Fluorescence correlation spectroscopy (FCS) can provide a
wealth of information about biological and chemical systems on a broad
range of time scales (<1 µs to >1 s). Numerical modeling of the FCS
observation volume combined with measurements has revealed, however,
that the standard assumption of a three-dimensional Gaussian FCS
observation volume is not a valid approximation under many common
measurement conditions. As a result, the FCS autocorrelation will
contain significant, systematic artifacts that are most severe with
confocal optics when using a large detector aperture and aperture-limited illumination. These optical artifacts manifest themselves in the fluorescence correlation as an apparent additional exponential component or diffusing species with significant (>30%) amplitude that can imply extraneous kinetics, shift the measured diffusion time by as much as ~80%, and cause the axial ratio to diverge. Artifacts can be minimized or virtually eliminated by using a
small confocal detector aperture, underfilled objective back-aperture,
or two-photon excitation. However, using a detector aperture that is
smaller or larger than the optimal value (~4.5 optical units) greatly
reduces both the count rate per molecule and the signal-to-noise ratio.
Thus, there is a tradeoff between optimizing signal-to-noise and
reducing experimental artifacts in one-photon FCS.
 |
INTRODUCTION |
Fluorescence correlation spectroscopy (FCS)
(Magde et al., 1972
, 1974
; Elson and Magde, 1974
) is an elegant and
sensitive technique for measuring dynamic processes that manifest
themselves in a fluorescent signal on the submicrosecond-to-second time
scales. With origins in quasielastic light scattering (Cummins and
Swinney, 1970
) and a basis in the statistical thermodynamics of
fluctuations in solution, FCS has been used to measure diffusion
coefficients, chemical kinetics, excited-state molecular dynamics,
picomolar concentrations (Eigen and Rigler, 1994
), and the dynamics of
the interaction of fluorescent molecules in vitro (Koppel et al., 1976
;
Magde et al., 1972
). Recently there has been rapidly increasing use of
FCS for biological applications (Hess et al., 2002
; Rigler and Elson,
2001
; Schwille et al., 1996
; Maiti et al., 1997
) including living
biological systems (Schwille et al., 1999a
; Brock et al., 1998
, 1999
;
Hink et al., 2000
; Wachsmuth et al., 2000
; Cluzel et al., 2000
; Politz
et al., 1998
).
What are the best optical operating conditions for FCS? That question
comprises the focus of this study. Many FCS systems allow changes in
the type and numerical aperture (NA) of the objective lens, the size of
the detector aperture, and the degree of underfilling of the
back-aperture of the objective. Previous theoretical work has
investigated the signal-to-noise ratio (S/N) for FCS systems (Koppel,
1974
; Qian and Elson, 1991
), and confocal microscopes (Gu and Gan,
1996
; Sheppard et al., 1991
; Sandison and Webb, 1994
; Gu and Sheppard,
1993
), but most previous work has made significant approximations in
the treatment of the focal volume (Rigler et al., 1993
), particularly
at high-NA (Sheppard and Matthews, 1987
) and in the treatment of
underfilling, except in the case of two-photon excitation (W. Zipfel,
personal communication). It is commonly assumed that the observation
volume in FCS is a Gaussian in three dimensions. However, it will be
shown that this is not sufficiently accurate under many measurement
conditions and may lead to inaccurate results. To attempt to improve
upon current FCS methodology we use a more sophisticated description of
the focal volume (Wolf, 1959
; Richards and Wolf, 1959
), which treats
the polarization of the excitation and is well-suited to high-NA
optics. Furthermore, we present theoretical predictions and
measurements together. We also attempt to determine the values of the
detector aperture and underfilling fraction that yield the highest S/N
and minimize artifacts in the FCS autocorrelation function that result
from the non-Gaussian nature of the observation volume.
 |
THEORY |
Although FCS is an excellent technique for extracting a variety of
quantitative information from biological systems, the influence of the
illumination and collection optics on the measured autocorrelation must
be considered to ensure accurate results. This section delineates the
relationship between the spatial profile of the focal volume and
parameters measured by FCS.
Introduction to FCS
Fluorescence correlation spectroscopy (Magde et al., 1972
), which
is based on fluctuation analysis of the fluorescence from an
observation volume on the order of a femtoliter containing a small
number (<1000) of molecules, provides quantitative physical and
chemical kinetic information on time scales from 10
7 to
>102 s. FCS has been demonstrated to be quite useful for
photophysical characterization of sparse fluorescent molecules (Mertz
et al., 1995
) and measurement of dynamics of those molecules that give rise to fluctuations in their fluorescence (Schwille et al., 1996
; Heikel et al., 2000
). Technological innovations, adapted particularly in the Rigler and Eigen groups (Eigen and Rigler, 1994
), have resulted
in a recent revival of the technique, the fundamentals of which were
first developed in the early 1970s in the Webb group. The basis of the
technique is the observation of the fluorescence F, produced
by dilute fluorescent species (~nM concentrations) that diffuse and
react chemically according to
|
(1)
|
where the concentration fluctuation
Cj
of the jth species from the mean (temporal average)
Cj
is given by
Cj(r, t) = Cj(r, t)
Cj(r, t)
, as a function of time (t), the diffusion coefficient Dj,
and the tensor Tjk, which expresses the chemical
reaction kinetics and stoichiometry between species.
The concentration fluctuations result in fluorescence fluctuations
F(t), which are related to the experimental illumination and collection profiles (now considering a single species) by
|
(2)
|
where IO is the illumination intensity
profile, S
I(rO)/I(rO = 0) is the normalized illumination profile,
is the collection
efficiency profile, rO is position in object
space, and q,
, and C are the effective
quantum yield, extinction coefficient, and concentration, respectively,
of the observed fluorescent species. Clearly based on Eq. 2, the
spatial dependence of the product of S and
, the
normalized observation volume profile
|
(3)
|
is crucial in determining the spatial distribution of fluorescence
fluctuations that are to be observed. In practice, kinetic information
is extracted from the fluorescence fluctuations by autocorrelation
(Webb, 1976
):
|
(4)
|
where
F(t) = F(t)
F(t)
, and
is time delay. G(
) contains diffusion, chemical kinetic,
molecular brightness, concentration, and photophysical information by
quantifying the temporal decay (as a function of
) of the
fluorescence fluctuations.
A solution to the autocorrelation function may be formulated in the
case of ideal solutions and concentration- and position-independent diffusion coefficients using Green's function for classical diffusion kinetics:
|
(5)
|
where
is time, N is the number of particles,
D is the diffusion coefficient, and r is
position. The diffusion autocorrelation function can be calculated
using a double integral:
|
(6)
|
Autocorrelation function assuming a Gaussian observation volume
The standard autocorrelation function for a single fluorescent
species freely diffusing in an ellipsoidal 3D-Gaussian observation volume O(rO) at low intensity (well below
saturation) has the solution (Schwille et al., 1996
):
|
(7)
|
where
D is the characteristic (diffusion) time
molecules spend on the average in the observation volume,
is the
axial ratio (ratio of axial to radial dimensions of the observation volume), and N is the average number of molecules.
Autocorrelation and volume are clearly related in the limit
0, where
|
(8)
|
and C(r) = C0 is a
spatially constant concentration of molecules in the observation volume
of size V. Note that both x and y
(z is axial) contribute a factor of (1 + D
/
2)
0.5 to
GD(
), where
is the transverse
1/e2 radius of O(rO), but that Eq. 7
is the special case where
x =
y, and
hence only two diffusion time constants exist:
D
2/4D,
' = (
)2/4D.
For anomalous diffusion, where
r2
=
t
and 4Dt
t
,
the standard fitting function is (Schwille et al., 1999b
)
|
(9)
|
which also assumes a Gaussian observation volume, where the
temporal exponent
has the possible range 0 <
< 1, and the transport coefficient
is generally analogous to
4Dt1
.
The autocorrelation of multiple diffusing species is just a linear
combination of the autocorrelations for each species separately, weighted by their fluorescence intensity (Hess et al., 2002
):
|
(10)
|
where
i and
Gi(
) are the average fluorescence and
autocorrelation for the ith species, respectively, and
m is the total number of species. Processes occurring faster
than the diffusion time
D can be resolved as additional
factors that generate shoulders added to G(
) at
<
D. Because time scales are involved that do not
depend on the concentration of fluorescent molecules, FCS can be used
as a "ratiometric" means of measurement, i.e., it can also extract
concentration-independent information from a system.
Consider now transitions of the form B*
B, between bright (B*) and
dark (B) states of a molecule or mobile object, which occur on time
scales faster than diffusion and are hence observable by FCS, as in
triplet-state intersystem crossing, molecular conformational changes,
or certain chemical reactions. If such a reaction has an equilibrium
constant K, forward and backward reaction rate constants
k+ and k
, respectively,
and a free energy
G, as related by
|
(11)
|
the Boltzmann constant kB, temperature
T, and respective concentrations [B*] and [B], then a
characteristic time constant (
B)
1 =
k+ + k
, and molecular dark fraction
FB = [B]/([B*] + [B]) may be defined.
These quantities
B and FB
describe the reaction and are introduced as an exponential factor
FB exp(
/
B) in the
autocorrelation G(
), resulting in
|
(12)
|
Exploiting the fact that
GD(0)
1 = N and
C0 = N/V, the physical volume of the
observation volume (V) can be determined from an FCS
measurement at known concentration C0, or for a
known observation volume an absolute concentration can be obtained.
For multiple independent chemical kinetic processes that occur on
well-separated time scales, the fitting function is a product of the
diffusion and m chemical kinetic factors (Hess et al., 2002
):
|
(13)
|
with amplitude Fi and time scale
i.
The above derivation assumes a Gaussian profile for O(r).
