| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
-Repressor Oligomerization Kinetics at High Concentrations Using Fluorescence Correlation Spectroscopy in Zero-Mode Waveguides


* School of Applied & Engineering Physics, Cornell University, Ithaca, New York 14853; and
Department of Molecular Biology, Princeton University, Princeton, New Jersey 08544
Correspondence: Address reprint requests to Harold Craighead, C. W. Lake Jr. Professor of Engineering, Applied Physics, 205 Clark Hall, Cornell University, Ithaca, NY 14853. Tel.: 607-255-8707; E-mail: hgc1{at}cornell.edu.
| ABSTRACT |
|---|
|
|
|---|
repressor protein at micromolar concentrations. The results agree with previous studies utilizing conventional techniques. Additionally, we demonstrate that the zero-mode waveguides can be used to assay biological activity by measuring changes in diffusion constant as a result of ligand binding. | INTRODUCTION |
|---|
|
|
|---|
Since its introduction several decades ago, florescence correlation spectroscopy (FCS) has been used extensively to study biological and chemical reactions both in vitro and in vivo (Cluzel et al., 2000
; Webb, 2001
). This technique has allowed the study of diffusive properties of fluorescent probes in systems ranging from polymer networks to subcellular compartments (Gennerich and Schild, 2000
; McCain et al., 2004
). FCS requires considerably less reagent than chromatography or mobility shift assays and relies on fluorescent tagging, thus alleviating the need for labeling with radioisotopes. Recent studies include characterization of protein folding dynamics as well as investigations into the intracellular mobility of macromolecules (Rischel et al., 2003
; Schmiedeberg et al., 2004
). FCS studies rely on fluorophore concentrations that yield around a single fluorescent molecule in the observation volume at a time. With typical diffraction limited systems restricted to femtoliter or larger observation volumes, fluorophore concentrations are limited to the pico- to nanomolar range.
Studies of systems that involve ligand binding or chemical change frequently require micromolar or higher reagent concentrations. For FCS to be applied to these systems, such as dimerization reactions that occur at high concentrations, it is necessary to reduce the observation volume. Attempts to work at higher concentrations include the use of near-field scanning optical microscopy and total internal reflection to produce an evanescent illumination profile (de Lange et al., 2001
). These techniques can effectively reduce the observation volume by an order of magnitude or more, and both have been used to study single molecules (Betzig and Chichester, 1993
; Mashanov et al., 2003
). Nanofabricated channels have also been used as a means of confining fluorescent probes and have the added benefit of increasing the rate of single molecule observations (Foquet et al., 2004
). Each of these techniques has increased the working concentration. However, none has achieved the three order of magnitude decrease in observation volume required to apply FCS to a multitude of systems that function at higher concentrations.
Zero-mode waveguides provide an excellent means of volume confinement for single molecule studies at high concentrations (Levene et al., 2003
). The waveguides have been implemented as small holes in a thin aluminum film on a fused silica substrate. Fluorescence is induced by a laser coupled to the waveguide via a confocal microscope in epi-illumination mode. The diameter of the metal waveguide, generally <50 nm, prevents any guided modes and hence produces a radially confined evanescent field with a total volume on the order of 1021 l (Fig. 1). This observation volume makes concentrations higher than 10 µM accessible to FCS experiments and allows one to apply this simple, versatile technique to previously inaccessible biosystems (Fig. 2).
|
|
repressor protein, CI. This protein controls the fate of infecting bacteriophage particles. If the CI concentration in the cell is not high enough, the viral genome will replicate, produce more bacteriophage, and lyse the host cell (the lytic cycle). The viral genome can also integrate into the Escherichia coli host genome, where it is switched off, lying dormant for an indefinite period (the lysogenic state; Ptashne, 1992
|
| THEORY AND MODELING |
|---|
|
|
|---|
The precise shape of the autocorrelation curve is closely tied to the shape of the observation volume. The observation volume is defined by both the shape and intensity of the excitation illumination and the spatial sensitivity of the detection apparatus (Hess and Webb, 2002
). Examining the profile of an evanescent wave below the cutoff frequency in a metallic waveguide yields the following relationship between the illumination profile and the waveguide diameter, d (Jackson, 1999
):
![]() | (1) |
is the wavelength of excitation illumination and
c = 1.7d is the waveguide's cutoff wavelength. The transverse field intensity is assumed to be uniform so that the intensity profile is given by I(x,y,z) = ez/
. If the fluorophore's coupling efficiency and quantum efficiency are assumed to be proportional to the intensity profile, then the zero-mode waveguide's observation profile can be approximated by an exponential decay in one dimension, S(z) = ez/L = e3z/
, where L sets the length scale for the decay. This yields an observation volume given by
![]() | (2) |
14 nm and VObs is
14 zl corresponding to a working concentration of
100 µM.
Because the transverse illumination is essentially uniform in small waveguides, diffusion of fluorescent species in the zero-mode waveguides is expected to appear one dimensional. Consequently, the autocorrelation curve is given by
![]() | (3) |
is the concentration correlation function, a hat denotes the Fourier transform, and S(z) is the observation profile approximated by ez/L (Elson and Magde, 1974
![]() | (4) |
An approximation yields a suitable fitting function. Let R be the ratio of the observation volume decay constant to the height of the waveguide, L/H, and let the diffusion time,
d, be defined as L2/D. A change of variables yields
![]() | (5) |
![]() | (6) |
![]() | (7) |
Experimental results as well as modeling (Levene et al., 2003
) indicate that for small waveguides (3070 nm), R is <0.15, whereas for larger holes, R rarely exceeds 0.25. The lower bound was found to be a markedly better approximation to numerically integrated values for Eq. 5 as shown in Fig. 4. This yields an approximation that is 97.7% accurate or better for small holes. Since the relative error from one fit to another is on the order of 10%, this approximation proves to be acceptable. The final fitting function is (Fig. 4):
![]() | (8) |
d, the average residency time in the observation volume, and the geometric ratio R. G0 is related to the average number of molecules in the focal volume, N, and the constant background signal, B (Koppel, 1974
![]() | (9) |
|
/4 in Eq. 8 is a normalizing factor required to smoothly combine the one-dimensional diffusion term and the correction term. The one-dimensional diffusion term results from integrating the first integral on the left-hand side of Eq. 5 and is consistent with previous models for one-dimensional diffusion (Starr and Thompson, 2001
d for each of the species and R, which is defined by the waveguide only. | MATERIALS AND METHODS |
|---|
|
|
|---|
Zero-mode waveguides
Zero-mode waveguide arrays were fabricated in a thin but opaque layer of aluminum deposited on a 170 µm fused silica cover slip. Film thicknesses were measured at the time of evaporation by the crystal monitor and again later by a profilometer. Both measurements indicated a film thickness of 100 nm. Electron beam lithography and anisotropic dry etching were used to create a variety of hole sizes. After fabrication, the zero-mode waveguides were attached to 12-µL glass wells using UV curable glue and exposed to a low power oxygen plasma to condition the surface. After a sample was placed in a well, it was sealed using a small polydimethylsiloxane slab.
Preparation of CI and OR1
CI was amplified from
DNA (New England Biolabs; Beverly, MA; EB N3011S) by polymerase chain reactionand cloned into the BamHI site of expression vector pPROTet.E133 (BD Biosciences Clontech, Palo Alto, CA to yield pPROTet133:CI. Plasmid DNA samples with inserts of the correct size were purified and checked for CI expression by standard immunity tests using superinfection of plasmid-carrying strains with wild-type
,
vir, and
i434 (Arber et al., 1983
). mRFP was then fused at the N-terminus of CI by cloning mRFP into pPROTet133:CI using standard methods (Sambrook and Russell, 2001
) and checked again for immunity. At this stage the fusion was also checked by DNA sequencing. pPROTet133 is an expression vector in which the cloned gene is fused to a histidine asparagine hexamer (HN6), and this tag at the N-terminus was used to purify the fusion protein on a cobalt column according to the manufacturer's protocol. The purified protein was then tested for in vitro activity by standard gel shift assays (Sambrook and Russell, 2001
) using a 200 basepair
fragment containing all three right operator sequences labeled at the 5' end with either 32P or Alexa-fluor 488. The band shift assay exhibited the expected dependence on protein concentration, and binding could be completed by excess unlabeled fragment but not by a random 19-mer of the same base composition. The 17 basepair OR1 sequence (TATCACCGCCAGAGGTA) and its complement were synthesized by standard methods, annealed, and purified by agarose gel electrophoresis.
Data analysis
The strategy for analyzing FCS curves acquired in the waveguides is to fit the data to Eq. 8 with G0, R, and
d as free parameters. Interestingly, from the definition of
d and R, it is easily seen that the value of L need never actually be determined, hence eliminating the need for a precise calibration of the waveguides. Instead it is sufficient to accurately measure the depth of the structures and calculate the diffusion constant from the equation
![]() | (10) |
d are empirical parameters and H is well known from the fabrication process. FCS autocorrelation functions are dependant on the diffusion time of the fluorescent species (Elson and Magde, 1974
Autocorrelation curves were fit to the model in Eq. 8 using Matlab's nonlinear curve fitting routines and the Levenberg-Mardquart algorithm. Free parameters were G0,dim,
d,dim, G0,tet,
d, tet, and R. The R2 (residual) values were all >.998, indicating good fits. The values obtained for R agree with theoretical predictions for the waveguides' behavior.
| RESULTS AND DISCUSSION |
|---|
|
|
|---|
1 µM solution of mRFP-CI was prepared and observed in the zero-mode waveguides. At this concentration, dimers and tetramers should be the dominant species, with the tetramers having the larger diffusion constant (Pray et al., 1998
d dim and
d tet, the dimer and tetramer diffusion times, and the geometric ratio R. The values obtained from one waveguide are summarized in Table 1. Equation 10 can be used to calculate the diffusion constants directly from the fit parameters; however, it is illuminating to plot the diffusion times as a function of the geometric factor R. As can be seen in Fig. 6, which shows the diffusion time and R parameters plotted for groups of FCS curves taken from several different waveguides, the data conform reasonably well to the expected quadratic relationship. Interestingly, the variations from curve to curve taken in a single waveguide conform to the same quadratic relationship. This suggests that there are small variations in the illumination profile as a function of time. And it seems likely that small shifts in stage or laser position give rise to these variations. Laser intensity was stable over the time period of the experiment. Despite the clear trend relating the diffusion time to the parameter R, neither G0,dim nor G0,tet displayed any notable dependence on R. This is unexpected, since an increase in diffusion time is assumed to be due to an increase in the total observation volume and a larger observation volume should admit more molecules.
|
|
|
|
0.4 microns in radius. Since the waveguides are on the 50 nm scale, and aluminum is an extremely reflective metal, a large fraction of the illumination must be reflected. The use of high quality fluorescence filters produces at least an optical density 6 reduction in the intensity of the excitation line. For traditional fluorescence techniques, this is more than sufficient and yields a good signal-to-noise ratio. When observing single molecules in zero-mode waveguides, however, the relatively weak signal may still be small compared to the transmitted excitation line. An additional source of background was optical transmission through the waveguides. The objective focuses as much as a milliwatt of laser power on a spot only a few microns across. Consequently, the local optical power density is high enough that the waveguide cannot totally attenuate the excitation line. The resulting weak transmission excited the concentrated solution of fluorophores in the reservoir above the waveguide and, as a result, a constant background was coupled back into the waveguide and detected by the photodiode. This background is proportional to the decay profile through the waveguide squared: once as the excitation passes through the guide and once again as the fluorescence passes back to the detector. In contrast, signal from the observation volume is proportional to the decay profile cubed: once for the excitation profile, once for the coupling interaction, and once for the inhibited emission behavior. As a result, the background should have a considerably weaker dependence on waveguide size than does the signal.
The zero-mode waveguide fabrication procedure included a dry chlorine process to etch the waveguides into the aluminum film. Chlorine is known to etch fused silica, albeit at a rate 10 times slower (Williams and Muller, 1996
) than aluminum. This etch is thought to produce small wells at the bottom of the waveguides. Any fluorophore that passes entirely through the waveguide into these subwaveguide glass wells was illuminated uniformly at full excitation intensity and suffers no coupling or radiation attenuation. Consequently, a small glass well beneath the waveguide will produce an intense background signal. Even a few seconds of etching can produce volumes an order of magnitude larger than the waveguide observation volume. Unlike the previous background sources, the fluorescence intensity from the wells should be strongly dependent on the size of the waveguide because the waveguide itself acted as an etch mask.
Though the background has no impact on the measurement of diffusion times, it depresses G0 from its expected value and must be characterized if meaningful kinetics measurements are to be made. A detailed and comprehensive analysis of the sources and intensities of the background would be a difficult task. However, the data can be background corrected without any specific knowledge of the intensity or source of the constant background signal. The equilibrium constant characterizing the tetramerization of CI should be determined wholly by the ratio Ntet/Ndim. Consequently, it is important to note that N is not a linear function of B (Eq. 9). If it were, then the background could be disregarded entirely, confident that the effects would cancel out. The correction can be accomplished by solving the following nonlinear system:
![]() | (11) |
The trends and values of the corrected data are entirely as expected. Whereas the original G0,dim and G0,tet values showed no significant dependence on R and hence the observation volume, the corrected data exhibit the expected linear relationship (Fig. 7 A). Fitting the data to a line yields the concentration of dimer or tetramer in solution. Additionally, the background, B, exhibits a dependence on R (Fig. 7 B). This is expected since the volume of the subwaveguide well depends on hole size. With the corrected data, the tetramerization constant can be calculated from the definition or a volume independent formula:
![]() | (12) |
G = RTlnK was 4 x 106 ± 1 x 106 M at 20°C, 25°C, and 30°C. The same study indicated that the rate limiting step in CI octamerization is the formation of tetramers (Pray et al., 1998
|
1 µM. Solutions containing 1 µM mRFP-CI, 2.4 uM OR1 and diluted in the standard Sauer binding buffer were prepared, and a series of 10 autocorrelation curves were acquired (Johnson et al., 1980
|
| CONCLUSION |
|---|
|
|
|---|
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
This work was supported by the Department of Eergy, by the Defense Advanced Research Projects Agency, and by the National Institutes of Health (NIH HG001506). Fabrication was done at the Cornell Nano-Scale Science and Technology Facility supported by National Science Foundation grant ECS-9731293, its users, Cornell University, and industrial partners.
Submitted on September 14, 2004; accepted for publication November 18, 2004.
| REFERENCES |
|---|
|
|
|---|
Betzig, E., and R. J. Chichester. 1993. Single molecules observed by near-field scanning optical microscopy. Science. 262:14221425.
Burz, D. S., and G. K. Ackers. 1996. Cooperativity mutants of bacteriophage lambda cI repressor: temperature dependence of self-assembly. Biochemistry. 35:33413350.[CrossRef][Medline]
Cluzel, P., M. Surette, and S. Leibler. 2000. An ultrasensitive bacterial motor revealed by monitoring signaling proteins in single cells. Science. 287:16521655.
de Lange, F., A. Cambi, R. Huijbens, B. de Bakker, W. Rensen, M. Garcia-Parajo, N. van Hulst, and C. G. Figdor. 2001. Cell biology beyond the diffraction limit: near-field scanning optical microscopy. J. Cell Sci. 114:41534160.
Dodd, I. B., A. J. Perkins, D. Tsemitsidis, and J. B. Egan. 2001. Octamerization of lambda CI repressor is needed for effective repression of P-RM and efficient switching from lysogeny. Genes Dev. 15:30133022.
Elson, E. L., and D. Magde. 1974. Fluorescence correlation spectroscopy. 1. Conceptual basis and theory. Biopolymers. 13:127.[CrossRef]
Foquet, M., J. Korlach, W. R. Zipfel, W. W. Webb, and H. G. Craighead. 2004. Focal volume confinement by submicrometer-sized fluidic channels. Anal. Chem. 76:16181626.[Medline]
Gennerich, A., and D. Schild. 2000. Fluorescence correlation spectroscopy in small cytosolic compartments depends critically on the diffusion model used. Biophys. J. 79:32943306.
Hess, S. T., and W. W. Webb. 2002. Focal volume optics and experimental artifacts in confocal fluorescence correlation spectroscopy. Biophys. J. 83:23002317.
Jackson, J. D. 1999. Classical Electrodynamics. Wiley & Sons, New York.
Johnson, A. D., C. O. Pabo, and R. T. Sauer. 1980. Bacteriophage lambda repressor and cro protein: interactions with operator DNA. Methods Enzymol. 65:839856.[Medline]
Koppel, D. E. 1974. Statistical accuracy in fluorescence correlation spectroscopy. Phys. Rev. A. 10:19381945.[CrossRef]
Levene, M. J., J. Korlach, S. W. Turner, M. Foquet, H. G. Craighead, and W. W. Webb. 2003. Zero-mode waveguides for single-molecule analysis at high concentrations. Science. 299:682686.
Lide, D. R. 2003. CRC Handbook of Chemistry and Physics, 84th ed. CRC Press, Boca Raton, FL.
Mashanov, G. I., D. Tacon, A. E. Knight, M. Peckham, and J. E. Molloy. 2003. Visualizing single molecules inside living cells using total internal reflection fluorescence microscopy. Methods. 29:142152.[CrossRef][Medline]
McCain, K. S., P. Schluesche, and J. M. Harris. 2004. Modifying the adsorption behavior of polyamidoamine dendrimers at silica surfaces investigated by total internal reflection fluorescence correlation spectroscopy. Anal. Chem. 76:930938.[Medline]
Pray, T. R., D. S. Burz, and G. K. Ackers. 1998. Cooperative non-specific DNA binding by octamerizing lambda cI repressors: a site-specific thermodynamic analysis. J. Mol. Biol. 282:947958.[CrossRef][Medline]
Ptashne, M. 1992. A Genetic Switch. Cell Press and Blackwell Science, Cambridge, MA.
Rischel, C., L. E. Jorgensen, and Z. Foldes-Papp. 2003. Microsecond structural fluctuations in denatured cytochrome c and the mechanism of rapid chain contraction. J. Phys. Condens. Matter. 15:S1725S1735.[CrossRef]
Sambrook, J., and D. W. Russell. 2001. Molecular Cloning. Cold Spring Harbor Press, Cold Spring Harbor, NY.
Schmiedeberg, L., K. Weisshart, S. Diekmann, G. M. Z. Hoerste, and P. Hemmerich. 2004. High- and low-mobility populations of HP1 in heterochromatin of mammalian cells. Mol. Biol. Cell. 15:28192833.
Starr, T. E., and N. L. Thompson. 2001. Total internal reflection with fluorescence correlation spectroscopy: combined surface reaction and solution diffusion. Biophys. J. 80:15751584.
Webb, W. W. 2001. Fluorescence correlation spectroscopy: inception, biophysical experimentations, and prospectus. Appl. Opt. 40:39693983.[CrossRef]
Williams, K. R., and R. S. Muller. 1996. Etch rates for micromachining processing. J. Microelectromech. Syst. 5:256269.[CrossRef]
This article has been cited by other articles:
![]() |
C. H. Reccius, S. M. Stavis, J. T. Mannion, L. P. Walker, and H. G. Craighead Conformation, Length, and Speed Measurements of Electrodynamically Stretched DNA in Nanochannels Biophys. J., July 1, 2008; 95(1): 273 - 286. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Korlach, P. J. Marks, R. L. Cicero, J. J. Gray, D. L. Murphy, D. B. Roitman, T. T. Pham, G. A. Otto, M. Foquet, and S. W. Turner Selective aluminum passivation for targeted immobilization of single DNA polymerase molecules in zero-mode waveguide nanostructures PNAS, January 29, 2008; 105(4): 1176 - 1181. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. T. Samiee, J. M. Moran-Mirabal, Y. K. Cheung, and H. G. Craighead Zero Mode Waveguides for Single-Molecule Spectroscopy on Lipid Membranes Biophys. J., May 1, 2006; 90(9): 3288 - 3299. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |