We report the application of pump-probe fluorescence microscopy in time-resolved polarization imaging. We derived the equations governing the pump-probe stimulated emission process and characterized the pump and probe laser power levels for signal saturation. Our emphasis is to use this novel methodology to image polarization properties of fluorophores across entire cells. As a feasibility study, we imaged a 15-µm orange latex sphere and found that there is depolarization that is possibly due to energy transfer among fluorescent molecules inside the sphere. We also imaged a mouse fibroblast labeled with CellTracker Orange CMTMR (5-(and-6)-(((4-chloromethyl)benzoyl)amino)tetramethyl-rhodamine). We observed that Orange CMTMR complexed with gluthathione rotates fast, indicating the relatively low fluid-phase viscosity of the cytoplasmic microenvironment as seen by Orange CMTMR. The measured rotational correlation time ranged from ~30 to ~150 ps. This work demonstrates the effectiveness of stimulated emission measurements in acquiring high-resolution, time-resolved polarization information across the entire cell.
| |
INTRODUCTION |
|---|
|
|
|---|
Fluorescence microscopy in biology
Fluorescence microscopy allows imaging of
specific cellular microstructures in great detail. There have been
different developments aimed at improving the spatial resolution in
microscopic imaging by reducing off-focal fluorescence. In particular,
confocal microscopy (Wilson and Sheppard, 1984
;
Wilson, 1990
) improves the spatial resolution to the
submicron range. In two-photon fluorescence microscopy, similar spatial
resolution and off-focal fluorescence reduction can be achieved by
inducing nonlinear excitation of the fluorescent sample (Denk et
al., 1990
). Compared to confocal microscopy, the localization
of excitation to the focal volume attained by two-photon excitation
significantly reduces photobleaching in off-focal regions. Although
spatial resolution much less than the wavelength of light can be
obtained with near-field microscopy (Betzig et al.,
1993
), this technique is limited to surface studies. Recently,
pump-probe (stimulated emission) microscopy has been demonstrated to
provide images with spatial resolution comparable to that of confocal
microscopy and at potentially very high temporal resolution
(Dong et al., 1995
). The main focus of this work is to
extend the pump-probe (stimulated emission) method to time-resolved anisotropy studies in microscopy. Our objective is to obtain image sections based on the rotational correlation time of a probe in a cell
and to accurately determine the characteristic time of the anisotropy decay.
Lifetime-resolved fluorescence microscopy
In fluorescence microscopy, lifetime-resolved images can provide
insights into functionally important processes in biological systems.
Dynamics of the chromophores in their local environments can be
characterized, and this can help to elucidate the functioning of
cellular components at the molecular level. Many applications of
fluorescence lifetime imaging have been demonstrated. Important cellular information such as calcium concentration or cytoplasm matrix
viscosity have been obtained by lifetime-resolved methods (Dix
and Verkman, 1990
; Keating and Wensel, 1990
;
Kao et al., 1993
). Measurements of the autofluorescence
lifetime have been used to monitor the mechanism responsible for cell
damage due to UVA exposure (Schneckenburger et al.,
1992
; Schneckenburger and Koenig, 1992
;
Koenig and Schneckenburger, 1994
). The fluorescence lifetime has also been used to assess antigen-processing stages in
mouse macrophage cells (Voss, 1990
). Furthermore, the
spatial distribution of chromophores with similar emission spectra but different lifetimes provides a new contrast mechanism (Draaijer et al., 1995
). Fluorescence lifetime-resolved studies have also been successfully conducted in confocal (Morgan et al.,
1991
; Buurman et al., 1992
) and two-photon
(Piston et al., 1992
; So et al., 1996
)
scanning microscopy.
Cytoplasmic viscosity of cells
Various fluorescence techniques have been used to probe the
(micro)viscosity of cytoplasm. Picosecond rotation of small
fluorophores has been studied to estimate fluid-phase viscosity in the
cell cytosol (Periasamy et al., 1991
; Fushimi and
Verkman, 1991
). Recently, cytoplasmic viscosity has been
probed, using the diffusion and rotation of the green fluorescent
protein (Swaminathan et al., 1997
). All of these studies
are essentially single point measurements inside cells, and they show
that for small molecules, the viscosity of cytoplasm relative to that
of water is between ~1 and 3. In our approach, we obtain
time-resolved polarization images of a cytoplasmic probe throughout the
cell. Imaging the entire cytoplasm may reveal the structural
heterogeneity that exists inside a cell.
