| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
The concept of FCS is based on the correlation analysis of fluorescence fluctuations in a confined observation volume. The sensitivity of the technique to detect binding of two or more components depends on the relative change in mass upon binding. For a multicomponent system consisting of reactants and products labeled with the same fluorescent dye, the only way of differentiating the product from the reactant is when the product has a molecular mass that differs from the reactants by a factor of at least 48 (12
). This in turn shifts the correlation curve to higher diffusion times by up to a factor of 2 given by the Stokes-Einstein equation for spherical diffusing particles (
). By separately labeling the reactants with differently emitting fluorophores, the labels can be simultaneously excited with two different laser lines and detected in separate channels. The signals from both detector channels can then be cross correlated and the doubly labeled products easily distinguished from the singly labeled reactants independent of their mass. Earlier cross-correlation systems have made use of light scattering or a combination with fluorescence to measure their cross-correlation functions and determine rotational diffusion and association-dissociation dynamics (13
,14
). In dual-beam fluorescence cross-correlation spectroscopy, the setup consisting of two spatially separated focal points has been applied to characterize flow systems (15
). Dual-color fluorescence cross-correlation spectroscopy was first experimentally realized by Schwille et al. to measure nucleic acid hybridizations (16
,17
). The potential of this technique to effectively measure biomolecular interactions has expanded its applications to detecting PCR complexes (18
), monitoring enzyme kinetics (19
,20
), measuring protein-DNA interactions (21
), and the analysis of live cells (22
,23
). Cross correlation has also been combined with FCS and Förster resonance energy transfer (FRET) for global data analysis (24
). Two-photon excitation laser sources have been used to overcome the difficulty of aligning two laser beams to the same confocal volume (25
). Increased axial resolution from a more confined focal spot reduces background fluorescence and photobleaching making it suitable for in vivo studies (26
,27
). Recently, two-photon excitation has achieved the excitation of up to three dyes simultaneously to perform triple-color coincidence analysis (28
). However, the high cost of a high power femtosecond laser source and lower emission rates limit its potential applications. A less expensive method is the excitation of two or more fluorescent dyes by one-photon excitation with single laser wavelength by using dyes with similar excitation spectra but spectrally different emission characteristics. With the recent advent of long Stokes shift fluorophores such as nanocrystal quantum dots (29
,30
) and tandem dyes (31
,32
) or MegaStokes dyes (www.dyomics.com), multicolor imaging using a single laser wavelength for excitation has been achieved (33
). Using single excitation wavelength in dual color fluorescence cross-correlation spectroscopy (SW-FCCS) was realized by Hwang and Wohland (34
) to detect ligand-receptor binding. The resolution of SW-FCCS was explored by using dyes with similar emission characteristics such as fluorescein and tetramethylrhodamine (35
).
In this article, we present an extension of this technique to multicolor SW-FCCS. Using a single laser wavelength to excite up to three differently emitting dyes simultaneously, we measured the binding of green ligand biotin-4-fluorescein (BF) and yellow ligand R-phycoerythrin biotin (BPE) to red receptor Alexa Fluor 647-R-phycoerythrin-streptavidin (AXSA). We formulated a theory to explain the pairwise cross correlations green x red (Ggr(
)), yellow x red (Gyr(
)), and green x yellow (Ggy(
)) for this system. For ways to extend the theory to take into account the sample impurities and labeling ratios, refer to Weidemann (36
) and Hwang (35
). It is shown that even with a higher amount of cross talk between three differently emitting fluorescent labels all excited at the same wavelength, SW-FCCS is capable of discriminating bound complexes from free reactants by more than 6 SD difference in the cross-correlation amplitudes. The capability of distinguishing trimers, dimers, and monomers regardless of their molecular weight, when performed with appropriate negative controls, opens up new possibilities of studying higher order interactions in complex molecular systems.
| THEORY |
|---|
|
|
|---|
![]() | (1) |
j for cross correlation of two channels.
In the following we assume we have a system consisting of R, a red fluorescent receptor with multiple binding sites for one ligand, and Lg and Ly, the ligand that is either labeled with a green or yellow emitting fluorophore. Considering a solution of receptor and ligands, free ligands Lf will bind with free receptors Rf to form complex RLn at equilibrium binding where n is the number of bound ligands on R. We assume here that each complex formed consists of one receptor with several ligands specifically bound, therefore excluding oligomerization of this receptor.
![]() |
![]() |
![]() |
![]() |
We assume that each binding site has the same affinity. If we disregard the multiplicity of the binding sites, the dissociation constant Kd for each individual binding site is then given by the concentration of reactants divided by the products,
![]() | (2) |
To take account of the multiple binding sites per receptor, binomial coefficients are introduced to describe the possibility of n ligands binding to nt binding sites (37
). The concentrations of free receptors and ligands are thus related to the total concentrations of receptor Rt and ligand Lt minus the sum of all bound receptors and ligands, respectively.
![]() | (3) |
![]() | (4) |
The concentrations of the complexes RLn, Lf, and Rf at binding equilibrium can then be numerically determined by simultaneously solving Eqs. 24.
The total concentration of ligand Lt consists of the ligands Lg and Ly. The probability of encountering either ligand Lg or Ly to form a complex with a receptor is given by their mol fractions
![]() | (5) |
![]() | (6) |
Consider a receptor with nt binding sites and n fluorescent ligands bound, of which ng are Lg ligands and ny are Ly ligands (ng
n
nt). In this case we have to account for the number of possibilities how to distribute firstly n ligands over nt binding sites and secondly ng ligands Lg to the n bound sites. The distribution of ny ligands Ly to the ny (= n ng) remaining binding sites has then only one possibility. The concentration of a complex with n bound ligands becomes
![]() | (7) |
The first binomial coefficient describes the distribution of n bound ligands over the total number of binding sites nt and the second coefficient is the distribution of Lg over the total number of bound ligands. Equations 37 will be used to calculate the cross-correlation amplitude as shown below.
The time dependent total fluorescence signal Fi(t) in detection channel i is the sum of all fluorescent species (s = L, R, RL) contributing to the signal. It is determined by their fluorescence yields
(often expressed as counts per molecule per second), and the time dependent number of particles NAVeffC(t) in the effective observation volume Veff. NA is Avogadro's number and C(t) represents the time-dependent values of the averages Rf, Lf, or RL(n,ng) as defined in Eqs. 3, 4, and 7, respectively. All possible species that contribute with
via cross talk into the detection channels are taken into account.
![]() | (8) |
The first term represents the total free ligands with different fluorescence yields QLg, QLy for ligands labeled with different fluorophores. The second term represents the free receptor and the third term denotes the complex itself with both types of ligands bound to the receptor where the fluorescence yield contribution of Lg and Ly are proportional to the number of ligands bound; i.e., ng x QLg and ny x QLy. Changes in fluorescence yields upon binding via processes such as quenching or FRET are taken into account by the factors qLg, qLy where q = 1.0 if there is no change in fluorescence yield. Assuming that the emission spectra do not undergo any shifts in wavelength, qLg, qLy are the same in all channels.
Because we are only interested in observing relative changes in the cross-correlation amplitudes we calculate the cross-correlation function at
= 0. The fluorescence yield factor
is obtained by the product of fluorescence yields in the cross-correlated channels. It determines the weighting factor contributing from various species to the cross-correlation amplitude, respectively
![]() | (9) |
![]() | (10) |
![]() | (11) |
By substituting Eq. 8 for two detection channels i x j (where i x j can be any combination pair of detection channels) into the cross-correlation function in Eq. 1, and assuming a three-dimensional (3D) Gaussian illumination intensity profile, the cross-correlation amplitude then becomes
![]() | (12) |
The cross-correlation function for the negative control does not include binding of ligand to receptor therefore only cross talk is contributing to the function
![]() | (13) |
Equation 12 is based on the assumption that both Lg and Ly bind to R to form a trimer. But in the case where only one type of ligand is bound to R and the other remains free, the cross-correlation amplitude will resemble the positive control function for the bound ligand and receptor and the negative control function for the free ligand. In this case we have two possible cases.
![]() |
In the case where all ligands binding to the red receptor R are green ligands Lg (n = ng) and the yellow ligands Ly remain free, the probability of binding Lg becomes 1. The concentration of complex RLg from Eq. 7 becomes
![]() | (14) |
All of the complexes formed consist of only RLg, therefore there is no fraction of Ly contributing to the concentration of free ligands Lf after binding (fLy = 0) nor to the formation of the complex RLg. Instead, all of Ly (= fLyLt) remains as completely free ligands but still contribute to the cross-correlation function between g x r via cross talk. These conditions are substituted into the cross-correlation function in Eq. 12 to obtain Gij(0) as a positive control for g x r
![]() | (15) |
The first two terms in the numerator denote free Lg and total nonbinding Ly, respectively. The third and fourth terms represent the contribution from free R and complexes RLg, respectively. The fluorescence yield factors
for species s (where s = Lg or Ly or R) are described by
![]() | (16) |
![]() | (17) |
Because there are no bound complexes formed between Lg and Ly or R and Ly, Ggy(0) and Gyr(0) represent the negative controls and any contribution from the RLg ligand-receptor complexes comes via cross talk.
![]() |
In the case where all ligands bound to red R are yellow Ly (n = ny) and green Lg remain free, the probability of binding Ly becomes 1. Equation 14 then refers to the concentration of complex RLy formed and the cross correlations can be derived from Eqs. 1517 by exchanging indices g and y.
Application of theory to biotin-streptavidin binding system
The biochemical system we present here consists of the red AXSA receptor R with up to four specific binding sites (nt = 4) for biotin ligand that is differently labeled with fluorescein (Lg) and R-phycoerythrin (Ly). In this case, we vary the number of ng and ny ligands bound to R from 0 to 4, such that the complex is always at full binding with all streptavidin binding sites occupied with biotin (see Materials and Methods). The cross-correlation functions for the positive and negative controls Gij can be any permutations of detection channels in the green, yellow, and red, corresponding to the colors at the emission maximum of the binding species.
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
30 min and all samples were prepared in PBS buffer pH 7.4 (Sigma-Aldrich Chemie GmbH, Buchs, Switzerland). | RESULTS AND DISCUSSION |
|---|
|
|
|---|
In this work, fluorescein, R-phycoerythrin, a 240-kDa phycobiliprotein, and Alexa Fluor 647-R-phycoerythrin, a tandem dye, were selected for SW-FCCS due to their overlapping excitation spectra and minimal cross talk. Tandem dyes are shorter wavelength emitting dyes such as phycobiliproteins linked to a red emitting dye, e.g., Alexa Fluor 647 or Cy5. They are excited at 488 nm and due to strong fluorescence energy transfer between the proteins and the red emitting dyes; emission is mainly detected in the red. Their molar extinction coefficients at 488 nm are shown in Table 1. The series of installed dichroic mirrors and bandpass filters effectively separates the emission wavelengths yet provide high count rates. Their absorbance and emission spectra are shown in Fig. 2, A and B. The fluorescence yields Q in each channel were calculated from the photon counts per second divided by the number of molecules determined from the amplitude of the autocorrelation function. The Q-values were corrected for background from Raman scattering of water in the yellow and Rayleigh scattering of the laser line (Table 1).
|
|
75%) due to stronger, faster, and noncooperative binding between the less hindered biotin-4-fluorescein and streptavidin (38
Other fluorophore combinations have also been considered for SW-FCCS. Organic dye pairs and quantum dots have been measured previously with this technique (34
,35
). In particular, quantum dots have become a convenient choice for multicolor detection due to high quantum yield and continuously tunable emission spectra that can all be excited with one laser line. In these experiments, organic dyes were selected instead of quantum dots due to the relative ease of control of binding ratios of biotin to streptavidin. Commercially available quantum dots are developed mainly for imaging purposes and usually have high protein to label conjugation (1015 streptavidin molecules to 1 quantum dot), making the binding concentrations difficult to manipulate between three binding partners. In addition, aggregation problem with quantum dots has been previously reported (34
). Thus although quantum dots are better in terms of photostability and brightness, their aggregation in solution makes it difficult to unambiguously determine interactions. Alternatively, tandem dyes have been widely used in flow cytometric applications for simultaneous detection of multiple fluorophores excited with a single laser. We use these bright dyes for the same advantages for the application to SW-FCCS.
The large sizes of phycobiliproteins may pose problems as labels for smaller-sized target biomolecules. Therefore, a range of other possible dyes that could be used for in vitro and in vivo SW-FCCS, including fluorescent proteins and Megastokes dyes, have been measured. Their fluorescence yields in the different channels are listed with their filter sets in the Supplementary Material.
