| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Biophys J, December 2002, p. 3177-3187, Vol. 83, No. 6

and
*Laboratory for Fluorescence Dynamics, Department of
Biochemistry, University of Illinois in Urbana-Champaign,
Urbana, Illinois 61801 USA
| |
ABSTRACT |
|---|
|
|
|---|
Adenylate kinase (AK) is a ubiquitous enzyme that
regulates the homeostasis of adenine nucleotides in the cell. AK1
(long form) from murine cells shares the same protein sequence as AK1 (short form) except for the addition of 18 amino acid residues at its
N-terminus. It is hypothesized that these residues serve as a signal
for protein lipid modification and targeting of the protein to the
plasma membrane. To better understand the cellular function of these AK
isoforms, we have used several modern fluorescence techniques to
characterize these two isoforms of AK enzyme. We fused cytosolic
adenylate kinase (AK1) and its isoform (AK1
) with enhanced green
fluorescence protein (EGFP) and expressed the chimera proteins in HeLa
cells. Using two-photon excitation scanning fluorescence imaging, we
were able to directly visualize the localization of AK1-EGFP and
AK1
-EGFP in live cells. AK1
-EGFP mainly localized on the plasma
membrane, whereas AK1-EGFP distributed throughout the cell except for
trace amounts in the nuclear membrane and some vesicles. We performed
fluorescence correlation spectroscopy measurements and photon-counting
histogram analysis in specific domains of live cells. For AK1-EGFP, we
observed only one diffusion component in the cytoplasm. For
AK1
-EGFP, we observed two distinct diffusion components on the
plasma membrane. One corresponded to the free diffusing protein,
whereas the other represented the membrane-bound AK1
-EGFP. The
diffusion rate of AK1-EGFP was slowed by a factor of 1.8 with respect
to that of EGFP, which was 50% more than what we would expect for a
free diffusing AK1-EGFP. To rule out the possibility of oligomer
formation, we performed photon-counting histogram analysis to direct
analyze the brightness difference between AK1-EGFP and EGFP. From our
analysis, we concluded that cytoplasmic AK1-EGFP is monomeric.
fluorescence correlation spectroscopy proved to be a powerful technique
for quantitatively studying the mobility of the target protein in live
cells. This technology offers advantages in studying protein
interactions and function in the cell.
| |
INTRODUCTION |
|---|
|
|
|---|
Adenylate kinase (AK) is a ubiquitous monomeric
enzyme that catalyzes the following reaction:
Mg2+-ATP + AMP
Mg2+-ADP
+ ADP (Noda, 1973
). The cell uses this reaction to convert
AMP to ADP (Glaser et al., 1975
), thereby regulating adenine nucleotide
levels. AK is also involved in other reactions, such as the
biosynthesis of phospholipids (Goelz and Cronan, 1982
). In vertebrates,
three isozymes of AK have been characterized: AK1 is cytoplasmic, AK2
is localized in the intermembrane space of mitochondria, and AK3 is
localized in the mitochondrial matrix. Recently, a long form of AK1
protein from murine cells, AK1
, was identified. Its amino acid
sequence was identical to that of the murine cytoplasmic AK1, except
for the addition of 18 amino acids (MGCCVSSEPQEEGGRKTG) at the
N-terminus. The transcription level of the AK1
gene was found to be
upregulated by p53, whereas that of the AK1 gene was not. This
observation suggested that AK1
might have some novel biological
functions other than its conventional function of regulating nucleotide
levels. For example, it may be involved in the cell cycle arrest
process (Collavin et al., 1999
). We are interested in the effect of the
additional 18 amino acid residues on the biological function of AK1
.
It was hypothesized that the 18 additional residues might serve as a
signal for protein lipid modification and targeting of the protein to
the plasma membrane, based on the N-myristoylation consensus motif.
Protein N-myristoylation is the result of the co-translational addition
of myristic acid to a Gly residue at the extreme N-terminus after
removal of the initiating Met (Utsumi et al., 2001
). In general, to
direct protein N-myristoylation, the N-terminal consensus motifs such
as Met-Gly-X-X-X Ser-X-X-X (Johnson et al., 1994
) or Met-Gly-X-X-X
Thr-X-X-X (Boutin, 1997
) are preferred.
Fluorescence correlation spectroscopy (FCS) was first introduced
by Webb and his coworkers in 1972 (Magde et al., 1972
). It has evolved
as a powerful method to study particle dynamics on a single-molecule
level in part because of recent technological advances (such as
confocal microscopy, multi-photon laser excitation). FCS uses the
autocorrelation function to characterize fluorescence intensity
fluctuations in the observation volume and it can be used to study
particle diffusion (Fahey et al., 1977
; Koppel et al., 1976
), chemical
kinetics (Haupts et al., 1998
; Starr and Thompson, 2001
), and molecular
aggregation (Palmer and Thompson, 1987
; Qian and Elson, 1990
). In
recent years, several laboratories have introduced FCS studies on live
cells. Fluorescently labeled particles were first loaded into
cells or labeled on the surface of the cells; the dynamics of the
fluorescent dyes in the cells were then studied with FCS. Successful
application of FCS in the cell provides a new experimental approach to
quantitatively study the diffusion rates of cellular molecules or
components (Berland et al., 1995
; Schwille et al., 1999a
; Brock et al.,
1998
; Politz et al., 1998
). With the discovery of green fluorescence
protein (GFP) and the development of molecular engineering, a protein of interest can be easily fused to GFP and expressed in the cell. The
dynamics of the target protein can then be followed with FCS (Wachsmuth
et al., 2000
; Brock et al., 1999
; Hink et al., 2000
). GFP is a 27-kDa
monomer first found in the jellyfish Aequorea (Shimomura et
al., 1962
). The gene for GFP has been isolated. It has become a useful
tool for making chimeric proteins, where GFP acting as a fluorescent
reporter is linked to proteins of interest. GFP tolerates N- and
C-terminal fusion to a broad variety of proteins. The fusion proteins,
which are fluorescent, usually maintain normal function and cellular
localization of the host protein (Tsien, 1998
). Enhanced GFP (EGFP) is
a GFP mutant with improved optical properties and folding stability.
Careful investigation of the spectral properties of EGFP (brightness,
resistance to bleaching, and long wavelength emission) suggests it is a
superior molecular indicator for quantitative fluorescence microscopy
studies of intra- and intercellularly tagged-proteins (Cinelli et al., 2000
).
In our system, we constructed AK1-EGFP and AK1
-EGFP chimera
proteins and then expressed them in HeLa cells. Two-photon excitation scanning fluorescence images of the transfected cells directly revealed
the localization of the chimera proteins in the cells. AK1
-EGFP
mainly localized on the plasma membrane, whereas AK1-EGFP was
distributed throughout the cell except for the nuclear membrane and
some vesicles. We demonstrated the power of fluorescence imaging in
differentiating two structurally and functionally similar proteins in
the cells. The mobility of AK1 and AK1
in live cells was quantified by FCS. We discussed in detail how to carry out FCS measurements in
live cells and our findings about the biological properties of the two
intracellular isoforms of AK. FCS can provide important information
about the microenvironment of the target protein and reveal any
possible protein-protein or other cellular interactions. In addition to
the FCS measurements, we also performed photon-counting histogram (PCH)
analysis on the FCS data, which provided molecular brightness
information and allowed for direct resolution of molecular heterogeneity (e.g., monomer and dimer mixtures). To characterize the
effect of the fusion protein on the optical properties of EGFP, we also
measured the fluorescence lifetime of AK1-EGFP and AK1
-EGFP in
transfected cells and compared then with EGFP-transfected cells.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Cell culture
HeLa cells obtained from American Type Culture Collection (Rockville, MD) were grown in a 5% CO2 humidified atmosphere at 37°C in Dulbecco's modified Eagle's medium (Gibco, Gaithersburg, MD) 4.5 g/L glucose, 2 mM L-glutamine, without phenol red, 25 mM Hepes (pH 7.4),10% fetal bovine serum, 100 U/ml penicillin, and 0.1 mg/ml streptomycin (Sigma Chemical Co., St. Louis, MO).
Generation of EGFP fusion proteins and transient transfected cells
The AK1 gene fragment obtained from Dr. Schneider's group
(Universidad Simon Bolivar) was inserted into the pEGFP-N1 plasmid (Clontech, Palo Alto, CA) between the EcoR1 and
BamH1 restriction sites. The stop codon for the AK1 gene
(TAA) was mutated to GAA using the Quick-Change Strategene mutagenesis
kit (Stratagene, La Jolla, CA). There was a seven-amino-acid-residue
linker (LDPPVAT) between AK1 and EGFP. The final plasmid construct
(AK1-pEGFP-N1) was sequenced to ensure that AK1 was in frame with EGFP
and there was no random mutation in the chimera protein. For transient
transfection, HeLa cells were seeded in an eight-chambered cover glass
system (Nagle Nunc International, Rochester, NY) at a confluency
of 20-30%. The next day, HeLa cells were transfected with the
AK1-pEGFP-N1 plasmid DNA using the Effectene transfection kit (Qiagen,
Chatsworth, CA). At 24 h after transfection, the cells were used
for fluorescence microscopy measurements. For FCS measurements, the
AK1-pEGFP-N1 plasmid was linearized with the AflII
restriction enzyme to reduce the expression level of AK1-EGFP. An
activity assay for AK (Huss and Glaser, 1983
) was performed on the
AK1-EGFP-transfected HeLa cell lysate. It confirmed that the chimera
protein was enzymatically active (data not shown).
The AK1
gene fragment was cloned into the pEGFP-N1 plasmid and
transfected into HeLa cells in the same manner. As for FCS measurements, it was not necessary to linearize the plasmid. The low
expression level of this chimeric protein in the cell was appropriate
for FCS measurements.
Two-photon excitation fluorescence microscopy instrumentation
The two-photon excitation scanning fluorescence microscope used in the experiments was assembled in our laboratory. The optical microscope was an Axiovert 100 inverted microscope (Zeiss, Oberkochen, Germany). The objective was a Zeiss Plan Neofluar 63× (1.25 N.A., oil). A mode-locked titanium-sapphire laser with 80-MHz, 100-fs pulse width (Tsunami, Spectra-Physics, Mountain View, CA) was used as the excitation light source. The laser was guided by an x-y galvano-scanner (model 6350; Cambridge Technology, Cambridge, MA) to achieve beam scanning in both x and y directions. A photomultiplier tube (Hamamatsu HC120-08, Somerville, NJ) was used for light detection in the photon-counting mode. A BG39 optical filter was placed before the photomultiplier for efficient suppression of IR excitation light. Data were processed and analyzed with software developed in our laboratory (SimFCS and Globals, Window Edition). At any designated area in the cell, fluorescence intensity imaging and FCS measurements can be acquired with the above-described instrumentation.
Two-photon excitation scanning fluorescence imaging
The excitation wavelength used in our study was 915 nm, where
the two-photon action cross section is maximized for EGFP and the
cellular autofluorescence is relatively low (Xu et al., 1996
). The
fluorescence images of transfected cells were scanned at the rate of 50 µs/pixel. Each frame had 256 × 256 pixels, and each image was
integrated for 10 frames.
FCS and PCH analysis
FCS is based on the principle that, when fluorescent
molecules are in equilibrium, there are always spontaneous microscopic fluctuations of this equilibrium in a small observation volume (sub-femtoliter). The fluorescence fluctuations can originate from the
diffusion of fluorescent molecules through the observation volume or
changes of the fluorescence quantum yield of the fluorophore due to
chemical reaction. Depending on the origin of the fluctuation, the
fluctuation contains information about the diffusion rate of the
molecule or the rate of chemical reaction (for an overview, theory, and
applications of FCS, see Rigler and Elson, 2001
). In a system where
particles undergo only translational diffusion, the analysis of these
fluctuations allows the determination of their diffusion coefficient
(D) and the average number of particles (N) in
the excitation volume (Palmer and Thompson, 1987
). Any changes in the
diffusion coefficient of a particle in solution would reflect changes
of its size (or shape) or the viscosity of the solution. Any changes in
the number of particles in the excitation volume would suggest particle
association or dissociation. Because the relative amplitude of the
fluctuations is inversely proportional to the number of molecules
simultaneously observed in the excitation volume, FCS measurements are
usually carried out at nanomolar sample concentrations, where the
fluctuations are observable above the average fluorescence signal.
From these spontaneous microscopic fluctuation measured directly by FCS
experiments, an autocorrelation curve can be calculated using the
autocorrelation function G(
). This function reflects the
time-dependent decay of the fluorescence intensity fluctuation. The
normalized autocorrelation function G(
) is defined as:
|
F(t) = F(t)
F
expresses the
fluctuation in fluorescence intensity at time t.
Experimental autocorrelation functions G(
) were fitted to
theoretical functions using a Gaussian-Lorentzian beam profile (Berland
et al., 1995
) to recover the diffusion coefficient and the number of
molecules in the excitation volume.
FCS measurements in our experiments were carried out on many
compartments of the cell (nucleus, cytoplasm, plasma membrane, etc.)
and in many different cells. The sampling frequency was 20 kHz, and
each measurement lasted less than 3 min. The laser power was 1 mW at
the sample. No photobleaching was observed for cytosolic measurements,
and only a small amount was observed for the membrane. The waist
(
0) of the excitation beam was calibrated with
EGFP in solution (D = 87 µm2/s)
(Swaminathan et al., 1997
) before each day's measurements. The typical
values of
0 are at the range of 0.3-0.35
µm.
Although the temporal behavior of fluctuation can be analyzed by
the autocorrelation function, the amplitude of the fluctuation is
characterized by its probability distribution (PCH). PCH is experimentally determined by the histogram of the detected photons per
sampling time for a fluorescence fluctuation experiment. The theoretical basis of PCH was described by Chen et al. (1999)
. Two
parameters, the molecular brightness
(in counts per second per
molecule, cpsm) and the average number of molecules per observation volume (N) can be extracted from the PCH analysis based on a
theoretical distribution function
(k; 
). PCH analysis is advantageous for detecting the presence of
molecule dimerization or aggregation based on the brightness of the
molecule when the changes in diffusion rate were too small to be
conclusive. For example, the brightness of a dimer will be twice as
bright as a monomer, whereas the diffusion rate of a dimer will be only
25% slower than that of a monomer. PCH analysis can resolve multiple
components, if their brightness contrast is sufficient (Muller et al.,
2000
).
Two-photon excitation fluorescence lifetime imaging
The fluorescence lifetime measurements were performed on a frequency domain instrument assembled in our laboratory, which is integral to the two-photon excitation microscope described above. Our instrument operates at a high cross-correlation frequency that rapidly provides lifetime information on a per pixel basis. A gain-modulated photo multiplier tube (PMT) detects the fluorescence signal from each position. A second gain-modulated PMT monitors the laser beam to correct for frequency and modulation drift of the laser pulse train. The analog outputs of the two PMTs are digitized by the data acquisition computer. The excitation light was modulated at 80 MHz, and the cross-correlation frequency was set at 2500 Hz. Data were acquired at eight points per cross-correlation period. The scanner was running at the rate of 400 µs/pixel. The lifetime reference was fluorescein in 0.1 M NaOH (4.05 ns).
| |
RESULTS |
|---|
|
|
|---|
Two-photon excitation scanning fluorescence imaging of transfected cells
The fluorescence of the tagged EGFP was directly visualized in
live HeLa cells with two-photon excitation scanning fluorescence microscope. Fig. 1 A shows
fluorescence images of several representative AK1-EGFP-transfected
cells. The AK1-EGFP protein was widely distributed throughout the cell,
except for the nuclear membrane and some vesicular structures. The
expression levels of AK1-EGFP vary from one cell to another under the
same transfection conditions. For example, there was at least a twofold
fluorescence intensity difference between the two adjacent cells in the
first image of Fig. 1 A. The fluorescence intensity of the
next image in Fig. 1 A was so high that it cannot be
displayed on the same fluorescence scale as the first image. Fig. 1
B shows fluorescence images of several typical
AK1
-EGFP-transfected cells. This fusion protein was mainly localized
on the plasma membrane. In some cells, fluorescence from AK1
-EGFP
was not homogeneously distributed on the plasma membrane. AK1
-EGFP
was also found in the cytoplasm and the nucleus of some cells, when the
overall protein expression level was high. Even though AK1
-EGFP can
be expressed in HeLa cells and the protein expression levels varied
significantly from cell to cell, the average expression level of
AK1
-EGFP was much lower than that of AK1-EGFP. As a control
experiment, the EGFP plasmid was transfected into HeLa cells. Its
corresponding fluorescence image is shown in Fig. 1 C.
|
FCS
FCS measurements were carried out on healthy transfected cells
with low protein expression levels. The cells shown in Fig. 1 (except
panel B) can only be used for fluorescence
imaging, because the average fluorescence is too high to resolve
fluctuation at a single-molecule level. For the FCS experiments, the
protein expression level must be relatively low, which can be achieved by linearizing the plasmid and optimizing the plasmid concentration. However, within the same batch of transfection, the fluorescence intensity of transfected cells still ranged from 5000 cps to millions of cps. Only cells with fluorescence intensity below 40,000 cps were
selected for FCS measurements. The fluorescence intensity of
nontransfected cells ranged from 300 to 1000 cps. Therefore, the
autofluorescence of the cell was usually less than 10% of the total
fluorescence intensity measured in the transfected cell. Fig.
2 shows some examples of typical
autocorrelation curves (displayed as
,
,
, *, and ×)
calculated from our FCS measurements. G(
) was normalized
to unity for ease of comparison. The lines running through the data
points
,
,
, *, and × represent the fitted autocorrelation curves. The calculated autocorrelation curves (displayed as
and
) for AK1-EGFP and AK1
-EGFP were almost identical when measured inside the cell (cytosol). Both curves could be
fit with a single-component model. The fitted curves are
superimposable. The recovered diffusion coefficient (D) of cytosolic AK1-EGFP and AK1
-EGFP was 13 µm2/s. The average residence time (
=
02/8D, for
two-photon excitation) of the chimeric proteins inside the excitation
volume was ~0.9 ms. When the excitation laser beam was focused on the
plasma membrane of the cell, the autocorrelation curve of AK1-EGFP was
similar to those measured in the cytoplasm. By contrast, the
autocorrelation curve of AK1
-EGFP (×) measured in the membrane
could be fitted only with a two-component model. The inset of Fig. 2
shows the two resolved components. The diffusion coefficient of the
fast component was similar to that of AK1
-EGFP in the cytoplasm,
whereas the second component was ~60 times smaller (D = 0.23 µm2/s). The ratio of these two
components depends on the position of the focus in the Z
direction. As a control experiment, the diffusion rates of EGFP in live
cells were also measured. The
in Fig. 2 represents the
autocorrelation curve of EGFP expressed in HeLa cells. The diffusion
coefficient of EGFP in the cell was approximately three times smaller
than that of EGFP (*) in solution (D = 87 µm2/s). The average residence time of cytosolic
EGFP in the excitation volume was 0.5 ms.
|
Live cells are complex organisms. The cellular structure, fluidity, and
protein function may vary from region to region, cycle to cycle, and
certainly from cell to cell. It is not surprising that the measured
diffusion rates of our proteins varied among cells and different
regions of the same cell. A useful way to represent the mobility of the
protein in the cell would be to construct a spatial map of diffusion
rates. Fig. 3 shows an example of such a
diffusion rate map. As shown in the image, multiple points along two
randomly selected lines were measured. The image, which is 15 µm in
size, displayed only one-third of a cell in its
X-Y plane. The full image of the cell is shown in
Fig. 1 B (middle panel). The light
green line on the upper part of the image corresponds to the plasma
membrane. The cytoplasm of the cell is below the line. The numbers
listed in the columns next to the images corresponds to the diffusion
coefficients (µm2/s) measured at the indicated
points across the cell. All the autocorrelation curves were fitted with
a single-component model unless the two-component model yielded a lower
2 value. The standard deviation of each fit is
~10% of its diffusion coefficient. The first row in each column
represents diffusion coefficients for the two components recovered from
the autocorrelation curves. This result was expected because the
excitation beam was focused on the plasma membrane and surrounding
areas. The other points of measurement do not involve the plasma
membrane, and they can be fitted with a single-component
autocorrelation function. The diffusion coefficients of AK1
-EGFP in
the cytosol range from 7.9 to 16 µm2/s, which
strongly demonstrated that the dynamics of one kind of protein in the
same cell could vary from region to region even though the fluorescence
image appeared homogeneous. This result could be due to the differences
in organelle fluidity, obstruction in diffusion, protein-protein
interaction, or a temporal dependence to these parameters. A large
sampling quantity is necessary to study the overall mobility of the
chimeric proteins in the cell. Fig. 4
shows a composite histogram of the diffusion coefficients of the
chimera proteins measured from dozens of cells and hundreds of
different cell positions. The diffusion coefficient of AK1-EGFP and
AK1
-EGFP measured inside the cell shared almost the same distribution with a peak around 13 µm2/s except
that a small fraction of AK1
-EGFP had a slow component (~0.1-0.5
µm2/s, not plotted in the graph). The slow
component of AK1
-EGFP measured on the plasma membrane ranges from
0.1 to 0.5 µm2/s with a peak at 0.1 µm2/s. The average diffusion coefficients of
EGFP and its fusion protein are listed in Table
1. The standard deviation of the diffusion coefficient of EGFP (in solution and in the cell) was ~10%, whereas the standard deviation of the diffusion coefficient of
EGFP fusion proteins (in the cell) was 30-40%. The large standard deviation of EGFP fusion protein diffusion measured in the cell was not
due to experimental error, but rather it reflected the dynamic
complexity of AK1-EGFP and AK1
-EGFP in the cells.
|
|
|
PCH analysis
PCH analyses were performed on the same FCS data set. Because the
brightness of a molecule is quadratically proportional to the input
laser power on the sample, the laser power was kept constant (1 mW on
the sample) for easy comparison of the brightness of the various
proteins. All histograms could be fit with a single component. Fig.
5 shows an example of a PCH graph and the
fitting of AK1
-EGFP measured in the cytosol of a HeLa cell. The
molecular brightness of AK1-EGFP, AK1
-EGFP, and EGFP in HeLa cells
are listed in Table 2. The fusion
proteins have the same brightness, which is 3600 cpsm, although the
corresponding fluorescence intensity is significantly different because
the expression levels vary greatly. EGFP in the cell and in solution
has a brightness of 4200 cpsm. The average number of EGFP or EGFP
fusion protein molecules in the excitation volume ranged from 3 to 10 depending on the protein expression level in each particular cell.
Based on the observation volume, their corresponding protein
concentration was 30-100 nM.
|
|
Fluorescence lifetime imaging microscopy
The fluorescence lifetime of a fluorophore is a
concentration-independent parameter that can be used to detect
variations in the optical properties of a fluorophore or its
environment. The fluorescence lifetime images of the EGFP, AK1-EGFP,
and AK1
-EGFP in viable HeLa cells were measured and compared with
EGFP in solution. In Fig. 6, the first
column shows fluorescence intensity images of EGFP-, AK1-EGFP-, and
AK1
-EGFP-transfected cells, respectively, the second column presents
the corresponding fluorescence lifetime images, and the third column
presents the pixel histogram of the lifetime images. From Fig. 6, it is
clear that the lifetimes of EGFP, AK1-EGFP, and AK1
-EGFP were
homogeneous throughout the cell. The average fluorescence lifetime of
EGFP and its fusion protein in the cells are listed in Table 2. The
lifetime of EGFP in solution (2.8 ns) was also measured, and it was
longer than that in live cells (2.45 ns). The fluorescence intensity of
the nontransfected cells was too low to allow for a fluorescence
lifetime determination of the autofluorescence of the cell.
|
| |
DISCUSSION |
|---|
|
|
|---|
The main objective of this study was to characterize the
localization of AK1 and AK1
in live cells and to demonstrate the applicability of FCS to quantitatively measure the dynamics of the
target proteins in cellular environments.
Two-photon excitation is the simultaneous absorption of two
photons by a molecule that is normally excited by a single photon with
twice the energy (Friedrich, 1982
). Two-photon excitation offers
advantages over one-photon excitation in many aspects. For example, the
excitation light can be easily separated from the fluorescence, so that
only molecules at the microscope focus are excited, therefore reducing
photo-damage to the sample outside the excitation volume, and the
inherent optical sectioning effect is useful for fluorescence imaging
and other microscopy measurements. The initial demonstration of
two-photon FCS measurements in cells was reported by Berland et al.
(1995)
. They investigated the diffusion characteristics of
fluorescently labeled latex spheres inserted into fibroblast cells by
electroporation. In recent years, a few reports on FCS applications in
live cells have appeared (Gennerich and Schild, 2000
; Kohler et al.,
2000
; Schwille et al., 1999a
; Wachsmuth et al., 2000
). FCS measurements
are generally carried out at nanomolar or sub-micromolar sample
concentrations, because the relative amplitude of the fluctuations is
inversely proportional to the number of molecules simultaneously
observed in the excitation volume. The upper limit for sample
concentrations in FCS measurements is usually micromolar (Thompson,
1991
). Sample concentration can be easily adjusted in solution;
however, the cellular protein concentration is more difficult to
adjust. The expression level of a protein in the cell can be lowered by
decreasing the plasmid concentration and linearizing the plasmid.
Lowering the protein's expression level was also beneficial for
studying the biological function of the protein in the cell, because it
more closely mimics physiological conditions. Under optimal conditions,
the concentration of the fusion protein in the cells was in the range
of ~30-100 nM.
Expression and imaging of EGFP fusion protein
The fluorescence images of AK1-EGFP- and AK1
-EGFP-transfected
cells (Fig. 1) revealed that AK1-EGFP is distributed throughout the
cell except for some vesicles in the cytoplasm and the nuclear membrane. By contrast, AK1
-EGFP is mainly localized on the plasma membrane. Immunoblotting experiments carried out by Collavin et al.
(1999)
showed that AK1
was found in the plasma membrane fraction of
the cell lysate. Our direct visualization of AK1
-EGFP in the cell
confirmed their findings, which suggested that the AK1
protein was
indeed targeted to the plasma membrane. In addition, we found that on
average, the expression level of AK1-EGFP in the cell was much higher
than that of AK1
-EGFP (Fig. 1), although the protein expression
level varied from cell to cell. The cells with high AK1
-EGFP
concentrations usually appeared unhealthy; they tended to round up or
die. This observation suggests that AK1
might be toxic to the cells
at higher concentration, whereas AK1 was not. The toxicity suggests
that AK1
might have additional biological functions other than
converting AMP and ATP to ADP. After all, the transcription level of
the AK1
gene was shown to be upregulated by p53 (Collavin et al.,
1999
).
In the cytoplasm and nucleus of AK1
-EGFP-transfected cells,
measurable levels of fluorescence were detected. Its origin could be
from AK1
-EGFP or truncated EGFP. There have been reports showing that truncated GFP were found in transfected cells (Brock et al., 1999
). A conventional method to discern AK1
-EGFP in the cytoplasm would be to run a Western blot to determine the presence of the fusion
protein. The blot would require a specific antibody for the protein.
With the application of FCS, the question can be easily resolved.
FCS and PCH analysis on EGFP fusion proteins
The fluorescence observed in the cytosol of an
AK1
-EGFP-transfected cell could also arise from autofluorescence.
Reduced pyridine nucleotide, flavin, and lipofuscin are the known
fluorescent molecules, which contribute to the autofluorescence in
mammalian cells (Andersson et al., 1998
; Croce et al., 1999
). Although
the autofluorescence of the cell was low at 915-nm excitation, it was
still possible to confuse it with the fluorescence from very low
concentrations of AK1
-EGFP. FCS measurements carried out on
nontransfected HeLa cells showed that there was no amplitude in the
autocorrelation curve and low signal-to-noise ratio from the
autofluorescence of the cell, suggesting that the autofluorescence came
from many dim heterogeneous molecules. This observation was in
agreement with reports from other laboratories (Schwille et al., 1999b
;
Wachsmuth et al., 2000
). In the autocorrelation analysis, the
contribution of each molecule is proportional to the square of its
brightness. Because the brightness of the autofluorescence molecule is
dim compared with EGFP, autofluorescence will contribute very little to
the autocorrelation curve. FCS measurements carried out on the
transfected yet dimly fluorescent cell showed robust autocorrelation
curves with a high G(0), indicating that the low fluorescence intensity was due to EGFP fusion protein expressed at low
levels. The autocorrelation curves of cytosolic EGFP and AK1-EGFP can
be fitted with a single component; no anomalous diffusion was observed.
Table 1 summarizes the averaged diffusion coefficients of EGFP
and its fusion proteins in aqueous solution and in the cell. The
diffusion coefficient of EGFP in the cytoplasm was approximately 3 times smaller than in solution. The ratio of D in the cytoplasm relative to that in solution
(DCyt/DH2O) is 0.26, which
matches the Verkman group's report (Seksek et al., 1997
). The cell
lines and experimental conditions used to transfect EGFP and AK1-EGFP were the same. The size of AK1-EGFP is twice as large as EGFP. Since
the diffusion coefficient of a particle should be proportional to the
inverse cubic root relative to its size (assuming spherical shape), the
calculated diffusion rate of EGFP should only be 25% faster than that
of AK1-EGFP. However, in the cytoplasm, the
DEGFP/DAK1-EGFP is 1.8, which is 50% more than one would expect from free diffusion of
AK1-EGFP. The unexpected slow diffusion rates of AK1-EGFP and AK1
-EGFP in the cell suggested that they were likely interacting with other proteins or forming oligomers in the cytoplasm. Our results
also excluded the possibility of a truncated EGFP in the AK1-EGFP
transfected cells.
FCS measurements not only confirmed the integrity of the fusion protein
AK1-EGFP but also suggested the existence of some protein interactions.
These interactions could be the oligomerization of the fusion proteins
or the interactions with other cellular proteins. The PCH analysis
provides information on the molecular brightness of the sample. For
example, when a monomeric protein is singly labeled, then its dimer
will appear twice as bright as the monomeric protein. Thus, by
analyzing the molecular brightness of EGFP versus AK1-EGFP, we should
be able to discern protein oligomerization by using molecular
brightness analysis. The fact that the brightness of AK1-EGFP and
AK1
-EGFP were roughly that of EGFP in the cell and in solution
(Table 2) suggested that the fusion proteins didn't form oligomers in
the cell. The slightly lower brightness in the fusion proteins, as
compared with EGFP alone, suggested that the formation of a chimeric
protein could slightly quench EGFP fluorescence. This observation of
quenching was consistent with the change in the measured fluorescence
lifetime (Table 2). Therefore, the smaller diffusion coefficient of
AK1-EGFP and AK1
-EGFP in the cell was not due to protein aggregation
but rather interactions with other cellular structures or proteins. One
candidate might be the potassium KATP channels.
There was a report that AK communicates cellular energetic signals to
ATP-sensitive potassium channels. Carrasco et al. (2001)
found that the
response of KATP channels to metabolic challenge
was regulated by AK, and increased AK activity was detected in
KATP channel immunoprecipitates. FCS measurements
confirmed that the fluorescent particles in the cytoplasm and the
nucleus of the AK1
-EGFP-transfected cell were AK1
-EGFP chimera
proteins as well. A few rationales for its appearance in those regions
include overexpression of the protein causing its nonspecific
localization in the cells, the interaction between the protein and the
membrane being not sufficiently strong to retain all proteins bound to
the membrane, and AK1
protein being recruited to those regions for
an unknown purpose.
Dynamics of AK1
-EGFP on the plasma membrane
Fluorescence images of AK1
-EGFP-transfected cells revealed that
AK1
-EGFP was targeted to the plasma membrane of the cell (Fig. 1).
Membrane proteins are classified as being either peripheral or
integral. Nearly all known integral membrane proteins span the lipid
bilayer. In contrast, peripheral proteins are bound to membranes
primarily by electrostatic and hydrogen-bond interactions. Many
peripheral membrane proteins are bound to the surfaces of integral
proteins or are anchored to the lipid bilayer by a covalently attached
hydrophobic chain, such as a fatty acid. FCS measurements of
AK1
-EGFP in the membrane enabled us to ascertain the dynamic interactions between AK1
-EGFP and the membrane. For a variety of
reasons, FCS measurements of membrane-bound protein are more challenging than those of cytoplasmic proteins. The diffusion rate of a
membrane-bound protein is much slower than that of a cytoplasmic
protein. The slow exchange rate of a protein in the laser focus tends
to result in more photo-damage to the membrane protein. The plasma
membrane is thinner (~10 nm) than the height of the focal volume
(~1 µm). If the fluorescent particles are present both on the
plasma membrane and in the cytosol, when the laser beam is focused on
the plasma membrane, it is not possible to isolate pure membrane
diffusion from intracellular diffusion. One expects that the recovered
autocorrelation curves should have multiple or distributed components.
Analysis of our experimental results showed that the autocorrelation
curves of AK1
-EGFP indeed have two components (Fig. 2
inset) when the laser was focused on the top of the cell
(plasma membrane). The ratio of the amplitude of two components
(fast:slow) is 1.2, and the ratio of the diffusion rates of the two
component (fast:slow) is 70. We attributed the slow component to the
lateral diffusion of AK1
-EGFP on the membrane, whereas the fast
component corresponded to the diffusion of cytosolic AK1
-EGFP. By
slightly changing the laser focal plane in the Z direction,
the ratios of the amplitude contribution of the two components
(fast:slow) were varied; they range between ~0.39 and ~6.8. In one
data set, when focusing at the very top of the cell, we were able to
measure 72% membrane-bound protein and 28% cytosolic protein in the
excitation volume. Besides diffusion, there was also a possibility of
an association/dissociation process of AK1
-EGFP between the membrane
fraction and the cytosolic fraction, which could contribute to the fast
component of the autocorrelation curve. The two components we observed
were not due to anomalous diffusion on the plasma membrane. First,
because changing the Z-direction focus affected the ratio of
the two components in the autocorrelation curve. Second, the diffusion
coefficient of the fast component matched that of cytoplasmic
AK1
-EGFP. There have been reports that fluorescently labeled
particles diffuse anomalously on cell membranes and inside the cell.
However, EGFP in cells can be described by assuming a single diffusing
species (Chen et al., 2002
; Schwille et al., 1999b
). It has been
suggested that the anomalous diffusion of the fluorescent dye in cells
may be due to interaction of the dye with intracellular structures or
with much larger soluble molecules.
The lateral diffusion rates of cellular components on the
membrane have traditionally been studied by fluorescence recovery after
photobleaching (Jacobson et al., 1976
). The reported diffusion rates of
membrane proteins measured by fluorescence recovery after photobleaching range from 10
3 to
100 µm2/s, depending on
their properties and location. For example, the diffusion coefficient
for Na,K-ATPase is 3.3 × 10
2
µm2/s (Paller 1994
), whereas the diffusion
coefficient of the platelet-derived growth factor
-receptors in
human fibroblasts can increase to 0.11 µm2/s
upon the addition of platelet-derived growth factor
(Ljungquist-Hoddelius et al., 1991
). The diffusion coefficient of the
slow component of AK1
-EGFP falls in the same range as that of other
lipid-anchored proteins (Niv et al., 1999
). This observation suggested
that some of the AK1
-EGFPs were indeed bound to the membrane.
Because AK1
-EGFP and AK1-EGFP differ only by the 18 amino acid
residues at the N-terminus, and AK1-EGFP doesn't bind to the plasma
membrane, it is likely that the 18 amino acid residues are interacting
with the inner surface of the plasma membrane. Alternatively, these residues provide a signal for protein modification that enables the
protein to bind to membrane or a combination of both possibilities. Dr.
Schneider's group, discoverers of this novel gene first suggested the
possibility of lipid modification on the second residue of the AK1
(Collavin et al., 1999
). Our in vitro studies of AK1
confirmed that
AK1
has a signal for protein myristoylation, and the fatty acid
chain enabled its binding to the lipid bilayer. The results of the in
vitro characterization of AK1
will be presented in another study.
We observed unevenly distributed fluorescence intensity on the plasma
membrane based on the fluorescence images of the
AK1
-EGFP-transfected cells (Fig. 1), suggesting the possibility of
some AK1
-EGFP membrane clustering. Our PCH analysis showed that
these clusters were not AK1
-EGFP aggregates but rather represented a
membrane region with a higher AK1
-EGFP concentration. The plasma
membrane presents an intriguing mix of dynamic process activities in
which components may randomly diffuse, be confined transiently to small
domains (lipid raft), or experience highly directed movements (Jacobson et al., 1995
). The diffusion coefficient of AK1
-EGFP we measured on
the membrane ranges between ~0.1 and ~0.5
µm2/s (Fig. 4). The wide distribution in
diffusion coefficient could be due to the complexity of the membrane organization.
Fluorescence lifetime imaging and molecular brightness
The fluorescence lifetime of a probe is a
concentration-independent parameter, which has been widely used to
study molecular interactions in the fluorescence resonance energy
transfer method. The fluorescence lifetime of EGFP in solution was
reported to be 2.9 ns and 2.6 ns in the cytosol (Swaminathan et al.,
1997
). We are interested in examining the effect of the fusion protein on the lifetime of EGFP and comparing the results with our PCH analysis. The fluorescence lifetime of EGFP in solution (pH 7.2) was
2.8 ns, which agrees with the reported value. The lifetime of EGFP
measured in the cell was 2.45 ns, which was shorter than the value
measured in solution. The lifetime of EGFP fusion proteins was slightly
smaller than that of EGFP in the cell, which was probably due to
quenching. These results appeared to be somewhat contradictory with the
brightness of EGFP obtained from PCH analysis, where EGFP in solution
and in the cell has identical brightness. The difference in lifetime
(2.8 ns vs. 2.45 ns) could be due to the cellular environment. We have
considered two factors: the pH and the background fluorescence. Control
experiments showed that the lifetime of EGFP was insensitive to pH
variation in the range of pH 5-9. Therefore, the pH difference between
the solution and the intracellular environment probably was not the
contributing factor. The background fluorescence and scattering were
the most likely contributing factors in the variation of lifetime in
the cell. Molecules that produced background fluorescence and the scattering light tend to have an apparent short lifetime, which could
shorten the average lifetime even though its fluorescence intensity
contributes less than 10% of the total fluorescence intensity measured
in the transfected cell. In the PCH analysis, the brightness of
background fluorescence molecule is much lower than that of EGFP (Chen
et al., 2002
; Schwille et al., 1999b
); therefore, it has little effect
on the brightness of EGFP and its fusion proteins in the cell. We
believe the brightness of EGFP measured in the cell better represents
its fluorescence property than the apparent lifetime.
| |
SUMMARY |
|---|
|
|
|---|
In summary, we constructed and expressed chimeric proteins of EGFP
and AK isoforms (AK1 and AK1
) in HeLa cells. We directly visualized
the distribution of two related fusion proteins in viable cells:
AK1-EGFP was distributed throughout the cell and AK1
-EGFP was mostly
localized on the plasma membrane. We quantitatively measured the
diffusion coefficients of the fusion proteins in different regions of
the cell. The diffusion of EGFP and AK1-EGFP in the cytoplasm was
described by assigning a single diffusing species, and no anomalous
diffusion was observed. There was some AK1
-EGFP distributed in the
cytoplasm and nucleus of the cell, including a portion that may
interact with intracellular membranes or other structures. AK1
-EGFP
showed a similar diffusion rate distribution as AK1-EGFP, suggesting
that their interactions with the intracellular environment are similar.
Multiple diffusion species were recovered from the autocorrelation
curves of AK1
-EGFP measured on the plasma membrane. Detection of the
slow-diffusion component suggested that AK1
-EGFP was bound to the
plasma membrane and underwent lateral diffusion on the membrane. The
fast-diffusing species could arise from the cytosolic AK1
-EGFP or
the exchange of AK1
-EGFP between its cytosolic and membrane-bound
fractions. The brightness of EGFP and its fusion proteins suggested
that they do not form aggregates in the cell. The apparent fluorescence lifetime of EGFP and its fusion protein in the cell was shorter than
that of EGFP in solution. This could due to the contribution of
background fluorescence. Our system (AK1-EGFP and AK1
-EGFP) provides
a good example of studying isozyme properties in the cell using
fluorescence microscopy. We particularly note the ability to
distinguish different localization and diffusion rates, in turn
reflecting functional differences. Of course, these techniques can also
be extended to study intrinsic protein interactions or those with drugs
in live cells.
| |
ACKNOWLEDGMENTS |
|---|
We thank Dr. Claudio Schneider, Department of Biologia Celular,
Universidad Simon Bolivar, for providing the original AK1 and AK1
plasmids. All the experiments reported in this study were performed at
the Laboratory for Fluorescence Dynamics (LFD) in the Department of
Physics of the University of Illinois at Urbana-Champaign (UIUC). The
LFD is funded by the National Institutes of Health (NIH RR03155) and UIUC.
| |
FOOTNOTES |
|---|
Address reprint requests to Dr. William W. Mantulin, 1110 W. Green St., Urbana, IL 61801-3080. Tel.: 217-244-5620; Fax: 217-244-7187; E-mail: mantulin{at}uiuc.edu.
Submitted May 23, 2002, and accepted for publication July 24, 2002.
| |
REFERENCES |
|---|
|
|
|---|
bioluminescent protein from the luminous hydromedusan, aequarea.
J. Cell. Comput. Physiol.
59:223-239.
Biophys J, December 2002, p. 3177-3187, Vol. 83, No. 6
© 2002 by the Biophysical Society 0006-3495/02/12/3177/11 $2.00
This article has been cited by other articles:
![]() |
J. Girouard, G. Frenette, and R. Sullivan Seminal Plasma Proteins Regulate the Association of Lipids and Proteins Within Detergent-Resistant Membrane Domains of Bovine Spermatozoa Biol Reprod, May 1, 2008; 78(5): 921 - 931. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Garcia-Marcos, S. A. Sanchez, P. Parada, J. Eid, D. M. Jameson, M. Remacha, E. Gratton, and J. P. G. Ballesta Yeast Ribosomal Stalk Heterogeneity In Vivo Shown by Two-Photon FCS and Molecular Brightness Analysis Biophys. J., April 1, 2008; 94(7): 2884 - 2890. [Abstract] [Full Text] [PDF] |
||||
![]() |
|