large enough to allow many diffraction patterns to be collected without severe radiation damage,
there is no need to further optimize the conditions to produce larger
crystals. These and other reasons for the development of microcrystal
diffraction cameras have been discussed recently by Cusack et al.
(1998)
. A possible, further benefit, yet to be tested, could be that
very small crystals can be frozen without the use of added
cryoprotectants, as is the case for protein crystals used in electron
diffraction studies (Taylor and Glaeser, 1974
).
Experimental work with protein microcrystals
defined here as crystals
that are only a few tens of micrometers on edge
can be facilitated by
focusing the x-ray beam to ~1/10 the usual size (Bilderback et al.,
1994
; Pebay-Peyroula et al., 1997
). Focusing to a smaller size
increases the x-ray flux density at the crystal, an important
consideration because the exposure required to produce adequate count
statistics in the diffraction spots scales inversely with the crystal
volume. Use of a narrow x-ray beam, comparable in size to the protein
microcrystal itself, also helps to reduce the background scattering
from material surrounding the crystal and from the air in the path
between the collimator and the beam stop.
Radiation damage ultimately places a limit on the smallest protein
crystal size that can be used for high-resolution data collection,
however. A given number of x-rays must pass through a crystal,
regardless of its size, to produce a desired number of counts in each
diffraction spot. Thus, as the number of unit cells in a crystal
decreases, the x-ray exposure per unit cell must increase
proportionately. Finally, when the crystal size is too small, radiation
damage will destroy the crystal before an adequate exposure has been
completed (Henderson, 1990
; Gonzales and Nave, 1994
).
In anticipation of work that will be done with protein microcrystals at
the Advanced Light Source at the Lawrence Berkeley National Laboratory,
we have further characterized the limitation on crystal size that is
imposed by radiation damage. We find that crystals of bacteriorhodopsin
are severely damaged at resolutions better than 3 Å (and significant
damage is apparent at lower resolution) after an exposure of
~1010 photons/µm2. The
damage "threshold" of 1010
photons/µm2 that we measure at ~3-Å
resolution is within a factor of 2 of the value that was estimated by
Henderson (1990)
from earlier measurements made in electron diffraction
experiments. As a conservative rule-of-thumb, we estimate that at least
one high-resolution diffraction pattern, covering 1° of rotation, can
be obtained from a protein crystal if the crystal size (in micrometers)
is 1/10 the unit cell size (in Angstroms). Data covering more than
100° of rotation can therefore be obtained if the crystal size (in
micrometers) is 1/2 the unit cell size (in Angstroms). The need to
minimize the air-scatter background becomes apparent as the crystal
size is reduced to very small dimensions, however.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Measurements of radiation damage on bacteriorhodopsin crystals
were carried out with equipment set up on beamline 7.3.3 of the
Advanced Light Source (ALS) at the Lawrence Berkeley National Laboratory (LBNL). The experiment is illustrated schematically in the
beamline layout shown in Fig. 1. The
toroidal mirror, located 16 m from the source, collected bending
magnet radiation and focused it to a spot in the sample at the 37-m
location. This mirror provides the full amount of horizontal focusing,
resulting in a convergence angle of 2.3 mrads. This mirror provides
only a small amount of vertical focusing, however. Most of the vertical
focusing is provided by the microfocus mirror located 12 cm in front of
the sample. This mirror was of a design similar to those used elsewhere
at the ALS (MacDowell et al., 1997
). Briefly, it consists of a flat, platinum-coated, fused silica substrate that is bent asymmetrically by
means of end couples that are applied by strip springs. The resulting
vertical convergence angle onto the sample was 2.5 mrads. A double
crystal Ge (111) monochromator, located 1 m in front of the
sample, was set to produce 11-keV photons with an energy bandpass of
~1:1000.
|
It should be noted that this temporary configuration was a test setup and is far from optimum. The flux density on the sample is reduced from the ideal value for two reasons. 1) The conjugates of the toroidal mirror were not set at 1:1 because of positional constraints. This results in an increase in the beam focus size and about an eightfold decrease in flux density at the sample. 2) The Ge (111) monochromator was poorly cooled and suffered from thermal strain that reduced throughput by about fivefold. Despite this, the focused spot size at the sample was measured to be 300 × 54 µm (FWHM). To further reduce the horizontal spot size in a controlled manner, motorized slits (Newport Corp.) were positioned 6 cm in front of the sample, followed by a 300-µm lead pinhole to act as a scatter guard.
A spot size of 130 × 54 µm (FWHM) with a total flux of 2.2 × 109 photons/s (400-mA ring current) was used for the measurements of radiation damage. An even smaller spot size of 48 × 54 µm was used to reduce the background scattering for diffraction from very small crystals. The x-ray spot size was measured by allowing the x-rays to illuminate a YAG phosphor. The visible light image of the beam was projected with magnification onto a CCD camera. The flux measurements were determined using a silicon photodiode (International Radiation Detectors). This device was calibrated against a scintillator detector (Bicron Corp) when operated with low-input flux (<50 kHz). The scintillator detector was considered to be 100% efficient with 11-keV photons at such low count rates. The cross-calibrated photodiode output current was extrapolated to higher currents for the high-flux measurements noted earlier.
A Newport single axis goniometer was mounted on a motorized stage that
was equipped with x and y drives to make the
spindle axis coincident with the x-ray beam. Additional motor drives
were provided on the goniometer head itself, making it possible to remotely position the protein crystals on the goniometer axis. An
Infinity K2 long working-distance microscope (Infinity Phot-Optical Co.) was used to view the sample during alignment. A MSC sample cooler
was operated at a specimen temperature of ~170 K. An ADSC Quantum 4 CCD camera (Area Detector Systems Corp.), borrowed from the ALS
Macromolecular Crystallography Facility, was used to record the
diffraction patterns. Diffraction patterns were indexed, and integrated
intensities were produced with the HKL-DENZO software suite (Otwinowski
and Minor, 1997
).
To demonstrate the principle for further reducing the air-scatter background, nitrogen gas was temporarily replaced by helium gas in the cold-gas stream of a protein-crystal sample cooler (Molecular Structure Corporation) on an R-axis IV diffractometer. To make this replacement, the external gas input line, which is normally used to flush the cooling coils with N2 gas during liquid nitrogen fills, was connected to a helium gas tank. The pressure regulator on the helium tank was adjusted to a value of 10-15 psi. At the same time the pump that is used to generate a steady flow of N2 gas from the liquid nitrogen reservoir was shut off, and the mode selector switch on the cold gas-stream controller was set to "fill," thus allowing the externally supplied helium gas to flow through the liquid nitrogen cooling coils. The gas flow controller gauges for the cold gas and for the outer (sheath) gas streams were set to values of 30 and 15, respectively. We estimate that two-thirds or more of the path length between the tip of the collimator and the beam stop is effectively "purged" by helium gas in this way.
Crystals of the F219L mutant of bacteriorhodopsin, grown by the lipidic
cubic phase technique of Landau and Rosenbusch (1996)
, were used for
our measurements of radiation damage. These crystals were already
available because of a program of experimental data collection that is
currently under way. Wild-type bacteriorhodopsin crystals grow from the
monoolein cubic phase as thin hexagonal plates
(p63; a = b = 60.8 Å, c = 108 Å), which in most cases are ~10-20
µm thick. While crystals will grow as large as 100 µm × 100 µm or more in area, our setups also contain an abundance of additional crystals in all sizes smaller than that.
| |
RESULTS |
|---|
|
|
|---|
The strategy for our measurement of radiation damage was to collect a 10° wedge of data as a series of 10 diffraction patterns, rotating the crystal by 1° for each image recorded on the CCD camera. At the end of such a series, the crystal was returned to its original orientation, and the same 10° wedge was recollected. This process was continued until considerable deterioration of the diffraction pattern became apparent at high resolution. The incident x-ray beam was oriented at ~45° to the crystal face. The small crystal thickness ensures that the decrease in intensity from the front to the back of the crystal, due to absorption, was negligible. Thus the radiation dose (energy deposited per unit volume) was nearly homogeneous throughout the volume of the crystal. The size of the x-ray beam was made slightly larger than the size of the protein crystal to ensure that none of the crystal would rotate outside of the beam for part of the exposure, thus sparing it from damage.
Fig. 2 shows a comparison of the
diffraction pattern recorded in frame 10 and the corresponding pattern
recorded in frame 70. The pattern obtained in frame 10 is shown in Fig.
2 A (central, low-resolution region) and in Fig. 2
C (example of the high-resolution region). Identical regions
of the diffraction pattern recorded in frame 70 are shown in Fig. 2,
B and D, respectively. It is clear that most of
the high-resolution diffraction spots that were originally present in
the first 10° wedge of data have been lost by the time that the
seventh wedge was recorded. The fading of diffraction spots due to
radiation damage is represented in more quantitative terms by the
intensity statistics shown in Table 1.
The number of reflections with measured intensities less than a given
factor times
, the error estimated by the program SCALEPACK (Otwinowski and Minor, 1997
), is shown in this table for different resolution shells. Table 1 A shows the intensity statistics
for the data in the first 10° wedge, and Table 1 B shows
the intensity statistics for data collected in the seventh cycle.
|
|
The same data are again shown in Fig. 3,
this time represented as the number of diffraction spots for which
I/
is greater than 5 for three different resolution
shells. The number of reflections in each shell is shown as a
percentage of the number present in the first 10° wedge of data; the
absolute number of reflections found in the first wedge of data is
shown in parentheses for each resolution shell. This representation of
the data is one in which the overall effects of radiation damage are
easily appreciated. The effects of radiation damage are also apparent
in a graph of the B-factor as a function of accumulated x-ray exposure
(data not shown), but the increase in B-factor is less dramatic than might be supposed because the intensity of the low-resolution diffraction spots also decreases with accumulated exposure, as can be
seen subjectively in Fig. 3.
|
It seems clear that, in most cases, one would not want to collect data
after an exposure of more than 1010
photons/µm2. Thus this value of exposure is
roughly the limit of what can be used to collect data at 3-Å
resolution. A resolution of 3 Å is also about the value obtained in
electron diffraction experiments with two-dimensional bacteriorhodopsin
crystals. The limiting exposure measured in our experiments with
three-dimensional crystals is indeed close to the value of 1.6 × 1010 photons/µm2,
estimated by Henderson (1990)
to be the maximum acceptable exposure on
the basis of electron diffraction data.
As is explained in the Discussion section below, our measurements of radiation damage imply that it would be possible to record at least a single high-resolution diffraction pattern from bacteriorhodopsin crystals whose volume is as small as ~(7 µm)3, containing ~109 unit cells. We have not been able to verify this experimentally, in part because the level of background scattering that existed in the prototype diffraction camera had not yet been optimized and partly because of the long exposure times that would still have been required with our prototype setup.
Data that we were able to obtain with a very small crystal, 25 µm × 25 µm × 10 µm = (18.4 µm)3, as measured with a high-power dissecting microscope, lend credibility to the above estimate, however, and at the same time demonstrate the need for improvements that can be made over the prototype camera. Fig. 4 A shows a 1° rotation photograph obtained with this small crystal with the experimental camera on beam-line 7.3.3, recorded with a 600-s exposure. The x-ray exposure used for this diffraction pattern was estimated to be 3.6 × 108 photons/µm2, and the x-ray beam size was ~50 µm × ~50 µm. It is evident that the air-scattering background is still a major limitation in the current setup. Fig. 4 B shows a similar, 1° diffraction photograph recorded from the same bacteriorhodopsin crystal in 80 s on the protein crystallography camera on beamline 5.0.2 at the ALS, using a 100-µm pinhole collimator. The increase in background, attributable to the less favorable ratio of x-ray beam size to crystal size, is quite conspicuous.
|
A further reduction of background scattering can be achieved simply by using cold helium gas rather than nitrogen gas, but at the same temperature, to cool the protein crystal. This fact was demonstrated on an R-axis IV diffraction camera equipped with the Yale mirrors, for which the x-ray beam size was ~300 µm in diameter. Fig. 5 shows a pair of diffraction patterns obtained with a single bacteriorhodopsin crystal, one photograph taken with nitrogen as the cold gas and the other with helium. The two diffraction patterns were recorded with identical exposure times and with all parameters of the x-ray source held constant. Fig. 5, A (nitrogen) and B (helium), shows the two diffraction patterns with gray scale and threshold parameters that are optimized separately for each image. This comparison underestimates the benefit of using helium gas in the cold stream, however. Fig. 5, C (nitrogen) and D (helium), shows how the two images compare when they are displayed with the same gray scale and threshold values, in this case chosen to not favor one image dramatically over the other. Quantitative measurement of the background in the two cases showed about a sixfold reduction achieved with helium at resolutions below 10 Å, a four- to fivefold reduction at resolutions between 7 Å and 10 Å, a three- to fourfold reduction at resolutions between 4 Å and 7 Å, and about a twofold reduction at resolutions between 2 Å and 3 Å. The relatively stronger reduction in the background scattering at low resolution is due, of course, to the fact that the larger diameter of the nitrogen molecule causes the scattering from N2 to be more strongly concentrated at small angle than is the case for He. If required, even further reduction of the background scattering would be achieved (at all angles) with a more sophisticated, helium-filled enclosure, which would more completely purge the space between the collimator and the beam stop.
|
| |
DISCUSSION |
|---|
|
|
|---|
The maximum x-ray exposure that can safely be given to a frozen
bacteriorhodopsin crystal has been measured to be no greater than
~1010 photons/µm2, if
data are to be collected to a resolution of at least 3 Å. This
measurement agrees very closely with the value of 1.6 × 1010 photons/µm2
predicted by Henderson (1990)
on the basis of radiation damage studies
that had been done earlier in electron crystallography. It is expected
that the radiation dose required to damage a frozen protein crystal
will be the same for x-ray exposures as it is for electron exposures,
of course, because nearly all of the energy deposited in the sample, in
both cases, is the result of inelastic electron-scattering events.
The value of "safe" x-ray exposure that we have measured, however,
is ~10 times lower than that estimated by Gonzales and Nave (1994)
.
(We have used the value of 8 × 1017
keV/mm3 deposited in the sample for a flux of
1011 photons/µm2 to
relate the calculated radiation dose given by Gonzalez and Nave to a
corresponding x-ray flux.) The difference may be attributed in part to
differences in criteria for judging how much exposure can be tolerated.
Damage to the protein exhibits no unique threshold, in that it occurs
progressively with increasing x-ray exposures, and a given degree of
damage occurs much earlier at high resolution than at lower resolution.
As noted by Gonzales and Nave, their value of the x-ray exposure that
could be tolerated would also tend to be an overestimate because of the
way in which they had to calculate the dose delivered by their beam of
"white" radiation. The use of monochromatic radiation and very thin
protein crystals in our experiments is thus an experimental refinement
that was not practical in the earlier measurement made by Gonzales and Nave.
How universal will the limiting, "safe" x-ray exposure prove to be for frozen protein crystals? We can again turn to experience in electron crystallography for an indication of what the answer will be. The maximum exposure that can be used for high-resolution electron diffraction experiments has not been found to differ for various proteins, at least to the accuracy with which it is normally measured. Because damage caused by x-ray exposures is really damage caused by electrons, one can also expect that the maximum tolerated x-ray exposure will not vary from one sample to another. An important caveat, however, is that the x-ray intensity must be kept low enough to prevent an appreciable rise in specimen temperature during the exposure. If x-ray flux densities are high enough to warm the sample, one can expect that damage will occur at even lower exposures than are reported here. Gonzales and Nave report, however, that there was no detectable dose-rate effect for beam intensities high enough to damage the sample after only a few minutes of exposure.
Padmore et al. (manuscript in preparation) have used heat-transfer calculations to further define the conditions under which sample heating can safely be ignored. For any currently practical rate of x-ray exposure, these calculations make it clear that the thermal impedance at the convective boundary between the frozen sample and the cold-gas stream determines how much heating will occur, while the thermal impedance of the sample itself (modeled as being pure ice) makes a negligible contribution to the rise in specimen temperature. For a ~12-keV x-ray flux density of 109 photons/s-µm2, high enough to limit the working time (total allowed exposure time) to ~10 s, one can expect the temperature of a 20-µm-diameter crystal to increase by less than 2 K from the temperature of the cold nitrogen gas, while the temperature of a 200-µm sample would increase by less than ~8 K. The latter value is probably closer to the correct estimate than the former, for small microcrystals if they are mounted within a 100-µm loop, surrounded by mother liquor. The amount of temperature rise can be further reduced by a factor of ~3 relative to the values quoted above if helium gas, cooled to ~100 K, is used in the cold stream. It is clear from these model calculations that specimen heating, which scales as the absorbed power, could become an important consideration if appreciably greater flux densities were used. However, it will rarely be the case that one would want to deliver a fully damaging exposure in times shorter than ~10 s.
