| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, October 2000, p. 1993-2001, Vol. 79, No. 4


*Department of Chemistry, Göteborg University,
Göteborg SE-412 96, Sweden and
Department
of Chemistry, University of Pittsburgh, Pittsburgh, Pennsylvania,
15260 USA
| |
ABSTRACT |
|---|
|
|
|---|
Electroporation of single NG108-15 cells with carbon-fiber microelectrodes was characterized by patch-clamp recordings and fluorescence microscopy. To minimize adverse capacitive charging effects, the patch-clamp pipette was sealed on the cell at a 90o angle with respect to the microelectrodes where the applied potential reaches a minimum. From transmembrane current responses, we determined the electric field strengths necessary for ion-permeable pore formation and investigated the kinetics of pore opening and closing as well as pore open times. From both patch-clamp and fluorescence microscopy experiments, the threshold transmembrane potentials for dielectric breakdown of NG108-15 cells, using 1-ms rectangular waveform pulses, was ~250 mV. The electroporation pulse preceded pore formation, and analyte entry into the cells was dictated by concentration, and membrane resting potential driving forces. By stepwise moving a cell out of the focused field while measuring the transmembrane current response during a supramaximal pulse, we show that cells at a distance of ~30 µm from the focused field were not permeabilized.
| |
INTRODUCTION |
|---|
|
|
|---|
During the last two decades, there has been a
tremendous growth in the number of experimental methods available for
the biochemical and biophysical investigations of single cells. Such
methods include 1) patch-clamp techniques for measuring transmembrane
currents through a single ion channel (Hamill et al., 1981
), 2)
scanning confocal and multiphoton microscopy for imaging and localizing bioactive components in single cells and single organelles (Maiti et
al., 1997
), 3) near-field optical probes for measuring pH in the cell
interior (Song et al., 1997
), 4) ultramicroelectrodes for monitoring
the release of single catechol- and indol-amine-containing vesicles
(Chow et al., 1992
, Wightman et al., 1991
), and 5) optical trapping and
capillary electrophoresis separations for analyzing the chemical
composition of individual secretory vesicles (Chiu et al., 1998
).
Although numerous high-resolution techniques exist to detect, image, and analyze the contents of single cells and subcellular organelles, few methods exist to control and manipulate the biochemical nature of these compartments. Most compounds of biological and medical interest are polar and therefore unable to cross cell membranes. Examples of such compounds are dyes, drugs, DNA, RNA, proteins, peptides, and amino acids. At present, it is extremely difficult, for example, to label an individual cell in a cell culture with a dye, or transfect it with a gene without labeling or transfecting its adjacent neighbor.
Electroporation is a frequently applied technique for introduction of
charged or polar molecules into intracellular domains of a plurality of
cells. The application of an external electric field across artificial
or biological phospholipid bilayer membranes results in increased
permeability and conductance of the membrane (Zimmermann, 1982
,
Zimmermann et al., 1974
).
The membrane voltage, Vm, at different
locations of a spherical cell membrane in a homogeneous electric field
for duration t, can be calculated from
|
(1) |
is
the angle in relation to the direction of the electric field, and
|
(2) |
The applied electric field will generate large depolarizing and
hyperpolarizing transmembrane potentials at the cathode- and anode-facing poles of the cell, respectively. When the membrane potential reaches a critical value, typically 0.2-1.5 V for mammalian cells, dielectric membrane breakdown will occur, which results in pore
formation. Pore density will follow the
Vm-gradient, and is highest at the
polarized (electrode-facing) ends of a cell. During the effective
pore-open time, cell-impermeant solutes added to the extracellular
medium can enter the cell interior (Weaver, 1993
).
The conductance caused by dielectric breakdown of membranes is
proportional to the amplitude and duration of the electric field
(Neumann et al., 1998
). Kinosita and Tsong, (1977a
,b
) reported an
effective pore diameter of 1 nm for human erythrocyte membranes and
concluded that the size of the pores was determined by field strength,
pulse duration, pH, and ionic strength of the suspending medium. Chang
and Reese, (1990)
observed pores with diameters between 6 and 240 nm in
human erythrocyte membranes by rapid-freeze fracture electron
microscopy. In charge-pulse studies on irreversible breakdown of planar
lipid bilayers, pores with diameters up to 400 µm were observed
(Wilhelm et al., 1993
).
