During micropipette aspiration, neutrophil leukocytes
exhibit a liquid-drop behavior, i.e., if a neutrophil is aspirated by a
pressure larger than a certain threshold pressure, it flows continuously into the pipette. The point of the largest aspiration pressure at which the neutrophil can still be held in a stable equilibrium is called the critical point of aspiration. Here, we
present a theoretical analysis of the equilibrium behavior and
stability of a neutrophil during micropipette aspiration with the aim
to rigorously characterize the critical point. We take the energy
minimization approach, in which the critical point is well defined as
the point of the stability breakdown. We use the basic liquid-drop
model of neutrophil rheology extended by considering also the
neutrophil elastic area expansivity. Our analysis predicts that the
behavior at large pipette radii or small elastic area expansivity is
close to the one predicted by the basic liquid-drop model, where the
critical point is attained slightly before the projection length
reaches the pipette radius. The effect of elastic area expansivity is
qualitatively different at smaller pipette radii, where our analysis
predicts that the critical point is attained at the projection lengths
that may significantly exceed the pipette radius.
 |
INTRODUCTION |
Micropipette aspiration is an important
experimental method in the research of the rheology of resting
neutrophil leukocytes. In this experiment, one studies the behavior of
a single neutrophil as it is controllably aspirated into a micropipette
narrower than its diameter. A commonly observed behavior of neutrophils
during the micropipette experiment is the following (Evans and
Kukan, 1984
): In the free state, a resting neutrophil is
spherical. If a small aspiration pressure is applied, the neutrophil
projects into the pipette and attains an equilibrium. In contrast, if
the applied aspiration pressure is above a certain threshold, the so-called critical pressure, the neutrophil does not reach an equilibrium but flows continuously into the pipette. Finally, if the
neutrophil is released from the pipette, it recovers its initial
spherical shape.
Evans and Kukan (1984)
were the first to recognize that
this behavior can be explained by a liquid-drop nature of neutrophils. They introduced the now generally accepted liquid-drop model of neutrophil viscoelasticity, in which the neutrophil is regarded as a
Newtonian fluid drop with a constant surface tension. The neutrophil
surface tension has been related to the neutrophil membrane cortex
(Evans and Kukan, 1984
), therefore it is usually denoted
as the cortical tension. A general discussion on the
validity of the basic liquid-drop model and a good basis for further
research of dynamics of micropipette aspiration was given recently by
Drury and Dembo (1999)
.
Evans and Yeung (1989)
recognized that the liquid-drop
model alone could not explain all neutrophil features and that the neutrophil could exhibit also some elastic behavior, as had been proposed in an earlier study by Schmid-Schönbein et al.
(1981)
. Many attempts were made to combine the liquid and the
elastic nature of neutrophils. Some of the extended models attributed elasticity to the whole neutrophil (Dong and Skalak,
1992
), whereas others focused more on the elasticity of the
membrane cortex. For example, the existence of a nonconstant cortical
tension was investigated (Needham and Hochmuth, 1992
),
and the importance of cortex bending rigidity was examined
(Zhelev et al., 1994
). However, it has not been shown
that any of these refined models completely describes both the dynamic
(i.e., the continuous flow) and the static equilibrium behavior of neutrophils.
The micropipette experiment has also been used in the research of the
possible underlying molecular mechanisms that govern neutrophil
mechanical properties. For example, research was carried out to
establish the dependence of neutrophil mechanical rigidity on the
polymerization rate of its actin filaments (Tsai et al., 1994
) and microtubules (Tsai et al., 1998
).