Personal experience has frequently been described to us by others that
seems not to be consistent with the belief that radiation damage for
frozen specimens must be independent of the type of crystal being
studied. We urge that two conditions should be met before coming to
such a conclusion. 1) Quantitative values of the x-ray flux density
should be known when the crystal lifetimes observed in one situation
are compared to those observed in another situation. 2) The rate at
which the protein is destroyed should be measured in units of
accumulated x-ray exposure (e.g., photons/µm2)
and not in units of the number of frames of data that could be
recorded. The number of frames of data that can be collected, while of
practical importance, will vary in the obvious way with unit cell size,
which determines the strength of diffraction intensities. In addition,
there is a danger
if quantitative values of the flux density are not
known
that much longer exposures than necessary are used when the flux
density is extremely high, resulting in many fewer frames of data being
recorded before severe damage is observed.
Our experience when collecting full data sets from the largest
available crystals of bacteriorhodopsin suggests that a crystal must
contain at least 1011 unit cells to record ~100
high-resolution diffraction patterns at 2-Å resolution, using 1° of
rotation for each photograph. The number of high-resolution diffraction
patterns that can be recorded for a given x-ray exposure will depend on
the signal-to-noise ratio (i.e., I/
) that one is willing
to accept, of course. The measured I/
, in turn, will
depend upon how high the background scattering is, the degree of
short-range order in the crystal, and the mosaic spread or other
factors that determine the ratio of diffraction spot size to detector
pixel size. Nevertheless, we suggest that a value of
109 unit cells is a moderately conservative
estimate for the smallest crystal that can be used to record just a
single diffraction pattern at 2-Å resolution.
Taking the value of 109 unit cells as the
smallest useful crystal size leads, in turn, to practical rules of
thumb that can be used to estimate how big a protein crystal should be
to record high-resolution diffraction data. The effective crystal size, L, can be expressed as the cube root of the crystal volume,
and the effective unit cell size, a, can be expressed as the
cube root of the unit cell volume. If L is then measured in
micrometers and a is measured in Angstroms, a crystal of
size
|
(1) |
|
(2) |
The rule of thumb written in Eq. 2 follows from that written in Eq. 1,
if one recognizes that the integrated diffraction intensity scales as
the number of unit cells in the crystal. It is obvious that this should
be the case, because the scattering cross section of the sample, i.e.,
the integrated diffraction intensity, must scale as the number of atoms
in the sample and thus as the number of unit cells in the crystal. The
same conclusion can also be derived from the familiar expression for
the integrated intensity, which states that
|
(3) |
Reduction in background scattering is an important, practical consideration when protein crystals become much smaller than about (40 µm)3. We have found a 100-µm pinhole collimator to be satisfactory for bacteriorhodopsin crystals as small as 70 µm × 70 µm × 15 µm, but the background scattering proved to be overwhelming when the crystal size was only 25 µm × 25 µm × 10 µm. For crystals in this size range or smaller it will be possible to use adjustable microfocusing in the vertical direction and adjustable slits in the horizontal direction to reduce the x-ray beam size to ~25 µm in diameter, ~1/4 the area of that used in Fig. 4 A. In addition, helium rather than nitrogen can be used in the cold gas stream. The combined effect would provide a large reduction in the background that is derived from air scattering. Some of the background scatter in Fig. 4 B comes from the surrounding mother liquor, however, and from the cryo-loop that is used to hold the crystal. Even these contributions could also be reduced by mounting very small crystals on thin fibers or on very thin films of the type used to support much smaller protein microcrystals in electron cryomicroscopy.
| |
ACKNOWLEDGMENTS |
|---|
We thank the ALS Macromolecular Crystallography Facility for use of the CCD detector, and we are grateful to Dr. Carl Cork and Dr. Gerry McDermott for assistance in the setup and calibration of the detector. We are pleased to thank Tom Alber, James Berger, Thomas Earnest, and Bing Jap for discussing various aspects of this work.
This work was initiated with support from the LBNL Laboratory Director's Research and Development Fund. Additional support was provided by National Institutes of Health grant GM51487 and by the U.S. Department of Energy, Office of Basic Energy Sciences, under contract DOE-AC03-765F00098.
| |
FOOTNOTES |
|---|
Received for publication 20 September 1999 and in final form 13 February 2000.