Calibration experiments
Calibration measurements were performed with Fluorescein (Invitrogen, 1 nM) in the green and yellow channels and AXSA in the red channel. Autocorrelation functions of BF, BPE, and AXSA were measured with increasing laser power from 50 to 500 µW to investigate the change of photon count rates per particle and triplet state population against excitation intensity. The diffusion times of the different molecules showed deviations at higher excitation intensities; however, this change depended on the molecular species and was minimal in our setup below 100 µW for all three species. The laser power of 50 µW was selected for minimal optical saturation and photobleaching of the dyes, as well as optimal count rates and low triplet fraction obtained between all three fluorescent dyes. Ten correlation functions measured for 10s were taken for all auto and cross-correlation functions. All correlation curves were fitted with the Levenberg-Marquadt fitting algorithm in Igor Pro (v4.0 Wavemetrics, Portland, OR). A fitting model for one-component diffusion model with triplet state (39
) was used for the autocorrelation curves of BF and AXSA. The BPE autocorrelation curves were fitted with the one-component diffusion model with two triplet states where the first decay corresponds to the singlet-triplet lifetime in the microsecond timescale (40
). The second decay in the tens of microseconds timescale could be due to other photodynamic process involved with R-phycoerythrin. The normalized autocorrelation functions and their fits are shown in Fig. 2 C. Fluorescein with a relative molecular weight of 376.3 Da and a reported diffusion coefficient D of 3.0 x 106 cm2/s (6
) was used as a standard dye to characterize the excitation volume. The beam waist radius wo of 0.29 µm is calculated from the equation
where the average diffusion time
d of 70.6 µs of fluorescein was determined from the fits of the autocorrelation functions. The diffusion coefficients of BF, BPE, and AXSA at 2.6 x 106 cm2/s, 2.2 x 106 cm2/s, and 1.7 x 107 cm2/s, respectively, are calculated from the beam waist and the respective diffusion times that are obtained from the fits in Fig. 2 C. The relative molecular weights of the molecules are then determined from the Stokes'-Einstein equation, which assumes spherical molecules, to be 547.6 Da, 964 kDa, and 2,100 kDa, respectively. The experimentally determined relative molecular weight of BF is similar to the literature value of 644.7 Da. However, the molecular weights of BPE and AXSA are much higher than the reported values of 240 and 294 kDa. This is most likely due to the nonspherical shapes of the molecules (41
) that the equation does not take into account. A deviation from the spherical shape will lead to a decrease in the diffusion coefficient (9
).
The blinking times of the triplet states for different labels are uncorrelated to each other despite being bound to the same complex. Thus, the triplet fractions that are detected in the autocorrelation functions are not detectable in the cross-correlation functions. The triplet state will reduce the count rate of the dye but the total number of molecules in the auto/cross-correlation functions remains constant. All the cross-correlation functions could be fitted sufficiently well with the one-component diffusion model and the structure parameter K (6
) was obtained as 1.02 ± 0.02 for Ggr(
), 1.06 ± 0.18 for Gyr(
), and 3.45 ± 1.45 for Ggy(
). The average K parameter was then fixed at 2 for all future cross-correlation fits.
Experimental results of biotin-streptavidin binding
In the following discussion, we will refer to AXSA as R, BPE as Ly, and BF as Lg. In general the cross-correlation functions exhibit the following trends. Under otherwise equal conditions the positive controls will have higher cross-correlation amplitudes due to complexes with multiple colors than the negative controls. The negative controls show only weak cross correlations due to the cross talk of the fluorophores into different channels. But both negative and positive controls will show decreasing amplitudes with increasing numbers of complexes or ligands and receptors.
Correlations of triple-color complexes
At any one time, three different components were mixed together in one sample aliquot and Ggr(
), Gyr(
), Ggy(
) were measured simultaneously. The cross-correlation functions and their fits for a ligand/receptor concentration ratio Lg/Ly/R = 2:2:1 are shown in Fig. 3, AC. The negative control amplitudes are due to cross talk between the respective channels but the positive control amplitudes are clearly higher due to the bound species. The amplitudes for each ligand/receptor ratio for positive and negative controls are plotted in Fig. 4, AC. Fig. 4 A shows Ggr(0) decreasing with 04 Lg and 40 Ly molecules bound to R due to the formation of complexes containing R and Lg (Eq. 12). In the case of the negative control (Eq. 13) where there is an absence of receptor-ligand complexes, the curve decreases sharply. The contribution to the amplitude is from cross talk, which is analogous to the autocorrelation curves. Likewise for Fig. 4 B, Gyr(0) decrease toward increasing concentration of complexes containing R and Ly. Although there is no direct binding between Lg and Ly, the binding through an intermediate receptor R gives rise to Ggr(
) as shown in Fig. 4 C. In this case, the positive control amplitude drops to a minimum toward the center of the curve where a maximum of complexes containing Lg and Ly is reached due to the presence of equal concentrations of Lg and Ly. As predicted, the correlation amplitudes are smaller for negative controls compared with positive controls in all cases.
|
|
|
2-value no more than 50% from the minimum best fit value. The negative control curves as shown by the shaded regions in Fig. 4, AC, are fitted (Eq. 12) to give Veff 1.12.1 femtoliters (Table 2). The fitted Veff values generally increase with the emission wavelengths detected from the fluorescent dyes; i.e., Veff (Ggy(0))
Veff (Ggr(0))
Veff (Gyr(0)). Positive controls are modeled with Eq. 12 to give the range of Veff and Kd values (Table 2) as shown by the shaded regions in Fig. 4.
The obtained Kd values are well above the predicted 1015 M. One reason for this is that the experiments were performed with sample concentrations in the nanomolar range (sensitivity limit of FCS), which makes it difficult to determine Kd values at six orders below this concentration limit. The Kd values determined from these fits, however, are close to FCS measurements done on the same binding system at similar concentration levels (42
). Another reason could be due to ligand and/or receptor impurities that cause the binding curve to alter its slope. Labeling ratios between protein and label is another possible factor affecting the slope of the binding curve. Having more than one label increases the brightness of the product and this contributes to the autocorrelation amplitude with the square of its fluorescence yield and the cross-correlation function with the product of the fluorescence yields. Here we assume that all labeling ratios for ligands and receptor are 1:1 as stated by the supplier, and we use the average photon counts per particle per second to model the curves.
Nevertheless, it is the magnitude of difference in amplitudes between the positive and negative control curves that resolves the binding of two components. To determine complex formation, we demand that the difference between the positive (+) and negative () control should be at least 6 SD; i.e.,
Factors that affect this difference include fluorescence yields, cross talk, and impurities (35
). Although Gyr(0) has a smaller difference because of larger cross talk between Ly and R from yellow emitting R-phycoerythrin molecules in R conjugates, the differences between all positive and negative control curves are more than 6 SD. Hence by measuring multiple cross-correlation curves with a single sample at one Lg/Ly/R concentration ratio, it is possible to determine binding between the different biomolecules.
The most significant difference between positive and negative controls are found when working at stoichiometric concentrations (see Supplementary Material). When measuring biotin/streptavidin ratios above 4:1 increasing free Lg molecules contribute larger background to the cross-correlation function, decreasing the amplitudes sharply toward the negative control, thus making binding irresolvable (35
).
Correlations of complexes with alternate ligand binding
The difference in amplitudes between the positive and negative controls of Fig. 4, A and B, show that binding occurs between both Lg and Ly ligands with receptors. However, this does not prove the existence of complexes formed between Lg, Ly, and R simultaneously. Only Ggy(
) confirms the existence of complexes containing Lg, Ly, and R. However, this conclusion is based on the assumption that the components are known beforehand and the nature of binding is identified. In this case, it is known that biotin binds specifically to streptavidin and does not dimerize with itself. In fact, Ggy(
) may even be sufficient to determine complexation between Lg, Ly, and R here (43
). In cases where the nature of binding is not known, additional negative controls will have to be performed to confirm that complexes RLgLy are formed. We have further performed these negative controls where only one ligand at a time is bound to the receptor and the binding of the second ligand is inhibited. The cross-correlation curves for a ligand/receptor concentration ratio of Lg/Ly/R = 2:2:1 are shown in Fig. 3, DF. The binding and nonbinding cases are clearly distinguishable for Ggr(
) and Gyr(
) (Fig. 3, D and E) where the interacting species possess the higher cross-correlation amplitudes. The similar cross-correlation curves for Ggy(
) (Fig. 3 F) demonstrate that the ligands are not complexed either directly or indirectly via streptavidin. The cross-correlation amplitudes over the whole range of ligand/receptor ratios are plotted in Fig. 5, AC.
|
Case 2: Binding between Ly and R with Lg inhibited shows the Gyr(0) values (Fig. 5 B, solid circles) eventually decreasing at higher Ly concentrations, as expected. Conversely, the Ggr(0) and Ggy(0) negative controls curves (Fig. 5, A and C, solid circles) decrease rapidly to lower amplitudes similar to the negative control curves in Fig. 4, A and C.
In Fig. 5 A the cross correlations Ggr(0) have the same amplitudes when no ligand Lg is present. The same effect can be observed in Fig. 5 B, where the cross-correlation amplitudes are similar when no ligand Ly is present. For all other cases the cross correlations representing the interacting molecules are always higher in amplitude than the cross correlation representing the noninteracting molecules. In Fig. 5 C, the Ggy(0) values are similar, no matter whether Ly or Lg is inhibited from binding. The curves are comparable to the negative control of Fig. 4 C because no complexes containing Lg and Ly simultaneously exist. In addition, it should be noted that if all three species are present, the amplitudes of the cross-correlation functions are always highest for the case of interacting molecules. For instance, when inhibiting Ly from binding (open circles) the highest amplitudes are found in Fig. 5 A, the Ggr(0) channel. Conversely, when inhibiting Lg from binding (solid circles) the highest amplitudes are found in the Gyr(0) channel (Fig. 5 B). The triple pairwise cross correlations directly show which molecules are interacting, thus substantiate the initial results from Fig. 4, AC, that trimers are indeed formed between both ligands and the receptor.
Fitting analysis of complexes with alternate ligand binding
Additional negative control curves with Ly or Lg binding inhibited are also modeled with Eq. 15 to give the best fit range of Veff and Kd within 50% of the lowest
2 (shaded regions, Fig. 5, AC). The exception is the Ggr(0) curve representing Ly binding and Lg inhibition (Fig. 5 A, solid circles), which could not be fitted to give Kd values within the limits of 1015106 M. This is due to the fact that Ly does have negligible cross talk into the green channel (see Table 1) and thus the RLy complexes do not contribute to the cross-correlation function and a determination of a Kd value is not possible. Therefore, the data points are fitted instead with Eq. 13 where cross talk from free Ly and RLy complexes into the green channel could be taken to be negligible. The fitting analysis yield Kd values of biotin-streptavidin binding from 1011 to 108 M (Table 2). The difference between Ggr(0) positive and negative control curve is more than 6 SD (Fig. 5 A). This excludes the first point that does not have any Lg present and consists of only background from RLy complexes. Gyr(0) on the other hand fulfills the condition for binding only at higher concentrations of Ly/R (Fig. 5 B). This is because at low Ly/R concentrations, free Lg molecules contribute to the cross correlation as background via cross talk, making binding indistinguishable. Both the negative controls with Ly or Lg inhibited have no contribution to Ggy(
) from simultaneous binding of Ly and Lg to R (Fig. 5 C). Therefore both curves at low amplitudes show little difference from each other and the contribution to the cross-correlation amplitudes come mainly from cross talk of the fluorophores.
Limitations of SW-FCCS
Influence of Kd on cross correlations
The effect of Kd on cross-correlation amplitudes were calculated from the models as a function of ligand/receptor concentration ratios. Kd values were varied from 1015 to 107 M at full binding conditions (Fig. 6, AC). The changes in cross-correlation amplitudes of the negative control curves are due to cross talk in both channels. The positive control curves decrease toward higher ligand concentrations for Fig. 6, A and B, but remain relatively constant for Fig. 6 C. At higher Kd values (107 M) where more free reactants contribute to the cross-correlation functions and fewer complexes are formed, the separations of amplitudes between the positive and negative control curves diminish. Thus the limit of measurable Kd is reached when the positive and negative control have a difference that is smaller than 6 SD. This in turn is dependent on the count rates of the different reactants and their cross talk into the different channels.
|
Stoichiometry determination
The determination of stoichiometry with SW-FCCS has been demonstrated previously for direct binding with dual-color biomolecules (34
,35
). In this case for triple-color cross correlations, the ligands bind indirectly over a common interaction partner. With higher background due to an additional third color, the stoichiometry can still be determined in a similar way depending on the Kd values of the ligands. By varying each ligand Lg and Ly across a range of concentrations while maintaining the receptor concentration constant, a plot of
with Lg and Ly will reveal the stoichiometry of the binding system. Various simulations of different stoichiometric ratios and further explanations are presented in the Supplementary Material.
Applications of multicolor SW-FCCS
The extension of FCCS to three colors diminishes the signal/noise ratio of the measurements because a narrower wavelength range is available for each channel and cross talk between the channels will be larger. Nevertheless, the extension to three colors is important for at least two reasons. Firstly, the biological variability between cells is often so high that any correlations between distinct biomolecules are hidden and can only be detected when all relevant molecules are observed simultaneously in a cell. Secondly, complex biochemical reactions in cellular systems involve higher order molecular interactions. These interactions consist of temporal association and dissociation reactions that multicolor SW-FCCS has the potential to detect and monitor. For instance, the detection of binding of the various proteins involved in signaling complexes in a cellular environment over time can only be followed when the different interaction partners are labeled. To be able to detect these intermediate complexes, the lifetimes of these complexes have to be longer than the time it takes for the complexed molecule to diffuse through the observation volume and the characteristic times of the interactions have to be of the same order or longer than our measurement time, which is limited by
1 s for FC(C)S. If that is the case, SW-FCCS measurements with three colors can differentiate between trimers, dimers, and monomers and can elucidate temporal sequence of biological interactions.
| CONCLUSION |
|---|
|
|
|---|
Multicolor SW-FCCS provides a fast and convenient method to offer yes or no answers to interacting biochemical systems, determine an upper Kd limit, and determine the stoichiometry of binding. Existing FCS optical setups can be easily modified to perform SW-FCCS by including three detectors at the detection pathway while keeping the excitation path unchanged with one CW laser. Multiple laser excitation setups, in contrast, involve the complicated alignment of several laser beams in 3D to the same excitation volume and suffer from artifacts of nonideal overlap of excitation volumes that arise because of chromatic aberrations. Compared to multiphoton FCCS, SW-FCCS utilizes one-photon excitation that not only uses less expensive lasers but also offers higher count rates per particle and a better signal/noise ratio (44
). In addition, recent advances in the setup of the detection pathways by using dispersive elements further simplify the setup and offer a simpler way of choosing wavelength ranges for detection and thus minimization of spectral cross talk (45
,46
).
SW-FCCS uses fluorophores that require similar excitation spectra but spectrally different emission characteristics with minimal cross talk. It has been shown to work with tandem dyes, quantum dots, and even with spectrally similar organic dyes. In Supplementary Material we give as well experimental count rates per particle for fluorescent proteins and small organic dyes with large Stokes shifts, both of which are potential fluorophores for this technique. The high sensitivity of FCS and its ability to probe spatial and temporal reactions coupled with the capability to detect multicolor labels simultaneously using single laser excitation, provides the opportunity to study higher order complex formation and molecular networks with a good signal/noise ratio in live cells. Potential applications include interactions involving membrane receptors, intracellular signaling proteins, and DNA transcription factors in signaling networks.
| SUPPLEMENTARY MATERIAL |
|---|
|
|
|---|
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
This work was supported by the Academic Research Fund of the National University of Singapore and funding from Ecole Polytechnique Fédérale de Lausanne.
Submitted on September 8, 2005; accepted for publication April 5, 2006.
| REFERENCES |
|---|
|
|
|---|
2. Elson, E. L., and D. Magde. 1974. Fluorescence correlation spectroscopy. I. Conceptual basis and theory. Biopolymers. 13:127.[CrossRef]
3. Koppel, D. E. 1974. Statistical accuracy in fluorescence correlation spectroscopy. Phys. Rev. A. 10:19381945.[CrossRef]
4. Qian, H., and E. Elson. 1991. Analysis of confocal laser-microscope optics for 3-D fluorescence correlation spectroscopy. Appl. Opt. 30:11851195.
5. Rigler, R., U. Mets, J. Widengren, and P. Kask. 1993. Fluorescence correlation spectroscopy with high count rate and low-background: analysis of translational diffusion. Eur. Biophys. J. 22:169175.
6. Thompson, N. L. 1991. Fluorescence correlation spectroscopy. In Topics in Fluorescence Spectroscopy, Vol. 1: Techniques. J. R. Lakowicz, editor. Plenum Press, New York. 337378.
7. Rauer, B., E. Neumann, J. Widengren, and R. Rigler. 1996. Fluorescence correlation spectrometry of the interaction kinetics of tetramethylrhodamin
-bungarotoxin with Torpedo califonica acetylchoine receptor. Biophys. Chem. 58:312.[CrossRef][Medline]
8. Van Craenenbroeck, E., and Y. Engelborghs. 1999. Quantitative characterization of the binding of fluorescently labeled colchicine to tubulin in vitro using fluorescence correlation spectroscopy. Biochemistry. 38:50825088.[CrossRef][Medline]
9. Wohland, T., K. Friedrich, R. Hovius, and H. Vogel. 1999. Study of ligand receptor interactions by fluorescence correlation spectroscopy with different fluorophores: evidence that the homopentameric 5-hydroxytryptamine type 3As receptor binds only one ligand. Biochemistry. 38:86718681.[CrossRef][Medline]
10. Gosch, M., H. Blom, J. Holm, T. Heino, and R. Rigler. 2000. Hydrodynamic flow profiling in microchannel structures by single molecule fluorescence correlation spectroscopy. Anal. Chem. 72:32603265.[Medline]
11. Widengren, J., and R. Rigler. 1998. Review: fluorescence correlation spectroscopy as a tool to investigate chemical reactions in solutions and on cell surfaces. Cell. Mol. Biol. 44:857879.[Medline]
12. Meseth, U., T. Wohland, R. Rigler, and H. Vogel. 1999. Resolution of fluorescence correlation measurements. Biophys. J. 76:16191631.
13. Kam, Z., and R. Rigler. 1982. Cross-correlation laser scattering. Biophys. J. 39:713.
14. Ricka, J., and T. Binkert. 1989. Direct measurement of a distinct correlation-function by fluorescence cross-correlation. Phys. Rev. A. 39:26462652.[CrossRef][Medline]
15. Brinkmeier, M., K. Dorre, J. Stephan, and M. Eigen. 1999. Two beam cross correlation: a method to characterize transport phenomena in micrometer-sized structures. Anal. Chem. 71:609616.[CrossRef]
16. Schwille, P., F. Meyer-Almes, and R. Rigler. 1997. Dual-color fluorescence cross-correlation spectroscopy for multicomponent diffusional analysis in solution. Biophys. J. 72:18781886.
17. Schwille, P. 2001. Cross-correlation analysis in FCS. In Fluorescence Correlation Spectroscopy: Theory and Applications. E. L. Elson and R. Rigler, editors. Springer, Berlin, Germany. 360378.
18. Rigler, R., Z. Foldespapp, F. J. MeyerAlmes, C. Sammet, M. Volcker, and A. Schnetz. 1998. Fluorescence cross-correlation: a new concept for polymerase chain reaction. J. Biotech. 63:97109.[CrossRef][Medline]
19. Kettling, U., A. Koltermann, P. Schwille, and M. Eigen. 1998. Real-time enzyme kinetics monitored by dual-color fluorescence cross-correlation spectroscopy. Proc. Natl. Acad. Sci. USA. 95:14161420.
20. Rarbach, M., U. Kettling, A. Koltermann, and M. Eigen. 2001. Dual-color fluorescence cross-correlation spectroscopy for monitoring the kinetics of enzyme-catalyzed reactions. Methods. 24:104116.[CrossRef][Medline]
21. Rippe, K. 2000. Simultaneous binding of two DNA duplexes to the NtrC-enhancer complex studied by two-color fluorescence cross-correlation spectroscopy. Biochemistry. 39:21312139.[CrossRef][Medline]
22. Bacia, K., I. V. Majoul, and P. Schwille. 2002. Probing the endocytic pathway in live cells using dual-color fluorescence cross-correlation analysis. Biophys. J. 83:11841193.
23. Saito, K., I. Wada, M. Tamura, and M. Kinjo. 2005. Direct detection of caspase-3 activation in single live cells by cross-correlation analysis. Biochem. Biophys. Res. Commun. 324:849854.
24. Eggeling, C., P. Kask, D. Winkler, and S. Jager. 2005. Rapid analysis of Förster resonance energy transfer by two-color global fluorescence correlation spectroscopy: trypsin proteinase reaction. Biophys. J. 89:605618.
25. Heinze, K. G., A. Koltermann, and P. Schwille. 2000. Simultaneous two-photon excitation of distinct labels for dual-color fluorescence crosscorrelation analysis. Proc. Natl. Acad. Sci. USA. 97:1037710382.
26. Berland, K. M., P. T. C. So, and E. Gratton. 1995. Two-photon fluorescence correlation spectroscopy: method and application to the intracellular environment. Biophys. J. 68:694701.
27. Schwille, P., U. Haupts, S. Maiti, and W. W. Webb. 1999. Molecular dynamics in living cells observed by fluorescence correlation spectroscopy with one- and two-photon excitation. Biophys. J. 77:22512265.