Lifetime and polarization measurements in the frequency domain
Time-resolved measurements in the frequency domain are performed
by means of modulated light sources (Spencer and Weber,
1969
; Gratton and Limkeman, 1983
). For
sinusoidal excitation at modulation frequency
=
/2
and
modulation me, the spatial (
)
and temporal (t) distribution of the excitation photon flux
I(
, t) can be written as I(
, t) = I(
)[1 + me sin(
t)]. The
number density of excited-state molecules N(
, t) of
fluorophores with a single lifetime
obeys the differential equation
|
(1) |
is the molecular absorption cross section, and
I(
, t)[c
N(
, t)] denotes the
number of ground-state molecules actually being excited by the light
source. The total fluorophore concentration c is assumed to
be constant. To the first order in which only a small fraction of
molecules are excited, the fluorescence response
F(
, t) is at the same modulation frequency as the
excitation light but is demodulated and phase shifted:
|
(2) |
is its phase shift with respect to the excitation
light. By measuring either the phase shift
or the
demodulation m = m'/me, the fluorescence
lifetime can be determined:
|
(3) |
p and
m denote the phase and
modulation lifetimes, respectively. More complex excitation modulation
functions, such as a laser pulse train, can be represented as a
superposition of Fourier components. The fluorescence signal can then
be represented as the sum of the response to each Fourier component
(Spencer and Weber, 1969Time-resolved anisotropy studies are performed by measuring the
fluorescence parallel and perpendicular to the excitation polarization.
For linearly polarized
-pulse excitation, the parallel (
) and
perpendicular (
) components of the fluorescence decay are given by
|
(4) |
|
(5) |
|
(6) |
is the rotational correlation time and
r0 is the initial anisotropy.
In the frequency domain the anisotropy is determined by measuring the
frequency response of the differential phase 
= 

and modulation ratio
Y between the parallel and perpendicular fluorescence
signals. For an isotropic rotator, the differential phase 
and
the (squared) modulation ratio Y2 are given by
|
(7) |
|
(8) |
, and
correspondingly, the fluorescence decay rate k is the
inverse of the fluorescence lifetime
, i.e., k = 1/
(Gratton and Limkeman, 1983| |
PUMP-PROBE FLUORESCENCE MICROSCOPY |
|---|
|
|
|---|
Overview of pump-probe techniques
Pump-probe spectroscopy has been applied successfully to the
investigation of ultrafast processes in biology, chemistry, and condensed matter. The temporal evolution of the process can be studied
by varying the timing between the pump and probe pulses (Evans,
1989
; Fleming, 1986
; Lytle et al.,
1985
). A common approach is to excite the sample with a strong
pulse of light (pump) and to sample the ground-state population with a
weak pulse of light (probe). The temporal separation between these
pulses is typically controlled by means of a mechanical delay line that
allows fine control of the laser beams' optical paths. The ultimate
temporal resolution of this type of pump-probe spectroscopy is
determined by the pulse width of the laser. In the visible and near-IR
wavelength ranges, the duration of light pulses is around a few
femtoseconds (10
15s) (Zhou et al., 1994
;
Xu et al., 1996
; Christov et al., 1996
). Ultrafast phenomena in many biological systems, including
photosynthetic reaction centers, rhodopsin, and heme proteins, have
been studied (Hochstrasser and Johnson, 1988
).
An alternative method in pump-probe techniques is the asynchronous
sampling version first proposed by the Lytle group (Elzinga et
al., 1987
). In this approach, two pulsed laser beams with
different repetition rates are focused on the same spot in the sample
to induce the pump-probe effect. The difference in repetition rates introduces a variable delay between the pump and probe pulses. The
effect of this delay is to repeatedly sample the excited-state population of the molecules at consecutive time intervals after the
pump laser pulses. A frequency domain analysis of the pump-probe signal
generated in this manner shows that the (high-) frequency components of
the fluorescence signal are shifted to the lasers' cross-correlation
frequency and its harmonics. These frequency components can be
analyzed to obtain time-resolved information. Harmonic content can be
explored in excess of 12 GHz (Berland et al., 1992
).
Instead of sampling the transient absorption of the probe pulse,
researchers can choose its wavelength to induce stimulated emission
from the excited-state population. The stimulated emission approach has
been demonstrated in spectroscopic studies (Lakowicz et al.,
1994
; Kusba et al., 1994
; Adachi et al.,
1995
).
Imaging microscopy based on the asynchronous sampling technique offers several advantages over the equivalent technique that uses a mechanical delay line. To acquire time-resolved images by means of a mechanical delay line, either all of the fluorescence properties are measured in one pixel by running one laser beam through a series of time delays before moving to the next pixel, or the entire image is scanned at one particular time delay and consecutively repeated for all delay times, yielding a corresponding series of time slice images. Within an image acquisition time of 60 s, 30 time slice images at most can be recorded, because it takes ~2 s to scan an image of 256 × 256 pixels (cf. section Scanning Optics). With respect to frequency domain, this is equivalent to acquiring 15 harmonics of the fluorescence signal in each pixel. To measure 15 harmonics in one pixel at once, the time delay has to be switched 30 times within the pixel dwell time of ~1 ms, which is not mechanically feasible.
With the asynchronous sampling technique, 15 harmonics can easily be recorded at once within the 1-ms pixel dwell time, assuming a corresponding base cross-correlation frequency of 1 kHz and a moderate signal digitization rate of 30 kHz. Hence, by simply increasing the digitization rate, the asynchronous sampling technique allows us to retrieve more harmonic components of the fluorescence signal than the mechanical delay line approach within the given image acquisition time of 60 s. Furthermore, if the cross-correlation frequency is also increased, the image acquisition time can be reduced without losing harmonic information. However, the major advantage in using a mechanical delay line to perform pump-probe experiments is in the autocorrelation capability of the laser pulses. In this setup, the jitter between the pump and probe pulses is greatly reduced, resulting in improved temporal resolution.
Principle of the pump-probe stimulated emission technique
The basic principle of the stimulated emission technique is illustrated in Fig. 1. The beams of two pulsed lasers are focused onto a common volume inside the specimen. The pulse repetition rates of the two lasers are offset slightly, and their wavelengths are chosen such that the pump beam excites the chromophores under study and the probe beam induces stimulated emission. The optical modulation of the excited-state population by the probe pulses results in a heterodyning of the fluorescence signal, i.e., the Fourier spectrum of the fluorescence decay is translated to the cross-correlation frequency range. The high harmonic content of pulsed lasers (THz range for femtosecond lasers) allows the acquisition of multiple harmonics of the decay spectrum at once. The fluorescence signal containing the cross-correlation harmonics can be isolated by means of optical filtering, because the emission is at wavelengths different from those of both excitation and stimulated emissions.
|
The timing diagram shown in Fig. 2
further illustrates the principle of the pump-probe (stimulated
emission) technique. The pulse train of the pump laser periodically
excites the chromophores and correspondingly generates a series of
fluorescence decays. At increasing numbers of excitation cycles, the
probe pulse is successively delayed with respect to the pump pulse. The
fundamental delay or an equivalent sampling time is determined by the
difference in the lasers' repetition rates. For instance, if the
lasers are operated at ~76.2 MHz and the offset frequency is 5 kHz,
the equivalent sampling time is ~860 fs. At the arrival of a probe
pulse, the correspondingly generated stimulated emission causes a drop
in the actual fluorescence because the stimulated emission photons propagate in the same direction as the probe pulse (Sargent et al., 1974
). The amplitude of the stimulated emission signal and the drop in fluorescence both depend on the number of excited-state molecules at the arrival time of the probe pulse. As a result, the
excited-state population is sampled by the probe pulses at different
delays after excitation. The corresponding time-resolved information
can be obtained by observing the stimulated emission (same direction as
the probe laser) or the residual fluorescence signal emanating in all
other directions.
|
Equations governing the pump-probe stimulated emission process
Our instrument acquires the residual fluorescence intensity
F(t) of the pump-probe stimulated emission process. The same
objective is used to focus and spatially overlap the two laser beams
and to collect the fluorescence (cf. Fig. 4). To obtain an analytical expression for F(t), first consider an exponentially
decaying species that is respectively excited and deexcited (via
stimulated emission) by means of a pump and probe laser. The
differential equation that governs the number density of excited state
chromophores N(
, t) now becomes
|
(9) |
, t) and I'(
, t)
describe the spatial and temporal intensity distributions of the pump
and probe laser, respectively. As in Eq. 1,
is the fluorophore
lifetime,
is the absorption cross section, and the total
fluorophore concentration c is assumed to be constant. The
additional term 
'I'(
, t)N(
, t) accounts for the number of excited-state molecules that return to the ground state via stimulated emission, and
' denotes the
(wavelength-dependent) stimulated emission cross section.
In our experiments the pump and the probe laser generate two periodic
pulse trains with slightly different periodicities T and
T' or repetition frequencies
= 2
/T
and
' = 2
/T', respectively. Thus Fourier analysis can
be applied to solve Eq. 9. The derivation of an analytical solution of
Eq. 9 and, correspondingly, a closed expression for the residual
fluorescence F(t) are shown in the Appendix. Here we only
present the final expression for the residual fluorescence emission
rate F(t) of the pump-probe stimulated emission process (cf.
Eq. A8):
|
(10) |
|
) and L'(n
) are the
Fourier coefficients of the temporal pulse profiles L(t) and
L'(t) of the pump and probe laser, respectively;
q is the quantum yield of the molecules; and the phase shift
n is given by
n = tan
1(n
).
