| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||


* Department of Biomedical Engineering, and
Department of Physics, Duke University, Durham, North Carolina
Correspondence: Address reprint requests to Ninita Brown, Department of Biomedical Engineering, 136 Hudson Hall, Duke University, Durham, NC 27708. E-mail: ninita.brown{at}duke.edu.
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
A ratiometric signal is produced by taking the ratio of two simultaneously collected optical signals from wavelengths at opposite sides of the dye emission spectrum. This method relies on the fact that the entire emission spectrum shifts toward lower wavelengths during an AP (27
), but the total emitted power stays constant. Knisley et al. applied ratiometry to optical recordings from the rabbit epicardium using a system that included a blue laser, a spectrograph, and a photodiode array (25
). Caldwell et al. applied both ratiometry and subtraction to recordings from a fiber-based system that included a blue laser, a dichroic mirror, and multiple discrete photodiodes (2
,3
). Byars et al. also performed recording using a fiber-based system consisting of a mercury lamp, a dichroic mirror, and a photodiode array (1
). The recordings from the fiber-based systems still contained MA that had to be further reduced using BDM.
The previously reported fiber-based ratiometric systems used a dichroic mirror to separate the dye emission light into two fixed wavelength bands (1
5
). The two bands were then detected individually using separate channels. The fluorescent intensity from each of these bands was not equal, causing incomplete cancellation of the MA (28
).
Our system uses a diffraction grating, which converts the wavelength-dependent shift during an AP into a spatial shift in the reflected intensity distribution. Changes in this intensity are detected by a split photodetector, where the split is centered near the peak of the emission spectrum. The increase in intensity at shorter wavelengths is measured on one photodiode cell, and the decrease in intensity at longer wavelengths is measured on the other photodiode cell. The specific wavelengths of detection are dependent on the location of the split photodetector with respect to the light reflected from the grating. By making small adjustments in the location of the split detector, the intensity variations due to motion can be minimized.
This work describes the simulations of the channel using the optical modeling software ZEMAX, the construction and alignment of the device with performance measurements, and the experimental testing in five rabbit studies. Our channel is compact and inexpensive compared to the previous systems. The total cost decreases by using a diffraction grating and a split photodetector, rather than a spectrograph and a detector array, respectively. This allows the channel to be expanded economically into a multichannel system.
| METHODS |
|---|
|
|
|---|
|
peak = 455 nm) was used because the absorption of the electrochromatic dye di-4-ANEPPS is greatest at blue wavelengths, close to the dye's absorption maximum (468 nm in bilayer) (29The fiber from the LED was terminated with a ferrule connector (FC), which was connected to the channel though a panel-mounted FC-FC connector (F1-FC6L; FIS, Oriskany, NY). This feature allowed for simple switching of the light source. Inside the panel, an additional FC-terminated fiber was connected to an aspheric lens collimation package (F230FC-A; Thorlabs, Newton, NJ), which collimated the light emerging from the fiber. Next, the light reflected off a dichroic mirror (515dcxr; Chroma Technology, Rockingham, VT), which transmitted light at wavelengths longer than 515 nm and reflected light at shorter wavelengths (B). The reflected illumination light was then focused by a second aspheric lens collimation package into another 125-µm core diameter tissue fiber (C). This optical fiber was terminated with an FC connector at the collimation package end but was left bare at the other end to insert into and illuminate the tissue. This light caused the dye, di-4-ANEPPS, to fluoresce from the cell membrane (D).
The detection pathway
The tissue fiber both transmitted the illumination light and received the emitted light. The bare end of the fiber was polished at a 45° angle to decrease back reflection from the excitation light (3
,32
) (E). The light collected by the fiber was transmitted through the dichroic mirror (F). The light was filtered to only transmit wavelengths above 530 nm using a long-pass filter (HQ530lp; Chroma Technology). This filter reduced reflected and scattered excitation light (G). The filtered light was reflected off an aluminum-coated diffraction grating (1200 grooves/mm, 43210; Edmund Industrial Optics, Barrington, NJ) (H). An additional aspheric lens (350330A; Thorlabs) focused the diffracted light onto a split photodetector (Spot2D; UDT Sensors, Hawthorne, CA) (I).
The highly sensitive split photodetector converted the emission light into two currents. The currents were converted into voltages using an ultralow bias current operational amplifier (OPA124U; Burr-Brown, Dallas, TX). Fig. 2 shows a schematic of the detection circuit.
|
feedback resistor (Rf) was chosen to optimize the tradeoff between gain and noise to achieve the required bandwidth (1
In stage 1, the photodiode current (I) was converted to voltage (V), as determined by
![]() | (1) |
![]() | (2) |
Ray tracing simulations
Ray tracing simulations (ZEMAX Development, Bellevue, WA) were performed to optimize the geometry of the components in the detection pathway. The diffraction grating separated the emission light into different wavelengths by reflecting the light at different angles. The angle of reflection
was determined by
![]() | (3) |
is the wavelength of the light, d is the grating constant,
i is the incident angle, and n is the diffraction order (35
i = 0) to maximize the angular dispersion.
The angular dispersion of a diffraction grating is determined by the inverse of the grating constant (1/d) or pitch. Commercially sold diffraction gratings at the dye emission wavelengths with a pitch of 600 grooves/mm or 1200 grooves/mm were simulated to select the optimal pitch. ZEMAX design software was used to simulate the emission pathway of the light by ray tracing a set of discrete wavelengths. We simulated 12 sample wavelengths spanning the emission spectrum of the dye. Each wavelength had an associated relative power, which was determined by the relative dye emission multiplied by the relative responsivity of the photodetector. The diffraction grating efficiency was not incorporated into the simulations. Fig. 3 shows the relative dye emission (36
), the relative responsivity of the photodetector, and the relative power.
|
System alignment
After the channel was built, a helium neon (HeNe) laser (05-LLR-811; Melles Griot, Carlsbad, CA) was used to test the alignment of the optics in the detection pathway. The laser was directed into the tissue fiber through a 632-nm band-pass filter (G42-081; Edmund Optics, Barrington, NJ) and two neutral density filters (ND 30A and ND 13A; Edmund Optics). The photodetector was centered such that each photodiode cell had an equal output voltage. The photodetector was moved with a micromanipulator (H LH; Line Tool, Allentown, PA) in the x and y directions (Fig. 1 shows the coordinate system). The photodetector was also moved in the z and x directions while maintaining equal output voltage after amplification. The slope of the z versus x displacements was the inverse tangent of the alignment angle. The optimal alignment angle was approximately zero when the light was centered and collimated.
Bandwidth calculations and measurement
The electrical bandwidth was also determined for the detector and amplifier circuit, both theoretically and experimentally. The 3-dB cutoff frequency, B, was determined theoretically by
![]() | (4) |
Experimentally, the bandwidth was measured using an intensity-modulated red LED (IF-E96; Industrial Fiber Optics, Tempe, AZ). The LED was powered using a waveform generator (33220A; Agilent, Palo Alto, CA), with a sine wave output that had an offset of 1 V and amplitude of 190 mV. The LED was positioned directly in front of the split photodetector to bring the signal produced by the detector circuit to an output peak-to-peak voltage (VPP) of 3 V. The VPP was measured at frequencies ranging 150 kHz for several settings of the variable capacitor. The final bandwidth of the circuit was adjusted to 150 Hz (38
) using the variable feedback capacitor. The minimal bandwidth was used to maximize the signal/noise ratio (SNR) (39
).
Noise calculations
The SNR is typically low in optical signals. Thus, the theoretical noise was also calculated to identify and minimize the dominant noise sources. The noise was a combination of the amplifier noise, the shot noise, and the Johnson noise. The shot noise (iI) was principally caused by the photodiode current (I). The mean-squared value of the shot noise was determined by
![]() | (5) |
The Johnson noise (iR) is due to the combination of the photodetector resistance and the feedback resistance of 100 M
(39
). The photodetector resistance includes both series and shunt resistances, which were ignored because of their magnitude (37
). The quadratic mean-squared value of the Johnson noise was determined by
![]() | (6) |
The amplification noise current was calculated based on the low bias current of 75 fA (33
). The noise associated with the second stage was also found using amplifier noise analysis techniques (40
). The total theoretical noise voltage was compared to the experimental noise voltage. The experimental noise voltage (NoiseRMS) was determined by measuring the standard deviation of the detector output during a 20-ms interval before each AP (1
).
Rabbit studies
These animal studies were performed under the guidelines of the Duke University Institutional Animal Care and Use Committee. Male Adult New Zealand White rabbits (n = 5) were anesthetized with intramuscularly injected ketamine (35 mg/kg) combined with xylazine (5 mg/kg). Heparin (100 mg/kg) and thiopental (20 mg/kg) were given intravenously. Both corneal eye and paw pinch reflexes were used as indicators of adequate sedation. The heart was rapidly excised via median sternotomy and submerged in cold, high potassium Tyrode's solution (24 mM KCl, 123 mM NaCl, 11 mM dextrose, 20 mM NaHCO3, 1.0 mM NaH2PO4·H2O, 1.1 mM MgCl2·H2O, and 1.8 mM CaCl2) oxygenated with 95%O2/5%CO2. The aorta was cannulated and Langendorff perfused. The heart was placed in a warm bath with normal Tyrode's solution (4.5 mM KCl) and perfused for 1 h to allow the heart to stabilize. Both the flow pressure and temperature were continuously monitored to maintain the physiological range of 6080 mmHg and 35°C38°C, respectively. Two electrodes connected to a differential amplifier (Iso-DAM8A; WPI, Sarasota, FL) were used to monitor the electrogram. A force transducer (FORT10; WPI) connected to a bridge amplifier (Bridge 8; WPI) was used to monitor the contraction-induced motion. Transmembrane voltage was measured using standard pulled glass microelectrodes filled with 3 mol/L KCL. The microelectrode signals were low-pass filtered at 500 Hz (Iso-DAM8A; WPI). The tissue was stained with 50 µM di-4-ANEPPS dissolved in ethanol added to the perfusate. Continuous recordings were taken from the base of the heart without BDM.
The atria were removed and the atrioventricular node ablated to prevent spontaneous activations. Baseline endocardial pacing was applied to the right ventricle via a unipolar electrode at a cycle length of 500 ms based on previous investigators (28
). The longer cycle length helps to reduce the tissue hypoxia associated with the nonblood Tyrode's perfusion. A shorter cycle length can accelerate the hypoxia due to the higher metabolic demand (41
). Control of stimulation and data collection were achieved with LabVIEW 7.0 software, a data collection card (AT-MIO-16E-2; National Instruments (Austin, TX) E-series PCI Card), and a voltage-to-period converter. The signals from the two optical detectors, the microelectrode, the force transducer, and the bath ECG were digitally oversampled at 2 kHz using a 16-bit analog-to-digital converter within a PC.
The signal from shorter wavelengths was divided by the signal from longer wavelengths (25
). The signals were normalized by subtracting the minimum value, then dividing the result by the maximum value minus the minimum value. A 20-point median filter was applied to the ratiometric signal to improve the SNR (42
). The SNR was determined by
![]() | (7) |
In addition, APD70 was measured for both the microelectrode and the optical signals in a single rabbit. The APD was calculated by measuring the interval between the AP upstroke where the voltage rose above 70% of peak amplitude and the repolarization time where the voltage fell below the 70% of the peak amplitude of the AP (1
). Recordings were made in the same region (base of the heart) to minimize spatial differences in APD (43
). The APD measured for the two conditions were compared using a two-tailed t-test. A p-value < 0.05 was considered significant.
| RESULTS |
|---|
|
|
|---|
|
|
The simulations also allowed us to optimize the distance between the diffraction grating and focusing aspheric lens by simulating the dispersion from the grating over the range of wavelengths encompassed by the dye emission. Fig. 6 A shows the ZEMAX simulation using the 12 sampled wavelengths and 1200 grooves/mm grating in the xz plane. The central wavelength was 655 nm for equal power distribution on the cells of the split photodetector. Fig. 6 C shows each wavelength focused into a circle at a distinct location in the xy plane. Fig. 6 C also shows an overlay of each detector (1.27 x 2.54 mm) with the detector gap of 0.127 mm. The top detector collected light from the "green" or lower wavelength side of the dye spectrum. The bottom detector collected light from the "red" or higher wavelength side of the dye spectrum. Thus, we found that a distance of 5 mm between the diffraction grating and focusing aspheric lens was optimal.
|
3.7 pW and 3.1 pW, respectively. This converts in the photodetector to a current of 1.50 pA and 1.49 pA (photodetector responsivity at 605 nm = 0.405 A/W, photodetector responsivity at 680 nm = 0.48). The amplification circuit converted the signal to a voltage according to Eq. 1 and amplified the signal according to Eq. 2, resulting in a theoretical voltage arising from the AP of
1.65 mV and 1.64 mV.
Laser measurements
Laser light was used to align the detector with respect to the diffraction grating. The narrow spectral wavelength bandwidth of the laser light produces a spot at a single location on the detector. The voltage from each detector was first measured while moving the micromanipulator left to right along the x direction. This measurement was repeated three times at a relative placement of 0 mm, 2.54 mm, and 5.08 mm along the z axis. At all three placements, the separation of the peak output voltages from the green and red detectors was 1.905 mm. The separation in the peaks for movement in the x direction was due to the element gap in the detector and misalignment of the aspheric lens.
Fig. 7 shows the voltage change as a function of the y displacement. The red and green detectors peaked at 650 mV and 645 mV, respectively. The peaks were separated by 0.635 mm. The separation in the peaks from movement in the y direction was due to the misalignment of the aspheric lens with respect to the photodetector. Fig. 8 shows measurements taken by moving the photodetector in the z and x directions to find the angular alignment. The angular alignment is approximately the inverse tangent of the slope. The slope was 0.0008 mm/mm, which corresponds to the detector being
0.05° off axis. This angle could be reduced with a more precise micromanipulator. Fine adjustments (<0.635 µm) were also made in the x location of the detector during the experiments to minimize the MA in the ratiometric signal.
|
|
caused the bandwidth cutoff to range from 160 Hz to 4 kHz. A red LED with varying intensity was used to measure the circuit bandwidth at various capacitor settings. Fig. 9 shows a plot of the voltage amplitude versus the frequency. The 3-dB roll-off frequency was found empirically to range from 110 Hz to 1.5 kHz. The difference between the theoretical and experiment bandwidth was likely due to parasitic capacitance from the traces on the amplifier board. Both detectors were set to have a 3 dB frequency of 150 Hz.
|
, which was greater than the feedback resistance of 100 M
. This difference decreased amplification of noise, voltage offset, and drift (39
The largest theoretical noise source was the shot noise from the detector with a mean-square value of 2.8 x 1026 A2. The mean-square noise currents from the amplifier and the Johnson noise were 3.6 x 1030 A2 and 2.5 x 1026 A2, respectively. The total theoretical value of mean-square current noise was 5.2 x 1026 A2 (21
). This noise was amplified through the first stage and second stage. There was additional amplifier noise Johnson voltage noise from the second stage of 3.0 µV and 52.3 µV (40
), respectively. This resulted in a total theoretical of
257 µV.
Experimentally, the NoiseRMS was 742 µV on the red detector and 724 µV on the green detector. Frequency analysis showed a peaking at 180 Hz, which was likely due to contributions from 60 Hz sources. The noise spectrum also showed significant contributions above 150 Hz, indicating noise from other sources, possibly the computer and data acquisition card. Both theoretical noise and experimental noise were smaller than the ZEMAX estimated signal size of
±1.65 mV.
In vitro recordings
Recordings were taken from the rabbit ventricle using the ratiometric channel. Fig. 10 shows the original red and green signals, the raw and filtered ratiometric signals, the force transducer signal, and the signals from the microelectrode. The microelectrode recordings exhibit some artifact during the repolarization phase of the AP. This is due to the tissue-contraction motion, which caused the microelectrodes to partially pull out of the cell membrane. The AP upstroke from the red and green detector was
± 3 mV. The offset voltages were 341 mV and 132 mV on the green and red portions of the photodiode, respectively. The green offset was higher because reflected and stray illumination light contributed more to this half of the spectrum. The mean total power minus the offset for the results shown was 557.3 pW and 483.2 pW for the red and green detectors, respectively.
|
Table 1 shows the average SNR and motion ratio for all rabbits. The number of continuous APs analyzed from each animal ranged 718. For rabbit 5, the average SNR from the optical recordings shown in Fig. 10 was 14.0 ± 2.5 (range 11.118.9, No. of APs = 7). The MA was most substantial 80 ms after the AP upstroke. It continued throughout the repolarization and during the diastolic interval. The average motion ratio before ratiometry for the red and green signals was 0.215 ± 0.01 (range 0.200.22, No. of APs = 7) and 0.132 ± 0.01 (range 0.1160.15, No. of APs = 7), respectively. The average motion ratio after ratiometry was 0.69 ± 0.02 (range 0.670.71, No. of APs = 7). The difference was statistically significant (p-value = 0.0001).
|
| DISCUSSION |
|---|
|
|
|---|
The ray-tracing simulations were useful for selecting the optical components and layout. The 1200 grooves/mm grating was used because the angular resolution was three times greater than the 600-grooves/mm grating. The larger angular resolution caused a more dramatic spatial dispersion of the emission light. The AP amplitude recorded in the tissue was almost double the AP amplitude predicted by ZEMAX, which assumed detection of 1.5 nW of light with only a 0.8% change in fluorescence. The small percentage change was used to determine if the system could detect a 1-nm shift in the spectrum above the theoretical noise. The experimentally measured signal had less total light detection but a larger percentage change in fluorescence. This fractional change can range 0.0%8.0% depending on the detection wavelengths (25
), experimental setup, and illumination source.
The laser alignment of the detector ensured that the optical components were optimally placed. If the laser spot showed misalignment by greater than the size of the lens (6.33 mm), the lens was not able to focus the light onto the photodiode. This resulted in a decrease in the voltage in both x and y displacement measurements. The angular misalignment was minimal, which helped to ensure the correct wavelengths fell on each half of the photodiode.
Each circuit had a measured 3-dB cutoff bandwidth of 150 Hz. This bandwidth was chosen to ensure that the signal contained the total energy of the AP without including higher frequency noise;
90% of the total energy of the AP is below 150 Hz (38
). The theoretical range of the bandwidth set by the transimpedance amplifier feedback capacitor exceeded the experimental range measured using the intensity varying light source. This difference is likely due to parasitic capacitance from the traces on the amplifier board. The theoretical noise was lower than the experimental noise, which can be additionally reduced with low-pass filtering in the second stage of the amplification circuit.
This system produced a relatively high SNR. These results were likely from illuminating the dye with a royal blue LED. Royal blue light has increased the dye absorption because it emitted at a wavelength that was close to the dye peak absorption. A change in membrane voltage causes a shift in both the absorption and emission spectrum. By exciting the dye at the absorption peak, the intensity changes in the detected signals were primarily due to the shift of the emission spectrum. Royal blue light was also easier to filter from the dye emissions, which is particularly important in ratiometric experiments (25
). Filtering out the excitation light reduced the voltage offset at the detectors to ensure an adequate dynamic range in the detector circuit. Royal blue LEDs have been shown to be low noise sources (34
) that are also more cost effective than lasers.
Previous work using a fiber-based system with a laser showed an SNR of 16 V/V (3
). This was only slightly higher than the recorded average unfiltered SNR of 14 V/V. The SNR could be improved by increasing the excitation light intensity, which will lead to a larger signal. The drawback of increasing the excitation light intensity is photobleaching. Photobleaching will decrease the signal size over time (25
).
The motion ratio was an average of 65% lower in the ratiometric signals than the raw signals. It was difficult to measure the motion ratio in the presence of substantial noise. The motion ratio is the upstroke voltage divided by the peak-to-peak voltage during the AP (25
). The motion ratio of a signal with no MA will not be close to one if the signal is noisy. This is because random noise increases the peak-to-peak range of voltages during the AP. A raw optical signal (acquired with 10 mM BDM) with almost no visible MA has a filtered SNR of 30.9 V/V but a motion ratio of 0.44 V/V. The motion ratio is useful for quantifying the change in the MA amplitude, but the motion ratio must be considered in the context of the SNR. Knisley et al. reported a better average motion ratio, but the signals also had substantially better SNR (25
). The calculated motion ratio for the raw red and green signals was approximately equal to the Knisley et al. results, whereas their ratiometric motion ratio was smaller due to the greater SNR (25
).
The channel demonstrated that MA could be almost entirely eliminated using ratiometry. The primary reason the system was successful in reducing MA was matching the two detection pathways. The two bands of emission light were transmitted through the same optical components and then captured by matching photodiodes. This allowed for common detection and amplification of the emission light. The optical signals were simply divided without normalization or background offset subtraction. These novel features allowed this channel to remove MA without the use of BDM.
Limitations
Optical recordings are a spatial average of AP upstrokes from multiple cells. Even at normal conduction velocities, the cells do not activate simultaneously, which increases the upstroke duration of the AP. An optical fiber with a diameter of 125 µm did not have the spatial resolution to overcome this limitation in cell culture studies (44
). Previous investigators measured the upstroke duration using 100 µm and 200 µm fibers and concluded that spatial resolution was not a limitation of their system (1
,3
). The APD measurements found in this study were essentially equal to the microelectrode APD measurements. This finding confirms that the bandwidth, noise, and spatial resolution were not limitations in the novel system.
The channel is a "proof-of-concept" design. The channel size must be further minimized for a multichannel system. In a multichannel system, the three-stage micromanipulator can be replaced by a single stage that can be moved in the x direction. Cheaper aluminum parts can also be used to replace the mountings and optical board. The diffraction grating is a simple and cost effective device that can easily be incorporated into a multichannel mapping system. For example, a grating can be used to replace the dichroic mirror for separating the emission spectrum of the dye in the previously described low cost optical system (45
). Though the diffraction grating does not have the same efficiency as the dichroic mirror, it was shown to be an effective addition to the channel.
| CONCLUSIONS |
|---|
|
|
|---|
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
We gratefully acknowledge the financial support of the National Institutes of Health under grant 1R01-HL-72831. N.H.B. acknowledges the financial support of the National Institutes of Health under grant HL072831-02S1 and F31 Eb003389-01A1. H.M.D. and D.J.G. acknowledge the financial support of the National Science Foundation under grant PHY-0243584.
| FOOTNOTES |
|---|
Submitted on November 18, 2006; accepted for publication February 22, 2007.
| REFERENCES |
|---|
|
|
|---|
2. Caldwell, B. J., I. J. LeGrice, D. A. Hooks, D. C.-S. Tai, A. J. Pullan, and B. Smaill. 2005. Intramural measurement of transmembrane potential in the isolated pig heart: validation of a novel technique. J. Cardiovasc. Electrophysiol. 16:10011010.[CrossRef][Medline]
3. Hooks, D. A., I. J. LeGrice, J. D. Harvey, and B. H. Smaill. 2001. Intramural multisite recording of transmembrane potential in the heart. Biophys. J. 81:26712680.
4. Krauthamer, V., H. Bryant, C. Davis, and T. Athey. 1991. Action potential-induced fluorescence changes resolved with an optical fiber carrying excitation light. J. Fluoresc. 1:207213.[CrossRef]
5. Neunlist, M., S. Zou, and L. Tung. 1992. Design and use of an optrode for optical recordings of cardiac action potentials. Eur. J. Physiol. 420:611617.[CrossRef][Medline]
6. Kodama, I., I. Sakuma, N. Shibata, S. B. Knisley, R. Niwa, and H. Honjo. 2000. Regional differences in arrhythmogenic aftereffects of high intensity DC stimulation in the ventricles. PACE. 23:807817.[Medline]
7. Wiggins, J. R., J. Reiser, D. F. Fitzpatrick, and J. L. Bergey. 1980. Inotropic actions of diacetyl monoxime in cat ventricular muscle. J. Pharmacol. Exp. Ther. 212:217224.
8. Li, T., N. Sperelakis, R. E. Teneick, and R. J. Solaro. 1985. Effects of diacetyl monoxime on cardiac excitation-contraction coupling. J. Pharmacol. Exp. Ther. 232:688695.
9. Efimov, I. R., and T. N. Mazgalev. 1998. High-resolution three-dimensional fluorescent imaging reveals multilayer conduction pattern in the atrioventricular node. Circulation. 98:5457.[Medline]
10. Dillon, S. M. 1991. Optical recordings in the rabbit heart show that defibrillation strength shocks prolong the duration of depolarization and the refractory period. Circ. Res. 69:842856.
11. Wu, J. S., M. Biermann, M. Rubart, and D. P. Zipes. 1998. Cytochalasin D as excitation-contraction uncoupler for optically mapping action potentials in wedges of ventricular myocardium. J. Cardiovasc. Electrophysiol. 9:13361347.[Medline]
12. Baker, L. C., B. London, B. R. Choi, G. Koren, and G. Salama. 2000. Enhanced dispersion of repolarization and refractoriness in transgenic mouse hearts promotes reentrant ventricular tachycardia. Circ. Res. 86:396407.
13. Baker, L. C., R. Wolk, B. R. Choi, S. Watkins, P. Plan, A. Shah, and G. Salama. 2004. Effects of mechanical uncouplers, diacetyl monoxime, and cytochalasin-D on the electrophysiology of perfused mouse hearts. Am. J. Physiol. Heart Circ. Physiol. 287:H1771H1779.
14. Bergey, J. L., J. D. McCallum, and K. Nocella. 1981. Antiarrhythmic evaluation of verapamil, nifedipine, perhexiline and skf 525-A in four canine models of cardiac arrhythmias. Eur. J. Pharmacol. 70:331343.[CrossRef][Medline]
15. Blanchard, E. M., G. L. Smith, D. G. Allen, and N. R. Alpert. 1990. The effects of 2,3-butanedione monoxime on initial heat, tension, and aequorin light output of ferret papillary muscles. Pflugers Arch. 416:219221.[CrossRef][Medline]
16. Gwathmey, J. K., R. J. Hajjar, and R. J. Solaro. 1991. Contractile deactivation and uncoupling of crossbridges: effects of 2,3-butanedione monoxime on mammalian myocardium. Circ. Res. 69:12801292.
17. Qin, H., M. W. Kay, N. Chattipakorn, D. T. Redden, R. E. Ideker, and J. M. Rogers. 2003. Effects of heart isolation, voltage-sensitive dye, and electromechanical uncoupling agents on ventricular fibrillation. Am. J. Physiol. Heart Circ. Physiol. 284:H1818H1826.
18. Cheng, Y., K. Mowrey, I. R. Efimov, D. R. Vav Wagoner, P. J. Tchou, and T. N. Mazgalev. 1997. Effects of 2,3-butanedione monoxime on atrial-atrioventricular nodal conduction in isolated rabbit heart. J. Cardiovasc. Electrophysiol. 8:790802.[Medline]
19. Banville, I., and R. Gray. 2002. Effects of action potential duration and conduction velocity restitution and their spatial dispersion on alternans and the stability of arrhythmias. J. Cardiovasc. Electrophysiol. 13:11411149.[CrossRef][Medline]
20. Kettlewell, S., N. L. Walker, S. M. Cobbe, and G. L. Smith. 2003. The electrophysiological and mechanical effects of 2,3-butane-dione monoxime and cytochalasin-D in the Langendorff perfused rabbit heart. Exp. Physiol. 89:163172.[CrossRef]
21. Biermann, M., M. Rubart, A. Moreno, J. Wu, A. Josiah-Durant, and D. P. Zipes. 1998. Differential effects of cytochalasin D and 2,3 butanedione monoxime on isometric twitch force and transmembrane action potential in isolated ventricular muscle: implications for optical measurements of cardiac repolarization. J. Cardiovasc. Electrophysiol. 9:13481357.[Medline]
22. Lee, M.-H., S.-F. Lin, T. Ohara, C. Omichi, Y. Okuyama, E. Chudin, A. Garfinkel, J. N. Weiss, H. Karagueuzian, and P.-S. Chen. 2001. Effects of diacetyl monoxime and cytochalasin D on ventricular fibrillation in swine right ventricles. Am. J. Physiol. Heart Circ. Physiol. 280:H2689H2696.
23. Jalife, J., G. E. Morley, N. Y. Tallini, and D. Vaidya. 1998. A fungal metabolite that eliminates motion artifact. J. Cardiovasc. Electrophysiol. 9:13581362.[Medline]
24. Efimov, I. R., D. T. Huang, J. M. Rendt, and G. Salama. 1994. Optical mapping of repolarization and refractoriness from intact hearts. Circulation. 90:14691480.
25. Knisley, S. B., R. Justice, W. Kong, and P. Johnson. 2000. Ratiometry of transmembrane voltage-sensitive fluorescent dye emission in hearts. Am. J. Physiol. Heart Circ. Physiol. 279:H1421H1433.
26. Montana, V., D. L. Farkas, and L. M. Loew. 1989. Dual-wavelength ratiometric fluorescence measurements of membrane potential. Biochemistry. 28:45364539.[CrossRef][Medline]
27. Fromherz, P., and A. Lambacher. 1991. Spectra of voltage-sensitive fluorescence of styryl-dye in neuron membrane. Biochim. Biophys. Acta. 1068:149156.[Medline]
28. Tai, D. C.-S., B. J. Caldwell, I. J. LeGrice, D. A. Hooks, A. J. Pullan, and B. H. Smaill. 2004. Correction of motion artifact in transmembrane voltage-sensitive fluorescent dye emission in hearts. Am. J. Physiol. Heart Circ. Physiol. 287:H985H993.
29. Fluhler, E., V. G. Burnham, and L. M. Loew. 1985. Spectra, membrane binding, and potentiometric responses of new charge shift probes. Biochemistry. 24:57495755.[CrossRef][Medline]
30. Loew, L., L. B. Cohen, J. Dix, E. N. Fluhler, V. Montana, G. Salama, and W. Jain-young. 1992. A naphthyl analog of aminostyryl pyridium class of potentiometric membrane dyes shows consistent sensitivity in a variety of tissue, cell and model membrane preparations. J. Membr. Biol. 130:110.[Medline]
31. Bullen, A., and P. Saggau. 1999. High-speed, random-access fluorescence microscopy: II. Fast quantitative measurements with voltage-sensitive dyes. Biophys. J. 76:22722287.
32. Liu, Q., and N. Ramanujam. 2004. Experimental proof of the feasibility of using an angled fiber-optic probe for depth-sensitive fluorescence spectroscopy of turbid media. Opt. Lett. 29:20342036.[CrossRef][Medline]
33. Burr-Brown Corporation. 1994. Application Bulletin: Designing Photodiode Amplifier Circuits with OPA128. Burr-Brown Corporation, Tucson, AZ.
34. Entcheva, E., Y. Kostov, E. Tchernev, and L. Tung. 2004. Fluorescence imaging of electrical activity in cardiac cells using an all-solid-state system. IEEE Trans. Biomed. Eng. 51:333341.[CrossRef][Medline]
35. Jenkins, F. A., and H. E. White. 1976. Fundamentals of Optics, 4th ed. McGraw-Hill, New York.
36. Molecular Probes Inc. 2006. SpectraDi-8-ANEPPS/Lipid. Eugene, OR.
37. Graeme, J., editor. 1996. Photodiode Amplifiers: Op Amp Solutions. McGraw-Hill, Boston.
38. Girouard, S. D., K. Laurita, and D. Rosenbaum. 1996. Unique properties of cardiac action potentials recorded with voltage sensitive dyes. J. Cardiovasc. Electrophysiol. 7:10241038.[Medline]
39. Donati, S. 2000. Photodetectors: Devices, Circuits and Applications. Prentice-Hall, Upper Saddle River, NJ.
40. Texas Instruments Inc. 1998. Noise Analysis in Operational Amplifier Circuits. Texas Instruments Incorporated, Dallas, TX.
41. Pitruzzello, A. M., W. Krassowska, and S. F. Idriss. 2007. Spatial heterogeneity of the restitution portrait in rabbit epicardium. Am. J. Physiol. Heart Circ. Physiol. 292:H1568H1578.
42. Witkowski, F. X., P. A. Penkoske, and L. Leon. 2001. Optimization of temporal filtering for optical transmembrane potential signals. In Optical Mapping of Cardiac Excitation and Arrhythmias. D. S. Rosenbaum and J. Jalife, editors. Futura Publishing, Armonk, NY. 7992.
43. Sicouri, S., and C. Antzelevitch. 1991. A subpopulation of cells with unique electrophysical properties in the deep subepicardium of the canine ventricle. The M cell. Circ. Res. 68:17291741.
44. Rohr, S., and J. Kucera. 1998. Optical recording system based on a fiber optic image conduit: assessment of microscopic activation patterns in cardiac tissue. Biophys. J. 75:10621075.
45. Idriss, S. F., and A. M. Pitruzzello. 2006. A low-cost high-efficiency fiber-optic coupler for recording action potentials within the myocardial wall. IEEE Trans. Biomed. Eng. 53:17081711.[CrossRef][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |