An externally applied electric field across vesicles
leads to transient perforation of the membrane. The distribution and lifetime of these pores was examined using
1,2-di-oleoyl-sn-glycero-3-phosphocholine (DOPC)
phospholipid vesicles using a standard fluorescent microscope. The
vesicle membrane was stained with a fluorescent membrane dye, and upon
field application, a single membrane pore as large as ~7 µm in
diameter was observed at the vesicle membrane facing the negative
electrode. At the anode-facing hemisphere, large and visible pores are
seldom found, but formation of many small pores is implicated by the
data. Analysis of pre- and post-field fluorescent vesicle images, as
well as images from negatively stained electron micrographs, indicate
that pore formation is associated with a partial loss of the
phospholipid bilayer from the vesicle membrane. Up to ~14% of the
membrane surface could be lost due to pore formation. Interestingly,
despite a clear difference in the size distribution of the pores
observed, the effective porous areas at both hemispheres was
approximately equal. Ca2+ influx measurements into
perforated vesicles further showed that pores are essentially resealed
within ~165 ms after the pulse. The pore distribution found in this
study is in line with an earlier hypothesis (E. Tekle, R. D. Astumian, and P. B. Chock, 1994, Proc. Natl. Acad. Sci.
U.S.A. 91:11512-11516
) of asymmetric pore distribution based
on selective transport of various fluorescent markers across electroporated membranes.
 |
INTRODUCTION |
External electric fields of short duration across
intact cells and vesicular systems leads to transient membrane pores
through which influx/efflux of impermeable molecules is believed to
occur (Zimmermann, 1986
; Neumann et al., 1989
; Chang et al., 1992
;
Potter, 1988
). These events are generally referred to as
electroporation or electropermeabilization and have led to a number of
other related processes such as electrofusion for the production of
somatic cell hybrids (Al-Atabee et al., 1990
) and electroinsertion for the specific incorporation of proteins in cellular membranes (Mouneimne et al., 1991
) as well as recent clinical applications in targeted transdermal drug delivery and tumor treatment (Prausnitz et al., 1993
;
Heller et al., 1996
). The induction of pores on the membrane surface of
some given number, size(s), and lifetime(s) is one common feature
underlying all these processes and ultimately governs the relative
success of the various methods used. Consequently, significant focus
has been directed toward a basic understanding of the underlying
mechanisms involved.
Field-induced membrane pores have been reported on a number of cell
types under a variety of experimental conditions, e.g., red blood cells
and ghosts (Chang and Reese, 1990
; Sowers and Lieber, 1986
; Sowers,
1987
), mammalian and plant cells (Tekle et al., 1990
, 1991
, 1994
;
Gabriel and Teissie, 1997
; Mehrle et al., 1985
, 1989
), sea urchin eggs
(Hibino et al., 1991
, 1993
; Kinosita et al., 1988
), and bacterial cells
(Miller et al., 1988
). Irrespective of the cell type examined to date,
by far the most consistent finding has been the eminent rupture of the
membrane provided a critical transmembrane potential,

c (~0.3-1 V), develops across the
membrane, where 
c refers to the minimum
amplitude required for macroscopically observable electroporation. For
non-porated membranes, the magnitude of the sub-critical transmembrane
potential, 
i, due to an external electric
field, E, is given by (Neumann and Rosenheck, 1973
)
|
(1)
|
where b is the outer radii of the cell membrane and
is the angle between the field direction and any point on the membrane. The maximum potential drop at the poles of the membrane facing the
electrodes (i.e., cos(
) = ±1 at
=
, 0) and the
cosine dependence have generally been confirmed by a number of studies employing potential-sensitive dyes (Ehrenberg et al., 1987
).
By contrast, the density of these pores as well as their size and
distribution on the perforated membrane surface is less well
understood, and their structure is even far less known. One clear
exception to the latter is an earlier study by Chang and Reese (1990)
where a rapid freezing electron microscopy was used to show
volcano-shaped pores in erythrocyte membranes exposed to an intense
electric pulse. These pore structures were visually detectable ~3 ms
after the pulse and resealed to their initial state in several seconds
with no observable residue. The spatial disposition of the pores
relative to the electroporating pulse was, however, not resolved.
Nevertheless, based on these early findings and further modeling
studies (Weaver and Barnett, 1992
; Kakorin et al., 1996
), it is
currently thought the initial electroporation event involves the
rearrangement of clusters of lipids into hydrophobic and hydrophilic
pore structures with minimum pore size in the order of a few
nanometers. Estimates of the number of pores per cell, derived from
conductance and transport studies, have ranged from just a few (<10)
(Saulis et al., 1991
), ~700 (Sowers and Lieber, 1986
) pores in ghost
cells to up to ~105 in mouse B cells (Neumann
et al., 1998
). Percolation of these nanometer-sized primary pores into
larger openings have also been envisaged (Sugar and Neumann, 1984
).
Much of the insight into how and where these pores are distributed on
the membrane surface has largely been inferred from transport studies
involving influx/efflux of probe molecules (Dimitrov and Sowers,
1990
; Tekle et al., 1990
; Mehrle et al., 1989
; Gabriel and Teissie,
1998
) as well as from optical signals of membrane-bound
potential-sensitive dyes (Hibino et al., 1991
, 1993
; Kinosita et al.,
1988
). Some features of the experimentally observed permeabilization
patterns have recently been shown in a series of modeling studies of
asymmetric electroporation (DeBruin and Krassowska, 1999a
,b
). However,
although influx/efflux and sieving experiments are powerful methods in
probing membrane pore dynamics, it is crucial to note these approaches
may not necessarily indicate the type of pores created or how they may have distributed at the membrane level, as the overall transport process could depend on the nature of the probe molecules used, the
mode of transport, and the kinetics of pore closure, all of which, or
in some combination, could lead to potential misinterpretations. This
possible anomaly, for instance, was originally anticipated by Saulis
(1993)
in a modeling study of asymmetric electroporation. Our earlier
study (Tekle et al., 1994
) had also shown selective transport of
various fluorescent markers on one side of the membrane despite the
fact that both hemispheres of cell membrane were permeabilized. This
earlier finding implied structurally different pores may have formed at
the opposite poles of the membrane, although no direct experimental
evidence was available in support of the hypothesis. It is thus clear
that current views on the structure, dynamics, and density of
field-induced membrane pores are limited, and further studies with a
particular focus to events occurring at the membrane level would be
useful. In this respect, vesicles offer a relatively less complex
membrane structure and have been used as model systems in a few earlier
studies (Teissie and Tsong, 1981
; Zhelev and Needham, 1993
), including
more recent ones (Tonsing et al., 1997
; Kakorin et al., 1998
; Kakorin
and Neumann, 1998
). Small unilamellar vesicles (SUVs, diameter ~100
nm) were employed in these studies, and no spatial information was thus obtained.
Here, giant vesicles (diameter >10 µm) of
1,2-di-oleoyl-sn-glycero-3-phosphocholine (DOPC) were used
to monitor the induction and distribution of pores at the membrane
level. A combination of membrane staining, electron microscopy, and
Ca2+ influx measurements were employed to study
the overall dynamics. The results revealed single visible pores (on the
order of ~7 µm) are predominantly formed at the cathode-facing
hemisphere and many small pores at the anode-facing hemisphere.
Analysis of the fluorescence pattern of the stained membrane as well as electron micrographs of negatively stained vesicles shows pore formation is accompanied by loss of lipid from the membrane surface. Despite the difference in the pore size distribution, however, the
overall porous area at both vesicle hemispheres was found to be
comparable. The pores are resealed within a few hundred milliseconds
after the pulse. These findings are compared and discussed with
currently held views on asymmetric electroporation.
 |
MATERIALS AND METHODS |
Reagents
DOPC in chloroform was purchased from Avanti Polar Lipids
(Alabaster, AL).
1-(3-sulfonatopropyl)-4-[B][2-(di-n-butylamino)-6-naphthyl vinyl] pyridinium betaine (Di-4-ANEPPS) and Fluo-3 penta-ammonium salt
were from Molecular Probes (Eugene, OR). CaCl2
was from Fisher (St Louis, MO).
Preparation of vesicles
Vesicles were prepared following the procedure of Angelova et
al. (1992)
using low-amplitude AC fields and modified to allow exchange
of hydrating liquid. A chamber was constructed using two conducting
indium-tin-oxide-coated microscope cover glasses (Delta Technologies,
Stillwater, MN), separated with a 1-mm-thick Teflon spacer and was
fitted with a simple flow system to allow exchange of hydrating
liquids. A 2-µl sample from a 20 mg/ml stock solution of DOPC in
chloroform was placed on the upper glass electrode and dried
extensively under a stream of nitrogen. The chamber was then
reassembled and the lipid hydrated using either deionized water alone
or a solution containing the desired dyes, i.e., di-4-ANEPPS or Fluo-3.
The initial hydration was done by directly injecting 200 µl of the
solution near the dried lipid with a micropipette. An AC field of ~70
V/cm, frequency 5 Hz, was then applied across the chamber from a
waveform generator (WaveTek 183, San Diego, CA), and the formation of
the vesicles was monitored under the microscope (Zeiss, Axiovert 100).
Within ~1 h, several layers of large vesicles, diameter ~25 µm,
were formed on the surface of the hydrated lipid. Inner layers
contained progressively smaller vesicles that grew with time up to ~2
h. Before the removal of the vesicles, the bathing solution was
replaced with dye-free medium to remove unincorporated dye. Membrane
labeling with di-4-ANEPPS was primarily done after the vesicles have
formed by bathing the vesicles in a solution containing the dye, unless
stated otherwise. The larger vesicles on the outer layer were finally
collected by gentle agitation of the bathing solution with a
micropipette. Typical yield of vesicles of diameter b, 20 µm < b < 30 µm, was 200-300 per 100 µl of
solution. These vesicle preparations were stable for several days at
4°C. In all the experiments reported here, the outside concentrations
of the reagents used were, 5 µM di-4-ANEPPS, 25 µM Fluo-3, and 2.5 mM CaCl2.
Electroporation and image acquisition equipment
The electroporation and image acquisition system is similar to
that described previously (Tekle et al., 1991
, 1994
). Electric pulses
of desired amplitude and a pulse width were supplied from a high-power
pulse generator (model 360, Velonex, Santa Clara, CA) and applied
across vesicles placed in an electroporation chamber. The chamber was
constructed with two polished stainless steel electrodes, fixed on a
75- × 25-mm microscope glass slide. The inter-electrode separation was
1 mm. The amplitude and pulse widths for each experiment were monitored
across the chamber on a digital oscilloscope (54502A, Hewlett-Packard,
Palo Alto, CA) using a 100:1 probe. The chamber is mounted on an
inverted fluorescence microscope (Zeiss, Axiovert-100), and the
excitation and emission wavelengths were set with appropriate bandpass
and cutoff filters at 520 and 610 nm for red emission and 480 and 520 for green emission, respectively. Fluorescence images were acquired
with an intensified camera (SIT-68, DAGE-MTI, Michigan City, IN) and
recorded onto a VHS tape (Panasonic model AG-7350). The intensity of
the fluorescence emissions was also monitored on an analog scope
(Hitachi, V-212) to protect the camera from excessive light exposure as
well as to monitor against saturation of the measured signals. Images were later processed using the National Institutes of Health Image 1.62 software (National Institutes of Health, Bethesda, MD) on a PowerMac
9500 computer. Because the electroporation events were fast and often
complete within only a few video frames, it was necessary to identify
t = 0 by synchronizing the video signal with the rising
edge of the electroporating pulse to eliminate timing uncertainty. A
useful pulse/video synchronization circuit was built for this purpose
(circuit diagram available upon request). The trigger signal is the
odd/even field signal from an LM 1881 sync separator (National
Semiconductor, Santa Clara, CA), which is driven by the composite video
signal from the video camera. When this flip-flop is triggered, it
provides an output to the pulse generator to start the experiment in
synchrony with the video signal. The circuit is further designed to
stamp a bright vertical bar near the upper left corner of the video
image and vanishes at the instant of the output trigger. This
triggering system accurately locates the video frame in which the
electric pulse is applied so that the time of electric pulse
application is included on the video frame recording of the experiment.