Both the amplitude
L(n
)L'(n
')/
and the phase
n of each harmonic term contain the
fluorescence lifetime
. For n = 1, this result is
consistent with Eq. 3. Thus, by measuring the signals' amplitudes and
phase spectra at the cross-correlation frequencies, the fluorescence
lifetime
can be determined from tan(
n) = n
and
L(n
)L'(n
')/
.
Note that the high-frequency response is limited by two factors: the
harmonic content of the pump and probe laser, respectively, given by
L(n
) and L'(n
'), and/or the lifetime modulation factor
.
Spatial resolution provided by pump-probe fluorescence microscopy
A key feature of the pump-probe method is the localization of the
cross-correlation signal within the overlapping integral
I(r)I'(r)d3r of the pump and probe
beam as shown in Eq. 10. The point-spread function (PSF) of pump-probe
techniques has been studied both theoretically and experimentally
(Dong et al., 1995
,
1997
). Spatial specificity is achieved because the pump-probe
signal is localized near the focal region where both the pump and probe
lasers have a high photon flux. As Eq. 10 shows, in a pump-probe
fluorescence microscope in which both the pump and probe processes
involve a one-photon transition, the PSF is given by
|
(11) |
|
(12) |
/
)sin2(
/2)z and
v = (2
/
)sin(
)r are the
dimensionless axial (z) and radial (r)
coordinates, respectively, and
is the wavelength of light focused
by a circular objective of numerical aperture sin(
) (Born and
Wolf, 1985Lifetime and time-resolved polarization imaging in pump-probe stimulated emission microscopy
Fluorescence lifetime measurements are generally performed at the
magic angle condition in which the emission polarizer is rotated to
54.7° relative to the excitation beam polarization; the lifetime
measured will then be free of the effect due to the chromophores'
rotational correlation time. Polarization measurements are achieved by
sampling the fluorescence at the parallel and perpendicular directions
relative to the polarization of the excitation light source by rotation
of the emission polarizer. In the pump-probe (stimulated emission)
technique, a different approach is used to satisfy the magic angle and
polarization measurement conditions. As illustrated in Fig.
3, the x/z plane polarized
pump pulse photoselects those molecules that have their absorption
transition dipoles (
A) preferentially
aligned with the z axis. The polarization of the probe pulse
is set at 54.7° relative to that of the pump pulse. Consequently, the
resulting modulation of the parallel (I
) and
perpendicular (I
) fluorescence components is proportional to the numerical factors cos2(54.7°) = 1/3 and sin2(54.7°) = 2/3, respectively. Because the
fluorescence was detected without an emission polarizer,
I
and I
are
acquired at a ratio of 1:2, thus effectively sampling the excited-state molecules at the magic angle. Similarly, to sample the population of
the excited-state molecules in directions parallel and perpendicular to
the pump beam's polarization, the probe beam's polarization is
rotated in parallel and perpendicular directions. Compared to standard
fluorescence measurements, the stimulated emission technique samples
the distribution of the excited-state molecules by changing the probe
beam polarization without the use of an emission polarizer.
|
Ultrafast molecular dynamics measured without high-speed photodetectors
Direct decay measurements are ultimately limited by the detector
frequency response in the study of ultrafast fluorescence phenomena in
both the frequency and time domains. A typical photomultiplier (Hamamatsu R928) used in a fluorescence lifetime instrument can be
gain-modulated to ~400 MHz. A faster detector such as a microchannel plate has been used up to ~10 GHz (Gratton and vandeVen,
1989
). On the other hand, a commonly available laser system can
easily generate picosecond pulses (FWHM) corresponding to a 3-dB
frequency bandwidth of 220 GHz. Because of the limited detector
bandwidth, the standard frequency domain fluorescence instrumentation
is not capable of exploring the high modulation frequency content available in picosecond pulsed laser systems. In contrast, pump-probe (stimulated emission) fluorescence microscopy allows lifetime-resolved imaging without the use of fast photodetectors. In Eqs. 3 and 10, we
showed that the phase and modulation of the cross-correlation signal at
|
'
| can be used to determine the fluorescence lifetime. In our experiments, |
'
| is chosen in the kHz range, which allows us to acquire images at fast scanning rates.
The main focus of this work is to demonstrate that pump-probe stimulated emission microscopy produces time-resolved (in particular, polarization-resolved) images of cells. While we demonstrate the feasibility of this technique, more important biological questions are addressed with appropriate systems. For example, appropriately labeled biomolecules may be introduced into cellular systems, and monitoring the time-resolved polarization properties of these biomolecules can elucidate functionally important motions such as aggregation or change in the local viscosity environment.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Sample preparation
The two imaging samples presented in this work are orange fluorescent latex spheres and labeled mouse fibroblast cells. Orange fluorescent latex spheres 15 µm in diameter (absorption maximum: 530 nm; emission maximum: 560 nm; Molecular Probes, Eugene, OR) were immobilized between a coverslip and a flat microscope slide with Fluoromount G mounting medium (Southern Biotechnology, Birmingham, AL). The mouse fibroblast cells were grown in Petri dishes on coverslips, using Dulbecco's modified Eagle's medium (DMEM) (D-6655; Sigma, St. Louis, MO) supplemented with 10% fetal bovine serum, 0.11 mg/liter sodium pyruvate, 1% MEM nonessential amino acids, and 0.5% penicillin/streptomycin. CellTracker CMTMR (5-(and 6)-(((4-chloromethyl)ben-zoyl)amino)tetramethyl-rhodamine (Molecular Probes) was used as the vital fluorescent label. The 1-mg aliquot supplied by the manufacturer was dissolved in dimethyl sulfoxide (DMSO) (Sigma), yielding a 10 mM stock solution. The final dye concentration was 5 µM. For staining, the culture medium was removed from the Petri dish, and the 5 µM CMTMR solution was deposited onto the coverslips' surfaces. After incubating for 30-45 min at 37°C, the coverslips were washed and immersed in fresh culture media and placed back into the incubator for the same period of time. For the microscopic measurements, the coverslips were mounted on well slides filled with the culture medium, and their edges were sealed with nail polish. All measurements were performed at room temperature immediately after the completion of the sealing procedure.
The experimental arrangement of our pump-probe (stimulated emission)
fluorescence microscope is shown in Fig.
4. A more comprehensive description of
the instrument can be found elsewhere (Dong et al.,
1995
).
|
Pump-probe light sources
The 10-MHz output of a master synthesizer is used to synchronize a mode-locked neodymium-YAG laser and a neodymium-YLF laser (Antares; Coherent, Santa Clara, CA). The pulse repetition rate of the two neodymium (Nd) lasers can be separately set by two frequency synthesizers (PTS500; Programmed Test Sources, Littleton, MA). Each PTS synthesizer drives a mode-locker unit and eventually the laser's acoustic-optical modulator (not shown in Fig. 4). The (frequency-doubled) 532-nm line of the Nd-YAG laser is used for sample excitation. To induce stimulated emission, the wavelength of the Nd-YLF pumped DCM dye laser (model 700; Coherent) is tuned to 640 nm. The pulse widths (FWHM) of the pump and probe laser are ~100 ps and ~10 ps, respectively. The power and polarization of each laser beam are separately controlled by two Glan-Taylor polarizers (03PTA001; Melles Griot, Irvine, CA). Typically, the average power of the pump and probe beams at the sample plane is less than 300 µW and 3 mW, respectively. For time-resolved fluorescence microscopy, the pump source is operated at 76.2 MHz + 2.5 kHz, and the probe laser's repetition frequency is 5 kHz away at 76.2 MHz + 2.5 kHz. For spectroscopic studies, the difference in repetition frequency is reduced to 210 Hz, with the pump and probe lasers operating at 76.2 MHz and 76.2 MHz + 210 Hz, respectively.
Scanning optics
The two laser beams are combined by a dichroic mirror D1 (Chroma Technology, Brattleboro, VT) before they reach the x-y scanner (model 6350; Cambridge Technology, Watertown, MA). The angular positions of the two scanning mirrors can be digitally set (DAC) by 16-bit binary numbers. However, only the middle eight bits are used for scanning, correspondingly yielding an image comprising 256 × 256 pixels. The closed-loop bandwidth of the scanner is in the audio range (settling time ~1.6 ms). Therefore, the fastest image acquisition time is ~2 s.
A beam expander is inserted into the microscope's excitation light
path to overfill the back-aperture of the objective and hence produce
diffraction-limited focusing. The expanded beams are reflected into the
microscope objective by a second dichroic D2 (Chroma Technology). We
have found that within the angular range used for scanning, the
combination of beam expander and objective linearly transforms the
angular deviation of the input laser beam into a lateral translation of
the focal point position at the object plane. Because the field
aperture plane is telecentric to the object plane of the microscope
objective, the movement of the focal point on the object plane is
proportional to the angular deviation of the scanned beam
(Stelzer, 1995
).
Because tight focusing increases the photon flux and localizes the pump-probe effect, a well-corrected, high numerical aperture objective is used (63× Plan-Neofluar, NA 1.25; Zeiss, Thornwood, NY). The fluorescence signal is collected by the same objective and transmitted through the dichroic D2 and two bandpass filters BP (600 nm ± 20 nm), before being refocused onto the detector (R3938 or R1104 photomultiplier tube; Hamamatsu, Bridgewater, NJ). For the 63× objective, the scanning range in the object plane is 35 × 35 µm, and the scanning raster is 140 nm.