Because these experiments rely on the quantitative measurements of the
neutrophil viscoelastic parameters, a good theoretical characterization
of these parameters is of great importance. One of the neutrophil
viscoelastic parameters, its cortical tension, is normally determined
by measuring the critical aspiration pressure. However, the notion of
the critical pressure in the literature is ambiguous. It is usually
based on the prediction of the basic liquid-drop model and the
approximation that the critical pressure is attained when the
neutrophil projection length in the pipette reaches the pipette radius
(Yeung and Evans, 1989
). This supposition should be
applied carefully, though. Because the critical pressure is the maximal
possible aspiration pressure at which the cell can still be held in a
static equilibrium, the equilibrium projection lengths longer than the
pipette radius are unstable and should not be observed. However, in
fact, it seems that the stable equilibrium of the aspirated neutrophil could be observed even for projection lengths much larger than the
pipette radius (Evans and Yeung, 1989
; Zhelev and
Hochmuth, 1994
). Although such behavior cannot be explained
within the basic liquid-drop model, it can be intuitively expected as a
consequence of neutrophil elasticity.
In this paper, we present a theoretical analysis of the critical
aspiration pressure behavior, based on the basic liquid-drop model
extended by also considering neutrophil elasticity. Specifically, we
are taking into account the nonconstant cortical tension as it was
observed by Needham and Hochmuth (1992)
. Accordingly,
the neutrophil is considered as a liquid drop with an apparent elastic area expansivity modulus. We focus on the equilibrium behavior of the
aspirated neutrophil and its stability. In this way, the critical
aspiration pressure can be identified as the aspiration pressure at
which stability of the aspirated neutrophil breaks and the neutrophil
starts to flow into the pipette. Consequently, the point of the
stability breakdown, i.e., the critical point of aspiration, can be
well characterized and the corresponding values of the critical
aspiration pressure and the critical projection length can be
calculated. Our analysis provides the predictions for the critical
projection length and the critical aspiration pressure and can
therefore be used to better evaluate the measurements of the neutrophil
cortical tension. In addition, the analysis can serve as a basis for
further measurements of the apparent neutrophil elastic area
expansivity modulus. The paper is organized as follows. We will first
discuss the significant terms of neutrophil energy that govern
neutrophil equilibrium behavior. Then we will apply the variational
approach to calculate the equilibrium states of the aspirated
neutrophil and determine stability of these equilibrium states.
Finally, we will present the predictions for the position of the
critical point of aspiration.
 |
THEORY |
Model: liquid droplet with an elastic surface
To understand the limitations of neutrophil mechanical properties
models, one should bear in mind the complex structure of neutrophils.
Free resting neutrophils are wrinkled spherical cells of ~8 µm in
diameter (Bessis, 1973
; Ting-Beall et al.,
1993
). When a neutrophil is deformed from the spherical shape,
the wrinkles are smoothing out and the apparent neutrophil area
enlarges. The wrinkles provide enough material for extensions up to
approximately two times the initial spherical area (Evans and
Yeung, 1989
). The investigation of the neutrophil interior
structure reveals many small granules and the cell nucleus floating in
the neutrophil cytoplasm. These organelles were estimated to comprise
around 37% of the total cell volume (Schmid-Schönbein et
al., 1980
). Another important neutrophil structure is the
membrane cortex, which supports the outer plasma membrane. It is
composed mainly of actin filaments meshwork and has an approximate
thickness of 0.1-0.2 µm (Sheterline and Rickard,
1989
; Zhelev et al., 1994
).
The analysis in the present paper is based on the basic liquid-drop
model extended by including the neutrophil elastic area expansivity.
The neutrophil is thus considered as an incompressible homogeneous
fluid with a cortical tension, which increases with the apparent
neutrophil surface area dilation (Needham and Hochmuth, 1992
). Within this model, the equilibrium behavior of an
aspirated neutrophil is governed exclusively by its cortical
energy, i.e., the energy of its membrane-cortex complex. In a
general case, this might be an oversimplification, but as long as the
deformations are not too large and the organelles do not touch the
membrane, and because we are focusing only on the equilibrium states of aspiration, this assumption is justified. In addition, during the
normal course of micropipette aspiration, the adhesion of resting
neutrophils to the pipette glass is negligible (Evans and Kukan,
1984
; Needham and Hochmuth, 1992
) and can be,
therefore, safely omitted from the analysis. Also, in our model, we do
not consider the bending rigidity of the cortex, which may become important in the case of aspiration into very narrow pipettes (Zhelev et al., 1994
).
The exact microscopical origin of the cortical tension is not known,
however, evidences exist that it is related to the actin filaments in
the cortex (Tsai et al., 1994
; Tsai et al.,
1998
). For example, the tension could be a consequence of some
active processes, which constantly drive the actin filaments one across another and contract the cortex. Another imaginable origin of one part
of the constant cortical tension could be attributed to a possible
difference of the surface energy of the lipid bilayer stocked in the
wrinkles and the bilayer stretched on the actin cortex (Svetina
et al., 1998
)
if the bilayer preferred to reside in the
wrinkles, the result would be a contractile cortical tension. Needham and Hochmuth (1992)
indicated that the cortical
tension might change with the activation of the neutrophil.
Because the exact physical origin of the tension is not known, the
cortical energy has to be described phenomenologically. To include the
nonconstant cortical tension, the cortical energy is expanded up to the
second power in the neutrophil area,
|
(1)
|
where
is the constant part of the cortical tension,
A is the neutrophil area, K is the area
expansivity modulus, and A0 is taken to be the
area of the free spherical neutrophil. The value of
A0 is fixed with the neutrophil volume
V0, which is constant during a typical
aspiration under stable osmotic conditions. The quadratic term in Eq. 1
can be interpreted as the neutrophil elastic area expansivity term.
Indications exist that the parameters
and K may depend
on the history of aspiration, i.e., the increase of the rate of actin
polymerization in the cortex may be induced by the applied stress
(Zhelev and Hochmuth, 1994
).
The effective cortical tension associated with the cortical energy
defined above is
|
(2)
|
For positive constants
and K, the cortical tension
is also positive and thus tends to minimize the cell area. When the neutrophil is undeformed, A = A0, the
cortical tension has the value of
. This definition of the
parameters K and
is consistent with the one proposed by
Needham and Hochmuth (1992)
.
To emphasize that our analysis applies to very idealized neutrophils,
we will use the term "elastic droplet" when referring to the
neutrophil within the extended liquid drop model, and "liquid droplet" when referring to the basic liquid-drop model.
Equilibrium states
Thermodynamically, the equilibrium of an aspirated droplet is
characterized as the state of the extreme thermodynamic potential of
the system. Within the presented model of micropipette aspiration, the
work done by aspiration is entirely transformed into cortical energy,
and the thermodynamic potential of an aspirated droplet can be written
as
|
(3)
|
where
P is the aspiration pressure and
Vp is the droplet volume aspirated into the
pipette. The aspiration pressure,
P, is defined as the
difference between the pressure of the solution outside the pipette and
the pressure in the pipette and is positive during aspiration.
The extrema of the thermodynamic potential defined in Eq. 3 can be
found by means of the calculus of variations. The extrema (i.e., the
equilibrium states of an aspirated droplet) are the states where the
variation of the thermodynamic potential with respect to all variables
and to the shape of the droplet is zero,
G = 0. It
turns out that the equilibrium shapes of an elastic droplet are the
same as in the case of a liquid droplet. This can be seen from the fact
that the variation of the potential
G can be written in
the same form as in the basic liquid drop model:
|
(4)
|
The variation of G depends on the effective cortical
tension. The variational procedure sees no difference between the
constant tension and the effective one, because, in both cases, the
tension is just a parameter of a given equilibrium state. Therefore,
the equilibrium states of an elastic droplet are exactly the same as
the equilibrium states of a liquid droplet.
Thus, two general results concerning the equilibrium states of a liquid
droplet, widely used in the literature, are valid also in the case of
an elastic droplet (the full derivation of these results is given in
the Appendix A):
| 1. |
The equilibrium shapes of an aspirated droplet are composed
of spherical and cylindrical parts (Fig.
1). The free droplet is a sphere with the
radius R0. As aspiration proceeds, the droplet projection in the pipette grows, the radius of the spherical part in
the pipette Rin decreases and the projection
length Lp increases. When
Rin (and Lp) reaches the
pipette radius Rp, the projection continues to
grow as a cylinder with a hemispherical cap of the same radius as the
pipette. The radius of the spherical part of the droplet outside the
pipette is denoted by Rout.
|
| 2. |
The equilibrium aspiration pressure is related to the cortical
tension by the law of Laplace,
|
(5)
|
|
| |
where is the effective cortical
tension given by Eq. 2.
|

View larger version (12K):
[in this window]
[in a new window]
|
FIGURE 1
Schematic representation of the droplet equilibrium
shape during aspiration into a micropipette. (A) The
projection length is less than the pipette radius and the shape is
composed of two spherical parts; (B) The projection length
exceeds the pipette radius and the projection is composed of a cylinder
with a hemispherical cap of the same radius as the pipette. The radius
of the outer spherical part is Rout, the radius
of the part in the pipette is Rin, the droplet
projection length in the pipette is Lp, and the
radius of the pipette is Rp. Note that, if
Lp Rp, the value
of Rin equals the pipette radius,
Rin = Rp.
|
|
Because of the simple geometry of the equilibrium states, the
overall droplet equilibrium shape is fully defined by only two geometrical variables, the radius Rout and the
projection length Lp. In terms of these two
variables, the equilibrium area and volume of a droplet are expressed
as
|
(6)
|
|
(7)
|
Furthermore, because the volume of the droplet during aspiration
is constant, Rout and Lp
are connected by the constant volume constraint V = V0. Therefore, only one geometrical variable is truly
independent, and the entire set of the equilibrium shapes can be
obtained and examined by varying only one geometrical variable. The
computational procedure to calculate the equilibrium states used in our
analysis was the following: for a given value of
Rp and Rout, the constant
volume constraint was used to calculate the corresponding
Lp. The corresponding value of
Rin was obtained from the geometrical relations
connecting Rin and Lp:
|
(8)
|
Then, the equilibrium aspiration pressure was determined by using
the law of Laplace (Eq. 5). Finally, Eq. 3 was applied to calculate the
corresponding thermodynamic potential. At a given pipette radius, the
whole set of the possible equilibrium states was obtained by varying
Rout from Rout = R0 (which corresponds to the free spherical
droplet at zero aspiration pressure,
P = 0) to
Rout = Rp (the
droplet is aspirated completely to a cigar shape and the corresponding
equilibrium aspiration pressure is again zero,
P = 0).
The radius Rin is a smooth function of the
projection length Lp. However, because there are
two distinct regimes of the projection growth in the pipette, the
function Rin(Lp) has a
discontinuous second derivative in the point where
Lp reaches Rp. As a
consequence, the same is true also for the equilibrium aspiration
pressure
P(Lp).
Introduction of the relative dimensionless quantities
Because the choice of the scale within the presented model does
not affect the behavior of the system, the analysis can be simplified
by the normalization of the involved quantities. The natural way to
normalize these quantities is to introduce relative dimensionless
quantities that have been scaled by the initial spherical size of the
droplet and the constant part of the cortical tension
. These
relative dimensionless quantities will be written in lowercase. The
radii and the projection length are normalized relative to the radius
of the initial sphere (rp = Rp/R0,
rin = Rin/R0,
rout = Rout/R0, and
lp = Lp/R0), the areas
relative to the initial droplet area (a = A/A0, and a0 = 1), and the
volume relative to the constant droplet volume
(vp = Vp/V0). In the
dimensionless form, all possible ratios between the pipette radius and
the neutrophil radius are obtained by varying the relative pipette
radius rp from 0 to 1.
The relative thermodynamic potential and relative aspiration pressure
are
|
(9)
|
|
(10)
|
The values in the parentheses represent typical energy and
pressure scales from the experiments. They are obtained by using a
typical neutrophil size (with R0 = 4 µm)
and a typical value of the cortical tension
= 30 µN/m; the
measured values of
lie in the range from 24 µN/m (Needham
and Hochmuth, 1992
) to 35 µN/m (Evans and Yeung,
1989
).
Using the relative dimensionless quantities, the thermodynamic
potential is written as
|
(11)
|
Note that the only parameter of the system material properties is
the ratio between the area expansivity modulus and the constant part of
the cortical tension K/
. The basic liquid-drop model is
obtained if this parameter is set to zero (K/
= 0), whereas the measured value of the ratio is on the order of unity (K/
~ 1), (Needham and Hochmuth,
1992
).
Stability of equilibrium states
Not all the equilibrium states of an aspirated droplet calculated
by the procedure described above are in a stable equilibrium. Thermodynamically, the stable equilibrium states are only the states of
the minimal thermodynamic potential, whereas the states of the maximal
thermodynamic potential are in an unstable equilibrium. If a state was
in an unstable equilibrium, a small fluctuation would drag it away to a
nearby stable equilibrium.
By choosing the relative projection length
lp as the only free geometrical variable of the
equilibrium states, the stability condition for the equilibrium states
of an aspirated droplet can be mathematically expressed as
|
(12)
|
It is possible and useful to translate this stability condition
into a condition connecting the observable quantities
p and lp. The stability condition can be thus
written as (for a detailed derivation see Appendix B)
|
(13)
|
Intuitively, this is a reasonable condition, analogous to the fact
that the bulk modulus in the standard thermodynamics is always
positive. If, for an equilibrium state, one needed to decrease the
aspiration pressure to increase the projection length, such a state
would be unstable.
 |
RESULTS |
Critical point
The critical point of micropipette aspiration can be characterized
as the point of the equilibrium state's stability breakdown. It can be
analyzed by considering the relation between the equilibrium aspiration
pressure
p and the corresponding projection length lp (via the law of Laplace, Eq. 5), and by using
the stability condition (Eq. 13). The dependence of the equilibrium
projection length on the aspiration pressure during a typical course of
aspiration is presented in Fig. 2. At the
beginning of aspiration, the projection length increases with
increasing aspiration pressure. These states are in a stable
equilibrium because the derivative
d
p/dlp is positive. The point
where the equilibrium aspiration pressure reaches its maximum (i.e.,
where d
p/dlp = 0) is the
critical point of aspiration. The corresponding aspiration pressure and projection length are the critical aspiration pressure
pcrit and the critical projection length
lp(crit). If the pressure is increased above the
critical pressure, the droplet starts to flow continuously into the
pipette and therefore enters the dynamic regime, which is not covered
by our analysis. The equilibrium states, with projection lengths
exceeding the critical projection length, have a negative
d
p/dlp and are in an unstable
equilibrium.

View larger version (9K):
[in this window]
[in a new window]
|
FIGURE 2
The ratio between the equilibrium projection length and
the pipette radius as a function of the relative aspiration pressure
for five different ratios K/ and at the relative pipette
radius rp = 0.5. (A) For
complete course of aspiration and (B) enlarged in the region
where the critical projection length is close to the pipette radius,
lp rp. The
values of the ratio are, (a) K/ = 0, (b) 0.5, (c) 1, (d)
2, and (e) 4. The stable states and unstable states are
represented by solid and dashed lines respectively. The point of the
maximal equilibrium aspiration pressure is the critical point. The
function p(lp) is smooth in the whole
interval, but has a discontinuous second derivative in the point where
the projection length becomes equal to the pipette radius. The relative
unit of the aspiration pressure corresponds to 23 Pa.
|
|
As can be seen from Fig. 2, the elasticity of the droplet strongly
affects the position of the critical point. For small ratios K/
, the critical projection length is near the pipette
radius, but it can increase significantly with higher K/
.
Note that, in the case of K/
= 1 and
rp = 0.5 (curve c in Fig. 2),
only a small aspiration pressure difference is needed to aspirate the droplet from lp = rp
to the critical projection length lp = lp(crit)
3rp.
The stability of the equilibrium states can be examined further by
considering the values of the equilibrium thermodynamic potential (Eq. 11). As expected, the unstable equilibrium states are at a higher
thermodynamic potential than the stable equilibrium states (Fig.
3). A droplet in an unstable equilibrium
state would either slightly withdraw from the pipette and reach the
stable equilibrium state at the same pressure or enter the continuous flow regime. Furthermore, because of the fluctuations, spontaneous transitions from a stable equilibrium state (over the maximum pertaining to the unstable equilibrium states) to the dynamic regime at
a fixed pressure could be theoretically possible before the critical
point. However, the energy barrier to be overcome is several orders of
magnitude larger than the thermal energy kT (compare the
energy scale in Fig. 3 and Eq. 9), so the spontaneous transitions to
the continuous flow regime at aspiration pressures less than the
critical are not likely to occur.

View larger version (13K):
[in this window]
[in a new window]
|
FIGURE 3
Thermodynamic potential g versus aspiration
pressure p for the equilibrium states at the relative
pipette radius rp = 0.5 and the elasticity
parameter K/ = 1. The solid line represents the
stable states and the dashed line the unstable ones. The level of the
aspiration is indicated schematically at four characteristic points.
The states aspirated beyond the critical point are in an unstable
equilibrium and are on a higher thermodynamic potential than the stable
equilibrium states.
|
|
Critical projection length
The position of the critical point can be determined by finding
the maximal possible equilibrium aspiration pressure and the corresponding projection length. Because the analytical relations connecting the geometrical variables are complicated, we calculated the
maximal equilibrium aspiration pressure numerically. The obtained predictions for the critical projection length as a function of the
relative pipette radius at five ratios K/
are presented
in Fig. 4. In the case of a liquid
droplet (K/
= 0), the critical projection length is
always a few percent smaller than the pipette radius (this result was
already calculated by Drury and Dembo, 1999
). However,
if the droplet is elastic, K/
0, the critical projection length increases significantly at small pipette radii.

View larger version (10K):
[in this window]
[in a new window]
|
FIGURE 4
The ratio between the critical projection length and
the pipette radius as a function of the relative pipette radius for
five values of the elasticity parameter K/ . Values of
K/ are the same as in Fig. 2. (A) Full scale
and (B) enlarged in the region where the critical projection
length is close to the pipette radius,
lp(crit)/rp 1. If K/ 0, the critical projection length at small
pipette radii exceeds the pipette radius. The critical projection
lengths are minimal around rp 0.92 for
all five values of K/ . The dashed line represents
projection lengths of a fully aspirated droplet and thus presents the
geometrical limit for critical projection lengths. The dotted line
represents projection lengths when the droplet area is increased to two
times the initial area of the spherical droplet and thus presents the
approximate practical limit of aspiration when all the membrane bilayer
is smoothed out (Evans and Yeung, 1989 ).
|
|
This behavior can be explained. If the cortical tension is constant, as
it is in the case of a liquid droplet, the equilibrium aspiration
pressure is proportional directly to the difference of curvatures of
the inner and the outer spherical part (a curvature is an inverse of a
radius), 1/rin
1/rout (Eq. 5). At the beginning of aspiration,
the inner curvature (1/rin) increases more
rapidly with lp than the outer one
(1/rout) and
p increases
(d
p/dlp > 0). When
lp is near rp, however,
p decreases
(d
p/dlp < 0). That is
because, when lp is near
rp, the inner curvature smoothly approaches its
final constant value 1/rp and thus begins to
increase more slowly with lp than the outer one.
It follows that the maximum equilibrium pressure
the critical
pressure
is attained before lp reaches
rp.
In the case of an elastic droplet, the cortical tension increases as
the droplet area increases during aspiration. The greater cortical
tension resists aspiration more strongly and the critical point of
aspiration is only attained at larger projection lengths. Once the
critical projection length becomes larger than the pipette radius, its
dependence on the pipette radius (and also on the ratio
K/
) is much stronger. As a consequence of the two
different regimes of the projection growth, the functions
lp(crit)(rp) have a
discontinuous first derivative in the point where
lp(crit) = rp. Note
that the elasticity only weakly affects the smallest possible value of
the critical projection lengths, which is around rp
0.92 for all reasonably expected
values of the ratio K/
(Fig. 4 B).
Whether the critical length exceeds or does not exceed the pipette
radius depends both on the ratio K/
and the relative
pipette radius. The values on the boundary between the two cases are
presented in Fig. 5. At smaller pipette
radii, even a small K/
has a significant effect on the
elongation of the critical projection length. In contrast, if the
relative pipette radius is larger than 0.9164, even extreme elasticity
(K/
) cannot significantly affect the position of
the critical point, and the critical projection length is always less
than the pipette radius. In the case of K/
~ 1,
which corresponds to the observed values of the parameter K/
(Needham and Hochmuth, 1992
), the
critical projection length exceeds the pipette radius for the relative
pipette radii under ~0.6. Therefore, the elongation of the critical
projection lengths above the pipette radius could be detected
experimentally.

View larger version (8K):
[in this window]
[in a new window]
|
FIGURE 5
The limiting values of the ratio K/ and
rp of the two regimes of behavior of the elastic
droplet. At large pipette radii and small ratios K/ the
droplet behaves as a liquid droplet with the critical projection length
slightly smaller than the pipette radius. At small pipette radii and
high ratios K/ , the critical projection lengths are
larger than the pipette radius. The vertical dotted line represents the
asymptotic value of rp 0.9164, beyond
which the critical projection length is less than the pipette radius
even in the limit K/ .
|
|
The measurements of the position of the critical point should be
carried out carefully. The reason for this is that the equilibrium projection lengths near the critical point can depend strongly on the
aspiration pressure (for the case of the relative pipette radius
rp = 0.5 and K/
= 1, see
Fig. 2). Thus, considering the limited precision of the aspiration
pressure set up in current experiments, it might be difficult to
measure the exact value of the critical projection length at all
relative pipette radii. In addition, the noninstantaneous response of
the neutrophil to the change of the aspiration pressure should be
considered. Therefore, to avoid missing the critical point, approaching
toward the critical point through the equilibrium states by increasing
the aspiration pressure should be performed slowly. Or, alternatively,
the procedure described by Zhelev and Hochmuth (1994)
can be used, in which the neutrophils were first driven in the
continuous flow regime and then relaxed to the equilibrium by a
decrease of the aspiration pressure.
Critical aspiration pressure
The approximation (based on the basic liquid-drop model) that the
critical point is reached when the projection length reaches the
pipette radius, leads to the estimation that the critical pressure of
aspiration is an inverse function of the pipette radius (Evans
and Yeung, 1989
). Our numerical calculation of the critical pressure as the function of the pipette radius for K/
= 0 shows that this estimation is accurate to a few percent. Even
for the pipette radius of 0.92, where the critical projection length
deviates most from the pipette radius (Fig. 4 B), the
difference is only about 15%. In the case of an elastic droplet, the
critical pressure increases with K/
, but the general
inverse relation to the pipette radius is conserved (Fig.
6). However, it should be stressed once more that, when K/
0 (curves b, c,
d, and e in Fig. 6), the critical pressures can occur
at larger projection lengths and can therefore not be directly compared
to the existing experimental data, where the critical pressure was
generally reported to be measured at lp = rp.

View larger version (12K):
[in this window]
[in a new window]
|
FIGURE 6
The relation between the relative critical pressure and
the relative pipette radius for five values of the elasticity parameter
K/ . Values of K/ are the same as in Fig. 2.
The critical pressure increases with increasing K/ .
|
|
The imprecise definition of the critical aspiration pressure used in
the literature should not affect the measurements of the cortical
tension. That is because the law of Laplace, which is used as the
relation between the aspiration pressure and the effective cortical
tension, is valid in all the equilibrium states and not only at the
critical point. So, if the law of Laplace is strictly used, the
cortical tension can be safely determined by measuring any of the
equilibrium aspiration pressures.
 |
CONCLUSION |
In this paper, we theoretically analyzed the stability of the
equilibrium states of micropipette aspiration with the aim to rigorously characterize the critical point of aspiration. Our analysis
shows that the critical point of aspiration does not necessarily occur
when the projection length in the pipette becomes equal to the pipette
radius. Specifically, it shows that, if the neutrophil
membrane-cortex complex has an elastic area expansivity, the neutrophil
can be held in a static equilibrium even if its projection length in
the pipette exceeds the pipette radius. This feature was, in fact,
observed (see, for example, Fig. 4 in Evans and Yeung,
1989
or Fig. 9 in Zhelev and Hochmuth, 1994
),
but could not be described within the basic liquid-drop model of
neutrophil rheology.
We emphasize that the notion "critical aspiration pressure"
has to be used with caution. Because the critical pressure is defined
as the largest possible aspiration pressure at which the neutrophil can
be held in a static equilibrium, it should not be generally attributed
to the aspiration pressure, which is attained when the projection
length reaches the pipette radius. Nevertheless, the imprecise
definition of the critical pressure should normally not affect the
measurements of the effective cortical tension. As long as the law of
Laplace is strictly used, the cortical tension can be reliably
determined by the measurements of the equilibrium aspiration pressure.
Finally, we suggest that the presented analysis serves as a basis for
further measurements of the neutrophil elastic properties. Although the
exact positions of the critical point of aspiration at all relative
pipette radii might be difficult to obtain, a comparison between the
presented predictions and the measured critical projection length as a
function of the relative pipette radius can yield new information on
the values of the neutrophil area expansivity modulus K and
the constant cortical tension
.
If experiments reveal that the equilibrium neutrophil behavior differs
significantly from the predictions of the presented analysis, the
simple neutrophil elastic area expansion rigidity could not be the main
neutrophil elastic property and at least two further options for the
neutrophil elasticity should be investigated. The first possibility is
that the neutrophil cytoplasm has significant elastic rigidity. The
second possibility is that the cortical tension is not only a function
of the area dilation, but depends also on the history of aspiration,
i.e., the cortical tension is governed by some active processes that
can be triggered by the applied stress.
In this appendix, we present a rigorous calculation of the
equilibrium states of an aspirated droplet. The equilibrium states are
obtained by the minimization of the system's thermodynamic potential.
In the case of the model used in this paper, the result could be also
derived from the force-balance equations. However, the energy
minimization approach enables us to perform the stability analysis of
the equilibrium states.
The equilibrium states of an aspirated droplet are the extrema of the
thermodynamic potential defined in Eq. 11. Because of the fixed droplet
volume constraint, the extrema of the thermodynamic potential
correspond to the stationary points of the functional
In the stationary point, the variation of the functional with respect
to shape is zero,
g' = 0. Written out, the variation of
g' is
The aspirated droplet is axisymmetric, therefore we can parametrize the
droplet shape in terms of its axisymmetric contour and express
the functional in terms of an integral over this contour. We get a
classical problem of the calculus of variations; the stationary shapes
correspond to the solutions of the Euler-Lagrange equations derived
for our system (for a general reference on the calculus of variations
see Elsgolc (1961)
, and, for the calculus of variations
applied to the calculation of equilibrium shapes of closed lipid
membranes, see Jülicher and Seifert (1994)
and Bo
i
et al. (1997)
).
The standard parameterization of a closed axisymmetric surface is done
via coordinates r(s), z(s), and
(s), where
r is the radial distance of the contour to the axis of
symmetry, z is the position along the axis,
is the
contour angle, and s is the arclength along the contour
(Fig. 7). The angle of the contour
(s) is defined by the equation tan
= dz/dr.
The variational procedure can be carried out in a relatively simple way
by using the Hamilton notation. We therefore define