We describe measurements of lateral diffusion in
membranes using resonance energy transfer. The donor was a rhenium (Re)
metal-ligand complex lipid, which displays a donor decay time near 3 µs. The long donor lifetime resulted in an ability to measure lateral diffusion coefficient below 10
8 cm2/s. The
donor decay data were analyzed using a new numerical algorithm for
calculation of resonance energy transfer for donors and acceptors randomly distributed in two dimensions. An analytical solution to the
diffusion equation in two dimensions is not known, so the equation was
solved by the relaxation method in Laplace space. This algorithm allows
the donor decay in the absence of energy transfer to be
multiexponential. The simulations show that mutual lateral diffusion
coefficients of the donor and acceptor on the order of
10
8 cm2/s are readily recovered from the
frequency-domain data with donor decay times on the microsecond
timescale. Importantly, the lateral diffusion coefficients and acceptor
concentrations can be recovered independently despite correlation
between these parameters. This algorithm was tested and verified using
the donor decays of a long lifetime rhenium lipid donor and a Texas
red-lipid acceptor. Lateral diffusion coefficients ranged from 4.4 × 10
9 cm2/s in
1,2-dimyristoyl-sn-glycero-3-[phospho-rac-(1-glycerol)] (DMPG) at 10°C to 1.7 × 10
7 cm2/s in
1,2-dioleoyl-sn-glycero-3-phosphocholine (DOPC) at 35°C. These results demonstrated the possibility of direct measurements of
lateral diffusion coefficients using microsecond decay time luminophores.
 |
INTRODUCTION |
Fluorescence spectroscopy has been widely used to
measure the dynamic properties of cell membranes (Stubbs et al., 1992
). Typically, the anisotropy decays of nanosecond decay time fluorophores are used to study the order and dynamics of the acyl side chain regions. Hence the anisotropy decays of nanosecond decay time membrane
probes reveals membrane dynamics over distances of 2 to 10 Å. However,
such studies provide no information on the lateral motions of lipids
and proteins in biological membranes (Tocanne et al., 1994
; Zhang et
al., 1993
; Walther et al., 1996
; Edidin et al., 1994
; Simson et al.,
1995
). Fluorescence recovery after photobleaching (FRAP) is presently
the most commonly used method to measure lateral diffusion (Perisamy
and Verkman, 1998
; Lippincott-Schwartz et al., 1999
; Valez and Axelrod,
1988
). When using FRAP one measures the rates at which fluorophores
repopulate a region of the membranes, which was photobleached by an
intense laser light pulse. Hence FRAP measures lateral motions of
lipids on proteins over large macroscopic distances ranging from 5000 to 30,000 Å. Such studies have shown that lipids or proteins do
diffuse in membranes and that the rates of diffusion depend on the
membrane lipid composition and the presence of membrane-bound proteins.
Additionally, such measurements have revealed the presence of mobile
and immobile fractions, the latter probably being proteins whose
motions are restricted by the cytoskeleton matrix.
An alternative approach is to use resonance energy transfer (RET)
to study the membranes (Fung and Streyer, 1978
; Wolber and Hudson,
1979
; Dewey and Hammes, 1986
; Hauser et al., 1976
; Estep and Thompson,
1979
), which can be expected to provide information over distances
comparable with the Forster distance (R0), which are typically in the range of 25 to 60 Å. Such studies provide information on the spatial distribution and/or distance between donor
(D) and acceptors (A). However, it is difficult to use RET to measure
lateral diffusion coefficients. This is because there is little
diffusive motion during decay times of the excited state, which are
typically near 10 ns. The available expression for RET in two
dimensions assumes a static distribution of donors and acceptors (Fung
and Streyer, 1978
; Wolber and Hudson, 1979
; Dewey and Hammes, 1986
;
Hauser et al., 1976
). To the best of our knowledge analytical
expressions for RET in two dimensions with diffusion have not been reported.
In the present report we describe the use of RET to perform direct
measurement of lateral diffusion in membranes. Such measurements are
possible using luminescent metal-ligand complexes with microsecond decay times, many of which have been developed in this laboratory over
the past several years (Terpetschnig et al., 1995
, 1997
; Szmacinski et
al., 1996
; Castellano et al., 1998
; Guo et al., 1997
). Because of the
long decay times one can predict the lipids will undergo significant
lateral motions during the excited state lifetime. It is well known
that donor-to-acceptor motions in three dimensions during the donor
decay time result in an increased efficiency of energy transfer
(Steinberg and Katchalski, 1968
; Stryer et al., 1982
; Thomas et al.
1978
). The availability of metal-ligand complex (MLC)-labeled lipids
with microsecond decay times thus suggests the possibility of measuring
lateral diffusion coefficients in membranes for the time-resolved
resonance energy transfer (RET) data. However, an analytical solution
for the diffusion equations in two dimensions is not known. Hence, we
developed a numerical algorithm that predicts the intensity decays of
the donors in the presence of acceptors randomly distributed in two dimensions with mutual donor-to-acceptor diffusion. We show by simulations and experimental data that lateral diffusion coefficients can be readily recovered from the intensity decay of long-lived donors.
These results demonstrate the possibility of measuring membrane
dynamics on the heretofor inaccessible microsecond timescale over
distances ranging from 20 to 100 Å.
 |
THEORY |
We now describe our method for simulations and analysis the
time-resolved donor decays in the presence of two-dimensional diffusion
of the donors and acceptors. Information about the static distribution
and mutual donor-to-acceptor diffusion coefficient D is
contained in the intensity decay of the donor. We assume that the donor
emission can be observed without contributions from autofluorescence of
the sample or from the fluorescent acceptor. We found that the
intensity decay of the MLC donor in membranes was more complex than a
single exponential even in the absence of acceptors. Hence, it was
necessary to use expressions that account for a multiexponential decay
law in the absence of energy transfer. The donor decays, in the absence
of acceptors, were analyzed using the multiexponential model
|
(1)
|
in which I
is the intensity of the
donor emission at time t = 0, and
Di are
the relative amplitudes (at t = 0) of the components
characterized by the decay times
Di in the absence of
acceptors. The factors
Di are normalized so that
i
Di = 1. Eq. 1 may be rewritten in
the form
|
(2)
|
in which
|
(3)
|
are the intensity decays of the components with each decay time.
Because the decay of the donor is multi exponential even in the
absence of acceptor, it is necessary to consider how energy transfer
affects each component in the decay. We assume that the components
behave as if they each have the same Förster distance (R0) for transfer. Thus for the ith
component of molecules containing a donor and an acceptor separated by
a distance r, the transfer rate is described by
|
(4)
|
This assumption is frequently used when using RET to measure
distance distributions and has not been found to introduce any difficulties (Cheung, 1991
). Using this assumption, the donor decay in
the presence of acceptors (Eq. 3) can be expressed as
|
(5)
|
in which C
is the concentration of
acceptor and Wi(t) is given by
|
(6)
|
We assume here that energy transfer and/or diffusion modify
only the t > 0 part of the donor fluorescence decay
curve and does not change the time-zero characteristics of the decay,
i.e., I
and
i. The form of
Eq. 4 does not depend on the dimensionality on the system. In contrast,
the form of the second order donor-acceptor transfer rates,
ki(t) are dimensionality dependent.
Within a two-dimensional model for energy transfer with diffusion, the
transfer rates ki(t) may be
calculated as
|
(7)
|
in which rmin denotes the distance of
donor-acceptor closest approach and yi(r,
t) satisfy the diffusion equation
|
(8)
|
with parameter D being a sum of the diffusion
coefficients of the donor and acceptor, respectively (D = DD + DA). Functions yi(r, t) have a meaning of ratios of
the mean concentration CAi(r, t) of
acceptor molecules at the distance r from the excited donor of the ith type to the bulk concentration of the acceptor
C
. The initial condition of Eq. 8 is
|
(9)
|
which means the donors and acceptors are randomly
distributed at t = 0. The inner and outer boundary
conditions are
|
(10)
|
|
(11)
|
in which rmin is the distance of closest
approach for the donor and acceptor. Eq. 11 can be understood as a
constant acceptor concentration at long distances from the donor. Eq. 10 is known as the reflection or specular boundary conditions,
which assure that donor-acceptor collisions at r = rmin do not influence the RET process except as due to
the dependence on distance according to Eqs. 4 and 8. That is, the RET
process is the only deactivation channel due to the acceptor.
An analytical solution of Eq. 8 is not known, so the numerical methods
were applied. To minimize the time of calculation of the fluorescence
decays with Eq. 5 we used an algorithm described previously (Kusba and
Lakowicz, 1994
). This algorithm allowed for variable mesh sizes on the
time axis and for exponential approximation of the decay in the
particular time intervals. To evaluate quantities Wi(t) in Eq. 5, we applied a method
similar to that described in (Kusba, 1987
). Each of the decays from Eq. 5 was evaluated for approximately n = 50 time points,
tik, (k = 1..n), inhomogenously distributed on the time axis. The intensity decays were calculated in
Laplace space and inverted using the Stedfest procedure (Lakowicz and
Gryczynski, 1991
). More detail concerning the fluorescence decay
evaluation can be found in the Appendix.
The calculated donor intensity decays were used to predict the phase
and modulation values (Stehfest, 1970
). For a given set of parameter
values, D,
, R0, and
rmin, the donor decays
IDi(t) obtained by the numerical
procedure were used for calculation of the quantities
|
(12)
|
and
|
(13)
|
at given modulation frequency
. The values of
N
and D
are needed
for calculation (c) of the phase angle (
) and modulation (m
) values, which are given by
|
(14)
|
|
(15)
|
The calculated phase (
c
) and modulation
(mc
) values are compared with the
experimental data to determine the diffusion (D) parameter
by the method of nonlinear least squares (Johnson, 1983
; Johnson and
Frasier, 1985
). The goodness-of-fit is characterized by
|
(16)
|
in which v is a number of degrees of freedom, and

= 0.4° and
m = 0.01 are the experimental
uncertainties in the measured phase angles (
) and
modulation (m
), respectively, assumed for
both the simulated and measured values. In the case of global analysis
the sum in Eq. 16 extends over both the modulation frequencies (
)
and over the multiple data sets used in analysis. Typically these data
are donor decays measured with different acceptor concentrations in the
membranes or data with a single acceptor concentration but several temperatures.
The frequency-domain data were also analyzed in terms of
multiexponential model Eq. 1. The fractional intensity of each
component to the steady-state intensity is given by
|
(17)
|
The mean lifetime is given by
|
(18)
|
The confidence intervals on a parameter were determined from the
dependence of 
on this parameter when it was held
constant at a value different from the optimal value. The other
parameters were then allowed to vary to minimize 
with the constant parameter held at its nonoptimal value, yielding

(par). This procedure yields the

surface on the contour plot for the parameter
with the minimum value 
(min) at the optimal
parameter value. The contour plots are usually presented as the ratio
|
(19)
|
The ranges of parameter values consistent with the data were taken
as the intersection of the 
surface with the
F
value for a given probability P,
which we took as P = 0.32,
|
(20)
|
In this expression, F(p, v, P) is the F-statistic with
p parameters and v degrees of freedom with a
probability of P. In our experiments a single frequency
domain data file contains about 40 measurements, 20 phase angles, and
20 modulation values. A global analysis is typically performed with
five data files or 200 measurements. For 40 degrees of freedom and one
variable parameters this value is F
1.03. For 200 degrees of freedom or five variable parameters
F
1.01.
 |
MATERIALS AND METHODS |
The syntheses of
[Re(4,7-Me2phen)(CO)3(4-COOHPy)](PF6),
where 4,7-Me2phen is 4,7-dimethyl-1,10-phenanthroline and
4-COOHPy is isonicotinic acid, and its phospholipid analogue (Re-PE)
serving as the energy donor were described in the previous reports (Li et al., 1999a
, b
). The energy acceptor, N-(Texas Red
sulfonyl)-1,2-dihexadecanoyl-sn-glycero-3-phosphoethanolamine (Tr-PE), was obtained from Molecular Probes (Eugene, OR) and used as
received. These chemical structures are shown in Fig.
1. The spectral properties of the donor
and acceptor resulted in R0 values near 35 Å,
depending on the temperature. Under our experimental conditions the
Förster distances (R0) from Re-PE to Tr-PE
energy transfer ranged from 28 to 37 Å.
1,2-dioleoyl-sn-glycero-3-phosphocholine (DOPC) and
cholesterol were from Sigma Chemical Co (St. Louis, MO) and
1,2-dimyristoyl-sn-glycero-3-[phospho-rac-(1-glycerol)] (DMPG) was obtained from Avanti Polar Lipids (Alabaster, AL). All
solvents and reagents were obtained from Aldrich and used without
further purification. Water was deionized with a Milli-Q purification
system.

View larger version (23K):
[in this window]
[in a new window]
|
FIGURE 1
Molecular structure of the energy donor (Re-PE) and
acceptor (Tr-PE). The lower panel shows the donor emission and acceptor
absorption spectra. The shaded area in the overlap intergral
corresponds to a R0 value near 35 Å.
|
|
Preparation of model membranes
Lipid vesicles were prepared by the usual procedure of mixing
and sonication. Appropriate amounts of the donor and acceptor phospholipids, DMPG, DOPC, and/or cholesterol in CHCl3 were
taken from stock solutions, and the solvent was removed by a stream of
argon. The molar ratio of Re-PE to Tr-PE was kept constant at 4.5:1
whereas the amount of unlabeled phospholipid was varied to obtain molar
ratios of Tr-PE to DOPC ranging from 0 to 0.02. Vesicles were prepared
by sonication under an atmosphere of argon in 0.1 M sodium phosphate
buffer, pH 7.2, at final lipid concentrations ranging from 0.5 to 7.0 mg/ml. Using this preparation procedure, the vesicle diameter is
between 200 and 250 Å as determined through anisotropy measurements
using the long lifetime ruthenium complex lipid (Li et al., 1997
). For
all simulations and analyses we assumed the minimum donor-to-acceptor
distance was rm = 7 or 8 Å as indicated in
the text. The area occupied per lipid molecules was assumed to be 74 Å2/lipid molecules for DOPC at all temperatures, 48, 55, and 62 Å/lipid molecule for DMPG at 10.23 and 35 ÅC, respectively,
and 55 Å/molecule for cholesterol at all temperatures (Marra, 1986
). The Förster distance (R0) was calculated
using the known equations and the uncorrected emission spectra of the
Re-PE labeled vesicles. Quantum yields of Re-PE were measured relative
to 3-aminofluoranthene in dimethyl sulfoxide with an assumed quantum
yield of 0.32 (Gryczynski et al., 1997
). For calculating the overlap
intergral we used the absorption spectrum of Tr-PE with a maximum
extinction coefficient of 109,000 M
1 cm
1 at
583 nm.
Instrumentation
Absorption and emission spectra were recorded on a HP 8453 diode
array spectrophotometer and a SLM AB2 fluorimeter under magic angle
polarization conditions, respectively. The frequency-domain fluorimeter
(ISS, Koala) used 325-nm excitation from a HeCd laser (Liconix, 20 mW).
This laser was passed through a Pockels cell operated from an ISS low
frequency amplifier (K2.LF), which provided modulated light from 3 kHz
to 2.5 MHz. Two PTS frequency synthesizers (PTS-500) were used to
modulate the Pockels cell and detection system. For fluorescence
intensity measurements, a 500-nm cutoff filter (500FH90-50S) and two
short-wavelength pass filters (550FL07-50S) from Andover (Salem, NH)
were used to isolate the donor emission from that of the acceptor.
 |
RESULTS |
Effect of two-dimensional diffusion on the transfer efficiency
Prior to considering the intensity decays of the donors it is
informative to examine the effect of two-dimensional diffusion on the
overall transfer efficiency. The transfer efficiency (E) was
calculated from the integral of the time-dependent donor decay in the
absence (ID(t)) and presence
(IDA(t)) of acceptors using
|
(21)
|
in which the superscript A indicates the presence of acceptor.
Fig. 2 shows the transfer efficiencies
for various assumed donor decay times and mutual diffusive
coefficients. The acceptor density was assumed to be 5 × 10
3 acceptors/lipid. For typical donor decay times near
10 ns even rapid lateral diffusion at 10
6
cm2/s will not effect the transfer efficiency. As the donor
decay time increases the transfer efficiency increases. For very long decay times and/or rapid lateral diffusion the transfer efficiency approaches a limiting value of 91.4% (Fig. 2, - - -). This is the
rapid diffusion limit at which the transfer efficiency is determined by
the distance of closest approach between the donor and acceptor. For
spherical donors and acceptors in two dimensions the diffusion limited
value of the rate of energy transfer kT is given
by
|
(22)
|
in which
A = 6.7 × 10
5
molecules/Å2 is the density of acceptors,
rmin = 7 Å is the distance of closest
approach, R0 = 25 Å, and
D = 3 µs is the donor decay time (Thomas and
Stryer, 1982
; Stryer et al., 1982
; Lakowicz, 1999
). Using these values
one can calculate the diffusion limited value of
kT to be 3.55 × 106
s
1 = 10.65/
D. The diffusion limited
transfer rate is ~10-fold larger than the donor-donor decay time.
This value of kT can be used to calculate the
transfer efficiency of 91.4% using
|
(23)
|
Additional information is available if the donor decay times are
intermediate between the static and rapid diffusion limit, where the
transfer efficiency depends on the mutual donor-to-acceptor diffusion
coefficient (Fig. 2). In particular, for decay times from 1 to 100 µs, lateral diffusion coefficients from 10
9 to
10
6 cm2/s result in increased transfer
efficiency. These results suggest that the time-dependent donor decays
can be used to measure the rate of donor-to-acceptor diffusion in
membranes.

View larger version (27K):
[in this window]
[in a new window]
|
FIGURE 2
Effect of the lateral diffusion coefficient and the
donor decay time on the steady-state RET transfer efficiency. For these
simulations we assumed rmin = 7 Å,
R0 = 25 Å, 75 Å2 per lipid molecule
and 5 × 10 3 acceptor per lipid molecule.
|
|
Simulated time-dependent donor decay with RET and two-dimensional
diffusion
We used our numerical algorithm for RET in membranes to simulate
the frequency-domain intensity decays. The frequency responses were
simulated using assumed values of R0, lipid
area, and acceptor density. Simulated data for unquenched donor decay
times (
D) of 3 and 30 µs are shown in Fig.
3. The solid and dashed lines show the
donor decays in the presence and absence of diffusion, respectively.
These frequency responses show that faster diffusion shifts the
response to higher frequencies and shorter mean donor decay times. For
the same diffusion coefficients longer donor decay times result in
larger shifts, as expected with the additional time for D-to-A
diffusion. Analytical expressions are available for RET in two
dimensions without diffusion (Wolber and Hudson, 1979
; Dewey and
Hammes, 1986
; Hauser et al., 1976
; Estep and Thompson, 1979
). We
confirmed that the frequency responses calculated using our algorithm
without diffusion (- - -) were equivalent to the known analytical
expressions.

View larger version (41K):
[in this window]
[in a new window]
|
FIGURE 3
Simulated frequency-domain intensity decays for donors
and acceptors randomly distributed in two dimensions. For the
simulation, R0 = 25 Å rmin = 7 Å, 75 Å2/lipid molecule,
and 5 × 10 3 acceptors/lipid molecules. For
D = 5 × 10 8 cm2/s the
solid lines show the simulated phase and modulation values. The dashed
lines show the expected donor decays without lateral diffusion, and the
dotted lines show the donor decays in the absence of acceptors. The
shaded area shows the contribution of lateral diffusion to increasing
the rate of energy transfer.
|
|
In an experimental analysis we wish to calculate the diffusion
coefficient from the time-resolved data. The ability to recover D from
the frequency-domain data is not obvious because increased diffusion
coefficients and increased acceptor densities both result in higher
amounts of energy transfer. Stated alternatively, one can expect the
diffusion coefficient and acceptor density to be correlated parameters
(Johnson, 1983
). Hence we analyzed the frequency-domain data simulated
with various acceptor densities and a mutual diffusion coefficient of
10
7 cm2/s (Table
1). One notices that the correct values
of both D and the A density were recovered over a range of acceptor
densities. The uncertainty in the recovered parameter is reasonably
small. These values were obtained with both the diffusion coefficients and acceptor concentrations as variable parameters. Because higher diffusion coefficients and higher acceptor concentrations both result
in an increase in the transfer efficiency these parameters are expected
to be correlated. More specifically, an increase in the diffusion
coefficient can be compensated for by a decrease in the acceptor
concentration and vice versa. Correlation between parameters results in
wider confidence intervals than for uncorrelated parameters (Johnson,
1983
; Johnson and Frasier, 1985
). Even with consideration of
correlation the confidence intervals are small, ~10% of the assumed
values. The ability to recover both the diffusion coefficients and the
acceptor density can be understood as the effects of diffusion on the
form of the intensity decay. Energy transfer to a static distribution
of acceptors, in 1, 2 or 3 dimensions, results in nonexponential decays
(Thomas and Stryer, 1982
). As the diffusion coefficient increases the
donor decays become shorter but more like a single exponential
(Lakowicz, 1999
). It is this dependence on the form of the intensity
decay on the diffusion coefficient, which allows recovery of both D and
A density from the time-resolved data.
We used simulations to estimate the lower limit of the diffusion
coefficient, which we expect to be detectable with a 3-µs decay time
donor (Fig. 4). The shaded areas indicate
the contribution of diffusion to increasing the donor decay rate. These
simulations show that diffusion coefficient as low as 5 × 10
9 cm2/s still visually alter the donor
decays. We performed additional simulations to more quantitatively
predict the lower limit of the diffusion coefficient detectable with a
3-µs decay time donor. This was accomplished by simulating data for
assumed diffusion coefficients and acceptor concentrations. The
simulated data were then analyzed to recover the 
values with the diffusion coefficient and acceptor density as variable
parameters (
) or with the diffusion coefficient
set to zero 
(D = 0). This
difference between these fits reflects the possibility of compensating
for a higher diffusion coefficient with a lower acceptor concentration.
In these analyses the acceptor concentrations were either held fixed at
the assumed value (Fig. 5, top) or taken
as a variable parameter (Fig. 5, bottom). When the acceptor
concentration is fixed the relative elevation of the

ratio indicates the overall contribution of
acceptor presence and diffusion on the donor decay. In this case one
sees that diffusion coefficients less than 10
9
cm2/s are detectable (Fig. 5, top). If the acceptor
concentration is a variable parameter the lowest detectable diffusion
coefficient is again near 10
9 cm2/s (Fig. 5,
bottom). With the acceptor concentration as a variable parameter the
increase in 
are ~10-fold less than with a known
acceptor concentration. Nonetheless, the relative 
values are significantly elevated, which reflect the effect of
diffusion on changing the shape of the frequency response as well as
its position on the frequency axis. In both cases somewhat lower
diffusion coefficients are detectable at higher acceptor
concentrations.

View larger version (24K):
[in this window]
[in a new window]
|
FIGURE 4
Simulated donor intensity decay for a donor decay time
0 = 3 µs and various lateral diffusion
coefficients. For these simulations R0 = 25 Å,
rmin = 7 Å, 75 Å2 per lipid
molecule, and 5 × 10 3 acceptors per lipid
molecule.
|
|

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 5
Relative  values for data
simulated with D-to-A diffusion and an acceptor-to-lipid ratio of 0.02. The simulated data were analyzed with and without D-to-A diffusion. The
simulated data were analyzed with the acceptor concentration fixed at
the known value (top) or as a floating parameter
(bottom). The assumed values are D = 3 µs, R0 = 25 Å, rmin = 7 Å, and 75 Å2 per lipid molecule. For D = 10 9 cm2/s the ratio of
 (D = 0)/
with the acceptor concentration fixed at 0.02 is 8.8 (top).
With the acceptor concentration floating the ratio is 3.4.
|
|
Experimental measurements of RET in membranes
Prior to measurement of RET in membranes we examined the
intensity decays of Re-PE in lipid bilayers in the absence of acceptor (Fig. 6 and Table
2). The Re-PE intensity decays were found
to be described by a double or triple exponential decay with mean decay
times ranging from 2.2 to 0.5 µs. The longest mean decay times were
observed in the absence of cholesterol at the lower temperature. These
intensity decays in Table 2 were used as fixed values when analyzing
the donor decays in the presence of the RET acceptors.
We next examined the donor decays in the presence of the Tr-PE
acceptor. These measurements were performed in unsaturated DOPC
vesicles (Fig. 7), saturated DMPG
vesicles (Fig. 8) and in DMPG vesicles
containing cholesterol (Fig. 9). In all
cases the donor frequency responses were adequately fit to our model
with the diffusion coefficient and the acceptor density as floating parameters (Table 3). The contributions
of diffusion-enhanced RET to the donor decays can be seen by comparing
the measured response with that calculated with D = 0
(- - -). In DOPC mutual donor-to-acceptor diffusion makes a substantial
contribution towards increasing the donor decay rate at all
temperatures from 10 to 35°C (Fig. 7). In the case of DMPG the effect
of diffusion is minimal below its transition temperature of 23°C, and
the effect of diffusion is evident at 35°C (Fig. 8). In the case of
DMPG vesicles, which also contain cholesterol the contribution of
diffusion is modest but visible from 10 to 35°C.

View larger version (30K):
[in this window]
[in a new window]
|
FIGURE 7
Re-PE donor decays in DOPC in the presence of a 0.02 mol fraction of Tr-PE acceptor. The solid line shows the best fit with
the diffusion coefficient and acceptor density as variable parameters.
The dash line shows the predictive response at the known acceptor
density and the diffusion coefficient set equal to zero.
|
|

View larger version (25K):
[in this window]
[in a new window]
|
FIGURE 8
Re-PE donor decays in DMPG in the presence of a 0.02 mol fraction of Tr-PE acceptor. The DMPG/cholesterol ratio is 4:1. See
Fig. 7.
|
|

View larger version (24K):
[in this window]
[in a new window]
|
FIGURE 9
Re-PE donor decays in DMPG/cholesterol (mol/mol) in the
presence of a 0.02 mol fraction of Tr-PE acceptor. See Fig. 7.
|
|
Fig. 10 summarizes the lateral
diffusion coefficients obtained from the RET data. These diffusion
coefficients were recovered from the global analysis at three
temperatures with the single acceptor concentration in the membrane as
a global parameter (Table 3). The largest diffusion coefficients were
observed in DOPC bilayers, but the effect of temperature was modest
(Fig. 10). Lower diffusion coefficients were observed in DMPG bilayers,
but these values were strongly dependent on temperature. The presence
of cholesterol in the DMPG bilayers resulted in an intermediate rate of
diffusion and in a weaker dependence of the lateral diffusion coefficient on temperature.

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 10
Temperature-dependent lateral diffusion coefficient in
membranes as observed using the Re-PE and Tr-PE donor-acceptor pair.
The diffusion coefficients are from the global analysis at three
temperatures and one lipid concentration (Table 3). The error bars
represent the confidence intervals obtained from the global analysis at
one acceptor concentration and three temperatures (Figs. 11 and 12).
|
|
We questioned the uncertainty in the recovered values of the diffusion
coefficients. As has been discussed previously, there is no
analytically correct method to calculate the uncertainty for parameters
recovered from nonlinear least squares analysis (Johnson, 1983
; Johnson
and Frasier, 1985
). However, these uncertainties can be obtained by
examination of the 
surfaces or contour plots. To
construct such a plot one repeats the least squares analysis with a
chosen parameter value fixed at a value near to but not at the value
yielding the minimum value of 
. The other
parameters are allowed to vary to minimize 
. This
new 
value represents that found for the fixed
parameter value with adjustment of all the other parameters to improve
the fit. Because the other parameters can vary this procedure accounts
for correlation between the parameters. If the parameters are
completely correlated, such as an product of two terms, then changes in
one parameter can be completely compensated by a change in the second parameter.
The 
surfaces for RET in the DOPC and DMPG
vesicles are shown in Figs. 11 and
12, respectively. For a single
temperature the 
surfaces for both the diffusion
coefficient and acceptor concentrations are well defined. This is an
important result, which indicates the diffusion coefficients and
acceptor concentrations are not completely correlated. Furthermore, the
range of diffusion coefficients and acceptor concentrations consistent
with the data are relatively small. For instance, for DOPC at 20°C
these values range from 7.7 × 10
8 to 6.1 × 10
8 cm2 and 0.0174 to 0.0192 acceptors per
lipid. These results mean that the RET data can be used to recover both
values. This possibility will become important in studies of isolated
cell membranes or intact cells where it is not always possible to know
the probe concentrations.

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 11
Resolution of the lateral diffusion coefficient and
acceptor concentration in DOPC vesicles as seen for the
 surfaces. The values  are
the  values normalized to the minimum
 value. The solid and darker dashed lines
represent the global and nonglobal fits, respectively. The dotted and
lighter dashed lines represents the upper value of the
 ratio consistent with the data.
|
|

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 12
Resolution of the lateral diffusion coefficient and
acceptor concentration in DMPG vesicles as seen for the
 surfaces. The values of  are
from the global analysis at three temperatures. The solid and darker
dashed lines represent the global and nonglobal fits, respectively. The
dotted and lighter dashed lines represents the upper value of the
 ratio consistent with the data.
|
|
The confidence intervals for the diffusion coefficient and acceptor
concentrations can be decreased by a global analysis. In these cases
(Figs. 11 and 12) we used the acceptor concentration as a global
parameter and allowed the diffusion coefficients to be different at
each temperature. The 
surfaces (
) were
calculated for changing just one of the three diffusion coefficients or
the acceptor concentration. The confidence intervals are smaller than
for a single data file mostly because of the lower

ratio (
) needed for significance
because of the larger number of data points. In these cases global
analysis results in an approximate twofold decrease in the confidence
interval. These ranges are shown in Fig. 10. It is interesting to
notice that the largest confidence interval was found for DMPG 10°C.
This can also be seen from the 
surface in Fig.
12. In this lipid the rate of diffusion below the phase transition
temperature is slow, even on the microsecond timescale. The small
contribution of diffusion to the donor decay results in a larger
uncertainty in this slower diffusion coefficient.
It is informative to question whether the data are sensitive to the
dimensionality of the system. More specifically, can the data
distinguish between donors and acceptors distributed randomly in two or
three dimensions. The frequency-domain data were analyzed in terms of a
model for diffusing donors and acceptors in three dimensions (Kusba et
al., 2000
). These analyses result in 
values,
which are somewhat elevated over our two-dimensional model (Table 3).
However, most of the 
values are reasonably low,
and the fits may be regarded as adequate. However, despite these low

values, the results of the analyses are
unacceptable because the answers are not reasonable. These analyses
result in acceptor concentrations of 12 to 80 mM and diffusion
coefficients ranging from 10
7 to 10
15
cm2/s. The acceptor concentrations are unacceptably large
and much higher than the known bulk concentration of the acceptors.
And finally we questioned whether the frequency-domain donor decays,
measured for multiple acceptor concentrations, would result in improved
resolution of the diffusion coefficients. The donor decays from a range
of acceptor concentrations in DOPG vesicles are shown in Fig.
13. The solid lines show the best fit
when the acceptor concentrations are variable and there is a
single global diffusion coefficient. Except for an acceptor density of
0.005, the data are well matched with a single diffusion coefficient. The 
surfaces shows little if any decrease in the
confidence intervals. Hence, there seems to be little advantage to
measuring the donor decay at multiple acceptor concentrations.

View larger version (33K):
[in this window]
[in a new window]
|
FIGURE 13
Frequency-domain intensity decay of Re-PE in DOPC with
various concentrations of acceptor. The solid line shows the global
analysis using a single diffusion coefficient. From left to right the
acceptor densities were 0.025, 0.005, 0.01, 0.015, and 0.02 acceptors
per lipid.
|
|
 |
DISCUSSION |
A variety of methods have been used to study lateral diffusion in
membranes (Tocanne et al., 1994
). The fluorescence methods include
quenching, pyrene excimer formation, and RET. Collisional quenching
requires close contact between the fluorophores and quencher. As a
result most fluorophores in membranes except pyrene are not
significantly quenched by reasonable quencher concentrations. Pyrene
excimer formation is useful but limited to pyrene and closely related
fluorophores, limited by the approximate 200-ns decay time of pyrene in
membranes, by the existence of preformed complexes, and by the
possibility that the efficiency of excimer formation in different
membrane environments. In contrast, RET seems to be preferred over
quenching and excimer formation. RET is a through-space interaction
that occurs over longer distances and is not affected by the local or
intervening environment. Additionally, longer lifetimes and a wide
range of spectral properties can be obtained from the extensive
literature on metal-ligand complexes (Kalayanasundaram, 1992
; Juris et
al., 1988
; Demas and DeGraff, 1997
). Still longer decay times near 0.5 to 3.0 ms can be obtained with lanthanide chelates (Sabbatini and
Guardigli, 1993
; Li and Selvin, 1995
; Martin and Richardson, 1979
;
Horrocks and Sudnick, 1981
; Chen and Selvin, 2000
). Given the
availability of theory to predict two-dimensional RET with diffusion,
we believe that RET with long decay time donors offers considerable
promise for studying lateral transport in membranes.
It is reasonable to question whether long lifetime donors can be used
in living cells. The use of cells raises two issues: uptake of the
labeled lipid and detectability of the emission. Although not easy,
cell membranes are frequently labeled with fluorescent fatty acids and
lipids (Fulbright et al., 1997
; Zucker, 2001
; Tocanne et al., 1994
;
Struck and Pagano, 1980
; Tanhuanpää and Somerharju, 1999
).
Because the metal-ligand complexes display good water solubility, it is
probable that labeling with MLC-lipids can be accomplished. However,
the MLC probes are less bright than typical organic fluorophores due to
their lower extinction coefficient and lower quantum yields. The low
extinction coefficients can be circumvented by the use of tandem
fluorophores (Tyson and Castellano, 1999a
,b
; Zhou et al., 2000
). These
probes contain a high extinction coefficient absorber, which donates
the energy to the MLC acceptor with high efficiency. Additionally, it
is now known that the effective quantum yield of MLCs can be increased
by RET to high quantum yield acceptors (Lakowicz et al., 2001
; Maliwal
et al., 2001
). Hence it appears likely that RET from long lifetime
donors can be used with living cells.
In the future one can imagine the use of microsecond and
millisecond decay time donors to study slower or more complex motions in membranes. As examples, it is known that membranes form domains of
solid and liquid phases (Thompson et al., 1995
; Jorgensen et al., 1996
;
Gheber and Edidin, 1999
; Matko and Edidin, 1997
). Also, lateral
diffusion of lipids is expected to depend on the presence of proteins,
which slow lipid diffusion or prevent the motions of proteins (Marguet
et al., 1999
; Kenworthy and Edidin, 1998
). Analysis of the motions in
such systems will require further development of the theory and
programs to extract the relevant
molecular information from the intensity decays.

View larger version (15K):
[in this window]
[in a new window]
|
FIGURE 14
 surfaces for the lateral
diffusion coefficient and acceptor concentrations in DOPC at 20°C.
The bold solid lines and bold dashed lines represent the global fit
with a single diffusion coefficient to the data at 5 or 1 acceptor
density, respectively. The lighter dotted and dashed lines indicate the
confidence intervals for the fits with 5 or 1 acceptor densities,
respectively.
|
|