The duration, strength, and wave function of the applied electric pulse
also determines whether the electropermeabilization of the membrane
will be reversible or irreversible or of a punch-through type (Weaver,
1993
). Reversible dielectric breakdown occurs when high-voltage,
short-duration pulses are applied, wherein the pores are formed and
resealed within fractions of a second (Benz and Zimmermann, 1980
;
Weaver, 1993
). Resealing of the membrane is a two-step process. Within
a few milliseconds after the pulse, the membrane conductance is
significantly decreased although the number of pores remain constant,
indicating that the radii of the pores has decreased. The decrease in
the number of pores is a much slower process because of an energy
barrier that prevents closure. The disappearance of pores takes seconds
(Glaser et al., 1988
). Also, formation of metastable pores after
reversible dielectric breakdown with resealing times from seconds to
hours has been observed (Kinosita and Tsong, 1977a
; Chernomordik et
al., 1987
; Lopez et al., 1988
).
In addition to bulk electroporation methods, instrumentation has been
developed that can be used for electroporation of a small number of
cells in suspension (Chang, 1989
; Kinosita and Tsong, 1979
; Marszalek
et al., 1997
) and for a small number of adherent cells grown on a
substratum (Teruel and Meyer, 1997
; Zheng and Chang, 1991
). These
electroporation devices create homogeneous electric fields across fixed
distances of 0.1-5 mm, several times larger than the size of a single
mammalian cell.
We recently reported on a highly spatially resolved electroporation
technique, miniaturized to perform experiments at the single-cell and
subcellular level (Lundqvist et al., 1998
). The electrodes are of
smaller dimension than a single cell and are controlled individually
with high-graduation micropositioners, which enables precise electrode
alignment in all three directions. This capability makes possible the
electroporation or electrofusion of a single selected cell in a
confluent cell culture (Lundqvist et al., 1998
; Strömberg et al.,
2000
). In addition to the high spatial resolution achieved by using
ultramicroelectrodes, this experimental arrangement avoids the use of
high-voltage pulse generators and complicated microchamber mounts. In
contrast to microinjection techniques for single cells and single
nuclei (Capecchi, 1980
), the present technique can be applied to
biological containers of sub-femtoliter (10
15 l)
volumes, that are less than a few micrometers in diameter. Electroporation has other advantages over microinjection techniques in
that it can be extremely fast, and well-timed (Kinosita et al., 1988
;
Hibino et al., 1991
), which is of importance in studying fast reaction phenomena.
This paper describes our studies on the mechanisms of microelectrode
electroporation using patch-clamp and fluorescence microscopy. Because
most characterization of electroporation has been performed in batch
mode with homogeneous electric fields, no in-depth characterization is
available for single-cell electroporation using inhomogeneous electric
fields. This characterization is necessary for finding optimal
protocols for the loading of single cells. In addition, a patch-clamp
electroporation method that enables probing of membrane properties
during and after exposure of a single cell to a DC electric field is a
valuable tool for studying membrane dynamics. Patch-clamp recordings
have previously been demonstrated as a powerful method to monitor
dielectric membrane breakdown in biological and synthetic membranes
under the influence of a homogeneous electric field (O'Neill and Tung,
1991
; Owen and Piotrowski, 1987
; Sharma et al., 1996
). NG108-15 cells
were patch clamped in the whole-cell configuration as described
elsewhere (Hamill et al., 1981
), and the electroporation electrodes
were positioned on opposite sides of the cell. The spatial distribution
of the electric field was characterized by stepwise moving the cell out
of the focused field. The potential needed for membrane breakdown was
determined by applying 1-ms rectangular waveform pulse sequences of
increasing voltages. The potential drop due to the solution resistance
was determined in separate experiments to estimate the effective
electric field strength. Furthermore, from the patch-clamp recordings, pore opening and resealing kinetics, and pore open times were determined.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
NG108-15 cells were grown in Dulbecco's minimum essential medium supplemented with 10% calf serum and 1% antibiotics at 37°C in a humidified 5% CO2, 95% air atmosphere. For patch-clamp experiments, the culture medium was replaced by a HEPES buffer (140 mM NaCl, 10 mM HEPES, 10 mM D-glucose, 5 mM KCl, 1 mM CaCl2, 1 mM MgCl2, and pH was adjusted to 7.4 with NaOH). The cells were mechanically removed from the petri dish in which they were cultured, and then separated from each other in a Pasteur pipette by shear forces and plated onto #1 circular coverslips just before experiments. The cell dishes were mounted on an inverted microscope (Leica DM IRB, Wetzlar, Germany) equipped with a Leica Fluotar 40× objective. All experiments were performed at room temperature.
Before experiments, cells were left to rest for 10-15 min. Recordings
were performed in the whole-cell configuration, and the inside of the
cell membrane was clamped at
50 mV with respect to the grounded bath.
Data were stored on videotape (digitized at 20 kHz) and were analyzed
later (sample frequency 2 kHz, filter frequency 1 kHz). The time delay
between the electroporation pulse and transmembrane current events, the
amplitude of the transmembrane current, and the time needed for
complete recovery were obtained from the traces. The high level of
noise in the traces was caused by laser flickering. The pipette
solution contained 100 mM KCl, 2 mM MgCl2, 1 mM
CaCl2, 11 mM EGTA, and 10 mM HEPES (pH was
adjusted to 7.2 with KOH).
For electroporation, we used 5-µm-outer-diameter carbon fiber microelectrodes (Pro CFE, Dagan Corporation, Minneapolis, MN) controlled by high-graduation micromanipulators (Narishige MWH-3, Tokyo, Japan). The carbon fibers were enclosed in a plastic tip filled with 3 M KCl that was in contact with a silver wire connected to the voltage-pulse generator (Digitimer Stimulator DS9A, Welwyn Garden City, U.K.). To study the potential needed for dielectric membrane breakdown, the two electrode tips were positioned 2-5 µm from the outer cell surface at an angle of 0-20°, and 160-180° with respect to the object plane, and 90° with respect to the patch pipette (Fig. 1). Cells were electropermeabilized by application of monophasic 1-ms rectangular DC voltage pulses of increasing field strengths from 1.1 to 8.1 kV/cm (not corrected for voltage drop due to solution resistance). By studying the current responses from the cells, the potential needed to cause pore formation in the plasma membrane could be determined. Figure 2 shows schematically how electroporation and patch clamp were performed on a single cell. The spatial distribution of the electric field was studied by stepwise withdrawal of a cell out of the focused field in steps of 2, 5, 8, 10, 11, 12, and 27 µm, and measuring the current response while applying a supramaximal pulse of 3.7 kV/cm. Permeabilization of the membrane and closure of pores could also be determined from the current responses during and after the pulse in the patch clamp experiments.
|
|
For fluorescence experiments, cells were grown for 1-2 days, and the culture medium was replaced with an intracellular buffer consisting of 135 mM KCl, 5 mM NaCl, 20 mM HEPES, 1.5 mM MgCl2, 10 mM glucose, and 5 µM fluorescein, pH was adjusted to 7.4 with NaOH. The cells were electroporated with ten pulses of 1-ms duration and electric field strengths of 1.3-6.5 kV/cm (values not corrected for potential drop), washed several times so no extracellular fluorescein remained in the buffer, and dye trapped inside the cells was detected. For excitation of fluorescein, an Ar+-laser (Spectra-Physics 2025-05, 488 nm), a 488 nm-line interference filter, a spinning disk to break the coherence, an inverted microscope (DMIRB, Leica) equipped with a 40× objective to focus and collect the laser light, and a fluorescein filter cube (I-3, Leica) were used. Images were recorded with a 3-chip color CCD-camera (C6157, Hamamatsu, Kista, Sweden) and stored at 25 Hz collection rate on a Super VHS (S-VHS AG-5700, Panasonic, Stockholm, Sweden). Fluorescence intensity was measured using an image processor (Hamamatsu). The CCD images were digitized from tape and processed for presentation.
