| ABSTRACT |
|---|
|
|
|---|
As our understanding of cellular organization increases, there is a need for methodologies to ascertain molecular concentration, dynamics, and organization at high temporal and spatial resolution. A common approach to study cellular dynamics uses fluorescently tagged molecules that can be observed by fluorescence wide field, total internal reflection fluorescence (TIRF), or confocal microscopy. Single-point fluorescence correlation spectroscopy (FCS) originally developed by Magde et al. for solutions (1
) and more recently by Berland et al. for cells (2
), provides high temporal resolution information about protein concentrations and dynamics within a small volume (<1 fL) of the cell. However, the limitations of this technique include a lack of spatial information, restricted information about processes that involve slower protein dynamics within large slowly moving structures, and lack of simultaneous cell imaging. Image correlation spectroscopy (ICS) as originally developed by Petersen et al. (3
,4
) provides the concentration, degree of aggregation of proteins, and the average number of aggregates in the cell membrane. In a recent variation, ICS was expanded to measure temporal correlations between images collected in a time sequence (5
) and to determine spatial correlations due to flow and other cellular processes (6
). The ICS theory was developed with the assumption that nothing moves on the timescale of the frame acquisition. To date it has been applied to fixed samples or slowly moving transmembrane or cytoskeleton associated proteins (6
). Overall, ICS provides a powerful method for systematically analyzing images for dynamic processes and molecular aggregation. The main idea of the ICS method is to calculate the image spatial autocorrelation function and to extract the number and size of aggregates from the analysis of the spatial power spectrum. For the time correlation in ICS, the autocorrelation function is calculated from stack of successive images.
We propose a novel approach to probe spatial correlations and previously inaccessible temporal windows. We exploit the time structure present in images obtained with a laser confocal microscope to spatially correlate adjacent pixels that are a fraction of a micron and a few microseconds apart along a line and a few milliseconds apart in successive lines. By exploiting this time structure we measure dynamic processes such as molecular diffusion in the microseconds to second timescale. We demonstrate this data analysis by measuring the diffusion of fluorescent beads, enhanced green fluorescent protein (EGFP) in solution, and EGFP inside the cytoplasm of live cells using raster scan images. This new analysis, termed raster image correlation spectroscopy (RICS), now allows one to bridge the timescales of FCS and ICS and provide spatially resolved information in the microsecond to seconds time range. To date, expensive and specialized instruments have been used for FCS experiments. A drawback of these instruments, when used for cell work, is that the correlation measurements and imaging are done at two different times. Because laser-scanning confocal microscopes are readily available in almost all life science labs, our method opens up scanning correlation microscopy to the general microscope user. In fact, we used an analog system (Olympus Fluoview 300, Melville, NY) with excellent results in a forthcoming article.
We consider signal fluctuations due to the diffusion of particles in a homogenous medium. In addition to translational diffusion, other processes at the molecular level such as conformational transitions, quenching associated with aggregation and molecular rotations can also cause fluorescence fluctuations. We focus on diffusion; as this is a common mode of transport of molecules from one location to another in the cell. The change in concentration of particles as a function of time due to diffusion in a uniform medium is described by the following relationship:
![]() | [1] |
If particles are either not moving or are moving very slowly, as we sample different locations in a raster scan image, the intensity at one point is correlated to the intensity of the adjacent points if there is superposition of the point spread function (PSF) at the adjacent points (Fig. 1, situation 1). In this case the two-dimensional (2-D) spatial correlation of an image just reflects the extent of superposition of the PSF in the adjacent points. To explain spatial correlation for fast-moving particles, consider two points separated by a given distance r. If this distance is less than the width of the PSF then there may be some correlation of the fluorescence between the two points (Fig. 1, situation 2) but the spatial correlation will be less than that for immobile particles. However, for points that are not superimposed by the PSF, there could be some correlation due to the diffusion of the particle to these points.
|
r (in the range 0.10.2 µm), the pixel residence time,
(in the range 5100 µs), and the pixel sequence number n. For points along a line n is 1. For points in different lines, n is equal to the number of pixels in a line (256 or 512) plus an equivalent number of pixels corresponding to retracing time (typically 64128).
![]() | [2] |
![]() | [3] |
The overall correlation function is given by
. For raster scan pattern the correlation of a series of images appears on three different timescales. Along the horizontal axis adjacent pixels are separated by microseconds, along the vertical axis pixels are milliseconds apart and pixels from successive image frames are typically seconds apart. We have performed extensive simulations of particles diffusing in a three-dimensional grid to test the validity of the expressions used to fit the spatial autocorrelation functions. These results are not reported in this letter due to space restrictions. We systematically varied the values of the diffusion coefficient, number of particles, pixel size, beam waist, and the other parameters of the experiments. In all cases the recovered parameters from the fit matched the simulated parameters.
For the measurements described in this letter we used the two-photon excitation scanning fluorescence microscope previously described (2
). Excitation wavelengths were 780 nm (for 10-nm polystyrene beads from Molecular Probes, Eugene, OR) and 910 nm (for EGFP). Due to the possible variation in the laser alignment from day to day, the beam waist w0 was calibrated before each day's measurement by measuring the autocorrelation curve for 10 nM fluorescein in 0.01 M NaOH, and was fit using a diffusion constant of 300 µm2/s. Data were collected at a pixel rate of 16 µs/pixel for the solution experiments and 32 µs/pixel for the CHOK1 cell. For solution and cell experiments the scanning area was 128 x 128 pixels corresponding to a 14-µm square. For the solution measurements 2000 frames were collected whereas 790 frames were collected for the CHOK1 cells expressing EGFP experiments. Although only one image is sufficient theoretically to obtain the autocorrelation function and the diffusion coefficient of freely diffusing particles, in practice we average the information of many frames to improve statistics. For the samples in solution each sample was also measured in two different modalities, point FCS and raster scan pattern. The diffusion coefficient recovered from the fit of the single-point FCS matched the parameters recovered from the fit of the 2-D spatial correlation. Fig. 2 shows experimental data for the solution systems investigated.
|
|
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
Submitted on February 22, 2005; accepted for publication March 21, 2005.
| REFERENCES |
|---|
|
|
|---|
2 Berland, K. M., P. T. So, and E. Gratton. 1995. Two-photon fluorescence correlation spectroscopy: method and application to the intracellular environment. Biophys. J. 68:694701.
3 Petersen, N. O. 1986. Scanning fluorescence spectroscopy. I. Theory and simulation of aggregation measurements. Biophys. J. 49:809815.
4 Petersen, N. O., P. L. Hödelius, P. W. Wiseman, O. Seger, and K. E. Magnusson. 1993. Quantitation of membrane receptor distributions by image correlation spectroscopy: concept and application. Biophys. J. 65:11351146.
5 Wiseman, P. W., J. A. Squier, M. H. Ellisman, and K. R. Wilson. 2000. Two-photon video rate image correlation spectroscopy (ICS) and image cross-correlation spectroscopy (ICCS). J. Microsc. 200:1425.[Medline]
6 Wiseman, P. W., C. M. Brown, D. J. Webb, B. Herbert, N. L. Johnson, J. A. Squier, M. H. Ellisman, and A. F. Horwitz. 2004. Spatial mapping of integrin interaction and dynamics during cell migration by image correlation spectroscopy. J. Cell Sci. 117:55215534.
7 Ruan, Q., Y. Chen, E. Gratton, M. Glaser, and W. W. Mantulin. 2002. Cellular characterization of adenylate kinase and its isoform: two-photon excitation fluorescence imaging and fluorescence correlation spectroscopy. Biophys. J. 83:31773187.
This article has been cited by other articles:
![]() |
M. A. Digman, C. M. Brown, A. R. Horwitz, W. W. Mantulin, and E. Gratton Paxillin Dynamics Measured during Adhesion Assembly and Disassembly by Correlation Spectroscopy Biophys. J., April 1, 2008; 94(7): 2819 - 2831. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Celli, S. Beretta, and E. Gratton Phase Fluctuations on the Micron-Submicron Scale in GUVs Composed of a Binary Lipid Mixture Biophys. J., January 1, 2008; 94(1): 104 - 116. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Saxton A Biological Interpretation of Transient Anomalous Subdiffusion. I. Qualitative Model Biophys. J., February 15, 2007; 92(4): 1178 - 1191. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y.-l. Wang, K. M. Hahn, R. F. Murphy, and A. F. Horwitz From imaging to understanding: Frontiers in Live Cell Imaging, Bethesda, MD, April 19-21, 2006 J. Cell Biol., August 14, 2006; 174(4): 481 - 484. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. K. Ramanujan, G. Biener, and B. A. Herman Scaling Behavior in Mitochondrial Redox Fluctuations Biophys. J., May 15, 2006; 90(10): L70 - L72. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Kolin, S. Costantino, and P. W. Wiseman Sampling Effects, Noise, and Photobleaching in Temporal Image Correlation Spectroscopy Biophys. J., January 15, 2006; 90(2): 628 - 639. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Digman, C. M. Brown, P. Sengupta, P. W. Wiseman, A. R. Horwitz, and E. Gratton Measuring Fast Dynamics in Solutions and Cells with a Laser Scanning Microscope Biophys. J., August 1, 2005; 89(2): 1317 - 1327. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |