Department of Biochemistry and Immunology, Cellular and Molecular
Sciences Group, St George's Hospital Medical School, University of
London, London SW17 0RE, United Kingdom
Surface charge in track-etched polyethylene
terephthalate (PET) membranes with narrow pores has been probed with a
fluorescent cationic dye (3,3'-diethyloxacarbocyanine iodide
(diO-C2-(3))) using confocal microscopy. Staining of
negatively charged PET membranes with diO-C2-(3) is a
useful measure of surface charge for the following reasons: 1) the dye
inhibits K+ currents through the pores and reduces their
selectivity for cations; 2) it inhibits
[3H]-choline+ transport and promotes
36Cl
transport across the membrane in a pH-
and ionic-strength-dependent fashion; and 3) staining of pores by
diO-C2-(3) is reduced by low pH and by the presence of
divalent cations such as Ca2+ and Zn2+.
Measurement of the time dependence of cyanine staining of pores shows
fluctuations of fluorescence intensity that occur on the same time
scale as do fluctuations of ionic current in such pores. These data
support our earlier proposal that fluctuations in ionic current across
pores in synthetic and biological membranes reflect fluctuations in the
surface charge of the pore walls in addition to molecular changes in
pore proteins.
 |
INTRODUCTION |
Certain properties of solutions, such as
viscosity (Granick, 1991
; Bhushan et al., 1995
), with the exception of
water (Raviv et al., 2001
), and ion conductance (Schwartz, 1962
;
Lakshminarayanaiah, 1969
; Manning, 1969
), deviate from the value in
bulk solution when measured near a surface. In the case of ion
conductance, an abnormally high value in a region that is within a
Debye length or so of a charged surface, termed surface conductance,
has been ascribed to an accumulation of counterions (Spitzer, 1984
). We have suggested that fluctuation of conductance in synthetic membrane pores (Lev et al., 1993
) and other materials (Sachs and Qin, 1993
) in
which surface conductance contributes significantly to total conductance, is due to oscillation in the concentration of counterions along the surface of the pore (Korchev et al., 1997
) and that this
results from a fluctuation in surface charge. Here we document this effect.
Track-etched polyethylene terephthalate (PET) membranes bear fixed
negative charges at their surface and within pores, due to the free
carboxyl groups that are generated by alkaline hydrolysis of ester
bonds during the etching process. We have calculated that there are
~1.6 such ionizable groups per nm2 of PET
(Korchev et al., 1997
); this density happens to approximate the number
of charged groups at the surface of a phospholipid bilayer. Such
charges attract positively charged counterions from solutions in which
the membrane is bathed. As a result, the conductance through narrow
pores in which the surface area of the pore is large in comparison with
its volume, is anomalously high (Korchev et al., 1997
). That is, the
conductance is made up of two quantities: the conductance of the bulk
solution plus the surface conductance. With 2-3-nm-diameter pores and
solutions of low ionic strength (<0.5 M) at pH 7, surface conductance
exceeds bulk conductance by more than 100-fold (Lev et al., 1993
);
under these conditions the transference number
t+ > 0.9 and the pore is cation
selective (Lev et al., 1993
). In wider pores, bulk conductance swamps
surface conductance, which then contributes relatively little to total conductance, so that t+ approaches 0.5 (equal contribution to current by cations and anions). Although we
accept that there is no theoretical relationship between conductance
and selectivity, or between conductance and permeability for that
matter, our results show that in PET membranes conductance and
selectivity (Korchev et al., 1997
), as well as conductance and
permeability (Rostovtseva et al., 1996
), do alter in concert. A
diminution of surface conductance relative to bulk conductance is also
seen in narrow pores at high ionic strength, at which the accumulation
of counterions by the fixed negative charges of the PET membrane is
saturated, or screened. Surface conductance in narrow pores is
depressed also at low pH due to protonation of the carboxyl groups, by
methylation of carboxyl groups (Pasternak et al., 1995
), and by the
presence of divalent cations that bind to carboxyl groups (Lev et al.,
1993
). Here we demonstrate that the binding of a counterion, namely,
the charged dye 3,3'-diethyloxacarbocyanine iodide, indeed fluctuates
in a narrow PET pore. This effect, in addition to that due to molecular changes in protein architecture, needs to be taken into account when
defining the origin of fluctuations in endogenous ion channels and in
pores induced across the plasma membrane of susceptible cells by
bacterial toxins (Bashford and Pasternak, 2000
).
 |
MATERIALS AND METHODS |
Sheets of PET membranes were exposed to a beam of high-energy
nuclei applied at right angles to create tracks through the material,
which is then etched in warm alkali to expand tracks into pores (Spohr,
1990
). The material we used was ~10 µm thick and had a pore density
of 1.1 × 106 cm
2.
This was measured by exposing one of many membranes being irradiated simultaneously to a long period of etching, so that pores become large
enough to be visible by negative-stain electron microscopy and can
therefore be counted. The pores had a nominal diameter of 20 nm as
measured by gas diffusion across dry membranes (P.Y. Apel, Joint
Institute for Nuclear Research, Dubna, Russia, 1993, personal
communication), but as assessed in the wet state by the diffusion of small molecules and ions (Rostovtseva et al., 1996
) the
average effective diameter was 2-3 nm. The cyanine dye
diO-C2-(3) (Fig. 1)
(Sims et al., 1974
) was obtained from Molecular Probes (Eugene, OR);
fluorescein (Fig. 1) was from BDH (Poole, U.K.).
To measure ion currents and ion selectivity PET membranes were clamped
across the orifice (diameter
0.5 cm) between two Teflon
chambers using high-vacuum silicone grease to prevent leakage. Solutions containing KCl and potassium phosphate buffers without or
with diO-C2-(3) were added to each chamber, and
ion current and ion selectivity were determined via Ag/AgCl electrodes
(Lev et al., 1993
). Reversal potentials were recorded directly by
operating the apparatus in (zero) current-clamp rather than
voltage-clamp mode. We used a fast integrative capacitor in the
feedback loop (Korchev et al., 1997
); the output voltage of the
integrator was used to control the membrane potential via virtual
ground of the head amplifier to keep zero current through the membrane
and to monitor the reversal potential continuously.
To measure the flow of radioactive ions across PET, the membrane was
placed in a flow dialysis chamber (Rostovtseva et al., 1996
). The upper
chamber, which was stirred continuously, contained 0.5 ml of the
solution containing radioactive ions; an identical solution without
radioactivity was pumped through the lower chamber at a constant rate
(~0.25 ml/min) and collected as 0.5-ml fractions. The area of the
membrane in contact with the solution in both chambers was 0.5 cm2. Radioactivity was assessed in each fraction
by liquid scintillation counting. Radioactivity in the upper solution
was sampled from time to time by removing 5-µl samples for liquid
scintillation counting.
For confocal microscopy, small pieces of PET were immersed for 5 min in
phosphate- or Hepes-buffered solutions containing dye (~5 µg/ml)
and adjusted to the requisite pH, placed on a glass slide, and sealed
beneath a coverslip with nail varnish. A Zeiss LSM 410 Invert confocal
microscope was used to examine the samples with a laser beam
transmitting at 488 nm and recording fluorescence above 515 nm. Samples
were first scanned in the xy plane to locate suitably
stained pores for study. A scan in the z plane was performed to check the penetration of pores by dye. An area in the xy
plane of 8 × 400 pixels was then chosen and exposed to a sequence
of 30 or 50 laser pulses of ~1 s duration at 3-s intervals. The
resulting images at a point half way through the membrane were analyzed in a time sequence as shown in the figures. Fluorescence intensity of
individual pores during the time sequence in the same optical section
was tabulated and their correlation coefficient computed by standard
methods (Excel 97). Flare from staining at either surface (in the case
of diO-C2-(3)) or from the solution (in the case
of fluorescein) prevented analysis of images at the surface of the membrane.
 |
RESULTS |
To validate the assumption that binding to PET membranes of the
cationic dye diO-C2-(3) mimics the attraction of
inorganic and other organic cations, we have examined the effect of the dye on the anomalously high flow of cations through narrow pores. First, we show that diO-C2-(3) indeed competes
with K+ in so far as it reduces ion current (Fig.
2 a) and ion selectivity (Fig.
2 b). Both results are compatible with the suggestion that dye binds to negatively-charged sites on the membrane, to which K+ is attracted. Another way to measure ion flow
through PET membranes is to assess the rate of equilibration of
radioactive ions across the membrane by the flow dialysis technique
(Rostovtseva et al., 1996
). Fig. 3
(panel i) shows that the flow of
[3H]choline+ is inhibited
by diO-C2-(3) whereas that of
36Cl
, which is more than
sixfold slower as anticipated from selectivity measurements (Fig. 2
b), is slightly increased. Raising ionic strength would be
expected to negate this effect (Lev et al., 1993
), which is what is
observed (Fig. 3, panel ii; note that in this
panel the ratio of 3H to
36Cl is plotted, to show the effect more
clearly). We conclude that diO-C2-(3) is a useful
cation with which to probe the movements of water-soluble ions.

View larger version (15K):
[in this window]
[in a new window]
|
FIGURE 2
Effect of diO-C2-(3) on ion currents and
reversal potential in track-etched PET membranes. (a)
The PET membrane contained 1.1 × 106
pores/cm2; nominal pore diameter was 0.02 µm.
Vapplied refers to the virtual ground side
of the chamber, which contained 0.01 M KCl and 0.001 M
K3PO4, pH 10.4. The other chamber contained 0.1 M KCl and 0.001 M K3PO4, pH 10.4, without or
with 20 µM diO-C2-(3) (panel
i) or with 100 µM diO-C2-(3)
(panel ii). (b) The
PET membrane contained 3 × 108 pores/cm2;
nominal pore diameter was 0.02 µm. Reversal potential was monitored
with the virtual ground side of the chamber containing 0.01 M KCl and
0.001 M K3PO4, pH 10.4, and the other chamber,
to which diO-C2-(3) was added to give the final
concentration indicated, containing 0.1 M KCl and 0.001 M
K3PO4, pH 10.4.
|
|

View larger version (13K):
[in this window]
[in a new window]
|
FIGURE 3
Effect of diO-C2-(3) on ion fluxes across a
track-etched PET membrane. The PET membrane contained 3 × 108 pores/cm2; nominal pore diameter was 0.02 µm. (Panel i) The medium contained 0.01 M choline chloride, 0.001 M K3PO4, pH 10.4. [3H]choline ( ) and 36Cl
( ) were added to the upper chamber and flow dialysis (2 min per
fraction) performed without or with 2 µM (i), 10 µM
(ii), 30 µM (iv), and 110 µM
(v) diO-C2-(3); or DMSO
(iii). (Panel ii) The
medium contained 0.01 M ( ) or 0.1 M ( ) choline chloride, 0.001 M
K3PO4, pH 10.4; or 0.01 M ( ) choline
chloride, 0.001 M KH2PO4, pH 4.5. [3H]choline and 36Cl were added
to the upper chamber and flow dialysis (2 min per fraction) performed
without or with 2 µM (i), 10 µM (ii),
30 µM (iv), and 110 µM (v)
diO-C2-(3); or DMSO (iii).
|
|
We next investigated the binding of diO-C2-(3) by
direct measurement of its fluorescence. The experiments were carried
out at different pH values and in the presence of divalent cations, as
these parameters are known to affect ion flow (Lev et al., 1993
). The
results are illustrated in Fig. 4,
a-e. The horizontal bands at the top and bottom
of each picture reflect the binding of the dye to the two surfaces of
the membrane; their width is due to the flare of the fluorescence at
high dye concentration. Staining is more intense at pH 11 (Fig. 4
a) than at pH 5 (Fig. 4 b) (the fluorescence of
diO-C2-(3) itself is unaffected by pH). Staining
within the membrane occurs only in certain regions, which are
interpreted to indicate the presence of pores; similar results are
obtained with negative-stain electron microscopy. The solution (the gap
between the membrane and the coverslip) is seen to be unstained, but
the dye binds to the coverslip (topmost part of Fig. 4 b) in
addition to both membrane surfaces. As anticipated, divalent cations
reduce the intensity of bound dye at the surface and within pores (Fig.
4, c-e: c, control; d, 0.1 mM ZnSO4; e, 5 mM
CaCl2). Note that the width of bound dye at
the surface, due to flare from a high concentration of dye, is
decreased by divalent cations. The dimmer staining and broadening of
the lower membrane/solute interface relative to the upper interface is
due, in part, to an optical effect (spherical aberration) arising from the greater separation of the objective lens from the lower interface. The results of Fig. 4 precisely mimic the effects of pH and divalent cations on surface conductance and give us confidence that the fluorescence of diO-C2-(3) is indeed a measure of
surface charge. It is reinforced by the fact that staining with a
negatively charged dye like fluorescein gives a pattern that is the
opposite of that generated by diO-C2-(3): there
is intense staining of the solution but little of the membrane; few
pores are visible (Fig. 4, f and g). At pH 11 (Fig. 4 f), staining of both solution and membrane is more
intense than at pH 5 (Fig. 4 g) because the fluorescence of
fluorescein itself is reduced at low pH.

View larger version (66K):
[in this window]
[in a new window]
|
FIGURE 4
DiO-C2-(3) and fluorescein staining of PET
membrane pores. PET membranes containing 1.1 × 106
pores/cm2 with diameters of ~20 nm were bathed in the
following solutions with 10 6 M diO-C2-(3):
0.1 M K3PO4, pH 11 (a); 0.1 M
KH2PO4, pH 4 (b); 0.1 M KCl,
0.005 M Hepes, pH 7.4 without (c) or with
10 4 M ZnSO4 (d) or 0.005 M
CaCl2 (e). The same membranes were bathed in
the following solutions with 5 µg/ml fluorescein in 1 M KCl: 0.01 M
K3PO4, pH 11 (f); 0.01 M
KH2PO4, pH 4 (g). Images were
obtained with a Zeiss LSM 410 invert confocal microscope using a
×63/1.4(NA) oil-immersion objective; fluorescence above 510 nm was
excited at 488 nm.
|
|
The effect of a timed sequence of laser pulses using
diO-C2-(3) is illustrated in Fig.
5, a-c. The upper
part of each panel shows the fluorescence when a narrow strip of
membrane in the xy plane is illuminated every 3 s. The
height of each peak corresponds to the fluorescence intensity half way
through the membrane in the z plane. Successive peaks, from
top to bottom, show the fluorescence at the end of each laser pulse.
The first thing to notice is that there is a clear difference between
the fluorescence along a row of peaks, corresponding to pores, and the
background fluorescence across the rest of the membrane, due to white
noise. The second point is that fluorescence does not decrease with
time (i.e., from top to bottom), indicating the absence of
photobleaching. The third point is that the height of successive peaks
is random. Finally, adjacent rows of peaks, in other words, adjacent
pores, do not fluctuate in concert. The correlations between 46 different pores stained with diO-C2-(3) under a
variety of conditions is presented in Table
1. Of the 58 correlations, three showed
significance at the 5% level (correlation coefficient > 0.36 or <
0.36 for 30 observations) and none showed significance at
the 1% level (correlation coefficient > 0.46 or <
0.46
for 30 observations). Taken together, these results make it unlikely
that fluctuations arise because of some artifact of the system. When
the fluctuations of a single pore are displayed as a time sequence
(lower part of each panel, from left to right), the similarity with
time sequences of current fluctuations (e.g., Lev et al., 1993
; Sachs
and Qin, 1993
; Korchev et al., 1997
) is apparent.


View larger version (195K):
[in this window]
[in a new window]
|
FIGURE 5
Fluctuations in DiO-C2-(3) or
fluorescein-stained PET membrane pores. PET membranes containing
1.1 × 106 pores/cm2 with diameters of
~20 nm were bathed with 10 6 M diO-C2-(3) in
the following solutions: 0.1 M KH2PO4, pH 4 (a); 0.1 M K2HPO4, pH 8 (b); or 0.1 M K3PO4, pH 11 (c). The same membranes were bathed with 5 µg/ml
fluorescein in 0.1 M KH2PO4, pH 4 (d). Images were obtained with a Zeiss LSM 410 invert
confocal microscope using a ×63/1.4(NA) oil-immersion objective;
fluorescence above 510 nm was excited at 488 nm. A box 400 pixels in
the x-direction and 8 pixels in the
y-direction was scanned 50 times at 3-s intervals. The
images are presented as a stack with the first time point at the top of
the page. The lower trace shows a time line at the
x-pixel indicated by the white arrow at the top of the
page; in c this image corresponds to a region where
there is no pore. Both the upper image and the lower trace
corresponding to one of the pores are shown as white signal
superimposed on arbitrarily chosen blue background.
|
|
A similar series of laser pulses, using fluorescein instead of
diO-C2-(3), is shown in Fig. 5 d. As
might be anticipated from the result of Fig. 4, f and
g, only a few pores can be distinguished above the
background noise, which indicates that most pores exclude the
negatively charged dye. What pores are visible fluctuate in intensity,
like the pores depicted in Fig. 5, a-c.
 |
DISCUSSION |
We interpret fluctuations of fluorescence to be a consequence of
protonation-deprotonation of carboxyl groups within the pore; it is to
the ionized form that cations are attracted. With
diO-C2-(3), increase of fluorescence reflects
entry of dye due to increase of ionization (deprotonation); with
fluorescein, increase of fluorescence also reflects entry of dye, in
this instance due to a decrease of ionization (protonation). We
conclude that fluctuations in the fluorescence of
diO-C2-(3) and of fluorescein in PET membrane pores reflect the same ionic changes that give rise to fluctuations of
cation current and anion current when a voltage is applied across a PET
membrane pore.
We believe that the fluctuations in fluorescence intensity we observe
in PET pores arise from time-varying fluctuations in surface charge
within the pore for the following reasons.
First, there are many more surface-charge-attracted dye molecules than
there are free dye molecules in the pore lumen. Under the conditions of
these experiments a 10-µm-long pore with a radius of 3 nm (as
determined by permeation of water; Rostovtseva et al., 1996
) will
contain, on average, two free diO-C2-(3)
molecules (assuming that the dye concentration in the pore is
10
5 M, as it is in solution), and 3 × 105 negative surface charges assuming that there
are 1.7 charges/nm2 (Wolf et al., 1995
). We note
that free solution between the membrane and coverslip has no detectable
fluorescence and conclude that the fluorescence signal we observe
arises from dye molecules attracted by negative surface charges. If the
confocal microscope samples 10% of the pore length (i.e., a 1-µm
slice), then that slice will contain 3 × 104 surface charges. If the dye occupies 3% or
more of those charged sites, then there will be at least 1000 dye
molecules to be detected; with a fluorescence lifetime of 1-10 ns and
a measurement time of about a second, this equates to
1011-1012 photons.
Second, diffusion is fast enough for dye molecules to enter and leave
the pores on the time scale of our fluorescence measurements, ~3 s
between each line scan. If we combine Ohm's law and Fick's law to
estimate flux across small pores as employed by Hille (1992)
and assume
that cyanine is present at 10
5 M and has a
diffusion coefficient of ~10
5
cm2/s, then ~100,000 dye molecules/s will
arrive at the mouth of a pore of radius 3 nm. If the pore is 10 µm
long, the rate of diffusion of a 10
5 M dye
solution through it is ~2000 molecules/s.
Third, the fluctuations represent a real signal, not noise due to the
size of the pore or the nature of the fluorescence detection system. A
priori the fluctuations we observe may represent variability of the
sample or variability of the detection system, or both. We note that
the day-to-day variability of fluorescence intensity of standard
objects (e.g., fluorescent beads) is less than 5% and often less than
1%. We have analyzed the statistical variation of fluorescence
intensity in the 46 pores that we investigated to see whether the
changes of intensity fluctuate together. Because fluorescence intensity
is proportional to the number of dye molecules in the image slice
selected by the confocal microscope we might expect that time
variations in intensity would follow a Poisson distribution. To compare
the distributions of all the pores it is necessary to transform each of
them into a form where this can be done simply. For Poisson
distributions this can be achieved by taking the square root of the
original values for each pore, finding the mean and standard deviation
for each pore and then subtracting the mean from each data point and
dividing by the standard deviation. The transformed data for each pore
have a mean of zero and a standard deviation of one. The distribution of all the data points, thus transformed (Fig.
6) is very close to a normal
distribution. This is consistent with data for each pore following a
Poisson distribution multiplied by a constant (the efficiency of
sampling in each case). We have also calculated the coefficient of
variation (mean/standard deviation) for the raw data for each of the 46 pores. If we plot coefficient of variation against mean fluorescence
intensity (Fig. 7) we find two classes of
pore (presumably as a result of the tracking and etching processes). The majority (40/46) have a coefficient of variation of 25%
irrespective of the mean intensity. The remainder (6/46) have low mean
intensity but very high coefficients of variation, greater than 100%.
These pores (13%) probably correspond to those that by
electrophysiological techniques show substantial fluctuations of ion
current; in our experience, the fraction of pores on any one day that
do this is ~10%. Since the two populations of pores come from
different pieces of PET, we assume that the high coefficient of
variation arises because of the nature of particular pores and not
because of the limitations of the detection system.

View larger version (34K):
[in this window]
[in a new window]
|
FIGURE 6
Distribution of fluorescence intensities among
different PET pores exposed to DiO-C2-(3). If we have many
observations from a Poisson variable, here 30 successive determinations
of fluorescence intensity in 46 different PET pores, their square root
should follow a normal distribution. This will also be true if the
Poisson variable is multiplied by a constant. First we took the square
root of the observations of fluorescence intensity. Then for each pore
separately we found the mean and standard deviation of the square root
transformed data. We then subtracted the mean and divided by the
standard deviation. This gives us a new variable (standardized
fluorescence intensity for each pore) with mean of 0 and standard
deviation of 1. We now combine the series into a single data set that
should follow a normal distribution if the distribution of the original
observations are a multiple of the Poisson distribution. The combined
data are shown as a histogram, and the line represents the normal
distribution.
|
|

View larger version (13K):
[in this window]
[in a new window]
|
FIGURE 7
Variability of fluorescence intensity of individual PET
pores exposed to DiO-C2-(3). We calculated the coefficient
of variation (mean/standard deviation) for 30 successive determinations
of fluorescence intensity (see, for example, Fig. 5,
a-c) of each of 46 different pores. Data
are presented as coefficient of variation versus mean fluorescence
intensity.
|
|
It could be argued that fluctuations of fluorescence of compounds like
diO-C2-(3) and fluorescein, which are partly
hydrophobic, are a consequence of changes in hydrophobicity of the
membrane surface. Because protonated carboxyl groups are more
hydrophobic than ionized carboxyl groups, the consequence of this
interpretation is the same as that given above, and we would not
quarrel with such a view. It could, furthermore, be maintained that
fluctuations of dye binding, as well as of ionic current, are due not
so much due to changes in ionization as to changes in architecture of the membrane between an open and a closed configuration. We have argued
elsewhere (Korchev et al., 1997
; Rostovtseva et al., 1996
) against such
an interpretation. Instead, we propose that our experiments show, for
the first time to the best of our knowledge, that the ionization of
fixed charges in a confined space, such as a narrow pore or the space
between two closely apposed surfaces, oscillates at a time scale many
orders of magnitude slower than that anticipated from the on and off
rate of protons in free solution. The reasons for such an enormous
decrease are a consequence of the ionic interactions between fixed
charges that are separated by no more than a few angstroms (Manning,
1969
; Korchev et al., 1997
).
Our model of a fluctuating surface charge is largely dismissed by Hille
(1992)
for biological ion channels on the grounds that in the one
instance where permeation of uncharged molecules has been measured, the
changes that affect ion current also affect permeability. In PET pores
the opposite situation holds, namely, that permeability of uncharged
molecules is unaffected by surface charge (Rostovtseva et al., 1996
).
The model is therefore unlikely to provide the major explanation for
the fluctuation of ion conductance in endogenous ion channels. But in
those instances where fixed charges may contribute to an anomalously
high conductance despite high selectivity, as in
Ca2+-activated K+ maxi
channels (Laver and Gage, 1997
), it is possible that fluctuations of
surface charge underlie the fluctuations of ion current that are
observed. Moreover, the oscillations in ion channels described as
open-channel noise (Sigworth, 1985
; Heinemann and Sigworth, 1991
),
which also occur in toxin-induced channels (Bezrukov and Kasianowicz,
1993
), may be partly due to fluctuations of surface charge. Although
they occur much more rapidly than the fluctuations in PET pores, the
difference in timescale is approximately of the same order as the
difference in length, around 1000-fold, and may reflect the timescale
of conformational/configurational changes of very differing space
dimension. In those toxin-induced pores in which the low-conductance
state has been shown to be still permeable to organic solutes (Korchev
et al., 1995
), fluctuations of surface charge do appear to account for
the observed oscillations in ion current (Korchev et al., 1997
). This
interpretation is likely to apply especially to pores that show
oscillations of conductance yet whose effective diameter is 10 nm or
greater; pores induced by toxins such as Clostridium
perfringens
toxin (Menestrina et al., 1990
) and immune
molecules such as activated complement (Bhakdi and Tranum-Jensen, 1984
)
or the cytolysin from cytotoxic T cells (Henkart, 1985
) are examples.
Finally, as nanopore technology (Bashford, 1999
; Rao et al., 1999
)
becomes applied to the fabrication of ion channel devices, the behavior
of fixed charges in a confined space that we have described will need
to be taken into account. Indeed it may lead to quite novel
applications in situations where normal transmission of nerve impulses fails.
We are grateful to Dr. Reimar Spohr and Dr. P. Y. Apel for the
gift of track-etched membranes and to Drs. Martin Bland, Donald Edmonds, Apolinario Nazarea, and Terry Poulton for much helpful advice.
This work was supported by the Biotechnology and Biological Sciences
Research Council and the Cell Surface Research Fund.
Address reprint requests to Dr. C. L. Bashford, Department of
Biochemistry and Immunology, Cellular and Molecular Sciences Group, St
George's Hospital Medical School, London SW17 0RE, UK. Tel.:
44-020-8725-5770; Fax: 44-020-8725-2992; E-mail:
l.bashford{at}sghms.ac.uk.