Solution resistance studies were performed using a waveform generator (Model DS345, Stanford Research Systems, Sunnyvale, CA) to apply pulses, and a digital oscilloscope (Model 9410, Lecroy, Chestnut Ridge, NY) with GPIB connection for computerized data collection. The current signal from the electrodes was amplified using a home-built current-to-voltage converter before going into the oscilloscope. Applied pulses were 1-ms long, with a rest period of 1.63 s between pulses. All experiments were performed in HEPES buffer solution (140 mM NaCl, 10 mM HEPES, 10 mM D-glucose, 5 mM KCl, 1 mM CaCl2, 1 mM MgCl2, and pH adjusted to 7.4 using NaOH). Commercial microelectrodes (Dagan Corp., Minneapolis, MN) were positioned using micromanipulators (Narishige Co. Ltd., Tokyo, Japan), and viewed under a stereo microscope (Fisher Scientific, Blawnox, PA).
Chemicals
HEPES (>99%), sodium chloride, potassium chloride, and sodium hydroxide (all suprapur), calcium chloride, magnesium dichloride, D-glucose and EGTA (Titriplex VI) (all pro analysis) were purchased from Merck (Darmstadt, Germany). Fluorescein (GC-grade) was obtained from Sigma (St. Louis, MO). Deionized water from a Milli-Q system (Millipore, Bedford, MA) was used.
| |
RESULTS |
|---|
|
|
|---|
Electrochemical reactions and spatial electric field focusing
The strength of the electric field over the cell in our setup is
not equal to the applied electric field strength because of initiation
of electrode reactions and the formation of an electric double layer at
the surface of the electrodes. The determination of the potential
gradient experienced by the cells has been investigated by Weaver and
co-workers (Bliss et al., 1988
; Gift and Weaver, 1995
; Pliquett and
Weaver, 1996
; Pliquett et al., 1995
; Pliquett et al., 1996
; Prausnitz
et al., 1993
; Weaver and Chizmadzhev, 1996
). In normal electroporation
experiments, there are several concerns, including the dynamics of the
voltage distribution and the fraction of the applied voltage that ends
up across the cell (Pliquett et al., 1996
). In the current experimental
setup, we are also concerned with these issues. The problem of the
amount of the applied voltage appearing across the solution is
particularly important for these microelectrode systems because the
voltage drop at the electrode-solution interface is significant
in comparison with the applied voltage.
For small applied voltages (<2 V), the potential gradient in solution is very small at long times (>100 µs). At short times, the convolution of the instrumental rise time and the electrode/electrolyte charging time gives rise to a peak-shaped field-versus-time curve. The maximum magnitude of the applied voltage in solution can be calculated by convoluting exponential functions for the instrument rise time and the charging time. For our particular case, with an instrument rise time of 8 µs, and a charging time in the range 20-40 µs, the maximum amplitude is 15-18% of the applied pulse amplitude. This maximum occurs ~10-15 µs after application of the pulse. For example, a 1.0-V pulse would result in a maximum of about a 150-mV potential difference between the electrodes, occurring ~10-15 µs after application of the pulse. After ~50-75 µs, the potential drop is effectively zero because all of the applied voltage appears at the electrode-solution interfaces.
We hypothesize that, at high voltages, the interfacial impedance
becomes negligible, and the current that passes between the microelectrodes is limited by the solution conductance. To determine the potential drop across the solution, 1-ms pulses were applied across
pairs of carbon fiber microelectrodes separated by 5-50 µm in a
HEPES buffer solution. The solution resistance was calculated by
plotting the current at the end of the pulse as a function of applied
potential between 9 and 10 V. Resistance was determined by linear
regression (the inverse of the conductance, see Fig. 3). The fact that the current is linearly
related to the applied voltage supports the idea that the medium is
resistive. This process was repeated for each electrode separation
distance. The solution resistance was found to be 27 ± 3 k
for
distances of 10 µm and greater. The constant value of the calculated
resistance reflects the nature of the field created in solution. The
field lines converge onto the electrodes, so most of the potential drop
occurs near the electrodes. Solution potential drops were calculated by
multiplying the current at 1 ms by the calculated resistance value. The
potential drop across the solution increases considerably with
increasing applied potential (see Fig.
4), but only slightly with increasing electrode separation.
|
|
In general, solution-resistance (conductance) measurements usually are
not done at DC because of the dual problems of electrolysis and
double-layer potential drop. With microelectrodes, however, AC
measurements cannot be used because of their small capacitance. If the
frequency used is high enough for the current to "see" a low
impedance in the double layer, then it is too high for commonly used
instrumentation. Also, we are not able to use a 4-probe type measurement (Pliquett et al., 1996
) in which a pair of measuring electrodes was placed between the two potential application electrodes, because of the small dimensions of our system. Thus, we are forced to
use the alternative approach described above.
Membrane potentials reported in this article will be given as calculated from Eq. 1 without voltage-drop correction for electrode reactions and solution resistance. Because Eq. 1 is written for homogeneous electric fields, this only gives an estimate of the magnitude of the fields in our system. Field strengths are reported as applied potential with respect to electrode separation.
The spatial distribution of the electric field was studied by stepwise
withdrawal of patch-clamped cells out of the center of the electrodes.
By measuring transmembrane currents at each position, at a constant
applied voltage pulse, the effect of the field on the cell could be
determined. Cells were moved out of the field in steps of 2, 5, 8, 10, 11, 12, and 27 µm, and, for each movement, a supramaximal pulse of
3.7 kV/cm and 1-ms duration was applied. Figure
5 shows that, at a distance of 27 µm
from the field focus, the current response was only
25 pA, which is 16% of the current response at the field focus. When patch-clamped cells were positioned 40 and 67 µm from the field focus, extremely high field strengths were needed to obtain a transmembrane current response, 5.6 kV/cm and 6.7 kV/cm, respectively.
|
Threshold membrane potential for dielectric breakdown in NG108-15 cells
From patch-clamp measurements, the transmembrane threshold
potential for pore formation of NG108-15 cells was determined to be
~2.5 V. When corrected for the voltage drop at the double layer, this
corresponds to ~250 mV, which is in good agreement with previous reports (Teissié and Rols, 1993
; Teissié
and Tsong, 1981
). Membrane-potential values higher than ~4 V resulted
in irreversible membrane breakdown, and the cells did not recover.
Figure 6 shows transmembrane current responses from a cell electroporated with increasing voltage. Dielectric membrane breakdown was observed from ~2.5 V, resulting in
currents in the range of
30 to
330 pA. Results presented in Table
1 indicate that membrane ion-permeability
is lower for potentials near the threshold value than for higher
potentials. The high conductance state is also more prolonged for
higher voltages than for threshold potentials. An increase of the field
strength can also result in morphological changes in the cells, such as chromatin condensation, swelling of organelles, and retraction of
processes to the cell body.
|
|
The results obtained in the fluorescence experiments using fluorescein (dianion, MW 376.3) as a marker are in agreement with the patch clamp data. Figure 7 shows fluorescence intensity as a function of applied voltage, and Fig. 8 shows fluorescence and brightfield micrographs of the electroporated cells. The threshold membrane potential required for a detectable difference in fluorescence intensity, obtained by comparing control and electroporated cells, was 3.0 V, and the fluorescence intensity increased with higher electric field strengths. Figure 8 also shows that cells electroporated at supramaximal field strengths had a swollen and granulated appearance.
|
|
Kinetics of pore-expansion and resealing
During the 1-ms pulse, a peak response is typically seen due to
capacitive charging of the patch pipette, and the response from the
electroporated membrane appears as an increase in the membrane
conductance immediately or a few milliseconds after the pulse. The time
for expansion/evolution of the pores ranges between 0 and 50 ms (Table
1) at a rate of 3.1-7.3 pA/ms. The high-conductance state of the
membrane lasted for 6-104 ms. Recovery times from high-conductance
states ranged from 10 to 620 ms. Cells did occasionally have recovery
times of seconds to minutes. These cells usually recovered from the
high-conductance state within 500 ms, but there seemed to be a
low-conductance state stable for longer times. These long time
scales are in accordance with previous reports where pores were shown
to be stable over very long times, minutes to hours (Chang and Reese,
1990
; Chernomordik et al., 1987
; Lopez et al., 1988
; Kinosita and
Tsong, 1977a
).
The rate of increase in transmembrane currents was estimated to be
3.1 to
6.1 pA/ms using linear regression.
| |
DISCUSSION |
|---|
|
|
|---|
Electroporation of selected single cells and subcellular structures in a culture requires ultrasmall electrodes, which produces inhomogeneous electric fields. By stepwise withdrawal of a patch-clamped cell out of the field focus, we have shown that the electric field is tightly focused, leaving adjacent cells almost unaffected. Cells were unperturbed at a distance of ~30 µm from the center of the field using supramaximal stimulation. At longer distances from the center of the focused field, 40 and 67 µm, extremely large field strengths were required, 5.6 and 6.5 kV/cm, respectively, for detecting a current response. Such field strengths would cause immediate cell death if the applied electric fields were homogeneous.
The critical field strength for producing a detectable current response
is approximately 3.2 kV/cm, which corresponds to a transmembrane
potential of ~2.5 V. When the cell membrane is permeabilized through
electroporation, transmembrane currents of
35 to
330 pA were
observed. Application of field strengths larger than the threshold
value caused transmembrane currents of increased magnitudes. Possibly
the number of pores increases with increasing applied field strength.
Alternatively, the same number of pores are created but expand in size
when a higher electric field is applied. Wilhelm et al., (1993)
argued
that the magnitude of the applied field does not affect the conductance
resulting from pore formation. Rather, pore formation is a chaotic
event where the probable number of pores is dependent upon applied
voltage. What happens to the pores formed is solely determined by the
properties of the membrane. In our study, we cannot determine what
process caused larger transmembrane currents, but it is evident that,
when potentials higher than the threshold value were applied,
transmembrane currents increased in magnitude. Well above the threshold
potential, the magnitude of the current did not change significantly
with increased potential.
Upon repeated electroporation of one cell with the same magnitude of
the electric field, the resulting transmembrane current diminished with
each subsequent pulse. Chernomordik et al., (1987)
reported this effect
for low-amplitude pulses, and the opposite effect for high-amplitude
and long-duration pulses, that is, the breakdown current becomes higher
for each pulse, and the membrane loses its ability to reseal.
In our fluorescence data, the trend is similar, as described above. Below the threshold value, no dye enters the cell. At ~3.0 V, pores are formed, which allows for entry of the dye into the cell; higher transmembrane potentials caused a corresponding increase in fluorescence intensity. The similarity of the curves in Fig. 4 and Fig. 7 shows that the fluorescence response from cells electroporated with increasing field strength in presence of fluorescein is proportional to the voltage drop in the solution.
In the patch-clamp experiments, the cells were lifted up in the electroporation media by the patch pipette, whereas, in the fluorescence experiments, the cells were adherent to the substratum during electroporation. This fact might have contributed to the observed difference in breakdown potential in these two experiments. The threshold potential could also be higher in the fluorescence experiments because they were less sensitive than the patch-clamp recordings.
Typically, pore formation or expansion was observed within a few
milliseconds after pulse application. In some experiments, however, a
delay of up to 30 ms was observed. This delay is probably caused by the
time it takes for the pores to expand to a size sufficient for ions to
pass. This time delay is consistent with previously reported data.
Chang and Reese (1990)
could detect volcano-shaped pores after 3 ms. In
their report, pore expansion occurred within 40-120 ms, which is in
good agreement with our observations: the conductance increased during
the first 50 ms and then reached its maximum state where it remained
stable for a few to 104 ms. A higher applied field seems to make this
state more extended in time. Teruel and Meyer (1997)
also have reported that small pores are formed initially, but only a few pores expanded locally to create entry sites large enough for macromolecules to pass.
The latter stage occurred over milliseconds. The rate of mass transport
is dependent on the concentration gradient and the electric-field
gradient across the cell membrane. Both of these factors can be
controlled by varying the applied voltage and the concentration of the
analyte outside the cell. The membrane was clamped at
50 mV, and,
when electropermeabilized inward currents were observed that were
reversed at +10 mV, indicating some sort of selection between ions.
Because of the delay between application of the electric pulse and
formation of pores, the applied field does not affect directly the
transport of ions over the membrane. This explanation is consistent
with the results presented by Neumann et al. (1998)
.
