Biophys J, July 2002, p. 491-501, Vol. 83, No. 1
and
Departments of *Biological Sciences,
Physics, and
Applied Physics, Stanford University, Stanford,
California 94305-5020 USA
| |
ABSTRACT |
|---|
|
|
|---|
We constructed a next-generation optical trapping instrument to study the motility of single motor proteins, such as kinesin moving along a microtubule. The instrument can be operated as a two-dimensional force clamp, applying loads of fixed magnitude and direction to motor-coated microscopic beads moving in vitro. Flexibility and automation in experimental design are achieved by computer control of both the trap position, via acousto-optic deflectors, and the sample position, using a three-dimensional piezo stage. Each measurement is preceded by an initialization sequence, which includes adjustment of bead height relative to the coverslip using a variant of optical force microscopy (to ±4 nm), a two-dimensional raster scan to calibrate position detector response, and adjustment of bead lateral position relative to the microtubule substrate (to ±3 nm). During motor-driven movement, both the trap and stage are moved dynamically to apply constant force while keeping the trapped bead within the calibrated range of the detector. We present details of force clamp operation and preliminary data showing kinesin motor movement subject to diagonal and forward loads.
| |
INTRODUCTION |
|---|
|
|
|---|
Optical trapping has been used extensively to
study the motion of individual motor proteins such as kinesin, myosin,
and RNA polymerase (Svoboda et al., 1993
; Finer et al., 1994
; Molloy et al., 1995
; Wang et al., 1998
; Warshaw et al., 2000
). In earliest use,
fixed traps simply provided a way to grab and hold motor-coated beads
while the motor substrate was moved with a manually operated microscope
stage. Bead motion was recorded at comparatively low spatiotemporal
resolution by video (Block et al., 1990
). Over time, these instruments
have become more sophisticated and versatile. Sensitive position
detectors, based on interferometry or quadrant photodiodes (QPDs), have
been added to track bead motion with subnanometer accuracy and high
bandwidth. The trapping beam can be steered by moving external lenses
or mirrors, either manually or through motorized control. Faster and
more precise steering is achieved with computer-controlled galvanometer
mirrors or acousto-optic deflectors (AODs) (Svoboda and Block, 1994a
).
AODs are especially versatile because they can also modulate the laser
intensity and generate multiple beams (Visscher et al., 1996
; Molloy,
1998
). The advantage of computerized trap-modulating and trap-steering devices is that they allow experiments to be automated with greater speed and precision. For example, rapid, on-the-fly calibration for
every bead is facilitated by the addition of an independent position
sensing system, such as a detection laser that is distinct from the
trap laser (Visscher et al., 1996
). Computer-controlled traps can also
be programmed to act as position clamps or force clamps using feedback
control. Position clamping, in which either the trap light intensity or
location is modulated to keep the bead at a fixed position, is useful
for studying motor stall forces (Finer et al., 1994
; Wang et al.,
1998
). Force clamping in which either the trap intensity or location is
varied to apply a constant load provides exceptionally clear records of
the motion of processive motors. Comparatively long records with
reduced Brownian motion can be collected because high loads can be
applied continuously. In addition, force clamping simplifies the
interpretation of results because it eliminates the need for
corrections arising from the series elastic compliance between the trap
and motor (Visscher and Block, 1998
; Rief et al., 2000
).
Biophysical insights provided by one-dimensional micromechanical
studies of single molecules may be expanded by extension into further
dimensions. For example, the reduced Brownian motion provided by a
two-dimensional (2D) force clamp may allow direct observation of the
stepping path of kinesin along its microtubule substrate. More
generally, multidimensional mechanical measurements supply additional
knowledge about how force is transmitted through proteins to modulate
their biochemistry. All motor proteins cycle through mechanical and
biochemical transitions that occur in three dimensions, such as
attachment-detachment from the substrate, nucleotide binding-unbinding,
chemical activation, or internal conformational changes. Because these
transitions may occur at multiple sites in the protein and along
various reaction coordinates, their rates may be differentially
affected by the direction of applied loads. Studying how these rates
depend on loading direction reveals information about the molecular
motions and chemical transformations underlying them, ultimately
providing rigorous tests for models of motor function. For certain
motility systems, such as kinesin and myosin, atomic structures have
been solved (e.g., Rayment et al., 1993b
; Sack et al., 1997
) and the
binding orientations between motor and substrate are known (e.g.,
Rayment et al., 1993a
; Rice et al., 1999
; Sosa et al., 2001
). The
filamentous substrate in these systems can be directly observed in the
microscope so that forces can be applied at a well-defined angle with
respect to the atomic structure of the motor-substrate complex. Thus
models of motor function that predict specific atomic motions are
testable with a 2D force clamp. Multidimensional force measurements may also prove useful in the study of other, nonmotor enzyme-substrate interactions, including receptor-ligand binding.
Motivated by these considerations, we built an optical trapping instrument with 2D feedback control capable of applying loads of variable magnitude and azimuthal direction. A key feature of the new instrument is that it incorporates a precision piezoelectric stage with capacitive position sensing. This addition facilitates accurate placement of each bead before a measurement and enables automated stage movements to keep the bead within the calibrated range of the position sensor once force is applied. A combination of piezo stage, AODs, and an independent detection laser provides great flexibility in experimental design. Further enhancements for stability and minimization of drift include 1) sound/vibration isolation and temperature control in the experiment room, and 2) fortification of the microscope body and optical layout. Moreover, the design does not interfere with the fluorescence capabilities of the microscope.
Here we include details of instrument design, testing, calibration, and performance. Automated experimental procedures are described for 2D calibration and bead positioning before each measurement and for force clamping in two dimensions. Records showing the motion of individual kinesin motor molecules subjected to diagonal loads (including force components both parallel and perpendicular to the direction of travel) and forward loads (parallel to the direction of travel) are presented.
INSTRUMENT DESIGN
The instrument is based on an inverted microscope (model Eclipse
TE200; Nikon Instruments Inc., Melville, NY) and incorporates three
lasers for trapping, position detection, and fluorescence excitation,
as shown in Fig. 1
A. The optical trapping and detection components are similar
to those described in Visscher and Block (1998)
, including AODs (model
DTD-274HA6; IntraAction Corp., Bellwood, IL) for computer steering of
the trap laser beam (1064 nm, Nd:YVO4, model BL-106C; Spectra-Physics
Lasers, Mountain View, CA), and a separate detection laser (828 nm,
model LDS-P-3-830-4-15-TE-P1501; Point Source, Hampshire, UK). Trap and
detector wavelengths were selected to minimize photodamage (Neuman et
al., 1999
) and to avoid standard fluorescence emission wavelengths. The
trapping laser delivers sufficient power to apply forces up to 100 pN
to a 0.52-µm diameter silica bead. In principle, the AODs are capable of very rapid beam steering (rise time of ~4 µs, limited only by
the ratio of the laser beam diameter to the speed of sound in the
crystal). But in our system, the update rate is limited to 8 kHz by the
LabVIEW 6i software (National Instruments Corp., Sunnyvale, CA) used to
change the AOD drive frequency. Although not implemented here, higher
rates (100 kHz or more) are achievable using a low-level programming
language (e.g., C++) to address the PCI-bus more
quickly (Visscher and Block, 1998
).
|
A key feature of the new instrument is the incorporation of a
three-axis piezo stage with capacitive position sensing (model P-517.3CD; Polytec PI, Tustin, CA). With this device, the position of
the specimen is digitally controlled in increments as small as 1 nm
over a 100 × 100 × 20 µm volume. Stage calibration data provided by the manufacturer (performed against a NIST-traceable standard) confirm subnanometer resolution and repeatability. Joystick control of the stage (implemented in our laboratory using LabVIEW software) allows the instrument to be operated remotely, which minimizes induced vibrations and drift. The 20-µm range along the
z axis (see axes in Fig. 1) is sufficient that all fine
focusing can be done using stage movement, although coarse focusing
requires adjustment of the objective height using the microscope knob. The time response of the stage includes both mechanical settling with
an exponential time constant of ~10 ms and an additional ~10-ms
delay due to communication through the GPIB interface. For
certain applications, this speed is sufficient to implement force or
position feedback control using the stage alone (Perkins et al., 2001
).
The microscope body was extensively modified for mechanical stability
and to accommodate the three lasers. The objective turret was replaced
with a custom mount attached to the focusing unit of the microscope
(Fig. 1 B), which improves upon an earlier design (Visscher
et al., 1996
) because it creates access directly above the fluorescence
filter cube for inserting the trap and detector beams. These beams
bypass the filter cube, so any fluorescence filter set can be
introduced without affecting the intensity, alignment, polarization, or
general quality of the beams. The injection mirror (D1, Fig.
1) transmits visible wavelengths between 450 and 750 nm while
efficiently reflecting the infrared light. The system thus allows
introduction of a third laser beam for TIR excitation and low
background single molecule fluorescence detection (Fig. 1). The custom
mount holds the lower Wollaston prism so that Nomarski DIC capability,
used here for video-enhanced observation of individual microtubules, is
preserved. The specimen stage was replaced with an aluminum platform to
support both a crossed roller-bearing stage for coarse positioning by
hand (model 750-MS; Rolyn Optics, Covina, CA) and the piezo stage for
fine positioning. The detection optics, including a dichroic mirror (D3, Fig. 1 A), one or more interference or color
filters (F2, Fig. 1 A), a lens (L7,
Fig. 1 A), and the quadrant photodiode (QPD,
model SPOT9-DM1, UDT, Hawthorne, CA), are mounted on a lightweight, rigid, cantilevered platform that extends horizontally from the condenser housing. The condenser assembly and detection system are
mounted on a heavy-duty fine focusing stage (Modular Focusing Unit;
Nikon Instruments Inc.), affixed to the illumination pillar by a thick
aluminum plate. An elevated breadboard next to the microscope allows
optics to be mounted on shorter posts for enhanced vibrational
stability. A plexiglass enclosure surrounds the optics to suppress
convection currents in air.
Signals from the four quadrants of the QPD are preamplified and passed
through a differential amplifier that supplies normalized x-
and y-position signals and a third signal representing the average intensity on all four quadrants of the QPD. The average signal
can be used to monitor vertical or z deflections of a bead (Pralle et al., 1999
; Neuman et al., 2001
). Because it is not normalized, it is more susceptible than the x and
y signals to noise arising from intensity fluctuations in
the detector laser. Position signals are antialias filtered to the
appropriate Nyquist frequency (i.e., one-half of the data collection
rate, using model SR640; Stanford Research Systems, Sunnyvale, CA)
before digitization by a 16-bit A/D board (model PCI 6052E; National
Instruments Corp., Austin, TX). Custom software was developed using
LabVIEW 6i for data acquisition and instrument control. Offline
analysis software was written using Igor Pro 4.01 (Wavemetrics, Inc.,
Lake Oswego, OR). The measured bandwidth of our detection system
overall, including the QPD, amplifying electronics, and the A/D board,
is 30 kHz (
3 dB). This value was measured by using the fast beam
steering capabilities of the AODs, as follows. Turning off the
detection laser and removing filters F2 in Fig. 1
A allow the trapping laser to impinge on the QPD. The
bandwidth was then determined by varying the laser beam position
sinusoidally over a stuck bead with the AODs while simultaneously
monitoring the normalized x- and y-deflection signals produced by the trapping beam, similar to Veigel et al. (1998)
.
To minimize drift and reduce acoustic noise, the microscope is situated
on an optical air table, inside a temperature-controlled, sound-proofed
clean room (class ~150,000 or better). The room temperature is
regulated to ±0.15°C peak-to-peak using a single wall-mounted
sensor. Feedback control is used to adjust the temperature of a gentle,
continuous airflow through the room. Temperature fluctuations appear to
be the dominant source of long-term drift, and with temperature
regulation the drift of the system can be reduced to <5 Å/min, as
shown in Fig. 2 A. Sound
couples strongly into the microscope and is another potential source of
instrument noise. Acoustically loud devices, especially those with
cooling fans such as laser power supplies, high-voltage amplifiers, and computers, are placed outside of the room. Background noise in the room
is below the NC30 rating. ("Noise Criterion" (NC) quantitates noise
levels over a range of frequencies relative to a standard family of
reference spectra. The NC30 sound level is roughly equivalent to a
quiet office or concert hall (Beranek, 1960
).) With these precautions,
detector noise is <2 Å/
Hz for frequencies above 0.6 Hz (Fig. 2
B). In addition to providing a dust-free environment, the
clean room can be darkened for low-light level experiments.
|
Precise alignment of the trap and detection beams, which is critical for obtaining accurate position detection with the highest possible resolution, is achieved with computer-controlled scanning of both the trap and stage. We outline three procedures useful in the alignment of our system. First, the detector and trap beams are made parallel to the z axis using the following method. The input angle of each beam is adjusted to minimize the QPD deflection signal while a bead stuck to the coverslip is scanned back-and-forth in the z direction through the beam waist using the piezo stage. In this way, both beams are aligned so that bead motions purely along the z axis produce less than 1% cross-talk in the x or y channels. Second, the z position of the detection beam waist is centered on a trapped bead by translating lens L4 (Fig. 1 A) axially, with a micrometer, until a trapped bead scanned in the x-y plane with the AODs produces deflection signals with maximal sensitivity (V/nm, the slope of the signal with position). Centering the detector beam waist to within ~100 nm of the z position of a trapped bead insures that slight vertical bead motions, which occur during a motility assay (<40 nm in our current kinesin experiments), result in changes of <1% in x- and y-signal sensitivity. Note that for unidirectional vertical motions, the position of the detector waist can sometimes be adjusted to maximize the useable portion of the range over which the sensitivity remains relatively flat. For example, in an experiment where only downward motion is produced, the waist might be positioned slightly below a trapped bead to accommodate this deflection with minimal sensitivity change. Third, lens L4 is also translated in x and y to align the detector beam with the center of the steerable range of the trap. The QPD is then translated, in a plane conjugate to the back aperture of the condenser, until the x and y signals are nulled.
AUTOMATIC 2D CALIBRATION AND BEAD POSITIONING
Positions of both the optical trap and the specimen stage are under precise computer control, allowing a number of fairly complex procedures to be performed automatically before each measurement. The 2D position calibration and subsequent trap stiffness determination rely on AOD steering of the trap, whereas the piezo stage is used for positioning of the bead before measurement. These procedures increase the level of control and reproducibility of the experiments.
2D position calibration
Accurate calibration of the QPD over the full detector range for
each individual bead is required for 2D force clamping experiments. Calibrating each bead avoids errors caused by bead size variations and
instrument drift. The position of a trapped bead is registered by
monitoring deflections of the detection beam caused by bead motion, as
described previously (Visscher et al., 1996
; Gittes and Schmidt, 1998
;
Visscher and Block, 1998
). One-time video tracking of a trapped bead in
two dimensions, calibrated against known stage motion, is performed to
verify the operation of the AODs and to obtain the conversion
parameters from AOD drive frequency (in MHz) to position (in nm). With
this calibration, the position response of the QPD for any trapped
object can be quickly mapped by using the AODs to raster-scan the
object over a matrix of known positions. Fig.
3, V1
and V2, show the response to a
0.52-µm diameter silica bead scanned over an 820 × 820-nm
square area. A 2D, fifth-order polynomial fit to the nonlinear response
is used to map QPD output voltages to x-y spatial
coordinates. The QPD can be used as a calibrated position sensor only
over the region where position as a function of voltage is single
valued, denoted by the dotted circles in Fig. 3,
V1 and
V2. Within this calibrated region
(~360-nm diameter), the residual error of the fit is <2 nm root mean
square, as shown in Fig. 3, Ex
and Ey. Note that the input diameter
of the detection beam (4 mm) is chosen as a compromise between the resulting sensitivity of the position detector (~40 mV/nm) and the
size of the calibrated range. A smaller laser beam input diameter generates a larger spot size in the specimen plane (Siegman, 1986
), which results in a larger range but reduced sensitivity (Allersma et
al., 1998
). We chose our detector beam spot size to achieve sufficient
positional sensitivity to observe small stepping events (as small as 1 nm) and sufficient detector range to measure processive events as long
as 360 nm. Similar methods, using the piezo stage to raster-scan a
stuck bead, can also be used for 2D or three-dimensional calibration
(Pralle et al., 1999
; Kuo and McGrath, 2000
; Neuman et al., 2001
).
|
2D stiffness calibration
For stiffness calibration in two dimensions, we use three
different methods: drag force, equipartition, and power spectrum (Visscher et al., 1996
; Visscher and Block, 1998
). Before these methods
can be applied, however, the shape of the 2D trapping potential must be
carefully mapped. Even with well-aligned beams, polarization-dependent
effects can cause the trap to be less stiff in one dimension than in
another, especially with overfilled, high-NA objectives (Wright et al.,
1994
; Rohrbach and Stelzer, 2001
). Added distortion may be introduced
by the Wollaston prism, which separates orthogonal polarization
components of the trap laser and creates two partially overlapping,
diffraction-limited spots in the specimen plane. Wollaston-induced
distortion is eliminated by orienting rotatable waveplate W1 (Fig. 1
A) so that the input polarization of the trapping beam is
aligned with the shear axis, and a single, diffraction-limited spot is
created. In our system, the resulting 2D trapping potential is
elliptical, with its principle axes aligned with the Wollaston axes, as
determined experimentally by the 2D position histogram for a trapped
bead. Measuring trap stiffnesses along the two principle axes allows
the vector restoring force in the plane of the coverslip to be computed
for any position in the trap. The trapping potential in our system is
harmonic (to better than 3%) for displacements out to 100 nm from the
trap center, and the two principle stiffnesses typically differ by 30%. Distortions in the detection laser beam are unimportant, because
they are compensated by the 2D position calibration procedure described earlier.
Precise bead placement relative to the substrate
We use two automated techniques, one of which is a variant of
optical force microscopy (Ghislain and Webb, 1993
; Florin et al.,
1997
), to increase accuracy and reproducibility in the placement of a
motor-coated bead near the motor substrate. As a vertical reference, we
find the position of the stage where the top surface of the coverslip
just contacts the trapped bead (Neuman et al., 2001
; Perkins et al.,
2001
). This point is determined by raising the coverslip through the
point of contact while continuously monitoring the average intensity at
the QPD (the z signal). At the point of contact, where the
moving coverslip begins to displace the bead from its trapped position,
the slope of the intensity signal changes abruptly (Fig.
4, A and B). This
inflection point (found by determining the intersection of lines fit to
the left and right of the contact point) provides a precise (±4 nm
RMS) fiducial reference for subsequent vertical positioning.
Alternatively, periodic artifacts caused by multiple reflections of the
detector beam between the bead and coverslip are present in the signal (Neuman et al., 2001
) and may also be used to determine a vertical reference point.
|
The lateral position of a motor substrate, such as a microtubule
stuck to the coverslip, can be found by scanning the object through the
detection beam. Before scanning, the AODs are used to steer the optical
trap (and any trapped bead) out of the way. The stage is then used to
move the microtubule through the detection beam (Fig.
5 inset). Like a bead, the
microtubule scatters the detection laser light, producing an
x-deflection signal that is similar in shape, but ~100×
weaker in magnitude. Fig. 5 shows the x-position signal for
a microtubule scanned perpendicular to its long axis. Fitting the scan
data to the function f(x) = A + B(x
x0)e(x
x0)2/
2
(Allersma et al., 1998
) using the Levenberg-Marquardt algorithm provides the center position, x0, for
the scanned object, to within ±3 nm RMS. A similar method using
optical force microscopy, where the trapped bead is used as a stylus,
may also be used for lateral positioning (Ghislain and Webb, 1993
;
Florin et al., 1997
).
|
APPLYING A 2D FORCE CLAMP TO KINESIN
Force clamp operation is preceded by an automated initialization sequence that is run on every bead, consisting of the 2D calibration and bead positioning procedures described above (Fig. 6). First, the surface height is found, so that the 2D position calibration and the remainder of the experiment can be performed at the same height above the coverslip. At the end of the sequence, which takes less than a minute, the calibration parameters are stored to disk, and the bead is positioned directly over the microtubule at the outer edge of the detection zone (Fig. 7 A). The computer triggers force-feedback operation when movement into the calibrated region is detected (Fig. 7 B). Because of the combined length of the motor protein-to-bead linkage and the bead radius (~350 nm), application of a sideways or forward load can easily move the tethered bead outside of the calibrated region. In our system, this difficulty is overcome by a compensatory stage motion to bring the bead back inside the detection zone, as depicted in Fig. 7 C. During subsequent clamping, the trap is steered in two dimensions with the AODs, to maintain the desired vector force (Fig. 7 D). Small bead-to-trap center separations can introduce force errors associated with small offsets or drifts in the position detection and Brownian motion of the bead. To minimize these errors, the trap stiffness is chosen so that a bead-to-trap separation of ~100 nm gives the desired force. Inaccurate force clamping can also result if the clamp update rate is insufficient to follow the bead motion. When applying purely longitudinal forces (parallel to the microtubule long axis), the clamp can be operated in a one-dimensional mode by confining the trap to move along the microtubule (see Fig. 9). When the bead either detaches from the microtubule or moves outside of the calibrated region, the feedback is terminated and the trap and stage are returned to their starting positions for the next run. Automatic repetition of this procedure by the computer allows for multiple runs from the same bead to be recorded with minimal intervention by the operator.
|
|
The accuracy of the position detection system and the level of force
clamping are best determined in control experiments that match the
assay conditions as closely as possible. As discussed, changes in
sensitivity are minimized by superposing the focal waist of the
detector beam with the center of the optical trap and by calibrating
the bead immediately above the coverslip surface using the vertical
fiducial reference. Even with these precautions, tiny vertical bead
motions or instrumental drift can introduce errors in position mapping.
Therefore, we performed control experiments with kinesin-labeled beads
tethered to microtubules in rigor using the nonhydrolyzable ATP analog,
AMP-PNP (Visscher and Block, 1998
). Tethered beads were moved using the
stage at predetermined velocities while operating the force clamp. The
resulting records confirmed that force is clamped accurately (within
±5% RMS) and that correct velocities are returned by the system (data
not shown).
To demonstrate the capabilities of the 2D force clamp, we performed
single molecule kinesin bead motility assays using native squid
kinesin, as described previously (Schnitzer and Block, 1997
). Figs.
8 and 9
show examples of typical kinesin-driven events for beads moving under
diagonal and forward loads, respectively, obtained with the instrument.
The position data of Fig. 8 were taken at a fixed load of 5.3 pN at an
angle of 45° to the microtubule long axis under saturating ATP
conditions (1.6 mM). A line fit to the clamped interval in Fig. 8
A yields an average velocity of 454 nm/s. Note that
individual 8-nm steps in this trace are obscured by Brownian noise, due
to a combination of the low stiffness of the kinesin-bead-trap linkage
along the direction of travel and the fast motor speed. The data of
Fig. 9 were taken under a constant forward load of 5 pN at low ATP
levels (4.2 µM). In this case, the stiffness along the direction of
travel is higher, whereas the average velocity is slower (25 nm/s), and
individual 8-nm steps can readily be seen. For detailed discussions of
the conditions under which kinesin steps can be observed, see Svoboda
et al. (1993)
, Svoboda and Block (1994b)
, Block and Svoboda (1995)
, and Visscher and Block (1998)
.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
Here we describe a new optical-trapping system designed to apply
fixed 2D loads to individual processive motor proteins. The instrument
expands and improves upon the capabilities of previous optical-trapping
force clamps (Visscher and Block, 1998
), originally applied to kinesin
(Visscher et al., 1999
) and later to myosin-V (Rief et al., 2000
),
which were implemented in one dimension with all loads being applied in
a direction opposing the motion. Specific new features include: 1) an
optical layout in which the detection and trapping laser beams do not
pass through the fluorescence filter cube, 2) a detection wavelength
that does not overlap with standard fluorescence colors, 3) enhanced
stability due to sound-proofing, temperature control, and mechanical
reinforcement, and 4) novel techniques for 2D calibration and
positioning of the bead relative to the substrate. A key improvement to
the instrument is the incorporation of a three-axis piezo stage capable
of subnanometer accuracy and precision. This, in combination with 2D
trap steering and an independent 2D position sensing system, creates a
highly adaptable instrument that can be programmed to perform many
aspects of a nanomechanical experiment automatically.
Combining automatic bead placement with on-the-fly position calibration maximizes repeatability, in addition to avoiding systematic errors that can occur with manual bead placement. For example, we use computer control of the specimen stage to position kinesin-coated beads relative to a microtubule with ±4 nm precision (RMS) in the vertical direction, and ±3 nm (RMS) in the lateral directions (Figs. 4 and 5, respectively). Trap steering is then used to calibrate the position sensor in two dimensions for each bead with better than 2 nm (RMS) accuracy (Fig. 3). Position calibration is performed at the same height as that for the remainder of the experiment, just sufficiently above the coverslip to allow motor binding to the microtubule.
One use of the automated capabilities of this instrument is the 2D force clamp. Our clamp relies on computer-controlled movements of the specimen stage to keep beads within the detection zone while constant loads are applied by steering the optical trap. The compensatory stage motions prevent the kinesin-tethered bead from swinging out of the detection zone when loads with forward or sideways components relative to the direction of motion are applied (Fig. 7). Force feedback, which can be operated with loop closure rates as high as 8 kHz, permits comparatively long records with reduced Brownian motion to be collected for detailed examination of molecular motion under fixed conditions. Force clamping also simplifies the interpretation of data by eliminating the need for corrections due to the compliance of the motor-bead linkage. Loads of up to 100 pN in any azimuthal direction within the plane of the coverslip can be maintained with high accuracy over the full 360-nm-diameter detection zone. Automatic software triggering initiates the stage moves, and starts and stops force clamping and data collection (Fig. 6). Many runs from each bead can be collected with little input from the user, quickly creating a representative ensemble from single molecule measurements. This ensemble can be used to determine bulk properties and to characterize protein variance/inhomogeneity.
Noise and drift in the position detection system are minimized by
buttressing the microscope body and optical layout and by housing the
entire instrument in a sound-proofed, temperature-controlled experimental room. Detector noise for frequencies above 0.6 Hz is <2
Å/
Hz, and long-term drift is <5 Å/min (Fig. 2). The overall detection bandwidth of 30 kHz corresponds to a rise time of ~12 µs.
This response is fast enough for measuring biochemical kinetics. The
optical layout preserves the fluorescence capabilities of the
microscope, because both trapping and detection beams do not pass
through the fluorescence filter cube (D1, Fig. 1).
The capabilities of the instrument are demonstrated in Figs. 8 and 9, which show records of kinesin-driven bead motion under the application of fixed forward and diagonal loads. To our knowledge, these are the first records of individual processive motors subject to constant forces that are not purely opposing the motion. The instrument is presently being used to study how the direction of loading affects kinesin motility by measuring ensemble-averaged velocities for single kinesins subject to various loading directions, forces, and ATP concentrations. Applying diagonal forces to moving kinesin molecules may also reduce the Brownian motion enough to allow direct observation of the 2D stepping path taken by the motor along its microtubule substrate.
Our instrument was designed with kinesin motility assays in mind, but
its capabilities are sufficiently flexible to find applications in a
variety of nanomechanical experiments. Applications requiring a larger
range of position detection but less sensitivity can be accommodated,
for example, by decreasing the diameter of the detection laser beam.
The size of the calibrated zone can be readily doubled with a
proportional decrease in sensitivity. For slower, processive motors,
the range of force clamping can be greatly extended by using the piezo
stage as part of the feedback control. Stage-based clamping is possible
over the full 100 × 100-µm range with loop closure rates of
50 Hz. Beyond clamping applications, this instrument could be used to
study enzyme-substrate or receptor-ligand interactions under various
loading directions.
Transitions in the cycle of a motor protein or enzyme take place in a
three-dimensional world. In principal, even the cycles of processive
motors that generate unidirectional motion, such as kinesin or
myosin-V, include mechanochemical transitions along reaction
coordinates that are not aligned with the overall direction of motion.
Forces applied in various directions therefore differentially affect
the rates of these transitions, causing measurable changes in motor
velocity when the affected transitions become rate limiting. Determining how the direction of force affects motor velocity is useful
for constraining models for motor-protein function. Such models are
becoming increasingly sophisticated as atomic structures and binding
orientations are discovered, sometimes even incorporating structural or
conformational changes at the atomic level (e.g., Vale and Milligan,
2000
). Understanding how atomic motions are affected by externally
applied loads is not straightforward, given uncertainties about which
parts of the structure bear load during the cycle. But certain features
of the proposed conformational changes, such as their direction
relative to the overall motion, provide testable predictions about how the motor speed changes with loading direction. The 2D capabilities of
our new instrument will facilitate studies of single molecule motility
from a new angle.
| |
ACKNOWLEDGMENTS |
|---|
The authors thank Koeu Visscher, Mark Schnitzer, and the other members of the Block Lab for helpful discussions.
M.J.L. is supported by the Jane Coffin Childs Memorial Fund for Medical Research. C.L.A. is supported by the Cancer Research Fund of the Damon Runyon-Walter Winchell Foundation Fellowship, DRG-1649. S.M.B. acknowledges the support of a grant from the National Institute of General Medical Sciences (the National Institutes of Health).
| |
FOOTNOTES |
|---|
Address reprint requests to Steven M. Block, Gilbert Bldg., Rm 109, Mail Code 5020, Stanford, CA 94305-5020. Tel.: 650-724-4046; Fax: 650-723-6132; E-mail: sblock{at}stanford.edu.
Submitted September 17, 2001, and accepted for publication March 20, 2002.
M. J. Lang, C. L. Asbury, and J. W. Shaevitz contributed equally to this work.
| |
REFERENCES |
|---|
|
|
|---|
Biophys J, July 2002, p. 491-501, Vol. 83, No. 1
© 2002 by the Biophysical Society 0006-3495/02/07/491/11 $2.00
This article has been cited by other articles:
![]() |
R. E. L. DeVille and E. Vanden-Eijnden Regular Gaits and Optimal Velocities for Motor Proteins Biophys. J., September 15, 2008; 95(6): 2681 - 2691. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Ferrer, H. Lee, J. Chen, B. Pelz, F. Nakamura, R. D. Kamm, and M. J. Lang Measuring molecular rupture forces between single actin filaments and actin-binding proteins PNAS, July 8, 2008; 105(27): 9221 - 9226. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. V. Desai, T. G. Bishop, L. Vicci, E. T. O'Brien Sr., R. M. Taylor 2nd, and R. Superfine Agnostic Particle Tracking for Three-Dimensional Motion of Cellular Granules and Membrane-Tethered Bead Dynamics Biophys. J., March 15, 2008; 94(6): 2374 - 2384. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Tsygankov and M. E. Fisher Mechanoenzymes under superstall and large assisting loads reveal structural features PNAS, December 4, 2007; 104(49): 19321 - 19326. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Pilizota, T. Bilyard, F. Bai, M. Futai, H. Hosokawa, and R. M. Berry A Programmable Optical Angle Clamp for Rotary Molecular Motors Biophys. J., July 1, 2007; 93(1): 264 - 275. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Knoner, B. E. Rolfe, J. H. Campbell, S. J. Parkin, N. R. Heckenberg, and H. Rubinsztein-Dunlop Mechanics of Cellular Adhesion to Artificial Artery Templates Biophys. J., October 15, 2006; 91(8): 3085 - 3096. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Brau, P. B. Tarsa, J. M. Ferrer, P. Lee, and M. J. Lang Interlaced Optical Force-Fluorescence Measurements for Single Molecule Biophysics Biophys. J., August 1, 2006; 91(3): 1069 - 1077. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Deufel and M. D. Wang Detection of Forces and Displacements along the Axial Direction in an Optical Trap Biophys. J., January 15, 2006; 90(2): 657 - 667. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T. X. Li, D. Collin, S. B. Smith, C. Bustamante, and I. Tinoco Jr. Probing the Mechanical Folding Kinetics of TAR RNA by Hopping, Force-Jump, and Force-Ramp Methods Biophys. J., January 1, 2006; 90(1): 250 - 260. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Fisher and Y. C. Kim Kinesin crouches to sprint but resists pushing PNAS, November 8, 2005; 102(45): 16209 - 16214. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Rosenfeld, J. Xing, G. M. Jefferson, and P. H. King Docking and Rolling, a Model of How the Mitotic Motor Eg5 Works J. Biol. Chem., October 21, 2005; 280(42): 35684 - 35695. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E.-M. Clemen, M. Vilfan, J. Jaud, J. Zhang, M. Barmann, and M. Rief Force-Dependent Stepping Kinetics of Myosin-V Biophys. J., June 1, 2005; 88(6): 4402 - 4410. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Nambiar, A. Gajraj, and J.-C. Meiners All-Optical Constant-Force Laser Tweezers Biophys. J., September 1, 2004; 87(3): 1972 - 1980. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. L. Asbury, A. N. Fehr, and S. M. Block Kinesin Moves by an Asymmetric Hand-Over-Hand Mechanism Science, December 19, 2003; 302(5653): 2130 - 2134. [Abstract] [Full Text] [PDF] |
||||
|
|