Successive video frames could then be time resolved at 33 ms per frame.
Electron microscopy
Vesicle preparations were negatively stained following a similar
procedure described by Moscho et al. (1996)
. Grids coated with a
parlodion/carbon substrate were treated with poly-D-lysine (1 mg/ml), rinsed, air dried, and glow discharged just before use to
make them hydrophilic. Vesicles were prepared as described earlier
here, and a drop of vesicle suspension was placed on Parafilm. The grid
was inverted on the drop for ~2 min. Excess liquid was removed by
blotting onto a filter paper at the periphery of the grid. The grid was
then negatively stained with 3 drops of 1% aqueous uranyl acetate.
Excess staining solution was similarly removed using filter paper.
Vesicle images were then obtained using a Philips-410 transmission
electron microscope at different magnifications.
 |
RESULTS |
Our primary aim in this study was to monitor the induction and
distribution of field-induced vesicle pores at the membrane level. For
this purpose, a widely used potential-sensitive dye, di-4-ANEPPS, was
used for its excellent membrane partitioning properties and
photo-stability. It should be noted, however, that this dye has been
used to monitor fast changes (microseconds time scale) in membrane
potential due to externally applied fields (Lojewska et al., 1989
), but
these optical responses are well below the time resolution
(milliseconds time scale) of our equipment here. In the present study,
the dye is used as a membrane stain to visualize the rupture events and
as a marker for lipid components lost from the vesicle membrane. This
stated use is justified given our experimental design and setup.
Fig. 1 shows a typical series of video
frames depicting the electroporation events before and after the
application of a 1.1-kV/cm (Fig. 1, A-C) or
0.920kV/cm in (Fig. 1, D-F) field and a pulse width of 700 µs. For all frames, the electrodes are positioned as
shown in Fig. 1 A. The membrane of the vesicle shown in Fig. 1, A-C, was labeled after the vesicle was
formed. For the vesicle shown in Fig. 1, D-F,
the dye was present during vesicle formation so that both the inner and
outer leaflets of the membrane were labeled to avoid any asymmetric
charge distribution. Fig. 1 A is the image before field
application, and Fig. 1 B shows the first frame (33 ms)
after the pulse. It shows that a large pore with a diameter ~6.2 µm
was formed at the negative electrode-facing hemisphere, whereas a rare
occurrence of a possible visible lipid aggregate loss was seen at the
hemisphere facing the anode. In the next frame (66 ms, Fig. 1
C), the membrane pore was resealed, partially leaving some
of the lipid on the outer surface. In Fig. 1,
D-F, the time sequence was the same as in Fig.
1, A-C, and the pattern of permeabilization is
representative of the successful experiments observed. Here again we
found a large pore, diameter ~7.5 µm (Fig. 1 E), on the
side facing the cathode, which subsequently resealed in the next frame,
(66 ms, Fig. 1 F). These data clearly show that symmetric
pores were not created in these vesicle systems. Further experiments
and analysis revealed that both sides of the membrane are
permeabilized, and a partial loss of membrane lipid occurs as a
consequence of pore formation. The asymmetry of electroporation is
reflected here in the size distribution of pores at the vesicle hemispheres facing the electrodes. Evidence supporting these events is
detailed below.

View larger version (45K):
[in this window]
[in a new window]
|
FIGURE 1
A select series of video frames showing the
electroporation events before and after electric field application.
E = 1.1 kV/cm in
(A-C), and E = 0.920 kV/cm in (D-F). The pulse width was 700 µs for both cases. The position of the electrodes is as indicated in
A for all frames. In A-C,
the membrane was labeled after the vesicle had formed. In
D-F, the dye was present during vesicle
formation so that both the inner and outer leaflets of the membrane are
labeled to avoid any asymmetric charge distribution. (A
and D) Images before field application;
(B and E) The first frames (33 ms) after
the pulse; (C-F) Images at 66 ms after
the pulse. The frame images are pseudo colored to enhance visualization
of the membrane pores and dye distribution after the field. Scale bar,
10 µm
|
|
Based on a number of image data as those shown in Fig. 1, the vesicle
dimensions (diameters) were carefully measured before (see for example,
Fig. 1, A and D) and immediately after (Fig. 1,
C and F) the electric field. Vesicle dimensions
were measured using contour templates from the vesicle images and are
the averages of two readings taken perpendicular to each other. These
measurements revealed that a significant reduction occurs in the size
of the vesicles after the field and clearly suggest appreciable amount of lipid must have been lost as a result of pore formation.
Surprisingly, however, we find that the loss of membrane lipid from the
observed pores at the cathode-facing hemispheres (Fig. 1, B
and E) could not account for the extent to which the vesicle
size is reduced when measured after the pulse (Fig. 1, C and
F). This discrepancy is elaborated below using the image
data in Fig. 1, D-F, as an example.
Figure. 2 shows a geometric
representation of the image data in Fig. 1 E, where
rp is the pore radius and the shaded
area, S = 2
r2(1
cos
), represents
the surface loss due to pore formation at the cathode-facing
hemisphere. If the reduction in vesicle size (Fig. 1 F) is
solely due to lipid loss at the cathode side, then
|
(2)
|
where r1 and
r2 are the radii of the vesicles
before (Fig. 1 D) and after (Fig. 1 F) the field,
respectively, and
is given by
|
(3)
|
The observed pore in Fig. 1 E should thus equal
rp, where the pore radius
rp (Fig. 2) is given by
|
(4)
|
The vesicle radii r1 (Fig. 1
D) and r2 (Fig. 1
F) were measured and found to be 13.4 and 12.7 µm,
respectively. Using Eqs. 3 and 4, we find
rp = 8.1 µm. However, the pore
radius (Fig. 1 E) is only ~3.8 µm, about half the value
of the computed pore radius, rp. Table
1 shows further
comparisons of the pore radii, rp,
calculated based on measured vesicle sizes before and after the
electric pulse with the pore radii observed,
robs, at the cathode-facing hemisphere for
different vesicle preparations. It is interesting to note that the
observed pore radii are consistently about half of that of
rp. In terms of total lipid lost from
the vesicle membrane, these data suggest at least two possibilities: either 1) additional lipid components must have been lost at other sites on the membrane surface or 2) the pore size at the cathode-facing hemisphere (Fig. 1 E) could have been larger at earlier
times than can be detected with the current equipment (note that images captured in Fig. 1, B and E, are ~32 ms after
the termination of the external pulse.). The latter possibility seems
highly unlikely as it would imply the vesicle would have to lose almost
half of its surface on the hemisphere facing the cathode to accommodate a pore radius of ~8.1 µm. More likely, many smaller pores are formed at the anode-facing hemisphere resulting in loss of lipid undetectable by the microscope setup.

View larger version (12K):
[in this window]
[in a new window]
|
FIGURE 2
Geometrical representation of a vesicle. The shaded
area represents the membrane pore as seen in Fig. 1, B
and E. x = r(1 cos ). The area of the shaded region is 2 rx,
where rp is the pore radius. The electric
field is as shown and is directed from left to right.
|
|
Lipid loss at the anode-facing hemisphere is in part implicated based
upon the asymmetric distribution of the membrane-staining dye. Before
field application, for example, the dye distribution is fairly
homogeneous as shown in Fig. 1, A and D, but a
marked asymmetry ensues in the images after the field with increased fluorescence at the anode-facing hemisphere (see Fig. 1, B,
C, E, and F). These distributions
remain for hundreds of seconds after the pulse and could not have been
due to potential redistribution around the vesicle as the external
field is absent in these time scales. A plausible cause for the uneven
dye signal could arise from redistribution of the lipid bilayer itself
as a consequence of pore formation such that small lipid aggregates
failing to reintegrate to the original bilayer remain attached to the
parent vesicle surface. This possibility is supported by electron
micrograph images of porated vesicles and explains the persistent
enhanced fluorescence long after the pulse has been terminated. The
data further point at the site origin of the missing lipid responsible for the observed reduction in the vesicle size shown earlier.
Fig. 3 shows negatively stained electron
micrographs of vesicles before (Fig. 3 A) and immediately
(~2 min) after (Fig. 2, B and C)
electroporation. Fig. 2, A', B', and
C' are magnified (zoomed) views of the membrane in Fig. 2,
A, B, and C, around the area indicated
by the arrows. Clearly, major rearrangement and undulation occurs on
the membrane surface as shown in Fig. 2, B' and
C'. All the images are from the same vesicle preparation. The field and pulse width were 5.7 kV/cm and 700 µs, respectively. The comparisons in the electron microscope data were gathered from a
subpopulation of vesicles with diameters ~1.5-5 µm. Many of the
giant vesicles (diameter
5 µm) were unavailable and possibly disintegrated into the lipid aggregates and other undefined structures seen under the electron microscope. This was observed in both the
control and pulsed samples and is due to the blot drying involved in
preparing the negatively stained grid samples (see Materials and
Methods).

View larger version (112K):
[in this window]
[in a new window]
|
FIGURE 3
Electron micrographs of vesicles stained with 1%
uranyl acetate before (A) and after (B,
and C) electric field application. E = 5.7 kV/cm; pulse width = 700 µs. A',
B', and C' are magnified views of the
area indicated by the arrow. Electron micrograph magnifications:
×24,000 (A), ×8700 (B), and ×4400
(C).
|
|
The results shown in Fig. 1 and Fig. 3 establish lipid loss at
both hemispheres and that a clear pore is observed at the
cathode-facing hemisphere. Whether pores were formed and subsequently
resealed at the anode-facing hemisphere was finally assessed by
monitoring Ca2+ influx from a medium containing
2.5mM CaCl2 into vesicles loaded with
Ca2+ indicator, Fluo-3. Intact vesicles are
impermeable to Ca2+ present in the suspending
medium. Fig. 4 shows both the rate of
Ca2+ influx and the spatial permeabilization
pattern after an electroporating field of 500 V/cm, 700-µs duration.
As shown in the inset of Fig. 4, Ca2+ influx
occurs at both the anode- and cathode-facing sites indicated by the
enhanced fluorescence of Ca2+-Fluo-3 complex. The
rate of Ca2+ influx is shown in the intensity
versus time plot of Fig. 4. The data points were obtained by
integrating the fluorescence intensity inside the vesicle from
successive video frames at the indicated times and shows a rapid rise
followed by a plateau within ~100-200 ms after the pulse. We should
mention here that the plateau was neither caused by saturation of the
indicator dye nor due to saturation of the camera's detection limit.
The observed time course could thus indicate the rate of pore closure
of the perforated membrane. The solid line in Fig. 4 is a fit to the
data assuming a simple diffusion of Ca2+ ions
through a time-dependent pore of radius r = r(t) (see Appendix).

View larger version (16K):
[in this window]
[in a new window]
|
FIGURE 4
Resealing of field-induced membrane pores. The vesicle
loaded with Fluo-3 was porated at 500 V/cm, 700-µs pulse duration, in
the presence of 2.5 mM CaCl2. The fluorescence intensity
from successive video frames as that shown in the inset was integrated
and plotted. The solid line is a two-parameter fit based on Eq. 8. The
time of resealing from the fit was ~164 ms.
|
|
Taken together, results presented here show that the pore distribution
at the membrane level is asymmetric with large pores at the cathode-
and small but numerous pores at the anode-facing hemisphere. The
creation of these pores is accompanied with loss of lipid from the
membrane surface. We should mention here that the data shown represent
results from the majority of the cases we have looked at. In general,
unilamellarity of the vesicles was not checked independently but was
assessed only by the sharpness of the membrane stain. Heavily
multi-layered vesicles, including some with patchy stained surfaces,
were easily identifiable, as were those containing other smaller
vesicles within them, and were not used for the experiments. In some
cases, vesicles simply disintegrated upon field application. However,
this does not appear to correlate with the strength of the applied
field used as many others were found to sustain much higher field
strengths. In general, experiments were carried out at ~1-1.5 V over
what we estimated to be a critical permeabilization potential,

c, of 648 mV. Fig. 5 shows a linear plot of Eq. 1, whose
slope gives 
c. The data points in the
figure were determined from the minimum field strength experimentally
required to observe Ca2+ influx into
Fluo-3-loaded vesicles. The electric field strength was incremented at
50 V/cm.

View larger version (8K):
[in this window]
[in a new window]
|
FIGURE 5
Determination of the critical permeabilization
potential,  c. The critical potential is defined as
the minimum potential required for observable pore formation. This
value was determined from three vesicle sizes based on the transport of
Ca2+ ions into Fluo-3-loaded vesicles. The electric field
was incremented at 50 V/cm. The field strength is plotted against the
inverse of the vesicle radius and considering cos( ) = 1 (see
Eq. 1). The slope of the fitted line gives  c = 648 mV.
|
|
 |
DISCUSSION |
The effect of an externally applied electric field on the
integrity of cell membranes in many cell types and vesicles, as well as
planar lipid bilayer systems, has been studied by means of a wide
variety of methods, including conductometric, optical, and rapid
freezing techniques. The most consistent finding in all systems
investigated has been the rupture of membranes provided sufficient
field strength is applied. Elegant studies using sea urchin eggs
(Hibino et al., 1993
) and hemispherical bilayer vesicles (Lojewska et
al., 1989
) have shown that the temporal and spatial variations of the
membrane potential due to external fields behaves as predicted by
theory. However, how and where the membrane ruptures to challenges of
an excess potential have produced differing results and interpretations.
Induction of membrane pores at some critical potential implies that
both hemispheres of the vesicle (cell) facing the electrodes be
permeabilized. To date, an apparent symmetric permeabilization has been
shown only when bipolar pulses were used (Tekle et al., 1991
). The
asymmetry often observed in many transport studies is believed to
originate due to a resting transmembrane potential in cells (negative
inside), thereby making the hemisphere facing the anode more
susceptible to rupture. However, this same hypothesis was used to
explain the susceptibility of the cathode-facing hemisphere, because
poration on one side can conceivably double the potential on the
opposite side. Teruel and Meyer (1997)
proposed unequal distribution of
charged phospholipid components at the membrane level to explain
asymmetric Ca2+ entry in depolarized cells, and
more recently ionic concentration gradients have been suggested to
influence the creation of asymmetric pore populations (DeBruin and
Krassowska, 1999a
). Despite these interesting propositions, the
asymmetry in electroporation still remains even under conditions where
large fields are applied that could easily overwhelm the relatively
small transmembrane potentials across cell membranes (Tekle et al.,
1990
) or even under conditions where, as shown in the present study,
neither a resting membrane potential, uneven lipid distribution, nor
ionic species exist in the vesicle preparations used. This last point
suggests other additional factors are probably involved in determining
the perforation pattern.
It is likely the pore distribution observed on the time scale of our
detection system results from a complex set of lipid rearrangements
that continues after the external field is turned off. One possible
initial source of asymmetry may, however, reside on how the external
electric field interacts with the intrinsic dipole potentials in the
inner and outer layer of the membrane bilayer (Loew, 1993
). The dipole
potential is the potential difference between the center of the bilayer
and the membrane/water interface, and its magnitude has been reported
to be of the order of ~300-450 mV (Gross et al., 1994
; Reyes et al.,
1983
). The electric field generated from these dipoles is directed from
the center of the bilayer toward the surface of the inner and outer
lipid layers. Considering that there would be more lipids on the outer
leaflet than on the inner leaflet due to packing constraints, it is
possible to imagine a net field could exist directed toward the vesicle surface. If this mechanism is operative, it could result in a larger
potential drop at the cathode-facing hemisphere compared with that of
the anode and may play a part in the induction of the large pores
observed at the negative electrode. Although this is currently
speculative, it should be possible to test this hypothesis by using
factors that could modulate the dipole potential. Additionally, faster
detection systems would also offer capturing events at earlier times
and shade light to the evolution of the pore structures observed here.
The processes leading to the induction of pores and the resulting
structure and distribution found here may in fact be common to other
previously studied cellular systems for a number of reasons. As shown
in Fig. 5, the magnitude of the critical potential required for
permeabilization (~0.6 V) measured by Ca2+
influx, is similar to that found in many cellular systems, typically in
the range of 0.5-1.0 V. Secondly, influx studies of a number of
fluorescent indicator dyes of varying charge and size has previously been shown to be selective (Tekle et al. 1994
; Gabriel and Teissie, 1997
). These selective transport patterns are consistent with the pore
distributions found here at the membrane level. In other systems where
asymmetry is observed, it is important to note that it may not
necessarily indicate the state of pore population at the membrane
level. Transport through a single large pore versus that which takes
place through many small pores spread over a wide region may lead to
the mistaken conclusion that one side is more porated than the
other. As is shown in the analysis of the lipid loss due to pore
formation in the present system, the total porous area at the two
hemispheres could be quite similar despite the fact that the size of
the pores is markedly different (Fig. 1 and Table 1). It is, however,
possible that such equivalence may not exist at times earlier than we
can detect with our system.
The other important finding in this study is the demonstration
that lipid loss is associated with pore formation (Figs. 1 and 3 and
Table 1). This has not been observed in cellular systems and may be due
to the presence of cytoskeletal and other protein components that may
serve as anchors to stabilize the lipid domains of the membrane. The
electron micrographs of Fig. 3 clearly show that massive restructuring
of the membrane occurs due to electroporation. Although some lipid
components are lost, other small cluster of aggregates remain attached
to the parent vesicle membrane. We should mention here that analogous
loss of lipid components due to external electric field has been
implicated in an earlier study of electric birefringence of a
three-component, sodium
bis(2-ethylhexyl)sulfosuccinate/isooctane/H2O, reverse micelle system (Tekle and Schelly, 1994
).
In summary, we have presented data on the induction and
distribution of membrane pores in DOPC vesicles showing asymmetric distribution of pores at the membrane level with large single pores at
the cathode- and many small pores at the anode-facing hemisphere, with
resultant partial loss of the membrane lipid. The porous areas at the
two hemispheres were, however, comparable, and the pores were shown to
reseal within a few hundred milliseconds after the pulse.
The overall pore closure time can be estimated based on Fick's
law of two-compartment diffusion (Batschelet, 1971
). In general, the
rate of mass accumulation m(t) inside an
enclosure through a time-dependent pore of radius r = r(t) can be written as
We thank Dr. Blair Bowers for taking the electron microscope data
Address reprint requests to Dr. E. Tekle, Laboratory of Biochemistry,
NHLBI, NIH, 50 South Drive, Room 2127, MSC 8012, Bethesda, MD
20892-8012. Tel.: 301-496-8390; Fax: 301-496-0599; E-mail:
ephrem{at}helix.nih.gov.