Signal detection and image processing
For imaging studies, the photomultiplier tube (PMT) signal is first amplified by an ac-coupled (f3dB = 300 Hz) current amplifier (SR570; Stanford Research, Sunnyvale, CA). Then the cross-correlation signal is isolated by a 5 kHz ± 1 kHz bandpass filter and postamplified (SR560, Stanford Research) before its analog-to-digital conversion (ADC) by a 12-bit sampling digitizer (A2D-160; DRA Laboratories, Sterling, VA). The ADC sampling clock is derived from the 10-MHz master line and hence is phase-locked to the lasers. According to the Shannon sampling theorem, at least two points per cross-correlation waveform have to be acquired. In our experiments, the signal is sampled four times per waveform to reduce harmonic noise. To further reduce the noise, several waveforms can be averaged. From such a four-point time record, the phase and amplitude of the 5-kHz cross-correlation signal are derived by fast Fourier transform algorithm. After the predetermined number of waveforms are acquired, the computer correspondingly drives the DAC to advance the scanner to the next pixel. The pixel-to-pixel movement is synchronized to the digitizer for phase coherence. When only one waveform of the 5-kHz cross-correlation signal is acquired, the pixel dwell time is 200 µs, and the corresponding frame acquisition time is 13 s. After digital processing, the amplitude and phase of the cross-correlation signal are displayed and stored by the data acquisition computer. Note that the phase and modulation data are acquired relative to the 10-MHz signal of the master synthesizer. Therefore, a separate measurement with a standard dye is required to calibrate the system.
For spectroscopic studies, the PMT signal is directly acquired by a spectrum analyzer (model 35665A; Hewlett Packard, Rolling Meadows, IL; not shown in the figure). Because of its limited bandwidth (~100 kHz), a smaller cross-correlation frequency is chosen to allow the acquisition of multiple harmonics of a fluorescence decay. The multiharmonic phase and modulation data can be fitted with the appropriate model function to derive the specimen's fluorescence lifetime or rotational correlation time. In contrast to imaging studies that provide only phase and modulation data at a single cross-correlation frequency, a standard dye is therefore not required in the spectroscopic mode. For the saturation studies described in the following section, the cross-correlation signal was set at 210 Hz.
| |
RESULTS AND DISCUSSION |
|---|
|
|
|---|
Saturation
We determined the appropriate power levels of the pump and probe lasers to avoid saturation of the cross-correlation signal in the focal region. Initially, the pump-probe (stimulated emission) signal increases linearly as either laser increases in power. With a further increase in the pump and probe laser powers, the ground-state and excited-state populations, respectively, in the focal region can become depleted. Consequently, fluorophores in the off-focal region start to contribute to the pump-probe signal, which results in a broadening of the point-spread function (PSF) and deterioration of image resolution.
As a characteristic fluorophore aqueous rhodamine B (280 µM) was used to determine the appropriate laser intensity values to operate in the linear region. The power levels of both lasers at the objective's focal point were measured by means of a transimpedance amplified solar cell that has been calibrated at the wavelengths of the pump and probe lasers. To mimic data acquisition conditions, the solar cell was placed at the specimen position with a layer of immersion oil between the solar cell and the objective. To enhance the dynamic range of the detection system, off-focal fluorescence was reduced by inserting a 600-µm pinhole at the focal plane of the emission tube lens. For the 63× objective used, the focal volume's radial dimension in the image plane of the tube lens is ~60 µm. Therefore, the 600-µm pinhole only eliminates off-focal fluorescence and does not interfere with the signal from around the focal volume.
For five distinct pump power intensities, i.e., 55, 127, 221, 400, and 510 µW, the first harmonic amplitude at 210 Hz was measured as the probe power was increased stepwise from 860 µW to 7.5 mW. The resulting data series are plotted in Fig. 5 a. For each of the five pump power series, the signal amplitude first increases linearly as a function of probe power. However, at higher probe power, the first harmonic amplitude starts to increase at slower rates, indicating the onset of signal saturation. The data series show that probe power saturation begins at ~3 mW regardless of the pump power used. To determine to onset of pump power saturation, the linear region of each data series was fitted. The slopes obtained are plotted in Fig. 5 b. Deviation from linearity of the slope indicates the onset of signal saturation due to the pump power level. Therefore, we conclude from Fig. 5 b that the signal saturation due to the pump power starts at ~300 µW.
|
Based on these saturation data, both pump and probe saturation levels of other chromophores can be estimated, provided their extinction coefficients, quantum yields, and stimulated emission cross sections relative to that of rhodamine B are known.
Point-spread function
The strong dependence of the optical cross-correlation signal on
the spatial overlap of the pump and probe beams provides optical
sectioning similar to that found in confocal and two-photon excitation
microscopy. For the case of one-photon pump excitation and one-photon
probe deexcitation, the point-spread function has been calculated and
experimentally verified. It was shown that the radial and axial
dimensions of our pump-probe (stimulated emission) microscopy are on
the order of 0.3 µm and 1.1 µm, respectively (Dong et al.,
1995
).
Lifetime and rotation of rhodamine B in water
As an application of the pump-probe (stimulated emission) technique, we measured the lifetime and rotational correlation time of rhodamine B (280 µM) in Tris buffer. The power spectrum of the fluorescence decay is shown in Fig. 6 a. The equivalent (after heterodyning) frequency response of our instrument easily extends into the GHz range, a significant improvement over the 400-MHz limit imposed by PMT modulation in a standard frequency domain fluorometer. Specifically, we can measure harmonics (peaks) of the fluorescence decay of rhodamine B up to 4.572 GHz before the noise significantly distorts the signal.
|
Because the finite pulse widths of the Nd-YAG and DCM dye lasers cause
an attenuation of the high-frequency components, a proper fit of the
harmonic decay must include factors accounting for the laser sources'
harmonic content. Of course, the Nd-YAG 100-ps pulses
(f3db
2.2 GHz) contribute more
significantly to the high-frequency harmonic distortion than the
shorter 10-ps dye laser pulses (f3db
22 GHz; cf. Eq. A10).
The appropriate frequency domain fitting function can be
derived from Eq. 10 by substituting
L(N
)L'(N
') with the corresponding Fourier coefficients
of a Lorentzian-shaped pulse profile (cf. Eq. A11),
|
(13) |
i.
The corresponding single exponential fit of the data yielded a
fluorescence lifetime of 1.54 ns for rhodamine B, in good agreement with the previously measured value of 1.5 ns (Dong et al.,
1995
). The lifetime fit is shown in Fig. 6 b. The
factor b was determined to be 61.8 ps, corresponding to a
FWHM of 123.6 ps. This value is reasonable because it accounts for both
the pulse widths and the relative timing jitter of the two lasers.
Fig. 7 shows the frequency domain differential polarization spectrum of rhodamine B in Tris buffer. The modulation ratio and differential phase were derived from the fluorescence between the parallel and perpendicular orientations of the probe beam with respect to the pump source. Using Globals Unlimited (LFD, Urbana, IL) software, a simple rotator model with a single exponential lifetime of 1.54 ns, and an initial anisotropy of 0.39, the rotational correlation time was determined to be 88 ps. Note that in both the lifetime and rotational measurements, we were able to explore the fluorescence harmonic content up to the GHz range.
|
Fluorescent latex spheres
To demonstrate the acquisition of time-resolved polarization
images, orange fluorescent latex spheres 15 µm in diameter were imaged. For the lifetime-resolved images, the spheres were scanned with
the probe beam oriented at 54.7° (magic angle) relative to the pump
beam. The first harmonic amplitude and phase data (at 76.2 MHz) of a
sphere are shown in Fig. 8 a.
The phase values were acquired relative to a reference compound
(aqueous rhodamine B,
= 1.5 ns), a common procedure in phase
fluorometry. Within a region inside the sphere (white
square), the averaged phase is 74.3°. The corresponding
fluorescence lifetime of 7.43 ns is derived by means of Eq. 3.
|
To obtain the time-resolved polarization images, the sphere was scanned
twice: first, with the probe beam polarizer oriented parallel (
) and
then perpendicular (
) to the pump beam polarization. The first
harmonic amplitude and phase data of the polarized images are shown in
Fig. 8 b. For the same region previously described (white square), the averaged phase for parallel
excitation and parallel deexcitation (
/
) is 68.3°, and that for
parallel excitation and perpendicular deexcitation (
/
) is
74.1°. The 5.8° phase shift between the two measurements can be
visualized by the histograms of the two phase images also shown in the figure.
The depolarization of fluorescence emission reflected by the phase
shift might be caused by energy transfer and/or rotation of dye
molecules inside the spheres. To differentiate between the two
phenomena, low-temperature (
10°C) fluorescence polarization measurements (
ex = 532 nm) of Fluormount G
suspended spheres were made in a frequency domain fluorometer (K2; ISS,
Champaign, IL). The measured low polarization value of 0.025 indicates
that energy transfer is most likely the depolarizing mechanism. In fact, fluorescent spheres from the same manufacturer have been shown to
be 95% efficient in excitation transfer efficiency (Roberts et
al., 1998
).
Mouse fibroblast labeled with CellTracker CMTMR
Using the pump-probe (stimulated emission) microscope, we
obtained time-resolved polarization images of mouse fibroblast cells labeled with CellTracker CMTMR (Molecular Probes), a rhodamine based,
cell-permeant probe. The CMTMR's chloromethyl group reacts with
intracellular thiols. Most likely, the dye enzymatically binds to the
glutathione (
-glutamylcysteinylglycine) commonly found in high
concentrations inside cells. The resulting CMTMR-glutathione conjugate
is cell-impermeant and is localized primarily in the cell's cytoplasm
(Molecular Probes product information sheet).
First, lifetime-resolved amplitude and phase images of a CMTMR-labeled mouse fibroblast were acquired under the magic angle condition. The first harmonic amplitude and phase data (at 76.2 MHz) are shown in Fig. 9 a. Again, the phase is determined relative to that of aqueous rhodamine B. The measured phase values are similar (violet-red color) over most of the cell, except for the two distinct regions I and III. For these two areas, the average fluorescence lifetimes are 1.93 ns (region I) and 1.85 ns (region III). For a region in between and an arbitrary chosen area at the top, the lifetime averages are 2.75 ns (region II) and 2.85 ns (region IV).
|
The first harmonic amplitude is nonuniform across the cell. The variation in amplitude can be due to local changes in both the concentration and the lifetime of fluorophores. From the phase data, it can be seen that the lifetime varies little across the cell. Therefore, the first harmonic amplitude image basically reflects the local concentration of the CMTMR label, except for the two 1.9-ns lifetime regions. The shorter lifetime in these two domains reflects the fact that the CMTMR-glutathione conjugate experiences a microenvironment different from that in the rest of the cell.
Second, time-resolved polarization images of the same mouse fibroblast
have been acquired in the same way as previously described (cf. section
Fluorescent Latex Spheres). The first harmonic amplitude and phase data
are shown in Fig. 9 b. The rotational correlation times are
derived from both the differential phase and modulation data by means
of Eqs. 7 and 8. Assuming r0 to be 0.39, the
rotational correlation times for the same four regions I-VI as marked
in the magic angle images (cf. Fig. 9 a) are 28, 107, 104, and 151 ps, respectively. The measured rotational rates are fast and
correspond roughly to the values obtained for unhindered rotation of
small dyes in an aqueous environment. The rotational properties of
CMTMR-glutathione conjugate are dominated by CMTMR, because the
fluorescent label is about two times larger than glutathione. Because
CMTMR is about twice as large as glutathione, such fast rotational
correlation times suggest that the cytoplasm acts more like an aqueous
than a gel-like environment for the CMTMR-glutathione complex. This is
in agreement with recent studies that also used small probes to
investigate the microviscosity of the cytoplasm (Periasamy et
al., 1991
; Swaminathan et al., 1997
).
| |
CONCLUSION |
|---|
|
|
|---|
We have applied the pump-probe stimulated emission approach to time-resolved polarization imaging. The system has been calibrated for pump-probe signal saturation at both laser powers. Lifetime imaging under the magic angle condition can be satisfied by rotating the probe beam polarization 54.7° relative to that of the pump beam. Similarly, polarization imaging can be achieved by rotating the probe beam polarization parallel and perpendicular to the pump beam. Time-resolved polarization images of fluorescent latex spheres show that the depolarization of fluorescence inside the spheres is most likely due to energy transfer. Lifetime and time-resolved polarization images of CellTracker CMTMR-stained mouse fibroblast cells show fast rotational correlation times of the glutathione-complexed probe ranging from 30 to 150 ps. This implies that a relatively low microviscosity exists inside the cytoplasm.
| |
APPENDIX: DERIVATION OF FREQUENCY DOMAIN, PUMP-PROBE FLUORESCENCE SIGNAL F(t) |
|---|
|
|
|---|
Based on Eq. 9, we derive an analytical expression for the
residual fluorescence intensity F(t) of the pump-probe
stimulated emission process. Equation 9 is a first-order linear
differential equation of the form y'(x) + p(x)y = g(x), the solution of which is given by
|
(A1) |
xp(t)dt), and
c1 is a constant specifying the boundary value
or initial condition (Boyce and DiPrima, 1977
|
(A2) |
|
(A3) |
= 2
is the angular frequency of the laser
pulses, and L(n
) is the Fourier coefficient of the
nth frequency component of the pulse train. The intensity
distribution of a laser pulse train I(
, t) focused
by a microscope objective can be written as the product
of the spatial distribution I(
) and the temporal profile L(t),
|
(A4) |
, t) and probe laser I'(
, t)
according to Eq. A4 into Eq. A2, N(
, t) becomes
|
(A5) |
|
|
, t) is the product of the molecules' quantum
yield q and the rate of change of the excited-state
population,
|
(A6) |
|
(A7) |
|
![]() |