Departments of Chemistry and Chemical Engineering, Texas A&M
University, College Station, Texas 77843-3122 USA
To investigate fluorescence lifetime spectroscopy in
tissue-like scattering, measurements of phase modulation as a function of modulation frequency were made using two fluorescent dyes exhibiting single exponential decay kinetics in a 2% intralipid solution. To
experimentally simulate fluorescence multiexponential decay kinetics,
we varied the concentration ratios of the two dyes, 3,3-diethylthiatricarbocyanine iodide and indocynanine green (ICG), which exhibit distinctly different lifetimes of 1.33 and 0.57 ns,
respectively. The experimental results were then compared with values
predicted using the optical diffusion equation incorporating 1)
biexponential decay, 2) average of the biexponential decay, as well as
3) stretched exponential decay kinetic models to describe kinetics
owing to independent and quenched relaxation of the two dyes. Our
results show that while all kinetic models could describe phase-modulation data in nonscattering solution, when incorporated into
the diffusion equation, the kinetic parameters failed to likewise
predict phase-modulation data in scattering solutions. We attribute the
results to the insensitivity of phase-modulation measurements in
nonscattering solutions and the inaccuracy of the derived kinetic
parameters. Our results suggest the high sensitivity of
phase-modulation measurements in scattering solutions may provide greater opportunities for fluorescence lifetime spectroscopy.
 |
INTRODUCTION |
Fluorescence lifetime spectroscopy is especially
advantageous for quantitative biomedical spectroscopy of analytes since
the measurement of fluorescence decay kinetics (rather than the
fluorescence intensity) eliminates the necessity for the knowledge of
the analyte-sensing fluorophore concentration. Frequency domain
techniques provide measurement of fluorescence lifetime (
) using
simple relationships of the phase-delay (
) and amplitude-attenuation
(M) of the reemitted fluorescence as a function of the modulation
frequency relative to intensity modulated excitation light. However,
the development of fluorescence lifetime spectroscopy for near infrared
(NIR) biomedical tissue diagnostics for sensing using systematically administered dyes (Hawrysz and Sevick-Muraca, 2000
; Weissleder et al.,
1999
) or implantable devices (Qing et al., 1997
; Russell et al., 1999
)
requires 1) deconvolving the influence of multiple scatter upon the
measured emission phase-delay and amplitude-attenuation measured in the
NIR wavelength region and 2) accounting for nonsingle exponential decay kinetics.
Most fluorophores capable of analyte sensing exhibit multiexponential
kinetics. For example, Ca2+ (Lakowicz and Szmacinski, 1992
)
and pH (Lakowicz and Szmacinski, 1993
) sensing with fluorescence
lifetime spectroscopy in dilute, nonscattering solutions requires
determination of multiexponential decay kinetics for accurate analyte
sensing. Accurate fluorescence lifetime spectroscopy in the presence of
tissue-like scattering requires a model that accounts not only for
photon propagation but also for multiexponential decay kinetics as
well. Indeed, the fluorescence resulting from ultraviolet excitation of
the normal and athlerosclerotic arterial wall also shows
multiexponential decay kinetics of its excitable constitutes
(Andersson-Engels et al., 1991
). Because ultraviolet light does not
multiply scatter in tissues, deconvolution of the influence of scatter
is not necessary. In contrast, failure to properly account for multiply
scattered NIR excitation and emission light in tissue or within a
scattering solution can result in erroneous identification of intrinsic
decay kinetics. Upon conducting phase-modulation measurements on a
solution of intralipid containing the NIR excitable fluorophore,
indocyanine green (ICG), Lakowicz and Abugo (1999)
did not incorporate
the propagation of light, yet attribute multiexponential decay kinetics to this dye, which typically exhibits single exponential decay kinetics.
Approaches to appropriately model the multiple scattering of NIR
excitation and fluorescence photons and to use diffusion models for
quantitative spectroscopy have been previously demonstrated (Hutchinson
et al., 1996
; Cerussi et al., 1997
; Mayer et al., 1999
) for dyes
exhibiting single exponential decay kinetics. Because most dyes exhibit
multiexponential decay kinetics or exist within two or more different
states, lifetime spectroscopy within tissues or other scattering media
must consider kinetics beyond simple, first order decay. Indeed, the
presence of scattering increases the sensitivity of frequency-domain
measurements and accurate kinetic models need to be considered.
However, to date, there has been no attempt to perform lifetime
spectroscopy of dyes exhibiting multiexponential decays in tissue-like
scattering media. Herein, we present a model-based, experimental study
of fluorescence involving two dyes with distinctly different lifetimes
combined together at various concentration ratios in a multiply
scattering medium. By varying the concentration ratios of short- and
long-lived dyes, we tested the models and identify a feasible approach
for quantitative characterization of multiexponential decay kinetics in
tissues as well as in implantable sensors.
Because fluorescent decay kinetics may also reflect quenching, we also
investigated incorporation of the stretched exponential decay kinetics
into the diffusion equation to predict experimental measurements. The
use of multi- and stretched exponential functions accounts for
intrinsic kinetics of the fluorescence decay, whereas the diffusion
equation accounts for the influence of excitation and emission light
propagation through scattering media. Upon combining the fluorescence
decay models with the diffusion equation describing light propagation,
we seek to describe the propagation and generation of emission light
from analyte-sensing fluorophores for quantitative spectroscopy on the
basis of first principles.
In the following, we briefly provide background and theory for
fluorescence decay kinetics, light propagation in scattering media, and
fluorescent light generation and propagation in scattering media. We
then describe the instrumentation and frequency-domain measurements of
scattering and nonscattering solutions containing mixtures of two
dyes (ICG and 3,3-diethylthiatricarbocyanine iodine (DTTCI))
exhibiting distinct single exponential decay times.
 |
BACKGROUND AND THEORY |
Multiexponential decay fluorescence measurement in nonscattering
media
The impulse response function I(t) for a
multiexponential decay model, fitted to a sum of n
exponentials, is described by (Lakowicz, 1983
)
|
(1)
|
in which aj is a preexponential factor
representing the fractional contribution to the time-resolved decay of
the component with a lifetime
j. The average lifetime
avg for a fluorophore experiencing multiexponential
decay kinetics is given by
avg =
j fj
j in
which fj represents the fractional steady-state
intensity of each component in the mixture and can be calculated if
aj and
j are available (Lakowicz,
1983
). The values of aj and
j for a given sample can be estimated from frequency-domain measurements of
phase-shift
(
) and modulation-ratio M(
) by
minimizing the error-weighted sum of the squared deviations between the
measured and predicted values (
2) (Gratton et al.,
1984
).
Stretched exponential decay function in nonscattering media
In many cases, decay profiles from complex systems can be fitted
more appropriately by stretched exponential function of the form
(Inokuti and Hirayama, 1965
; James and Ware, 1985
; Nemzek and Ware,
1975
):
|
(2)
|
in which
and
are functions of diffusion coefficients of
the fluorophore and quencher, their encounter distance, and the quencher concentration. The type of decay profile in Eq. 2 has also
been studied to model diffusion controlled reactions as applied to a
first-order (simplest case) fluorescence quenching kinetics (Nemzek and
Ware, 1975
). Herein we use the stretched exponential function to
account for potential fluorescence quenching of one dye by another as
well as between the two dyes whose fluorescence is collectively
measured. In the former case the intrinsic decay kinetics from
simultaneously emitting dyes is expressed as I(t) = a1 exp(
t/
1) + a2 exp(
t
t0.5),
and in the latter case, by I(t) = a1
exp(
1t
1t0.5) + a2
exp(
2t
2t0.5).
Hirayama et al. (1990)
showed that the stretched exponential
function in Eq. 2 could be approximated by the sum of n
exponential terms.
|
(3)
|
If a set of aj and
j values
are available from a sufficient number, n, of exponential
terms, they can be used to generate a decay profile from which the
parameters
and
can be extracted using the relationship given in
Eq. 3 (Hirayama et al., 1990
). In this investigation, we similarly
extract stretch exponential decay parameters from the multiexponential
decay model containing the optimal number of components, n,
which fit the nonscattering data with the greatest degree of correlation.
Optical diffusion equation
Whereas the intrinsic decay kinetics provides the information for
fluorescence lifetime spectroscopy, the kinetics must be used and
derived within the context of the diffusion equation to appropriately
account for the influence of multiple light scattering. The diffusion
approximation to the radiative transfer equation describing the
time-dependent transport of light through a highly scattering medium is
given by
|
(4)
|
in which
|
(5)
|
U(r, t) is the density of photons at position
r and time t; v is the speed of light within the
medium; D is the optical diffusion coefficient;
µa is the linear absorption coefficient (i.e., the inverse of the mean free path for photon absorption with units of
inverse distance); µs is the linear scattering
coefficient (i.e., the inverse of the mean free path for photon
scattering); g is the average of the cosine of the
scattering angle; and q0(r, t) is the photon
source. The term µs(1
g) is often
referred as the isotropic scattering coefficient
µ's.
Eq. 4 has been solved for an infinite medium to yield an equation
describing transport of light through a scattering medium (e.g.,
tissue) for the case of a sinusoidally modulated point source of light
(frequency-domain) (Fishkin and Gratton, 1993
). The development of
time- and frequency-domain photon migration technique for
characterization of tissue absorbance and scattering has been well
established (Sevick et al., 1991
).
Generation of fluorescence light from a dye exhibiting single,
multi-, and stretched exponential decay
The photon density of fluorescent light,
Uf, which is generated and propagates within the
multiple scattering medium, can also be described by Eq. 4, provided
that its source emission kinetics is properly modeled. When a
fluorophore is distributed uniformly in a random media, the photon
density of excitation light, Ux, experiences
absorption owing to the fluorophore, in which µafx is the
absorption coefficient proportional to the fluorophore concentration.
The radiative relaxation of the activated fluorophore serves at a
distributed source of emission light within the random media. The
assumptions are: 1) the fluorophore concentrations are dilute so that
the probability that a fluorophore will absorb emission light and
refluoresce (i.e., secondary fluorescence) is negligible; and 2)
photobleaching does not occur. The emission source
qm in frequency domain for single, multi-, and
stretched exponential decay kinetics are:
|
(6)
|
|
(7)
|
|
(8)
|
in which Ux(r,
),
, and
m are excitation photon density, quantum efficiency of
the fluorophore, and the detection efficiency factor of the system at
the emission wavelength (which contains the system spectral response
and the fluorophore spectral emission efficiency (Lakowicz and
Szmacinski, 1993
)). The sources of emission light from a mixture of
fluorophores undergoing single-, multiple-, or stretched exponential
decays are simply a combination of the above expressions.
Upon using Eqs. 6 to 8 as the source term in Eq. 4, the generated
fluorescence photon-density Uf can then be
solved for single, multi-, and stretched exponential decay kinetics:
|
(9)
|
|
(10)
|
|
(11)
|
in which S is the fluence of the source (photons per
second); A is the modulation of the source; i is
a complex number (i =
);
is the
angular modulation frequency of the source; Dx
and Dm are the diffusion coefficients at
excitation and emission wavelengths, respectively. The terms
and
are functions of optical properties (µa and
µs), v, and
. In the case of
multiexponential decay kinetics, an additional fluorescence photon
density equation can be derived by incorporating the
avg, which results in an equation similar to that of Eq. 9.
|
(12)
|
For the mixture of two fluorophores of which one is quenched by
the other, the measured fluorescence should be predicted by a
combination of Eqs. 9 and 11.
The photon density can be related to the experimentally measured
phase-shift (
) and modulation-ratio (M) as a function of modulation frequency (Lakowicz et al., 1984
). The relationship between
fluorescence photon density to the measured fluorescence phase-shift
f and modulation-ratio Mf for a
dye undergoing single exponential decay kinetics in multiply scattering
solutions has been obtained (Mayer et al., 1999
). The measured
fluorescence phase-shift
f and modulation-ratio
Mf for a dye exhibiting multiexponential decay
kinetics can also be obtained using similar relationship. In the case
in which the decay kinetics are described by stretched exponential
decay kinetics, numerical integration needs to be performed in solving
the integration in Eq. 11 prior to attaining a useful relationship
between the fluorescence photon density to the measurable quantities.
 |
MATERIALS AND METHODS |
Dyes and solutions
Two fluorescence dyes with distinct lifetimes, ICG (Aldrich
Inc., Milwaukee, WI) and DTTCI (Acros Organics
Fisher Scientific, Fair
Lawn, NJ) were chosen in this study because they can both be excited at
780 nm, and their emission at 830 nm can be observed simultaneously,
giving rise to apparent multiexponential decay kinetics if relaxation
of the dyes occur independently. If quenching of at least one of the
two dyes occurs, stretched exponential decay kinetics should be
evident. The dyes were dissolved in a mixture (1:1 volumetric ratio) of
water and ethanol (EtOH, A406 P, Fisher Scientific, Fair Lawn, NJ) and
added to aqueous intralipid solution (Abbott Laboratories, North
Chicago, IL) to mimic the scattering properties of biological tissue.
Instrumentation and data acquisition
The data acquisition system schematic used for frequency domain
photon migration (FDPM) measurements is illustrated in Fig. 1 and briefly discussed in the following.
Two photomultiplier tubes (PMT) (Model H6573, Hamamatsu, Tokyo Japan)
were gain modulated by an amplified (ENI Amplifier, Model 403 LA,
Rochester, NY) radiofrequency (RF) signal from a frequency
synthesizer (Marconi Instruments Signal Generator 2022A, Mountain View,
CA), that was phase-locked to a second master oscillator used for laser
diode source modulation. The detectors were outfitted with
neutral-density filters (CVI Laser, Albuquerque, NM), narrow-bandpass
interference filters (10 nm full width half maximum, CVI Laser),
and focusing lens assemblies. In addition to neutral-density filters at
the photocathode, a neutral-density filter wheel was also used in the
setup to aid in adjusting DC levels of the detected PMT signals. The
PMTs were modulated at the same frequency as the source
(
0, on the order of 10 to 100 MHz), plus a small offset
frequency (
, 100 Hz). The result of mixing at the detector (i.e.,
heterodyning) was a sinusoidal signal, which contained frequencies of

as well as higher frequencies. These higher frequencies were
removed by transimpedance-amplifiers (TI-Amp, Model 70710, Oriel,
Stratford, CT), which acted as low pass filters, and are shown in Fig.
1 as TI-Amp. Consequently, the mixed signal, after passing through TI-Amp, was collected via standard data acquisition software (Labview 5.0, National Instruments, Austin, TX) and was processed to extract phase and modulation information. Depending upon the measurement (see
below), the source was either a 828-nm diode (40 mW, Thorlabs Inc.,
Newton, NJ) or a 778-nm diode (25 mW, DFS series; Melles Griot,
Boulder, CO) that was modulated by the master oscillator.
Measurement of fluorescence lifetime in nonscattering solution
Fig. 2 A illustrates
the incident and detected light schemes typical of fluorescence
measurements in nonscattering solutions (Lakowicz, 1983
). The incident
excitation light is directed to a beam splitter, which directed
approximately approximately one-third of the light to PMT1
via 1000-µm optical fiber (HCP-M1000T-08, Spectran, Avon, CT),
whereas the remaining portion was delivered to the sample. A second
fiber collected the fluorescent signal generated within the sample and
directed the signal to PMT2. A 780-nm interference filter
was used at PMT1 and an 830-nm interference filter was used
at PMT2. The phase-shift and modulation-ratio measured at
PMT2 for each sample were taken as a function of modulation frequency (10-120 MHz), and reported relative to that measured at
PMT1. As described in Appendix A, lifetime determination
from phase-shift and modulation-ratio measurements in nonscattering media required a reference measurement.

View larger version (18K):
[in this window]
[in a new window]
|
FIGURE 2
(A) Source and fiber-optically coupled
detection schematic for fluorescence measurements in dilute,
nonscattering solutions. (B) Fiber-optically coupled source
and detectors placement for fluorescence, excitation, and emission
measurements in scattering media. Fluorescence measurements used 778-nm
source, a 780-nm interference filter at PMT1, and an 830-nm
interference filter at PMT2 to collect fluorescent light.
When multiplexing was performed, the interference filters were switched
as well. Excitation measurements used 778-nm source and 780-nm
interference filters at both PMTs. Emission measurements used 828-nm
source and 830-nm interference filters at both PMTs.
|
|
Measurement of fluorescence lifetime in scattering solution
The incident and detected light configurations for measurements
in scattering solutions are depicted in Fig. 2 B. Three
different measurements were performed on fluorescent scattering
samples, 1) the fluorescence measurement, 2) the excitation
measurement, and 3) the emission measurement. The fluorescence
measurement of phase-shift and modulation-ratio (
f and
Mf) includes photon migration at the excitation
wavelength, generation of fluorescence, propagation at the emission
wavelength, as well as the fluorescence decay kinetics. The excitation
measurement of phase-shift and modulation ratio (
x and
Mx) and emission measurement of
(
m and Mm) are due solely to
photon migration at excitation and emission wavelengths.
For the case of fluorescence measurements in scattering solutions, an
optical fiber carrying the incident light (778 nm) was submerged inside
the sample so the solution to the diffusion equation for infinite media
was not violated. A second fiber, at a distance r1 from the source fiber, collected the
attenuated optical signal and directed the signal to PMT1.
At a distance r2 from the source fiber, a third
optical fiber collected the further attenuated signal and directed it
to PMT2. Measurement was conducted with a 780-nm
interference filter at the PMT that enabled measurement of the
attenuated optical signal at the excitation wavelength and an 830-nm
interference filter at the one that measured the signal at fluorescence
wavelength. The phase-shift and modulation-ratio data for each sample
was taken at a range of modulation frequencies (10-120 MHz).
The sample systems for excitation measurement and emission measurement
in scattering solutions were similar. The only difference was the
wavelength of the light source. Whereas light of 778 nm was used for
excitation measurement, 828 nm light source was used for emission
measurement. Interference filters were used in both PMTs to select
780-nm light for excitation measurements and 830-nm light for emission
measurements. For both measurement cases (excitation and emission), the
light was directed to a beam splitter (not shown). The smaller portion
of the light was directed to PMT1 via optical fiber
(not shown), whereas the remaining portion was delivered into the
sample. A second fiber, at varying distance (r2
ro), collected the attenuated optical signal from
inside the sample and directed the signal to PMT2.
Excitation and emission measurements were conducted to provide optical
property measurements at the two wavelengths.
The data acquired consisted of AC intensity
(Iac) and DC intensity
(Idc) measured at both PMTs, as well as the
relative phase (
rel) between the two PMTs, as a function
of modulation frequency. For each measurement at a given modulation
frequency, 20,480 of data sets (each set consists of one value
Iac, Idc, and
rel) in a total of 10 cycles that were collected in
approximately 2 s. The DC intensity was determined from the
average AC-intensity value of collected signals. The value of the DC
intensity was then subtracted from each detected value, and a fast
Fourier transform analysis was performed to provide AC and
.
Multiplexing and calibration
The phase and amplitude of the detected signal is subject
to the detection system's response functions. The effective response functions (i.e.,
instr and
Minstr), which are defined as the combined
response function effects of the two measurement channels, are mainly
due to differences in timing characteristics of the PMTs, TI-Amps, and
analog data cables. The differences in timing characteristics of the
two measurement channels may corrupt the phase and modulation (i.e.,
Iac/Idc) information
during the signal processing. Therefore, it is important that
corrections that are outlined in the Appendices A and B for these
response function effects to be performed.
 |
RESULTS AND DISCUSSIONS |
The sample identification and compositions used in the
experimental study are summarized in Table
I. The excitation and emission spectra of
0.4 µM ICG and 0.5 µM DTTCI in 50% (volume) EtOH are shown in Fig.
3. The spectra were obtained by SPEX
FLUOROLOG-2 (Model F212I, Jobin Yvon Spex, Edison, NJ) and recorded by
SPEX DM3000F. In the case of scanning the excitation spectrum, the emission wavelength was stationed at 780 nm, whereas in the case of
emission spectrum scanning, the excitation wavelength was stationed at
830 nm.

View larger version (21K):
[in this window]
[in a new window]
|
FIGURE 3
Excitation (emission at 830 nm) and emission
(excitation at 780 nm) spectra of 0.4 µM ICG and 0.5 µM DTCCI
solution in 50% EtOH.
|
|
Fluorescence multi- and stretched exponential decay in
nonscattering media
Measurements in nonscattering solutions was performed at various
ICG-DTTCI concentration ratios; SAMPLE A (0.05 µM ICG and 0.5 µM
DTTCI), SAMPLE B (0.15 µM ICG and 0.5 µM DTTCI), and SAMPLE C (0.4 µM ICG and 0.5 µM DTTCI). Each sample was diluted in 50% (volume)
EtOH. The dye concentrations were determined to obtain adequate signal
strength and resolution at various ICG-DTTCI concentration ratios.
The sample used as the reference was 0.5 µM of DTTCI in EtOH and
provided measurements of 
and M
as outlined in Appendix A. The
lifetime of DTTCI in EtOH is known to be 1.33 ns (Mayer et al., 1999
). The measurement data of phase-shift
rel and modulation
ratio Mrel are substituted into Eqs. A5 and A6
(see Appendix A), and the values of
and M after
correction for the instrument function are plotted in Fig.
4. The phase-shift and modulation-ratio
data at each concentration ratio was used to obtain the parameters aj and
j at the minimum value of
2 using the Nelder-Mead method (Matlab Optimization
Toolbox) (Press et al., 1992
). The frequency domain data were analyzed
for aj and
j in terms of a
biexponential model as well as an average lifetime model
both
correspond to independent radiative relaxation of the two dyes. The
results are listed in Table 2a. The
fractional steady-state parameter of species 1 (f1) increases with ICG concentration as one
would expect for independent relaxation of the two dyes.

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 4
Corrected values of fluorescence phase-shift
(A) and modulation-ratio (B) versus modulation
frequency for pure DTTCI (DTTCI-dilute), pure ICG (ICG-dilute), as well
as for ICG-DTTCI solution mixture at various concentration ratios in
nonscattering solution (50% EtOH). Samples were excited at 778 nm, and
the emission was observed at 830 nm.
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 2
Decay analysis at various mixture ICG-DTTCI concentrations
in nonscattering 50%-EtOH solutions based on (a)
biexponential, (b) stretched exponential, and
(c) bistretched exponential models
|
|
The lifetimes
1 and
2 are expected to be
constant at different concentration ratios, if negligible quenching
between ICG and DTTCI occurs. The variation of
1 and
2 with concentration ratio may be due to the error
associated with the optimization. An optimization study was also
performed using synthetic data generated with some randomly distributed
error model (data not shown for brevity). The study shows that
2 optimization of synthetic phase-shift and
modulation-ratio data with 1% random error in the
phase-modulation data within the frequency range as high as 120 MHz resulted in as much as 50% and 5% deviations from the true values
for lifetime for ICG and DTTCI. For similar synthetic phase-modulation
data at modulation frequency as high as 350 MHz, the deviations were
found to be as much as 3% and 1% from the true values for ICG and
DTTCI lifetime. When 5% error was considered, the optimization failed
to produce results from low frequency phase-modulation data limited to
120 MHz. Hence, phase-modulation values measured at high frequencies in
nonscattering solution are crucial for successful recovery of lifetime,
especially for subnanosecond decay kinetics. In conclusion, the results
presented in Table 2a may be consistent with independent relaxation
kinetics for the two dyes as modeled by an average and biexponential
decay model.
However, if the decay kinetics of ICG and DTTCI are not independent and
quenching occurs, then the stretched exponential decay functions may be
used to describe the decay kinetics. Originally, the multiexponential
model was used as an approximation to the stretched exponential decay
kinetics (Hirayama et al., 1990
), whereby an infinite number of
components could provide adequate representation. Because of the
complexities associated with deconvolution procedures, particularly for
those working in time-domain, recent studies in fluorescence lifetime
tend to use multiexponential rather than stretched exponential models.
Many even further simplify multiexponential into global
multiexponential model in which
j in Eq. 1 are assumed
to be constants. In this study, we investigated stretched exponential
decay kinetics to find a justifiable model, especially as the simple
two component multiexponential function fails in presence of scattering
(see below) and as the presence of scattering increases the range of
phase and amplitude measurements, which consequently enhances the
sensitivity to intrinsic kinetics.
The multiexponential decay parameters aj and
j(j = 1, n) were then used to generate a
decay profile from Eq. 1 and then fitted to the kinetic models of 1)
stretched exponential with a single exponential decay component (Table
2b) as well as 2) bistretched exponential decay (Table 2c). We fit the
scattering data to the maximal number of exponential terms, which did
not result in an increase in sum of square errors. In our case, the
maximum number of exponential terms required to provide the best fit
was n = 3. In the model using stretched exponential
decay kinetics, ICG presumably acts as a quencher to DTTCI as indicated
by the lifetime of the single exponential decay component of the model.
To explore the possibility that the two dyes are mutually quenched, the
bistretched exponential model was fit to the scattering data. It is
noteworthy that the results for the bistretched exponential decay
kinetics model also indicates that ICG does not act as the donor as the parameter estimate of
associated with the inverse lifetime (or
)
corresponding to ICG is zero.
In conclusion, the results from measurements in nonscattering solutions
show the inability to completely discriminate decay kinetic models at
moderate modulation frequencies (10-120 MHz) for the two dyes with
lifetimes separated by a factor of 2.
Fluorescence multiexponential decay in scattering media
The variations of ICG-DTTCI concentration ratio investigated in
scattering measurements were similar to that in the nonscattering measurements. Each 500-mL sample contained 2% (volume) of intralipid in a water-EtOH mixture (1:1 volumetric ratio) solvent (Table I). The
multiplexed fluorescence phase-shift and modulation-ratio data were
substituted into Eqs. B5 and Eq. B6 (see Appendix B), and the corrected
values of
and M are plotted in Fig.
5. Comparison of Figs. 4 and 5 shows that
scattering contributes significantly to the phase-shift and
modulation-ratio values.

View larger version (18K):
[in this window]
[in a new window]
|
FIGURE 5
Corrected values of fluorescence phase-shift
(a) and modulation-ratio (b) versus modulation
frequency for pure DTTCI (DTTCI-scatter), pure ICG (ICG-scatter), as
well as for ICG-DTTCI mixture at various concentration ratios in
scattering solution (2% intralipid). Samples were excited at 778 nm,
and the emission was observed at 830 nm. The distances
r1 and r2 in Fig. 2
b were 1 cm and 1.8 cm, respectively.
|
|
Model predictions of multiexponential lifetime decay in scattering
If the optical properties and decay kinetics information of a
scattering sample containing a fluorescence dye are known, the phase-shift
f and modulation-ratio
Mf due to the fluorophore lifetime and the
fluorescence light propagation can be predicted. In this study,
f and Mf are modeled as the phase
shift and modulation ratio arising from the propagation of excitation
and the generated emission light and from the independent relaxation of
the two dyes, or the quenching of one or both of the dyes. In the first case, the intrinsic kinetic models incorporated into the diffusion model were 1) biexponential decays and 2) the weighted average of the
biexponential decay. In the second case, the intrinsic models used
within the diffusion model were either 1) a stretched exponential
function to describe the donor fluorescence and a single exponential
decay component or (2) bistretched exponential decay kinetics to
describe mutual quenching between the two dyes. For the case that the
decay kinetics are described by the stretched exponential decay,
numerical integration was performed in mathematica. Problems such as
cut-off effects were not encountered in the range of frequency such as
that obtained in the data acquisition owing to the sharp decay of the
stretched exponential functions (short lifetime).
The optical properties of the scattering samples at various ICG-DTTCI
concentrations are shown in Table III.
The values were obtained from emission and excitation measurements and
from Eq. C1 (see Appendix C), in which KDC and
K
were the slopes obtained by linear
regression (Sun et al., 2002
) at a fixed modulation frequency and
varying source-detector separation (r2
ro) in Fig. 2 B. The values of
µa,x increase with dye concentrations was as expected.
The slight variation of µa,m,
µs,x, and µs,m values at
different concentration ratios may be due to the error associated with
the regression.
Fig. 6 shows the dependence of
fluorescence phase shift and modulation ratio upon modulation frequency
for SAMPLE E in 2% intralipid solution as measured
experimentally and predicted by the solution to the diffusion equation
that incorporated the four different kinetic models, 1) biexponential
decay, 2) single exponential decay representing average lifetime of the
bi-exponential decays, 3) stretched exponential decay with a single
exponential decay component, and finally 4) bistretched exponential
decay models. Specifically, we used the parameter estimates of the four
kinetic models obtained from measurements in nonscattering solutions
and presented in Table 2. Although all four intrinsic kinetic models provided comparable fit with values of
f and
Mf measured in nonscattering solution, in the
presence of scattering, the intrinsic kinetic models incorporated into
the diffusion equation could not provide comparable fits. Our results
show that the values of
f and Mf predicted from the diffusion equation incorporating the four
exponential models exhibit the same trend as that exhibited by the
experimental data for the various concentration ratios of the two dyes
as illustrated in Fig. 5.

View larger version (20K):
[in this window]
[in a new window]
|
FIGURE 6
Values of fluorescence phase-shift (a) and
modulation-ratio (b) as a function of modulation frequency
for corrected experimental measurements on SAMPLE E (black
diamond), and that predicted by the propagation model
incorporating biexponential decay kinetics (BE) (bold black
line), average bi-exponential decay (ABE) (bold gray
line), the stretched exponential decay with a single exponential
decay component (StE-SE) (dotted black line), and the
bistretched exponential decay (BStE) (thin gray line).
Values from Table 2 provide the parameter for predictions. The
refractive index of the medium was assumed to be that of 50%-EtOH
(1.345).
|
|
Fig. 6 illustrates the experimental values of phase shift and
modulation ratio measured from SAMPLE E and those that are predicted from the diffusion equation incorporating the four decay models. The
bistretched exponential model and the stretched exponential model
incorporating a single decay component provide identical results that
predict well the phase-shift data but less accurately predict
modulation data. On the other hand, the biexponential decay model fails
to predict either the phase-shift or modulation data suggesting that
independent relaxation of the two dyes does not occur. However,
the single decay model that incorporates an average lifetime provides
the best match to both the phase-shift and the modulation data. Similar
results are obtained from the remaining samples with differing
fluorophore concentration ratios.
Whereas the nonscattering data provided no discrimination between
the intrinsic kinetic models, the measurement data made in the presence
of scattering and predicted by the diffusion equation show that the
intrinsic kinetic models may very well be discriminated from one
another. Yet, owing to the difficulty in extracting decay kinetics
directly from the measurement data made in the presence of scattering,
our results cannot be used to definitively determine which decay model
fits the scattering data best. Whereas the lack of agreement with the
biexponential model prediction indicates the relaxation processes of
the two dyes are indeed dependent, the stretched exponential and
bistretched exponential models more closely predict the phase data
obtained in the presence of scattering. Most curiously, the average or
single exponential decay model fits best.
Fig. 6 also shows that in all of the decay models evaluated, the values
of modulation ratio are not well predicted. One reason for the lack of
fitting may be the difficulty in obtaining precise modulation data,
especially in absence of scattering, that critically depends on the
stability of source intensity, consistency of detection apparatus,
presence of ambient light, or electrical disturbances, etc.
Due to the fact that the fluorescence data measured in presence
of scattering are in good agreement with the average lifetime model,
fluorescence lifetime sensing of dyes exhibiting multiple decays may be
performed by simply obtaining the average lifetime. Table
4 shows the average lifetime values


that was recovered from phase-shift data acquired in presence
of scattering, at a fixed frequency of 80 MHz. The values in Table 4
are in good agreement with
avg shown in Table 2a. In
absence of scattering, subnanosecond lifetime detection would be
difficult at such low modulation frequency (as discussed in the
optimization study). The presence of scattering however, magnifies the
measured phase-shift value, and hence increases the sensitivity of
lifetime sensing, given that the optical properties of the scattering
solution can be accurately predicted.
View this table:
[in this window]
[in a new window]
|
TABLE 4
Average lifetime values calculated from phase-shift data
in presence of scattering, assuming multiexponential decay kinetics at
a fixed frequency of 80 MHz
|
|
 |
CONCLUSIONS |
In this contribution, we have extended fluorescence lifetime
spectroscopy in scattering solution to fluorophores that exhibit intrinsic decays not attributed to first order kinetics. Specifically, we used the diffusion equation to describe the propagation of excitation and emission light in scattering media and incorporated the
intrinsic decay kinetics described by average, multi-, and stretched-exponential decay models to predict frequency-domain measurements of multiply scattered emission light and demonstrate fluorescence spectroscopy in scattering solution.
The fluorescence decay kinetics of two emitting dyes in nonscattering
solutions could be equally well predicted by 1) biexponential decay, 2)
a stretched exponential with a single exponential component, and 3) a
bistretched exponential decay model. However, when the parameter
estimates from those models were used to predict frequency-domain measurements in scattering solutions using the optical diffusion equation, there was little agreement in the moderate frequency range
used. Yet the average lifetime obtained from the parameter estimates in
the biexponential model fit the data obtained in the scattering
solutions comparably better.
These results suggest that frequency-domain measurements of fluorescent
decay kinetics may be augmented with scatter to probe kinetics more
sensitively than in nonscattering solutions. Future work involves
assessing the multiple scattered data to directly obtain parameters
estimates of the kinetic models, as done with
avg in the
current study.
The instrument response function for the FDPM measurements in
scattering media can be corrected without the use of reference dye. The
correction can be obtained by multiplexing the optical signals of the
two detector fibers at two different positions in the sample (Mayer et
al., 1999
). The two detector fibers shown in Fig. 2 B are of
the same length, to ensure equal optical path lengths of the two signals.
The measured relative phase shift and modulation reflects light
propagation (and fluorescence generation, in the case of fluorescence measurements) in the sample as well as the instrument function
Theoretically, the multiplexing method described in previous
section can be performed for excitation and emission measurements as
well, and hence single distance (multifrequency) nonlinear regression
(Fishkin et al., 1995
) can be performed to obtain optical properties of
the medium. But a recent study in our laboratory showed that
measurements made at multidistances enable linear regression of
parameters (Fishkin et al., 1995
) and results in the most precise
optical property information (Sun et al., 2002
).
Address reprint requests to Dr. Eva M. Sevick-Muraca, Texas A&M
University, Chemical Engineering Department, 337 Zachry Engineering
Center, College Station, TX 77843-3122. Tel.: 979-458-3206; Fax:
979-845-6446; E-mail: sevick{at}che.tamu.edu.