Recovery of the membrane usually occurs within 0.5-1 s. The recovery
process was longer at higher voltages (Table 1). There seemed to be two
time constants involved in membrane recovery: a fast recovery process
(milliseconds) as mentioned above, which returns the conductance to the
approximate prepulse level, but with a noisier baseline, and a slow
process (seconds or longer), which returns this low-conductance state
into a state with closed pores. Neumann et al., (1998)
have reported
dye uptake experiments that indicate pore resealing in kinetically
resolved steps for mouse B cells, and Benz and Zimmermann (1981)
for
planar lipid bilayers. The time scale for recovery is also in good
agreement with other studies (Chang and Reese, 1990
; Glaser et al.,
1988
; Teruel and Meyer, 1997
). Some of the cells were not fully
recovered within 5 min, especially those cells that had experienced
higher fields, which indicates that some of the pores have remained
stable over minutes. This observation agrees with previous reports, in which pores were shown to be stable over minutes to hours (Chernomordik et al., 1987
; Kinosita and Tsong, 1977a
; Lopez et al., 1988
).
In the field of biophysical and bioanalytical sciences, a tremendous development toward single-cell- and single-organelle-based technologies has emerged. Our electroporation technique describes the ability to load substances, such as genes, dyes, and pharmaceuticals into single selected cells in a confluent cell culture. This represents, to our knowledge, the highest spatial resolution offered for electroporation and provides, in combination with patch clamp, the opportunity for detailed kinetic and mechanistic studies. This microelectrode system can be developed further for altering the biochemical state of subcellular structures such as organelles, for accessing sensing systems, including receptors, ion channels, and enzymes in single-cell biosensor applications, or for drug targeting a specific group of cells in vivo. For these applications, further optimization studies with regards to voltage pulse parameters (e.g., pulse profile, voltage amplitude, pulse time), electrode materials, and electrode design, including topographically complex microelectrodes, are needed.
| |
ACKNOWLEDGMENTS |
|---|
We are grateful to Susanne Orwar for digital image editing and photographic work. This work was made possible through grants by the Swedish Natural Science Research Council (grants 10481-305, 308, 309), the Swedish Foundation for Engineering Sciences, the Swedish Foundation for Strategic Research, and The Swedish Foundation for International Cooperation in Research and Higher Education (STINT) for a visiting scientist scholarship to D.T.C.
| |
FOOTNOTES |
|---|
Received for publication 13 March 2000 and in final form 10 July 2000.
Address reprint requests to Owe Orwar, Department of Chemistry, Göteborg University, Göteborg SE-412 96, Sweden. Tel.: +46-31-772-2778; Fax: +46-31-772-2785; E-mail: orwar{at}amc.chalmers.se.
Dr. Kent Jardemark's present address is Department of Psychiatry and Behavioral Science, State University of New York at Stony Brook, Stony Brook, NY 11794-8790.
Dr. Daniel T. Chiu's present address is Department of Chemistry and Chemical Biology, Harvard University, Cambridge, MA 02138.
| |
REFERENCES |
|---|
|
|
|---|
Biophys J, October 2000, p. 1993-2001, Vol. 79, No. 4
© 2000 by the Biophysical Society 0006-3495/00/10/1993/09 $2.00
This article has been cited by other articles:
![]() |
A. Bruckbauer, P. James, D. Zhou, J. W. Yoon, D. Excell, Y. Korchev, R. Jones, and D. Klenerman Nanopipette Delivery of Individual Molecules to Cellular Compartments for Single-Molecule Fluorescence Tracking Biophys. J., November 1, 2007; 93(9): 3120 - 3131. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Zudans, A. Agarwal, O. Orwar, and S. G. Weber Numerical Calculations of Single-Cell Electroporation with an Electrolyte-Filled Capillary Biophys. J., May 15, 2007; 92(10): 3696 - 3705. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Pavlin, M. Kanduser, M. Rebersek, G. Pucihar, F. X. Hart, R. Magjarevic, and D. Miklavcic Effect of Cell Electroporation on the Conductivity of a Cell Suspension Biophys. J., June 1, 2005; 88(6): 4378 - 4390. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Pavlin and D. Miklavcic Effective Conductivity of a Suspension of Permeabilized Cells: A Theoretical Analysis Biophys. J., August 1, 2003; 85(2): 719 - 729. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |