| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, September 2002, p. 1691-1700, Vol. 83, No. 3
and
*Department of Biomedical Engineering, Johns Hopkins University,
Baltimore, Maryland 21205, and
Department of
Biomedical Engineering, Rice University, Houston, Texas 77005 USA
| |
ABSTRACT |
|---|
|
|
|---|
Optical scatter imaging (OSI), a technique we developed recently, was used to measure the ratio of wide-to-narrow angle scatter (OSIR) within endothelial cells subjected to calcium overload (1.6 mM) after permeabilization by ionomycin. Within a few minutes of calcium overload, the mitochondria, which started as elongated organelles, rounded up into spherically shaped particles. This change in morphology was accompanied by a statistically significant 14% increase in OSIR in the cells' cytoplasm. Mitochondrial rounding and OSIR increase were suppressed by cyclosporin A (25 µM), implying that the observed geometrical and scattering changes were directly attributable to the mitochondrial permeability transition. The angular scattering properties of a long mitochondrion rounding up were approximated by numerical simulations of light scatter from an ellipsoid rounding up into a sphere. The simulations predicted a relative increase in OSIR comparable to that measured experimentally for the case where the shape transition takes place with little or no volume increase. The simulations also suggested that mitochondrial refractive index changes could not account for the OSIR changes observed. Our data show that changes in OSIR correlate with mitochondrial morphology change in situ. OSI provides a new tool for subcellular imaging and complements other microscopy methods, such as fluorescence.
| |
INTRODUCTION |
|---|
|
|
|---|
Alterations in mitochondrial morphology have been
associated with mitochondrial metabolic state (Lehninger, 1959
; Packer, 1960
, 1967
; Hackenbrock, 1966
; Harris et al., 1969
; Hunter et al.,
1976
; Halestrap, 1994
; Territo et al., 2001
) as well as ischemic and
apoptotic cell death (Herdson et al., 1964
; Pastorino et al., 1995
;
DiLisa et al., 1998
; Vander-Heiden et al., 1997
; Martinou et al., 1999
;
Bernardi et al., 2001
; Mootha et al., 2001
; Scorrano et al., 2002
).
Whereas change in mitochondrial morphology could be assessed by
electron microscopy, dynamic studies of viable mitochondria typically
use light scattering to study this organelle, whose size is close to
the optical resolution of microscopes. Light scattering is sensitive to
changes in the size and shape of particles with dimensions on the order
of the wavelength. The light scattering measurement may be carried out
either in a spectrophotometer with the mitochondria suspension
contained in a regular cuvette or by flow cytometry. Measurements of
light transmission or angular light scattering at 90° from a
suspension of isolated mitochondria have long been correlated with the
morphology of mitochondria in the orthodox and condensed states
(Hackenbrock, 1966
; Hunter et al., 1976
). Since these early studies,
light scattering has become the technique of choice to detect
mitochondrial size change. Light scattering techniques have proved
essential in studying the mitochondrial permeability transition (Hunter
and Haworth, 1979
; Bernardi et al., 1992
; Bernardi, 1992
; Petronilli et
al., 1994
; Hoek et al., 1995
; Constantini et al., 1996
; Scorrano et al., 1997
; Kristal and Dubinsky, 1997
). More recently, the same light
scattering measurements were applied to the detection of mitochondrial
morphology change during apoptosis using flow cytometry (Vander-Heiden
et al., 1997
) and by measuring changes either in 90° scatter or
absorbance in a spectrophotometer (Zamzami et al., 1996
; Jurgensmeier
et al., 1998
; Narita et al., 1998
; Finucane et al., 1999
). In these
studies, mitochondria are usually isolated from the cell before the
light scatter measurement can be carried out.
The goal of the present study is to show whether light scattering could
be used to track relative changes in mitochondrial morphology within
living cells. We use a recently developed optical scatter imaging (OSI)
technique, which can track particle size in situ (Boustany et al.,
2001
). The OSI technique consists of a combination of light scatter
spectroscopy and microscopy. Optical scatter images of monolayers of
cells in culture are generated. These images directly encode the ratio
of wide-to-narrow angle scatter within the full field of view. Although
the cell cytoplasm may contain many several potential scatterers
consisting of the different organelles, earlier results have shown that
mitochondria should provide a very significant scattering signal
(Beauvoit et al., 1994
, 1995
).
We had previously validated the OSI technique on polystyrene sphere
suspensions, and shown that the measured scatter image ratio (OSIR)
decreases monotonically with sphere diameter as predicted by theory
(Boustany et al., 2001
). Here, we validate OSI in a biological system.
We use calcium injury as a model of mitochondrial size/shape change and
demonstrate how the OSIR correlates with calcium-induced alterations in
mitochondrial morphology in situ. Furthermore, we compare the scatter
changes observed to theoretical light scattering predictions from
ellipsoids and spheres. Both the experiments and the theoretical
modeling are used to quantify the morphological response of
mitochondria to calcium overload.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Cell culture
Bovine aortic endothelial cells (Clonetics, Walkersville, MD) were cultured on glass coverslips in Dulbecco's modified Eagle's medium (DMEM) supplemented with 10% fetal bovine serum (FBS), 2 mM L-glutamine, 20 U/ml heparin, 0.012 mg/ml bovine brain extract, 40 U/ml penicillin, and 40 µg/ml streptomycin. DMEM, FBS, L-glutamine, penicillin, and streptomycin were from Life Technologies (Rockville, MD). Heparin was from Sigma Chemical Co. (St. Louis, MO). Bovine brain extract was from Clonetics (Walkersville, MD). The cells were kept in culture at 37°C in a 5% CO2 in air atmosphere.
Calcium experiments
Each coverslip with attached live cells was mounted by means of
a steel plate onto the stage of the inverted microscope. Just before
mounting onto the microscope's stage, the growth medium was replaced
with a salt solution containing calcium and magnesium (solution 1 in
Table 1). While still on the microscope
stage, the cell membranes were permeabilized for 15 min in a
calcium-free salt solution (solution 2 in Table 1) supplemented with 20 µM of the calcium ionophore, ionomycin (Calbiochem, San Diego, CA). After permeabilization, solution 2 was replaced by a final salt solution containing calcium, which can now pass through the cell membranes. Two different calcium concentrations (solution 3 or 4 in
Table 1) were used after permeabilization to probe the response of
mitochondria to high calcium. In the data presented, solution 4 (0.8 mM
CaCl2) was used for the cell depicted in Figs. 2
and 3, whereas solution 3 (1.6 mM CaCl2) was used
for the cells discussed in Figs. 4 and 5. To assess whether the
mitochondrial changes observed were caused by the mitochondrial
permeability transition, cyclosporin A (CsA) was used to block the
mitochondrial permeability transition, (Broekemeier et al., 1989
;
Zoratti and Szabo, 1995
). In this case, 25 µM cyclosporine (Novartis,
East Hanover, NJ) was added to all solutions, before, during, and after
ionomycin permeabilization. All calcium experiments were conducted at
room temperature.
|
In selected cases, the live cells were pre-labeled with the fluorescent mitochondrial probe Mitotracker Green (Molecular Probes, Eugene, OR) to visualize specifically the mitochondria. For this, the cells were incubated for 45 min in growth medium supplemented with 100 nM Mitotracker Green before transferring the coverslips to the microscopy setup. Mitotracker labeling was done at 37°C in the 5% CO2 in air atmosphere.
Apparatus and data acquisition
The OSI technique was described previously in detail
(Boustany et al., 2001
). Fig. 1 shows the
microscopy setup. The specimens were mounted on the stage of an
inverted microscope (Nikon, Eclipse T300, Melville NY), which was also
fitted with epifluorescence and differential interference contrast
(DIC) imaging capabilities. The microscope condenser was adjusted to
central Kohler illumination, with a condenser numerical aperture (NA)
of 0.03 (condenser front aperture closed). For illumination, light from
the microscope's halogen lamp was filtered to yield an incident red
beam,
= 630 ± 5 nm. The images were collected with a
×60 oil immersion objective, NA = 1.4, and displayed on a charge
coupled device camera (Sensicam, Cooke Corp., Auburn Hills, MI). In a
Fourier plane conjugate to the back focal plane of the objective, a
beam stop was placed in the center of an iris with variable diameter.
As the inset in Fig. 1 shows, the variable iris collected light
scattered within a solid angle, bound by 2° <
< 10° for low
NA, and 2° <
< 67° for high NA. The diameter of the beam
block (0.5 mm), the smallest iris diameter, and the NA of the
microscope objective defined the angles 2°, 10°, and 67°,
respectively. The angles 2° and 10° were measured from the
diffraction pattern of a grid of lines with spacing a = 10 µm.
|
For each specimen studied, two sequential dark-field images, collected
20-30 s apart, were acquired at high and low NA by manually adjusting
the diameter of the variable iris. A solvent sample consisting of the
salt solution in which the sample was prepared served to provide the
background scatter signal due to the microscope optics. This background
signal was subtracted from each image. The background-subtracted
dark-field images were binned into pixels that, depending on the final
magnification, corresponded to regions in the object between 2 × 2 µm2 and 2.5 × 2.5 µm2. The OSIR was then measured at each pixel
bin by dividing the background-subtracted high NA image by the
background-subtracted low NA image. Given the aperture settings used,
OSIR can be defined as:
|
(1) |
,
) gives the intensity of the light
scattered in a given direction defined by the angles
and
.
is the angle between the scatter direction and the direction of
propagation of the incident light, and
is the azimuthal angle of scatter.
Light scatter simulations
Simulations of light scattering from spheres and ellipsoids were
carried out utilizing the finite-difference time-domain technique (Dunn
et al., 1996
; Drezek et al., 1999
). These simulations were used to
calculate the scatter intensity function F(
,
) in Eq. 1. The
scatter ratio, OSIR, was calculated after numerical integration of the
scatter intensity function F(
,
) according to Eq. 1. In these
simulations, the incident light was naturally polarized, and the
incident wavelength was
= 630 nm. The cytoplasm refractive index was 1.35. The initial mitochondria refractive index was 1.4.
| |
RESULTS |
|---|
|
|
|---|
Effect of calcium treatment on mitochondria and light scattering
In Fig. 2, differential interference
contrast (Fig. 2, A and C) and fluorescence (Fig.
2, B and D) images depict the sudden change in
the shape of mitochondria within a few minutes of calcium overload.
Mitotracker Green, which was used to stain the mitochondria, shows a
fluorescent pattern in Fig. 2 B consistent with other studies of endothelial cell mitochondria stained with functional dyes
such as fura 2 and rhodamine 123 (Steinberg et al., 1987
), rhodamine 6G
(Siemens et al., 1982
), or a fluorescent antibody toward a
mitochondrial protein (Lewis et al., 1991
). The mitochondria (arrows),
which are the elongated organelles before calcium injury (Fig. 2,
A and B) round up in response to the sudden
increase in cytoplasmic calcium (Fig. 2, C and
D). Fig. 3 shows the
dark-field images and ratiometric scatter images for the cell depicted
in Fig. 2, before and after the calcium insult (upper and lower panels, respectively). Fig. 3, A and D, shows the
dark-field images collected with the high NA setting of the variable
iris before (Fig. 3 A) and after (Fig. 3 D) the
calcium overload. Fig. 3, B and E, shows the
dark-field images collected with the low NA setting of the iris. In
these dark-field images, the transmitted light has been blocked out by
the beam stop in the center of the variable iris (see Fig. 1), and the
intensity is directly proportional to the intensity of the light
scattered by the particles (organelles) within the cell. By comparing
Fig. 3, A and D, with Fig. 2, B and
D, we find that a considerable number of scatterers within the cytoplasm of endothelial cells correspond to mitochondria (arrows
in both Figs. 2 and 3).
|
|
In these dark-field images, certain scattering effects can also be observed. For example, one would expect from scatter theory that the larger a particle, and the larger the difference between its refractive index and that of the surrounding medium, the larger its scattering cross section, and the brighter it will appear in this central dark-field imaging configuration. Moreover, the ratio of wide- to narrow-angle scatter will generally decrease with increasing particle size. These scattering effects can be observed in Fig. 3, A and B, keeping in mind that index differences among organelles are relatively small. For example, the nucleoli (e.g., particle at grid coordinate 110 (row) and 140 (column) in Fig. 3 A) appear brighter than the smaller organelles in the cytoplasm because of their larger size. Moreover, the difference between the nucleoli and organelle intensity is accentuated at low NA (Fig. 3 B). Larger particles are expected to have a more forward-directed scatter. Thus, the intensity of small organelles becomes more significantly attenuated than that of larger organelles when the iris diameter is reduced at F' in the optical setup (see Fig. 1).
The OSI technique consists of taking the ratio of the high to low NA
image after binning, which ensures that each pixel bin includes most of
the scatter intensity from at least one particle (see Materials and
Methods). The ratiometric scatter images are shown in Fig. 3,
C and F. The change in the morphology of the mitochondria previously shown in Fig. 2 is accompanied by a significant intensity increase in the optical scatter image (Fig. 3,
C
F). The subcellular regions exhibiting the
largest change in OSIR correspond to the regions in which the
mitochondrial shape changes result in the most pronounced changes in
angular scatter properties. These regions may not always correspond to
the regions of most intense Mitotracker fluorescence signal, which
itself does not depend on angular scatter properties but rather on
mitochondrial concentration and retention of Mitotracker in the
mitochondrial matrix. The OSIR increase depicted in Fig. 3 was
reproducible. After segmenting the images, the OSIR was averaged over
the cytoplasm of each tested cell, excluding the nucleus. The mean OSIR
values per cell thus generated were subsequently normalized, then
averaged and plotted in Fig. 4 (filled
circles) for all cells (n = 70) subjected to calcium
injury. The average OSIR values before and a few minutes after calcium
addition are listed in Table 2. The mean
OSIR increased by 14% and went from 1.47 to 1.68 upon calcium addition. Adding the mitochondrial permeability transition inhibitor CsA resulted in no significant scatter ratio change (Fig. 4, open squares, n = 64) and no mitochondrial rounding (an
illustrative example is shown in Fig. 5).
The suppression of OSIR increase by CsA shows that mitochondrial
changes resulting from the permeability transition fully account for
the scatter changes observed. Using solution 4 (0.8 mM calcium) instead
of solution 3 (1.6 mM calcium) after permeabilization (see Table 1)
resulted in 63% of the cells exhibiting no mitochondrial changes at
0.8 mM calcium, compared with 6% at 1.6 mM. However, at either of the
two calcium concentrations, whenever the cells responded to the calcium
insult, they exhibited the same mitochondrial and OSIR changes.
|
|
|
Light scatter simulation of mitochondrial rounding
To assess whether the rounding of elongated scatterers does in
fact result in an increase in scatter ratio, we conducted simulations of light scattering by ellipsoidal and spherical particles. Ellipsoids were used to approximate mitochondrial shape before the calcium injury;
spheres were used to model the mitochondria after the injury. Based on
published morphometric measurements in liver parenchymal cells (Loud
,1968
), we find that the short axis of mitochondria may vary between
0.3 µm and 0.6 µm, whereas the length of the mitochondria may vary
from 2 µm to as much as 11 µm. This variation in mitochondrial
length was reflected in our endothelial cells. Some endothelial cells
seemed to have long mitochondria (the cell in the center of the field
in Fig. 2 B), whereas others (such as the cells adjacent to
the one in the center of the field in Fig. 2 B) had shorter
mitochondria. In addition, mitochondria may appear artificially long in
certain cases where long strands may actually correspond to two shorter
and overlapping mitochondria in different focal planes. In light of the
difficulty in obtaining an accurate mitochondrial length estimate
directly from our images, and the fact that mitochondrial width is
below the resolution of our microscope, we relied on the mitochondrial
measurements published by Loud to estimate the initial mitochondrial
dimensions. The average short and long axes diameters of the
mitochondria before the calcium insult were taken to be 0.5 µm and 4 µm, respectively. Three sphere diameters were used to model the
mitochondria after the calcium overload. As shown in Table
3, upon rounding into a sphere of the
same volume (case A
B), the scatter ratio increased by 10%, in close
agreement with the 14% relative OSIR increase that was measured.
However, if the sphere diameter keeps increasing (cases
B
C
D), OSIR decreases. Thus, the modeling results suggest that
OSIR increase results from a mitochondrial shape change, which
consisted primarily of rounding with little volume increase, or
swelling.
|
It is important to reemphasize that the scattering response modeled by the ellipsoid in Table 3 corresponds to an average representative case that can result in a relative OSIR increase as predicted by scatter physics. As was discussed above, the mitochondria in our experiment may have varying lengths and diameters. In general, at any given diameter between 0.3 and 0.5 µm, and up to a length of ~5 µm, our model predicts that OSIR will increase upon particle rounding at constant volume. Rounding of ellipsoids longer than 5 µm will, however, result in OSIR decrease according to our model. This is because when the length of the ellipsoid is larger than 5 µm, the angular scattering properties are already dominated by the width of the ellipsoid, and the OSIR is independent of the ellipsoid's length. Thus, at any given width between 0.3 and 0.5 µm, ellipsoids longer than 5 µm will have the same OSIR as the 5-µm-long ellipsoids, even though their volume is larger. At the same time, the OSIR will continue to decrease for spheres with corresponding increasing volumes. Thus, very long ellipsoids with high aspect ratios will typically round up into spheres with lower OSIR, whereas ellipsoids with lengths smaller than 5 µm round up into spheres with higher OSIR. Because the average OSIR increased by 14% in our study, then based on the model prediction, this increase would result from the rounding of particles that are 4-5 µm in length (for diameters between 0.3 and 0.5 µm). Based on the model, we then conclude that the mitochondria are behaving on average like 4-5-µm-long particles. This average 4-5-µm length is consistent with the observation in our images and with the lengths of mitochondria reported in the literature.
To investigate whether changes in refractive index could account for the OSIR increase, we varied the refractive index of the initial ellipsoid (cases A' and A" in Table 3). The OSIR changes resulting from variations in refractive index were negligible. Although changes in mitochondrial index may have occurred as a result of the calcium overload, these changes could result only in intensity changes in the original dark-field images. In particular, a change in organelle refractive index may be responsible for the intensity increase in the initial dark-field images going from Fig. 3 A to Fig. 3 D upon calcium overload. However, the scatter predictions suggest that once the ratio of high NA to low NA signal is taken, the intensity of the ratiometric OSI images (Fig. 3, C and F) is itself insensitive to refractive index change. The theoretical predictions therefore imply that the change in OSIR is primarily caused by the mitochondrial shape change.
Fig. 6 shows the scattering intensities
as a function of the angles
and
(see Eq. 1) for cases A and B
in Table 3. The scattering
-
plane depicted in Fig. 6 is
equivalent to a diffraction pattern. The scattering angle
increases
in the radial direction, with
= 0 corresponding to forward
scatter (center of graph). The azimuthal scatter angle
increases
counterclockwise. In Fig. 6, the scatter intensity integral at high NA
is represented by the area bound by the dashed white circle and center
stop (2° <
< 67°), whereas the low NA integral is the area
bound by the dotted black circle and center stop (2° <
< 10°). These high and low limits also illustrate the diameter of the
variable iris in the plane F' of the optical setup (Fig. 1). Thus, the
OSIR calculated in Eq. 1 is a measure of angular scatter anisotropy in
the
direction, represented by the ratio of side to forward angle
scatter intensity. Anisotropy in the
direction is averaged out by
integrating
from 0° to 360°. For spheres, for which there is no
scatter anisotropy in the
direction (Fig. 6 B), the OSIR was previously shown to decrease monotonically for diameters
D, 0.2 µm < D < 1.5 µm, at 630 nm (Boustany
et al., 2001
). In contrast, nonspherical elongated particles exhibit
both
and
anisotropy (Fig. 6 A). The present
theoretical predictions show that particle shape change results not
only in alteration in
anisotropy but also in a measurable change in
anisotropy (OSIR change between cases A and B in Table 2).
|
| |
DISCUSSION |
|---|
|
|
|---|
The optical scatter imaging technique, OSI, presented here
complements current microscopic methods by providing images that directly encode angular scatter data from subcellular organelles. The
intensity of the ratiometric scatter images corresponds to the ratio of
side-to-forward scattering intensity measured at each pixel. This
scatter ratio, OSIR, correlates with subcellular morphological change.
In this paper, OSIR was shown to increase within the cytoplasm of cells
subjected to calcium injury. In these cells, there was a concomitant
rounding of mitochondria, which appeared initially elongated. The
calcium-induced mitochondrial rounding and scatter changes that ensued
were suppressed by the mitochondrial permeability transition inhibitor,
CsA. Thus, the measured scatter changes were directly attributable to
mitochondria and the mitochondrial permeability transition. These
results also suggest that the mitochondria are a significant
subcellular scatterer, and corroborate previous data underlining the
large contribution of mitochondria to the scattering of biological
tissue (Beauvoit et al., 1994
, 1995
).
Mitochondrial content may vary greatly from cell to cell (Beauvoit et
al., 1995
). Comparing endothelial cells with hepatocytes for example,
we find that the volume fraction of mitochondria in hepatocytes is
28.3%, whereas it is only 4.3% in endothelial cells (Blouin et al.,
1977
). Because we have been able to successfully correlate our scatter
signal with changes in mitochondrial morphology in endothelial cells,
which have a relatively low mitochondrial content, we would expect that
detecting changes in light scatter by mitochondria should be possible
in cell types where the volume fraction of mitochondria is at least
4%. Cells with higher mitochondrial content, such as hepatocytes,
would be expected to provide better mitochondrial scattering signals.
It is also important to note that by being able to spatially exclude
the nucleus in our analysis, we have eliminated a significant
scattering component of the cell. This may explain why even though the
volume fraction of mitochondria is low in endothelial cells, we can
still detect a significant mitochondrial signal. Additional
experimental studies are, however, necessary to show how the technique
will actually work in other cell types. In particular, changes in
angular scatter by certain cells could potentially originate from
alterations in a highly scattering subcellular structure in addition to
mitochondria. In this case, and as for this paper, attributing the OSIR
changes to mitochondria will depend on proper microscopic
observation, biochemical manipulation, and optical scatter modeling.
Very inhomogeneous cells may also be analyzed by considering the OSIR
in specific subregions of the cell, as opposed to averaging the OSIR
over the whole cytoplasm as was done in our study. Conversely, the ability of the technique to detect general morphological change may
make it amenable to studying biological problems involving organelles
other than mitochondria.
In our system, the suppression of the scatter changes by CsA in
combination with microscopic evidence served to identify the mitochondrial permeability transition as the biological process underlying our measurement. On the other hand, the theoretical simulations of light scattering from an ellipsoid rounding up served to
elucidate the optical basis behind the measured increase in OSIR. The
approximation of the mitochondria as ellipsoids rounding up after
calcium overload resulted in predicting a relative increase in the
OSIR, on the order of 10%, which was in close agreement with the
relative 14% increase in OSIR that was measured in cells. In addition,
the simulations showed that the calculated OSIR was not significantly
sensitive to refractive index changes. This suggests that changes in
OSIR can be solely attributed to changes in particle size or shape. We
had previously shown that the scatter ratio is independent of
refractive index for spheres with diameter D, 0.2 < D < 1.5 (with a wavelength of 630 nm) (Boustany et al., 2001
).
Here, this result is extended to the case of an ellipsoid.
The approximation of mitochondria as ellipsoids rounding up into
spheres in response to large calcium concentration sheds light on the
angular scattering properties of a particle subjected to a sudden shape
change. The current simulations of light scattering by mitochondria are
consistent with experimental observation to the extent that they show
that mitochondrial rounding can result in a relative OSIR increase.
Nonetheless, the actual average OSIR value measured in the cell
cytoplasm went from 1.47 to 1.68 (Table 2) upon mitochondrial rounding
compared with the calculated OSIR values, which went from 3.33 to 3.67 for ellipsoids rounding up into spheres (case A
B in Table 3). The
discrepancy in absolute OSIR values could be because of the fact that
mitochondria start out like rounded cylinders rather than ellipsoids
and round up into spheres of larger volume than estimated. This could
result in lower OSIR values than those presently predicted. To take
into account the exact level of final swelling in situ and the more subtle shape/size changes that may accompany the gross change in shape
will require more elaborate models of the initial and final
mitochondrial matrix geometry. Ultimately, geometrical models of the
mitochondria could be refined to take into account the precise
mitochondrial ultrastructure (Frey and Mannella, 2000
). Simulations may
also become available to predict the dark-field image of
three-dimensional scatterers on a pixel-by-pixel basis (Ovryn and Izen,
2000
), thus taking the imaging properties of the optical system also
into account.
The mitochondrial morphological changes observed in situ in this study
differ from those observed in isolated mitochondrial suspensions.
Isolated mitochondria are initially spherical in shape as opposed to
elongated rounded cylinders. Studies of the mitochondrial permeability
transition in isolated mitochondria do not show rounding of the
mitochondria from an initially elongated shape. Instead, the
mitochondrial permeability transition in isolated mitochondria is
typically accompanied by swelling of the spherically shaped
mitochondria and transition from an initial aggregated matrix state to
the noncondensed orthodox matrix configuration (Hunter et al., 1976
;
Pfeiffer et al., 1976
; Beatrice et al., 1984
; Petronilli et al., 1993
).
One exception is the study by Reed and Savage (1995)
who show that
CsA-dependent permeabilization of the mitochondrial inner membrane can
be achieved with or without swelling of isolated liver mitochondria.
Still in that study, the mitochondria are initially spherical. Thus,
the cytoplasmic environment and the presence of cytoskeletal proteins
may play an important role in maintaining the shape of mitochondria in situ. One study shows in situ calcium-induced mitochondrial rounding from an initial rod-like cylindrical shape, in agreement with our
observation of significant shape change (Kristal and Dubinsky, 1997
).
Nevertheless, in that study of mitochondria in cultured astrocytes, the
rounded mitochondria are reportedly swollen after calcium injury, with
a final diameter exceeding their initial measured length. This very
large amplitude swelling, which may accompany the shape change, was not
present in our experiments (see Fig. 2). Additional studies will be
necessary to assess how cell type and the cytoplasmic environment may
affect the morphological nature of the mitochondrial permeability
transition as observed in situ.
In this investigation, our transmission microscope was set up in the
central dark-ground configuration, and wide and narrow angles of
scatter were selected by varying the diameter of the aperture in a
Fourier plane conjugate to the objective's back focal plane. The
measured scatter ratio, OSIR, represented the wide-to-narrow scatter
intensity as an indicator of particle size/shape and was directly
mapped onto the full field of view. As depicted in Fig. 6, OSIR is a
measure of scatter anisotropy in the
direction. Previous scattering
studies of mitochondrial suspensions have used the measurement of light
transmission (or absorbance) or 90° scatter intensity to track
changes in mitochondrial volume. Although under certain conditions,
these measurements correlate well with mitochondrial morphology (Hunter
and Haworth, 1979
, Pfeiffer et al., 1976
; Petronilli et al., 1993
),
they could present some shortcomings. The general relationship between
transmitted light or light scattered at one single angle and particle
volume is not always monotonic (Bryant et al., 1969
; Latimer and Pyle, 1972
). Moreover, changes in refractive index also contribute to the
change in light scatter in addition to morphology change, thus
confounding data interpretation. A study by Knight et al. (1981)
shows
how changes in light scattered at 90° may not necessarily correlate
with mitochondrial volume change and points at the difficulty in
interpreting single-angle scatter data. By comparison, the OSIR
measured in this study has the advantage of decreasing monotonically over a large range of sphere diameters from 0.2 µm to 1.5 µm (see Boustany et al., 2001
). Moreover, the OSIR is not significantly sensitive to refractive index changes for spheres in this size range as
well as for ellipsoids with dimensions comparable to those of
mitochondria in situ.
In the present setup, the scatter intensity was averaged in the
azimuthal scatter direction
. Scatter anisotropy in the
direction may give information about particle shape for nonspherical particles and could be measured by redesigning the iris to take the
ratio of two perpendicular scatter directions, such as horizontal to
vertical scatter with respect to the
-
plane. This ratio would be
a function of the particle orientation with respect to the optical
axis. To provide particle shape anisotropy, the iris will need to be
rotated in the
direction while several ratiometric images are
collected and later processed. Alternatively, particle shape and
orientation may be probed by varying the polarization of the incident
beam. In general, our method consists of directly mapping
selective scatter information into the imaging plane by use of a
spatial filter in the Fourier plane. Although we have chosen to first
look at the ratio of wide-to-narrow-angle scatter, other Fourier plane
apertures could easily be designed to select different scattering parameters.
To date, most light-scatter-based studies on mitochondrial morphology
have been conducted in preparations of isolated mitochondria. Although
studies on the mitochondrial permeability transition can now be
conducted in situ by observing the movement of the fluorescent dye
calcein-AM across the mitochondrial inner membrane (Lemasters et al.,
1998
; Petronilli et al., 1999
), changes in the actual morphology of
mitochondria have rarely been directly assessed from an optical
measurement made in situ. One example is the study by Kristal and
Dubinsky (1997)
, where mitochondria in cultured cortical astrocytes are
classified according to their shape (rod-like or rounded) and counted
after labeling with the fluorescent dye JC1. Another is the study of
mitochondrial volume in situ using measurements of light scattering by
hepatocytes in suspension (Quinlan et al., 1983
). As shown in Fig. 2,
large amplitude mitochondrial shape changes could be detected by
differential interference contrast (DIC) and fluorescence. However, to
quantify the observed changes, one would have to analyze the DIC and
fluorescence images in great detail. OSI gives a rapid quantification
of the data. The particles being probed need not be individually
resolved and measured by traditional morphometric methods, thus
avoiding a tedious process of image recognition, particle sizing, and
counting. OSI complements current microscopic methods, allows real-time monitoring of the same cells, and could be easily automated and extended to high-throughput screening.
| |
ACKNOWLEDGMENTS |
|---|
We thank Prof. Rebecca Richards-Kortum from the University of Texas, Austin, for graciously giving a copy of the finite-difference time-domain software developed in her laboratory.
This work was supported by National Institutes of Health grant R21-RR15264. N.N.B. was also partially supported by a fellowship from the Whitaker foundation.
| |
FOOTNOTES |
|---|
Address reprint requests to Dr. Nada N. Boustany, Johns Hopkins University School of Medicine, Department of Biomedical Engineering, 720 Rutland Ave., Traylor 701, Baltimore, MD 21205. Tel.: 410-955-0077; Fax: 410-955-0549; E-mail: nnboustany{at}alum.mit.edu.
Submitted January 16, 2002, and accepted for publication May 1, 2002.
| |
REFERENCES |
|---|
|
|
|---|
Biophys J, September 2002, p. 1691-1700, Vol. 83, No. 3
© 2002 by the Biophysical Society 0006-3495/02/09/1691/10 $2.00
This article has been cited by other articles:
![]() |
A. A. Gerencser, J. Doczi, B. Torocsik, E. Bossy-Wetzel, and V. Adam-Vizi Mitochondrial Swelling Measurement In Situ by Optimized Spatial Filtering: Astrocyte-Neuron Differences Biophys. J., September 1, 2008; 95(5): 2583 - 2598. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-Y. Zheng, Y.-C. Tsai, P. Kadimcherla, R. Zhang, J. Shi, G. A. Oyler, and N. N. Boustany The C-Terminal Transmembrane Domain of Bcl-xL Mediates Changes in Mitochondrial Morphology Biophys. J., January 1, 2008; 94(1): 286 - 297. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Dobrucki, D. Feret, and A. Noatynska Scattering of Exciting Light by Live Cells in Fluorescence Confocal Imaging: Phototoxic Effects and Relevance for FRAP Studies Biophys. J., September 1, 2007; 93(5): 1778 - 1786. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Arifler, I. Pavlova, A. Gillenwater, and R. Richards-Kortum Light Scattering from Collagen Fiber Networks: Micro-Optical Properties of Normal and Neoplastic Stroma Biophys. J., May 1, 2007; 92(9): 3260 - 3274. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Wilson, C. E. Bigelow, D. J. Calkins, and T. H. Foster Light Scattering from Intact Cells Reports Oxidative-Stress-Induced Mitochondrial Swelling Biophys. J., April 1, 2005; 88(4): 2929 - 2938. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||