The EPR oximetry technique is based on the paramagnetic distinctiveness of molecular oxygen (10
). Using a small amount of sample (
10 µl), it allows one to determine molecular oxygen from the oxygen-induced EPR line broadening of a suitable paramagnetic probe such as lithium phthalocyanine (LiPc), nitroxides, lithium octa-n-butoxy-substituted naphthalocyanine (LiNc-BuO), etc. (3
,11
,12
). The origin of this technique was based on the principle of the Heisenburg spin-spin interaction between the molecular oxygen and another spin probe. This spin-spin interaction between the molecular oxygen and the spin probe will shorten the relaxation time of the spin probe, causing an increase in the peak-to-peak width in the first derivative EPR spectrum. The line widths of the EPR spectra correlate with oxygen concentrations for soluble probes and pO2 for particulate probes such as LiPc (13
). EPR oximetry can provide real-time, noninvasive, as well as nonperturbing measurements of oxygen. It is distinctive in its potential for high sensitivity in local oxygen measurements (14
17
).
When it was first introduced for oxygen measurements, several advantages of this method were highlighted including: the fact that the cells and the spin probe(s) are homogenously distributed throughout the sample tubes and the spin probes do not consume oxygen, allowing low concentrations of O2 in solutions to be measured (17
). To date, this method has been shown to be a useful tool for accurate, repeated measurements of local concentrations of molecular oxygen in the ischemic heart, tumors, cerebral interstitial regions of the brain, and in cell suspensions (4
,15
,18
). When considering accurate EPR oximetry measurements, careful planning is necessary to select the most suitable probe. Both soluble nitroxide-based oximetric probes and insoluble particulate-based probes can be used. A limitation with the soluble nitroxides is that these probes are susceptible to cellular reduction and a progressive decay in signal intensity with time during measurements (19
). Although signal intensity is not a major limitation in EPR oximetry where only the line width is used to measure oxygen concentrations, the redox reactions may change the redox equivalence in cells and may even alter the respiration. On the other hand, particulate probes such as LiPc (11
,18
), chars (20
,21
), LiNc, and its derivatives (3
,22
,23
) are inert and noninteractive with cells, allowing them to be useful for oxygen measurements. Specifically, LiPc is stable in tissues for extended periods of time. It is insoluble in aqueous solvents and enables high resolution pO2 measurements in tissues (11
,14
,15
). Typically, the response of the LiPc line shape to pO2 is not altered under various environments such as pH, temperature, redox conditions, etc. However, the oxygen sensitivity may change or even be lost in some tissues (24
,25
).
Several studies have demonstrated that the oxygen consumption rate in cellular respiration can be measured using particulate probes such as LiPc in EPR oximetry (3
,17
,18
). Typically, EPR oximetry measurements are performed for a short period of time, where a linear variation of pO2 with time is observed. The slope of that region is used to quantify the oxygen consumption rate (3
). However through other techniques, it is known that there are three distinct phases of cellular respiration (2
). Thus far, it has not been established whether EPR oximetry is capable of distinguishing these phases. This is partially due to a lack of proper models to interpret the EPR oximetry data. This poses a serious limitation in applying this technique and extracting meaningful respiration parameters. A major problem in this technique is the O2 concentration gradient from the cell surface to the bulk (surrounding medium). During EPR oximetry measurements, the sample is kept stationary, which causes oxygen diffusion zones to form as the oxygen is depleted by the respiration. In this work, a new procedure to analyze EPR oximetry data is used where we demonstrate that these diffusion zones have considerable implications for the measured data. Also, we demonstrate that by correcting the diffusion impacts of oxygen, one can extract more significant results.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Cell culture
Bovine aortic endothelial cells (BAECs) were obtained from Cell Systems (Kirkland, WA). The BAECs were cultured in CS-C Complete medium (Cell Systems). The growth medium contained JetFuel formulated with growth factor, 10% serum, and antibiotic (gentamycin sulfate). Cells were grown in regular 150-cm2 culture dishes coated with attachment factor. Cells were trypsinized and used in experiments when the cultures reached 8090% confluence.
Cell viability
The cell viability was determined by a NucleoCounter system (New Brunswick Scientific, Edison, NJ) comprising the NucleoCounter automatic cell counter, the NucleoCassette, a cell preparation lysing buffer and a stabilizing buffer, and NucleoView software. Two aliquots of the cell suspension, for nonviable count and the total cell count, were taken. For the total cell count, an equal amount of the lysing buffer and the stabilizing buffer were added to the cell suspension. Each sample was loaded into the NucleoCassette and placed into the NucleoCounter cell counter for analysis. The nonviable count was determined first, followed by the total cell count. Using the NucleoView software, the nonviable, total cell count, viable cell count, and viability were determined.
Oximetry probe
For oximetry measurements, we have used
2030 µm LiPc microcrystals as the oximetry probe. Since the LiPc microcrystals were comparatively larger in the oximetry measurements, the particulates remained in the bulk volume and there was no ingestion by the cells. The probe was synthesized electrochemically, using the established procedure (26
). The synthesized microcrystals were subjected to various physicochemical characterizations such as x-ray diffraction (XRD), EPR, microscopy, etc. to ensure the purity of the obtained material. These microcrystals were found to be in the pure
isoform (or equivalently known as x form), which has been characterized to yield a pO2-dependent EPR line width.
Measurement of cellular respiration
The oxygen measurements in this study were performed using EPR oximetry (11
,14
). The EPR line width versus pO2 calibration curve was constructed using known ratios of premixed O2 and N2 gases. The slope of the calibration curve was 5.8 mG/mmHg. Although this calibration curve was constructed using gas mixtures, we have previously demonstrated that this curve is applicable in aqueous solutions as well (11
). Thus, by measuring the EPR line width, the pO2 in the solution can be obtained at any given time.
EPR oximetry experimental setup
The respiration studies have been carried out using an X-band (9.7 GHz) EPR spectrometer, fitted with a TM110 microwave cavity. A 50-µl microcapillary tube was used to hold the cells in the horizontal EPR cavity. In a typical experiment, the cell suspension of the required cell density was saturated with room air (pO2
160 mmHg). The cell suspension was incubated for 10 min in a 37°C water bath, and 20 µg of LiPc microcrystals were added to the cells and sampled into 50-µl capillary tubes. The tube was then closed off at both ends using tube sealing clay (Chase Scientific Glass, Rockwood, TN). While sealing, care was taken to ensure that there were no air gaps present inside the tube since such a gap may act as an additional O2 source. The tube was placed inside the horizontal microwave cavity, and EPR spectral acquisitions of the LiPc were immediately started. During measurements, the modulation amplitude was adjusted to always be less than one third of the line width to avoid modulation-induced broadening. From the EPR line width, the pO2 in the cell suspension was determined using the standard curve.
Experimental protocol for drug treatments
For each drug treatment, the appropriate concentrations of rotenone (20 µM) or stigmatellin (20 µM) were added to the cell culture and incubated for 30 min. The cells were trypsinized and isolated for respiration measurements. At the time of measurements, cell suspensions with the appropriate concentrations of each drug were incubated at 37°C in a water bath for 10 min and then drawn into the glass capillary tubes for oxygen measurements. The cell densities in this study are expressed as the viable cell counts per milliliter, estimated after the addition of the drug.
Curvefit and data analysis
The EPR spectra, collected during the cellular respiration measurements, were analyzed by fitting the spectral data into a Lorentzian line shape by a least-squares algorithm, using custom-designed software. The correlation coefficient of 0.99 was set as the standard of acceptance of the results. The decrease in pO2 data was converted into the rate of oxygen depletion by using four-point piecewise differentiation with respect to time. The four-point differentiation was performed using MATLAB 6.1 (The MathWorks, Natick, MA). The simulations and curve fit of the experimental data were carried out using MATLAB 6.1 Curvefit PC (The MathWorks). Data are presented as mean ± SD. Statistical analysis was performed using Student's t-test and one-way ANOVA. The general acceptance level of significance was P < 0.05.
| RESULTS |
|---|
|
|
|---|
30 mG. To ensure that the oxygen depletion being measured was associated with the mitochondrial respiration, mitochondrial electron transport chain (ETC), inhibitors were used to display their effect on cellular respiration. Rotenone is a complex I blocker, which augments electron leakage from the ETC, and stigmatellin is a complex III inhibitor. Fig. 1 also shows the EPR spectra acquired for
2 h in the presence of rotenone (20 µM); stigmatellin (20 µM) had similar results (spectra not shown). In the case of the ETC blockers, there was no significant change in the EPR spectra, suggesting that the oxygen consumption by the cells is inhibited by these agents (Table 1). Within 60 min, there was only an approximate decrease of 10 mmHg for each inhibitor. Thus, it is confirmed that the oxygen consumption measured by EPR oximetry is essentially due to mitochondrial respiration.
|
|
|
![]() | (1) |
In the above equation, VO2 is the oxygen consumption rate at a given pO2, VO2max is the maximum oxygen consumption rate (steady state), p50 is the concentration at which the VO2max is reduced to 50%, and the p0 is the equilibrium pO2 where pO2 is very low. The p50 [O2] is analogous to the Km values in purified enzymatic reactions. It is an inverse of a measure of the oxygen affinity to cytochrome c oxidase in complex IV of the mitochondrial ETC (28
); therefore, the lower the p50, the greater the oxygen affinity.
The data in Fig. 2 C were fit to Eq. 1 and the parameters VO2max, p50, and p0 were determined (Table 1). By utilizing the above equation, we were able to quantify levels of respiration for various cell densities. Based on the data we acquired, we noticed considerable differences in comparison to previously reported respiration data. The VO2max, determined from this approach, agree with the values determined from the slopes of the linear portion. However, although the VO2max values were comparable to the reported values (2
,27
), the p50 values were unusually higher in each of the cell densities used. For example, the p50 for human umbilical vein endothelial cells and human neuroblastoma cells has been determined to be 0.6 mmHg and 0.8 mmHg, respectively, for 1.5 million cells (27
,29
). Also, the p50 for heart mitochondria is 0.26 mmHg, and liver mitochondria have been stated to be 0.43 mmHg (28
). Moreover, the oxygen affinity measured in this work has shown a strong dependence on the cell densities used as shown in Table 1 and Fig. 2 D. These two discrepancies illustrate the need for the refinement of Eq. 1 to be suitable for EPR oximetry.
A diffusion model for oxygen consumption in EPR oximetry
One important difference in the case of EPR oximetry compared to respirometry methods is the mass transport limitation in EPR oximetry. In HRR methods, constant stirring of the solution is applied. Thus Eq. 1, originally developed for HRR, does not include any mass transfer rate limitation. However, in the case of EPR oximetry, such stirring is not possible and hence, the mass transport may limit the availability of oxygen near the cell surface. Thus, the above model needs to be modified accordingly to take the O2 diffusion into account.
Fig. 3 illustrates the phenomenological conception of the oxygen diffusion limitation. Three different scenarios are depicted in Fig. 3. Fig. 3 A illustrates a condition of low cell density with a higher consumption rate. Thus, diffusion zones develop, but the effective overlap of diffusion zones is negligible. As the time increases, a concentration gradient is developed from the cell surface to bulk, establishing a diffusion zone with radius, r, which increases with time and radial velocity,
t, as defined below (30
):
![]() | (2) |
r3) can be defined as follows:
![]() | (3) |
|
ex, covered by such diffusion zones is then defined by including the number of cells (N):
![]() | (4) |
The
ex is the fraction of volume covered under the diffusion zones, assuming no overlap is occurring. However, when the cell density is considerably high and/or the rate of consumption is elevated, the diffusion zones develop very fast and these zones overlap, as illustrated in Fig. 3 B. Under such circumstances, the volume under diffusion control can be defined by the Avrami theorem (31
) as follows:
![]() | (5) |
![]() | (6) |
Under these conditions, the radial flux of the oxygen delivered to the cell surface is limited by the linear diffusion of O2. The conservation of mass in the diffusion zones implies that the amount of oxygen coming into the diffusion zone is consumed by the cell, causing the consumption rate to be dependent on the diffusion rate. Thus, Eq. 1 can be modified, taking into consideration
, as:
![]() | (7) |
After the appropriate modifications, Eq. 7 was used to simulate normalized pO2 versus time curves for various VO2max values. Fig. 4 A shows the effect of VO2max on the change of pO2 with time. It can be seen that the increase in the VO2max changes the shape of the pO2 versus time curve. At very low VO2max values, a linear variation in pO2 is observed, whereas at high VO2max values a curve is shown. Similarly, an increase in the N value also alters the consumption rate (data not included) because it is directly related to
. The dependence on
is better illustrated when these plots are converted into rate versus time plots as shown in Fig. 4 B. As noticed in Fig. 4 B, the consumption rate remains relatively constant when the VO2max is loweven at longer periods of time. However, when VO2max is high, the rate decreases with time, clearly demonstrating the VO2max-dependent consumption rate with respect to time. When compared to HRR, such reliance is not expected because there is no diffusional impact; thus, it is associated with EPR oximetry and other magnetic resonance based oximetry techniques where stirring is not applied. Furthermore, the dependence of VO2max on pO2 was also demonstrated as shown in Fig. 4 C. At low VO2max values, the rate remains the same at low pO2 ranges. However at high VO2max values, the rate decreases as the pO2 decreases. An important implication of this result is that the p50 (and the oxygen affinity to complex IV of ETC) is indirectly associated with the VO2max, as illustrated in Fig. 4 C. The dependence is uniquely due to the diffusion limitation with EPR oximetry, unlike the polarographic methods where this effect is removed by effective stirring.
|
0.98 in all cases, the new values for the p50, VO2max, and p0 were acquired. The VO2max, determined from the new approach at all cell densities, was not significantly different from the values determined either from the slope or from the initial method of fit. However, there was a significant change in p50 and hence the mitochondrial oxygen affinity (Table 1).
|
50% decrease in the p50 utilizing this new approach. Upon fitting each data set into the new method, there was a distinct shift to the left (i.e., toward the lower pO2 side) for each cell density (Fig. 5). Such a shift was obvious at higher cell densities (Fig. 5, A and B). This clearly implies that the diffusion limitation has a direct effect on the p50, while allowing the maximum rate of consumption to remain unchanged. Furthermore, HRR methods (2
|
| DISCUSSION |
|---|
|
|
|---|
20 µl) and makes it nearly impossible to stir. Conversely, this restriction is neglected in respirometry experiments because the large volume of sample is continuously stirred.
The majority of the reported cellular respiration studies with EPR oximetry show a linear variation of pO2 with time (3
,17
,18
). This is due to the respiration not being measured until the oxygen is completely depleted. Even any noticeable deviation from this linearity was either neglected or qualitatively interpreted (5
,34
,35
). Although this may be sufficient for determining the oxygen consumption rate, the entire course of oxygen consumption is necessary to calculate other relevant parameters such as the p50. Typically, EPR oximetry experiments are performed for a specific time interval and the rate of oxygen consumption is determined based on the slope of the linear region (3
,33
). However, we have demonstrated that cellular respiration can be followed using EPR oximetry while attaining the complete consumption of oxygen in the system and many significant parameters. There are three apparent regions in the pO2 versus time plots, i.e., a linear portion, followed by a curved part, and then a constant value, similar to the results obtained in respirometry methods. These particular areas are easily detected in the dpO2/dt versus pO2 plots as illustrated in Fig. 2 for various cell densities. The consumption rates (VO2max), calculated using the conventional approach, i.e., from the slopes of Fig. 2 A, were found to enhance with an increase in cell density. These values correspond to the plateau regions represented in Fig. 2 C. As the pO2 decreased, the rate remained relatively the same until it approached around 30 mmHg. At pO2 values <30 mmHg, the consumption rate decreases, illustrating the pO2-dependent respiration rates. This is the region reflected as a curve in the pO2 versus time plots and is described as the pO2-dependent respiration in cells (Fig. 2 A). Although this type of behavior has been reported in the HRR measurements of oxygen (2
,27
), we report for the first time that the consumption rate dependence on prevailing pO2 can be demonstrated in EPR oximetry as well.
Due to the fact that we obtained the rate versus pO2 data from EPR oximetry, we extended the method to estimate other significant parameters, as achieved with HRR. In the case of the HRR technique, specific phases of cellular respiration are analyzed and numerous parameters such as the VO2max, p50, and p0 have been determined based upon Michaelis-Menten kinetics (2
). Similarly, we utilized an approach to determine those relevant parameters. Nonetheless, there were some discrepancies in comparison to the established values. The differences were found to arise from the diffusion limitation of oxygen to the cell surface. As pointed out, the theory adopted in this work was developed based on the approaches used for electrocrystallization of metal ions, where the metal ions diffusion was found to influence the measured current features (30
). Upon accounting for the diffusion limitation, the values for the respiration parameters turned out to be in agreement with the reported values, illustrating the importance of such a correction. Therefore, the developed method for EPR oximetry is an accurate and alternative way to compute levels of respiration.
In summary, this study utilizes a diffusion model based algorithm to quantify various phases of cellular respiration. Through this technique, we acquired the maximum oxygen consumption rate, the oxygen affinity to cytochrome c oxidase of the ETC (1/p50), and the mitochondrial equilibrium pO2. To ensure that this respiration was due to the mitochondria, ETC blockers were used, which showed a substantial reduction in the respiration rate (Fig. 2 C and Table 1). Likewise, the approach here provides reliable and reproducible measurements. EPR oximetry offers a dependable, convenient way to measure pO2 while requiring only a tiny amount of sample, and the probes do not consume oxygen. When utilizing this method, the effect of the diffusion problems and the concentration gradient due to the diffusion layer must be considered and corrected. However, there are some limitations that should be taken into account. Due to the stationary condition of the samples, gravity may cause the cells and the particulate probes to settle down. In such a situation, the cells would be concentrated at the bottom of the microtube lying flat in the resonator, where semispherical diffusion zones may develop into bulk volume instead of spherical zones considered in the approach here.
To address this concern, we performed similar EPR oximetry experiments where the capillary tube was rotated 180° (upside down) within a few seconds between each acquisition, and each acquisition was
20 s. Upon analysis of the data, we found that there is no difference between the stationary sample and the sample that was rotated. Thus, settling down did not affect the respiration behavior. However, this may be a potential problem for larger cells such as cardiomyocytes. Second, the oxygen data measured by particulate-based EPR oximetry is from whole volume, as illustrated in Fig. 3. The approach used here does not quantitatively differentiate between the data obtained from the particles present within the diffusion zones and the remaining volume. This may also pose a problem in interpreting the results in some experimental conditions. Thus, further refinement of the approach here may be required, depending upon specific experimental conditions. Despite these limitations, the advancement here highlights the benefits of EPR oximetry and several improvements. This oximetry methodology can be especially valuable to measure p50 and maximum consumption rate in the presence of competitive molecules such as nitric oxide (NO), which is known to reduce the oxygen affinity in the mitochondria. With the suggested corrections and improvements, EPR oximetry provides more meaningful results in cellular respiration studies.
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
We acknowledge support from National Institutes of Health grants R21 EB004658 and R01 HL078796-02.
Submitted on May 30, 2006; accepted for publication September 8, 2006.
| REFERENCES |
|---|
|
|
|---|
2. Gnaiger, E. 2001. Bioenergetics at low oxygen: dependence of respiration and phosphorylation on oxygen and adenosine diphosphate supply. Respir. Physiol. 128:277297.[CrossRef][Medline]
3. Pandian, R. P., V. K. Kutala, N. L. Parinandi, J. L. Zweier, and P. Kuppusamy. 2003. Measurement of oxygen consumption in mouse aortic endothelial cells using a microparticulate oximetry probe. Arch. Biochem. Biophys. 420:169175.[CrossRef][Medline]
4. O'Hara, J. A., H. Hou, E. Demidenko, R. J. Springett, N. Khan, and H. M. Swartz. 2005. Simultaneous measurement of rat brain cortex PtO2 using EPR oximetry and a fluorescence fiber-optic sensor during normoxia and hyperoxia. Physiol. Meas. 26:203213.[CrossRef][Medline]
5. Ilangovan, G., T. Liebgott, V. K. Kutala, S. Petryakov, J. L. Zweier, and P. Kuppusamy. 2004. EPR oximetry in the beating heart: myocardial oxygen consumption rate as an index of postischemic recovery. Magn. Reson. Med. 51:835842.[CrossRef][Medline]
6. Ardenkjaer-Larsen, J. H., I. Laursen, I. Leunbach, G. Ehnholm, L. G. Wistrand, J. S. Petersson, and K. Golman. 1998. EPR and DNP properties of certain novel single electron contrast agents intended for oximetric imaging. J. Magn. Reson. 133:112.[CrossRef][Medline]
7. Krishna, M. C., S. English, K. Yamada, J. Yoo, R. Murugesan, N. Devasahayam, J. A. Cook, K. Golman, J. H. Ardenkjaer-Larsen, S. Subramanian, and J. B. Mitchell. 2002. Overhauser enhanced magnetic resonance imaging for tumor oximetry: coregistration of tumor anatomy and tissue oxygen concentration. Proc. Natl. Acad. Sci. USA. 99:22162221.
8. Kodibagkar, V. D., W. Cui, M. E. Merritt, and R. P. Mason. 2006. Novel (1)H NMR approach to quantitative tissue oximetry using hexamethyldisiloxane. Magn. Reson. Med. 55:743748.[CrossRef][Medline]
9. Robinson, S. P., and J. R. Griffiths. 2004. Current issues in the utility of 19F nuclear magnetic resonance methodologies for the assessment of tumour hypoxia. Philos. Trans. R. Soc. Lond. B Biol. Sci. 359:987996.[CrossRef][Medline]
10. Froncisz, W., C. S. Lai, and J. Hyde. 1985. Spin-label oximetry: kinetic study of cell respiration using a rapid-passage T1-sensitive electron spin resonance display. Proc. Natl. Acad. Sci. USA. 82:411415.
11. Ilangovan, G., H. Li, J. L. Zweier, and P. Kuppusamy. 2001. Electrochemical preparation and EPR studies of lithium phthalocyanine. Part 3: Measurements of oxygen concentration in tissues and biochemical reactions. J. Phys. Chem. B. 105:53235330.
12. Strzalka, K., T. Walczak, T. Sarna, and H. M. Swartz. 1990. Measurement of time-resolved oxygen concentration changes in photosynthetic systems by nitroxide-based EPR oximetry. Arch. Biochem. Biophys. 281:312318.[CrossRef][Medline]
13. Swartz, H. M. 2004. Using EPR to measure a critical but often unmeasured component of oxidative damage: oxygen. Antioxid. Redox. Signal. 6:677686.[CrossRef][Medline]
14. Ilangovan, G., J. L. Zweier, and P. Kuppusamy. 2000. Electrochemical preparation and EPR studies of lithium phthalocyanine. Part 2: Particle size dependent line broadening by molecular oxygen and its implications as an oximetry probe. J. Phys. Chem. B. 104:94049410.
15. Ilangovan, G., J. L. Zweier, and P. Kuppusamy. 2004. Microximetry: an EPR spectroscopy based method for simultaneous measurements of oxygen utilization and free radicals production in enzymatic reactions. Meth Enzym. 381:747762.[Medline]
16. Kuppusamy, P., and J. L. Zweier. 2004. Cardiac applications of EPR imaging. NMR Biomed. 17:226239.[CrossRef][Medline]
17. Lai, C., L. E. Hopwood, J. S. Hyde, and S. Lukiewicz. 1982. ESR studies of O2 uptake by Chinese hamster ovary cells during the cell cycle. Proc. Natl. Acad. Sci. USA. 79:11661170.
18. Ilangovan, G., S. Osinbowale, A. Bratasz, M. Bonar, A. J. Cardounel, J. L. Zweier, and P. Kuppusamy. 2004. Heat shock regulates the respiration of cardiac H9c2 cells through upregulation of nitric oxide synthase. Am. J. Physiol. Cell Physiol. 287:C1472C1481.
19. Samuni, A., A. J. Carmichael, A. Russo, J. B. Mitchell, and P. Riesz. 1986. On the spin trapping and ESR detection of oxygen-derived radicals generated inside cells. Proc. Natl. Acad. Sci. USA. 83:75937597.
20. Gallez, B., R. Debuyst, F. Dejehet, K. J. Liu, T. Walczak, F. Goda, R. Demeure, H. Taper, and H. M. Swartz. 1998. Small particles of fusinite and carbohydrate chars coated with aqueous soluble polymers: preparation and applications for in vivo EPR oximetry. Magn. Reson. Med. 40:152159.[Medline]
21. Zweier, J. L., M. Chzhan, U. Ewert, G. Schneider, and P. Kuppusamy. 1994. Development of a highly sensitive probe for measuring oxygen in biological tissues. J. Magn. Reson. B. 105:5257.[CrossRef][Medline]
22. Ilangovan, G., A. Manivannan, H. Li, H. Yanagi, J. L. Zweier, and P. Kuppusamy. 2002. A naphthalocyanine-based EPR probe for localized measurements of tissue oxygenation. Free Radic. Biol. Med. 32:139147.[CrossRef][Medline]
23. Pandian, R. P., N. L. Parinandi, G. Ilangovan, J. L. Zweier, and P. Kuppusamy. 2003. Novel particulate spin probe for targeted determination of oxygen in cells and tissues. Free Radic. Biol. Med. 35:11381148.[CrossRef][Medline]
24. Liu, K. J., P. Gast, M. Moussavi, S. W. Norby, N. Vahidi, T. Walczak, M. Wu, and H. M. Swartz. 1993. Lithium phthalocyanine: a probe for electron paramagnetic resonance oximetry in viable biological systems. Proc. Natl. Acad. Sci. USA. 90:54385442.
25. Matsumoto, A., S. Matsumoto, A. L. Sowers, J. W. Koscielniak, N. J. Trigg, P. Kuppusamy, J. B. Mitchell, S. Subramanian, M. C. Krishna, and K. Matsumoto. 2005. Absolute oxygen tension (po2) in murine fatty and muscle tissues as determined by EPR. Magn. Reson. Med. 54:15301535.[CrossRef][Medline]
26. Ilangovan, G., J. L. Zweier, and P. Kuppusamy. 2000. Electrochemical preparation and EPR studies of lithium phthalocyanine: evaluation of the nucleation and growth mechanism and evidence for potential-dependent phase formation. J. Phys. Chem. B. 104:40474059.
27. Steinlechner-Maran, R., T. Eberl, M. Kunc, R. Margreiter, and E. Gnaiger. 1996. Oxygen dependence of respiration in coupled and uncoupled endothelial cells. Am. J. Physiol. Cell Physiol. 271:C2053C2061.
28. Gnaiger, E., B. Lassnig, A. Kuznetsov, and R. Margreiter. 1998. Mitochondrial respiration in the low oxygen environment of the cell. Effect of ADP on oxygen kinetics. Biochim. Biophys. Acta. 1365:249254.[Medline]
29. Rumsey, W., C. Schlosser, E. Nuutinen, M. Robiolio, and D. F. Wilson. 1990. Cellular energetics and the oxygen dependence of respiration in cardiac myocytes isolated from adult rat. J. Biol. Chem. 265:1539215399.
30. Gunawardena, G., G. Hills, I. Montenegro, and B. Scharifker. 1982. Electrochemical nucleation. Part I. General considerations. J. Electroanal. Chem. 138:225239.[CrossRef]
31. Avrami, M. 1939. Kinetics of phase change. I General theory. J. Chem. Phys. 7:11031112.[CrossRef]
32. Vanderkooi, J., G. Maniara, T. Green, and D. F. Wilson. 1987. An optical method for measurement of dioxygen concentration based upon quenching of phosphorescence. J. Biol. Chem. 262:54765482.
33. James, P. E., S. K. Jackson, O. Y. Grinberg, and H. M. Swartz. 1995. The effects of endotoxin on oxygen consumption of various cell types in vitro: an EPR oximetry study. Free Radic. Biol. Med. 18:641647.[CrossRef][Medline]
34. Sentjurc, M., M. Cemazar, and G. Sersa. 2004. EPR oximetry of tumors in vivo in cancer therapy. Spectrochim. Acta A Mol. Biomol. Spectrosc. 60:13791385.[CrossRef][Medline]
35. Zhao, X., G. He, Y. Chen, R. Pandian, P. Kuppusamy, and J. L. Zweier. 2005. Endothelium-derived nitric oxide regulates postischemic myocardial oxygenation and oxygen consumption by modulation of mitochondrial electron transport. Circulation. 111:29662972.
This article has been cited by other articles:
![]() |
S.-R. A. Hussain, D. M. Lucas, A. J. Johnson, T. S. Lin, A. P. Bakaletz, V. X. Dang, S. Viatchenko-Karpinski, A. S. Ruppert, J. C. Byrd, P. Kuppusamy, et al. Flavopiridol causes early mitochondrial damage in chronic lymphocytic leukemia cells with impaired oxygen consumption and mobilization of intracellular calcium Blood, March 15, 2008; 111(6): 3190 - 3199. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||