However, if the profile is non-Gaussian,
GC(
), GD(
), and G(
) will all be distorted, and the kinetic parameters
obtained from measurements will be subjected to significant systematic errors. An understanding of the effects of the optics on the functional form of G(
) are thus crucial. Therefore, we consider the
optical problem of calculation of O(r) that depends on the
illumination and collection point spread functions of the FCS system,
and which allows us to calculate G(
) from
O(r).
Calculation of the point spread function
The point spread function for a lens system is defined as the 3D
image of a point source. It is also the intensity distribution in the
vicinity of the focal plane resulting from a point source of
monochromatic light in the image plane of the lens system. The paraxial
approximation (Born and Wolf, 1991
) is used to simulate confocal
optical systems (Sheppard et al., 1991
; Sandison and Webb, 1994
),
which is adequate for sin
< 0.8 (Sandison and Webb, 1994
), or
roughly NA = 1.06 in water, where
= sin
1
(NA/n) is the half-angle of opening of the objective lens in radians, and n is the refractive index. Here we consider a
non-paraxial, high-NA description (Richards and Wolf, 1959
), which is a
better approximation for
> 0.8 and ideal lenses (no aberrations).
The dimensionless optical coordinates used are
and u,
corresponding to radial and axial coordinates, respectively:
|
(14A)
|
|
(14B)
|
where k is wavenumber in the medium,
is wavelength,
and r2 = x2 + y2. The diffraction theory by Richards and Wolf uses
the complex integral representation:
|
(15)
|
for the energy density (intensity):
|
(16)
|
where A0 is a constant,
Ji are Bessel functions of ith order,
i are the integrals (Eqs. 15),
is the
integration angle, and
is the azimuthal angle around the
longitudinal axis (
= 0 corresponds to the x-axis).
Illumination optical geometry: the underfilling fraction
Underfilling the back-aperture of the objective can be used to
dramatically elongate and enlarge the illumination profile at the focus
of the objective and to reduce the effective NA of illumination, which
produces a nearly Gaussian illumination profile. Specifically, for a
Gaussian beam with 1/e2 intensity radius
r0 and objective back-aperture radius
rBA, the underfilling fraction
= rBA/r0 is just the ratio of the
back-aperture radius to the beam radius. For
< 1 the
objective is "overfilled," and for
> 1 the objective is
"underfilled." The apodization function in Eqs. 15
|
(17)
|
(W. Zipfel, unpublished data) is the factor that accounts for the
effect of
on the illumination profile.
Collection optical geometry and observation volume profile
After illumination with confocal optics, the second factor
determining the optical geometry in FCS is the collection profile, which accounts for the confocal detector aperture and diffraction of
collected fluorescence propagating from the object space into the image
space, and is combined with the illumination profile to calculate the
observation volume profile. The relative spatial collection efficiency
can be expressed as the contribution to the measured fluorescence
signal as a function of position in object space. Therefore we consider
a point rO in object space and the contribution
to the image it produces (a point spread function centered at the
corresponding image point ri, weighted by the
illumination at rO). The collected portion of
the fluorescence emitted from this point gives the collection efficiency for this point relative to other points in the object space.
We assume that the detector collects all of the fluorescence that
passes through the confocal detector aperture, which is by definition
in the image plane.
In the case of two-photon excitation, where there is typically no
confocal detector aperture, the observation volume is determined exclusively by the squared illumination intensity:
|
(18)
|
where S(rO) is the normalized
illumination intensity point spread function, which depends on the
illumination wavelength, intensity, and objective properties.
In confocal FCS, the collection through the detector aperture must also
be considered. The observation volume is defined not only by the
illumination profile factor S(rO), but
also by a collection profile factor
(rO)
resulting from the objective and tube lens properties, fluorescence
emission spectrum, and the confocal detector aperture in the image
plane, assuming incoherent emission (Born and Wolf, 1991
; Sandison et
al., 1995
). The observation volume is thus defined as:
|
(19)
|
where ri is the position of the element of
area in the image plane,
(rO) is the
centrosymmetric collection point spread function, M is the
total magnification of the system, and rO is the
object-space position. Note that an infinitesimal detector is
represented by a delta-function at ri = 0 and in this case the integral reduces to
|
(20)
|
This form reduces to the point spread function of a two-photon
(2P) microscope with no detector aperture, where
(rO )
S(rO),
yielding O(rO)
[S(rO)]2. The detector aperture
size in optical units is calculated using
|
(21)
|
where Rd is the detector aperture radius in
real space, and rd is its radius in
dimensionless optical units.
Calculation of collected fluorescence for incoherent emission
The time-averaged collected fluorescence
F
is
given by an integral over the object space
|
(22)
|
where O(r) is the observation volume spatial profile,
and
is a constant proportional to overall detection efficiency, dye
fluorescence excitation cross section, and quantum yield.
Observation volume
We define the volume of an illumination or observation profile
(note that volume is to be distinguished from profile) by an integral
over the appropriate space:
|
(23)
|
where W(r) is the spatial profile normalized to
unity at its maximum, i.e., W(r) = O(r)/O(r = 0). This definition is similar to
established definitions (Thompson, 1991
; Mertz et al., 1995
; Xu and
Webb, 1997
). Equation 23 is used to calculate the effective volume of
the calculated observation volumes.
Signal-to-noise ratio in FCS
Forming a correlation requires the observation of correlated
photons. The S/N ratio in FCS depends simply on being able to observe
pairs of photons emitted by a single molecule within a correlation time
. There must be an abundance of photons within the time window
for the correlation to have a reasonable S/N ratio, and hence the count
rate per molecule (
= F/N) has been shown to be
directly related to signal-to-noise in FCS (Koppel, 1974
). In the
regime (
)
1, the S/N ratio is proportional to
(Koppel, 1974
). The average fluorescence F is measured
routinely during FCS, and the zero-time value of the autocorrelation
function, G(0), gives the inverse of the mean number of
fluorescent molecules in the focal volume (N), if properly
corrected for background fluorescence. Corrections to N and
for measured (noncorrelating) background (Koppel, 1974
) are made
using
|
(24)
|
Therefore,
is an accessible experimental parameter that is
commonly used to optimize the S/N ratio. Fortunately,
is also accessible to our calculations once we have obtained fluorescence and
number of molecules (i.e., the physical volume of the observation volume). Because the maximization of S/N with minimal experimental artifacts is of great interest, the dependence of
and artifacts on
and rd are calculated and examined under the
same conditions.
Dependence of count rate per molecule on detector aperture
Small detector aperture limit
The value of
is greatly reduced by using a pinhole smaller
than the optimal size. This reduction is dictated by the rapid decrease
in collected fluorescence as the detector gets very small (F
is proportional to r
). For small pinhole values, however, the size of the observation volume does not go to zero
as the pinhole becomes infinitesimally small; instead, the profile of
the observation volume approaches the illumination profile squared.
Thus, F/V
0 as rd
0.
Large detector aperture limit
When the observation volume is non-Gaussian, the count rate per
molecule
also declines for very large rd.
This is clear if one considers
with spatially uniform concentration
C0:
|
(25)
|
where
is a constant (see above), I0 = I(rO = 0), and
W(rO) = O(rO)/O(rO = 0).
Note that O(rO = 0) is proportional to the
light collected from a point source at the origin as a function of
detector aperture, which reaches a maximum (constant) value at very
large aperture. Furthermore, as the pinhole is opened, the observation
profile more and more closely approaches the illumination profile. In
the presence of diffraction fringes in the observation volume at large
r and z, the integral of W will grow
more quickly than the integral of W2, which
contains the profile to the second power, which greatly reduces the
influence of the dim regions. Hence, at large detector apertures, the
product of O|r=0 ·
W2(r)dr/
W(r)dr declines, resulting in decreased count
rate per molecule. The best detector aperture (to give the maximum
)
is therefore a compromise between reduced collected fluorescence for
small aperture values and large observation volumes, which occur at
large aperture values.
No peak in
(rd) with a 3D-ellipsoidal
Gaussian focal volume
If a 3D-ellipsoidal Gaussian profile is assumed for illumination
and collection functions, as well as a Gaussian profile for convolution
with the pinhole, then the observation volume is Gaussian. With a
Gaussian observation volume and spatially constant concentration C0 the ratio
W2(r)dr/
W(r)dr
is constant, and hence the count rate per molecule
(rd) is proportional simply to
O(rO = 0), which is proportional to the
collected fluorescence from a point source at the origin (object plane
focus) and will only increase with increased detector aperture.
Therefore, using a Gaussian approximation for the observation volume
will never give a peak in count rate per molecule as a function of
detector aperture.
 |
MATERIALS AND METHODS |
Numerical implementation
The integrations in Eqs. 15 are done numerically using Bessel
functions J0(x), J1(x), and
J2(x) evaluated from x = 0 to
200 in steps of 0.02 and interpolated linearly. Illumination profiles agree well at low NA with paraxial results in the literature (data not
shown) and with paraxial calculations (Sandison and Webb, 1994
) at
intermediate NA. Furthermore, results as a function of
at high NA
agree with independent calculations by the method of Richards and Wolf
and measurements (W. Zipfel, personal communication, 2001).
Numerically, the observation volume profile (Eq. 19) is calculated as a
sum:
|
(26)
|
where r = (x, y, 0) and n is the
number of steps in the grid approximating the detector face, which is
contained in the x-y plane. The theta function excludes
values of r outside the detector radius. To accelerate
calculations, the values for the normalized illumination intensity
S(u,
) and collection efficiency
(u,
)
are calculated once and stored in 1000 × 1000 element
floating-point arrays. Values are then interpolated from the array
values for use in the observation volume calculations. Typical
parameters used are
= 1.12 (=1.2 NA in water),
from 0 to
200, u from 0 to 200, 1000 steps in both u and
,
= 0, 
= 0.002, and
1 for overfilled or
> 1 for underfilled. Typically a 20 × 20 detector grid
is used.
Calculated correlation curves
The double integral in Eq. 6 is solved numerically as a
convolution, followed by a product, followed by a single integral in
three dimensions, using a discrete three-dimensional array of
(128)3 = 2,097,152 elements. The theoretical
G(
) in Eq. 6 then becomes (for discrete elements)
|
(27)
|
where Tijk is a three-dimensional array
representing the result of the first spatial integral, calculated from
a product of the discretized spatial observation profile
Oijk, and the result of the convolution,
Aijk. The subscripts i, j, k denote spatial position along the three Cartesian axes. The convolution is between the
concentration function Cijk and a second
(identical) spatial observation profile,
O'ijk. Cijk is the
solution to the diffusion equation from Eq. 5 evaluated at the discrete
grid positions defined by the three orthogonal Cartesian coordinates i,
j, and k. Note that the Einstein summation convention is not
used below.
|
(28)
|
|
(29)
|
where
|
(30)
|
The theoretical GT(
) is calculated at
50 logarithmically spaced
values spanning ~4 orders of magnitude
in time. GT(
) can also be used to obtain a
diffusion coefficient by fitting GT(
) to the
measured G(
), using the diffusion coefficient as a
fitting parameter. The theoretical GT(
) is
normalized such that G(
0) = 1. The two-photon
autocorrelation was calculated using the square of the illumination
profile for Oijk and
O'ijk, and using the same
Cijk as in the one-photon case. Because there is
usually no detector aperture in two-photon FCS, the illumination profile is the only factor determining the profile of the observation volume.
Experimental methods
Autocorrelation curves were measured with two different systems.
One-photon FCS with aperture
A Confocor (Carl Zeiss, Jena, Germany) FCS
instrument used the 488 nm line of a 20 mW argon ion laser (Zeiss),
focused by a Zeiss C-Apochromat infinity-corrected 1.2 NA 40× water
objective (
~ 1) and either the standard 8-well plastic
sample holders designed for use with the system or a deep-well slide
covered with a no. 11/2 coverslip. The fluorophore was rhodamine
green (D-6107, Molecular Probes, Eugene, OR) of known concentration
between 1 and 300 nM. Fluorescence was collected through the same
objective and directed through FITC filters (480/10 excitation filter,
510 LP dichroic, 540/40 emission filter) and a variable-diameter
aperture (25-200 µm) onto an avalanche photodiode (EG&G, Salem, MA,
provided by Zeiss). The measured overall magnification was 81. The
fluorescence signal is digitized and autocorrelation curves thereby
calculated by a PC with onboard correlator card (ALV Laser, Hamburg,
Germany). Software provided with the Confocor was used to do numerical
fits to the data and obtain diffusion time (
d), number
of molecules N, average intensity
F
, and
count rate per molecule
. Multiple correlation curves were obtained
for each detector aperture rd on different days.
Measured parameters were corrected for measured background from a
distilled water blank under the same conditions. Unattenuated laser
power at the back-aperture of the objective was 12 mW, but the
intensity was always further reduced by neutral density filters to
typically 10-100 µW. Measurements of the number of molecules were
done at the highest concentrations (~100-300 nM) to eliminate
dye-glass sticking artifacts and to enable measurement of the
absorbance of the sample accurately.
Two-photon FCS
A ~100 fs pulsed-IR Ti:sapphire laser
(Tsunami, Spectra-Physics, Palo Alto, CA) pumped by an 8 W argon ion
laser (Beamlok, Spectra-Physics) mode-locked at 980 nm was guided into
an IM35 microscope (Zeiss, Jena, Germany) and reflected by an
appropriate dichroic into the back-aperture of a 1.2 NA 60× objective
(Zeiss). Fluorescence was passed by the same dichroic and focused by
the tube lens into a multimode optical fiber (200 µm ~ 52 ou
nominal diameter). The fiber carried the fluorescence to the detector, an avalanche photodiode (model SPCM-AQ-141-FC, EG&G) whose active area
was large enough not to restrict the observation volume in the absence
of light scattering (Schwille et al., 1999a
). The TTL pulses from the
detector are autocorrelated in a PC by an autocorrelator card
(ALV/5000, ALV Laser). Typically, 20 runs of 30-60 s each were
averaged. FCS on rhodamine green or Alexa488 (Molecular Probes) at
~10 nM concentration in a deep-well slide with a no. 11/2
coverslip was performed for 20 scans of 60 s each, per intensity.
Power at the sample was 5-10 mW, while the back-aperture of the
objective was underfilled (
~ 3) or overfilled (
< 1). One-photon FCS results using the second setup, 488 nm illumination from an argon ion laser, and various fiber diameters agreed with the
results from the first setup.
The concentration of rhodamine green was measured by absorbance in a
spectrophotometer (HP 8451A, Hewlett Packard, Palo Alto, CA) in either
a 3 × 3 mm or 10 × 10 mm quartz cuvette. Between 5 and 15 independent wavelength scans were averaged. We used the peak absorbance
(at
~ 490 nm) subtracting the baseline at long wavelength
(
~ 650 nm) in the equation C = A/
L (with
= 5.8 × 104 M
1 cm
1,
A = absorbance, L = 0.3 or 1.0 cm) to
calculate concentration.
 |
RESULTS AND DISCUSSION |
Calculated illumination and observation volume profiles
are non-Gaussian and will lead to artifacts
We delineate the effects of the observation volume profile
O(rO) on the FCS autocorrelation function and
the resulting consequences on the measured dynamics and analytical FCS
parameters. Here we will present calculated illumination and
observation profiles for a high-NA objective, as a function of the
back-aperture underfilling (
) and the confocal detector aperture
(for observation profiles only), first as two-dimensional pseudocolor
intensity bitmaps, then as plots with attempted fits to analytic
functions. It will be shown that the observation profile is predicted
to be non-Gaussian under many experimental conditions. The
detector-aperture dependence of the count rate per molecule, which is
directly related to the signal-to-noise ratio, provides direct evidence
for the non-Gaussian nature of the observation volume.
Next, the effect of a non-Gaussian observation volume on the FCS
autocorrelation function is presented, demonstrating that inaccuracy or
artifacts frequently arise as a result of using the standard fitting
function (i.e., Eq. 7). The form and magnitude of these artifacts are
then explored; their effects on the measured diffusion time, diffusion
coefficient, axial ratio, and exponential fraction(s) and time
constant(s) are shown. Methods for avoiding artifacts are outlined.
Finally, the collected fluorescence and volume of observation (number
of molecules at a known concentration) are analyzed as a function of
and rd and provide evidence for an actual
observation volume that is larger than predicted.
Calculated illumination and observation volume profiles
It is known that the intensity profile at the focus of an
underfilled, thin, low-NA lens is a Gaussian in the radial direction and a Lorentzian in the axial direction (Self, 1983
). FCS typically uses a high-NA objective lens for intense illumination and efficient fluorescence collection, and the illumination and observation profiles
are typically assumed to be 3D-ellipsoidal Gaussians. However, it will
be shown that this assumption is not sufficiently accurate for precise
analysis under many common measurement conditions.
Fig. 1 visually depicts FCS observation
profiles as logarithmically scaled pseudocolor bitmaps, with various
optical parameters. Here one can readily see evidence for the
non-Gaussian profiles and recognize the dependence of the observation
volume on the key experimental variables
, objective underfilling
fraction, and rd, confocal detector aperture
radius. Most importantly, the observation profile for an overfilled
back-aperture (
1), commonly used in FCS (top row),
has visible oscillatory aperture-limited diffraction fringes,
especially at large rd, where the observation profile is unconstrained and approximately equal to the illumination profile (left side). These fringes clearly indicate
non-Gaussian profiles, as a Gaussian would decay monotonically to zero
without oscillation. Second, closing the detector aperture (moving from right to left in the figure) constrains the volume significantly, reduces fluorescence collection from regions far from the focus (i.e.,
the fringes), and as will be shown, results in a more nearly Gaussian
observation profile. Third, when the objective is underfilled (
= 1.25; second row) and severely underfilled (
= 4;
third row), the illumination profile and observation volume
are elongated in the axial direction, and somewhat in the radial
direction, but this effect is significant only for
> 1.25. Underfilling also reduces or eliminates the diffraction fringes and
results in a smoother, more nearly Gaussian observation volume (shown in the bottom row).

View larger version (42K):
[in this window]
[in a new window]
|
FIGURE 1
Observation profile images in log-scale pseudocolor,
for a 1.2 NA objective, plotted for various values of the detector
aperture rd and underfilling fraction .
Images show the axial direction (u) horizontally and the
radial direction (v) vertically away from the focus. The
logarithmic color scale emphasizes the low-amplitude fringes away from
the focus. Closing the detector aperture (from left to
right) constrains the observation volume, reducing fluorescence
collection from the tails of the illumination profile that are
particularly periodic and non-Gaussian for an overfilled back aperture
(top row). Underfilling the back aperture (from top to
bottom) elongates and smoothes the illumination profile and hence
the observation volume, but the effect is not drastic until > 1.25. Using small rd and an underfilled back
aperture ( > 1) results in a more nearly Gaussian observation
volume. An exactly Gaussian profile is shown (bottom row
left) with a 20 ou scale bar (center) and intensity
color scale (right) for reference.
|
|
To be more quantitative, we now move from the images to plots of the
corresponding profiles of the illumination and observation volume along
the axial and radial directions, which will show functional dependence.
First we consider the functional form of the axial illumination
profiles, which corresponds approximately to the limiting case of the
observation profile with very large rd and
overfilled back-aperture (top left image of Fig. 1). Figs. 2 and
3 show axial and radial plots,
respectively, of the illumination profile for an overfilled 1.2 NA
objective, which indicates that the functional form is neither Gaussian
nor Lorentzian in the axial direction. Plotting {
Ln[O(u,
0)/O(0, 0)]}0.5 emphasizes the deviation from
Gaussian because an actual Gaussian, A
exp(
Bu2), plotted this way is
{
Ln[Ae
Bu2/A]}0.5 = {Bu2}0.5 = u(B0.5) which is linear in u. In general,
the Gaussian fits well at short distances from the focus, but decays
too quickly at large distances. The deviation begins at distances near
or slightly less than the 1/e2 waist and worsens at larger
distances. The overfilled radial illumination profile is also neither a
Lorentzian nor a Gaussian beyond the 1/e2 waist (see
oscillations in Fig. 1, top left).

View larger version (19K):
[in this window]
[in a new window]
|
FIGURE 2
Using conventional analytical functions to fit
calculated axial illumination profiles is ineffective. The calculated
profile for an overfilled 1.2 NA water objective (solid
line) is fit with a Lorentzian (dashed line) and
Gaussian (dotted line) on a linear (A) and
logarithmic (B) intensity scale.
|
|

View larger version (18K):
[in this window]
[in a new window]
|
FIGURE 3
Deviation of calculated radial illumination profile
from a Gaussian profile. The calculated illumination profile for an
overfilled 1.2 NA water objective is shown along the radial axis
(A and D), with a linear intensity scale
(top) and a nonlinear scale (bottom),
[ log(intensity)]0.5, which should yield a straight line
for all positions if the profile is Gaussian. Note that the profile
begins to deviate from Gaussian (B and C,
red dashed line) at less than the 1/e2 radial
width (dotted black line).
|
|
Because the observation volume is a function of the illumination
profile (Eq. 3), it is not surprising that the observation profile is
also non-Gaussian, as it is shown plotted with attempted Gaussian fits
along the radial direction (see Fig. 4,
left) and axial direction (see Fig. 4, right).
Again, the Gaussian is only a reasonable approximation very close to
the focus. Furthermore, because the fringes are at such large distances
from the focus, they occupy a significant volume due to the approximate
axial symmetry of the system, and can have a strong influence on the measured volume and autocorrelation in FCS. Therefore, in many circumstances, no simple analytic function will be able to describe the
FCS observation volume accurately.

View larger version (20K):
[in this window]
[in a new window]
|
FIGURE 4
Attempt to fit calculated observation profiles for an
overfilled 1.2 NA water objective with a 4.7 ou detector aperture
(corresponding to the top center panel in Fig. 1) using analytical
functions. Left: The radial profile O(r) = O(u, ) = O(0, ) is shown on a linear scale
(A) with Gaussian fit (B), and on a nonlinear
scale (C) with the same fit (D).
Right: The axial profile O(r) = O(u,
) = O(u, 0), (top right,
E) and its Gaussian fit (F), is also shown with
nonlinear vertical axis (G) and fit (H). The
nonlinear axis transforms any Gaussian function into a straight line.
|
|
An experimental quantity derived directly from the calculated
observation volumes is the count rate per molecule (
), which is
directly related to the S/N in FCS. While a typical FCS fit using
N and
D as free parameters does give
information about the observation volume, in the absence of
fluorescence saturation and photobleaching the aperture-dependence of
provides a direct experimental test of whether
O(rO) is Gaussian (see also Eq. 25), because a
Gaussian observation volume will not produce a peak in
as a
function of detector aperture rd. Hence,
important physical properties of the observation volume can be
extracted from the dependence of
on rd.
Measurements confirm observation volume is non-Gaussian
Peak in count rate per molecule versus detector aperture
An important result is the dependence of the count rate per
molecule (
) on confocal detector aperture size
(rd). A peak in
is observed, which
demonstrates that 1) there is an optimal (maximal S/N) set of
conditions for FCS, and 2) that the observation volume must be
non-Gaussian.
Fig. 5 shows the strong maxima of the
calculated and measured
(rd), with
diffraction theoretical curves shown for overfilled and underfilled
back-aperture and a theory curve assuming a Gaussian observation
volume. The curves are normalized such that their peak value is unity.
The best fit of the aperture dependence is for the overfilled
back-aperture, and the peak position does not change significantly for
< 1.6. The peak does shift to slightly larger values when the
objective is underfilled, because as the focal volume is made larger,
the pinhole that restricts the observation volume enough to optimize
will also be larger.

View larger version (23K):
[in this window]
[in a new window]
|
FIGURE 5
Experimental evidence for a non-Gaussian observation
volume. Measurements and diffraction-theory-calculated count rate per
molecule ( ) versus detector aperture (rd) for
a 1.2 NA 40× water objective as a function of underfilling fraction
( ). A Gaussian observation profile predicts no maximum in
(rd) at finite rd
(thin green curve).
|
|
Previous studies on confocal microscopy also discuss optimization of
signal to noise (Webb et al., 1990
; Sandison and Webb, 1994
; Sandison
et al., 1995
). However, signal-to-noise optimization in confocal FCS is
different than in confocal microscopy because FCS has a different
optimization criterion. The best detector aperture for FCS is different
from the best aperture size for maximal S/N in a confocal microscope
(using a paraxial diffraction theory) of 2.4-3.3 ou for an overfilled
objective (Sandison and Webb, 1994
). This difference is expected
considering that the above S/N optimization for FCS does not explicitly
consider the effects of fluorescence background, and is therefore
defined differently. A 3D-ellipsoidal Gaussian observation volume
predicts no peak in
(rd); however, a peak is
observed in both our experimental and diffraction theory results,
evidence that a Gaussian observation volume is not consistent with the
measured results.
The peak in count rate per molecule (at rd ~ 4.5 ou for a 40 × 1.2 NA water immersion objective) also
signifies that there is, under many typical measurement conditions, an
optimal detector aperture that maximizes
, and hence S/N.
Underfilling reduces the maximum
by a factor of ~2 for
= 2.2, but does not change the location of the peak significantly for
< 1.6. However, as will be shown, using an overfilled
back-aperture and the detector aperture that gives optimal S/N will
result in a non-Gaussian observation volume and artifacts in the
autocorrelation. Therefore, great care must be taken by the FCS user
who requires the smallest possible observation volume, optimal S/N, and
a Gaussian observation profile.
Underfilling the objective decreases the peak value of the count rate
per molecule. Fig. 6 shows calculated
versus detector aperture for different underfilling fractions, this
time keeping the integrated rate of excitation in the x-y
plane constant, i.e., constant illumination power. Note the greatly
reduced magnitude of
with underfilling, due to decreased intensity
at the focus. The effect of underfilling becomes particularly
pronounced for
> 1.6. The peak becomes less pronounced,
indicating that the volume is more nearly Gaussian for an underfilled
back-aperture. The effect of rd on
is
negligible for rd > 7 ou when
> 4.

View larger version (27K):
[in this window]
[in a new window]
|
FIGURE 6
Underfilling reduces count rate per molecule. The
effect of underfilling fraction ( ) and detector radius
rd on count rate per molecule ( ) for a 1.2 NA
40× water objective. The curves are normalized such that the total
rate of excitations in the x-y (focal) plane is constant.
|
|
The detector-aperture-dependence of
shows clear experimental
evidence of a non-Gaussian observation volume (see Eq. 25). A
non-Gaussian illumination profile such as shown in Fig. 1 with an
overfilled detector aperture will have fringes far from the focus,
which contribute significantly to the volume when the detector aperture
is large, increasing the number of weakly fluorescent molecules and
reducing the average fluorescence per molecule and the value of
.
The peak in
is most pronounced for an overfilled back-aperture
where the observation volume is most non-Gaussian due to the
diffraction fringes. At larger values of
, where the objective is
underfilled, the peak is less pronounced or nonexistent (see Fig. 6).
Because the measured FCS diffusion autocorrelation is a function of the
observation profile (Eq. 6), deviations from Gaussian behavior in the
observation profile will result in deviations from the analytic form
for the diffusion autocorrelation (Eq. 7), which assumes Gaussian
behavior. Consequently, analysis of the effects of these deviations on
the autocorrelation was performed by simulation of the autocorrelation
function using calculated non-Gaussian observation profiles (see Fig.
7). Simulated autocorrelation functions
also offer a means to measure absolute values of the diffusion
coefficient of a molecule.

View larger version (29K):
[in this window]
[in a new window]
|
FIGURE 7
Comparison of measured and calculated autocorrelation
versus detector aperture. (Top) Measured autocorrelation
(points) and diffraction theory fits (lines) for
1P-FCS (left column) with detector aperture diameter
(A) 2.4 ou, (B) 3.9 ou, (C) 5.8 ou,
and for 2P-FCS (right column) with no detector aperture
(D). Residuals are shown below for fits with calculated
autocorrelation (middle row) and with Eq. 7 (Gaussian
volume; bottom row). Residual curves (E-G)
correspond to measured data in A-C, respectively. Residuals
for 2P-FCS are shown on the middle and bottom right plots. The number
of molecules and diffusion coefficient were used as free parameters for
the theoretical curve-fitting.
|
|
Simulation of the autocorrelation function using diffraction-based
observation profiles
Comparison of calculations with measured autocorrelation and
diffusion coefficients
This section demonstrates the degree of agreement between
calculations using the predictions of diffraction theory and measured FCS results, including 1) autocorrelation functions at high NA and 2)
diffusion coefficients, which are a measure of similarity of the
experimental and calculated observation volumes. The measured autocorrelation of rhodamine green (aperture setup) is fit using the
calculated autocorrelation for a 1.2 NA 40× water objective, excitation wavelength
x = 488 nm. Fig. 7 shows the
measured autocorrelation, fit using the diffraction-calculated
autocor