| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
The study of cell-material interactions is a broad field of research, including cell adhesion, migration, proliferation, and differentiation, as well as the traction forces generated by adherent cells. In the past decades, a variety of substrates have been used for the study of cell traction forces. Traction forces in chick heart fibroblasts cultured on silicone rubber membranes were quantified based on the morphology of deformations observed in the films (3
), and this method was later refined for use with keratocytes (4
). The usage of deformable cell culture substrates further progressed with the use of synthetic hydrogels like polyacrylamide coated with collagen (5
), as well as with embedded fluorescent particles for for measurement of substrate deformations by the cell (6
,7
). Other hydrogels that have been used include agarose (8
), fibrin (9
), and collagen (10
). Variation of total gel composition, component concentration, or extent of cross-linking could be used to produce substrates of varying stiffnesses (5
,6
,11
), often with some surface modification to facilitate cell adhesion. The modulus of such gel substrates needs to be low enough to enable detectable substrate deformation within the range of forces produced by attaching, spreading, and migrating cells.
A major drawback of using flexible substrates for cell force detection is that the force indicators, whether wrinkles or embedded particles, are mechanically coupled. Advanced computer modeling is required to convert the deformations into forces, and these force measurements may only be reliable at cell level. Mapping of focal adhesions can increase the accuracy (12
), but ultimately the measurements are still coupled, both within a single cell, and between neighboring cells. Hence, quantification of cell-substrate interactions on these types of continuous substrates is complicated. As an alternative, systems of discrete cantilevers, henceforth termed "pillars", were developed to obtain truly independent measurements and simplify the force calculations.
In the last decade, pillared substrates have become a popular alternative to the continuous substrates. A sensor was developed using micromachined silicon wafers, incorporating horizontally oriented pillars with independent force-sensing attachment pads (13
). The force measurements (up to 100 nN under chick embryo fibroblasts) were still limited because forces could not be measured along the axis of the embedded cantilever. Moreover, it was impossible to decouple force direction and magnitude. Other studies have utilized vertical pillars, which do not have such limitations as they can freely bend into all directions and are not coupled to each other. Most commonly used are poly-dimethyl-siloxane (PDMS) elastomer pillars, as they are easy to produce by replication from a micromachined mold. PDMS is an elastic, biocompatible material that is optically transparent. PDMS microneedles, functionalized with fibronectin to facilitate cell attachment, could be used to measure forces in the range of 590 nN in smooth muscle cells (14
). Primary neonatal rat cardiomyocytes were cultured on the flat surface surrounding PDMS pillars, grown to confluence, and then observed deflecting the pillars with forces of up to 3.5 µN (15
). Given the elasticity of this material, vertical pillars do not need to have a high aspect ratio to obtain the desired spring constant, and may have heights on the order of a cell, or less. It is therefore a risk that cells reach down to the base of the pillars. Cell force measurements performed with vertical pillars might also influence cell biological reactions by the specific topography provided by the pillar arrays. As cells should attach mainly to the tops of the pillars, they are presented with a limited attachment area that might significantly affect morphology and functions of anchorage dependant cells (16
). Thus geometry of the pillars and their distribution has to be taken into account when designing vertical pillar-based force sensors.
Previously, we have also shown that vertical pillars micromachined in a stiff biocompatible material, such as silicon, can be used to measure cell forces (17
). These sensors have improved optical contrast compared with PDMS, due to their flatness and high substrate reflectivity, theoretically allowing higher precision force measurements to be achieved. When micromachining in silicon, it is possible to manipulate both the organization and aspect-ratio of pillars, allowing subcellular force measurement, and pillars of varying spatial arrangements and elastic properties. This type of manufacturing also ensures the production of well-controlled topographical substrates with the possibility for enlarged cell attachment area on the top of pillars. An interesting possibility exists of incorporating additional features to influence the migration of cells. Since the patterning of substrate ridges is known to affect cell shape and motion (18
), we predict these substrates may be designed to influence cell movement, a property desirable for modern biomaterials. Contact guidance (19
) is a phenomenon whereby a cell that senses a step, in this case the trenches containing pillars adjacent to ridges, will tend to follow the direction of the step, becoming elongated. The magnitude of topographical alignment is mainly dependent on the depth of the trenches (20
). Even with the presence of an opposing (discontinuous) chemical pattern, MC3T3-E1 cells were seen to be influenced by the underlying topographical pattern of 4 µm (21
). Though in the case of BHK cells, whereas both topography and chemistry could induce alignment, a continuous chemical pattern was more dominant (22
). Hence cellular reaction to chemical and topographical substrate cues is dependent on cell type.
In this article we explore further the application of micromachined silicon substrates patterned with arrays of pillars in between solid ridges. Such substrates are intended to observe cellular traction force distribution and dynamics during cell spreading, migration, and topographical contact guidance. Human vascular endothelial cells (EC) and fibroblasts were able to attach, spread, and migrate on the top of microfabricated silicon substrates and, as a result, deflect the pillars underneath the cells. As multiple pillars were positioned under each cell, the observed deflections were used for the detection of forces distributed with subcellular lateral resolution in the nanonewton magnitude range.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Atomic force microscopy (AFM; Dimension 3000 NanoScope, Veeco Instruments, Woodbury, NY) in force calibration mode was used to determine stiffness of the microfabricated pillars by measuring their spring constant. The substrate was cleaved and mounted on the microscope stage so that its cross section was exposed in a horizontal position. An AFM tip with a rectangular cantilever of known spring constant was used to record force curves on the rigid part of the cleaved substrate and on the free end of the pillar, which enabled the determination of the pillar's spring constant. The spring constant was determined by applying the equation kp = kt cos
Zf / (Zf Zr), where kp is the spring constant of the pillar, kt is the spring constant of AFM tip cantilever (kt = 0.283 N/m, CLFC-NOBO calibration tip from ThermoMicroscopes, Sunnyvale, CA),
is the inclination angle of the AFM tip cantilever (
= 10° for Dimension 3000 NanoScope), and Zr and Zf are the vertical displacement of the AFM tip when the same force is applied on the rigid part and the flexible end of the microfabricated pillar, respectively (17
,23
). Four pillars were measured on a force sensor chip taken from the same wafer as the chips used in the cell experiments. The mean spring constant and standard deviation was calculated from these four measurements.
Cell culture
Passaged primary human vascular endothelial cells (EC) were obtained from the Wallenberg Laboratory for Vascular Research (Department of Surgery and Vascular Surgery, Sahlgrenska University Hospital, Gothenburg, Sweden). They were routinely cultured in T25 flasks (Greiner, Bio-One North America Inc., Monroe, NC; 1:2) in a standard tissue culture incubator at 37°C, 100% humidity, and 5% CO2. Medium M199 (Gibco BRL, Gaithersburg, MD) was supplemented with 25 mM HEPES, 20% fetal calf serum (PAA Laboratories GmbH, Pasching, Austria), 2 mM L-glutamine, penicillin, streptomycin, 0.15 mg/ml ECGF, and 5000 IU/ml heparin (Sigma, St. Louis, MO). The EC cells used in this study were from the passage 5.
Passaged primary human fibroblasts (Wallenberg Laboratory for Vascular Research) were routinely cultured in T25 flasks (Greiner, 1:4) in a standard tissue culture incubator at 37°C, 100% humidity, and 5% CO2. High-glucose Dulbecco's modified eagle medium (DMEM) (PAA Laboratories) was supplemented with 10% fetal calf serum (PAA Laboratories). For time-lapse microscopy, cells were detached with 1x trypsin-EDTA solution (Sigma, St. Louis, MO) and plated, as per routine passage. The fibroblasts used in this study were from the passage 7.
The EC were allowed to attach for an hour before observation, the fibroblasts were observed
30 min after plating.
Digital microscopy of live cells
For live cell observation, microfabricated silicon substrates were placed in 35-mm tissue culture dishes (Nalge Nunc International, Rochester, NY) and sterilized by immersion in 70% ethanol, washed in Dulbecco's phosphate buffered saline (PBS) (PAA Laboratories) and equilibrated for a few minutes in complete medium before seeding of cells. Images were acquired every 3 and 5 min (for fibroblasts and EC, respectively) using a Zeiss x40, NA 0.80 Achroplan immersion objective on a Zeiss Axioplan 2E imaging microscope equipped with differential interference contrast (DIC) optics, Zeiss Axiocam digital camera, and Zeiss Axiovision software (Carl Zeiss Light Microscopy GmbH, Göttingen, Germany). The use of reflected DIC was necessary due to the substrate's opacity. A thin film of sterile mineral oil (Sigma) was placed on top of the media to prevent evaporation. The culture was maintained at 37°C by an on-stage incubator (PDMI-2, Harvard Apparatus, Holliston, MA) and heating controller (TC-202A, Harvard Apparatus).
Image analysis
Image analysis was performed on stacks of uncompressed 8-bit grayscale TIFF images. Custom Matlab codes were used to process time-lapse sequences in Matlab 7.1 (The MathWorks, Natick, MA) with the following steps:
Generation of an "ideal lattice"
The "ideal lattice" represents the imaged array of undeflected pillars in the absence of cells. The image background was calculated using an opening operation and removed. The pillars were then identified in the background-subtracted image, using a thresholding operation. The center-to-center pitch between adjacent pillars in the x-direction was calculated by measuring the distance between pillars for all neighbors in a direction specified by the user. The pitch in the y-direction was calculated similarly, with a shift in angle of 90°. Using these pitches, a square ideal lattice was created. To correct for distortions in the real pillar lattice due to sample tilt or uneven illumination, pillars in the image were compared to the ideal lattice. Displacements in the x- and y-direction were separately mapped onto the z axis to visualize the displacements as a surface. A cubic surface was fitted to each of these mappings, and the pillar locations in the ideal lattice were shifted to obtain best agreement between the ideal lattice and the pillars in the first image in the sequence.
Corrections for substrate drift and rotation between images
Drift between subsequent images in a time-lapse sequence was corrected by tracking the median displacements of all pillars in the field of view (FOV). Rotation was corrected by minimizing the summed absolute displacement of all pillars due to rotation of the sample.
Measurement of pillar deflections and calculation of force vectors
Deflections were measured as the lateral displacement of the center of mass for each pillar. The force vectors were calculated assuming the pillars as standing cantilevers that behave as ideal springs, as described by Hooke's Law, F = kx, with a spring constant k and lateral deflection x. The displacement of each pillar in each image of the sequence was calculated by comparing the center of mass of the pillar in the current image with the center of mass of the same pillar in the ideal lattice. The force vectors were then superimposed on each image of the sequence. Contrast and brightness corrections were used after analysis to enhance cell detail for Figs. 3 and 4.
|
|
Orientation of force vectors
Force vectors were further classified with respect to the direction of cell spreading. The center of a cell was manually marked; the vectors were designated either positive or negative, by the sign of the comparison of the distance between the cell center and the shifted pillar location, and the distance between the cell center and the original pillar location. A positive vector is thus oriented away from the center of the cell, and similarly, a negative vector is oriented toward the cell center.
Determination of contact guidance and cell speed
EC alignment was derived from the 5-min interval time-lapse images. Cells were manually outlined in enhanced images. Elongation was calculated as the ratio of the distance between the foci and the major axis length of an ellipse with the same second-moments as the outlined cell. Elongation values fall between 0 (circle) and 1 (line). Alignment angle was then defined as the absolute value of the angle between the ridge orientation and the major axis of that ellipse. Movement angle was calculated as the absolute value of the angle between the ridge orientation and the displacement vector of the center of mass for each cell for the 5-min interval. Cell speed was calculated as the length of that displacement vector. A cell was classified as contact guided if the following criteria were met: cell alignment angles of <10° compared to the ridge direction and, where appropriate, movement angles of <10° for a minimum duration of 15 min.
| RESULTS |
|---|
|
|
|---|
3 µm2 attachment pads at the top of the pillars (Fig. 1). Rows of pillars were located in trenches between solid ridges of 6 or 8 µm in width for EC and fibroblast cell samples, respectively. Correspondingly, the pillar array had pitch patterns of 4 x 12 or 4 x 14 µm2. The spring constant of the pillars was 116 ± 10 nN/µm, as determined from measured AFM force curves (17
|
80° (observed from 2060 min) to
0° (by 90 min), indicating that the cell body was beginning to align with the ridges.
|
We observed dynamic motion of pillars under migrating and stationary EC, and forces up to 57 nN were recorded (Fig. 6). Multiple pillars were positioned and deflected under each cell, and their deflections were dynamic and independent. Many force vectors were detected at the leading and trailing edges of the cells, this being especially the case for the contact-guided cell EC 2. The forces generated by the contact-guided EC 2 showed a lower maximum (31 nN) than EC 1 (57 nN), though the force distributions were otherwise similar (data not shown). Naturally, fewer forces were measured under EC 2 compared with EC 1, due to the fivefold difference in area (mean area 3371 and 674 µm2, respectively). One of the pillars (Fig. 3, third pillar row from left, 13 pillars from patterned edge) shows a spurious deflection due to a mismatch with the ideal lattice, which is altered once the cell is overlying it. (This type of error could be selectively excluded from the analysis if desired.)
|
During the entire 360-min sequence, of which a selected area is depicted in Fig. 4, we observed that the majority of force vectors generated by the cell at the center of the field of view were oriented in the direction of cell spreading (i.e., radially outwards) during the beginning of the sequence, but then switched to be oriented opposite to the direction of spreading (i.e., radially inwards) with time. Pillars with force vectors oriented toward the cell center (i.e., inwards) have been circled in Fig. 4, AE. The number of inwardly directed force vectors under this cell increased with time. Note that pillars located under the portion of the cell, which is in contact with the other cell in the field of view (at upper right), have been neglected from this analysis. A cartoon to describe this observation is shown in Fig. 5 A. It illustrates a cell spreading from both a top and side view. The dotted line represents the cell perimeter at early time points, whereas the solid line represents the cell perimeter at later time points in the spreading process. Dark gray arrows show the forces oriented in the direction of cell spreading. Light gray arrows indicate forces oriented away from the direction of cell spreading.
|
125 min into the sequence.
Fig. 5 D shows a simplified version of the same plot, indicating the forces measured on two sets of six pillars located in adjacent rows. The first set (pillars numbered 114119) represents those pillars positioned under the cell (dark gray), and the second set (pillars numbered 140145) those not under the cell (light gray), at start of the time-lapse acquisition. The solid black line in Fig. 5 D represents a single pillar, number 162, which was never in contact with either of the cells. Forces measured in the first
150 min of the sequence were exclusively those pillars located under the cell, and were oriented with the direction of cell spreading; whereas those forces measured in the latter half of the sequence are for pillars that the cell had spread onto and were oriented against the direction of spreading. We have observed outward spreading forces in other spreading cells in the full FOV, however, the time for transition and clarity of this transition varied between cells.
In addition to the observation of a transition in force orientation, the plot in Fig. 5 D also demonstrates the elasticity of the pillars as many resumed their original resting positions when the spreading cell had advanced beyond them and begun deflecting pillars in the adjacent row. Likewise, the pillars in the adjacent row show no deflections until the cell is in contact with them, though a noise band of the range of ±5 nN is apparent in the absence of any force thresholding.
Summary of traction forces
Fig. 6 contains a histogram of all forces detected in the fibroblast sequence (solid line) and EC sequence (dashed line). Data are obtained from all pillars in the full time-lapse sequences shown in Fig. 2 (EC) and Fig. 4 (fibroblasts). Note that Figs. 4 and 5 contain a selected portion of the full field of view of the fibroblast time-lapse sequence data included in Fig. 6. The data is represented as the fraction of total force measurements falling into each bin, with a bin size of 4 nN. The EC data contains 33,869 force measurements obtained from an FOV containing an average number of five cells on the force sensing area throughout the time course of the sequence. This data were collected over a sequence of 100 frames taken at 5-min intervals. The fibroblast data contains 113,027 force measurements obtained from an FOV containing an average number of 19 cells on the force sensing area throughout the time course of the sequence. This data was collected over a sequence of 121 frames taken at 3-min intervals. A logarithmic scale on the y axis was used to place emphasis on the higher forces measured under the cells with respect to the noise, which would otherwise dominate the plot because there are more pillars around than under cells.
To estimate nonspecific forces due to image noise detected by the analysis routines, a portion of the force sensor chip void of cells was also processed. The distribution of forces detected in a portion of the FOV from the fibroblast sequence containing no cells is shown in Fig. 6 (dotted line). A total of 1818 measurements were included in this data set. The mean force was 3.29 nN (SD 2.23 nN); 95% of the forces were lower than 7.2 nN (99% < 8.7 nN, 99.9% < 11.2 nN). Thus we claim to have a sensor sensitivity of 7.2 nN. Forces of larger magnitude than the measured nonspecific forces were observed for both cell types. The fibroblasts exerted a greater range of forces, up to 138 nN, than did the EC, up to 57 nN.
| DISCUSSION |
|---|
|
|
|---|
For EC, a microgrooved pattern can (24
) and, on our ridged sensor chips, did induce changes in cell shape and motility associated with contact guidance. One cell (EC 1) approached perpendicular to the pattern and remained spread as it moved across the rows of pillars, whereas the second cell (EC 2) approached parallel to the patterned ridges, became highly polarized, and remained aligned to the pattern as it moved along the ridge. The measured alignment and movement angles of 1.7° and 0.4°, respectively, for the latter cell were clearly parallel to the pattern and well within the limits set for contact guidance. The elongation value of 0.98 for this cell indicated how extreme the morphology change was once the cell contacted the pattern. This provided further validation of the alignment angle because a cell that is almost round has no alignment value of interest. Differences in the forces measured for the contact-guided and non-contact-guided EC need to be confirmed by additional measurements to allow proper statistical analysis. It is possible to reason that there should be differences in cell traction forces between contact-guided and non-contact-guided cells. Using the example of EC 1 and EC 2 observed in this study, there are differences in cell-substrate contact area (fivefold) as well as duration of contact. The first, more spread cell (EC 1) was, during the entire period of observation, overlying more pillars than the elongated contact-guided cell (EC 2). There could additionally be a contribution from differing migration speeds, because a slow moving cell has more opportunity to develop mature focal adhesions. As an example, in a comparison of force measurements performed on a slow cell type, like fibroblasts (6
), with those on a fast cell type, keratocytes (25
), a 60-fold increase in mean migratory speed equated to a 10-fold decrease in stress on the surface.
Our observations on the initial stages of fibroblast spreading show a noticeable and unexpected feature. Initially, pillars located under a round cell appeared deflected outward and oriented with the direction of spreading. Later (at
150 min into the image sequence for the cell shown in Figs. 4 and 5 and
180 min after seeding) as the cell became flattened, a transition in force orientation occurred and the pillars were then observed to be deflected inward, against the direction of spreading. The fibroblasts we observed attached directly on the pattern. During the course of the sequence, they spread and contacted other cells, but did not undergo any major morphological changes or migration. Fibroblasts have previously been shown to exhibit contact guidance when grown on ridged substrates (26
,27
). For those fibroblasts imaged on our sensor that were entirely within the FOV, mean alignment angle was 39° (SD 29°) and mean elongation was 0.66 (SD 0.17), which does not indicate any alignment trend. Analysis on the distribution of force vector angles was performed (data not shown) and it was concluded that the direction of cell traction forces associated with spreading did not appear to be influenced by ridge orientation. At the cell seeding density used in this study, influence from cell-cell contacts cannot be excluded from the interpretation of the results and can play a role in the behavior of the fibroblast cell analyzed in Figs. 4 and 5. However, all fibroblasts found to have a rounded morphology similar to the cell shown in Figs. 4 and 5 exhibited initially outward spreading forces. The time for a force direction transition and clarity of this transition varied between these cells, most likely due to the cells being in different stages of the attachment/spreading process. Again, the influence of cell-cell contacts cannot be excluded. Our current experimental setup does not allow immediate imaging of cells directly at time of seeding. However, an improvement to this setup would include the use of microinjection pipette to place an individual cell under the field-of-view after the start of the time-lapse sequence. In this way we would both avoid cell-cell contacts and observe the entire attachment and spreading events. Similarly, the use of an isotropic cell force sensor design, such as a chip without a ridge pattern or overriding orientation, would be more suitable for future studies of changes in force direction for attaching and spreading cells.
In previous studies of initial EC attachment and spreading, traction forces were evident already in early spreading and typically pointed inward throughout the entire process of spreading (28
). Cells were capable of exerting significant forces before either notable focal adhesion or stress fiber formation (28
) and these early forces were attributed to other mechanisms of force transduction. Maximum stress fiber content in EC was observed to peak at
150 min after cell seeding. The occurrence of new binding events of ligand-receptor interactions was assumed to be completed by that time. Our observations with early spreading fibroblasts show a change in force direction, and which might be attributed to a similar maturation in the focal contact-cytoskeletal machinery allowing contractile forces to be generated. The use of live fluorescent labels of the cytoskeletal and focal contact machinery would allow us to correlate these events with observed forces. We did not observe any transition in force orientation for EC because the cells were already spread at start of image acquisition. This discrepancy between directions of forces for attaching and spreading EC reported in the study by Reinhart-King et al. (28
) and our early spreading fibroblasts could be attributed to the different cell types having different behavior; however, the differences in force-sensing substrates cannot be excluded. The magnitudes of cell traction forces measured (28
,29
) were somewhat lower than those we report, however, it is not straightforward to make a comparison between discrete forces, as we measure, and a force average over an entire cell.
Passaged human EC and fibroblast cells attached to and spread on the silicon dioxide-covered silicon wafer substrates and deflected the microfabricated pillars. ECs were observed to align with the ridge and groove pattern as well as exhibit contact guidance in migration. However, fibroblast cells were not observed to elongate, align, or exhibit contact guidance under the conditions tested. The different cell types thus showed markedly different morphology and behavior on the ridged force sensing substrate. Although the ridge width used in the culture of EC was 2 µm narrower than used for testing fibroblasts, the pillar dimensions were the same. It is possible that the larger ridge width prevented fibroblasts from aligning and experiencing contact guidance, as fibroblasts have previously been reported to be sensitive to ridge width, height, and pitch on microgrooved substrates (18
,30
,31
) The five EC and 19 fibroblasts included in the presented data produced an almost continuous range of forces up to 57 and 138 nN, respectively. The magnitude of force observed for fibroblasts was higher than in several previously reported results (6
,12
,13
) but, similarly to the EC result, the difference in methods makes the comparison unreliable. Given that the force data are a range and do not have a normal distribution, a mean value and standard deviation is not as representative as showing the distribution in its entirety, and rather, may be deceptive in the information it appears to give. Therefore we do not include this type of analysis, opting instead to show the full force distributions.
The dimensions and sensitivity of the patterned, pillared, force sensing substrates were such that multiple pillars could be deflected independently under each cell. Tracking a group of pillars over time showed that pillar deflections were elastic, returning to their resting positions when not actively deflected by the cell. In this study, forces were measured in a range up to 138 nN. Nonspecific force vectors, i.e., vectors measured in the absence of cells, were presented in Fig. 6. The sensitivity of 7.2 nN obtained from this data set, combined with the spring constant of the pillars, and the scaling factor of the image results in the spurious noise being <0.25 pixels. We believe this noise level is quite acceptable and is indicative of the quality of our images and analysis routines. Therefore, to increase the sensitivity of the sensor, pillars with lower spring constants could be used. This would be theoretically possible because the highest measured force of 138 nN is clearly lower than the total force needed for a collision of two neighboring pillars (232 nN). The individual force measurements are affected by a nonspecific error of mean value 3.29 nN but, because the errors are evenly distributed in all directions, the effective measurement error would be lower. Factors contributing to the measurement error could include imperfect correction of shift, debris, image distortion, and artificial vectors on pillars imperfectly aligned with the ideal lattice. Forces exerted on standing silicone (PDMS) pillars have previously been reported to be up to 60 nN, with individual pillars reaching even 90 nN for smooth muscle cells, and this substrate was shown to have a standard deviation of spurious forces of 12 nN (14
). The authors therefore claim to resolve forces >12 nN. Another group using pillared silicone substrates reported a range of forces from 10 to 30 nN for human fibroblast cells, claiming a resolution of 4 nN with an error of at least 20% (12
). Thus our rigid silicon pillar sensor performs as well as, if not better than, other pillared cell traction force sensors described in the literature.
The advantages of using patterned silicon substrates as cell force sensors include excellent optical contrast, a large attachment area at the top of each pillar, batch processing, and robust structure. The pillars are mechanically isolated so forces exerted on one pillar will not influence the deflection of others unless they touch due to extremely large deflections. Such large deflections have not been observed and, due to the stiffness of the pillars, are not expected. The disadvantages of the patterned silicon substrates are that calibration is labor and cost intensive and that their fabrication requires both clean-room facilities and specialized equipment. In addition to the choice of material, the use of DIC microscopy enabled acquisition of high-contrast images of cell-substrate interactions over time on the nontransparent sensors. The images had a very low noise level, which is essential for both cell force measurement and image enhancement. By using time-lapse microscopy, the sampling rate can be tuned to capture the process(es) of interest and there is no artifactual force detection, such as shrinkage from cell fixation procedures. The image enhancement routine developed in this work, based on intensity shifts in pixels over time, allows for better visualization of cell perimeters and detection of boundaries between neighboring cells compared with standard brightness and contrast enhancements. This is due to the fact that the most rapid intensity shifts in the DIC image are in the ruffling membrane, where the cytoskeleton is actively remodeled. Standard brightness and contrast enhancements have limited effect in cells where the membrane is spread thin. However, with contrast being generated by differences in refractive index, it could still be advantageous to image cell and substrate in separate optical channels. In the absence of postprocessing, flattened, spreading cells may be difficult to distinguish from the background, whereas imaging thicker cells would move the focus away from the surface. The use of a fluorescence-labeled cell line, focal adhesion protein, or labeled pillar tops would allow imaging of cells and pillars separately, and would therefore provide a more detailed perspective of the relationship between cell (shape) and traction forces. Furthermore, this would also elucidate whether the cell is in fact attached to the tops, or penetrating the etched trenches to partially engulf the pillarsa factor of particular interest for pillared sensors with lower aspect ratios, such as those in PDMS, where cells might have access to the basal surface.
In summary, the technique of using micron-sized pillars as force sensors in cell traction studies and the use of micropatterned ridges to influence cell elongation, alignment, migration, and contact guidance have both been shown previously. In this pilot study we present the possibility of combining these two established techniques in a rigid material and, as an example, measuring cell forces in two physiologically relevant cell models. The observations presented here indicate proof-of-function of the ridged cell force sensor to induce contact guidance, and that the pillared cell force sensor constructed in rigid silicon has the necessary sensitivity to detect differences in traction force vectors between different cell phenotypes and morphologies. With the possibility to incorporate force sensing with topographical and chemical cues, we predict that these substrates may be designed to influence movement and mechanical substrate interaction for various adherent cell types.
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
Financial support was received from the "Biocompatible Materials" program of the Swedish Foundation for Strategic Research (SSF), the Swedish Council for Research (VR), and the Chalmers Bioscience Program (CBP).
| FOOTNOTES |
|---|
Submitted on August 1, 2006; accepted for publication February 27, 2007.
| REFERENCES |
|---|
|
|
|---|
2. Wong, J. Y., J. B. Leach, and X. Q. Brown. 2004. Balance of chemistry, topography, and mechanics at the cell-biomaterial interface: issues and challenges for assessing the role of substrate mechanics on cell response. Surf. Sci. 570:119133.[CrossRef]
3. Harris, A. K., P. Wild, and D. Stopak. 1980. Silicone rubber substrata: a new wrinkle in the study of cell locomotion. Science. 208:177179.
4. Lee, J., M. Leonard, T. Oliver, A. Ishihara, and K. Jacobson. 1994. Traction forces generated by locomoting keratocytes. J. Cell Biol. 127:19571964.
5. Pelham, R. J. Jr., and Y. Wang. 1997. Cell locomotion and focal adhesions are regulated by substrate flexibility. Proc. Natl. Acad. Sci. USA. 94:1366113665.
6. Dembo, M., and Y. L. Wang. 1999. Stresses at the cell-to-substrate interface during locomotion of fibroblasts. Biophys. J. 76:23072316.
7. Pelham, R. J., Jr., and Y. Wang. 1999. High resolution detection of mechanical forces exerted by locomoting fibroblasts on the substrate. Mol. Biol. Cell. 10:935945.
8. Balgude, A. P., X. Yu, A. Szymanski, and R. V. Bellamkonda. 2001. Agarose gel stiffness determines rate of DRG neurite extension in 3D cultures. Biomaterials. 22:10771084.[CrossRef][Medline]
9. Dubey, N., P. C. Letourneau, and R. T. Tranquillo. 2001. Neuronal contact guidance in magnetically aligned fibrin gels: effect of variation in gel mechano-structural properties. Biomaterials. 22:10651075.[CrossRef][Medline]
10. Freyman, T. M., I. V. Yannas, R. Yokoo, and L. J. Gibson. 2002. Fibroblast contractile force is independent of the stiffness which resists the contraction. Exp. Cell Res. 272:153162.[CrossRef][Medline]
11. Chicurel, M. E., C. S. Chen, and D. E. Ingber. 1998. Cellular control lies in the balance of forces. Curr. Opin. Cell Biol. 10:232239.[CrossRef][Medline]
12. Schwarz, U. S., N. Q. Balaban, D. Riveline, L. Addadi, A. Bershadsky, S. A. Safran, and B. Geiger. 2003. Measurement of cellular forces at focal adhesions using elastic micro-patterned substrates. Mater. Sci. Eng. C-Biomimetic Supramol. Syst. 23:387394.
13. Galbraith, C. G., and M. P. Sheetz. 1997. A micromachined device provides a new bend on fibroblast traction forces. Proc. Natl. Acad. Sci. USA. 94:91149118.
14. Tan, J. L., J. Tien, D. M. Pirone, D. S. Gray, K. Bhadriraju, and C. S. Chen. 2003. Cells lying on a bed of microneedles: an approach to isolate mechanical force. Proc. Natl. Acad. Sci. USA. 100:14841489.
15. Tanaka, Y., K. Morishima, T. Shimizu, A. Kikuchi, M. Yamato, T. Okano, and T. Kitamori. 2006. Demonstration of a PDMS-based bio-microactuator using cultured cardiomyocytes to drive polymer micropillars. Lab Chip. 6:230235.[CrossRef][Medline]
16. Petronis, S., C. Gretzer, B. Kasemo, and J. Gold. 2003. Model porous surfaces for systematic studies of material-cell interactions. J. Biomed. Mater. Res. A. 66A:707721.[Medline]
17. Petronis, S., J. Gold, and B. Kasemo. 2003. Microfabricated force-sensitive elastic substrates for investigation of mechanical cell-substrate interactions. J Micromech Microeng. 13:900913.[CrossRef]
18. Dalby, M. J., M. O. Riehle, S. J. Yarwood, C. D. Wilkinson, and A. S. Curtis. 2003. Nucleus alignment and cell signaling in fibroblasts: response to a micro-grooved topography. Exp. Cell Res. 284:274282.[Medline]
19. Weiss, P., and B. Garber. 1952. Shape and movement of mesenchyme cells as functions of the physical structure of the medium: contributions to a quantitative morphology. Proc. Natl. Acad. Sci. USA. 38:264280.
20. Clark, P., P. Connolly, A. S. Curtis, J. A. Dow, and C. D. Wilkinson. 1990. Topographical control of cell behaviour. II. Multiple grooved substrata. Development. 108:635644.
21. Charest, J. L., M. T. Eliason, A. J. Garcia, and W. P. King. 2006. Combined microscale mechanical topography and chemical patterns on polymer cell culture substrates. Biomaterials. 27:24872494.[CrossRef][Medline]
22. Britland, S., H. Morgan, B. Wojiak-Stodart, M. Riehle, A. Curtis, and C. Wilkinson. 1996. Synergistic and hierarchical adhesive and topographic guidance of BHK cells. Exp. Cell Res. 228:313325.[CrossRef][Medline]
23. Tortonese, M., and M. Kirk. 1997. Characterisation of application specific probes for SPMs. SPIE. 3009:5360.[CrossRef]
24. Dike, L. E., C. S. Chen, M. Mrksich, J. Tien, G. M. Whitesides, and D. E. Ingber. 1999. Geometric control of switching between growth, apoptosis, and differentiation during angiogenesis using micropatterned substrates. In Vitro Cell. Dev. Biol. Anim. 35:441448.[Medline]
25. Doyle, A., W. Marganski, and J. Lee. 2004. Calcium transients induce spatially coordinated increases in traction force during the movement of fish keratocytes. J. Cell Sci. 117:22032214.
26. Brunette, D. M. 1986. Fibroblasts on micromachined substrata orient hierarchically to grooves of different dimensions. Exp. Cell Res. 164:1126.[CrossRef][Medline]
27. Walboomers, X. F., H. J. Croes, L. A. Ginsel, and J. A. Jansen. 1999. Contact guidance of rat fibroblasts on various implant materials. J. Biomed. Mater. Res. 47:204212.[CrossRef][Medline]
28. Reinhart-King, C. A., M. Dembo, and D. A. Hammer. 2005. The dynamics and mechanics of endothelial cell spreading. Biophys. J. 89:676689.
29. Reinhart-King, C. A., M. Dembo, and D. A. Hammer. 2003. Endothelial cell traction forces on RGD-derivatized polyacrylamide substrata. Langmuir. 19:15731579.[CrossRef]
30. Walboomers, X. F., W. Monaghan, A. S. Curtis, and J. A. Jansen. 1999. Attachment of fibroblasts on smooth and microgrooved polystyrene. J. Biomed. Mater. Res. 46:212220.[CrossRef][Medline]
31. Loesberg, W. A., X. F. Walboomers, J. J. van Loon, and J. A. Jansen. 2006. The effect of combined hypergravity and micro-grooved surface topography on the behaviour of fibroblasts. Cell Motil. Cytoskeleton. 63:384394.[CrossRef][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||